|
|
||||||||
Departments of 1 Anaesthesia and 3 Respiratory Medicine, Austin and Repatriation Medical Centre, Heidelberg 3084; and 2 Department of Anaesthesia and Pain Medicine, The Alfred, Prahan 3181, Melbourne, Victoria, Australia
| |
ABSTRACT |
|---|
|
|
|---|
Ventilation-perfusion (
A/
)
inhomogeneity was modeled to measure its effect on overall gas exchange
during maintenance-phase N2O anesthesia with an inspired
O2 concentration of 30%. A multialveolar compartment computer model was used based on physiological log normal
distributions of
A/
inhomogeneity. Increasing
the log standard deviation of the distribution of perfusion from 0 to 1.75 paradoxically increased O2 uptake
(
O2) where a low mixed venous partial
pressure of N2O [high N2O uptake
(
N2O)] was specified. With
rising mixed venous partial pressure of N2O, a threshold was observed where
O2 began to
fall, whereas
N2O began to rise with increasing
A/
inhomogeneity. This
phenomenon is a magnification of the concentrating effects that
O2 and
N2O have on each other in
low
A/
compartments. During
"steady-state" N2O anesthesia,
N2O is predicted to
paradoxically increase in the presence of worsening
A/
inhomogeneity.
alveolar-arterial difference; oxygen uptake
| |
INTRODUCTION |
|---|
|
|
|---|
THE EFFECT ON GAS
EXCHANGE of inhomogeneity of ventilation-to-perfusion ratios
(
A/
) in the lung has been explored by
previous authors (1). Analysis of log normal
distributions of expired alveolar ventilation (
A;
AE) and blood flow (
) by West (17, 18) and Kelman (8) predicted reduced gas exchange
for any gas species. For the sake of simplicity, this early modeling
assumed the presence of no soluble accompanying gases in the inspired mixture.
However, two-compartment modeling of the effect of increasing
inhomogeneity of ventilation and
predicted a paradoxical increase in the uptake of one gas when mixtures of two soluble gases
are administered, such as is the case during inhalational anesthesia
with oxygen and nitrous oxide (N2O). These findings are
presented in the accompanying paper (12).
To determine the clinical relevance of these findings, we have extended
this study using a computer model of physiological distributions of
ventilation and
to investigate the predicted effects of
differing degrees of
A/
inhomogeneity on gas
exchange in the presence of a typical inspired mixture of
O2 and N2O. The model first employed
theoretical log normal distributions. Because distributions of
AE and inspired
A
(
AI) may produce different effects, results based on
each were compared. In addition, measured distributions of
AE and
published previously by other authors using the multiple inert-gas elimination technique (6)
were modeled to see whether the patterns of widening of distributions measured in patients are expected to produce the same effects on gas
exchange as widening of the smooth theoretical distributions.
| |
METHODS |
|---|
|
|
|---|
A computer model was used to calculate the exchange of multiple
gases across the alveolar-capillary membrane according to principles of
mass balance for each gas. The model assumes that, within a
compartment, arterial and alveolar partial pressures for each gas
species are identical. The independent variables for the calculation of
alveolar partial pressures and gas exchange for each gas are its
inspired fractional concentration, the mixed venous content [or
partial pressure and Ostwald solubility coefficient (
)], and
AE or
AI and
for that
compartment. The structure and data flow of this model are outlined in
more detail in the accompanying papers (12, 13). The
distributions of ventilation and
given to the model were
obtained as follows.
Theoretical Log Normal Distributions
Log normal distributions of
and ventilation were
generated. The log SD of the distribution is the index of its spread, varying between 0 (homogeneous lung) and 1.75. West (17)
showed that, for any given mode and log SD, identical results are
obtained with a primary distribution of either
or ventilation.
Log normal distributions of either
AE or
AI can be nominated.
When a log normal distribution of
AI was nominated,
the effect of absorption atelectasis was modeled as follows. No
inspired ventilation was distributed to compartments where
AI/
was below a critical value at which
AE was calculated to be less than zero. This is
based on similar assumptions to those made by Dantzker et al.
(5) that such compartments would suffer collapse. Given
that steady-state gas exchange was being modeled, it was assumed that
perfusion of such compartments was shunt, with an end-capillary gas
content identical to that of mixed venous blood. Both
AE and
AI for these compartments
were made zero, and the inspired ventilation from them was
redistributed to the remaining compartments by multiplying each by a
scaling factor to restore total
AI to its nominated
value. Modifications incorporated by Dantzker et al. to simulate the
effect of hypoxic pulmonary vasoconstriction on the distribution of
were included. Once again, perfusion of all compartments was
scaled so that total
remained at the nominated value. An
iterative approach is required for these modifications so that final
distributions obtained were consistent with all of the input variables.
West (17) demonstrated that 10 compartments are adequate
to obtain maximal precision of results for output variables from such a
model. It was found, however, that when collapse of compartments with
critically low
AI/
was incorporated,
50 compartments were required to avoid noticeable quantization error
because of inclusion or exclusion of compartments with
AI/
values near the critical value.
Measured Distributions
Three pairs of distributions of ventilation and
were
taken from the previously published paper by Dueck et al.
(6). These were distributions of their subjects
6, 7, and 8. Each pair consisted of a
narrower and a wider distribution (taken before and after induction of
anesthesia in Dueck's subjects) whose log SD (as given by Dueck) is
listed in Table 1.
|
Analysis Performed
Theoretical log normal distributions.
A scenario typical of the maintenance phase of an inhalational
anesthetic was modeled involving administration of an inspired mixture
of 30% O2 and 70% N2O. For purposes of
comparison, regardless of whether a distribution of
AE
or
AI was nominated, overall
AE was held at 4.1 l/min and
was 4.8 l/min.
Parameters examined in the primary analysis were uptakes of
O2 (
O2), CO2
(
CO2), and N2O
(
N2O) on a global basis and
by compartment. Analyses were performed with either specified mixed
venous partial pressures or specified gas uptakes (or combinations of
these for different gases). Where specified,
O2 and
CO2 were set at 250 and 200 ml/min, respectively.
Measured distributions.
The analysis was repeated using the three pairs of measured
distributions. Gas exchange was calculated and compared for each pair
of distributions using the input variables listed above. For purposes
of standardization, the distributions were scaled so that overall
AE and
(including shunt and dead space), inspired concentrations, and
O2 and
CO2 were set at the values given above.
The mixed venous N2O partial pressure
(P
N2O of 100 ml/min
for the narrower distribution of each pair.
| |
RESULTS |
|---|
|
|
|---|
Theoretical Log Normal Distributions
Paradoxical augmentation of steady-state
N2O.
A P
N2O for
a homogeneous lung (0 log SD) of 100 ml/min. Increasing the log
SD of the distribution of
(increasing inhomogeneity of
A/
matching) produced an increased arterial
partial pressure of N2O
(PaN2O) and
N2O. This occurred in the presence of either fixed O2 and CO2 exchange or
fixed mixed venous partial pressures of these gases. The predicted
N2O at a log SD of 1.75 was
more than twice the value predicted at 0 log SD.
AE or
AI, although the results
are quantitatively different for each. The increase in
N2O at a given
P
AI/
. Paradoxical
augmentation in fact peaked at a log SD of ~1.25 and then declined
using this model. However, incorporation of the effect of hypoxic
pulmonary vasoconstriction increased uptake somewhat by reducing shunt
fraction, particularly at more severe levels of
A/
inhomogeneity.
|
Rising P



N2O was simulated,
PaN2O and
N2O fell progressively with
increasing
A/
inhomogeneity. This fall became
less steep as P

N2O increased with worsening
inhomogeneity. At a P
300 Torr, paradoxical augmentation of
N2O occurs.
|

N2O is calculated to be
negative for the homogeneous lung and still positive for the lung with
significant
A/
inhomogeneity, despite
identical inspired and mixed venous gas content.
O2 exchange.
O2 was calculated simultaneously with
N2O at increasing
P
O2 [and thus arterial
O2 partial pressure (PaO2)] occurred as
the log SD of the distribution of perfusion was increased from 0 to
1.75. The augmentation effect was similar in nature but reciprocal to
that found for N2O, in that it diminished as P


N2O),
O2 and PaO2 fell with
worsening
A/
inhomogeneity, as is normally
expected, and was demonstrated in the accompanying paper
(13) modeling a low "maintenance-phase" level of
N2O.
|
Measured Distributions
The results of modeling using the three pairs of measured distributions are summarized in Table 1. The log SD of the distributions as given by Dueck et al. (6) are listed in the first row, and the associated values for true shunt fraction and dead space fraction are shown. The calculated percent change in
N2O, moving from the
narrower to the wider distribution of each pair, incorporating the
effects of changes in shunt and dead space, are given in the bottom
row. It can be seen that predicted
N2O doubled in subject 7 as the log SD of
increased from 1.42 to 2.37.
N2O in subject 8 increased by 22% as the log SD increased from 1.17 to 1.95, despite a
doubling of dead space ventilation from 25 to 50% and an increase in
true shunt. Predicted
N2O in
subject 6 was 5% lower as the log SD increased from 0.86 to
1.67. This change was seen in the presence of an increase in true shunt
from 2 to 27%.
| |
DISCUSSION |
|---|
|
|
|---|
Paradoxical Increase in Inert-gas Uptake
This study applies distributions of ventilation and
, which
are more physiologically realistic than those used in the simple two-compartment modeling of the accompanying paper (12).
It confirms that, between certain extremes, increase of uptake of one
gas is predicted with worsening
A/
inhomogeneity with inspired mixtures of two soluble gases. The clinical
relevance of this phenomenon needs to be explored, as previous authors
have demonstrated an increase in the spread of
AE/
throughout the lung after induction of
anesthesia (3, 6, 7, 10, 11, 14).
At first glance, this increase in gas uptake seems somewhat
counterintuitive, given that it has been well demonstrated by computer
modeling by previous authors (8, 17, 18) that the expected
result of increasing inhomogeneity in
A/
matching is a reduction in gas exchange for all gas species. Modeling
of gas elimination showed that this reduction is maximal for gases with
an Ostwald blood-gas partition coefficient
equal to the overall
AE/
of the lung (4). In the
presence of an overall
AE/
of 0.86, exchange
of a gas such as N2O with a
of 0.47 would be expected
to be significantly reduced by
A/
inhomogeneity. Where gas uptake was modeled (18), this
value of
was higher, but inert-gas exchange was still predicted to
be decreased at any level of inhomogeneity compared with that in a
perfectly homogeneous lung.
The important limitation of some of these early studies was that they
assumed an insoluble vehicle gas as the balance of the inspired
mixture. More physiologically realistic models have since been applied
that allow for the interdependent exchange of multiple alveolar gases,
including N2O. Dantzker et al. (5) applied a
modified form of such a computer model, involving a log normal distribution of
AI to demonstrate that lung units
with very low
A/
may suffer collapse where
gas uptake exceeds
AI. These later models of
multiple gas exchange (5, 6), which looked at
O2-N2O mixtures, did not explore the
relationship of global gas exchange to
A/
inhomogeneity.
Examination of the exchange of individual gas species on a
compartment-by-compartment basis provides some insight into the mechanism for paradoxical augmentation. Figure
4 plots the distribution of exchange of
individual gas species in a heterogeneous lung. For simplicity, the
exchange of O2 and CO2 is shown as a single plot representing net respiratory gas exchange
(
O2-
CO2),
where
O2 predominates in the lower
A/
compartments and
CO2 in the better ventilated areas. It
shows that
O2 and
N2O occur predominantly
within low
A/
compartments and largely follow the distribution of
. This is consistent with the
perfusion-limited nature of uptake of these gases.
|
Uptake of one gas within the lung will raise the concentration of other
gases in the alveolar gas mixture. The net exchange of O2
and CO2 governs the magnitude of the concentrating effects on alveolar N2O within each compartment. N2O
exchange will also exert similar effects on O2 and
CO2. The asymmetric nature of respiratory gas exchange
across the distribution of
A/
shown in Fig. 4
results in an asymmetric concentrating effect on alveolar N2O. This is demonstrated by Fig.
5, which shows the different distributions of N2O partial pressures in two lungs of
differing degrees of inhomogeneity. In the moderately low
A/
compartments, substantial
O2 concentrates N2O in the
alveolus and drives
N2O there. These compartments have the greatest
, and here the
concentrating effect on alveolar N2O is most powerful. Lung
units with moderately low
A/
receive a
substantial proportion of total pulmonary
and thus would be
expected to have a dominating influence on the content of these
perfusion-limited gases in arterial blood and thus on total exchange of
these gases.
|
Figure 6 demonstrates quantitatively the
process of paradoxical augmentation by comparing two lungs of differing
A/
homogeneity. For both the more uniform and
the more inhomogeneous lung, the area under the N2O curve
is the overall
N2O. As the
SD of the distribution increases, there is redistribution of a greater
proportion of total
to lower
A/
areas.
N2O increases
markedly in these compartments as this occurs, and this increase
outweighs the reduced uptake in higher
A/
lung units that occurs as their
is reduced. Figure 6 shows
that, for the overall lung, the concentrating effects of
O2 on alveolar N2O are
magnified by greater inhomogeneity.
|
O2 and
N2O produce competing
concentrating effects, predominantly in these low
A/
compartments, the balance of which is
determined by their solubilities, inspired concentration, and mixed
venous content. The mutual enhancement of uptake that O2 and N2O possess is limited only by the approach of their
alveolar partial pressures to the mixed venous for each gas within
these compartments and to a lesser extent by dilution of alveolar gas by CO2.
An interesting manifestation of paradoxical augmentation is that when
PaN2O is very close to
P
A/
inhomogeneity, despite identical inspired
and mixed venous gas content. Examination of gas exchange in such an
inhomogeneous lung on a compartment-by-compartment basis shows that
N2O is being taken up by low
AE/
units and eliminated by high
AE/
units, with
the balance determining the direction of global gas exchange. This
"flow through" of gas within the lungs has been predicted to occur
for N2 in the presence of
A/
inhomogeneity, and raised urinary N2 has been used as a
test for the severity of chronic pulmonary obstructive disease
(1, 9). It can be seen for other inert gases at particular
ranges of inspired and mixed venous partial pressures. At high degrees
of inhomogeneity,
N2O in low
A/
units is magnified to a considerably
greater extent than N2O exchange in other compartments
(Fig. 7).
|
Dueck et al. (6) measured
N2O in their subjects during
maintenance-phase anesthesia. However, there is no data available in
the literature measuring
N2O
and changes in
A/
distributions simultaneously. Most studies employing the multiple inert-gas elimination technique in anesthetized patients have measured
distributions before and after induction of anesthesia to generate
widening of these distributions. There are obvious difficulties in
directly measuring changes in N2O exchange that occur in
patients who are not receiving anesthesia when the first distribution
is taken. For this reason, we have used computer modeling to simulate
gas exchange with given
A/
distributions.
Where this was applied to Dueck's published distributions, the
predictions of modeling of both two-compartment (12) and
theoretical log normal distributions were confirmed; widening
inhomogeneity of
A/
increases
N2O where maintenance-phase
levels of inert-gas exchange are modeled. This also casts some light on
the physiological factors that will limit paradoxical augmentation.
These measured distributions were collected from patients with
significant chronic lung disease and also incorporate varying degrees
of true shunt and dead space ventilation. The calculated increase in
N2O was greatest for the
distributions of subject 7, where there was no significant change in true shunt or dead space moving from the narrower to the
wider distribution. Using the distributions of subject 8, an
increase in
N2O was still
predicted, despite a large increase in dead space ventilation i.e., a
doubling of the amount of ventilation that was effectively wasted. The
greater uptake by perfused lung units still overcame the effects of
reduced
A to them. Similarly, the distributions of
subject 6 predicted only a small reduction in
N2O, despite the
superimposition of a large proportion of true shunt accompanying the
wider distribution. Thus augmentation of
N2O in ventilated lung units
largely compensated for the reduction in effective pulmonary capillary
imposed by the increased shunt.
It is of interest that, in the study by Dueck et al. (6),
the measured
N2O was highest
(325 ml/min uptake after 85 min of N2O anesthesia) in
subject 7, who demonstrated the most severe degree of
inhomogeneity of pulmonary
. Examination of the distributions measured for this subject show that there was a significant proportion of lung units with low
AE/
at the time of the
measurement. We have shown that it is in these compartments that the
most powerful concentrating effects occur, which drive the paradoxical
increase in inert-gas exchange.
At higher levels of
N2O,
augmentation of
O2 with increasing
A/
inhomogeneity is predicted (Fig. 3). This
may produce clinically significant increases in PaO2
compared with the homogeneous lung. The corresponding clinical scenario
is the immediate postinduction phase. The concentrating and second gas
effects of rapid
N2O early
in an inhalational anesthetic were described by Stoelting and Eger
(16). They demonstrated that high
N2O raises the alveolar
concentration of accompanying gases, including O2, by contraction of alveolar volume with or without indrawing of further inspired gas to replace the lost volume. However, Fig. 3 shows that a
more pronounced postinduction increase in PaO2 is
expected at more severe levels of
A/
inhomogeneity.
The accompanying paper (12) shows that competing
concentrating effects of uptake of one gas on the other are inevitably present in any lung compartment where two gases are being taken up.
Logically, the net concentrating effect can only be in one direction at
any given time. The position of balance of these processes and the
direction of the net concentrating effect are determined by the
relative inspired concentrations and uptakes of the two gases. In the
more complex physiological model presented here, the point of balance
is complicated by a number of factors not encompassed in the simplified
treatment given by the accompanying paper (12). These are
the presence of alveolar CO2 exchange, the differing
solubilities of the gases involved, and the alinear nature of the
dissociation curve for O2, which cause these gases to be
taken up unequally in lung compartments of different
A/
.
As shown by Figs. 2 and 3, the direction of the overall concentrating
effect changes at a certain
P
N2O. Inhomogeneity simply
magnifies these existing concentrating effects, and thus paradoxical
augmentation can only be seen for one gas or the other at any time.
Clinical Considerations: Inert-gas Exchange
Paradoxical augmentation of gas uptake with worsening
A/
inhomogeneity is predicted, regardless of
the nature of the distributions used to generate the data. It is
predicted by simple two-compartment models and more physiologically
realistic log normal distributions. Given the increase in
A/
inhomogeneity seen normally in
anesthetized patients, paradoxical augmentation of
N2O is likely to be
prevalent in any patient during maintenance-phase N2O anesthesia.
It can be seen from Fig. 2 that the phenomenon of paradoxical
augmentation of gas uptake only commences when the mixed venous partial
pressure has risen above a certain threshold (in this scenario 300 Torr). Thus it is a product of a relatively low inspired-to-mixed venous partial pressure gradient and commences when uptake of the gas
has declined below a certain value. Consideration of changing gas
exchange over time shows that this threshold
P
N2O is at ~550 ml/min.
Severinghaus (15) states that the
N2O (ml/min) at time
t (min) after introduction of an inspired N2O
concentration of 70% is given by
|
N2O is predicted to increase
with worsening
A/
inhomogeneity.
One of the physiological factors that might be expected to minimize
paradoxical augmentation of
O2 and
N2O is incomplete denitrogenation of the body. This is due to accumulation of the poorly
soluble N2 in low
AE/
lung
compartments as shown by Dantzker et al. (5), where its
presence dilutes the most potent concentrating effects of
O2 and
N2O. The effect of this was examined by repeating the analysis of Fig. 1, superimposing an excretion of N2 of 10 ml/min from the lungs. This level of
N2 elimination is consistent with the findings of Beatty et
al. (2) for patients during maintenance-phase anesthesia.
Figure 8 shows that
N2O is reduced in the
presence of N2 elimination to a greater extent at higher
levels of
A/
inhomogeneity, but that
paradoxical augmentation still clearly occurs for
N2O despite this.
|
The lack of effect of N2 retention on paradoxical
augmentation of
N2O may not
hold true for
O2, however. Paradoxical
augmentation of
O2, if it is a clinical
reality, occurs during the early postinduction phase of rapid
N2O, when incomplete
denitrogenation of the alveolar volume is likely. The presence of
significant levels of retained alveolar N2 throughout the
lung is likely to ablate much of the early concentrating effect on
alveolar O2.
Conclusion
A model of physiological distributions of
A/
values confirms that concurrent
administration of O2-N2O mixtures results in
competing second gas and concentrating effects modulating uptake of
both gases. During steady-state conditions typical of the
maintenance-phase of a N2O anesthetic, this results in a
paradoxical increase in
N2O
with increasing
A/
inhomogeneity. This
paradoxical augmentation is due to increasing concentration of alveolar
N2O by
O2 in a greater
proportion of lung compartments with low
A/
ratios.
At higher levels of
N2O,
such as are seen in the early phases after anesthetic induction (and in
the absence of retained alveolar N2), paradoxical
augmentation of
O2 with worsening
A/
inhomogeneity is seen instead. The effect
on O2 is not seen concurrently with that for
N2O. This is because the effect of inhomogeneity is to
magnify the existing concentrating effects of uptake one gas on the
other. These can only operate in one direction or the other at a time,
depending on the relative uptakes and inspired concentrations of the gases.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: P. J. Peyton, Dept. of Anaesthesia, Austin & Repatriation Medical Centre, Heidelberg 3084, Melbourne, Australia (E-mail: phil{at}austin.unimelb.edu.au).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 4 April 2000; accepted in final form 17 January 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Aksnes, E,
and
Rahn H.
Measurement of total gas pressure in blood.
J Appl Physiol
10:
165-173,
1957
2.
Beatty, PCW,
Kay B,
and
Healy TEJ
Measurement of the rates of nitrous oxide uptake and nitrogen excretion in man.
Br J Anaesth
56:
223-232,
1984
3.
Bindislev, L,
Hedenstierna G,
Santesson J,
Gottlieb I,
and
Carvallhas A.
Ventilation-perfusion distribution during inhalation anesthesia.
Acta Anaesthesiol Scand
25:
360-371,
1981[ISI][Medline].
4.
Colburn, WE,
Evans JW,
and
West JB.
Analysis of effect of the solubility on gas exchange in nonhomogeneous lungs.
J Appl Physiol
37:
547-551,
1974
5.
Dantzker, DR,
Wagner PD,
and
West JB.
Instability of lung units with low
A/
ratios during O2 breathing.
J Appl Physiol
38:
886-895,
1975.
6.
Dueck, R,
Young I,
Clausen J,
and
Wagner PD.
Altered distribution of pulmonary ventilation and blood flow following induction of inhalational anesthesia.
Anesthesiology
52:
113-125,
1980[ISI][Medline].
7.
Hedenstierna, G,
Lundh R,
and
Johansson H.
Alveolar stability during anesthesia for reconstructive vascular surgery in the leg.
Acta Anaesthesiol Scand
27:
26-34,
1983[ISI][Medline].
8.
Kelman, GR.
A new lung model: an investigation with the aid of a digital computer.
Comput Biomed Res
3:
241-248,
1970[ISI][Medline].
9.
Klocke, FJ,
and
Rahn H.
The arterial-alveolar inert gas ("N2") difference in normal and emphysematous subjects, as indicated by the analysis of urine.
J Clin Invest
40:
286-294,
1961.
10.
Lundh, R,
and
Hedenstierna G.
Ventilation-perfusion relationships during anesthesia and abdominal surgery.
Acta Anaesthesiol Scand
27:
167-173,
1983[ISI][Medline].
11.
Lundh, R,
and
Hedenstierna G.
Ventilation-perfusion relationships during halothane anesthesia and mechanical ventilation. Effects of varying inspired oxygen concentration.
Acta Anaesthesiol Scand
28:
191-198,
1984[ISI][Medline].
12.
Peyton, PJ,
Robinson GJB,
and
Thompson B.
Ventilation-perfusion inhomogeneity increases gas uptake: theoretical modeling of gas exchange.
J Appl Physiol
91:
3-9,
2001
13.
Peyton, PJ,
Robinson GJB,
and
Thompson B.
Effect of ventilation-perfusion inhomogeneity and N2O on oxygenation: physiological modeling of gas exchange.
J Appl Physiol
91:
17-25,
2001
14.
Rehder, K,
Knopp TJ,
Sessler AD,
and
Didier EP.
Ventilation-perfusion relationship in young healthy awake and anesthetized-paralyzed man.
J Appl Physiol
47:
745-753,
1979
15.
Severinghaus, JW.
The rate of uptake of nitrous oxide in man.
J Clin Invest
33:
1183-1189,
1954.
16.
Stoelting, RK,
and
Eger EI.
An additional explanation for the second gas effect: a concentrating effect.
Anesthesiology
30:
273-277,
1969[ISI][Medline].
17.
West, JB.
Ventilation-perfusion inequality and overall gas exchange in computer models of the lung.
Respir Physiol
7:
88-110,
1969[ISI][Medline].
18.
West, JB.
Effect of slope and shape of dissociation curve on pulmonary gas exchange.
Respir Physiol
8:
66-85,
1969[ISI][Medline].
This article has been cited by other articles:
![]() |
J. S. Yem, M. J. Turner, A. B. Baker, I. H. Young, and A. B. H. Crawford A tidally breathing model of ventilation, perfusion and volume in normal and diseased lungs Br. J. Anaesth., November 1, 2006; 97(5): 718 - 731. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. W. Hahn and A. D. Farmery Gas exchange modelling: no more gills, please Br. J. Anaesth., July 1, 2003; 91(1): 2 - 15. [Full Text] [PDF] |
||||
![]() |
P. J. Peyton, G. J. B. Robinson, and B. Thompson Ventilation-perfusion inhomogeneity increases gas uptake: theoretical modeling of gas exchange J Appl Physiol, July 1, 2001; 91(1): 3 - 9. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |