|
|
||||||||
3,1 Harvard Medical School, Department of Anesthesiology, Perioperative and Pain Medicine, Brigham & Women's Hospital, 2 Physiology Program, Harvard School of Public Health, and 3 Department of Biomedical Engineering, Boston University, Boston, Massachusetts 02115; and 4 Harvard College and 5 Harvard Extension School, Cambridge, Massachusetts 02138
| |
ABSTRACT |
|---|
|
|
|---|
At fixed lung volume (VL), alterations in surface tension change alveolar surface area (S) and lung recoil (PL). Wilson (26), using data from fixed lungs (1, 9), quantified the isovolume change in S with PL. We reexamined this question in fresh excised rabbit lungs, with two important differences. First, we measured fractional changes in S by using diffuse light scattering, avoiding the potential upset of the balance of tissue and surface forces during fixation. Second, we altered surface tension by ventilating the lungs with nebulized polydimethylsiloxane, with much less residual fluid compared with lavage. We found that S decreased at low and mid VL (treatment surface tension > control) by about half of Wilson's estimates and was nearly unaffected by treatment at high VL. This suggests that with increased surface tension there is 1) greater septal retraction in lungs fixed by vascular perfusion compared with unfixed lungs and 2) a greater increase in PL and less loss of S than would have been predicted.
lung; mechanics; lung recoil
| |
INTRODUCTION |
|---|
|
|
|---|
IT IS GENERALLY AGREED
THAT over a wide range of lung volumes (VL), the
architecture of the mammalian lung displays approximate geometric
similarity. However, the design of the lung includes an internal degree
of geometric freedom in that the division of acinar volume between
alveoli and alveolar ducts is determined by a balance of forces:
inward-acting tissue forces in the alveolar entrance ring and
outward-acting forces arising from septal tissue forces in parallel
with surface forces at the air-liquid interface (3, 12,
27). This implies that alveolar surface area (S) can change
independently of VL as a consequence of changes in the
surface tension (
) acting at the air-liquid interface of the
alveolar septal surface. There are wide variations in
, both in
normal lungs (especially true for large-volume maneuvers but also true
for tidal breathing) and in diseased lungs (such as in acute
respiratory distress syndrome), and therefore it is important to assess
the magnitude of variations in surface area associated with these
changes and their physiological consequences to lung recoil
(PL) and gas exchange.
The interplay between surface forces and tissue forces on the one hand
and radially inward- and outward-acting forces on the other hand is
shown diagrammatically in Fig. 1. The
effect of increased
is seen to have two effects. First, overall
PL is increased, and second, S of the alveolar septa,
especially those with free edges, is decreased. It follows that the
tissue forces within those septa (excluding the free edge) are
correspondingly reduced and the tissue forces in the free borders of
the septa are increased; the net effect is that the increase in
PL is less than that which would have occurred had there
been no change in S. Furthermore,
itself is not an independent
variable insofar as it is strongly dependent on both the surface area
of the surfactant film as well as the inflation-deflation history. Thus
the simplest measure of lung mechanics, i.e., the pressure-volume (P-V)
curve, depends on the simultaneous solution to the geometric response to changes in
, the response of
to changes in S, and the
interaction with tissue forces both within the septa and at their free
edges.
|
In this study, we determined the fractional change in S (
S/S), at
the same VL, between control conditions and after changing
. We then determined the relationship between
S/S and the change in PL (
P) caused by changes in
. This was done over a
range of VL from functional residual capacity (FRC) to
total lung capacity (TLC, the volume at 30 cmH2O distending
pressure). All measurements were made in fresh excised rabbit lungs, by
use of diffuse light scattering technology that, because the lungs are
not fixed, permits repeated examination of an individual lung both at
different volumes, and before and after altering
.
can be
altered by lavage with a variety of fluids, followed by reinflation
with air [e.g., Smith and Stamenovic (20), Bachofen et
al. (1), and Hoppin et al. (13)], but this
is known to leave a significant amount of fluid in the lung
(4). The septa of partially filled alveoli are thus not
equivalent to two-dimensional membranes with (altered)
at the
air-liquid interface. Thus, in the lavage preparation, the alveolar
architecture is changed owing both to the direct effects of changes in
and to partial alveolar flooding. We therefore chose to alter
by ventilating the lung with nebulized polydimethylsiloxane (hereafter abbreviated as siloxane), the volumetric dose of which was
several orders of magnitude less than techniques employing lavage. We
found that the shifts in the pressure-volume characteristics with
exposure to nebulized siloxane were remarkably similar to those found
in air-filled lungs following lavage with the same fluid
(20), a technique generally thought to effect an
approximately constant
preparation (13, 28).
Using these techniques, we found that
S/
P systematically
decreased with increasing VL. Near FRC it was about half
that concluded by Wilson (26), using data of Bachofen et
al. (1) and Gil et al. (9) obtained from
fixed lungs. At high VL,
S/
P was very close to zero.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Two groups of New Zealand White rabbits were used for these
experiments: one group (n = 9) for characterizing the
effects of nebulized siloxane on pulmonary mechanics (pressure/volume curves, residual siloxane, and regional variability); the other group
[n = 7, surface-to-volume ratio (S/V) study] for
measuring the effect of alterations in
, as quantified by changes in
PL, on pulmonary surface area.
Effects of nebulized siloxane.
The deflation P-V characteristics were studied after three different
types of exposure to constant-
siloxane. In exposure type
1 (repeat nebulized siloxane, n = 3), lungs were
excised and control P-V data were collected. The lungs were then
ventilated with nebulized siloxane, and posttreatment P-V data were
collected. Ventilation with nebulized siloxane and P-V data
collection were then repeated two more times. In exposure type
2 (nebulized siloxane vs. lavage, n = 3), excised
lungs were studied under control conditions, after a single nebulized
siloxane treatment, and after siloxane lavage and reinflation with air.
In exposure type 3 (nebulized siloxane in vivo,
n = 3), anesthetized animals were ventilated with
nebulized siloxane, the lungs were excised, and postnebulization P-V
data were collected.
Animal preparation. Each of the 16 animals (mean body wt 3.7 kg) was anesthetized with ketamine (30-50 mg/kg im) and xylazine (1-3 mg/kg im). Additional pentobarbital sodium (10-20 mg/dose iv) was given via an ear vein as needed to maintain adequate anesthesia. Thirteen rabbits (treatment types 1 and 2 and animals used for the S/V study) received heparin (1,000 units/kg) and were exsanguinated via the carotid arteries. They were then intubated, the trachea was clamped at positive airway opening pressure (Pao), and the lungs were excised. Three rabbits (treatment type 3) were intubated below the larynx through a midline incision and ventilated with nebulized siloxane. The animal then received heparin (1,000 units/kg iv) and was exsanguinated via its carotid arteries. The lungs were then excised.
Experimental setup and ventilatory histories.
Each pair of rabbit lungs was suspended by the trachea, and the
endotracheal tube was connected to an adjustable constant-pressure source for control of airway opening pressure (Pao). The lungs were
kept moist with 0.9% saline. Between measurements, the lungs were
ventilated at 20 breaths/min (Harvard respiration pump, Harvard Apparatus, Millis, MA), and during measurements they were held at a
constant airway pressure. The ventilator was operated as a "pressure
ventilator" with the addition of pop-off valves to its inspiratory
and expiratory limbs. The pop-off valve on the ventilator's
inspiratory limb was set to open at Pao
25 cmH2O, whereas its passive expiratory limb's valve was set to open at positive end-expiratory pressure of 2-3 cmH2O
(pretreatment) or 7-10 cmH2O (posttreatment). This
increase in positive end-expiratory pressure posttreatment was done to
prevent atelectasis and maintain FRC near control levels, because the
siloxane treatment raises the lung's
and recoil pressure,
especially at low VL. Before each measurement, the lung was
cycled three times between Pao of 3 and 30 cmH2O and then
set to the target Pao on the lung's deflation limb. During
measurements, Pao was held constant by connecting the lung to an air
source with a "T" side branch submerged to a measured depth in a
water column. Because there is no gas flow during measurements, Pao is
a direct measure of PL. At each target Pao ranging from 2.5 (pretreatment) and 5 (posttreatment) to 30 cmH2O, both
VL and S were measured, as described below.
VL measurement.
At PL ~10 cmH2O, three pleural markers (short
pieces of black suture) were cemented to the middle portion of the
costal pleural surface of one diaphragmatic lobe (S/V experiments) or
both left and right lobes (siloxane-treatment experiments) by use of a
small drop of cyanoacrylate glue, ~1.5 cm apart in a triangular
pattern. The edge lengths were measured either with calipers (for the
S/V experiments) or from photographs of the pleural markers together with a ruler held parallel to the lung surface (siloxane-treatment experiments). The area of the triangle was computed from the edge lengths. VL was calculated as the 3/2 power of the area
inscribed by the suture markers (14), normalized to its
value at TLC pretreatment. (TLC is defined as VL at
Pao = 30 cmH2O). Note that VL as thus defined is an estimate of the sum of gas and tissue volume. Pressure and volume data were fitted by least squares to the exponential expression of Salazar and Knowles (18)
|
S measurement.
Fractional changes in S, or, equivalently, surface-to-volume ratio,
were determined by using diffuse scattering of laser light (693 nm).
These changes were measured, during control conditions, with TLC used
as the reference state. After nebulized siloxane treatment,
measurements were repeated, using pretreatment isovolume points as the
reference states. For a description of the theories, techniques, and
interpretations involved when using diffuse light scattering in
measuring lung architecture, see Butler et al. (6, 7),
Suzuki et al. (21), Miki et al. (16), and
Topulos et al. (22). Briefly, one optical fiber was used
to place a point source of light normal to the pleural surface of the
lung lobe, and three receiving fibers detected the backscattered light
intensity at fixed and known distances from the optical source. In the
diffuse far field of scattering, the intensity falls off with distance r from the source proportionate to
(kdiffLopt/r2)exp(
kdiffr),
where kdiff is the diffuse extinction
coefficient and Lopt is the optical mean free
path for scattering. Fitting a linear function to the logarithm of the
r2-weighted intensity suffices to determine
kdiff, the logarithm of
kdiffLopt, and hence the
fractional changes in Lopt from any reference
state. These can be converted to fractional changes in surface area by
observing that Lopt is proportional to
Lm2 (21), where
Lm is the mean linear intercept of air-liquid
interfaces in the lung and that Lm is
proportional to (S/V)
1, i.e., the inverse of
S/V.
Treatment of the lungs with nebulized siloxane.
After control measurements, the lung was ventilated for 5 min at 20 breath/min. Treatment was then begun with nebulized siloxane (5 ml
polydimethylsiloxane; viscosity 20 cS, molecular weight = 2,000,
20.6 dyn/cm; Nye Lubricants, New Bedford, MA) in the inspiratory limb
of the ventilator by using a continuous flow of ~13 l/min compressed
air into the nebulizer (Misty-Neb, Allegiance, McGraw Park, IL). The
time necessary for nebulizing the entire 5 ml of siloxane was between 5 and 10 min. Gas leaving the circuit during nebulization passed through
HEPA filters (OF612HE Omega filter, Atrix International, Burnsville,
MN) to avoid contaminating the room with siloxane aerosol. The size
distribution of the siloxane aerosol leaving the nebulizer was
characterized (Aerosizer, model MACCII, API, Amherst, MA). The count
median aerodynamic diameter of the mist leaving the nebulizer was
1.8 ± 1.3 µm (mean ± geometric SD). After treatment, the
lung was ventilated for either 5 min (siloxane-characterization
measurements) or 30 min (S/V measurements).
Siloxane lavage. After control data and data after a single nebulized siloxane treatment were obtained (as above), the lungs of rabbits in the type 2 exposure group (nebulized siloxane vs. lavage) were lavaged with the same fluid. The lungs were filled by connecting the trachea (starting from a PL of 0) to a siloxane-filled reservoir held 30 cm above the lung. Filling required 80-88 ml of siloxane and took ~15 min. To drain the siloxane from the lungs of two animals, the reservoir was then lowered to ~100 cm below the level of the lungs. The siloxane was withdrawn from the lungs of the third animal by aspiration with a syringe and passive drainage. After 15-20 min, 22-25 ml of siloxane remained in the lungs. The lungs were then ventilated with air for ~5 min, and then postlavage P-V data were collected as described above.
Siloxane dose deposited. In separate experiments, the volume of siloxane deposited per gram of lung tissue was determined in three excised rabbit lungs and one excised dog lobe ventilated with nebulized siloxane and in one rabbit lobe ventilated in vivo with nebulized siloxane. The mass of elemental Si in each of 23 samples (1.3 ± 0.5 g) was determined by atomic absorption spectroscopy (Galbraith Laboratories, Knoxville, TN). Also analyzed were untreated control lungs prepared in an identical manner, to which known volumes of siloxane were added after homogenization (used to calculate Si recovery efficiency). The volume of siloxane deposited per gram of lung tissue was computed from recovery efficiency (~0.63), the density and composition of the siloxane, and the mass of each sample.
| |
RESULTS |
|---|
|
|
|---|
Effects of siloxane exposure on the P-V curves.
The P-V characteristic of the lung changed consistently after all of
the exposures to siloxane, with no important differences between a
single nebulization, multiple nebulizations, or lavage. Figure
2 shows a comparison of P-V curves under
control conditions; one, two, or three treatments with nebulized
siloxane; and after lavage. Elastic recoil increased between
~40-90% TLC. At higher VL there was less change in
the P-V curves. Regardless of the exposure method, the magnitude of the
change in recoil at 50% TLC was always ~5 cmH2O. These
P-V findings are similar to those of Smith and Stamenovic
(20) in rabbit lungs inflated with air after lavage with
the same siloxane, as shown in Fig. 3.
|
|
1 (9)], would result in a siloxane
film layer ~0.01 µm thick at TLC and about twice that at FRC.
Equivalent results were found in the dog lobe. This thickness is a
negligible fraction of alveolar diameter and on the order of magnitude
of the 0.1-µm thickness of the normal surfactant layer
(12). There is, however, considerable uncertainty
regarding the nature and degree of spreading (see DISCUSSION).
Pooled pressure-volume data (mean ± SD) for the S/V experiments,
both pre- and posttreatment, together with their exponential fits, are
shown in Fig. 4.
|
P. At any VL below
~95% TLC, PL increased posttreatment (
P > 0),
and conversely above this volume. This volume defines a crossover point
at which the posttreatment PL is the same as control, i.e.,
the point at which
after siloxane is the same as that in the normal
lung (see DISCUSSION for comments on film spreading). This
is consistent with the idea (20) that, at this crossover
volume, the in vivo
is similar in magnitude to that of the siloxane
and should therefore induce little geometric distortion from the
control state. Our results are also consistent with other observations
of the volume at which normal
is ~20 dyn/cm, which range from mid
80% TLC (19, 20) to mid 90% TLC (26, 13).
After treatment with siloxane, it was not possible to determine
VL by pleural markers below Pao = 5 cmH2O
because at such low distending pressures areas of the lung began to
collapse, thus distorting its pleural surface. The gross appearance of
such collapsed areas posttreatment was strikingly similar to that of a
degassed lung.
Alveolar S, VL, and
P.
Pretreatment S varied strongly as a function of VL. Figure
5 shows a log-log plot of S vs.
VL (mean ± SE). Control data points are shown,
together with the regression line (constrained to go through the origin
because both S and VL are normalized to their values at
TLC). The slope is 0.65, which is not significantly different from

|
P (control S points
are shown on the
P = 0 axis).
|
P. Also shown in both Figs. 6
and 7 are estimates of S and
S vs.
P from the work of Wilson
(26). These lines were derived from Fig. 1 of that work.
The intercepts at
P = 0 were shifted slightly to coincide with
our measurements; the slopes were taken from the chord slopes of the
air-filled and detergent-rinsed preparations below TLC and from the
tangent slope of the line at TLC.
|
; this in turn is necessarily associated with a
decrease in surface area. Above 80% TLC, differences were not significant.
| |
DISCUSSION |
|---|
|
|
|---|
Our measurements of the change in parenchymal surface area, at
fixed volumes when
is artificially changed, are similar to but
quantitatively smaller than that seen by others [Bachofen et al.
(1) and Gil et al. (9) as synthesized by
Wilson (26)]. There are two differences in the
experimental preparations that may account for this. First, our
measurements were performed in fresh, unfixed lungs by use of diffuse
light scattering, whereas the classical measurements of surface area
have been done in fixed and sectioned lungs by using stereological
principles and morphometric measurements made under light microscopy.
Second, our intervention to change the lung's
involved ventilation
with nebulized siloxane, which resulted in a deposited volume
significantly smaller than is seen with previous lavage techniques. We
remark on each of these. After these methodological issues, we give
some interpretations of our findings in terms of their physiological
and pathophysiological implications.
How are forces "fixed" when chemically preserving a lung?
The inwardly acting forces in the tissue comprising the alveolar
entrance rings are essentially at all times balanced by the combination
of outwardly acting tissue forces in the alveolar septa together with
the
present at the air-liquid interface. For the resulting geometry
(fractionation of volume between ducts and alveoli, or, equivalently,
the determination of the alveolar septal area at that given
VL) to remain unchanged during fixation requires that the
forces that are mechanically in series change quantitatively in an
identical fashion over time. If this is not the case, then there will
be a distortion of the geometry during fixation before the final state
is reached when sections can be cut. The change in the force/length
characteristics of connective tissue (primarily elastin and
collagen) as well as smooth muscle during aldehyde fixation is not well
known. Similarly, there are no studies to our knowledge of the time
course of changes in
in the lung (be they associated with native
surfactant or with any of the perturbing fluids or detergents used to
artificially change
) during vascular perfusion with osmium
tetroxide or glutaraldehyde. Moreover, leakage of any fluid into the
alveolar space may transiently increase
before fixation and result
in a larger decrease in surface area. The question of whether there is
geometric distortion during fixation, particularly as manifested in
changes of surface area relative to that present in vivo, thus remains
open. It is important to note that we are not questioning the value of
classical fixation techniques in preserving the ultrastructure at the
cellular and subcellular level, nor do we question the appropriateness of using stereological techniques to infer three-dimensional properties such as S/V from morphometric measurements made on sections. Rather, it
would seem important in the future to characterize the temporal evolution of stiffening of proteins and surface films in the presence of aldehyde perfusates. This would resolve the possibility that differences in our data, obtained on fresh, unfixed lungs, from classical studies are due to architectural alterations in the lung
parenchyma associated with perfusion fixation of prestressed tissues.
Diffuse light scattering.
In addition to the uncertainties associated with chemical fixation,
there are also a number of unanswered questions regarding the use of
diffuse light scattering to obtain stereological information. Chief
among these is the relationship between fractional changes in the
optical mean free path (which we feel confident we can measure) and the
corresponding fractional changes in the geometric mean free path (which
is the desired quantity, inversely proportional to S/V). The
relationship between the optical and geometric mean free path was
established by Suzuki et al. (21), by comparison with
microscopic morphometry on preparations fixed by vascular glutaraldehyde and osmium tetroxide perfusion, assuming that the fixed
preparation is reasonably faithful to the in vivo architecture (17). It is important to note that the measurements we
report here are expressed as fractional changes in S/V, not absolute values. We therefore cannot make any direct comparison with that portion of Wilson's work that quantifies the specific relationship between
and S/V. Furthermore, we have explicitly extrapolated the
calibration of Suzuki et al. (21) to the posttreatment
preparation; the magnitude of the error involved here is unknown, but
we suspect that it is small, because films significantly less than a
wavelength of light in thickness have little effect on the optical
properties of surfaces. On the other hand, an upper bound on this
effect can be calculated in a manner similar to that in Butler et al. (6), treating the coated septa as an optical composite
(siloxane-tissue-siloxane; index of refraction 1.4:1.33:1.4). The
scattering cross sections with and without the siloxane film present
amount to ~5%, which we take as negligible. In summary, even in
light of the above uncertainties, diffuse light scattering has the
singular advantage of being applicable to fresh, unfixed lungs, and, in
consequence, individual preparations can be used as their own controls,
either for comparison at different VL (see Fig. 5) or for
comparison at a variety of isovolume points before and after treatment
(see Figs. 6 and 7).
Altering the lung's
.
There is little doubt that treatment with nebulized siloxane
significantly alters the
in the lung. This could result from the
siloxane spreading into a thin and continuous film at high VL, where the surfactant
is greater than 21 dyn/cm,
and, through unknown mechanisms, remaining spread at low
VL, even though that is not the energetically favored
state. In this case, the lung would have an approximately constant
,
independent of VL. By contrast, it is also possible that
the siloxane contaminates the entire surfactant film sufficient to
raise
well above normal values at low VL, although not
necessarily to the siloxane-air interfacial value. This possibility is
supported by the observation that contamination from protein leakage in
diseases characterized by increased capillary permeability can lead to
significant surfactant dysfunction, with elevated
. In this case
there is no reason to believe that
posttreatment is constant. We
have no experimental evidence to resolve this question, which must
therefore remain open. Nevertheless, it is important to recognize that,
whether or not
is constant, exposure to nebulized siloxane results
in consistent changes in PL, and the values we report here
for the change in surface area per change in recoil pressure would
still be valid, insofar as this ratio is an estimate just of the
tangent
S/
PL|VL. In
particular, the specific value of
does not enter into our
calculations; in this sense one can consider the "dose" in our
treatment to be the change in PL, albeit through the agency
of the nebulized siloxane.
after treatment with nebulized
siloxane, we are confident that the resulting change in acinar architecture is close to that after a pure
alteration. In contrast to the lavage preparation, the nebulized technique leaves a much smaller volume of siloxane in the lung. Compared with the 0.0045 ml/g
lung mass we found with the nebulization method, Smith and Stamenovic
(20) report their lungs retained 0.4 ml/g after siloxane lavage, and Bachofen et al. (1) reported a similar amount
(5 ml retained in the lungs of 3-kg rabbits) after detergent lavage. These amounts are ~10% of alveolar volume at FRC and ~100 times greater than the volume of liquid retained by using the nebulization method described here. With any substantial amount of remaining fluid, there can be a significant alteration of the local alveolar architecture. Figure 7 in Bachofen et al. (4) is a
striking example of this. See also Bachofen et al. (5) for
a discussion of the alterations of alveolar architecture in
fluid-filled and fluid-rinsed preparations.
Physiological significance.
In his investigations into the relationships among PL,
surface area, and
, Wilson (25, 26) concluded that at
any given VL, increased
leads to an increase in
PL through both its direct effects [quantitatively given
by P
= (
S/V] and its
indirect effects via a net increase in ductal tissue forces due to
their serial arrangement with alveolar septa. (This is true independent
of parenchymal geometric distortion, but note that decreases in S will
reduce tissue forces within the septa, because they are in parallel
with the septal surface.) An equivalent expression of this phenomenon
is to observe that, in the absence of geometric distortion, an increase
in
would lead to increased PL in excess of that which
is observed; the presence of distortion allows the lung to relax to a
new equilibrium state at a lower PL. The extent to which
this occurs can be quantified by the change in surface area with
changes in PL under isovolume conditions. The fact that we
found less area change with changes in recoil then implies that
alterations in
such as found in disease would be expected to have a
more serious impact on PL than would have been predicted.
Indeed, the smaller the change in area with PL, the closer
P
is to the observed
P.
. Missing from his formulation is the term
associated with the energy of adsorption or desorption of molecules
from the liquid lining layer when
is changed. The effect of this
term on Wilson's results is unknown to us but invites future investigation.
Our control measurements of the dependence of surface area on
VL show that S varies roughly as
VL
is simply increased local recoil. The loss of area with abnormally
high
may compromise gas exchange as well, because of both the
simple loss of available area as well as increased tissue thickness and
decreased conductance for diffusive gas transport. However, our
observation that area loss is less than previously thought suggests
that the lung is less at risk for this to occur, although the cost for
this protection is a commensurately increased recoil pressure, as
remarked above. Changes in
are also present in normal lungs,
especially during large volume excursions but even during tidal
breathing. Our finding that there is a relatively smaller amount of
area distortion with changes in
is consistent with the observations
of Butler et al. (7), who found the expected pressure
relaxation and recovery but essentially no geometric adaptation after
step changes in volume, and the observations of Miki et al.
(16), who found essentially no hysteresis in surface area
during normal tidal breathing in live rabbits. At high VL,
our observation of a negligible change in surface area is consistent
with the observations of Bachofen et al. (2), who found
little hysteresis in S at high VL and concluded that tissue
forces play a dominant role.
In summary, our experiments demonstrate that a very small amount of
aerosolized siloxane results in significant, systematic, and
reproducible changes in the lung's pressure-volume relationship. The
increased PL at all but the highest VL are
consistent with the
of the siloxane-exposed lung being greater than
the
in the normal lung. With increased
, we found a reduction in
surface area under isovolume conditions, but of a magnitude
significantly less than that found by Wilson (26). We
speculate that the major source of this discrepancy lies in the
difference between the balance of forces that determine surface area in
chemically fixed lungs and in freshly excised lungs. Physiologically,
we conclude that increased
associated with respiratory diseases or
with industrial exposure to siloxane aerosols in the workplace has a
larger effect on PL and a smaller effect on S than would
have been predicted from the work of Wilson (26).
| |
ACKNOWLEDGEMENTS |
|---|
We are grateful to Dr. Krishna Gazula of the Harvard School of Public Health, who determined the physical characteristics of the nebulized aerosol.
| |
FOOTNOTES |
|---|
This work was supported by National Heart, Lung, and Blood Institute Grant HL-55569.
Address for reprint requests and other correspondence: G. P. Topulos, Dept. of Anesthesia, Brigham & Women's Hospital, 75 Francis St., Boston, MA 02115 (E-mail: topulos{at}zeus.bwh.harvard.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
February 8, 2002;10.1152/japplphysiol.00126.2001
Received 9 February 2001; accepted in final form 27 December 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Bachofen, H,
Gehr P,
and
Weibel ER.
Alterations of mechanical properties and morphology in excised rabbit lungs rinsed with a detergent.
J Appl Physiol
47:
1002-1010,
1979.
2.
Bachofen, H,
Schürch S,
Urbinelli M,
and
Weibel ER.
Relations among alveolar surface tension, surface area, volume, and recoil pressure.
J Appl Physiol
62:
1878-1887,
1987.
3.
Bachofen, H,
Schürch S,
and
Urbinelli M.
Surfactant and alveolar micromechanics.
In: Basic Research on Lung Surfactant, edited by von Wichert P,
and Müller B.. Basel: Karger, 1990, p. 158-167. (Prog Respir Res 25)
4.
Bachofen, H,
Schürch S,
Michel RP,
and
Weibel ER.
Experimental hydrostatic edema in rabbit lungs.
Am Rev Respir Dis
147:
969-996,
1993.
5.
Bachofen, H,
Schürch S,
and
Possmayer F.
Disturbance of alveolar lining layer: effects on alveolar microstructure.
J Appl Physiol
76:
1983-1992,
1994.
6.
Butler, JP,
Suzuki S,
Oldmixon EO,
and
Hoppin FG, Jr.
A theory of diffuse light scattering by lungs.
J Appl Physiol
58:
89-96,
1985.
7.
Butler, JP,
Miki H,
Suzuki S,
and
Takishima T.
Step response of lung surface-to-volume ratio by light scattering stereology.
J Appl Physiol
67:
1873-1880,
1989.
8.
Carney, DE,
Bredenberg CE,
Schiller HJ,
Picone AL,
McCann UG, II,
Gatto LA,
Mailey G,
Fillinger M,
and
Nieman GF.
The mechanism of lung volume change during mechanical ventilation.
Am J Respir Crit Care Med
160:
1697-1702,
1999.
9.
Gil, J,
Bachofen H,
Gehr P,
and
Weibel ER.
Alveolar volume-surface area relation in air- and saline-filled lungs fixed by vascular perfusion.
J Appl Physiol
47:
990-1001,
1979.
10.
Goerke, J,
and
Clements JA.
Alveolar surface tension and lung surfactant.
In: Handbook of Physiology. The Respiratory System. Mechanics of Breathing. Bethesda, MD: Am. Physiol. Soc, 1986, sect. 3, vol. III, pt. 1, chapt. 16, p. 247-261.
11.
Gregory, TJ,
Longmore WJ,
and
Moxley MA.
Surfactant chemical composition and biophysical activity in acute respiratory distress syndrome.
J Clin Invest
88:
1976-1981,
1991.
12.
Hoppin, FG, Jr,
and
Hildebrandt J.
Mechanical properties of the lung.
In: Bioengineering Aspects of the Lung, edited by West JB.. New York: Marcel Dekker, 1977, p. 96.
13.
Hoppin, FG, Jr,
Stothert JC, Jr,
Greaves I,
Yih-Loong L,
and
Hildebrandt J.
Lung recoil: elastic and rheological properties.
In: Handbook of Physiology. The Respiratory System. Mechanics of Breathing. Bethesda, MD: Am. Physiol. Soc, 1986, sect. 3, vol. III, pt. 1, chapt. 13, p. 195-216.
14.
Lehr, JL,
Butler JP,
Westerman PA,
Zatz SL,
and
Drazen JM.
Photographic measurement of pleural surface motion during lung oscillation.
J Appl Physiol
59:
623-633,
1985.
15.
Lewis, JF,
and
Jobe AH.
Surfactant and the adult respiratory distress syndrome.
Am Rev Respir Dis
147:
218-233,
1993.
16.
Miki, H,
Butler JP,
Rogers RA,
and
Lehr JL.
Geometric hysteresis in pulmonary surface-to-volume ratio during tidal breathing.
J Appl Physiol
75:
1630-1636,
1993.
17.
Oldmixon, EH,
Suzuki S,
Butler JP,
and
Hoppin FG, Jr.
Perfusion dehydration fixes elastin and preserves lung airspace dimensions.
J Appl Physiol
58:
105-113,
1985.
18.
Salazar, G,
and
Knowles JH.
An analysis of pressure-volume characteristics of the lungs.
J Appl Physiol
19:
97-104,
1964.
19.
Schürch, S,
Goerke J,
and
Clements JA.
Direct determination of surface tension in the lung.
Proc Natl Acad Sci USA
73:
4698-4702,
1976.
20.
Smith, JC,
and
Stamenovi
D.
Surface forces in lungs. I. Alveolar surface tension-lung volume relationships.
J Appl Physiol
60:
1341-1350,
1986.
21.
Suzuki, S,
Butler JP,
Oldmixon EH,
and
Hoppin FG, Jr.
Light scattering by lungs correlates with stereological measurements.
J Appl Physiol
58:
97-104,
1985.
22.
Topulos, GP,
Brown RE,
and
Butler JP.
Influence of lung volume on pulmonary microvascular pressure-volume characteristics.
J Appl Physiol
89:
1591-1600,
2000.
23.
Topulos, GP,
Brown RE,
and
Butler JP.
Increased surface tension decreases pulmonary capillary volume and compliance.
J Appl Physiol
93:
1023-1029,
2002
24.
Ware, LB,
and
Matthay MA.
The acute respiratory distress syndrome.
N Engl J Med
342:
1334-1349,
2000.
25.
Wilson, TA.
Relations among recoil pressure, surface area, and surface tension in the lung.
J Appl Physiol
50:
921-926,
1981.
26.
Wilson, TA.
Surface tension-surface area curves calculated from pressure-volume loops.
J Appl Physiol
53:
1512-1520,
1982.
27.
Wilson, TA,
and
Bachofen H.
A model for mechanical structure of the alveolar duct.
J Appl Physiol
52:
1064-1070,
1982.
28.
Wilson, TA,
Anafi RC,
and
Hubmayr RD.
Mechanics of edematous lungs.
J Appl Physiol
90:
2088-2093,
2001.
This article has been cited by other articles:
![]() |
G. P. Topulos, R. E. Brown, and J. P. Butler Increased surface tension decreases pulmonary capillary volume and compliance J Appl Physiol, September 1, 2002; 93(3): 1023 - 1029. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |