|
|
||||||||
Cardiovascular Biophysics Laboratory, Washington University School of Medicine, St. Louis, Missouri 63110
| |
ABSTRACT |
|---|
|
|
|---|
A chamber
stiffness (KLV)-transmitral flow (E-wave)
deceleration time relation has been invasively validated in dogs with the use of average stiffness [(
P/
V)avg].
KLV is equivalent to kE,
the (E-wave) stiffness of the parameterized diastolic filling model.
Prediction and validation of 1) (
P/
V)avg
in terms of kE, 2) early
rapid-filling stiffness [(
P/
V)E] in terms of
kE, and 3) passive (postdiastasis)
chamber stiffness [(
P/
V)PD] from A waves in terms
of the stiffness parameter for the Doppler A wave
(kA) have not been achieved. Simultaneous
micromanometric left ventricular (LV) pressure (LVP) and transmitral
flow from 131 subjects were analyzed. (
P)avg and
(
V)avg utilized the minimum LVP-LV end-diastolic
pressure interval. (
P/
V)E utilized
P and
V from
minimum LVP to E-wave termination. (
P/
V)PD utilized atrial systolic
P and
V. E- and A-wave analysis generated
kE and kA. For all
subjects, noninvasive-invasive relations yielded the following
equations: kE = 1,401 · (
P/
V)avg + 59.2 (r = 0.84) and kE = 229.0 · (
P/
V)E + 112 (r = 0.80). For subjects with diastasis (n = 113),
kA = 1,640 · (
P/
V)PD
8.40 (r = 0.89). As predicted, kA
showed excellent correlation with (
P/
V)PD; kE correlated highly with
(
P/
V)avg. In vivo validation of average, early, and
passive chamber stiffness facilitates quantitative, noninvasive
diastolic function assessment from transmitral flow.
diastole; echocardiography; hemodynamics; ventricles; mathematical modeling
| |
INTRODUCTION |
|---|
|
|
|---|
IN ACCORDANCE WITH THE LAWS of fluid mechanics, left ventricular (LV) diastolic transmitral flow is generated by the atrioventricular pressure gradient (21, 30). In response to the rapid (~100 ms) development of the maximal pressure gradient, normal mitral valve (MV) leaflets separate widely as early rapid filling of the ventricle ensues, and the Doppler E wave is generated. Diastolic filling occurs in three phases. The first phase is early rapid filling, a phase initiated by a period of mechanical suction (dP/dV < 0, where P is pressure and V is volume) in all hearts (11), generated by the transient dominance of intracellular [titin (10)] and extracellular compartment [connective tissue matrix (4, 26)]-driven mechanical recoil over progressive myocardial relaxation. The magnitude of transmitral flow velocity can be modulated by numerous extracardiac factors, such as preload and afterload. The second phase is diastasis; if heart rate is typically <75 beats/min, this quiescent phase is characterized by an overall balance of forces (e.g., chamber wall, pericardial, thoracic, and similar) manifesting as absence of the atrioventricular pressure gradient and no change in volume. We emphasize that balance of forces during diastasis does not imply absence of forces (24). The third phase, atrial contraction, ensues when contraction of atrial myocardium pulls the MV annulus and aortic root toward the mediastinum, generating a positive atrioventricular pressure gradient, forcing blood into the ventricle (dP/dV > 0) and retrogradely into the pulmonary veins. Atrial contraction causes a simultaneous increase in LV pressure (LVP) and volume, and the relative duration of antegrade (transmitral) vs. retrograde (pulmonary vein) flow is related to atrioventricular compliance and pressure (1).
By definition, the slope (dP/dV) of the LVP-volume curve is the LV chamber stiffness (19). The diastolic pressure-volume relation is curvilinear (concave upward) in shape, and its slope (i.e., chamber stiffness) increases as the LV end-diastolic pressure (LVEDP) and volume increase. Classically, measurement of this chamber property has required catheterization-based measurement of LVP and the simultaneous change in volume.
Doppler-derived indexes of diastolic function have utilized the pattern of LV filling as seen on transmitral Doppler echocardiographic images (16). Selected indexes derived from this method include peak velocities of early rapid filling (Doppler E wave), atrial contraction (Doppler A wave), and the E-to-A ratio, as well as the acceleration time (tacc) and deceleration time (tdec).
Diastolic dysfunction has been associated with three types of E-wave
patterns: delayed relaxation, pseudonormalized, and restrictive. Delayed relaxation is characterized by prolonged isovolumic
relaxation time (IVRT), lower-than-normal E-wave and higher-than-normal
A-wave amplitudes, and prolonged tdec. The
pseudonormalized pattern features increased E-wave amplitude due to
elevated atrial pressure. Because this generates an increased
atrioventricular pressure gradient, it "normalizes" the E-to-A
ratio to >1. The restrictive pattern is characterized by short IVRT, a
tall, narrow E wave with high peak velocity, short
tdec (<160 ms), and a small A wave (2, 29), and it is usually associated with a third heart sound (S3) (18). The effects of compensatory mechanisms may mask the
effects of diastolic dysfunction on the LV filling pattern and indicate a lack of specificity in the ability of a transmitral flow
velocity-derived index, such as the E-to-A ratio, to quantitate
specific physiological attributes such as chamber compliance or
stiffness. For routine interpretation of transmitral Doppler waveforms,
the differentiation between the effect of relaxation and compliance has
been emphasized (2). In an effort to relate measures
derived from the Doppler E wave to specific physiological chamber
properties, Little et al. (17) proposed a method
for measurement of LV chamber stiffness. In their study in dogs
(17), average LV chamber stiffness
[(
P/
V)avg] was defined as the pressure change
(
P) from minimum LVP to LVEDP divided by the change in LV volume
during the same time interval. E-wave deceleration was modeled via the
equation of motion (Newton's law) for an undamped harmonic oscillator.
When applied to the deceleration portion of the E wave, its solution
related the rate of flow deceleration directly to atrial pressure and
MV area and inversely to the sum of left atrial and LV stiffness, as
previously modeled (20, 28). Solution of the differential
equation predicted that the deceleration portion of the E wave should
be well fit by a cosine function, the argument of which contained
tdec. The final result obtained by Little et al.
predicted that chamber stiffness (KLV) could be
calculated from tdec of the Doppler E wave as
follows
|
(1) |
is the density of blood, L is the effective,
mitral "plug-flow" length, and A is the (constant)
effective MV area. To validate the Doppler-predicted
tdec-KLV relation,
(
P/
V)avg was measured using micromanometric LVP
values. The change in volume (
V) associated with
P was based on
an ellipsoidal model of the LV with axes corresponding to
anterior-posterior, septal-lateral, and long-axis dimensions. A strong
relationship was observed between the
tdec-predicted KLV and
(
P/
V)avg invasively measured in eight dogs. Thus
tdec of the Doppler E wave was shown to be a
noninvasive index of average chamber stiffness for E waves, for which
(concave downward) the deceleration portion was well fit by a cosine
function (17). Recently, using the same conceptual
approach, Garcia et al. (5a) estimated LV operating
stiffness from tdec in 18 subjects undergoing
open heart surgery.
For quantitative diastolic function assessment from Doppler
echocardiography, a parameterized diastolic filling formalism has been
developed and previously validated (6, 7, 12). This model
accounts for the mechanical suction attribute of the heart (11,
26) via a damped, simple harmonic oscillator as a paradigm for
filling. The Doppler E-wave contour is the analog of the solution for
velocity as a function of time to the equation of motion of the
oscillator. According to Newton's law, the equation of motion for the
E wave is given by
|
(2) |
t)], the analog of atrial systole, and utilizes
the parameters F0 and
(the magnitude and
angular frequency of the forcing function, F). This model-based image-processing (MBIP) strategy for analysis of Doppler E and A waves
eliminates the need to digitize the contour by hand or determine its
attributes such as maximum velocity, E-to-A ratio, tacc, or tdec by eye
(Fig. 1).
|
The methods employed by Little et al. (17) indicated that
the deceleration portion of the E wave (starting at the peak of the E
wave) should be well fit by a cosine function, whereas the solution of
Eq. 2 predicts that the entire E wave (starting at its
onset) is well fit by the product of a sine function and a decaying
exponential. It has been shown (14) that, for small values
of the exponential decay term, the MBIP-predicted E-wave contour (sine
wave, starting at the onset of the E wave) is equivalent to the cosine
(starting at the peak of the E wave) fit to the deceleration portion of
the E wave proposed by Little et al. The equivalence of sine to cosine
(shifted by
/2 rad) permits derivation of an exact algebraic
relationship relating tdec (and
KLV) as employed by Little et al. to
kE of the parameterized diastolic filling model.
For validation, transmitral Doppler images were subjected to MBIP, and
the predicted linear relationship between KLV
and kE has been previously verified
(14).
To validate the predicted linearity of the (E-wave-generated)
kE-(
P/
V)avg relation,
simultaneous LVP and transmitral flow data were obtained in 131 subjects exhibiting a broad range of E-wave patterns. The method also
permitted determination of chamber stiffness during early rapid filling
(
P/
V)E and its correlation to
kE. In those having a period of diastasis,
passive chamber stiffness, kA, was determined
from the A wave and compared with postdiastasis stiffness
[(
P/
V)PD].
Glossary
| LV | Left ventricular |
| KLV | Chamber stiffness computed by Little et al. (17) (mmHg/cm3) |
| tdec | Deceleration time (s) |
( P/ V)avg |
Average chamber stiffness (mmHg/cm3) |
( P/ V)E |
Chamber stiffness of early rapid filling (mmHg/cm3) |
( P/ V)PD |
Passive (postdiastasis) chamber stiffness (mmHg/cm3) |
| MV | Mitral valve |
| m | Mass (g) |
| c | Damping constant (g/s) |
| k | Spring constant (linear form; g/s2) |
| x | Displacement (cm) |
| x0 | Initial displacement (displacement at t = 0; cm) |
| MBIP | Model-based image processing |
| kE | Stiffness parameter for the Doppler E wave (g/s2) |
| kA | Stiffness parameter for the Doppler A wave (g/s2) |
| TVI | Time-velocity integral |
| MSE | Mean square error |
| MVE | Maximum velocity envelope |
| LVP | LV pressure (mmHg) |
| LVEDP | LV end-diastolic pressure (mmHg) |
| IVRT | Isovolumic relaxation time (s) |
| |
METHODS |
|---|
|
|
|---|
Clinical data acquisition. Simultaneous transmitral Doppler and high-fidelity hemodynamic (LVP) data were obtained in 131 patients undergoing elective cardiac catheterization at the request of their referring physician. Informed consent was obtained in accordance with Washington University Medical Center Human Studies Committee guidelines. All subjects were in sinus rhythm. None of the subjects had any significant valvular abnormality. Simultaneous hemodynamic and Doppler echocardiographic data were acquired before injection of iodinated contrast for selective coronary angiography and left ventriculography. Subjects were sedated with diazepam (2.5-5.0 mg iv). Left ventriculography was performed in the conventional 30° right anterior oblique view using a power injector (Mark V, MedRad) via a 7-Fr pigtail catheter (Millar, Houston, TX).
Short-axis parasternal views for effective MV area determination and transmitral pulsed-Doppler flow velocity images were obtained according to standard criteria (3). The sample volume was located at the mitral leaflet tips in the four-chamber apical view. Imaging was performed using an echocardiographic imaging system (model C256, Acuson Sequoia, Mountain View, CA) equipped with a 2-MHz transducer. Between 20 and 30 consecutive beats were digitally stored in the imager's memory and downloaded to magnetooptical disks. Simultaneous micromanometric LVP values were obtained using a 7-Fr pigtail catheter (model 747P, Millar). LVP and electrocardiogram data were subjected to analog-to-digital conversion at a sampling rate of 1 kHz with a digitizing board (model MIO-16XE, National Instruments, Austin, TX) and were acquired in LabView resident on a 200-MHz Power Macintosh 7200 computer with 128 megabytes of random access memory and 500 megabytes of hard drive. Pressure and transmitral Doppler image data were synchronized by a square-wave fiducial marker. Additionally, simultaneous electrocardiogram traces were recorded as part of pressure and flow data sets. Analysis was performed off-line in the Cardiovascular Biophysics Laboratory.MBIP. Transmitral Doppler E and A waves were subjected to MBIP as previously described (6, 7, 12, 14) for model parameter determination (Fig. 1). We also previously showed that averaging of parameters from individual images is superior to averaging of images, with an average variation of <10% for all parameters (8, 9). Accordingly, individual beats were selected from the original transmitral digital Doppler image data set. To account for beat-to-beat variability, at least five beats per subject underwent MBIP, and the resulting parameters were averaged (8, 9).
Examples of model fit to a range of transmitral Doppler images with corresponding MBIP-derived parameters are displayed in Fig. 2. For E waves, model fit is not restricted to concave-downward deceleration segments only (cosine shapes) but includes E-wave patterns having an inflection point in the deceleration portion. Independently, tdec was determined for each of the E waves by two methods. The first used the conventional method (7) of extending the deceleration portion from the peak of the E wave as a straight line to the velocity = 0 axis. The second, automated method used the MBIP-predicted contour and computed the elapsed time from its peak to its intersection with the velocity = 0 axis. The tdec was used to compute KLV using Eq. 1.
|
Average chamber stiffness.
Average chamber stiffness was defined as the ratio of the change in
pressure to the change in volume [(
P/
V)avg] during
the time interval from minimum LVP to LVEDP, as defined by Little et
al. (17). Briefly, (
P)avg was defined as
the difference between minimum LVP and LVEDP for the same beats
subjected to MBIP. An example of the micromanometric LVP trace from a
single individual and the simultaneous transmitral Doppler image are shown in Fig. 3. The pressure portion
utilized for determination of (
P)avg is shown in Fig.
3C. (
V)avg was calculated by using the sum of
the time-velocity integral (TVI) for the E and A waves commencing from
the time of minimum LVP to LVEDP. To obtain the corresponding change in
volume, the two-dimensional echocardiographic effective MV area was
multiplied by the TVI.
V as a function of time is shown in Fig.
3B. The effective MV area is the echocardiographic annular
area multiplied by the (dimensionless) factor R (R < 1), where R
is the ratio of annular area E- and A-wave TVI to leaflet tip E- and
A-wave TVI. The Doppler E-wave (MBIP)-derived parameter kE was plotted vs. the invasively obtained
(
P/
V)avg for each subject. To facilitate direct
comparison with the method of Little et al. (17),
tdec was measured directly, allowing
determination of KLV. This allowed experimental
validation of the theoretically predicted (14) linear
relationship of kE to
KLV.
|
Early rapid-filling stiffness.
Stiffness during the E-wave interval (E-wave stiffness) was defined as
the change in pressure divided by the change in volume [(
P/
V)E] for the interval from minimum LVP to LVP
at diastasis. The E-wave-derived stiffness parameter
kE was plotted vs. (
P/
V)E [see Fig. 3 for graphical definition of (
P)E and
(
V)E].
Passive chamber stiffness.
Passive chamber stiffness was determined for subjects having a period
of diastasis. Presence of diastasis indicated a period of no (or
insignificant) transmitral flow, i.e., abolition of the
atrioventricular pressure gradient with concomitant equilibration of
residual atrioventricular forces. Passive chamber stiffness after
diastasis was defined by the ratio of change in pressure to change in
volume due to atrial systole [(
P/
V)PD; Fig. 3]. In
analogy to the relationship of kE to
E-wave-determined stiffness, we sought to validate the prediction that
the A-wave-derived parameter kA is linearly
related to the passive chamber stiffness. Accordingly, the relationship
of kA to (
P/
V)PD was compared.
Linear least-squares best fit and corresponding r value were
determined using DeltaGraph for echo-derived vs. invasively derived
definitions of stiffness for the above-defined average, early
rapid-filling, and passive (postdiastasis) phases of filling.
| |
RESULTS |
|---|
|
|
|---|
Group attributes.
LV ejection fraction (LVEF) as determined by ventriculography for the
entire group was normal: 70 ± 12% (SD). Additional
characteristics of the sample include gender (85 men and 46 women), age
(53 ± 10 yr), heart rate (68 ± 12 beats/min), and LVEDP
(17 ± 5 mmHg). MBIP-derived fits of the predicted velocity
contours to the clinical Doppler E-wave images showed excellent
agreement. For all (n = 131) subjects, the average mean
squared error (MSE) was 7.23 × 10
4 for the E wave
and 6.14 × 10
4 for the A wave. The ability to
compute a measure of goodness of fit in the form of MSE is an inherent
advantage of the MBIP approach. Figure 2 illustrates model-predicted
transmitral flow velocity superimposed onto selected clinical E and A
waves. Least-squares linear best fit and correlation coefficient in
this human study were determined in analogy to the study by Little et
al. (17) in their canine model. In addition, the method
and the data set permitted noninvasive prediction and invasive
validation of stiffness during selected phases of diastole.
Specifically, the Doppler-derived early rapid-filling stiffness,
kE, could be compared with
(
P/
V)E, and (postdiastasis) chamber stiffness,
kA, could be compared with (
P/
V)PD. In analogy to Little and Downes
(16), Fig. 4 displays the
data for the echocardiographically determined parameter
kE and its relation to the simultaneous,
invasively determined average stiffness:
KLV = (
P/
V)avg. Linear
regression, with least MSE best fit (n = 131), yielded
kE = 1,402 · (
P/
V)avg + 59.2 with strong
correlation (r = 0.84). The E-wave stiffness parameter kE is compared with KLV
of Little and Downes as determined by tdec of
the Doppler E wave in Fig. 5.
Linear regression yielded kE = 1.110 · (A/
L) · KLV + 55.8 (r = 0.85). Figure
6 depicts kE vs.
(
P/
V)E. Linear regression yielded
kE = 229.0 · (
P/
V)E + 112 (r = 0.80). Figure 7 depicts the Doppler
A-wave-derived parameter kA vs.
(
P/
V)PD. Linear regression yields
kA = 1,640 · (
P/
V)PD
8.40 (r = 0.89). Figure 8
combines Figs. 4, 6, and 7 and indicates the following order: slope of
the linear regression relation for early stiffness < slope for
average stiffness < slope for postdiastasis stiffness. This is
consistent with the concave-upward shape of the LV diastolic
pressure-volume curve.
|
|
|
|
|
| |
DISCUSSION |
|---|
|
|
|---|
The quest for Doppler-derived physiological indexes of diastolic function has remained an area of active investigation (22). Progress has been made through mathematical modeling of diastolic physiology, which provided quantitative insight into how Doppler E-wave contours are determined by the transmitral atrioventricular pressure gradient (15). In turn, how the atrioventricular pressure gradient depended on numerous physiological variables, such as atrial stiffness and capacitance as well as MV inertance and chamber properties, has been also described in detail (19, 21, 24, 27, 30). One inherent limitation of the (nonlinear) physiological modeling approaches is their inability to use the clinical Doppler contour as input to generate unique model parameters as output (23). Our approach circumvents this limitation by approximating the filling process by the use of a linear model for filling. This requires the lumping of all physiological determinants of transmitral flow for the E wave into three parameters to account for its amplitude (xo), width (kE), and rate of decay (c). The goodness of the model-predicted fit to in vivo contours underscores the utility of the approach as a practical tool for analysis and for facilitation of clinical decisions by accommodating the full range of Doppler flow velocity patterns encountered in vivo (13, 25).
As a result of the pioneering work of Little et al. (17) in a canine model, additional progress toward physiological interpretation of E-wave attributes has been achieved by their theoretical derivation and validation of a relationship between E-wave tdec and chamber stiffness. On the basis of their work, we previously established a direct linear relationship between the E-wave-determined parameter kE and their tdec-derived chamber stiffness KLV (14). A minor difference between the study of Little et al. and this study is that the entire E wave is used as input to the model, instead of only the tdec (2 points of the entire contour). This has the advantage of using more information to fit the model-predicted contour, thereby providing a more powerful analysis. On the other hand, it has the shortcoming of requiring computation that cannot be simply performed by inspection of the Doppler contour. Testing the analogous (kE-average stiffness relation) hypothesis in humans, extending it to include E-wave patterns having an inflection point in the deceleration portion, and being able to consider the early rapid-filling and the late, postdiastatic (passive) stiffness phases of diastole represent the focus of the present in vivo work. One key result (Fig. 4) relates invasive and noninvasive determination of (average) chamber stiffness in vivo in humans (n = 131). Figure 4 is the analog of the experimentally determined relationship observed in a canine model (n = 8) by Little et al. (17). In agreement with the canine study, a linear correlation was found.
The MBIP strategy eliminates the need to quantify the transmitral
Doppler contour or determine its attributes (e.g.,
tacc, tdec, E-to-A ratio)
"by eye." Figure 2 illustrates typical E-wave images demonstrating
excellent agreement between model prediction and the clinical velocity
contour. Furthermore, the predicted linear correlation between the
noninvasively determined parameter kE and the
invasively determined (
P/
V)avg, between
kE and KLV, between
kE and early rapid-filling stiffness, and
between kA and postdiastatic stiffness has been
validated in a large in vivo sample having a broad distribution of
Doppler flow velocity patterns.
Although a strong correlation is observed indicating that kE is a clinically viable noninvasive estimate of average chamber stiffness and that kA is an excellent estimate of passive chamber stiffness, some limitations remain.
Limitations. The limitations encountered in this study fall into two categories: methodological and conceptual.
One minor methodological limitation is the lack of individual pressure or volume alteration in the subjects studied. Ideally, to determine the diastolic pressure-volume relation and its slope
P/
V at any point
along the pressure-volume curve for an individual subject, manipulation
of pressure by pharmacological means (pressors or nitroglycerine) or
alteration of volume (by volume loading or inferior vena cava
occlusion) could be performed. These manipulations permit determination
of the entire diastolic pressure-volume relation for an individual
subject. Although these manipulations were not part of our protocol for
P/
V determination in the subjects undergoing elective diagnostic
cardiac catheterization, measurement of (
P/
V)avg as
performed by Little et al. and determination of
(
P/
V)E and (
P/
V)PD permitted their
respective determinations for each subject at the operational point
along each individual's pressure-volume curve. This is justified,
since our goal was not determination of the entire pressure-volume
relation for each subject but determination of the
echocardiographically determined average, early rapid-filling, or
passive chamber stiffness for a single diastole and its relation to the
invasively determined
P/
V for the same diastole. The large number
of subjects studied (n = 131) and inclusion of E-wave patterns having an inflection point in their deceleration portions extended the range of tdec and stiffnesses encountered.
Another minor methodological limitation concerns the reproducibly of
locating the micromanometric pressure transducer at the mid-LV level.
Because diastolic intraventricular pressure gradients exist
(5), LVP contours depend somewhat on the base-to-apex location of the transducer. Our primary criterion for mid-LV transducer location was modulated by our desire to prevent catheter-induced ectopy. Variation in pressure transducer location introduces some variability into our measurement of
P, particularly during the E
wave and likely contributes to the slightly lower (r = 0.84) correlation observed in this study compared with the correlation (r = 0.92) observed in the canine study by Little et
al. (17). During passive filling due to atrial systole,
intraventricular gradients are diminished; hence, reproducible catheter
location has less effect on determination of
P, likely contributing
to the higher correlation we observed (r = 0.89).
Our study may be subject to the concern that we did not select, sort,
or analyze our data according to clinical subset criteria such as
hypertension, diabetes, or ischemic heart disease or group them
according to the presence of various pharmacological therapies (e.g.,
angiotensin-converting enzyme inhibitors,
-blockers,
diuretics). Treating all subjects as a single group is justified by our
goal to validate the predicted physiological relationship between
invasively determined average, early rapid-filling, or passive
stiffness and the simultaneously determined noninvasive chamber
stiffness, kE or kA, in a
broad in vivo sample. This relationship is derived from the biophysical
rules of filling, which all hearts obey, rather than the type of
pathology or medication that may or may not be present. We wish to
emphasize that our findings in this group of patients have yet to be
further validated in specific pathophysiological subsets having
abnormally low LVEF or abnormally high LVEDP. Determination of the
relationship of average, early rapid-filling, or passive chamber
stiffness as a function of pathological state and pharmacological
intervention is an interesting and justified subject for future studies.
A conceptual limitation relates to the definition of average stiffness
as (
P/
V)avg and the recognition that
P (defined as
the difference between LVEDP and minimum LVP) utilized in this study
and in the prior study by Little et al. (17) does not correspond to the entire duration of filling, because it omits the
period of filling from MV opening to minimum LVP. When ventricular filling (Doppler E wave) begins, LVP continues to diminish
(dP/dt < 0) as LV volume rapidly increases
(dV/dt > 0) (18, 29). By definition,
dP/dV (stiffness) during this interval (of mechanical suction) is
negative. Negative stiffness in and of itself is not a cause for
conceptual difficulty because it indicates that the recoiling chamber
is doing external work (mechanical suction) on atrioventricular blood.
After minimum LVP is reached, stiffness changes sign and becomes
positive, indicating that the fluid, having reached maximum velocity,
is doing external (inertia generated) work on the chamber by distending
it and raising its pressure. Because (
P/
V)avg was
defined as starting at minimum LVP (16), the
suction-initiated (negative stiffness) portion of early filling is
neglected. In addition,
V was calculated commencing at the corresponding portion (minimum LVP) of the E wave and including all the
A wave. The advantage of opting for the above choice for average
stiffness is that it confines analysis to the
P/
V > 0 portion of the diastolic interval (Fig. 3). The physiological and
conceptual limitation is that it combines (suction initiated, inertia
and relaxation modulated) stiffness during the latter half of the E
wave with (passive) postdiastasis stiffness during the A wave. We were
able to address this conceptual limitation in this study by considering
the kE-(
P/
V)E and the
kA-(
P/
V)PD relations
separately. Our measurements showing a high correlation of
kE to (
P/
V)avg but an even
higher correlation of kA to
(
P/
V)PD suggest that lumping early and late diastolic
stiffness into a single averaged stiffness parameter introduces a
slight variation. The separation of early and passive chamber stiffness
manifests as a nonzero intercept of the linear regression relation in
Fig. 6 compared with the (essentially) zero intercept in Fig. 7. The nonzero y-intercept for early rapid filling reflects the
effect of determining stiffness while filling is already in progress. Note that stiffness changes sign from
P/
V < 0 (from MV
opening to minimum LVP) to
P/
V > 0 (from minimum LVP to the
end of the E wave). In comparison, for passive stiffness, flow velocity
starts from zero and ends at zero (duration of the A wave), and the
sign of
P/
V remains > 0 throughout. As expected on
conceptual grounds, maintaining the distinction between E wave
(suction-initiated) and postdiastasis A wave (passive) stiffness yields
a better correlation between noninvasive prediction and invasive validation.
Conclusion.
This study validates the model-predicted linear relationship between
the MBIP-determined, E-wave-based kE and the
simultaneous, invasively obtained (
P/
V)avg. Our
results in humans are in concert with the work of Little et al.
(17) in a canine model. We found the relationship to be
kE = 1,402 · (
P/
V)avg + 59.2 with good correlation (r = 0.84). Such validation has not been
previously performed in a large sample (n = 131) of
human subjects having normal LVEF. The model-predicted linear
relationship of kE to (
P/
V)E
was also validated and observed to be kE = 229.0 · (
P/
V)E + 112 with good
correlation (r = 0.80). In addition, the prediction that postdiastasis A-wave-based kA is linearly
related to (
P/
V)PD was validated:
kA = 1,640 · (
P/
V)PD
8.40 with excellent
correlation (r = 0.89). These results establish new
tools by which average, early rapid-filling, and passive
(postdiastasis) chamber stiffness may be noninvasively attained on a
beat-by-beat basis in vivo. The findings underscore the role that all
hearts play as mechanical suction pumps in early diastole and
strengthen our understanding of the relationship between Doppler
echo-derived parameters and their in vivo physiological analogs.
| |
ACKNOWLEDGEMENTS |
|---|
Helpful discussions with Stephanie Eucker, Mark Sessoms, and Michael Courtois are appreciated. Echocardiographic data acquisition would not have been possible without the skilled contributions of Peggy Brown. The assistance of the nursing and technical staff of the Barnes-Jewish Hospital Cardiac Catheterization Laboratory is gratefully acknowledged.
| |
FOOTNOTES |
|---|
This work was supported in part by the Whitaker Foundation (Roslyn, VA), National Heart, Lung, and Blood Institute Grants HL-54179 and HL-04023, and the Alan A. and Edith L. Wolff Charitable Trust (St. Louis, MO).
Address for reprint requests and other correspondence: S. J. Kovács, Cardiovascular Biophysics Laboratory, Box 8086, Barnes-Jewish Hospital at Washington University Medical Center, 660 South Euclid Ave., St. Louis, MO 63110 (E-mail: sjk{at}howdy.wustl.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 29 December 2000; accepted in final form 5 March 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Appleton, CP.
The hemodynamic determinants of Doppler pulmonary venous flow velocity components: new insights from studies in lightly sedated dogs.
J Am Coll Cardiol
30:
1562-1574,
1997[Abstract].
2.
Appleton, CP,
Firstenberg MS,
Garcia MJ,
and
Thomas JD.
The echo-Doppler evaluation of left ventricular diastolic function: a current perspective.
In: Diastolic Function and Dysfunction, edited by Kovács SJ.. Philadelphia, PA: Saunders, 2000, p. 513-546.
3.
Appleton, CP,
Jensen JL,
Hatle LK,
and
Oh JK.
Doppler evaluation of left and right ventricular diastolic function: a technical guide for obtaining optimal flow velocity recordings.
J Am Soc Echocardiogr
10:
271-291,
1997[ISI][Medline].
4.
Caulfield, JB,
and
Janicki JS.
Structure and function of myocardial fibrillar collagen.
Technol Health Care
5:
95-113,
1997[Medline].
5.
Courtois, MR,
Kovács SJ, Jr,
and
Ludbrook PA.
The diastolic atrioventricular relationship is a function of the site of pressure measurement in the left ventricle (Abstract).
Circulation
76:
IV-426,
1987.
5a.
Garcia, MJ,
Firstenberg MS,
Greenberg NL,
Smedira N,
Rodriquez L,
Prior D,
and
Thomas JD.
Estimation of left ventricular operating stiffness from Doppler early filling deceleration time in humans.
Am J Physiol Heart Circ Physiol
280:
H554-H561,
2001
6.
Hall, AF,
Aronovitz JA,
Nudelman SP,
and
Kovács SJ.
Automated method for characterization of diastolic transmitral Doppler velocity contours: late atrial filling.
Ultrasound Med Biol
20:
859-869,
1994[ISI][Medline].
7.
Hall, AF,
and
Kovács SJ.
Automated method for characterization of diastolic transmitral Doppler velocity contours: early rapid filling.
Ultrasound Med Biol
20:
107-116,
1994[ISI][Medline].
8.
Hall, AF,
Nudelman SP,
and
Kovács SJ.
Beat averaging alternatives for transmitral Doppler flow velocity images.
Ultrasound Med Biol
7:
971-979,
1998.
9.
Hall, AF,
Nudelman SP,
and
Kovács SJ.
Evaluation of model-based processing algorithms for averaged transmitral spectral Doppler images.
Ultrasound Med Biol
24:
55-66,
1998[ISI][Medline].
10.
Helmes, M,
Trombitás K,
and
Granzier H.
Titin develops restoring force in rat cardiac myocytes.
Circ Res
79:
619-626,
1996
11.
Katz, LN.
The role played by the ventricular relaxation process in filling the ventricle.
Am J Physiol
95:
542-553,
1930.
12.
Kovács, SJ,
Barzilai B,
and
Pérez JE.
Evaluation of diastolic function with Doppler echocardiography: the PDF formalism.
Am J Physiol Heart Circ Physiol
252:
H178-H187,
1987
13.
Kovács, SJ,
Rosado J,
Manson-McGuire AL,
and
Hall AF.
Can transmitral Doppler E-waves differentiate hypertensive hearts from normal?
Hypertension
30:
788-795,
1997
14.
Kovács, SJ,
Setser R,
and
Hall AF.
Left ventricular chamber stiffness from model-based image processing of transmitral Doppler E-waves.
Coron Artery Dis
8:
179-187,
1997[ISI][Medline].
15.
Kovács, SJ,
Yellin E,
and
Meisner J.
Modeling of diastole.
In: Diastolic Function and Dysfunction, edited by Kovács SJ.. Philadelphia, PA: Saunders, 2000, p. 459-490.
16.
Little, WC,
and
Downes TR.
Clinical evaluation of left ventricular diastolic performance.
Prog Cardiovasc Dis
32:
273-290,
1990[ISI][Medline].
17.
Little, WC,
Ohno M,
Kitzman DW,
Thomas JD,
and
Cheng CP.
Determination of left ventricular chamber stiffness from the time of deceleration of early left ventricular filling.
Circulation
92:
1933-1939,
1995
18.
Manson, AL,
Nudelman SP,
Hagley MT,
Hall AF,
and
Kovács SJ.
Relationship of the third heart sound to transmitral flow velocity deceleration.
Circulation
92:
388-394,
1995
19.
Mirsky, I.
Assessment of diastolic function: suggested methods and future considerations.
Circulation
69:
836-841,
1984
20.
Myreng, Y,
and
Smiseth OA.
Assessment of left ventricular relaxation by Doppler echocardiography: comparison of isovolumic relaxation time and transmitral flow velocities with time constant of isovolumic relaxation.
Circulation
81:
260-266,
1990
21.
Nikolic, S,
Yellin EL,
Tamura K,
Vetter H,
Tamura T,
Meisner JS,
and
Frater RW.
Passive properties of canine left ventricle: diastolic stiffness and restoring forces.
Circ Res
62:
1210-1222,
1988
22.
Nishimura, RA,
and
Tajik AJ.
Evaluation of diastolic filling of the left ventricle in health and disease: Doppler echocardiography is the clinician's Rosetta stone.
J Am Coll Cardiol
30:
8-18,
1997[Abstract].
23.
Nudelman, S,
Manson AL,
Hall AF,
and
Kovács SJ, Jr.
Comparison of diastolic filling models and their fit to transmitral Doppler contours.
Ultrasound Med Biol
8:
989-999,
1995.
24.
Omens, JH,
and
Fung YC.
Residual strain in the rat left ventricle.
Circ Res
66:
37-45,
1990
25.
Rich, MW,
Stitziel NO,
and
Kovács SJ.
Prognostic value of diastolic filling parameters derived using a novel image processing technique in patients
70 years of age with congestive heart failure.
Am J Cardiol
84:
82-86,
1999[ISI][Medline].
26.
Robinson, TF,
Factor SM,
and
Sonnenblick EH.
The heart as a suction pump.
Sci Am
254:
84-91,
1986[ISI][Medline].
27.
Sun, Y,
Sjoberg BJ,
Ask P,
Loyd D,
and
Wranne B.
Mathematical model that characterizes transmitral and pulmonary venous flow velocity patterns.
Am J Physiol Heart Circ Physiol
268:
H476-H489,
1995
28.
Thomas, JD,
Newell JB,
Choong CY,
and
Weyman AE.
Physical and physiological determinants of transmitral velocity: numerical analysis.
Am J Physiol Heart Circ Physiol
260:
H1718-H1731,
1991
29.
Thomas, JD,
and
Weyman AE.
Echocardiographic Doppler evaluation of left ventricular diastolic function: physics and physiology.
Circulation
84:
977-990,
1991
30.
Yellin, EL,
Hori M,
Yoran C,
Sonnenblick EH,
Gabbay S,
and
Frater RW.
Left ventricular relaxation in the filling and nonfilling intact canine heart.
Am J Physiol Heart Circ Physiol
250:
H620-H629,
1986
This article has been cited by other articles:
![]() |
W. Zhang and S. J. Kovacs The diastatic pressure-volume relationship is not the same as the end-diastolic pressure-volume relationship Am J Physiol Heart Circ Physiol, June 1, 2008; 294(6): H2750 - H2760. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. S. Chung and S. J. Kovacs Physical determinants of left ventricular isovolumic pressure decline: model prediction with in vivo validation Am J Physiol Heart Circ Physiol, April 1, 2008; 294(4): H1589 - H1596. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Riordan and S. J. Kovacs Elucidation of spatially distinct compensatory mechanisms in diastole: radial compensation for impaired longitudinal filling in left ventricular hypertrophy J Appl Physiol, February 1, 2008; 104(2): 513 - 520. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Shmuylovich and S. J. Kovacs E-wave deceleration time may not provide an accurate determination of LV chamber stiffness if LV relaxation/viscoelasticity is unknown Am J Physiol Heart Circ Physiol, June 1, 2007; 292(6): H2712 - H2720. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Riordan and S. J. Kovacs Absence of diastolic mitral annular oscillations is a marker for relaxation-related diastolic dysfunction Am J Physiol Heart Circ Physiol, June 1, 2007; 292(6): H2952 - H2958. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Riordan and S. J. Kovacs Stiffness- and relaxation-based quantitation of radial left ventricular oscillations: elucidation of regional diastolic function mechanisms J Appl Physiol, May 1, 2007; 102(5): 1862 - 1870. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wu, Y. Yu, and S. J. Kovacs Contraction-relaxation coupling mechanism characterization in the thermodynamic phase plane: normal vs. impaired left ventricular ejection fraction J Appl Physiol, April 1, 2007; 102(4): 1367 - 1373. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wu and S. J. Kovacs Frequency-based analysis of the early rapid filling pressure-flow relation elucidates diastolic efficiency mechanisms Am J Physiol Heart Circ Physiol, December 1, 2006; 291(6): H2942 - H2949. [Abstract] [Full Text] [PDF] |
||||
|
|