INNOVATIVE METHODOLOGY
1Department of Medical Informatics and Engineering, University of Szeged, H-6720 Szeged, Hungary; and 2Telethon Institute for Child Health Research and Centre for Child Health Research, University of Western Australia, Perth 6008, Australia
Submitted 1 February 2003 ; accepted in final form 27 May 2003
| ABSTRACT |
|---|
|
|
|---|
20 s. Zrs was averaged for 0.125 s and fitted by a model featuring airway resistance (Raw) and inertance, and tissue damping and elastance (H). Lower airway conductance (Glaw) was linearly related to volume above functional residual capacity (V) between 0 and 75-95% maximum V, with a mean slope of dGlaw/dV = 13.6 ± 4.6 cmH2O-1 · s-1. The interdependence of Raw and H was characterized by two distinct and closely linear relationships for the low- and high-volume regions, separated at
40% maximum V. Comparison of Raw with the highest-frequency resistance of the total respiratory system revealed a marked volume-dependent contribution of tissue resistance to total respiratory system resistance, resulting in the overestimation of Raw by 19 ± 8 and 163 ± 40% at functional residual capacity and total lung capacity, respectively, whereas the lowest frequency reactance was proportional to H; these findings indicate that single-frequency resistance values may become inappropriate as surrogates of Raw when tissue impedance is changing. airway resistance; tissue elastance; forced oscillations; volume dependence; total lung capacity
Continuous estimation of impedance has been applied in the characterization of transient changes in lung mechanics during a constrictor challenge (2-4); slow partial volume inflations in anesthetized, mechanically ventilated subjects and conscious volunteers (19); and deep inhalation in healthy and asthmatic subjects (5, 18). In these studies, either forced oscillations of a single frequency of 3 Hz (19), 6 Hz (3), or 8 Hz (5, 18) were used to estimate the resistance of the total respiratory system (Rrs) or the lungs (RL) on a cycle-by-cycle basis, or resistance and elastance values were tracked at two (2) or three frequencies (4) applied simultaneously. The ease of collection of impedance data at a high temporal resolution is nevertheless associated with problems in the interpretation of the mechanical properties manifested at a single frequency, the most important question being to what extent and under what conditions a single-frequency value of Rrs or RL can be regarded as the resistance of the airways (Raw) (2-5, 18, 19).
The separate assessment of airway and tissue mechanical parameters from low-frequency input impedance data requires a number of simultaneously applied frequency components in an appropriate range that permits the manifestation of the mechanical properties of both the airways and tissues in the particular species, and the use of a validated model in the evaluation of the impedance data (14, 16, 17, 21, 22, 24, 26). In the present report, we employ pseudorandom forced oscillations in anesthetized mice in the frequency range of 4-48 Hz and demonstrate for the first time that the airway and tissue parameters can be obtained reliably by fitting the instantaneous spectra of total respiratory input impedance (Zrs) collected during a slow continuous inflation-deflation maneuver. The multitude of Zrs parameter values obtainable from a single maneuver enables us 1) to show the characteristic patterns of pressure and volume dependences derived for the airways and tissues, 2) to explore the relationship between the Newtonian (airway) resistance and tissue elastance, which breaks down into two distinctly different linear segments in the low- and high-volume ranges, and 3) to demonstrate that single-frequency resistance values can seriously overestimate Raw because of the involvement of tissue resistance, which varies systematically with lung volume.
| MATERIALS AND METHODS |
|---|
|
|
|---|
40-60-min intervals, as required. After the surgical anesthesia, a tracheostomy was performed and a polyethylene cannula (10 mm long, ID = 0.81 mm) was inserted. Mechanical ventilation (flexiVent, Scireq, Montreal, Canada) was maintained with a tidal volume of 8 ml/kg, a rate of 450 min-1, and a positive endexpiratory pressure of 2 cmH2O. The animal handling and study protocol conformed to the guidelines of the Australian National Health and Medical Research Council and were approved by the Animal Ethics Committee of the Institute for Child Health Research.
Zrs was determined with pseudorandom forced oscillations in the frequency range of 4-38 Hz, by using the wave-tube technique (12, 27) adapted to small animals (21, 24). The oscillatory signal consisted of nine components, which were relative primes, i.e., noninteger multiples of each other (the 2-Hz fundamental was of zero amplitude), to minimize the effects of nonlinear harmonics production (9, 23). The component amplitudes decreased hyperbolically between 4 and 22 Hz and were then kept constant, in accordance with the expected frequency dependence of the magnitude of Zrs and the signal-to-noise values. The phase angles of the components were optimized to find the minimum peak-to-peak excursion of the composite signal at the specified total energy (11). All these considerations relating to nonlinearity and signal-to-noise ratios were particularly important because the measurements were not made under strictly apneic conditions despite the absence of spontaneous breathing (the mean flow changed according to the volume maneuver), there was interference by the cardiac activity, and moreover Zrs was not in a steady state. Zrs was determined as the load impedance on a wave tube (12, 27)
![]() |
are the characteristic impedance and wave number of the tube, respectively; L is the tube length; and P1 and P2 are the lateral pressures at the outlet and inlet of the tube, respectively. Z0 and
can be expressed as
![]() |
Before the oscillatory measurements, the mouse was hyperinflated by superimposing three tidal volume inspirations and was then ventilated normally for 1 min. The respirator was next stopped, and the tracheal cannula was switched to the wave tube and the loudspeaker box. The measurement started with an
2-s baseline recording, during which the mean values of P1 (P1mean) and P2 (P2mean) were zero, and was followed by a slow (
20 s), nearly linear inflation of the whole system from an O2 source up to 20 cmH2O; the inflation was then stopped and the system was connected to an expiratory line, the resistance of which was chosen to result in quasi-exponential decay in P1mean and P2mean to nearly atmospheric pressure in
20 s. The length of the total recording was 45 s, after which the tracheal tube was switched back to the respirator, and the ventilation was resumed.
The signals were filtered with fifth-order Butterworth anti-aliasing low-pass filters at 50 Hz and sampled at a rate of 256 s-1. Zrs was determined after digital high-pass filtering of P1 and P2 at 3.5 Hz, which eliminated the slow pressure components of the total lung capacity (TLC) maneuver. To further suppress the effects of nonoscillatory pressure components, P2/P1 was computed from the cross spectra between P1 and U, and P2 and U, where U is the noise-free driving signal of the loudspeaker (10). The time window of fast Fourier transformation was 0.5 s, and the spectra estimated by moving sample-by-sample along the recording were averaged for 0.125 s; hence, these Zrs data represented the average impedance for a 0.625-s interval, and the successive overlapping estimates were obtained every 0.125 s. These Zrs spectra were then fitted to a four-parameter model (16) to estimate the Raw and airway inertance (Iaw) and the constant-phase tissue damping (G) and elastance (H)
![]() |
is angular frequency and
= (2/
) · arctan(H/G). Strictly speaking, the parameters Raw and Iaw, respectively, include the Newtonian components of tissue resistance and tissue inertance. However, measurements in intact and open-chest rats (17, 22) demonstrate that the contributions of the tissues to Raw and Iaw can be neglected. The model fitting was accomplished by minimizing the sum of squared differences, normalized by |Zrs|, between the measured and model data. A smoothing procedure (a Gaussian weight function and quadratic fit on 5% blocks of the entire data sequence) was then applied to the time sequences of the parameters, which performed a resampling reduction of each parameter set to 100 data points per maneuver. P1mean and P2mean were obtained by low-pass filtering P1 and P2, respectively, at 3.5 Hz, which removed the oscillatory components from the pressure signals. P2mean was used as the estimate of Prs in the pressure-volume relationship. The volume above functional residual capacity (FRC) (V) was obtained by integrating the nonoscillatory flow computed as the ratio of the pressure difference P1mean - P2mean and the flow resistance of the wave tube. The setup was regularly checked by applying the usual inflation-deflation protocol to measure a mechanical lung model which consisted of a 20-ml syringe attached to the end of the tracheal tube used in the mouse experiments.
| RESULTS |
|---|
|
|
|---|
|
Typical recordings of P2 and V, and the temporal changes in the model parameters during the inflationdeflation maneuver, are depicted in Fig. 2. The oscillatory excursions of P2 ranged from ±2 cmH2O (at mid-inspiration) to ±3.5 cmH2O (at TLC), depending on the actual value of Zrs. The change in V was mirrored by those in Raw and Iaw, although the courses of the latter were noisy and sometimes inconsequential. Both tissue parameters decreased initially, then increased steeply until the maximum in V, and displayed the opposite changes with decreasing V. The volume-dependent changes in H exceeded those in G. These temporal patterns are in accordance with the qualitative changes in Zrs illustrated in Fig. 1. The fitting error was on average
5%, its variations following the rate of change in Zrs: the high values of fitting error (F%) coincided with intervals in which H, the major determinant of Zrs, changed rapidly in either direction (Fig. 2).
|
As expected, the parameters obtained in the mechanical model were fairly time invariant, i.e., independent of the applied pressure (Fig. 2). The "tissue" parameters of the model are related to the gas compressibility in the syringe, which produced a value of H comparable with the average H during the maneuver in a mouse, and a small value of G resulting from the polytropic gas compression process; accordingly, their estimations serve only as the validation and quality check of the measurement and evaluation techniques. However, as the values of Raw estimated in the mechanical model (i.e., the resistance of the tracheal cannula with connections) comprised a constant part (0.275 cmH2O · s · ml-1) of the Raw measured in the animals, it can be subtracted from the latter to obtain the lower airway resistance (Rlaw) of the mice, which varied as a function of P2 or V. Although the same applies for the inertance values, Iaw was a minor parameter in determining Zrs at the relatively low oscillation frequencies (which were well below the resonant frequency) and was easily biased by other parameters in the model fitting; its correction for the tracheal tube component (0.00056 cmH2O · s2 · ml-1) would have resulted in unrealistic values, and hence this was not done.
In one animal, the tracking estimates of model parameters were compared with those obtained at constant levels of Prs. After the recordings of two TLC maneuvers, P2mean was set successively to 0, 2, 5, 10, 15, and 20 cmH2O, four 6-s measurements were collected at each level, and then another two TLC maneuvers were performed. The parameter values from the individual tracking estimations were interpolated to a common sequence of Prs data and then averaged separately for inflation and deflation (Fig. 3). Although slight but systematic loopings were formed by the inspiratory and expiratory limbs in the Raw, G, and H vs. Prs graphs, the agreement between the tracking estimates and the steady-state parameter values was good, and the variabilities in the courses of the parameters were comparable to those in the steady-state recordings at most Prs levels.
|
Examples of the estimates of Raw, G, H, and tissue hysteresivity,
= G/H (13), plotted against Prs and V, are presented in Fig. 4. The dependences of the parameters on Prs and V exhibit similar patterns, although the inspiratory and expiratory limbs of the courses of Raw, H, and
are closer to each other if plotted as functions of V. The values of the parameters at end expiration remained slightly below the corresponding values at the beginning of the maneuver in the case of G (3.9 ± 1.9%, P < 0.001) and H (6.6 ± 4.2%, P < 0.02), whereas this decrease was more marked in Rlaw (18.1 ± 8.1%, P < 0.005). The lower airway conductance (Glaw = 1/Rlaw) is plotted against V for each animal in Fig. 5. The Glaw vs. V relationships are fairly linear and similar for the inspiratory and expiratory limbs up to 75-95% of the maximum V and become much steeper above those levels. From the regression Glaw = a + bV calculated for the visually selected linear ranges, fairly uniform conductance values at FRC (a = 2.53 ± 0.21 ml · cmH2O-1 · s-1) and high correlation coefficients (r2 = 0.93 ± 0.01) and somewhat more variable coefficients of volume dependence (b = dGlaw/dV = 13.6 ± 4.6 cmH2O-1 · s-1) were obtained. When Glaw is plotted against Prs, the relationships become more linear, and the regression Glaw = c + dPrs applied to the whole data range in each mouse results in values with similar variability as in the volume dependence of Glaw (c = 2.32 ± 0.23 ml · cmH2O-1 · s-1, d = 0.41 ± 0.14 ml · cmH2O-2 · s-1, and r2 = 0.91 ± 0.03).
|
|
When Raw is plotted against H during a TLC maneuver, a characteristic pattern of interdependence between the airway and tissue properties is disclosed (Fig. 6). In the early phase of the maneuver, both parameters decrease, but H later begins to rise whereas Raw continues to decrease. During expiration, the reverse changes occur in the Raw vs. H data, which closely follow the same trajectories as in inspiration, although the end-expiratory slope of the relationship is slightly but systematically lower than the early inspiratory slope. The two distinct phases form a low-volume and a high-volume region, separated by the minima of H in the inspiratory and expiratory directions, which take place at a mean V of 0.175 ± 0.038 and 0.174 ± 0.028 ml, respectively. These on average identical values of V correspond to 40% of the maximum V and to values of Rlaw amounting to 44% of those at FRC. The Raw vs. H relationships are characterized by high correlation coefficients of 0.95 ± 0.03 and 0.84 ± 0.06, respectively, for the low-volume and high-volume regions.
|
Model parameters and single-frequency impedance values during the TLC maneuver were compared to establish whether the values of the highest frequency Rrs (Rrs38) and the magnitude of the lowest-frequency Xrs (Xrs4) followed the changes in Raw and H, respectively. As illustrated in Fig. 7, a very high correlation coefficient was obtained for both relationships. Raw and Rrs38 displayed a similar dependence on V, with an upward shift for Rrs38, indicating a tissue resistance component in the latter. However, with decreasing Raw, the change in Rrs38 gradually became smaller as a result of an increased contribution of the tissue resistance to Rrs38 with increasing V (cf. the change in G in Fig. 4). For all animals, Rrs38 overestimated Raw by 19 ± 8 and 163 ± 40% at FRC and TLC, respectively. By contrast, the proportionality between Xrs4 and H was much better; the mild curvilinearity apparent at high values of H can be explained on the basis of the change in
with V (see DISCUSSION).
|
| DISCUSSION |
|---|
|
|
|---|
The reproducibility of the Zrs parameters in subsequent TLC maneuvers was good (see Fig. 3) and was comparable with that obtained from the more time-consuming measurements at a sequence of constant Prs levels in one animal in the present study and in the previous investigation (24). The errors of model fitting to Zrs data ranged from 2 to 10% during the maneuvers and averaged
5%, roughly twice as high as the values of F% obtained invariably at all Prs levels (24). The reasons for this difference in fitting quality are 1) the involvement of cardiac artifacts, which were detectable via the increased scatter in the Zrs data at certain frequencies in the repeated measurements at the same Prs and which were omitted from the fitting, and 2) the increase in F% in phases of the TLC maneuver in which the change in Zrs was relatively fast. In these phases, however, the impairment in fitting quality was caused by increased fluctuations in the Zrs vs. frequency data (this is also apparent in Fig. 1) rather than systematic differences in the fitting.
The impedance parameters obtained from the tracking and constant-Prs estimations were in good agreement and exhibited comparable variabilities in subsequent measurements; the patterns of the dependences on Prs were also similar to those observed previously (24). Raw decreased monotonically with Prs, covering a wide range between 0 and 20 cmH2O, especially if the resistance of the tracheal tube was removed. The changes in G and H were also significant: the initial decreases turned into steep elevations at medium Prs values, which slowed down more in the case of G, resulting in a gradual fall in
. Whereas the data obtained during stepwise increases in Prs did not differ from those collected in the opposite direction (24), within the 45 s of the TLC maneuvers the inspiratory and expiratory limbs were slightly but clearly separated in all parameters except the weakly determined Iaw (Figs. 3 and 4). However, most of the loopings disappeared when the parameters were plotted as functions of V, and the plateaus in Raw and
that were formed at the highest Prs were less apparent when plotted against V (Fig. 4).
Patterns of volume dependence. The dependences of the mechanical parameters on Prs [or the transpulmonary pressure (Ptp)] or V were qualitatively similar to those obtained in previous investigations in rats (17, 22), and in rodents including rats and mice (14), although the ranges of the pressure levels used in those studies were narrower than the span of Prs in the present measurements: 0-9 cmH2O (17), 0-7 cmH2O (14), and 3-11 cmH2O (22). Accordingly, the fall in Raw with increasing Prs, Ptp, or V was generally milder, and the elevations in G and H after the medium-pressure minima of these parameters were less apparent (14, 17, 22, 25) than in the present study; a clear decrease in
was also observed at the highest Ptp levels (17, 22). Among these patterns of pressure dependence, the decreases in Raw with increasing Prs or Ptp obviously reflect the changes in airway diameters, and the increases in G and H at higher lung volume are consistent with the stiffening of the respiratory tissues and particularly that of the lungs (1, 17). It is less straightforward, however, to interpret the increases in G and H toward low values of Prs: whereas the elevations in pulmonary G and H with decreasing Ptp can readily be explained by the progressive closure of lung units (17, 22), perplexingly similar patterns of chest wall G and H have been reported in dogs (1) and rats (17) and have been attributed to the complexity of the chest wall tissues (1). At low lung volumes, artifactual increases in tissue parameters resulting from inhomogeneous constriction of the peripheral airways, such as those observed during constrictor challenge (20, 21), might also have been involved.
Airway resistance vs. lung volume. As concerns the mechanical parameters of the respiratory system, the volume dependence of the airway resistance has been of primary interest. The inverse relationship between Raw and V (7, 8) reflects an interdependence between the tissue and bronchial systems and can be regarded as one of the principal determinants of the airway tone. Jordan et al. (19) applied 3-Hz forced oscillations in mechanically ventilated patients during slow inflations and calculated the specific lower airway conductance (sGlaw = dGlaw/dV) from the Rrs data as a function of volume. After the removal of a resistance component, which was assumed to be independent of volume, by fitting a hyperbolic model between Rrs and V, those authors obtained linear relationships between Glaw and V, which were different in normal subjects and in those with chronic pulmonary disease and which changed after bronchodilator treatment. They argued that sGlaw reflects changes in airway distensibility due to a change in bronchomotor tone, and hence it provides an index of airway tone that is independent of V. In the present study, we also established fairly linear Glaw vs. V relationships for most of the covered volume range; however, unlike the relationships in the above-mentioned work (19), the range did not decrease below FRC but included much larger inflation volumes relative to the end-expiratory volume in this species. As TLC was approached, a sudden rise in Glaw was observed in all mice. Although it may be possible that at extreme lung volumes the Glaw vs. V relationship is governed by mechanisms different from those acting at moderate distensions of the respiratory system, it is also possible that measurement inaccuracies were involved. When very low values of Rlaw are reached, even small errors in the estimation of the frequency-independent resistance, which includes those that do not change with V (e.g., the endotracheal tube) or depend on it in a different manner (such as the Newtonian component of tissue resistance), may distort the linear relationship. However, it should be pointed out that the linear Gaw (1/Raw) vs. V relationship is a simplified concept, and it was never a uniform finding in the earliest work on the volume dependence of plethysmographic Raw (6, 8, 15, 28). In fact, Butler et al. (8) obtained sigmoid-shape Gaw vs. V curves with an upward swing in Gaw at the highest values of V, a pattern similar to that found in the present study (Fig. 5). Those authors also observed that the Gaw vs. Ptp relationships remained fairly linear and, on the basis of measurements obtained during chest strapping, they concluded that a greater pull on the airway wall results in an additional airway enlargement at the same high volume level.
Single- vs. multiple-frequency tracking. The continuous monitoring of respiratory impedance is the most appropriate tool with which to explore the effects of changing lung volume (5, 18, 19) or the temporal dynamics of constrictor responses (2-4). Although such single-frequency measurements are relatively easy to implement, the choice of the oscillatory frequency and the interpretation of the impedance values at that rate are not straightforward. The use of high frequencies offers a good temporal resolution and is usually assumed to provide an estimate of Raw (2, 3, 5, 18), although it is important to assess the contributions of the (negatively) frequency-dependent and Newtonian tissue resistance components to the measured resistance, especially if they are expected to change via volume-dependent stretching or tissue constriction. The values of RL at 8 and 6 Hz, used in previous studies in humans (5, 18) and dogs (2, 3), respectively, were probably acceptable surrogates for Raw, whereas the selection of the 3-Hz values of Rrs in humans was clearly more critical, despite the relatively small inflation volume (<1 liter) applied above FRC (19). Our results show that, at a higher frequency, although in a much smaller species, Rrs includes only an
20% contribution from the tissue resistance at low lung volumes, whereas this component increases progressively with increasing V and results in an overestimation of Raw by 160% at TLC. Because the magnitude of the constant-phase tissue component decreases with increasing frequency (proportionally to
-
), an increase in oscillation frequency by factors of 2 and 10 would result in overestimations of Raw at TLC by 87 and 21%, respectively, on the assumption of
= 0.2 (which corresponds to
= 0.875) for simplicity and provided that Raw remains the same at all frequencies. This simple calculation indicates the need for a much higher tracking frequency for a single-frequency estimation of Raw in the mouse, which has been demonstrated to have a relatively low Raw for its size (14). By contrast, the elastic properties of the respiratory system can be assessed appropriately on the basis of low-frequency Xrs or lung reactance, which are uninfluenced by the inertial forces and severe peripheral inhomogeneities, such as the Xrs4 values in the present study, and the reactance values at 0.4 Hz (1) or 1 Hz (2) in dogs.
Raw vs. H: the elbow plot. The display of the data pairs of two parameters obtained from the TLC maneuver that characterize the overall bronchial geometry and the tissue elasticity, respectively, reveals a biphasic mechanical relationship not explicitly containing Prs or V (Fig. 6). The existence of the elbow in the Raw vs. H plot, i.e., the sudden change in the arrangement of data from one trajectory to another, is striking, because both parameters vary fairly smoothly with Prs and V in this volume region, and in the usual presentation of the data the biphasic relationship between Raw and H therefore remains concealed. Although this pattern could have been observed in previous investigations in which the Newtonian resistance and the tissue elastance were estimated at different levels of Prs or transpulmonary pressure (14, 22, 24), in consequence of both the wide pressure range and the multitude of Zrs estimates the present study is the first to document it. The utility of these plots nevertheless depends on the interpretation of the relationships displayed. In the high-volume region, the Raw vs. H relationship expresses the coupling between the stiffness of the tissues of the chest wall and the lungs and the effective cross-sectional area of the airways embedded in the latter, and it is therefore connected with the relative distensibility of these compartments. The underlying mechanisms are more complex in the low-volume range, where both continuous processes (changes in airway diameters and tissue nonlinearities) and discrete events (closure and reopening of airways and alveoli) may be involved in the parallel changes in Raw and H. Extension of the deflation part of the maneuver to negative mean values of Prs, such as those maintained in other studies (1, 19), and separate measurement of the pulmonary and chest wall compartments (1, 17) may help clarify the contributing processes. All these concurrent mechanisms result in remarkably linear Raw vs. H relationships in this region, which is separated from the "high-volume" set of data by the points of minimum elastance. It is also interesting to note that these points, corresponding to 40% of maximum V, are located around the maximum slope, i.e., the inflexion points of the pressure-volume loops (data not shown); thus they represent the joint minima of the dynamically measured (oscillatory) and quasi-static elastance values.
In conclusion, the continuous estimation of respiratory mechanical parameters from the tracking of multiple-frequency Zrs spectra during slow TLC maneuvers is feasible in anesthetized and intubated, but nonparalyzed mice. This method allows the fast revelation of the dependences of the mechanical parameters on Prs and V, and their patterns of interdependence, and hence it is well suited to study the alterations in respiratory mechanics during challenges of the bronchopulmonary system.
| DISCLOSURES |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
D. On the imperfect elasticity of lung tissue. J Appl Physiol 67: 2408-2419, 1989.
1-Antitrypsin determines the pattern of emphysema and function in tobacco smoke-exposed mice. Parallels with human disease. Am J Respir Crit Care Med 166: 1596-1603, 2002.
This article has been cited by other articles:
![]() |
G. R. Zosky, V. Cannizzaro, Z. Hantos, and P. D. Sly Protective mechanical ventilation does not exacerbate lung function impairment or lung inflammation following influenza A infection J Appl Physiol, November 1, 2009; 107(5): 1472 - 1478. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. T. Bates, M. Rincon, and C. G. Irvin Animal models of asthma Am J Physiol Lung Cell Mol Physiol, September 1, 2009; 297(3): L401 - L410. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Petak, G. Albu, E. Lele, Z. Hantos, D. R. Morel, F. Fontao, and W. Habre Lung mechanical and vascular changes during positive- and negative-pressure lung inflations: importance of reference pressures in the pulmonary vasculature J Appl Physiol, March 1, 2009; 106(3): 935 - 942. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Hantos, A. Adamicza, T. Z. Janosi, M. V. Szabari, J. Tolnai, and B. Suki Lung volumes and respiratory mechanics in elastase-induced emphysema in mice J Appl Physiol, December 1, 2008; 105(6): 1864 - 1872. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Habre, A. Adamicza, E. Lele, T. Novak, P. D. Sly, and F. Petak The Involvement of Histaminic and Muscarinic Receptors in the Bronchoconstriction Induced by Myorelaxant Administration in Sensitized Rabbits Anesth. Analg., December 1, 2008; 107(6): 1899 - 1906. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. R. Zosky, T. Z. Janosi, A. Adamicza, E. M. Bozanich, V. Cannizzaro, A. N. Larcombe, D. J. Turner, P. D. Sly, and Z. Hantos The bimodal quasi-static and dynamic elastance of the murine lung J Appl Physiol, August 1, 2008; 105(2): 685 - 692. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. T. Bates Point:Counterpoint: Lung impedance measurements are/are not more useful than simpler measurements of lung function in animal models of pulmonary disease J Appl Physiol, November 1, 2007; 103(5): 1900 - 1901. [Full Text] [PDF] |
||||
![]() |
C. von Garnier, M. E. Wikstrom, G. Zosky, D. J. Turner, P. D. Sly, M. Smith, J. A. Thomas, S. R. Judd, D. H. Strickland, P. G. Holt, et al. Allergic Airways Disease Develops after an Increase in Allergen Capture and Processing in the Airway Mucosa J. Immunol., November 1, 2007; 179(9): 5748 - 5759. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. L. Martin, T. A. Sheikh, K. J. Leco, J. F. Lewis, and R. A. W. Veldhuizen Contribution of alveolar macrophages to the response of the TIMP-3 null lung during a septic insult Am J Physiol Lung Cell Mol Physiol, September 1, 2007; 293(3): L779 - L789. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Bozanich, T. Z. Janosi, R. A. Collins, C. Thamrin, D. J. Turner, Z. Hantos, and P. D. Sly Methacholine responsiveness in mice from 2 to 8 wk of age J Appl Physiol, August 1, 2007; 103(2): 542 - 546. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Huang, W. Mitzner, R. Rabold, B. Schofield, H. Lee, S. Biswal, and C. G. Tankersley Variation in senescent-dependent lung changes in inbred mouse strains J Appl Physiol, April 1, 2007; 102(4): 1632 - 1639. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Z. Janosi, A. Adamicza, G. R. Zosky, T. Asztalos, P. D. Sly, and Z. Hantos Plethysmographic estimation of thoracic gas volume in apneic mice J Appl Physiol, August 1, 2006; 101(2): 454 - 459. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Petak, Z. Hantos, A. Adamicza, H. Gality, and W. Habre Development of bronchoconstriction after administration of muscle relaxants in rabbits with normal or hyperreactive airways. Anesth. Analg., July 1, 2006; 103(1): 103 - 109. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Bozanich, R. A. Collins, C. Thamrin, Z. Hantos, P. D. Sly, and D. J. Turner Developmental changes in airway and tissue mechanics in mice J Appl Physiol, July 1, 2005; 99(1): 108 - 113. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||