|
|
||||||||
1 Department of Emergency Medicine, The University of Alabama at Birmingham, Birmingham 35249; and 2 Department of Physics, The University of Alabama in Huntsville, Huntsville, Alabama 35899
| |
ABSTRACT |
|---|
|
|
|---|
The objective was to test
calibration of an eye oximeter (EOX) in a vitiligo swine eye and
correlate retinal venous oxygen saturation (SrvO2),
mixed venous oxygen saturation (SvO2), and cardiac
output (CO) during robust changes in blood volume. Ten anesthetized
adult Sinclair swine with retinal vitiligo were placed on stepwise
decreasing amounts of oxygen. At each oxygen level, femoral artery
oxygen saturation (SaO2) and retinal artery oxygen saturation (SraO2) were obtained. After equilibration
on 100% O2, subjects were bled at 1.4 ml · kg
1 · min
1
for 20 min. Subsequently, anticoagulated shed blood was reinfused at
the same rate. During graded hypoxia, exsanguination, and reinfusion, SraO2 and SrvO2 were measured by
using the EOX, and CO and SvO2 were measured by using
a pulmonary artery catheter. During graded hypoxia,
SraO2 correlated with SaO2
(r = 0.92). SrvO2 correlated with
SvO2 (r = 0.89) during exsanguination
and reinfusion. SvO2 and SrvO2
correlated with CO during blood removal and resuscitation (r = 0.92). Use of vitiligo retinas improved the
calibration of EOX measurements. In this robust hemorrhage model,
SrvO2 correlates with CO and SvO2
across the range of exsanguination and resuscitation.
retinal oximetry; noninvasive monitoring; swine; calibration
| |
INTRODUCTION |
|---|
|
|
|---|
SEVERAL KEY PARAMETERS ARE indicators of blood loss; these include central venous oxygen saturation, cardiac output, serum lactate, and gastric pH (2, 32, 40, 53). Perhaps the most important of these in the situation of acute blood loss is cardiac output (8, 19, 32, 38). However, cardiac output alone can be deceptive as is demonstrated by increased cardiac output during septic and spinal shock (32). Indeed, when autoregulation fails, cardiac output may increase as the patient's condition worsens. However, in the situation of acute blood loss, as normally seen in trauma, autoregulation is maintained and cardiac output is a principal indicator of decreasing blood volume associated with blood loss (32).
The clinician faced with a patient in shock or impending shock attempts to improve oxygen delivery, especially to the central organs, by improving cardiac output or blood oxygen-carrying capacity (40, 54). Several outcome studies have shown that central venous oxygen saturation is a reliable index of oxygen delivery, facilitating these clinical decisions (1, 36, 39, 41, 50). Because obtaining mixed venous oxygen saturation (SvO2) requires invasive monitoring and has associated complications (7, 39, 40), a noninvasive, rapidly applicable technique that provides a comparably reliable index of oxygen delivery from early blood loss to profound shock would be a valuable adjunct to patient management (1, 39, 41, 54).
The large vessels of the retina are a potential source of noninvasive perfusion data and are relatively easily observed (10, 11, 25). Previous attempts at retinal vessel oximetry were able to detect changes in oxygen saturation as small as ±4% (6, 10, 12, 26). These devices were not used to monitor retinal saturation changes during shock states and were not reduced to clinical use. Studies of the retinal circulation have shown a strong correlation between retinal perfusion and regional cerebral blood flow (21, 30). The blood flow to the central circulation including the retina (30) is relatively preserved during shock states (40, 41), and we have demonstrated that retinal venous oxygen saturation (SrvO2) is sensitive to early blood loss in anesthetized swine (14, 15).
Over the last seven years, we have developed an experimental, noninvasive eye oximeter (EOX). The device is used to noninvasively measure the optical density of the large retinal vessels, arteries, and veins (44) and uses a spectroscopic model to calculate their oxygen saturation.
In a pilot test of the device in which swine were bled at 0.4 ml · kg
1 · min
1
to a total of 16 ml/kg, there was a strong correlation between blood
loss and SrvO2 (r =
0.93)
(14). In another study, SrvO2 correlated
with the rate of blood loss during three intermediate rates of early
blood loss and subsequent reinfusion and was more sensitive to blood
loss than vital sign measurements in the same animals
(15).
Published studies using an EOX to measure SrvO2 during exsanguination used a mild amount of blood removal (20% of total blood volume) and did not demonstrate calibration of measurements (14, 15, 44, 45). We believe that calibration of data is desirable because it may allow for a single measurement to be used when making decisions about resuscitation or triage. Trending data can be used to monitor patients, but this requires repeated measures over time in the initial period of resuscitation when decisions must be made rapidly if a patient is to survive (2, 54). After an extensive review of the literature, we are unable to find any reports comparing SrvO2 to cardiac output, SvO2, or blood volume across the range of profound blood removal and subsequent reinfusion. It is unknown what changes will occur in SrvO2 measurements during more life threatening blood removal and during resuscitation from blood removal.
Studies using a model eye and the human eye have demonstrated that the highly absorbing fundus and pigment variation in the retina make calibration of oximetry measurements difficult (3, 17, 46). We have performed experiments using a model eye with a highly reflective background that allowed for calibration in this model (13). We have also performed experiments in a model eye and in the human eye that utilized a detection pathway filter to correct for fundus pigmentation (17, 46). However, in this study, we used a device without a detection pathway filter to study swine from Sinclair Research that spontaneously develop melanoma and then reject the tumor during adolescence. These swine sometimes develop vitiligo of the retina, making them an ideal test subject because increasing fundus reflectivity increases the effective light pathlength and simplifies our data reduction (33). We hypothesized that utilizing an animal model that had retinal vitiligo would increase retinal reflectivity similar to our model eye (i.e., have a highly reflective background) and consequently improve the calibration of our data. We also hypothesized that SrvO2 changes would correlate with cardiac output and SvO2 during profound exsanguination (40% of total blood volume) and subsequent resuscitation.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
The EOX scans low-power lasers across the retinal vasculature. The light scattered and reflected back out of the eye is collected and analyzed, and the optical density of the blood within the vessels is determined from the collected signals. These optical density measurements are made at multiple wavelengths, and a spectroscopic model is used to calculate the oxygen saturation of the blood within the vessels (6, 10, 26, 44, 52).
Through an eyepiece, the EOX provides an image of the subject's ocular fundus to the operator. The operator then targets a retinal artery or vein and initiates the measurement procedure. A full data set is acquired within 0.1 s. Typically, 8-16 data sets are averaged to comprise a single saturation determination (44). A detailed description of the device used for these experiments is provided elsewhere (45). This study adhered to National Institutes of Health guidelines for the use of laboratory animals and was approved by the Institutional Animal Care and Use Committee.
Ten mature Sinclair swine (2-6 yr old) with retinal vitiligo,
weighing 55-95 kg, were fasted overnight but allowed water ad libitum. On the morning of the experiment, the animals were given intramuscular preanesthetic ketamine 50 mg/kg and xylazine 2 mg/kg. The
swine were placed in the supine position, intubated endotracheally, and
placed on a ventilator. The swine were placed on 2-4% isoflurane during the surgical procedures, and the depth of anesthesia was monitored by using web space stimulation. An esophageal temperature probe was used to monitor core body temperature, and continuous electrocardiographic monitoring was utilized. The eyes were treated with two drops of 1% cyclopentolate hydrochloride. At the beginning of
the surgical preparation, the animal was given a bolus of 1,000 ml of
normal saline. A solution of 5% dextrose in half normal saline with 10 milliequivalents of KCl per liter was infused at 80-110 ml/h. A
celiotomy was performed by using an infraumbilical approach, the
bladder was exposed, and a Foley catheter was placed in the bladder via
cystotomy. The abdominal wall was closed around the bladder catheter. A
femoral cut down was performed, a 7.0-Fr catheter was placed in the
femoral artery, and an 8.0-Fr introducer was placed in the femoral
vein. The femoral artery catheter was connected to a Hewlett-Packard
78203 physiological pressure monitoring system, and a 7.5-Fr Abbott
continuous mixed SvO2 monitoring catheter was placed
in the central circulation via the introducer in the femoral vein. The
distal port of the central venous catheter was connected to a
Hewlett-Packard 78203 physiological pressure monitoring system.
Placement of the central venous catheter was verified by waveform. The
catheter oximeter calibration was verified by using mixed venous blood
obtained from the distal port. All blood gas analysis was performed by
use of an IL 482 CO-oximeter system. The eyelids were sutured open, and
sutures were placed in the conjunctiva to hold the eye in place. A
catheter, attached to a 60-ml syringe filled with buffered 0.9% saline
(pH = 7.0), was sutured to the periocular skin and used to bathe
the eye every 30-45 s to maintain corneal hydration throughout the
experimental protocol. When the surgical preparation was completed, the
isoflurane was decreased to 1.0-1.5% as needed to maintain
anesthesia. The respiratory rate was adjusted such that arterial
CO2 tension was between 36 and 44 Torr and the blood pH was
between 7.35 and 7.45. The end-tidal CO2 was measured
continuously by using a Datex 254 airway gas monitor. The central
venous catheter placed in the pulmonary artery was used to record
continuous mixed SvO2, and the thermodilution
technique was used to measure cardiac output at 2-min intervals during
exsanguination and reinfusion. The EOX was aimed at a large artery near
the optic disk, and the swine were placed on stepwise decreasing
amounts of oxygen. At each level of oxygen, femoral artery oxygen
saturation and retinal artery oxygen saturation were obtained. After
the arterial study was complete, the animals were placed on 100%
oxygen and allowed to equilibrate for 20 min. The EOX was then aimed at
a large vein near the optic disk, and the SrvO2 was
measured every minute for 5 min to obtain baseline data. After the
baseline data were acquired, each animal was exsanguinated at 1.4 ml · kg
1 · min
1
until a total of 28 ml/kg had been removed. Shed blood was
anticoagulated by using anticoagulant citrate phosphate dextrose (ACD)
solution. When the exsanguination was complete, the animal was
resuscitated by reinfusing the anticoagulated shed blood at 1.4 ml · kg
1 · min
1.
After the exsanguination and reinfusion, the retina was examined for laser damage by using direct ophthalmoscopy. At the conclusion of the experimental protocol, the anesthetized swine were euthanized by using supersaturated potassium chloride.
| |
RESULTS |
|---|
|
|
|---|
Retinal artery oxygen saturation (SraO2)
correlated with femoral artery oxygen saturation during graded hypoxia
(r = 0.92). SrvO2 correlated with
SvO2 (r = 0.89), blood volume
(r = 0.9), and cardiac output (r = 0.92) during the baseline blood loss and resuscitation periods of the
experiment. Figure 1 shows how
SraO2 correlated with femoral artery oxygen saturation
during graded hypoxia. The linear best fit for the conglomerate data
from all the swine tested for arterial calibration is shown as the
heavy line on the figure (r = 0.92, slope = 0.93, intercept = 0.03). The data set from each individual swine is
shown by using a different symbol, and the linear best fit for each
animal is shown as a fine line. The average residual retinal artery
saturation for this calibration plot was 1.5 ± 8.9%. Figure
2 shows the tight correlation of
SrvO2 and normalized cardiac output (the cardiac output at the beginning of the baseline period divided into each subsequent measurement in that swine) measured during blood removal and
resuscitation. For comparison purposes, Fig.
3 shows SvO2 and
normalized cardiac output. Figure 4 shows
the correlation plot of SrvO2 with normalized cardiac
output during exsanguination and reinfusion; the data shown are
extracted from Fig. 2 and are the average cardiac output and
SrvO2 measured concurrently with the cardiac output
during exsanguination and resuscitation. Figure 5 shows the correlation between average
SrvO2 and SvO2 during exsanguination
(r = 0.97) and resuscitation (r = 0.82). All error bars shown and reported ranges are the standard
deviation from the mean.
|
|
|
|
|
| |
DISCUSSION |
|---|
|
|
|---|
We performed model eye experiments by using a highly reflective background that demonstrated improved calibration in our instrument. On the basis of these results, we expected improved calibration in this in vivo retinal vitiligo model. The need for calibration to a range of 3 to 4% is evident from our exsanguination and reinfusion experiments in which a change of 10% oxygen saturation correlated with a 5% change in blood volume (15). The hypothesis that retinal vitiligo would improve calibration of our measurements is supported by our data. In previous studies using Yorkshire swine with normal retinal pigmentation, we had significant variations in slopes and intercepts of our calibration plots (14), and the overall correlation coefficient when all data from all swine were plotted together was significantly inferior to our results here (with the Yorkshire swine r = 0.646, slope = 0.57, intercept = 0.235). However, the average residuals from our calibration experiment had an error of ±8.9% saturation; thus we were still unable to achieve our goal of 3 or 4% error in calibration. As noted in the introduction, our laboratory utilized detection pathway filters and scanning systems (46) to address this concern (17).
There have been several devices advocated as noninvasive systems for the direction of trauma resuscitation. Previously, noninvasive blood oxygenation measurements have been attempted (29, 35). Various limitations exist with these approaches. Near infrared spectroscopy is potentially erroneous because differences in skull and scalp thickness alter pathlength (29). Pulse oximetry is sometimes inaccurate as a result of optical shunting (presence of light that alters the true reading) (28, 48) and is sometimes associated with thermal injury (34, 43, 47). The device measures peripheral arterial oxygenation only (42) and is of limited efficacy in patients with anemia or hypoxia (27, 42). More recently, gastric tonometry, retinal oximetry, and sublingual capnometry have been tested in animal and human models (17, 35a, 45, 53).
In this swine model, we have demonstrated a strong correlation between cardiac output and SrvO2 across the spectrum of survivable blood loss and resuscitation (40% of total blood volume). This is important because vital signs may be maintained until large amounts of blood have been lost. To be used for clinical care, a monitoring tool must be superior to present technology and must change predictably across the breadth of the clinical spectrum to allow the clinician to make accurate decisions about care. Also, calibration is important because a single measurement, in conjunction with vital signs and clinical assessment, may be all that is available for critical decision making during time-sensitive trauma resuscitation (2).
The change in SrvO2 and central venous oxygen saturation seen during resuscitation demonstrates the monitoring capacity of these tools (see Figs. 2 and 3). Both indicated a rapid response to initial reinfusion and a marked flattening of this response after ~5-6% of total blood volume had been infused. The nonlinear response to the reinfusion of autologous blood seen in this model has been described in a study using autologous blood transfusions in swine (20) and is similar to changes that we have seen during resuscitation from mild blood loss (14, 15). During resuscitation, SrvO2 increased more dramatically than SvO2 and behaved more like cardiac output than SvO2. In Fig. 5, where SrvO2 and SvO2 are correlated, the bimodal grouping of data is evident as SrvO2 and SvO2 remain correlated but the slope changes during resuscitation. This is probably a result of total body autoregulation shunting flow to the central organs and may represent an advantage in monitoring SrvO2 because the resuscitation of trauma victims before definitive repair using traditional end points can lead to overresuscitation and increased mortality in swine and humans (4, 9).
In conclusion, we have demonstrated improved calibration of a spectroscopic EOX achieved by use of an in vivo model of increased retinal reflectivity by using swine with retinal vitiligo. We have also demonstrated the use of an experimental EOX during profound blood loss and resuscitation in a specialized anesthetized swine model. Changes in SrvO2 correlated with blood volume, SvO2, and cardiac output during profound blood removal and subsequent resuscitation. The response to changes in blood volume was nearly linear during exsanguination. There was a nonlinear response to the reinfusion of autologous blood seen in this model. This nonlinear response has been described elsewhere and is probably a physiological response to autologous blood transfusions and physiological hysteresis in swine (20). Use of a calibrated EOX to noninvasively monitor trauma patients for unrecognized blood loss and during resuscitation from exsanguinating hemorrhage warrants further study.
| |
ACKNOWLEDGEMENTS |
|---|
We thank The University of Alabama at Birmingham Injury Control Research Center for support. We also thank Sharon Tyra, Senior Research Associate, UAB Department of Surgery, for technical support.
| |
FOOTNOTES |
|---|
This work was supported by Department of the Army Grant DAMD 17-98-1-8007 and Office of Naval Research Grant ONR N00014-99-1-0226 and by the Centers for Disease Control Grant R49/CCR403641.
Address for reprint requests and other correspondence: K. R. Denninghoff, Dept. of Emergency Medicine, The Univ. of Alabama at Birmingham, JTN 266, 619 South 20th St., Birmingham, AL 35233-7013 (E-mail: kdenning{at}uabmc.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
10.1152/japplphysiol.01197.2001
Received 4 December 2001; accepted in final form 14 October 2002.
| |
REFERENCES |
|---|
|
|
|---|
1.
Abou-Khalil, B,
Scalea TM,
Trooskin SZ,
Henry SM,
and
Hitchcock R.
Hemodynamic responses to shock in young trauma patients: need for invasive monitoring.
Crit Care Med
22:
633-639,
1994[Web of Science][Medline].
2.
Advance Trauma Life Support Student Manual (6th ed.). Chicago, IL: American College of Surgeons, 1997, p. 89-91, 97-100.
3.
Beach, JM,
Schwenzer KJ,
Srinivas S,
Kim D,
and
Tiedeman JS.
Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation.
J Appl Physiol
86:
748-758,
1999
4.
Bickell, WH,
Bruttig SP,
Millnamow GA,
O'Benar J,
and
Wade CE.
The detrimental effects of intravenous crystalloid after aortotomy in swine.
Surgery
110:
529-536,
1991[Web of Science][Medline].
5.
Carneiro, JJ,
and
Donald D.
Blood reservoir function of dog spleen, liver, and intestine.
Am J Physiol Heart Circ Physiol
232:
H67-H72,
1977
6.
Cohen, AJ,
and
Laing RA.
Multiple scattering analysis of retinal blood oximetry.
IEEE Trans Biomed Eng
23:
391-400,
1976[Web of Science][Medline].
7.
Connors, AF, Jr,
Speroff T,
Dawson NV,
Thomas C,
Harrell FE, Jr,
Wagner D,
Desbiens N,
Goldman L,
Wu AW,
Califf RM,
Fulkerson WJ, Jr,
Vidaillet H,
Broste S,
Bellamy P,
Lynn J,
and
Knaus WA.
The effectiveness of right heart catheterization in the initial care of critically ill patients.
JAMA
276:
889-897,
1996
8.
Convertino, VA.
Gender differences in autonomic functions associated with blood pressure regulation.
Am J Physiol Regul Integr Comp Physiol
275:
R1909-R1920,
1998
9.
De Guzman, E,
Shankar MN,
and
Mattox KL.
Limited volume resuscitation in penetrating thoracoabdominal trauma.
AACN Clin Iss
10:
61-68,
1999.
10.
Delori, FC.
Noninvasive technique for oximetry of blood in retinal vessels.
Appl Optics
27:
1113-1125,
1988.
11.
Delori, FC,
Gragoudas ES,
Francisco R,
and
Pruett RD.
Monochromatic ophthalmoscopy and fundus photography. The normal fundus.
Arch Ophthalmol
9:
861-868,
1977.
12.
Delori, FC,
and
Pflibsen KP.
Spectral reflectance of the human ocular fundus.
Appl Optics
28:
1061-1077,
1989.
13.
Denninghoff, KR,
and
Smith MH.
Optical model of the blood in large retinal vessels.
J Biomed Opt
5:
1-4,
2000.
14.
Denninghoff, KR,
Smith MH,
Chipman RA,
Hillman LW,
Jester PM,
Hughes CE,
Kuhn F,
and
Rue LW.
Retinal large vessel oxygen saturations correlates with early blood loss and hypoxia in anesthetized swine.
J Trauma
43:
29-34,
1997[Web of Science][Medline].
15.
Denninghoff, KR,
Smith MH,
Hillman LW,
Redden D,
and
Rue LW.
Retinal venous oxygen saturation correlates with blood volume.
Acad Emerg Med
5:
577-582,
1998[Web of Science][Medline].
16.
De Schaepdrijver, L,
Simoens P,
Pollet L,
Lauwers H,
and
Laey JJ.
Morphologic and clinical study of the retinal circulation in the miniature pig. B: fluorescein angiography of the retina.
Exp Eye Res
54:
975-985,
1992[Web of Science][Medline].
17.
Drewes, J,
Smith MH,
Denninghoff KR,
and
Hillman LW.
An instrument for the measurement of retinal vessel oxygen saturation. Optical Diagnostics of Biological Fluids IV.
Proc SPIE
3591:
114-120,
1999.
18.
Dries, DJ,
and
Waxman K.
Adequate resuscitation of burn patients may not be measured by urine output and vital signs.
Crit Care Med
19:
327-329,
1991[Web of Science][Medline].
19.
Engelke, KA,
Doerr DF,
and
Convertino VA.
Application of acute maximal exercise to protect orthostatic tolerance after simulated microgravity.
Am J Physiol Regul Integr Comp Physiol
271:
R837-R847,
1996
20.
Filos, KS,
Vagianos CE,
Stavropoulos M,
Tassoudis V,
Patroni O,
Fligou F,
Goudas LC,
and
Androulakis J.
Evaluation of the effects of autotransfusion of unprocessed blood on hemodynamics and oxygen transport in anesthetized pigs.
Crit Care Med
2:
855-861,
1996.
21.
Harris, A,
Arend O,
Kopecky K,
Caldemeyer K,
Wolf S,
Sponsel W,
and
Martin B.
Physiological perturbation of ocular and cerebral blood flow as measured by scanning laser ophthalmoscopy and color Doppler imaging.
Surv Ophthalmol, Suppl
38:
S81-S86,
1994.
22.
Hartwig, H,
and
Hartwig HG.
Structural characteristics of the mammalian spleen indicating storage and release of red blood cells. Aspects of evolutionary and environmental demands.
Experientia
41:
159-163,
1985[Web of Science][Medline].
23.
H.C.I.A.: Hospital Inpatient Charges. Baltimore, MD: Healthcare Knowledge Resources, 1993.
25.
Hickam, JB,
and
Frayser R.
Studies of the retinal circulation in man: observations on vessel diameter, arteriovenous oxygen difference, and mean circulation time.
Circulation
33:
302-316,
1966
26.
Hickam, JB,
Frayser R,
and
Ross J.
A study of retinal venous oxygen saturation in human subjects by photographic means.
Circulation
27:
375-385,
1963
27.
Jay, GD,
Hughes L,
and
Renzi FP.
Pulse oximetry is accurate in acute anemia from hemorrhage.
Ann Emerg Med
24:
32-35,
1994[Web of Science][Medline].
28.
Kelleher, JF,
and
Ruff RH.
The penumbra effect: vasomotion-dependent pulse oximeter artifact due to probe malposition.
Anesthesiology
71:
787-791,
1989[Web of Science][Medline].
29.
Kurth, CD,
Steven JM,
Benaron D,
and
Chance B.
Near- infrared monitoring of the cerebral circulation.
J Clin Monit
9:
163-170,
1993[Web of Science][Medline].
30.
Laughlin, MH,
Witt WM,
and
Whittaker RN.
Regional cerebral blood flow in conscious miniature swine during high sustained +Gz acceleration stress.
Aviat Space Environ Med
50:
1129-1133,
1979[Medline].
31.
Luna, GK,
Eddy AC,
and
Copass M.
The sensitivity of vital signs in identifying major thoracoabdominal hemorrhage.
Am J Surg
157:
512-515,
1989[Web of Science][Medline].
32.
McNutt, S,
Denninghoff KR,
and
Terndrup T.
Shock: rapid recognition and appropriate ED intervention.
Emerg Med Pract
2:
1-24,
2000.
33.
Misfeldt, ML,
and
Grimm DR.
Sinclair miniature swine: an animal model of human melanoma.
Vet Immunol Immunopathol
43:
167-175,
1994[Web of Science][Medline].
34.
Murphy, KG,
Secunda JA,
and
Rockoff MA.
Severe burns from a pulse oximeter.
Anesthesiology
7:
350-352,
1990.
35.
Poets, CF,
and
Southall DP.
Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern.
Pediatrics
93:
737-746,
1994
35a.
Povoas, H,
Weil MH,
Tang W,
Moran B,
Kamohara T,
and
Bisera J.
Comparisons between sublingual and gastric tonometry during hemorrhagic shock.
Chest
118:
1127-1132,
2000
36.
Rady, M,
Rivers EP,
Martin GB,
Smithline H,
Appelton T,
and
Nowak RM.
Continuous central venous oximetry and shock index in the emergency department.
Am J Emerg Med
10:
538-541,
1992[Web of Science][Medline].
37.
Rasanen, J.
Supply-dependent oxygen consumption and mixed venous oxyhemoglobin saturation during isovolemic hemodilution in pigs.
Chest
101:
1121-1124,
1992
38.
Sather, TM,
Goldwater DJ,
Montgomery LD,
and
Convertino VA.
Cardiovascular dynamics associated with tolerance to lower body negative pressure.
Aviat Space Environ Med
57:
413-419,
1986[Medline].
39.
Scalea, TM,
Hartnett RW,
Duncan AO,
Atweh NA,
Phillips TF,
Sclafani SJ,
Fuortes M,
and
Shaftan GW.
Central venous oxygen saturation: a useful clinical tool in trauma patients.
J Trauma
30:
1539-1543,
1990[Web of Science][Medline].
40.
Scalea, TM,
Holman M,
Fuortes M,
Baron BJ,
Phillips TF,
Goldstein AS,
Sclafani SJ,
and
Shaftan GW.
Central venous blood oxygen saturation: an early, accurate measurement of volume during hemorrhage.
J Trauma
28:
725-732,
1988[Web of Science][Medline].
41.
Scalea, TM,
Simon HM,
Duncan AO,
Atweh NA,
Sclafani SJ,
Phillips TF,
and
Shaftan GW.
Geriatric blunt multiple trauma: improved survival with early invasive monitoring.
J Trauma
30:
129-136,
1990[Web of Science][Medline].
42.
Severinghaus, JW,
and
Spellman MJ.
Pulse oximeter failure thresholds in hypotension and vasoconstriction.
Anesthesiology
7:
532-537,
1990.
43.
Sloan, TB.
Finger injury by an oxygen saturation monitor probe.
Anesthesiology
68:
936-938,
1988[Web of Science][Medline].
44.
Smith MH, Denninghoff KR, Hillman LW, et al. Technique for
noninvasive monitoring of blood loss via oxygen saturation measurements
in the eye. Invited paper in Optical Diagnostics of Biological
Fluids II, Proc SPIE 2982: 46-52, 1997.
45.
Smith, MH,
Denninghoff KR,
Hillman LW,
and
Chipman RA.
Oxygen saturation measurements of blood in retinal vessels during blood loss.
J Biomed Opt
3:
296-303,
1998.
46.
Smith, MH,
Denninghoff KR,
Lompado A,
Woodruff JB,
and
Hillman LW.
Minimizing the influence of fundus pigmentation on retinal vessel oximetry measurements. Ophthalmic Technologies XI.
Proc SPIE
4245:
135-145,
2001.
47.
Sobel, DB.
Burning of a neonate due to a pulse oximeter: arterial saturation monitoring.
Pediatrics
89:
154-156,
1992
48.
Southall, DP,
and
Samuels M.
Inappropriate sensor application in pulse oximetry.
Lancet
340:
481-482,
1992[Web of Science][Medline].
50.
Swan, H,
Sanchez M,
Tyndall M,
and
Koch C.
Quality control of perfusion: monitoring venous blood oxygen tension to prevent hypoxic acidosis.
J Thorac Cardiovasc Surg
99:
868-872,
1990[Abstract].
51.
Trouwborst, A,
Tenbrinck R,
and
van Woerkens E.
Blood gas analysis of mixed venous blood during normoxic acute isovolemic hemodilution in pigs.
Anesth Analg
70:
523-529,
1990
52.
Van Assendelft, OW.
Spectrophotometry of Haemoglobin Derivatives. Assen, The Netherlands: Royal Van Gorcum, 1970.
53.
Weil, MH,
Nakagawa Y,
Tang W,
Sato Y,
Ercoli F,
Finegan R,
Grayman G,
and
Bisera J.
Sublingual capnometry: a new noninvasive measurement for diagnosis and quantitation of severity of circulatory shock.
Crit Care Med
27:
1225-1229,
1999[Web of Science][Medline].
54.
Wo, CJ,
Shoemaker WC,
Appel PL,
Bishop MH,
Kram HB,
and
Hardin E.
Unreliability of blood pressure and heart rate to evaluate cardiac output in emergency resuscitation and critical illness.
Crit Care Med
21:
218-223,
1993[Web of Science][Medline].
This article has been cited by other articles:
![]() |
M. Shahidi, J. Wanek, N. P. Blair, and M. Mori Three-Dimensional Mapping of Chorioretinal Vascular Oxygen Tension in the Rat Invest. Ophthalmol. Vis. Sci., February 1, 2009; 50(2): 820 - 825. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Hardarson, A. Harris, R. A. Karlsson, G. H. Halldorsson, L. Kagemann, E. Rechtman, G. M. Zoega, T. Eysteinsson, J. A. Benediktsson, A. Thorsteinsson, et al. Automatic Retinal Oximetry Invest. Ophthalmol. Vis. Sci., November 1, 2006; 47(11): 5011 - 5016. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Mellem-Kairala, A. E. Elsner, A. Weber, R. B. Simmons, and S. A. Burns Improved Contrast of Peripapillary Hyperpigmentation Using Polarization Analysis Invest. Ophthalmol. Vis. Sci., March 1, 2005; 46(3): 1099 - 1106. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |