1 Bioengineering Institute, University of Auckland, Auckland, New Zealand
2 Electrical and Computer Engineering, University of Auckland, Auckland, New Zealand
3 Physiology, University of Auckland, Auckland, New Zealand
4 Physiology, University of Auckland, New Zealand; Bioengineering Institute, University of Auckland, New Zealand
* To whom correspondence should be addressed. E-mail: s.malpas{at}auckland.ac.nz.
We report the development of a novel technology that enables the wireless transmission of power to implantable physiological devices. The system involves a primary unit generating the magnetic field and a secondary pickup unit deriving power from the magnetic field and a power conditioner. The inductively coupled system was able to supply a minimum of 20mW at all locations and pickup orientations across a rat cage, although much higher power of up to 10 W could be achieved. We hypothesized that it would be possible to use this technology to record a high fidelity ECG signal in a conscious rat. A device was constructed in which power was utilized to recharge a battery contained within a telemetry device recording ECG signal sampled at 2000 Hz in conscious rats (200-350 gm) living in their home cage. Attributes of the ECG signal (QT, QRS, and PR interval) could be obtained with a high degree of accuracy (<1 ms). ECG and heart rate changes in response to treatment with the beta blocker propranolol and the proarrhythmic alkaloid aconitine were measured. Transmitters were implanted for up to 4 months. Such technology allows potentially lifetime monitoring without the need for implant refurbishment. The ability to provide suitable power levels to implanted devices without concern to the orientation of the device and without causing heating provides the basis for the development of new devices to record or influence physiological signals in animals or humans over significantly longer time periods than can currently be accommodated.
This article has been cited by other articles:
![]() |
S.-J. Guild, C. J. Barrett, F. D. McBryde, B. N. Van Vliet, and S. C. Malpas Sampling of cardiovascular data; how often and how much? Am J Physiol Regulatory Integrative Comp Physiol, August 1, 2008; 295(2): R510 - R515. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| Visit Other APS Journals Online |