Vol. 93, Issue 4, 1515-1526, October 2002
INNOVATIVE TECHNIQUES
Phonospirometry for noninvasive measurement of ventilation:
methodology and preliminary results
Cheng-Li
Que1,
Christof
Kolmaga1,
Louis-Gilles
Durand2,
Suzanne M.
Kelly1, and
Peter T.
Macklem1
1 Meakins-Christie Laboratories, Montreal Chest
Institute, Royal Victoria Hospital, Montreal, Quebec H2X 2P2;
2 Institut de Recherches Cliniques de
Montréal, University of Montreal, Montreal, Quebec,
Canada H2W 1R7
 |
ABSTRACT |
We measured tracheal flow from
tracheal sounds to estimate tidal volume, minute ventilation
(
I), respiratory frequency, mean inspiratory flow
(VT/TI), and duty cycle (TI/Ttot).
In 11 normal subjects, 3 patients with unstable airway obstruction, and
3 stable asthmatic patients, we measured tracheal sounds and flow
twice: first to derive flow-sound relationships and second to obtain
flow-volume relationships from the sound signal. The flow-volume
relationship was compared with pneumotach-derived volume. When subjects
were seated, facing forward and with neck rotation, flexion, and
standing, flow-volume relationship was within 15% of
pneumotach-derived volume. Error increased with neck extension and
while supine. We then measured ventilation without mouthpiece or nose
clip from tracheal sounds during quiet breathing for up to 30 min.
Normal results ± SD revealed tidal volume = 0.37 ± 0.065 liter, respiratory frequency = 19.3 ± 3.5 breaths/min,
I = 6.9 ± 1.2 l/min,
VT/TI = 0.31 ± 0.06 l/s, and TI/Ttot = 0.37 ± 0.04. Unstable airway obstruction had large
I due to
increased VT/TI. With the exception of
TI/Ttot, variations in ventilatory parameters were closer
to log normal than normal distributions and tended to be greater in
patients. We conclude that phonospirometry measures ventilation
reasonably accurately without mouthpiece, nose clip, or rigid postural constraints.
minute ventilation; pattern of breathing; breath sounds; tidal
volume; breathing frequency
 |
INTRODUCTION |
PRECISE MONITORING
OF VENTILATION is usually achieved by having a subject breathe
through a mouthpiece or face mask attached to a pneumotachygraph or
spirometer. Although these devices permit the accurate measurement of
ventilation and its parameters, they also alter the pattern of
breathing and minute ventilation (
I) (2, 12,
13, 37). They are not useful for monitoring ventilation in any
circumstance in which keeping a mouthpiece and nose clip in place is
too difficult or impossible.
Magnetometry and inductance plethysmography record the movements of the
rib cage and abdomen during respiration and, by the use of a suitable
calibration procedure, convert the summed thoracic and abdominal
motions into a volume signal (30, 34, 35). Although these
devices eliminate the need for a mouthpiece and nose clip and, as a
result, permit measurement of the volumes, flows, and timing of normal
breathing, they are unfortunately sensitive to changes in posture,
particularly xyphi-pubic distance (24, 34, 35, 38, 39).
Even when the latter is controlled, the rib cage and abdomen do not
always act as compartments with a single degree of freedom (7,
14, 21, 36), and this introduces a further source of error.
Motion analysis by optoelectronic plethysmography works extremely well
(5) but is quite expensive and constrains the subject
because markers on the trunk must be visualized by fixed videocameras.
There is a need for other devices that can measure ventilation cheaply
and accurately without a mouthpiece and nose clip.
Phonospirometry, the estimation of ventilation parameters from
measurements of tracheal breath sounds, provides a simple alternative to motion analysis systems and may prove to be more versatile. Since
the invention of the stethoscope by René Laennec in 1819, auscultation has provided the clinician with a quick, but crude, assessment of pulmonary ventilation. Objective measurements of breath
sounds were first made more than 25 yr ago (17) and have been used as a qualitative assessment of ventilation during sleep (8); but it is only with advances in computer technology
and the wide application of digital signal analysis that recording and
analysis of respiratory sounds has accelerated.
Normal breath sounds are primarily generated by turbulence within
large- and medium-sized airways. Airflow velocity and airway dimensions
influence the generation of turbulent flow, whereas the intensity of
the detected sound is influenced by the sound transmission
characteristics of the tissue between the regions of sound generation
and the point where measurements are made. As a result, for a given
subject and microphone position, sound amplitude is proportional to the
flow rate (10, 15, 17, 32, 33), suggesting the possibility
of deriving flow estimates from sound amplitude (33). In
this paper, we present a new method to measure ventilation by
converting tracheal breath sound intensity to ventilatory flow. By
integration, tidal volume (VT), respiratory frequency (f),
I, and other ventilatory parameters are obtained.
Many investigators have measured airflow by acoustical techniques
(1, 6, 10, 18, 19, 27, 28). However, their interest has
lain primarily in the derivation of parameters related to the frequency
spectrum of the sound signal, or the mechanism of sound production,
rather than in the use of the sound signal as a surrogate for flow.
With the exception of Soufflet et al. (33), the studies
quantifying the relationship between sound intensity and flow
(10, 32) have focused exclusively on flows of >0.5 l/s.
Our purpose was to estimate flow from breath sound intensity during
quiet breathing and, therefore, almost exclusively on flows of <0.5
l/s. This presented particular problems of an adverse signal-to-noise ratio (SNR) and the threshold of flow below
which flow-derived sound cannot be detected. These problems have not previously been satisfactorily addressed. We have solved these problems
by measuring the threshold of detection of flow-derived sounds (~0.3
l/s) and interpolating from the threshold to zero-flow points. Thus we
were able to measure VT,
I, f,
inspiratory (TI) and expiratory (TE) durations,
respiratory cycle time (Ttot), duty cycle (TI/Ttot), and
mean inspiratory flow (VT/TI) during quiet
breathing from the envelope of the sound signal with only a microphone
placed on the skin over the thyroid cartilage in the neck, without the
use of a mouthpiece or nose clip except for a calibration procedure. In
this paper we describe this technology and present preliminary results.
 |
METHODS |
Subjects.
A total of 11 nonsmoking normal volunteers with no history of pulmonary
disease or recent respiratory tract infection, 3 asymptomatic mild
asthmatic patients, and 3 patients with unstable airway obstruction attending the day hospital of the Montreal Chest Institute were recruited for the study. Of these, seven normal subjects and one asymptomatic asthmatic subject participated in the early studies in
which we validated phonospirometry as a method to measure ventilation, estimated the errors by a Bland-Altman analysis, and tabulated the
ventilatory pattern shown in Table 1. SNR
and the power spectrum of normal breath sounds were measured in one of
these subjects and in four additional normal volunteers. Effects of
posture were studied in 5 of the 11 normal subjects. Subsequently five
additional patients, two with mild asthma and three with unstable
airways obstruction, attending the day hospital of the Montreal Chest Institute, were studied to measure their ventilatory pattern and its
variations. These patients were studied primarily to get a feeling for
how applicable the technique is to disease rather than to make
conclusions as to the effect of disease on breathing pattern.
Calibration procedure.
Airflow at the mouth and tracheal breath sounds were measured
simultaneously during two separate 30-s intervals. The setup is
illustrated in Fig. 1. Subjects were
seated, wore a nose clip, and breathed quietly on a mouthpiece. An
electret microphone (model 1306, Armaco, Vancouver, Canada), coupled to
a custom-designed preamplifier with a 30-fold amplification, was placed
over the trachea at the level of the thyroid cartilage, a region that
had previously been determined to provide the best sound signal. The microphone was inserted in a conical aluminium housing 5.5 mm in
diameter around the microphone and 13 mm in diameter at the opening
that was placed on the skin. The distance between the microphone and
the opening was 3.9 mm. Frequency response of the electret microphone
was flat (±3 dB) between 20 Hz and 8 kHz. Flow signal was obtained by
using a pneumotachygraph (model no. 1A, Fleisch, Lausanne, Switzerland)
and a piezoelectric pressure transducer. Both flow and sound were
amplified and filtered before analog-to-digital conversion. For flow,
the cutoff was set at 50 Hz; for the sound signal, cutoff was at 1,000 Hz. Both signals were sampled at 3,000 Hz by using a commercially
available software package (ORIGIN, MICROCAL, Northampton, MA) and a
12-bit analog-to-digital converter (model DT-2801, Data Translation,
Marlboro, MA).

View larger version (27K):
[in this window]
[in a new window]
|
Fig. 1.
Diagram of data acquisition system. Tracheal breath
sounds are measured with an electret microphone (Mic), and flow is
measured with a pneumotach and pressure transducer. Both signals are
amplified and low-pass filtered (flow, 50 Hz; sound, 1 KHz) before
digitization and storage.
|
|
For one 30-s period, the relationship between flow and sound was
determined (see below). This relationship was then used to derive flow and volume from the amplitude of the sound signal of the
second measurement period. We assessed the accuracy of this acoustic
method by comparing the volume obtained from integration of the
measured flow (Vm) to the flow estimated from the sound signal (Ve).
This calibration procedure was done individually for each subject in
the seated position. When it was successfully completed, breath sounds
alone were used to estimate flow and the ventilatory parameters were
derived from this.
Effect of neck rotation and body posture on volume measurements.
Because head movements and changes of posture are to be expected during
long-term measurements of ventilation, we examined the correspondence
between Vm and Ve signals during systematic changes of head and body
position. First, in five normal subjects and one asymptomatic asthmatic
subject (SK), flow and sound signals were obtained during up-and-down
and right-to-left movements of the head. Subjects were asked to breathe
quietly for 10 s while facing forward, for 20 s while facing
left and right, respectively, and then for a final 10 s facing
forward for a second time. Second, maximum up-and-down movements of the
head were studied in a similar fashion, i.e., a subject breathed
quietly for 10 s with the head positioned centrally, for 20 s
each with the neck maximally extended then flexed, then finally for
10 s facing foward. At a f of 15 breaths/min, 20 s would be
sufficient to record about five breaths; in fact we found that f was
closer to 20 breaths/min so that with rotation, extension, and flexion
we measured between six and seven breaths.
To determine whether the calibration made seated could be used in
different postures, recordings of sound and flow were obtained during
two 30-s periods in each of three postures: sitting, standing, and
supine. We determined how accurately VT calculated from the sound envelope using the calibration made while seated estimated VT measured in the standing and supine postures by
integrating flow. In addition, the reproducibility of the calibration
was verified by repeating the measurements in the seated position at
the end of the experiment.
Data analysis.
Figure 2 is an example of the data
obtained from one subject during a 30-s calibration period. Only a part
of the record is illustrated. Shown from above downward are flow,
corresponding sound signal, filtered sound signal, and sound envelope.
During a single breath, two major sound bursts are observed (Fig.
2B), one during inspiration, the other during expiration.
Sounds in phase with the heartbeat are seen as sharp transients. The
digitized sound signal was band-pass filtered between 200 and 1,000 Hz
to remove heart and muscle sounds, which are typically <200 Hz, and the high-frequency noise, which is >1,000 Hz. Specifically, the digitized data was transformed to the frequency domain by using a
discrete Fourier transform (DFT). Then, all negative frequencies
1,000 Hz, all positive frequencies >1,000 Hz, and frequencies between
200 and 200 Hz were set to zero by using a three-point roll-off function. To provide a better attenuation in the rejection bands of the DFT-based band-pass filter, the three first frequency coefficients at the upper and lower boundaries of the pass bands were
weighted by the following coefficients: 0.022, 0.23, 0.70 (11). After band-pass filtering (Fig. 2C),
there was a marked attenuation of both the background noise and the
cardiac artifact, whereas the breath sound signal was well preserved.

View larger version (36K):
[in this window]
[in a new window]
|
Fig. 2.
A
segment of quiet breathing from 1 subject. A: flow measured
by the pneumotachygraph. B: tracheal sound signal.
C: sound signal after band-pass (200-1,000 Hz)
filtering. D: sound envelope obtained after filtering and
Hilbert transform.
|
|
To obtain sound intensity, we integrated the sound signal by applying
the Hilbert transform to our digital data (23). This was
done in conjunction with the frequency-domain filtering mentioned above
and required that frequencies between
200 and
1,000 Hz also be set
to zero. After an inverse DFT, the resulting envelope signal was
filtered in two steps to further reduce noise. First, data was
subsampled by averaging the 100 closest neighbors at every 50th data
point. Second, a 20-point second-order least-squares smoothing filter
was applied. The resulting signal is illustrated in Fig. 2D.
The flow signal was also subsampled by a factor of 50 to maintain the
same frequency as the sound signal.
In Fig. 3, the relationship between sound
amplitude and flow in eight normal subjects is plotted for all breaths
of a single measurement period. As found by others (33)
below ~0.3 l/s, sound amplitude did not exceed the background noise.
A threshold sound value above which the relationship between sound
amplitude and flow could be determined was thus defined. For each
subject, the threshold was set so as to exceed both the background
noise and the transition where tracheal sounds are just apparent. As a
result, there was an approximately linear relation between flow and
tracheal breath sound amplitude. Because this relationship was
different during inspiration and expiration, these intervals were
analyzed separately, and, for each of inspiration and expiration, the
relationship between sound amplitude and flow was determined by least
squares linear regression.

View larger version (31K):
[in this window]
[in a new window]
|
Fig. 3.
Relationship between flow and sound amplitude during
quiet breathing in 8 subjects (identified by 2-letter initials). Note
that, for a given flow, there is a wide between-subject variation in
sound amplitude. In addition, at low flows, the sound signal does not
exceed the background noise. Transfer function for conversion of sound
to flow was calculated for inspiration and expiration separately over
the quasilinear portion of the relationship.
|
|
Because the start and end of each inspiration can easily be
determined from the sound amplitude signal (Fig. 2D), flow
below the threshold value was estimated by linear interpolation between zero flow points and the first time points when flow could be obtained
from the sound signal (Fig. 4).
Interpolation fit the data well during inspiration but not during
expiration. Therefore, VT was estimated from inspiratory
sounds. As seen in Fig. 4, zero flow points were properly identified.

View larger version (21K):
[in this window]
[in a new window]
|
Fig. 4.
Comparison of measured (broken line) and calculated
(solid line) flows from the data segment in Fig. 2. The thinner solid
line is the interpolated flow. Note the good correspondence between the
estimated and the actual flow signals during inspiration (positive
deflections) and the early part of expiration (negative
deflections).
|
|
From the VT estimates as a function of time and the zero
flow points, we calculated f,
I, TI,
TE, VT/TI and TI/Ttot.
We constructed frequency distributions of all ventilatory parameters before and after log transformation. We measured means ± SD (µ and
for log-transformed data). Frequency was normalized by
expressing it as a fraction of the total number of measurements and for
choice of bin width, thus obtaining the probability density functions (PDF). We estimated skewness and kurtosis of the gaussian and log-normal distributions. We used an unpaired t-test to
compare ventilatory patterns between normal subjects and patients.
 |
RESULTS |
The relationship between sound intensity and flow in seven
normal subjects and an asthmatic subject (SK) is illustrated in Fig. 3.
The general shape is similar for all subjects, although there were
variations in sound intensity at a given flow from subject to subject
as well as in the flow value above which tracheal sounds could be
detected. In Fig. 4, flow derived from the sound signal (solid lines)
is compared with the measured flow signal (dotted line). The thick
segments represent the calculated flows, and the thin solid lines
represent the interpolated values. During inspiration, calculated flow
is quite close to the measured value and even reflects some of its
details. Expiratory flows are less well estimated from the sound signal
and, in this case, underestimate the true value. Similarly, the
interpolated values are more accurate during the entire inspiration
interval and the beginning of expiration, but the concave shape of the
actual flow curve during the latter part of expiration is poorly
estimated. This error was greatest when subjects had a long
end-expiratory pause.
To determine more precisely how sound intensity changed with flow, in
five of the normal subjects shown in Fig. 3 and in SK, we compared the
power spectrum of the sound signal obtained during a 15-s breath hold
to the spectra obtained during constant flow inspirations between 0.07 and 0.3 l/s. Data from a representative subject is illustrated in Fig.
5A. In this subject,
inspiratory sounds were detectable above the background noise at 0.15 l/s and progressively increased in amplitude as flow was increased. Similar results were seen in the other five subjects, but the threshold
for inspiratory sound detection differed from individual to individual.
SNR was calculated by dividing, at every frequency, the sound amplitude
during inspiration by the sound amplitude during the breath hold and
then averaging the values for all frequencies between 200 and 1,000 Hz.
These values, plotted as a function of the mean flow rate in Fig.
5B, show the intersubject variation in the threshold flow
rate above which sound could be detected above background. The
threshold value was generally specific to each subject. Above it was a
quasi-linear relationship between SNR and flow in all but one subject.

View larger version (25K):
[in this window]
[in a new window]
|
Fig. 5.
Signal-to-noise ratio (SNR) of the sound signal.
A: power spectrum of sound signals in a single subject
during a breath hold (dotted line) and during a series of inspiratory
flow rates. B: SNR in 6 subjects as a function of
inspiratory flow. Note that flow had to exceed a threshold value before
tracheal breath sounds could be detected. mean AR, mean amplitude
ratio.
|
|
Measured (solid lines) and calculated (dotted lines) flows for the
eight subjects in Fig. 3 are shown in Fig.
6. In subjects CL, QZ, SK, QL, EB, and
SW, the two flow signals were very close during both inspiration and
expiration, whereas in the remaining subjects, there were significant
differences between inspiration and expiration. These differences were
greatest during expiration, and, in all subjects, the two signals were
very similar during inspiration.

View larger version (39K):
[in this window]
[in a new window]
|
Fig. 6.
Comparison of measured (solid lines) and estimated (dotted lines)
flow signals in all subjects (identified by initials). Note that, in
general, inspiratory flow is better estimated than expiratory flow.
|
|
Validation of transfer functions.
Because we were more interested in volume than in flow, we used the
error in volume estimation as our criterion of accuracy for the
technique. The relationship between Ve and Vm is shown in Fig.
7A. The solid line represents
the line of identity. All data points fall near the identity line,
which indicates a good correspondence between Ve and Vm. The
breath-by-breath error (Ve
Vm) in all subjects as a function of
the average of Ve and Vm is illustrated in Fig. 7B. The mean
difference between Ve and Vm (bias) was 0.009 ± 0.046 liters and
was independent of VT. Only two subjects, JS and SW, had
values that spanned the entire confidence interval; the other
subjects' values were more tightly grouped, although some of them had
a positive (QL) or negative (SW) bias.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 7.
Comparison of estimated and measured tidal volumes (VT)
during quiet breathing in 8 subjects. Each symbol represents data from
a single subject. A: solid line represents the identity
relationship; note that all data points fall close to the identity
line, indicating a good correspondence between measured and estimated
VT measurements. B: plot of the difference in
volume measurement vs. means of the 2 measurements. Mean (±SD)
difference was 0.009 ± 0.046 liter.
|
|
Effect of head movements on volume estimation.
In Fig. 8, the flow signal derived from
tracheal breath sound amplitude (dotted line) is compared with measured
flow (solid line) during head movements in a single subject. Figure
8A shows data obtained when the head was moved from side to
side. Figure 8B illustrates up-and-down movements of the
head. In this subject, the correspondence between the measured and
estimated signal was good no matter what the position of the head was.
The breaths during the transition from one position to another tended
to be larger than the preceding and following breaths and could often be readily identified; nevertheless, they were reasonably well described by the estimated flow signal. Bland-Altman analysis revealed
a slight positive bias (+0.015 ± 0.071 liters) with neck flexion
and extension and a slight negative bias (
0.019 ± 0.061 liters) with rotation. No bias was introduced by VT size.
Standard deviation (SD) was somewhat larger than that measured (0.046 liters) with the subject facing forward. The percent error of the
VT estimates are shown in Table 1. They show that, with one
exception, the error in VT estimation was
16% with neck
rotation. The error was not increased by full neck flexion. With full
neck extension, the error was considerably greater.

View larger version (40K):
[in this window]
[in a new window]
|
Fig. 8.
Estimated (dotted lines) and measured (solid lines) flows
in different head positions in a single subject. A: head
movement from left to right. B: up-and-down head movement.
Horizontal lines indicate breaths in a given position. Note the good
correspondence between estimated and measured flows in this subject
even during the transition between different positions.
|
|
Effect of body posture on volume estimation.
Figure 9 shows the effect of posture on
the estimation of volume. Volumes obtained from the sound signal were
similar to those obtained from the pneumotach signal when subjects were
standing, and there was no change in the SD of the difference between
the two values (0.06 liters sitting; 0.07 liters standing). The bias increased from
0.005 (sitting) to 0.009 liter (standing), but this
change was small. The biggest disparity between the estimated and the
measured VT was seen in the supine position. In the
identity plot (Fig. 9C, left) the majority of the
data points fell above the line of identity, indicating an
overestimation of VT by the sound signal. The bias in the
Bland-Altman plot (Fig. 9C, right) was
0.123
liter compared with 0.005 liter when the subjects were seated. In
addition, the SD of the difference between the two measurements
increased to 0.113 liter. In Table 2, the
percent error of VT estimates in the three positions is
given; these were large in the supine position.

View larger version (33K):
[in this window]
[in a new window]
|
Fig. 9.
Estimated vs. measured VT in different body positions.
The calibration equation was obtained from the flow-sound relationship
in the seated position. A: values obtained in seated
position. B: values obtained in standing position.
C: values obtained in supine position. Note that the
difference between estimated and measured volumes was not increased by
standing but that there was a systematic overestimation of
VT in the supine position. Mean differences in
VT were 0.005 ± 0.061 liter (seated), 0.009 ± 0.069 liter (standing), and 0.123 ± 0.113 liter (supine).
|
|
Reproducibility.
Because measurements in the forward-facing, seated position were
repeated after neck movement and postural changes in the six subjects,
reproducibility of the estimation of VT could be estimated.
The two volume signals remained comparable despite the intervening
movement on the part of the patient. The mean difference between the
two measurement techniques was 0.001 liter, and the SD of the
difference was 0.06 liter.
Ventilatory pattern.
Table 3 gives values of
VT, f,
I, TI/Ttot, and
VT/TI in the normal subjects and the patients.
It is noteworthy that normal f averaged 19.3 breaths/min,
VT averaged only 0.37 liter, and VT/TI averaged only 0.31 l/s.
I was >2 SD greater than normal in the
patients with unstable airways obstruction and in one
of the asthmatic subjects. An increased
VT/TI accounted for the
increased
I in these patients. One
asthmatic patient had an unusually low f and a high
VT.
In Fig. 10, the PDF of
ln
I (Fig. 10A) and VT (Fig.
10B) are shown. Dashed lines are normal subjects, and solid
lines are patients. Patients tended to have larger
I
and VT. In Table 4, mean ± SD of
I, VT, f,
VT/TI, and TI/Ttot and the mean of
their log transformed values are given along with the values of
skewness and kurtosis for both the normal and log normal PDF.
For each ventilatory parameter, the top row gives the normal
values and the lower row the values in the patients. As judged by
skewness and kurtosis, log normal PDF provided a better fit to the data with the sole exception of TI/Ttot, in normal subjects.

View larger version (25K):
[in this window]
[in a new window]
|
Fig. 10.
Log-normal (ln) probability density functions (PDF) of
minute ventilaton ( I; A) and
VT (B) in normal subjects (dashed lines) and
patients (solid lines).
|
|
 |
DISCUSSION |
Derivation of flow from sound.
Continuous noninvasive measurement of flow and volume has both clinical
and research applications. At present, the most widely used noninvasive
technique for respiratory measurements is respiratory inductive
plethysmography. However, with this technique, measured VT
is affected by changes in posture and/or the shape of the chest wall
(24, 35, 38, 39), limiting its usefulness in long-term monitoring. For this purpose, other noninvasive methods need to be
developed. In this study, we have explored the feasibility of deriving
ventilatory flow from tracheal sound signals and using this signal to
estimate VT.
The relationship between flow at the mouth and the amplitude of breath
sounds has been under investigation for a number of years. Breath
sounds obtained from microphones positioned at different places on the
chest wall (3, 11, 17, 29, 32) and over the trachea
(6, 9, 10, 33) have been examined. Some of these
investigators have quantified the relationship between breath sound
amplitude and tracheal flow by using mathematical functions. Banaszak
et al. (3) described a linear relationship between flow
and breath sound amplitude measured over the chest wall. A quadratic
function was used by Shykoff et al. (32) and a power
function by Gavriely and colleagues (10, 11). These functions have all provided a good fit to the experimental data; however, only in one paper (33) was there an attempt to
use the derived mathematical relationship to obtain a flow signal from
the sound signal. To our knowledge, Soufflet et al. (33) is the only paper that attempts to quantify tracheal breath sounds at
flows of <0.5 l/s, a necessity if one wishes to record flows during
quiet breathing. We confirmed their finding that there is a threshold
flow that needs to be exceeded in order for sound to be present, and
that even at flows greater than threshold the adverse SNR (Fig. 5) can
present a major problem. Although the investigators recognized the
threshold problem and the adverse SNR (evident from their data), they
do not state how these major problems were solved.
Calculating respiratory flow from the breath sound signal is more
complex than simply determining the relationship between the two
signals. Because the breath sound amplitude at low flow rates does not
exceed the background noise (Fig. 5), there exists a range of flows
where no relationship between breath sound amplitude and flow can be
obtained. Therefore, we derived the relationship between sound
amplitude and flow only above the threshold value of sound that
exceeded the background noise. The attempts to quantify the
relationship between sound intensity and flow taken by others (3,
10, 11, 32) have not included this threshold and forces the
relationship through the origin. This is clearly an oversimplification
if one wishes to record the low flow rates used for quiet breathing.
Although above the threshold value the relationship between flow and
sound might be better fit by a curve rather than a straight line,
recent work (L.-G. Durand, unpublished observations) indicates that the
scatter of the data relating flow to sound intensity (Fig. 3) does not
allow curve fitting to provide a better estimate of VT than
the linear approximation we used, which provided a reasonably accurate
estimation of VT (Fig. 7). Thus we used a simple
least-square linear regression to calculate the transfer function, and
this was sufficient for a first attempt at deriving flow from breath
sound amplitude.
During the period when flow was below the threshold value and the
transfer function could not be used, we obtained flow values by linear
interpolation from the lowest computed flow value to the end of either
inspiration or expiration and then from these zero-flow points to the
next recorded value of flow. Because the transitions between
inspiration and expiration occupy only a small fraction of the total
time of breathing and the flows are low during these periods,
interpolation is quite accurate when the timing is distinguished
correctly. It can be seen from Figs. 4 and 6 that Ve and Vm timing
matched quite well despite the high threshold values used for the
analysis. This was particularly true during the transition between
inspiration and expiration due to the fast decelerations and
accelerations during this part of the breathing cycle. Correspondence
was not so good during the latter part of expiration and the
transition to inspiration because some subjects had an end-expiratory
pause. Although further refinements of the algorithm might enable us to
better fit this part of the breathing cycle, accurate expiratory flows
are not needed to obtain most ventilatory parameters of interest. Thus we calculated VT, f,
I, TI,
TE, TI/Ttot, and mean
VT/TI rate simply by measuring inspiratory flow
alone as a function of time.
Sources of error.
When we compared VT derived by integration of measured flow
with the volume obtained by integration of estimated flow, in subjects
in the seated position, we found a good correlation between the two
values (Fig. 7A). The mean difference between Ve and Vm (0.001 liter) was insignificant, but individual breaths could vary by
up to 0.1 liter (Fig. 7B). However, the error spanned the
entire confidence interval in only two subjects (EB and JS). In subject
SW, Ve was consistently higher than Vm, and, in JS, Ve was lower than
Vm. A bias such as this can be characterized and accounted for when
making measurements in an individual subject. As a result, the percent
error in estimating VT was <15% in all subjects when they
were seated and facing forward (Table 1).
One explanation for errors in estimating Ve is breath-to-breath
variation in the relationship between sound amplitude and flow (Fig.
3). Because the derived regression represents the best fit over a
number of breaths, its application to individual breaths will result in
differences compared with the volumes derived from integration of flow.
Another source of variation could be filtering. Because most of the
power in the tracheal sound signal is below 800 Hz (11),
we chose 1,000 Hz as the low-pass cutoff for eliminating high-frequency
noises and a sampling frequency of 3,000 Hz to exceed the Nyquist
frequency. Tracheal sound frequencies can exceed 1,000 Hz
(16), and these were not included in the sound envelope we
measured. Nevertheless, our measurements of VT were
reasonably accurate. In some subjects, estimation of zero flow points
could pose a problem, and distinguishing inspiration from expiration might be difficult. Future refinements of the methodology should take
these potential sources of error into account.
Effect of head movement.
The estimated flow signal was not greatly affected by changing the head
position apart from the breaths during the transition period (Fig. 8).
Because these breaths tend to have higher than average
VT/TI, they are easily distinguished. For our
analyses, we eliminated one breath at the transition; nevertheless, it
is interesting to note that both the actual and the estimated flow signals were similarly altered, suggesting that inclusion of these breaths in the data analysis will not increase the error of our estimation of VT.
Side-to-side changes in head position did not alter the accuracy of the
volume estimation (Fig. 8A), but there was an increase in
Ve
Vm with neck extension (Fig. 8B; Table 1).
Estimated volumes tended to exceed measured volumes likely because
extending the neck tightened the skin and decreased the distance
between the microphone and the trachea. Because in this experiment the neck was maximally extended, it is likely that, in mobile subjects, such errors will be less than we estimate here, as the neck is not
usually fully extended in normal daily activities.
It is also possible that neck extension caused a cephalad displacement
of the microphone, but we would expect, in that case, that random
rather than systematic changes in volume would have resulted.
Effect of posture.
Moving from the seated to the standing position had little effect on
VT estimation (Fig. 9; Table 2). This could be considered an advantage over inductance plethysmography, which changes calibration with posture and changes in xiphipubic distance. We have not
systematically measured the effect of thoracoabdominal configuration on
the relationship between sound and flow in the trachea but see no a
priori reason why it should change. Almost certainly, there was a
significant change in xiphipubic distance between sitting and standing
postures in at least some of our subjects. They were not instructed on how to sit during the recording period and merely chose a position that
was comfortable. We subsequently measured changes in xiphipubic distance in normal subjects between seated and standing posture while
maintaining a rigid upright spinal configuration and found changes
between 5 and 8 cm.
The unacceptable error in changing posture from seated to supine
indicates that the transfer function obtained in the seated position
cannot be extrapolated to all positions. If this technology is to be
used in the supine position, calibration of the sound signal needs to
be done in that posture. This will be important because we have not yet
shown that phonospirometry can be used supine, yet many of its clinical
applications will require supine measurements. Other positions have yet
to be investigated. Movement to the standing position did not affect
the measurements; therefore, for longer measurements, a single
calibration will probably suffice for standing and seated positions
even with moderate head rotation flexion and extension.
Ventilatory pattern.
In the preliminary results we obtained in normal subjects, we found a
ventilatory pattern that differed in important ways from those
summarized by Shephard (31). In contrast to the results shown in Table 3, he gives normal values taken from several studies of
f that ranged from 10.1 to 15.8 breaths/min with a mean value of 13.2 breaths/min. Our value of 19.3 breaths/min is 46% higher than this.
McCool and colleagues (20, 21, 24, 25) measured the normal
ventilatory pattern noninvasively from body surface measurements and
found values for f that were only 10% higher than values obtained by
spirometry. Our mean value for
I of 6.9 l/min is
identical to the mean value of the studies summarized by Shephard,
although the range of his data (4.4-11.4 l/min) was considerably
greater than the range we found (31). However, the
VT we measured of 373 ml is considerably less than the mean value of 508 ml (range of 213-751 ml) reported by Shephard
(31) and close to the value reported by Ashkanazi et al.
(2) of 362 ml. They used a canopy system to make their
measurements. Our value for VT/TI of 0.31 l/s
is similar to the value of 0.298 l/s calculated from their data
(2). They found that the primary effect of a mouthpiece
and nose clip was to increase VT without change in
TI so that VT/TI increased to 400 ml/s.
It has been shown by several studies that a mouthpiece and nose clip
alters breathing pattern. Although in some studies a decrease in f has
been observed when breathing through a mouthpiece and nose clip
(12, 13, 20, 27), this has not been found in other studies
(2, 37). However, in reports where no significant effect
on f was observed, the f on mouthpiece and nose clip was high, e.g.,
19.1 breaths/min in Ashkanazi's study (2). It seems clear
that the normal f is considerably higher, whereas mean
VT/TI, VT, and TI are
less than that frequently measured with a mouthpiece and nose clip.
Although the numbers are too small to draw any firm conclusions, Table
3 indicates that
I was abnormally increased in the three patients with unstable airway obstruction, and this was primarily
due to an increase in mean VT/TI. Two stable
asthmatic patients also had an increased VT/TI,
one of whom had a high
I. The third asthmatic
patient had an abnormally low f. The fact that all six patients had one
or more abnormalities in ventilatory pattern suggests that
phonospirometry will prove useful in detecting and quantifying abnormal
ventilatory patterns in disease.
Ventilatory variability.
We found considerable breath-to-breath variation in all ventilatory
parameters. Nonrandom variations in ventilatory parameters are well
described (see Ref. 4 for a review). Davis and Stagg (8a) found variations in VT and TI
with constant VT/TI, whereas Ashkanazi et al.
(2) found constant TI and TI/Ttot
with variations in VT and VT/TI. We
found variations in all these parameters. With the exception of
TI/Ttot, these were better described by log normal than by
normal PDF.
Noninvasive measurements of ventilation.
In this study, we have shown that, with proper calibration,
VT/TI rate can be derived from a tracheal sound
signal and that VT can be estimated noninvasively to an
accuracy within 15% of the measured volume signal. Because
phonospirometry has the advantage of being easy to calibrate and is
insensitive to postural change from sitting to standing, it has
advantages over other noninvasive methods of measuring ventilation,
such as inductance plethysmography and magnetometry (21, 30,
34). Its VT accuracy is equivalent to these
techniques (20). It probably is not as accurate as optoelectronic plethysmography but is cheaper, easier to use, and less
intrusive and does not require the subject's position and posture to
be restricted so that the thorax is continuously visualized by videocameras.
 |
ACKNOWLEDGEMENTS |
This study was supported by the Medical Research Council of Canada
and the Respiratory Health Network of Centres of Excellence.
 |
FOOTNOTES |
Address for reprint requests and other correspondence:
P. T. Macklem, PO Box 25, Lansdowne, ON, Canada K0E 1L0
(E-mail: ptm01{at}hotmail.com).
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
June 14, 2002;10.1152/japplphysiol.00028.2002
Received 14 January 2002; accepted in final form 6 June 2002.
 |
REFERENCES |
1.
Ajmani, A,
Mazumdat J,
and
Jarvis D.
Spectral analysis of an acoustic respiratory signal with a view to developing an apnoea monitor.
Australas Phys Eng Sci Med
19:
46-52,
1996.
2.
Ashkanazi, J,
Silverberg PA,
Foster RJ,
Hyman AI,
Milic-Emili J,
and
Kinney JM.
Effects of respiratory apparatus on breathing pattern.
J Appl Physiol
48:
577-580,
1980.
3.
Banaszak, E,
Kory RC,
and
Snider GL.
Phonopneumography.
Am Rev Respir Dis
107:
449-455,
1973.
4.
Bruce, E.
Temporal variations in the pattern of breathing.
J Appl Physiol
80:
1079-1087,
1996.
5.
Cala, SJ,
Kenyon CM,
Ferrigno G,
Carnevali P,
Aliverti A,
Pedotti A,
Macklem PT,
and
Rochester DF.
Chest wall and lung volume estimation by optical reflectance motion analysis.
J Appl Physiol
81:
2680-2689,
1996.
6.
Charbonneau, G,
Sudraud M,
and
Soufflet G.
Une methode d'evaluation du debit à partir des soins pulmonaire.
Bull Eur Physiopath Respir
23:
265-270,
1987.
7.
Chihara, K,
Kenyon CM,
and
Macklem PT.
Human rib cage distortability.
J Appl Physiol
81:
437-447,
1996.
8.
Cummiskey, J,
Williams TC,
Krumpe PE,
and
Guilleminault C.
The detection and quantification of sleep apnea by tracheal sound recordings.
Am Rev Respir Dis
126:
221-224,
1982.
8a.
Davis, JN,
and
Stagg D.
Interrelationships of the volume and time components of individual breaths in resting man.
J Physiol
245:
481-498,
1975.
9.
Gavriely, N.
Measurement of tracheal lung sounds.
J Appl Physiol Respir Environ Exercise Physiol
56:
817-818,
1984.
10.
Gavriely, N,
and
Cugell DW.
Airflow effects on amplitude and spectral content of normal breath sounds.
J Appl Physiol
80:
5-13,
1996.
11.
Gavriely, N,
Paltri Y,
and
Alroy G.
Spectral characteristics of normal breath sounds.
J Appl Physiol
50:
307-314,
1981.
12.
Gilbert, R,
Auchincloss JH, Jr,
Brodsky J,
and
Boden W.
Changes in tidal volume frequency and ventilation induced by their measurement.
J Appl Physiol
33:
252-254,
1972.
13.
Hirsch, JA,
and
Bishop B.
Human breathing patterns on mouthpiece or facemask during air, CO2, or low O2.
J Appl Physiol
53:
1281-1290,
1982.
14.
Kenyon, CM,
Cala SJ,
Yan S,
Aliverti A,
Scano G,
Duranti R,
Pedotti A,
and
Macklem PT.
Rib cage mechanics during quiet breathing and exercise in humans.
J Appl Physiol
83:
1242-1255,
1997.
15.
Kraman, SS.
The relationship between airflow and lung sound amplitude in normal subjects.
Chest
86:
225-229,
1984.
16.
Kraman, SS,
Pasterkamp H,
Kompis M,
Takase M,
and
Wodicka GR.
Effects of breathing pathways on tracheal sound spectral features.
Respir Physiol
111:
295-300,
1998.
17.
Leblanc, P,
Macklem PT,
and
Ross WR.
Breath sounds and distribution of pulmonary ventilation.
Am Rev Respir Dis
102:
10-16,
1970.
18.
Lessard, CS,
and
Wong WC.
Correlation of constant flow rate with frequency spectrum of respiratory sounds when measured at the trachea.
IEEE Trans Biomed Eng
33:
461-463,
1986.
19.
Makarenkov, AP,
and
Rudnitskij AG.
Diagnosis of lung pathologies by two-channel processing of breath sounds.
Akusticheskii Zurnal
41:
272-277,
1994.
20.
McCool, FD,
Kelly KD,
Loring SH,
Greaves IA,
and
Mead J.
Estimates of ventilation from body surface measurements in unrestrained subjects.
J Appl Physiol
61:
1114-1119,
1986.
21.
McCool, FD,
and
Paek D.
Measurements of ventilation in freely ranging subjects.
Res Rep Health Eff Inst
59:
1-17,
1993.
23.
Oppenheim, AV,
and
Schafer RW.
Chapter 7.
In: Digital Signal Processing. Englewood Cliffs, NJ: Prentice-Hall, 1975, p. 337-375.
24.
Paek, D,
Kelly KB,
and
McCool FD.
Postural effects on measurements of tidal volume from body surface displacements.
J Appl Physiol
68:
2482-2487,
1990.
25.
Paek, D,
and
McCool FD.
Breathing patterns during varied activities.
J Appl Physiol
73:
887-893,
1992.
26.
Pasterkamp, H,
Kraman SS,
and
Wodicka GR.
Respiratory sounds. Advances beyond the stethoscope.
Am J Respir Crit Care Med
156:
974-987,
1997.
27.
Perez, W,
and
Tobin MJ.
Separation of factors responsible for change in breathing pattern induced by instrumentation.
J Appl Physiol
59:
1515-1520,
1985.
28.
Ploysongsang, Y,
Ilyer VK,
and
Ramamoorthy PA.
Characteristics of normal lung sounds after adaptive filtering.
Am Rev Respir Dis
139:
951-956,
1989.
29.
Ploysongsang, Y,
Pare JA,
and
Macklem PT.
Correlation of regional breath sound with regional ventilation in emphysema.
Am Rev Respir Dis
126:
526-529,
1982.
30.
Sackner, MA,
Watson H,
Feinerman D,
Suarez M,
Gonzalez G,
Bizousky F,
and
Krieger B.
Calibration of respiratory inductive plethysmograph during natural breathing.
J Appl Physiol
66:
410-420,
1989.
31.
Shephard, RJ.
Respiration and circulation.
In: Biological Handbook, edited by Altman PL,
and Dittmer DS.. Bethesda, MD: FASEB, 1971, p. 42-43.
32.
Shykoff, BE,
Ploysongsang Y,
and
Chang HK.
Airflow and normal lung sounds.
Am Rev Respir Dis
137:
872-876,
1988.
33.
Soufflet, G,
Charbonneau G,
Polit M,
Attal P,
Denjean A,
Escourrou P,
and
Gaultier C.
Interaction between some tracheal sounds and flow rate: a comparison of some different evaluations from lung sounds.
IEEE Trans Biomed Eng
37:
384-390,
1990.
34.
Stradling, JR,
Chadwick GA,
Quirk C,
and
Phillips T.
Respiratory inductance plethysmography: calibration techniques, their validation and the effects of posture.
Bull Eur Physiopath Respir
21:
317-324,
1985.
35.
Stromberg, NO,
Dahlback GO,
and
Gustafsson PM.
Evaluation of various models for respiratory inductance plethysmography calibration.
J Appl Physiol
74:
1206-1211,
1993.
36.
Ward, ME,
Ward JW,
and
Macklem PT.
Analysis of human chest wall motion using a two-compartment rib cage model.
J Appl Physiol
72:
1338-1347,
1992.
37.
Weissman, C,
Ashkanazi J,
Milic-Emili J,
and
Kinney JM.
Effect of respiratory apparatus on respiration.
J Appl Physiol
57:
475-480,
1984.
38.
Whyte, KF,
Gugger M,
Gould GA,
Molloy J,
Wraith PK,
and
Douglas NJ.
Accuracy of respiratory inductive plethysmograph in measuring tidal volume during sleep.
J Appl Physiol
71:
1866-1871,
1991.
39.
Zimmerman, PV,
Connellan SJ,
Middleton HC,
Tabona MV,
Goldman MD,
and
Pride N.
Postural changes in rib cage and abdominal volume-motion coefficients and their effect on the calibration of a respiratory inductance plethysmograph.
Am Rev Respir Dis
127:
209-214,
1983.
J APPL PHYSIOL 93(4):1515-1526
8750-7587/02 $5.00
Copyright © 2002 the American Physiological Society