|
|
||||||||
1INSERM/ERIT-M 0207 Motricite-Plasticite Laboratory, Faculty of Sports Sciences-University of Bourgogne, BP 27877, 21078 Dijon; and 2PPEH Research Unit, Jean Monnet University, 42000 Saint Etienne, France
Submitted 7 April 2004 ; accepted in final form 11 July 2004
| ABSTRACT |
|---|
|
|
|---|
low- and high-frequency electrical stimulation; central activation; gender; body fat percentage
Tetanic stimulation of the motor nerve is effective in recruiting almost maximally the knee extensor muscle group (KE) (7, 22) but induces discomfort so that one cannot use this method with patients or aged subjects. However, neurostimulation (NES) can be considered as the best currently available method to study LFF. Rutherford et al. (25) previously addressed this issue of transcutaneous vs. motor nerve stimulation of the quadriceps for the interpolated twitch technique and did not find any difference between the two methods. As far as we know, this comparison has never been made for the evaluation of LFF. Therefore, the purpose of this experiment was to compare the results provided by these two methods of electrical stimulation, i.e., ES and NES, after eccentric-type exercise to test the validity of submaximal elicited torques in the evaluation of LFF. A secondary objective was to examine any gender differences in the responses to ES and NES, because the higher percentage body fat observed in female subjects may alter the response to ES.
| METHODS |
|---|
|
|
|---|
|
MVCs.
MVC testing involved three conditions. Two trials per condition were required with a 1-min rest between each trial. For each condition, the best result was used for further analysis. In the first condition, the contractions were performed without electrical stimulation. These two MVC attempts were only performed before the fatiguing exercise because the best value was used to determine ES stimulation intensity (see below). The second and third conditions involved MVC with a double pulse (time interpulse: 10 ms) superimposed to the isometric plateau, by either ES or NES, in a randomized order. The stimulation intensity was the same as that set for the electrically evoked torque measurements (see below). The double pulse superimposition technique, based on the interpolated-twitch method (20), enabled us to estimate the KE maximal voluntary activation level (%VA). The ratio of the amplitude of the superimposed double pulse over the size of a double pulse in the relaxed muscle (control doublet) was then calculated to obtain %VA as follows:
![]() |
Electrically evoked torque measurements. After MVC measurements, electrical stimulation of resting KE was performed by using either ES or NES. Square-wave pulses with a width of 0.5 ms at a maximal voltage of 400 V were produced by two stimulators (both were Digitimer DS7, Welwyn Garden City, United Kingdom).
ES was delivered percutaneously via two 5.1 cm x 10.2 cm gel pad electrodes (Compex SA, Ecublens, Switzerland) placed proximally and distally on the anterolateral thigh. The stimulation intensity, ranging from 36 to 112 mA, was set by progressively increasing the stimulus intensity of a 500-ms tetanus at a stimulation frequency of 80 Hz until the stimulation train was sufficient to evoke 50% of the subject's MVC on the day of testing. This 500-ms tetanus duration is sufficient to reach a plateau during the tetanus (Fig. 1).
|
The electrically evoked torque measurements comprised three single twitches and two 0.5-s tetani at a frequency of 80 and 20 Hz for each stimulation method. The stimulations were applied in the order presented in Fig. 2. A P20-to-P80 ratio (P20/P80) was also calculated; any decrease in this ratio is commonly interpreted as an index of low-frequency fatigue. The three twitches were averaged and the resulting mean responses were considered for the comparison between rest, fatigue (Post), and recovery (Post30) conditions. The following parameters were obtained from the mean twitch response: peak torque, twitch contraction time, half-relaxation time, maximal rate of torque development, i.e., maximal value of the first derivative of the mechanical signal divided by peak torque, maximal rate of torque relaxation, i.e., maximal value of the first derivative of the mechanical signal divided by peak torque. These twitch characteristics were measured to assess any difference in motor unit recruitment between genders and/or stimulation method.
|
![]() |
![]() |
Electromyogram recording.
The EMG signal of the right vastus lateralis was recorded by using bipolar silver chloride surface electrodes of 10-mm diameter (type 0601000402, Contrôle Graphique Medical, Brie-Comte-Robert, France) during MVC and electrical stimulation. The recording electrodes were fixed lengthwise over the muscle belly with an interelectrode distance of 25 mm. The position of the electrodes was marked on the skin to fix them at the same place postexercise. The reference electrode was attached to the opposite patella. Low impedance (Z) at the skin-electrode surface was obtained (Z < 5 k
) by abrading the skin with emery paper and cleaning with alcohol. Myoelectrical signals were amplified (custom-made amplifier: common mode rejection ratio = 90 dB, Z input = 1,000 M
, gain = 1,000) with a bandwidth frequency ranging from 1.5 Hz to 2 kHz and simultaneously digitized online (Tida ITC 16, Heka Elektronik, Lambrecht/Pfalz, Germany; sampling frequency 2,000 Hz). The root mean square (RMS) value of the M-wave (RMSM) obtained with NES was determined during the maximal twitches. The window size was adjusted to the size of the M-wave, to take into account the distinct phases of the M-wave, i.e., all the modifications of the resting potential. In addition, an RMS value was calculated during the MVC trials (RMSMVC) over a 0.5-s period after the torque had reached a plateau and before the surperimposed stimulation was evoked. All mechanical and EMG data were stored with commercially available software (Tida 5.0, Heka Elektronik).
Stimulation-induced pain. Pain associated with 20- and 80-Hz stimulation using either ES or NES was evaluated by use of visual analog scales consisting of a 100-mm horizontal line with an item at each extremity: from "no pain" to "very very sore." Subjects were asked to put a vertical mark on the horizontal line to describe the pain associated with each frequency and each type of stimulation. The distance between the beginning of the line and the vertical mark was used as the pain score.
Body fat percentage. Body fat percentage, based on skinfold thickness measurements, was assessed by using a Harpenden skinfold caliper (Harpenden skinfold caliper, British Indicators, Burgess Hill, West Sussex, Great Britain). Skinfolds measured at bicipital, tricipital, subscapular, and suprailiac sites (5) were used to determine body density from which body fat percentage was calculated (27). All skinfolds were the average of three measurements assessed on the right side of the body, always by the same observer.
Statistics. All descriptive statistics presented are mean values ± SD. Normal distribution was checked by using a Shapiro-Wilk test of normality. Each study variable was then compared between rest, Post, and Post30 conditions by a three-way (gender x stimulation type x time) ANOVA with repeated measures. Newman-Keuls post hoc tests were applied to determine between-means differences if the ANOVA revealed a significant main effect for time or interaction of stimulation type x time or gender x stimulation type x time. Morphological variables were compared with a single-tailed Student's t-test. For all statistical analyses, a P value of 0.05 was accepted as the level of significance.
| RESULTS |
|---|
|
|
|---|
For the 80-Hz-stimulation at rest, stimulation efficiency displayed significant gender x stimulation efficiency interaction (P < 0.05). Whereas no gender difference was observed for ES (male: 1.11 ± 0.25 vs. female: 1.00 ± 0.41 N·m/mA), male subjects displayed greater stimulation efficiency for NES compared with female subjects (2.26 ± 0.51 vs. 1.65 ± 0.63 N·m/mA; P < 0.05). The same tendency was observed for the 20-Hz stimulation, but the interaction failed to reach the significance level (P = 0.09). These tendencies disappeared when stimulation efficiency was normalized to lean body mass. Normalized stimulation efficiency only displayed a stimulation method effect: NES proved to be more efficient in evoking torque than ES, both for the 80-Hz (0.037 ± 0.011 vs. 0.020 ± 0.008 N·m·kg lean body mass1·mA1; P < 0.001) and the 20-Hz stimulation (0.028 ± 0.01 vs. 0.014 ± 0.006 N·m·kg lean body mass1·mA1; P < 0.001).
Voluntary contractions. MVC data displayed a significant time effect (P < 0.001). Torque decreased at all points in time postexercise with no significant recovery between Post and Post30 (Table 2). This significant impairment occurred concurrently with a significant reduction of %VA and RMSMVC over time (P < 0.001 and P < 0.01, respectively; see Table 2). At Post30, RMSMVC decline just failed to reach the significance level (P = 0.06). None of these variables showed any gender difference when the relative decrease was considered. In addition, estimation of %VA did not differ according to the method of stimulation.
|
|
|
|
Stimulation discomfort. As shown in Fig. 5, the stimulation-induced perceived discomfort demonstrated significant stimulation method effect (P < 0.001) and gender x frequency interaction (P < 0.01). ES was 1.5 to 2 times more comfortable than NES. For men, low-frequency stimulation was less painful than high frequency. This difference was not apparent in female subjects.
|
| DISCUSSION |
|---|
|
|
|---|
Stimulation-induced discomfort. ES appeared less painful than NES. On the one hand, this cannot be the result of stimulation intensity, because ES and NES did not differ in the intensity level set to evoke muscle contractions. On the other hand, the current density is more diffuse with larger electrodes. Thus the larger size of ES stimulation electrodes, compared with NES, could have contributed to the minimization of electrically induced pain (1).
Exercise-induced fatigue: comparison with the literature. The 30-min downhill run induced a 15% MVC decrease together with a slight decline of %VA and RMSMVC. The fatiguing exercise also induced a significant reduction in the P20/P80 over time, together with a general impairment of electrically evoked torques. M-wave characteristics were also altered after the downhill run. These results are in line with previous work on neuromuscular fatigue after eccentric-type exercise involving KE (19). In this study, we demonstrated that both central, i.e., decline of voluntary activation level, and peripheral, i.e., low-frequency fatigue, components contributed to the 19.6% KE MVC decline observed after an intermittent one-legged downhill run. If low-frequency fatigue is commonly observed after eccentric exercise (8, 23), the experimental evidence for the presence of central fatigue is rather scarce (2, 18, 19, 21). However, it is generally believed that the contribution of this central component to MVC decline may be weak. Peripheral fatigue, i.e., low-frequency fatigue and decreased sarcolemmal excitability, explains most of the MVC decrement in the present study. It is also worth noting that the estimation of %VA was independent of gender and stimulation method. This result is consistent with previously published literature on twitch-interpolation methodology (25).
Electrically evoked contractions. To our knowledge, the present report is the first to compare ES and NES in the evaluation of low-frequency fatigue. Previous attempts were aimed at comparing torques evoked at different voltages using ES (4, 6, 9, 12). However, all these studies have only used submaximal torques (<70%MVC). As stated in the introduction, this choice may involve a number of limitations.
In the present study, the comparison of transcutaneous vs. motor nerve stimulation of the quadriceps revealed that the P20/P80 behavior under fatigued and recovery conditions was influenced neither by the method employed to elicit muscle contraction nor by gender. As a result, low-frequency fatigue appeared to be accurately assessable using ES. This result can be linked to the report by Edwards et al. (6) on sternomastoid muscle in which the P20/P50 ratio was demonstrated to be stable across a range of submaximal voltages under fatigued conditions. Hanchard et al. (12) drew the same conclusions for the tibialis anterior muscle using torque levels comprising between 10 and 33% MVC. However, conflicting results were obtained by Davies and White (4) on the plantarflexor muscles. The reasons for such a discrepancy are unclear, but it is possible that voltage (or intensity) dependency may differ between muscle groups (12).
If both methods similarly detected the presence of low-frequency fatigue, they also revealed a general alteration of electrically evoked torques. Qualitatively, both stimulation methods detected an impairment in P20 and P80 but were not in agreement on the magnitude of these decreases. Overall, ES tended to detect a greater impairment at Post in male and a lesser one in female subjects at both Post and Post30. The reasons for these differences are unclear. A possible explanation may be that different motor unit pools are recruited when male and female subjects are stimulated by ES. However, the analysis of the twitch characteristics does not support this hypothesis. Another possibility is that the greater subcutaneous fat content in women may have influenced their response to ES stimulation. It is worth noting that stimulation intensities did not differ between genders. One could suppose that the higher subcutaneous fat content in female subjects may have altered stimulation efficiency in this group of subjects. However, it must be remembered that men have a greater muscle mass. Because cross-sectional area was not measured in the present study, we normalized evoked torque to lean body mass. When evoked torque was expressed in newton-meters per kilogram of lean body mass, there was no longer any difference between genders for a given stimulation intensity, suggesting that the stimulation efficiency did not differ between men and women, whatever the stimulation method. According to Lieber and Kelly (17), the ability to evoke muscle torque via transcutaneous stimulation is not simply a matter of applying high currents to small and lean individuals. Geometric factors related to the location of the recruited motor units with respect to the stimulating electrodes as well as superficial patterns of motor nerve branching could also influence muscle response to ES (3, 14). In the present study, differences in muscle architecture could have contributed to gender and stimulation method differences in the evaluation of P20 and P80 decreases after the fatiguing exercise, although this is not supported by any scientific evidence. In any case, it is important to note that these differences did not affect the P20/P80, which did not differ, either between ES and NES or between genders.
In conclusion, the results presented here suggest that, after a 30-min downhill run, relative decrements in low-to-high frequency ratios evoked via transcutaneous and motor nerve stimulation were not different. Thus submaximal torques evoked via ES proved to be valid in evaluating low-frequency fatigue. In estimating evoked torque decrements, both ES and NES were able to detect fatigue effects, although results from the two methods were not in agreement on the magnitude of these reductions. The reasons for this discrepancy remain unclear, but differences in muscle architecture might be involved. Because ES proved to be less painful that NES, these results may have important practical implications for the evaluation of low-frequency fatigue in sensitive populations such as patients or aged subjects.
| GRANTS |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Verges, N. A. Maffiuletti, H. Kerherve, N. Decorte, B. Wuyam, and G. Y. Millet Comparison of electrical and magnetic stimulations to assess quadriceps muscle function J Appl Physiol, February 1, 2009; 106(2): 701 - 710. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Laghi Advancing femoral nerve stimulation into the stage of science J Appl Physiol, February 1, 2009; 106(2): 356 - 357. [Full Text] [PDF] |
||||
![]() |
L. A. Burnes, S. J. Kolker, J. F. Danielson, R. Y. Walder, and K. A. Sluka Enhanced muscle fatigue occurs in male but not female ASIC3-/- mice Am J Physiol Regulatory Integrative Comp Physiol, April 1, 2008; 294(4): R1347 - R1355. [Abstract] [Full Text] [PDF] |
||||
![]() |
R B. Keeton and S. A Binder-Macleod Low-Frequency Fatigue Physical Therapy, August 1, 2006; 86(8): 1146 - 1150. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |