Departments of Medicine and Physiology/Biophysics, University of Calgary, Calgary, Alberta, Canada T2N 4N1
| |
ABSTRACT |
|---|
|
|
|---|
Previously, we developed a balloon transducer to measure the constraint of the pericardium (i.e., pericardial pressure) on the surface of the heart. It was validated physiologically in that it was shown to measure a pressure equal to the difference between the left ventricular end-diastolic pressure measured before and after pericardiectomy at the same left ventricular volume. To define its static operating characteristics, we loaded the balloon nonuniformly with weights that covered fractions of the balloon surface and found that the balloon accurately recorded the average stress if the stress was applied over at least 23% of its surface. To test its performance when curved, we placed it in large and small cylinders (minimum diameter 31 mm) and found that the balloon accurately recorded the stress. To define its dynamic operating characteristics, we applied sinusoidal stresses and found that its frequency response was limited only by that of the connecting catheter. When better dynamic response is required, we introduce a micromanometer-tipped catheter to obtain a unity-gain frequency response that is flat to 200 Hz.
pericardium; pericardial pressure; transducer; frequency response; epicardial stress
| |
INTRODUCTION |
|---|
|
|
|---|
A CENTURY AGO, Barnard (5) concluded that the pericardium is a significant constraint to filling of the heart. Many early investigators also observed that the parietal pericardium exerts a significant stress on the epicardium, but various transducers and catheters used to measure this constraint have yielded conflicting results (14, 17, 20, 24, 30).
Scalar terms such as "pressure" have been used to describe the stresses that the pericardium imposes on the epicardial surface. If one considers the tangential stress (shear stress) at the interface of these two structures to be minor, then the remaining radial stress might be considered to be scalar in nature, because the total vector stress on the epicardium will be normal to both surfaces, emulating liquid behavior over small elemental areas. The same situation can exist if a sufficient amount of liquid is present in the pericardial space to separate the epicardial and pericardial surfaces such that the distribution of radial stresses behaves hydrostatically (25).
Attempts have been made to measure pericardial pressure directly by using open-end catheters (4, 14, 20, 30), air-filled balloons (9, 17, 27), and liquid-filled balloons (12, 25). As measured by these devices, pericardial pressure has ranged from zero to nearly the cavitary pressure of the chamber being investigated, indicating the need for a standard in transducer design and calibration.
In 1985, Smiseth et al. (25) showed that pericardial pressure was strikingly dependent on the method of measurement used. They confirmed that the conventional open-ended catheters used to measure pericardial constraint in earlier studies registered a negligible pressure; however, they also showed that a balloon adapted from Holt et al. (12) recorded a hemodynamically significant pressure. (Since that time, the design and specifications of the balloon have not been changed.) The pressures measured from these flat balloons in surgical patients (7, 10, 29) were later shown to follow the blood volume-induced changes in right atrial pressure, in both direction and magnitude. As Tyberg et al. (28) originally proposed, the magnitude of pericardial constraint may be determined by comparing the (vertical) shift in the left ventricular (LV) end-diastolic pressure-volume relationships before and after pericardiectomy. This rationale for determining the true magnitude of pericardial constraint has been adopted by several other investigators (2, 3, 11).
The purpose of this investigation was to provide a systematic description of the fabrication technique and the static and dynamic operating characteristics of the balloon transducer used in several previous studies (6, 10, 15, 25, 29). In addition, we will describe a simple modification of the balloon that provides higher frequency response and fewer motion-induced artifacts.
| |
METHODS |
|---|
|
|
|---|
Balloon fabrication.
The pericardial balloon is fabricated by using a 7 cm × 4 cm flat
Silastic sheet (0.25 mm thick; compound AR131; Armet Industries, Concord, Ontario, Canada) wrapped around a 14-cm length of medical grade Silastic tubing (1.58 mm ID; 3.18 mm OD; Dow Corning, Midland, MI, catalogue no. 602-285). The tubing has three 2.0 mm × 1.5 mm
holes, spaced 5.0 mm apart, cut into its side starting 7.5 mm from the
end (see Fig. 1). The Silastic sheet is
then wrapped around the tubing, and the opposing surfaces are bonded by
using silicone adhesive (Dow Corning, catalog no. 891). The fabrication procedure is standardized by using a metal mold (schematic diagram available from the National Auxiliary Publications Service) that allows
the balloon to be cured under slight pressure (10 mmHg). This method
standardizes the internal volume and geometry. The opposing surfaces to
be bonded are cured at 50°C for 24 h and then removed from the
mold and left to stand at room temperature and out-gas for another
24 h. The edges of the balloon are trimmed to standard dimensions
(30 mm × 30 mm; a smaller, 14- × 12-mm balloon has also been
fabricated). The advantages of using Silastic sheets and silicone
adhesive are the long shelf life after fabrication, nonreactivity to
body tissues (8), flexibility, and resistance to fatigue
and repetitive gas sterilizations. For most experiments, the balloon is
attached to a length of 7-Fr angiographic catheter material, cut as
short as convenient. Recently boiled room-temperature water is
introduced into the balloon, and air bubbles are purged. Operating
volumes range from 0.70 to 0.90 ml.
|
Static calibration: Uniform stress application.
For convenience and reproducibility, we designed a compact balloon
calibrator to apply stresses normal to the balloon surface using a
thin, unstressed, compliant Silastic membrane (Fig.
2). The thin membrane (the same material
used in the pericardial balloon) was attached to a Plexiglas cylinder
(12.7 cm OD; 11.4 cm ID) so that it was unstressed but did not present
any fold or redundancy to the balloon or to its supporting end plate.
The other end of the cylinder was sealed with a Plexiglas end plate
(15.2 cm, diameter; 1.27 cm, thickness). By placing the Silastic
membrane over the pericardial balloon, which lay on the flat end plate,
the desired configuration was achieved. The end plate was held against
the thin membrane by use of three brass screws. A small trough was machined into the end plate, permitting passage of the tubing for
connection to a pressure transducer. The pressure inside the calibrator
could be manipulated and measured through a side port. Because the
Plexiglas was clear, it was possible to observe the balloon to ensure
its proper position. The end plate had several small holes drilled
through its surface to allow air to escape. This ensured that the only
stress applied to the balloon was through contact with the Silastic
membrane and the supporting end plate. To permit testing under negative
pressure, an alternate end plate was fabricated. This end plate had
only one hole, which allowed air to escape during positive
pressurization; the hole was then sealed and the pressure in the test
chamber was lowered to expose the balloon to negative stresses. Details
of the calibration procedure are given in APPENDIX A.
|
Static calibration: Nonuniform stress application. A benchtop experiment was performed to investigate whether the pericardial balloon would accurately record an average stress, even if the load had been applied nonuniformly. The effects of nonuniform loading were examined by placing rectangular blocks of different areas (1.0, 2.1, 4.0, 6.1, and 8.0 cm2) at the center of the balloon's transducing contact surface (see Fig. 1) while the balloon was lying on a flat surface. These blocks were loaded with weights that were adjusted to apply identical stresses (i.e., force per area of the block). Arbitrarily, we chose stresses equal to 18 mmHg (23,940 dynes/cm2 = 24.4 gm/cm2) and 32 mmHg (42,560 dynes/cm2 = 43.4 gm/cm2). For example, to achieve a stress equivalent to 32 mmHg, the 2.1 cm2 block was loaded with 91.1 (43.4 × 2.1) gm, the 4.0 cm2 block was loaded with 173.6 (43.4 × 4.0) gm, and so on. The pressures measured by the balloon were plotted against the areas of the blocks when so loaded. If the measured pressure was proportional to the area of the balloon surface that was covered by the loaded block, it was concluded that the balloon had spatially integrated (i.e., averaged) the applied stress appropriately.
Static calibration: Stress applied over a curved surface.
Bending a balloon might introduce Laplacian stresses on the membranes,
thereby exaggerating the recorded pericardial pressure. The calibration
was verified by placing the pericardial balloon within different
Plexiglas cylinders (internal diameters 31, 38, and 50 mm) and
inflating a thin-walled cylindrical balloon (i.e., a condom) to
compress it against the inside wall of the cylinder (see Fig.
3). The surfaces were flushed with water,
and the condom was inflated to different measured pressures. According
to the pleural-pressure measurement technique developed by Lai-Fook et al. (16, 31), the stress that the condom exerted on the
inner surface of the cylinder was measured by use of a pressure
transducer (Model P23 Db, Statham-Gould) connected by a liquid-filled
tubing to a hole in the wall of the cylinder at the same hydrostatic level as the pericardial balloon. The pressure measured by the pericardial balloon was compared with the condom pressure and that
measured by the wall transducer. Later, the unconstrained condom was
inflated and the distending pressure was plotted as a function of
condom diameter (measured using vernier calipers), to allow estimation
of the transmural pressures required to achieve condom diameters equal
to the inside diameters of the cylinders.
|
Dynamic calibration. To determine the balloon's frequency response, we constructed a dynamic calibration chamber using a high-fidelity speaker (model SW6025G Alpha Sub-Woofer, Heco-Hennel and, Germany). The speaker was driven by two 100-W, direct-current coupled amplifiers arranged in a class-B push-pull configuration (model SM-100, OPAMP Labs, Los Angeles, CA), which amplified a sinusoidal input signal (model 180 LF sweep generator, Wavetek, San Diego, CA). The static calibrator was modified to accommodate the speaker, which imposed a time-varying pressure onto the membrane overlying the pericardial balloon (schematic diagram available from the National Auxiliary Publications Service). The oscillating chamber pressure was measured by using a micromanometer-tipped catheter (model PR-279 8F Mikro-tip, Millar Instruments) introduced through a side port. The signal from the sweep generator was compared with the pressure measured in the chamber by using specially designed circuitry to create an error signal that was used for negative feedback to the power amplifier of the speaker. This sinusoidal pressure-generating system kept total harmonic distortion of the chamber-pressure waveform to <5% over frequencies ranging from 2 to 200 Hz (peak sinusoidal pressures up to 20 mmHg). The calibration chamber was designed to be pressurized to 60 mmHg (i.e., the mean chamber pressure could be 40 mmHg on which a ±20 mmHg sinusoidal pressure could be modulated). To measure the stresses effectively being imposed on the balloon by the membrane, a load cell (diaphragm diameter 25 mm; model P23-BB venous pressure transducer, Statham, Hato Rey, Puerto Rico) was placed flush with the surface of the balloon-support end plate, immediately beneath the position to be occupied by the balloon. The load-cell output was also measured with the balloon in place to assess the dynamic effect of the presence of the balloon. If the relation of chamber pressure to load-cell "pressure" (i.e., stress) was unaffected by the presence of the balloon, it was concluded that the balloon did not alter the dynamic characteristics of the system and balloon pressure would track chamber and load-cell pressures. The pressure recorded from the balloon and that from the load cell were compared with the chamber pressure in terms of amplitude and phase lag.
| |
RESULTS |
|---|
|
|
|---|
Static calibration: Uniform stress application.
More than 90% of the balloons fabricated using the mold were found to
pass the static calibration procedures. A typical calibration plot is
shown in Fig. 4, which illustrates the
linearity of these devices. The difference between the calibrator and
balloon pressure never exceeded 1 mmHg over the range of pressures.
When properly calibrated, 10 mold-fabricated balloons were found to
have volumes of 0.775 ± 0.054 ml (means ± SD) at which they
were unstressed.
|
Static calibration: Nonuniform stress application.
Figure 5 shows the results of the
nonuniform loading test. When rectangular blocks of different sizes
were loaded with weights such that stresses equivalent to 32 or 18 mmHg
were applied, the pressures recorded from the balloon were proportional
to the areas of the blocks. When a block smaller than 23% of the
transducing contact surface (e.g., 11%) was used, the balloon
underestimated the applied stress. However, the balloon transducer
accurately recorded the average stress when that stress was applied
over at least 23% (2.1/9.0) of its transducing contact surface.
|
Static calibration: Stress applied over a curved surface.
Figure 6 shows that, when the pericardial
balloon was placed within cylinders of various sizes, the recorded
pressure increased 1:1 with the inflation pressure of the condom. The
X-axis offset can be approximately accounted for by the fact
that the unconstrained condom required pressure to attain the diameters
of the cylinders: 1, 5, and 10 mmHg for the 31-, 38-, and 50-mm
cylinders, respectively. The figure also shows that the balloon
recorded a pressure that was equal to that recorded through an open
side hole at the same level (16). These results confirm
that the pericardial balloon measures radial stress accurately when
confined within spaces with radii of curvature similar to the
pericardial space. (Although there is an unexplained ~1 mm difference
between pressures recorded using the balloon and the side hole when the
50-mm cylinder was used, we concluded that this did not represent an
effect of curvature because it only occurred in the largest cylinder.)
|
Dynamic calibration.
In Fig. 7 (top), the amplitude
ratio (pericardial balloon pressure/chamber pressure) is plotted
against the frequency of the sinusoidal input function. A typical
second-order, underdamped response with a peak resonance and a 180°
phase shift at ~25 Hz is seen (the falloff was linear when amplitude
ratio and frequency were plotted on logarithmic scales). Because there
was always a 1:1 relationship between the chamber pressure and the
load-cell "pressure" (data not shown), it was clear that the
calibrator membrane did not modify the time-varying chamber pressure.
The frequency response of the 7-Fr catheter itself (data not shown) was
also determined and found to be virtually identical to that of the
pericardial balloon. Because the frequency response of the
high-fidelity balloon was flat to 200 Hz (only data to 100 Hz shown),
we concluded that the limitation in the frequency response of the
pericardial balloon was determined by the catheter. As shown in Fig. 7,
bottom, the phase response for the pericardial balloon
shifted from 0 to 180°. Typical of a second-order resonant system, a
90° phase shift occurred at the frequency at which there was maximum
resonance.
|
| |
DISCUSSION |
|---|
|
|
|---|
The simple tests reported in this paper demonstrate that the pressure recorded from a flat, liquid-filled, pericardial balloon transducer accurately reflects the stress applied to its surface. This is true not only when a static stress is applied uniformly to the balloon but also when the stress is applied over as little as 23% of its area. It is also true when the balloon is deformed by placing it on a surface more curved than the free wall of the canine LV. Although the frequency response of the standard balloon transducer is limited by that of the connecting catheter, the standard balloon can be modified by inserting a micromanometer-tipped catheter inside the balloon, extending its frequency response to 200 Hz, and reducing its susceptibility to catheter motion-induced artifacts. Therefore, in addition to the fact that the balloon transducer records a pressure equal to the constant-volume shift in LV end-diastolic pressure (28), these results indicate that there is no reason to doubt that the balloon transducer can be used to measure epicardial radial stress (i.e., "pericardial pressure") reliably throughout the cardiac cycle.
Although we and others have filled pericardial balloons with water or saline (7, 12), other pericardial investigators have filled their balloons with air (3, 11, 24), following the precedent of those who used balloons to measure pleural or esophageal pressure (13, 19). Any fluid (air, water, oil, etc.) can be used, but, because of compressibility, air-filled balloons perform somewhat less well than water-filled balloons in terms of nonuniform loading and, because of viscosity, oil-filled balloons perform less well in terms of frequency response (data not shown). However, as discussed in detail in APPENDIX B, liquids with bloodlike specific gravities are very advantageous because they allow accurate assessments of transmural pressure, regardless of the position of the balloon.
Conventional static calibration techniques (18) load the balloon with a uniform stress distribution. We developed a more rigorous test in which loading was applied nonuniformly, probably much less uniformly than could occur within the pericardium. We applied equal, known stresses over varying fractions of the transducing surface of the balloon. We reasoned that, if a given stress were applied over half the surface, for example, the proper response would be one-half the pressure that would be registered if that same stress had been applied uniformly over the entire surface. We found that the balloon performed this spatial integration very well, probably because of its construction. The loaded rectangular block would have caused Laplacian stresses to develop in both surfaces. If those stresses had not been opposed, a pressure less than the predicted pressure would have been recorded from the balloon. Because the measured pressure was equal to the predicted pressure, those Laplacian stresses must have been opposed by the rigidity of the borders of the balloon, which caused pressure to increase to the predicted level. [The borders of a pericardial balloon must be rigid enough so they resist the buckling that would tend to occur because of the tension in the Silastic membranes created by nonuniform loads, yet be thin enough to be flexible (1, 18).]
It has been suggested that, because of the curvature of the LV surface, the balloon must record an artifactually high pressure, even though it had been calibrated to be accurate when flat (21). The surface of the LV is hemiellipsoidal and, therefore, curved in two directions. We made several attempts to create a test chamber with an elliptical surface. However, we abandoned the effort because we were unable to specify the applied stress unequivocally, because of the complexity of the stresses. We then chose an alternative strategy: to demonstrate that the balloon performed accurately in a cylindrical calibrator with a relatively extreme (albeit one-dimensional) curvature in which the imposed stresses could be defined (16). We observed that the balloon was accurate even in the cylinder with the smallest radius of curvature; 15.5 mm is less than either the short- or long-axis radius of curvature of the LV free wall of a 20-kg dog and much less than those dimensions in the adult human. Thus we conclude that there is no significant error due to curvature when the balloon transducer is used to measure pericardial pressure over the LV free wall.
Predictably, we found that the frequency response of the standard
pericardial balloon was not better than that of the connecting 7-Fr
catheter. Figure 7 indicates that the 20-Hz component of the pressure
waveform was approximately doubled in magnitude. We also evaluated the
output of the high-fidelity balloon to assess the frequency response
required to measure the fastest events in the cardiac cycle (i.e., the
early systolic peak approximately coinciding with the c
wave). Using low-pass digital filters with sharp cutoffs, we filtered
the pericardial pressure recorded from a dog with a heart rate of 94 min
1 and evaluated the attenuation of the peak. Excluding
frequencies above 40, 30, 20, and 10 Hz (equivalent to excluding
harmonics greater than ~25, 19, 13, and 6) resulted in respective
attenuations of 2, 7, 24, and 44% (see Fig.
8). Because pericardial pressure does not
change rapidly at end diastole, that pressure was not changed by even
10-Hz filtering and, thus, we conclude that the frequency response of
the standard pericardial balloon transducer is sufficient to measure
end-diastolic transmural pressure accurately. However, when faster
events are to be studied, the high-fidelity adaptation of the
transducer is satisfactory and convenient.
|
In addition to allowing the study of faster events, the insertion of a micromanometer-tipped catheter into the balloon minimizes motion-induced artifacts (i.e., "catheter-whip"). Although the connecting catheter might still generate such artifacts, the micromanometer in the balloon will not detect them because the mechanical impedance of the balloon is very low. We were able to demonstrate that "artifacts" generated in the catheter by tapping, which were conspicuous when measured using the external transducer, were undetectable using the micromanometer-tipped catheter. The problem of micromanometer baseline instability can be overcome by recording pressure through the lumen and adjusting the mean values to be equal either manually, before the data are recorded, or retrospectively, during analysis, as we did.
In conclusion, we have performed static and dynamic calibrations of a Silastic, liquid-filled, pericardial balloon transducer. Static performance was satisfactory, even when loading was markedly nonuniform or when the balloon was curved. Dynamic performance was limited only by the transmission characteristics of the connecting catheter. When higher frequency response is required or motion-induced artifacts are to be minimized, a convenient high-fidelity modification is available.
| |
APPENDIX A. DETERMINATION OF BALLOON VOLUME |
|---|
|
|
|---|
Recently boiled room-temperature water is introduced into
the balloon, and all air bubbles are purged. The balloon is positioned in the calibration chamber, and chamber pressure is increased to 5 mmHg
to expel the free air from the space between the membrane and the end
plate. The calibration chamber pressure is then returned to atmospheric
pressure. Zero balloon volume is established when liquid is withdrawn
until the balloon surfaces form a dimple over the side ports of the
internal tubing (suction of approximately
5 mmHg is required for this
to occur). The chamber pressure is increased to ~30 mmHg, and the
volume of liquid introduced into the balloon is adjusted until the
balloon pressure equals the chamber pressure within ±0.5 mmHg. The
balloon volume is fixed at the value determined by the previous step,
and balloon error is measured over chamber pressures ranging from
40
to +50 mmHg. If the balloon error is found to be >1 mmHg, the balloon
is considered flawed and is discarded.
| |
APPENDIX B. COMPARISON OF AIR- AND LIQUID-FILLED BALLOONS |
|---|
|
|
|---|
Assuming that blood has a specific gravity of 1 gm/cm3, that the hydrostatic equivalent of LV diameter is
equal to 4 mmHg (5.44 cmH2O = 4 mmHg) and that the
true transmural LV end-diastolic pressure is 2 mmHg at every point on
its circumference, Fig. 9 shows that air-
and liquid-filled balloons yield different values of transmural
pressure, depending on the position of the balloon. If absolute LV
intracavitary pressure is assumed to be 10 mmHg at the mid-LV reference
plane, the absolute pressure must be 8 mmHg at the top of the cavity
and 12 mmHg at the bottom, because of the weight of the blood. An
air-filled balloon at the top of the LV (A in Fig. 9) would
record 6 (8
2) mmHg, at the midplane (B) 8 (10
2) mmHg, and at the bottom (C) 10 (12
2) mmHg. On the other hand, a liquid-filled balloon would record 8 mmHg at every
location. This is because the siphon effect of the column of water
contained in the connecting catheter automatically adds or subtracts an
appropriate hydrostatic correction after the transducer has been zeroed
at the mid-LV reference plane. Because it is difficult to ascertain the
position of the balloon precisely and at all times, it is difficult to
correct the measurements from an air-filled balloon and, thus, to
achieve accurate estimations of transmural pressure at any time in the
cardiac cycle.
|
| |
ACKNOWLEDGEMENTS |
|---|
We thank Cheryl Meek, Rozsa Sas, and Gerald Groves for technical assistance and Dr. I. Belenkie for criticism. Dr. Jiun-Jr Wang provided frequency-response analysis of pericardial pressure.
| |
FOOTNOTES |
|---|
The study was supported by operating grants from the Alberta Heart and Stroke Foundation (Calgary, Alberta) to J. V. Tyberg. J. V. Tyberg is a Medical Scientist of the Alberta Heritage Foundation for Medical Research (Edmonton, Alberta), which also supported the study with an Establishment Grant.
Address for reprint requests and other correspondence: J. V. Tyberg, Medicine and Physiology/Biophysics, Univ. of Calgary Health Sciences Centre, 3330 Hospital Dr. NW, Calgary, Alberta, Canada T2N 4N1 (E-mail: jtyberg{at}ucalgary.ca).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 18 February 2000; accepted in final form 22 November 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Agostoni, E.
Mechanics of the pleural space.
Physiol Rev
52:
57-128,
1972
2.
Applegate, RJ,
Santamore WP,
Klopfenstein HS,
and
Little WC.
External pressure of undisturbed left ventricle.
Am J Physiol Heart Circ Physiol
258:
H1079-H1086,
1990
3.
Assanelli, D,
Lew WYW,
Shabetai R,
and
LeWinter MM.
Influence of the pericardium on right and left ventricular filling in the dog.
J Appl Physiol
63:
1025-1032,
1987
4.
Banchero, N,
Schwartz PE,
Tsakiris AG,
and
Wood EH.
Pleural and esophageal pressures in the upright body position.
J Appl Physiol
23:
228-234,
1967
5.
Barnard, HL.
The functions of the pericardium.
J Physiol (Lond)
22:
43-47,
1898.
6.
Belenkie, I,
Dani R,
Smith ER,
and
Tyberg JV.
Ventricular interaction during experimental acute pulmonary embolism.
Circulation
78:
761-768,
1988[ISI][Medline].
7.
Boltwood, CM,
Skulsky A,
Drinkwater DC,
Lang S,
Mulder DG,
and
Shah PM.
Intraoperative measurement of pericardial constraint: role in ventricular diastolic mechanics.
J Am Coll Cardiol
8:
1289-1297,
1986[Abstract].
8.
Boyd, IA,
and
Pathak CL.
The comparative toxicity of silicone rubber and plastic tubing.
Scott Med J
9:
345-351,
1964[Medline].
9.
Freeman, GL,
and
LeWinter MM.
Determinants of intrapericardial pressure in dogs.
J Appl Physiol
60:
758-764,
1986
10.
Hamilton, DR,
Dani RS,
Semlacher RA,
Smith ER,
Kieser TM,
and
Tyberg JV.
Right atrial and right ventricular transmural pressures in dogs and humans. Effects of the pericardium.
Circulation
90:
2492-2500,
1994[ISI][Medline].
11.
Hoit, BD,
Dalton N,
Bhargava V,
and
Shabetai R.
Pericardial influences on right and left ventricular filling dynamics.
Circ Res
68:
197-208,
1991[Abstract].
12.
Holt, JP,
Rhode EA,
and
Kines H.
Pericardial and ventricular pressure.
Circ Res
8:
1171-1180,
1960[Abstract].
13.
Hoppin, FG,
Green ID,
and
Mead J.
Distribution of pleural surface pressure in dogs.
J Appl Physiol
27:
863-873,
1969
14.
Kenner, HM,
and
Wood EH.
Intrapericardial, intrapleural, and intracardiac pressures during acute heart failure in dogs studied without thoracotomy.
Circ Res
19:
1071-1079,
1966[Abstract].
15.
Kingma, I,
Smiseth OA,
Frais MA,
Smith ER,
and
Tyberg JV.
Left ventricular external constraint: relationship between pericardial, pleural and esophageal pressures during positive end-expiratory pressure and volume loading in dogs.
Ann Biomed Eng
15:
331-346,
1987[ISI][Medline].
16.
Lai-Fook, SJ,
Price DC,
and
Staub NC.
Liquid thickness vs. vertical pressure gradient in a model of the pleural space.
J Appl Physiol
62:
1747-1754,
1987
17.
Mann, D,
Lew W,
Ban-Hayashi E,
Shabetai R,
Waldman L,
and
LeWinter MM.
In vivo mechanical behavior of canine pericardium.
Am J Physiol Heart Circ Physiol
251:
H349-H356,
1986.
18.
McMahon, SM,
Permutt S,
and
Proctor DF.
A model to evaluate pleural surface pressure measuring devices.
J Appl Physiol
27:
886-891,
1969
19.
Milic-Emili, J,
Mead J,
Turner JM,
and
Glauser EM.
Improved technique for estimating pleural pressure from esophageal balloons.
J Appl Physiol
19:
207-211,
1964
20.
Morgan, BC,
Guntheroth WG,
and
Dillard DH.
Relationship of pericardial to pleural pressure during quiet respiration and cardiac tamponade.
Circ Res
16:
493-498,
1965[Abstract].
21.
Nakamoto, T,
Li K,
Santamore WP,
and
Johnston WE.
Inherent sensitivity of pericardial balloon catheters to surface curvature (Abstract).
FASEB J
3:
A1174,
1989.
22.
Scott-Douglas, NW,
Traboulsi M,
Smith ER,
and
Tyberg JV.
Experimental instrumentation and left ventricular pressure-strain relationship.
Am J Physiol Heart Circ Physiol
261:
H1693-H1697,
1991
23.
Shabetai, R.
Measuring pericardial constraint.
J Am Coll Cardiol
7:
315-316,
1986[ISI][Medline].
24.
Shabetai, R.
Pericardial and cardiac pressure.
Circulation
77:
1-5,
1988[ISI][Medline].
25.
Smiseth, OA,
Frais MA,
Kingma I,
Smith ER,
and
Tyberg JV.
Assessment of pericardial constraint in dogs.
Circulation
71:
158-164,
1985[ISI][Medline].
26.
Smiseth, OA,
Refsum H,
and
Tyberg JV.
Pericardial pressure assessed by right atrial pressure: a basis for calculation of left ventricular transmural pressure.
Am Heart J
108:
603-605,
1983.
27.
Takata, M,
Mitzner W,
and
Robotham JL.
Influence of the pericardium on ventricular loading during respiration.
J Appl Physiol
68:
1640-1650,
1990
28.
Tyberg, JV,
Misbach GA,
Glantz SA,
Moores WY,
and
Parmley WW.
A mechanism for the shifts in the diastolic, left ventricular, pressure-volume curve: the role of the pericardium.
Eur J Cardiol
Suppl 7:
163-175,
1978.
29.
Tyberg, JV,
Taichman GC,
Smith ER,
Douglas NWS,
Smiseth OA,
and
Keon WJ.
The relation between pericardial pressure and right atrial pressure: an intraoperative study.
Circulation
73:
428-432,
1986[ISI][Medline].
30.
Tyson, GSJ,
Maier GW,
Olsen CO,
Davis JW,
and
Rankin JS.
Pericardial influences on ventricular filling in the conscious dog: an analysis based on pericardial pressure.
Circ Res
54:
173-184,
1984[Abstract].
31.
Wiener-Kronish, JP,
Gropper MA,
and
Lai-Fook SJ.
Pleural liquid pressure in dogs measured using a rib capsule.
J Appl Physiol
59:
597-602,
1985
This article has been cited by other articles:
![]() |
T. N. Hobson, J. A. Flewitt, I. Belenkie, and J. V. Tyberg Wave intensity analysis of left atrial mechanics and energetics in anesthetized dogs Am J Physiol Heart Circ Physiol, March 1, 2007; 292(3): H1533 - H1540. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Gibbons Kroeker, S. Adeeb, J. V. Tyberg, and N. G. Shrive A 2D FE model of the heart demonstrates the role of the pericardium in ventricular deformation Am J Physiol Heart Circ Physiol, November 1, 2006; 291(5): H2229 - H2236. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Mitchell, W. A. Whitelaw, R. Sas, E. R. Smith, J. V. Tyberg, and I. Belenkie RV filling modulates LV function by direct ventricular interaction during mechanical ventilation Am J Physiol Heart Circ Physiol, August 1, 2005; 289(2): H549 - H557. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Wang, F. Jalali, Y.-H. Sun, J.-J. Wang, K. H. Parker, and J. V. Tyberg Assessment of left ventricular diastolic suction in dogs using wave-intensity analysis Am J Physiol Heart Circ Physiol, April 1, 2005; 288(4): H1641 - H1651. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Hamilton, R. Sas, and J. V. Tyberg Atrioventricular nonuniformity of pericardial constraint Am J Physiol Heart Circ Physiol, October 1, 2004; 287(4): H1700 - H1704. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. H. Hollander, G. M. Dobson, J.-J. Wang, K. H. Parker, and J. V. Tyberg Direct and series transmission of left atrial pressure perturbations to the pulmonary artery: a study using wave-intensity analysis Am J Physiol Heart Circ Physiol, January 1, 2004; 286(1): H267 - H275. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. G. Kroeker, N. G. Shrive, I. Belenkie, and J. V. Tyberg Pericardium modulates left and right ventricular stroke volumes to compensate for sudden changes in atrial volume Am J Physiol Heart Circ Physiol, June 1, 2003; 284(6): H2247 - H2254. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. D. Moore, M. P. Frenneaux, R. Sas, J. J. Atherton, J. A. Morris-Thurgood, E. R. Smith, J. V. Tyberg, and I. Belenkie Ventricular interaction and external constraint account for decreased stroke work during volume loading in CHF Am J Physiol Heart Circ Physiol, December 1, 2001; 281(6): H2385 - H2391. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. deVries, D. R. Hamilton, H. E. D. J. Ter Keurs, R. Beyar, and J. V. Tyberg A novel technique for measurement of pericardial pressure Am J Physiol Heart Circ Physiol, June 1, 2001; 280(6): H2815 - H2822. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |