|
|
||||||||
1 Meakins-Christie Laboratories, McGill University, Montreal, Quebec, Canada H2X 2P2; 2 TIMC IMAG Laboratory, UMR CNRS 5525, Grenoble, France; and 3 Vermont Lung Center, Department of Medicine, University of Vermont, Burlington, Vermont 05446
| |
ABSTRACT |
|---|
|
|
|---|
The beating heart naturally oscillates the lung because of the close juxtaposition between these organs producing cardiogenic oscillations in flow that can be measured at the mouth when the glottis is open. Correspondingly, if the mouth is occluded, the same phenomenon produces cardiogenic pressure oscillations that can be measured just distal to the site of occlusion. The Fourier-domain ratio of these oscillations in pressure and flow constitutes what we call cardiogenic respiratory impedance (Zc). We calculated Zc between about 1.5 and 10 Hz in relaxed normal subjects at functional residual capacity with open glottis. Zc was insensitive to heart rate changes induced by exercise and had an imaginary part close to zero at all frequencies investigated. Its real part was similar to or smaller than resistance determined by the forced oscillation technique. We speculate that Zc measures the flow resistance of the central and upper airways of the lung. Zc may be useful as a means of obtaining information about lung mechanics without the need for an external source of flow perturbations.
input impedance; airway resistance; reactance; heart sounds; forced oscillation technique
| |
INTRODUCTION |
|---|
|
|
|---|
THE MECHANICAL PROPERTIES of the respiratory system are conveniently encapsulated in terms of its mechanical impedance (Z), a complex function of frequency that accounts for both the conservative and dissipative properties of the system. Z is usually determined by the so-called forced oscillation technique in which an external generator, such as a loudspeaker or piston pump, generates a broad-band flow into the lungs via the mouth while pressure is measured at the mouth simultaneously. Input impedance (Zin) is then obtained as the Fourier-domain ratio of pressure to flow (16). Alternatively, a transfer impedance of the respiratory system may be obtained from the Fourier-domain relationship between pressure oscillations applied to the body surface and flow measured at the mouth or vice versa (15). It is also possible to obtain regional lung impedance if the applied pressures and flows can be confined to some local region of the lung, such as the alveolar input impedance provided by the alveolar capsule oscillator technique in dogs (3). Zin, transfer impedance, and alveolar input impedance are not equivalent quantities, but rather give complementary information about respiratory mechanics, each with its own particular advantages.
Thus, in general terms, some Z relating to the mechanical properties of the respiratory system can be obtained from any oscillatory source capable of generating corresponding pairs of pressure and flow signals whose relationships are somehow affected by respiratory system mechanics. To date, virtually all endeavors of this nature have utilized external power sources to generate the necessary oscillations in pressure and flow. This allows the spectral contents of the applied signals to be optimized for the application at hand but also requires a certain amount of instrumentation that may, in some cases, be cumbersome and impractical.
In the present study, we utilize the fact that the beating heart naturally oscillates the lung because of the close juxtaposition between the organs. This results in so-called cardiogenic oscillations (CO) in flow that can be measured at the mouth when the glottis is open. Correspondingly, if the mouth is occluded, the same phenomenon leads to CO in pressure that can be measured just distal to the site of occlusion. The Fourier-domain relationship between these flow and pressure oscillations constitutes a cardiogenic output impedance (Zc), according to the classic notion of a Thevinin equivalent circuit in which an idealized pressure source (due to the heart) acts on a series source output impedance. Exactly what Zc corresponds to physiologically is unknown, but it has the potential advantage of not requiring an external oscillator to produce perturbations in flow. In contrast, whereas Zin does require an external oscillator, its physiological interpretation is well understood. Therefore, the purpose of the present study was to elucidate the physiological interpretation of Zc by investigating how it relates to Zin in normal human volunteers.
| |
METHODS |
|---|
|
|
|---|
We studied five healthy human volunteers (2 women, 3 men) with no history of lung disease. The study was approved by our Institutional Review Board, and written, informed consent was obtained from each subject.
Resting measurements of Zc and Zin were made first (see below). The subjects then exercised on a cycle ergometer in 3-min stages and from 10 to 50 W (depending on the subject). At the end of each stage, the subject got off the bicycle and immediately sat down at the measurement apparatus, and Zc and Zin recordings were again taken.
Measurement of Zc.
Subjects wore a nose clip and sat upright in a straight-backed chair
while breathing through a mouthpiece (Fig.
1A). Measurements were made at
functional residual capacity (FRC) over a period of up to 10 s
during which subjects attempted to relax all respiratory muscles while
supporting the cheeks with the hands and keeping the glottis open.
During the first half of the measurement period, cardiogenic flow
(
c) at the mouth was measured with a pneumotachograph (Fleisch
no. 1) connected to the mouthpiece. A piezoresistive differential
pressure transducer (SensorTechnics HCXPM002D6V) was connected by the
shortest possible length of flexible tubing to the two ports of the
pneumotachograph. During the final half of the measurement period, the
outflow port of the pneumotachograph was occluded, and cardiogenic
pressure (Pc) was measured with a gauge differential pressure
transducer (Fujikura, FPM-02PG) positioned just proximal to the
mouthpiece.
c and Pc were low-pass filtered at 50 Hz with 6-pole
Bessel antialiasing filters and were sampled at 128 Hz with a 12-bit
analog-digital converter (DT EZ-01, Data Translation, Marlborough, MA)
before being stored on a PC computer using the LABDAT data acquisition
software (RHT-InfoDat, Montreal).
|
c are not measured simultaneously because the
oscillations in
c arise before the airway opening is occluded
whereas those in Pc occur afterward. However, to calculate Zc, these
signals need to be phase matched, which requires an independent
reference signal related to the mechanical activity of the heart and
which can be measured regardless of whether the airway is occluded or
not. This reference was provided by the heart sound signal measured by
placing a microphone over the chest (Fig. 1). The first heart sound,
which had the largest amplitude and identifies the onset of ventricular
systole, was used as the phase reference for each CO in either Pc or
c.
Measurement of conventional Zin.
Conventional forced oscillatory measurements were made using a 50-ml
piston oscillator (Fig. 1B), which generated an 8-s
broad-band volume perturbation consisting of the superposition of seven
discrete sinusoidal components having mutually prime frequency between 0.5 and 10.25 Hz. The amplitude (A) of each sinusoidal
component was chosen according to the formula
|
c. Subjects were connected to the oscillator through the same
mouthpiece as used for the CO measurements (see above). During the
application of the 8-s forced oscillations, the subject remained
relaxed and apneic with open glottis at FRC.
The pressure signal required for calculation of Zin was measured with
the same transducer system as used for determining Zc (see above). Data
acquisition was again performed by using LABDAT. The piston oscillator
was controlled by the computer via a digital-analog converter (DAC-02,
Keithley Metrabyte, Cleveland, OH). Pressure (P) and flow (
)
signals were low-pass filtered at 50 Hz (6-pole Bessel filter), sampled
at 128 Hz, and stored for further analysis.
Data processing.
Data analysis was carried out by using the Matlab 5.3 mathematical
software (The Mathworks, Natick, MA). To calculate Zc, we first used
the heart sound signal to divide
c and Pc into individual heart
cycles. Figure 2 shows typical recordings
c and Pc, for a subject at rest, together with the heart sound
signal that identifies each cycle. Next,
c and Pc were divided
into individual cycles, and each cycle was resampled by use of linear interpolation to have exactly an integer power of two data points (usually 128). The individual
c and Pc cycles were then
ensemble-averaged (Fig. 3). Finally, Zc
was calculated by taking the ratio of the Fourier transform of the
averaged Pc to the transform of the averaged
c at those
frequencies having power equal to 2% or more of the maximum. This
assumes that the Pc and
c signals were stationary, i.e., the
only difference between successive cycles was random noise. To test the
validity of this assumption, we calculated Zc for subject 1 using each possible pair of individual Pc and
c cycles. Figure
4 shows the mean and a standard deviation
of the individual real parts (Rc) and imaginary parts (Xc) of Zc, and
demonstrates that for most frequencies the variation in individual Zc
is small, which supports our assumption of stationarity for
c
and Pc.
|
|
|
and P were divided into 4-s time windows overlapping by
50%, and the first window in each signal was discarded. Next, the fast
Fourier transform of each remaining window was calculated, and the
average auto- and cross-power spectrum between
and P was
calculated (G
and GP
,
respectively). Finally, impedance was calculated as
|
|
2
0.90 were accepted.
| |
RESULTS |
|---|
|
|
|---|
Figure 5 shows Zc and Zin for all
subjects studied at rest. In two (subjects 3 and
5), Rc and Rin agree well over most of the frequency range,
whereas in the three remaining subjects Rc is markedly lower than Rin.
The reactances are quite different in four of the subjects, with Xin
being considerably lower than Xc at all frequencies. Only in
subject 5 are Xin and Xc similar. The most striking finding,
however, is that Xc is essentially zero at all frequencies in all
subjects.
|
Figure 6 shows the effects of exercise,
and hence of increased heart rate (which reached levels 20-40%
greater than baseline), on Rc. Values of Rc at rest and two exercise
levels are shown for all subjects, normalized to the value at rest at 2 Hz. None of the values at any frequency or exercise level are
significantly different (ANOVA, P < 0.05). The mean
values of Xc at the same frequencies, for all exercise levels, are
given in Table 1. None was significantly
different from zero (one sample t-distribution, P < 0.05). These results show that increasing heart
rate had no effect on the estimated Zc.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
Each beat of the heart produces an asymmetric deformation of the lung that causes the fluctuations in airway flow (or pressure) known as CO. The amplitude of these oscillations increases with increases in cardiac output and filling pressure (10), both of which increase the mechanical deformation applied to the lung. Conversely, CO in flow are diminished or absent when airway resistance is increased (4). These observations lead to the notion that CO may harbor useful information about heart and/or lung function. Despite this, relatively little has been done to try and extract any such information. A few studies (7, 11, 19, 20) have investigated CO as a means of tracking changes in stroke volume. CO have also been used as a qualitative index for differentiating between obstructive and central apneas during sleep (1, 9, 12). However, in most situations, CO are regarded as a nuisance and attempts are frequently made to eliminate them from recordings of airway opening pressure and flow and esophageal pressure (17, 18). CO have also been associated with troublesome autotriggering of mechanical ventilators (10).
Our goal was to investigate the use of CO in airway flow and pressure
as means for determining a mechanical output impedance related to lung
function. This involved first obtaining and processing Pc and
c
in an appropriate way and then interpreting the resulting Zc in
physiological terms. The data acquisition and processing steps involved
a number of assumptions. First, we assumed that the oscillations in
c measured before airway occlusion corresponded to the
oscillations in Pc measured afterward and that each bore the same phase
relationship to the heart sounds measured throughout the recording
period. This assumption seems reasonable because we do not
expect that occluding the airway of a relaxed apneic subject will have
any immediate effect on the activity of the heart. Another issue is
that a period of apnea would be expected to cause a progressive change
in cardiac activity, and possibly lung mechanics, as a result of neural
effects and changes in blood gases. There was a slight decrease in
heart rate throughout the breath-hold maneuver, but it was small enough
to be considered negligible, and the individual cycles of Pc and
c were highly reproducible (Fig. 3). Furthermore, when Zc was
calculated with different combinations of the individual cycles in Pc
and
c, the resulting Zc showed good reproducibility (Fig. 4).
CO have interested physiologists for a long time, particularly in terms
of their manifestations in expired gas concentrations (4).
CO in flow measured at the mouth are probably due to a direct action of
the heart on the lung parenchyma (4), although exactly how
is complicated and still somewhat obscure. The heart has an irregular
shape and contracts with a twisting action, producing localized
transient inflations and deflations of the lung. This produces
transient redistribution of gas throughout the lung (4), so there is no simple relationship between movement of the heart and
the size of the CO in flow seen at the mouth. Also, the total heart
volume varies very little throughout its beat (8), so the
CO in flow presumably have little to do with stroke volume. It seems
inevitable that CO in both pressure and flow should be modulated by
anatomic or physiological factors (5, 11, 19), such as
exercise, lung volume, and body posture. In the five subjects we
studied, the peak-peak swings in Pc at rest were 0.377 ± 0.073 cmH2O. During the two levels of exercise, these swings were
0.423 ± 0.227 and 0.492 ± 0.297 cmH2O,
respectively. The corresponding swings in
c were 0.0350 ± 0.0278 ml/s at rest and 0.0346 ± 0.0376 and 0.0314 ± 0.0287 ml/s, respectively, at the two levels of exercise. The magnitudes of
the CO in both pressure and flow were, thus, perhaps surprisingly,
essentially unaffected by the exercise even though there was
considerable variation between subjects. We also observed differences
in the morphology of the oscillations in the same subject when repeated
measurements were made on different days.
An important aspect of our study was thus to ascertain the extent to which Zc is reproducible with variations in the nature of the heartbeat. We investigated this issue by determining Zc at rest and after low and moderate exercise (we did not attempt severe exercise because the subjects had to remain relaxed and apneic for 8 s immediately afterward for the measurements to be made). The subjects were all young and healthy, so minimal exercise-induced bronchodilation and changes in respiratory mechanics were expected. We found that Zc was highly reproducible regardless of the changes in heart rate and stroke volume produced by the exercise (Fig. 4), which supports the notion that Zc does indeed reflect a well-defined quantity.
To elucidate the physiological interpretation of Zc, we compared it to
Zin measured by the forced oscillation technique. Obviously, we could
not determine Zin at exactly the same frequency as the heart rate using
the same amplitude of the flow as the CO because the very presence of
the CO would have interfered with the determination of Zin. Therefore,
we identified Zin over a range of frequencies bracketing the heart rate
and its first few harmonics. We also used flow amplitudes that were
similar to those of the CO in an attempt to make Zc and Zin comparable,
because Zin is known to depend to a certain extent on flow amplitude
(16). However, this resulted in a poorer
2
than would have been obtained with a larger amplitude flow
perturbation. We attempted to achieve a compromise between requiring an
adequate
2 and retaining sufficient data to make an
effective comparison between Zin and Zc by setting our acceptance
cutoff at
2
0.90.
The most striking aspect of the comparison between Zin and Zc (Fig. 5) is that Xc is essentially zero at all frequencies whereas Xin is mostly negative and increases with frequency. Some of the subjects (subjects 3 and 4) exhibited quasi-hyperbolic Xin as described in numerous previous studies (16) whereas the remaining subjects had Xin that varied less with frequency. These differences may be due to differences in the degree of relaxation among the subjects, because any respiratory muscle tone would be expected to have a large effect on respiratory tissue stiffness and hence on Xin. Nevertheless, the absence of any appreciable Xc indicates that Zc was essentially purely resistive. This conclusion is further supported by the observation that Xc remained essentially zero despite increases in heart rate (Table 1).
The agreement between Rc and Rin was good in subjects 3 and
5 (Fig. 5), whereas in the remaining three subjects Rin was
uniformly higher than Rc. One possibility for such discrepancies is
that the spectral content of the applied flow oscillations used to obtain Zin was not the same as that producing Zc. In particular, the
harmonics of
c fell off quickly with increasing frequency, whereas the components of flow used to produce Zin did not.
Because respiratory impedance depends on many factors, including flow amplitude (6, 14, 20), it is possible that imperfect
matching of measurement conditions could have accounted for the
discrepancies between Rc and Rin. However, it is also possible that Rc
and Rin reflect different quantities. Rin is a measure of the
resistance of the total respiratory system and, even at high
frequencies, contains components from both the airways and the chest
wall tissues (2). The fact that Zc is purely real suggests
that Rc is a measure of an airway flow resistance only, without any
contribution from the respiratory tissues (which are viscoelastic and
would give rise to a significant reactive component to Zc). Therefore, we speculate that Rc is a measure of the resistance of the central and
upper airways. In other words, when the mouth and glottis are open, the
beating heart acts as a flow generator by compressing the lung tissue
at the base of the airway tree. This produces
c at the mouth.
When the mouth is occluded, the pressure oscillations producing
c are transmitted to the mouth to produce Pc. The ratio of Pc to
c then yields a measure of the output impedance of the
respiratory system, which in this case is the flow resistance of the
conduit between the site of origin of the oscillations and the
measurement point at the mouth.
It is perhaps somewhat curious that there is no appreciable imaginary part to Zc. Given that the movement of the heart is known to produce local redistribution of flow in the lungs (4), one might expect there to be some effect of a local shunt compliance. In such a situation, the imaginary part of Zc would reflect this compliance. It is possible that the shunt compliance was small and not discernable above the noise in our measurements. Interestingly, however, the real parts of Zc tend to decrease with frequency (Figs. 4 and 5). The only way this can happen, barring nonlinear effects, is if there is a reactive component to the system. Some of the Zc imaginary parts in Fig. 5 have a tendency to increase very slightly with frequency. Although these trends are not significant, they may hint at a small finite reactive component hidden in the noise.
As a further test of this interpretation of Zc, we repeated its
measurement in two of our subjects before and after the addition of an
external resistance between the mouth and the pressure and flow
transducers. The resistor had a value (1.5 cmH2O · s · ml
1) comparable
to that of normal human airway resistance. Four measurements were made
under resting conditions in each configuration. In one subject, the
real part of Zc at 2 Hz with the added resistance had a mean value of
7.29 ± 0.89 cmH2O · s · ml
1, whereas
without the added resistance the mean value was 5.32 ± 1.13 cmH2O · s · ml
1. In the other
subject, the mean value was 4.94 ± 0.84 cmH2O · s · ml
1 with the
added resistance and 3.85 ± 1.29 cmH2O · s · ml
1 without the
added resistance. The added resistance thus made a difference in the
real part of Zc for the two subjects of 1.97 and 1.09 cmH2O · s · ml
1,
respectively. These differences are not precisely equal to the added
resistance of 1.5 cmH2O · s · ml
1, but this
could easily be due to slight differences in measurement conditions
(e.g., lung volume, glottic aperture) between the measurements made
with and without the added resistor. However, the values of resistance
obtained with the added resistance bracket that of the added resistance
itself, which supports the notion that Zc gives a measure of the
resistance due to flow of gas through airways.
To be able to measure CO in Pc and
c reliably, it was necessary
for our subjects to remain relaxed with an open glottis throughout the
measurement period. This is not an easy thing to do and requires a
considerable degree of subject cooperation. Untrained subjects either
tend to close the glottis shortly after they suspend breathing or
cannot discern whether the glottis is open or closed (6,
13). This problem is almost certainly exacerbated with dyspnea
after exercise. Our subjects became quite practiced at sitting relaxed
with open glottis, and the measurements of Zin and Zc were made as
close together in time as possible. We also observed that the signals
measured during the first 4 s of the forced oscillation
measurements were similar to those obtained during the second 4 s,
which suggests that glottic aperture was consistent during these
measurements. Nevertheless, it is still possible that our comparisons
of Zin and Zc were affected by differences in glottic aperture.
In summary, we calculated Zc between ~1.5 and 10 Hz using CO in airway pressure and flow measured in relaxed normal subjects at FRC with open glottis. Zc was insensitive to heart rate changes induced by exercise and had an imaginary part close to zero at all frequencies investigated. Rc was similar to or smaller than Rin determined by the forced oscillation technique. We speculate that Zc provides a measure of the flow resistance of the central and upper airways of the lung. Zc may thus be useful as a means of obtaining information about lung mechanics without the need for an external source of flow perturbations.
| |
ACKNOWLEDGEMENTS |
|---|
We acknowledge the financial support of the Medical Research Council of Canada, the Fonds de la Recherche en Sante du Quebec, and the JT Costello Memorial Research Fund.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: J. H. T. Bates, Colchester Research Facility, Univ. of Vermont, 208 South Park Drive, Suite 2, Colchester, VT 05446 (E-mail: jhtbates{at}zoo.uvm.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 2 November 2000; accepted in final form 16 April 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Ayappa, I,
Norman RG,
and
Rapoport DM.
Cardiogenic oscillations on the airflow signal during positive airway pressure as a marker of central apnea.
Chest
116:
660-666,
1999
2.
Bates, JHT,
Brown KA,
and
Kochi T.
Respiratory mechanics in the normal dog determined by expiratory flow interruption.
J Appl Physiol
67:
2276-2285,
1989
3.
Davey, BLK,
and
Bates JHT
Regional lung impedance from forced oscillations through alveolar capsules.
Respir Physiol
91:
165-182,
1993[Web of Science][Medline].
4.
Engel, LA.
Dynamic distribution of gas flow.
In: Handbook of Physiology. The Respiratory System. Mechanics of Breathing. Bethesda, MD: Am. Physiol. Soc, 1986, sect. 3, vol. III, pt. 2, chapt. 32, p. 588-590.
5.
Engel, LA,
and
Macklem PT.
Gas mixing and distribution in the lung.
Respir Physiol
14:
37-82,
1977.
6.
Hantos, Z,
Daroczy B,
Suki B,
Galgoczy G,
and
Csendes T.
Forced oscillatory impedance of the respiratory system at low frequencies.
J Appl Physiol
60:
123-132,
1986
7.
Heckman, JL,
Stewart GH,
Tremblay G,
and
Lynch PR.
Relationship between stroke volume and pneumocardiogram.
J Appl Physiol
52:
1672-1677,
1982
8.
Hoffman, EA,
and
Ritman EL.
Heart-lung interaction: effect on regional lung air content and total heart volume.
Ann Biomed Eng
15:
241-257,
1987[Web of Science][Medline].
9.
Idiong, N,
Lemke RP,
Lin YJ,
Kwiatkowski K,
Cates DB,
and
Rigatto H.
Airway closure during mixed apneas in preterm infants: is respiratory effort necessary?
J Pediatr
133:
509-512,
1998[Web of Science][Medline].
10.
Imanaka, H,
Nishimura M,
Takeuchi M,
Kimball WR,
Yahagi N,
and
Kumon K.
Autotrigggering caused by cardiogenic oscillation during flow-triggered mechanical ventilation.
Crit Care Med
28:
402-407,
2000[Web of Science][Medline].
11.
Johnson, WK.
The dynamic pneumocardiogram: an application of coherent signal processing to cardiovascular measurement.
IEEE Trans Biomed Eng
28:
471-475,
1981[Medline].
12.
Lemke, RP,
Al-Saedi SA,
Alvaro RE,
Wiseman NE,
Cates DB,
Kwiatkowski K,
and
Rigatto H.
Use of a magnified cardiac airflow oscillation to classify neonatal apnea.
Am J Respir Crit Care Med
154:
1537-1542,
1996[Abstract].
13.
Morrell, MJ,
Badr MS,
Harms CA,
and
Dempsey JA.
The assessment of upper airway patency during apnea using cardiogenic oscillations in the airflow signal.
Sleep
18:
651-658,
1995[Web of Science][Medline].
14.
Oosteven, E,
Peslin R,
Gallina C,
and
Zwart A.
Flow and volume dependence of respiratory mechanical properties studied by forced oscillation.
J Appl Physiol
67:
2212-2218,
1989
15.
Peslin, R,
and
Duvivier C.
Partitioning of airway and respiratory tissue mechanical impedances by body plethysmography.
J Appl Physiol
84:
553-561,
1998
16.
Peslin, R,
and
Fredberg JJ.
Oscillation mechanics of the respiratory system.
In: Handbook of Physiology. The Respiratory System. Mechanics of Breathing. Bethesda, MD: Am. Physiol. Soc, 1986, sect. 3, vol. III, pt. 1, p. 145-178.
17.
Schuessler, TF,
Gottfried SB,
and
Bates JHT
A model of the spontaneously breathing patient: applications to intrinsic PEEP and work of breathing.
J Appl Physiol
82:
1694-1703,
1997
18.
Schuessler, TF,
Volta CA,
Goldberg P,
Gottfried SB,
Kearney RE,
and
Bates JHT
An adaptative filter for the reduction of cardiogenic oscillations on esophageal pressure signals.
In: Ann Int Conf IEEE Eng Med Biol Soc 17th Montreal, Canada, 1995, p. C-291.
19.
Wessale, JL,
Bourland JD,
Babbs CF,
Milewski RC,
Rockenhauser ME,
and
Geddes LA.
Correlation of the cardiogenic air flow in the respiratory airway (i.e. the pneumocardiogram) with left ventricular stroke volume in dogs.
Jpn Heart J
26:
777-785,
1985[Medline].
20.
Wessale, JL,
Bourland JD,
and
Geddes LA.
Relationship between tracheal air flow and induced changes in intrathoracic volume. A basis for calibration of pneumocardiogram.
Jpn Heart J
29:
99-106,
1988[Medline].
This article has been cited by other articles:
![]() |
J. Fontecave Jallon, E. Abdulhay, P. Calabrese, P. Baconnier, and P.-Y. Gumery A model of mechanical interactions between heart and lungs Phil Trans R Soc A, December 13, 2009; 367(1908): 4741 - 4757. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lichtwarck-Aschoff, B. Suki, A. Hedlund, U. H. Sjostrand, A. Markstrom, R. Kawati, G. Hedenstierna, and J. Guttmann Decreasing size of cardiogenic oscillations reflects decreasing compliance of the respiratory system during long-term ventilation J Appl Physiol, March 1, 2004; 96(3): 879 - 884. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |