Vol. 90, Issue 3, 1025-1030, March 2001
Mathematical assessment of qualitative diagnostic calibration
for respiratory inductive plethysmography
Anne
De Groote,
Manuel
Paiva, and
Yves
Verbandt
Biomedical Physics Laboratory, Université Libre de Bruxelles,
1070 Brussels, Belgium
 |
ABSTRACT |
We present a critical
assessment of qualitative diagnostic calibration (QDC), which claims to
provide a relative calibration of respiratory inductive plethysmography
during natural breathing (Sackner MA, Watson H, Belsito AS, Feinerman
D, Suarez M, Gonzalez G, Bizousky F, and Krieger B. J Appl
Physiol 66: 410-420, 1989). QDC computes the calibration
factor (K) by considering breaths of constant tidal volume
(VT) and provides a criterion to select breaths when
VT is unknown. We applied QDC on uncalibrated data constructed from simulated sets of thoracic and abdominal volumes, with
a predefined K. As expected, QDC yields a correct
K when applied to breaths at constant VT. In
breathing at quasi-constant VT, the criterion for breath
selection is shown to bias the results toward K = 1. For spontaneous breathing, the calculated K deviates from
its predefined value and depends heavily on the selection criterion. We
conclude that QDC will only provide a correct calibration factor when
applied to an entire set of breaths with constant or quasi-constant
VT. More generally, physiological conclusions based on QDC
should be critically evaluated on a case-by-case basis.
thoracoabdominal partitioning; simulation; Konno-Mead model
 |
INTRODUCTION |
SINCE THE WORK OF
Konno and Mead (6) modeling the chest wall movements with
two degrees of freedom and introducing the isovolume maneuver, other
methods have been proposed to calculate tidal volume (VT)
from measurements of the movements of abdomen and rib cage. They are
mostly based on a simultaneous recording of the breathing volume by a
spirometer or a pneumotachograph (4, 5, 7, 9, 12). More
recently, Sackner et al. (10) described the qualitative
diagnostic calibration method (QDC), which is based on spontaneous
breathing, and Banzett et al. (2) used standard
calibration ratios to determine the appropriate gains for the rib cage
and abdominal signals. These calibration methods differ in the
information they provide: an absolute calibration against the volume
(4, 5, 7, 9, 12) or a relative weighting of the thoracic
and abdominal signals (2, 6, 10). Furthermore, the
complexity of their implementation varies significantly from the
straightforward use of standard gains (2) to the isovolume maneuver requiring active cooperation of the subject (6).
From this practical point of view, QDC is very attractive because it is
a single-posture method that is carried out during spontaneous breathing without need for a mouthpiece or a facemask. However, this
method has been subjected to criticism: Sartene et al.
(11) proposed a more rigorous formulation of the QDC
coefficient, Thompson (13) concluded from mathematical
considerations that QDC is not reliable, and Brown et al.
(3) indicated that "a critical evaluation of the
mathematics underlying the QDC method, including computer simulations
of the potential errors that could arise under circumstances that
violate the underlying assumptions, would be extremely valuable."
In this paper, we analyze the principles and the mathematics underlying
the QDC method, we test the approximations proposed by its authors to
fulfill its implicit assumptions, and we evaluate its limits of application.
 |
QDC THEORY |
We will briefly recall the theory underlying QDC
(10). The two-compartment model of the respiratory system
of Konno and Mead (6) expresses the change of volume
measured at the mouth (
Va) as the sum of a rib cage (
Vrc) and an
abdominal (
Vab) contribution
|
(1)
|
These volume changes are obtained indirectly by measuring the
variations of two representative thoracic and abdominal dimensions by
inductive plethysmography, magnetometers, or strain gauges. Using the
notations of Sackner et al. (10), Eq. 1 can be
rewritten as
|
(2)
|
where the calibration factor K weights the relative
contribution of the uncalibrated (u) electrical signals
uVrc and
uVab from the thoracic and abdominal compartments and M
scales the sum (K
uVrc +
uVab) to
Va. The QDC
as well as the isovolume calibration are methods yielding K.
The isovolume calibration consists of having the subject shift volume
between the thoracic and abdominal compartments while the upper airways
are occluded. Hence,
Va = 0 and Eq. 2
becomes
|
(3)
|
The QDC method (10) works at constant VT
instead of
Va = 0. It is based on a claimed observation that
the breath-to-breath variations of
Vrc and
Vab are normally
distributed (10). With breathing at constant
VT, calculating the standard deviation (SD) of each term of
Eq. 2 and using SD(VT) = 0 would yield
according to Sackner et al. (10)
|
(4)
|
Because, during natural spontaneous respiration, breathing at
constant VT is not possible, Sackner et al. propose to
approximate a constant VT by collecting a large number of
breaths and by excluding those whose sums show large deviations from
the mean sum µ(
uVrc +
uVab). The range to be considered is
thus
|
(5)
|
where x defines the interval width around the mean.
According to Sackner et al., this chosen parameter should be between 0.6 and 1.
 |
METHODS AND DATA |
Our approach is based mainly on simulated data following
Eq. 1 and consists of two parts. First, we analyze the
mathematics underlying the QDC method. Second, we study in some detail
the approximation of Eq. 5, which selects the breaths at
constant VT. For this analysis, we start from simulated
sets of thoracic and abdominal volumes and construct uncalibrated
signals using a predefined value of K. The procedure is illustrated in
Fig. 1: the sets of thoracic and
abdominal volumes, {
Vrci} and
{
Vabi}, are obtained from a
two-dimensional normal distribution, with major axis at angle
with
respect to the abscissa and centered around the mean (
Vrc,
Vab). [
Vrci and
Vabi correspond to the ith breath
(i = 1, ... , n)]. The sets are then
divided by two predefined calibration factors Krc and
Kab, which yield {
uVrci} and
{
uVabi}. In this operation,
Krc/Kab represents the calibration value
K that should be found by the QDC procedure (Kab
is equal to M in Eq. 2). The simulated data are
illustrated in Fig. 2
and Fig. 3,
in terms of distributions and Konno-Mead plot (abdominal
vs. rib cage), respectively, for three different cases corresponding to
breathing at constant VT (the ideal case of QDC theory,
Figs. 2A and 3A), breathing at quasi-constant
VT (Figs. 2B and 3B), and spontaneous
breathing (Figs. 2C and 3C). The constant and quasi-constant breathings are characterized, respectively, by no or
minor variations in VT. In these cases,
necessarily
equals
45°; i.e., the points (
Vrci,
Vabi) are aligned along an isovolume line
(Fig. 3A) or are symmetrically distributed between two
closely spaced isovolume lines (Fig. 3B). There is a
negative correlation between the thoracic and abdominal volumes and a
highly variable thoracoabdominal partitioning, which reflects the
variable breathing strategy used to achieve the constant or quasi-constant VT. The spontaneous breathing is defined by
minor changes in the relative abdominal contribution [
=
Vab/(
Vab +
Vrc)] and simulates the breathing pattern observed
in normal adult subjects during quiet breathing (2, 11). A
mathematical calculation based on
= constant shows that the
points (
Vrci,
Vabi)
are positively correlated and define a general orientation (tan
)
equal to
/(1
) (Fig. 3C). In this
simulation, larger variations are observed for VT even
though the thoracic and abdominal volumes show the same amplitude of
variations as in the previous cases. The sets of parameters [
,
SD(
Vrc), SD(
Vab)] used to construct each simulation are defined
in Table 1. In these examples, we assumed
a mean VT of 0.75 liters and a mean relative abdominal
contribution
of 1/3 (8). Krc and
Kab were chosen at 0.7 and 1, respectively. This choice
corresponds to M = 1 and K = 0.7. We
considered sets of 150 simulated breaths corresponding to the 3- to
10-min calibration period recommended by Sackner et al.
(10).

View larger version (16K):
[in this window]
[in a new window]
|
Fig. 1.
Construction of the simulated breathing data: the
pairs of thoracic and abdominal volumes
( Vrci, Vabi) are
obtained from a two-dimensional normal distribution, centered around
the mean [µ( Vrc),µ( Vab)], with major axis at angle with
respect to the abscissa. The subscript i indicates the
breath number in the data set. The sets are divided by two predefined
calibration factors Krc and Kab, which yield the
uncalibrated thoracic and abdominal volumes
( uVrci, uVabi). In
this operation, Krc/Kab represents the
calibration value K.
|
|

View larger version (24K):
[in this window]
[in a new window]
|
Fig. 2.
Thoracic and abdominal volumes (Vrc and Vab,
respectively) generated by random sampling from a two-dimensional
normal distribution (see Fig. 1) in 3 different cases. A:
simulation of constant tidal volume (VT). B:
simulation of quasi-constant VT. C: simulation
of spontaneous tidal breathing. Top: calibrated volumes
( V). Bottom: uncalibrated volumes ( uV). We assumed for
all simulations a mean VT of 0.75 liters and a mean
abdominal relative contribution of 1/3. The sets of parameters used to
construct the simulations are given in Table 1. Krc and
Kab were chosen at 0.7 and 1, respectively. Every set
contains 150 data points.
|
|

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 3.
Calibrated V ( ) and uncalibrated uV
(circles) thoracic and abdominal volumes of Fig. 2, plotted in a
Konno-Mead representation. A: simulation of constant tidal
volume VT. B: simulation of quasi-constant
VT. C: simulation of spontaneous tidal
breathing. The boundary lines with a slope of ( 1) delimit the
selected breaths ( ) corresponding to the interval
[µ( uVrc + uVab) ± 0.6 SD( uVrc + uVab)].
, Breaths outside the selection interval. Values of the
different slopes are indicated by . See DISCUSSION for
details.
|
|
 |
RESULTS |
Analysis of the mathematics of the QDC method.
It is shown in the APPENDIX that
|
(6)
|
is actually the solution of Eq. 2 when the pairs
(
uVrci,
uVabi)
belong to a straight line. Note that there is no minus sign in
Eq. 6. In this case, K is interpreted geometrically as the absolute value of the slope of the line defined by
the set of uncalibrated volumes (
uVrci,
uVabi). If the pairs
(
uVrci,
uVabi) do
not belong to a straight line but are strongly correlated, K
represents an approximation of the absolute value of the slope of the
regression line calculated through the set
(
uVrci,
uVabi). This approximation is all the better when the points are strongly correlated.
Approximation of Eq. 5 for simulated breathing at constant
VT.
Figure 2A shows the distributions of thoracic, abdominal and
volumetric data for the simulation, which corresponds to a respiration pattern at perfectly constant volume. The distribution of
{VTi} is thus an infinitesimally
narrow peak, generally referred to as a Dirac distribution. Note
that the set of uncalibrated sum {
uVrci +
uVabi} does not follow a Dirac
distribution. Figure 3A shows the calibrated and
uncalibrated data in a Konno-Mead plot. Because VT is
constant, these data define two straight lines with slopes equal to
1
and
0.7, respectively (Eqs. 1 and 2). The
selection of breaths according to x = 0.6 is shown in
Fig. 3A by the two dotted lines of slope
1. The proportion
of breaths that are eliminated corresponds to (1
p)
where p is the probability of being inside the interval
(µ
x SD, µ + x SD).
For x = 0.6, p equals 0.45. Hence, more than
half of the total number of breaths (55%) are eliminated although they
are, by construction, at constant VT. Nevertheless, the
points that remain define always the same straight line with the slope
of
0.7. This slope thus provides a correct value for K in
the particular case x = 0.6. Figure
4 shows K as a function of the
selection of breaths defined by x. This curve was obtained
by repeating the simulation procedure 100 times and averaging the
results. When breathing is at constant VT, the ratio of SD
always provides a correct value for K, regardless the value
of x, and thus a fortiori in the range 0.6
x
1.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 4.
The calibration factor K as a function of
x that determines the width of the selection interval
(Eq. 5) for the 3 examples of Figs. 2 and 3, i.e., at
constant VT, at quasi-constant VT, and for
spontaneous (irregular) breathing. K = 0.7 is the
correct value, by construction of the data. Each curve was obtained by
averaging 100 simulations.
|
|
Approximation of Eq. 5 for simulated breathing at quasi-constant
VT.
In the second simulation, some dispersion is added to VT
with respect to the ideal case of Fig. 2A. Figure
2B shows the distributions of the simulated data for this
case, and Fig. 3B shows the corresponding Konno-Mead plot.
It can be seen that {VTi} is no
longer a Dirac distribution. Furthermore, the points are not perfectly aligned along a straight line of slope
1. Again, the selection of
breaths according to the criterion of Sackner (Eq. 5 with
x = 0.6) yields many breaths (55%) at quasi-constant
VT that are eliminated whereas breaths corresponding to a
slope of
1 remain. This is highlighted in Fig. 3B by the
two dotted lines. According to the value of x and thus to
the selected breaths, K will take different numeric values,
which tend to 1 when x decreases (Fig. 4). In the present
simulation, a selection of breaths with 0.6
x
1 yields a corresponding value K ranging from 0.938 and 0.864 instead of 0.7. Figure 5 shows
K as a function of x for the four different cases
of Table 2, i.e., with decreasing
standard deviation of the rib cage and of the abdomen volume
variations. Here the same averaging procedure as in Fig. 4 was
performed. It can be seen that if the variability of the contributions
of thoracic and abdominal compartments decreases, the accuracy of K decreases for constant x.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 5.
The calibration factor K as a function of
x (Eq. 5) for a decreasing variability of the
thoracic and abdominal contributions in the model of quasi-constant
volume (Figs. 2B and 3B). K = 0.7 is the correct value. We assumed a mean VT of 0.75 liters
and a mean abdominal relative contribution of 1/3. The sets of
parameters used to construct the different simulations are defined in
Table 2. See Table 2 for definition of symbols. Each curve was obtained
by averaging 100 simulations.
|
|
Approximation of Eq. 5 for simulated spontaneous breathing.
In the third simulation, we consider spontaneous breathing (Figs.
2C and 3C). In that case, the slopes of the
calibrated and uncalibrated data in the Konno-Mead plot are positive.
The values of K obtained from QDC according to x
are shown in Fig. 4. Without any selection, K equals
[0.7 µ(
Vab)/µ(
Vrc)]. For 0.6
x
1, the values of K range from 0.492 to 0.42. K
tends to 1 when x decreases.
 |
DISCUSSION |
Underlying theory and basic assumptions.
In this paper, we use the two-compartment model of Konno and Mead
(Eq. 1), and we assume the validity of the observation of Sackner et al. (10) that
Vrc and
Vab are normally
distributed. This allowed us to focus on the mathematics underlying
QDC. The idea of working at constant VT by analogy with the
isovolume calibration is ingenious. We show in the APPENDIX
that the formula for K in the original paper
(10) is correct although the mathematical demonstration
can be criticized (3, 11, 13). In fact, we were intrigued
by the negative sign of Eq. 4 and by the simplicity of the
mathematical derivation of K by taking the SD of each term of Eq. 2 and stating that the SD of a sum of statistical
variables is equal to the sum of the SD of each variable. From a
physiological point of view, it is clear that K should be
positive. In fact, when the thoracic and abdominal compartments are
modeled by two cylinders and
uVab and
uVrc are considered as
variations of their cross-sectional areas, K represents the
ratio of the heights of each compartment, which are positive values. In
the APPENDIX, it is shown that the QDC-calculated
K value can geometrically be interpreted as the
approximation of the absolute slope of the regression line through the
set of uncalibrated thoracic and abdominal volumes plotted in the
Konno-Mead representation.
Breath selection for constant VT.
The critical point of the QDC procedure is the hypothesis of constant
VT in situations in which this parameter is unknown. Sackner et al. (10) proposed to collect the breaths that
fall in an interval around the mean µ(
uVrc +
uVab). We
demonstrated that this selection is not needed for breathing at
constant or quasi-constant VT (Figs. 3, A and
B, and 4). In these cases, the selection procedure even
biases the results by favoring in the Konno-Mead plot breaths that
correspond to a slope of
1 instead of breaths that are at constant or
quasi-constant VT. When the respiratory pattern is
spontaneous, K takes a value that depends on the selection
criterion and that tends to 1 with decreasing x. The latter
observation is due to the selection procedure, which favors breaths
falling on a line with a negative unitary slope. Hence, the breaths
selected define a specific constant volume VT =
uVrc +
uVab related to the situation of K = 1, which is not the actual value of K. Furthermore, it was
shown that if the variability of the contributions of thoracic and
abdominal compartments decreases, the accuracy of K
decreases for constant x. In fact, the selected points
become more focused and more easily define a straight line with a slope
corresponding to K = 1 instead of its real value.
Finally, the limitation of the method used to keep breaths at constant
VT is highlighted by the fact that multiplying the gain of
one channel (rib cage or abdomen) does not induce the same variation of
the QDC-calculated K. This observation is explained by the
fact that the modification of gain induces a rotation of the points
(
uVrci,
uVabi) in the Konno-Mead plot while the selection interval of slope (
1) remains identical. Hence,
this operation modifies the breaths selected and thus the calculation
of K.
Applications of QDC.
In spite of these serious reservations, QDC was demonstrated to yield,
in adults in supine posture, results equivalent to the other
calibration procedures, namely isovolume and multilinear regression
methods (10, 11). The best results were obtained when
estimating VT during uninstructed tidal breathing. Other situations, such as natural preferential thoracic or abdominal breathing and voluntary changes of VT, yielded less
accurate VT estimations. These results can be explained by
the observation that the variability of thoracoabdominal partitioning
in normal adult subjects is very small during quiet breathing (2,
11). Hence the rib cage signal is proportional to the abdominal
one. In that case,
uVrc =
uVab and Eq. 2
becomes
|
(7)
|
It can be seen in Eq. 7, where
is a proportionality
constant, that a correct VT will be obtained with an
appropriate choice of M, regardless of the value of
K. This can explain why all the calibration techniques,
including the one using standard ratios (2), provide a
good evaluation of the VT for quiet breathing, whereas
greater discrepancies in the evaluation of VT are observed when changes in breathing pattern occur. In the latter case, two degrees of freedom are used and, hence, the accuracy of K
becomes more critical.
Adams et al. (1) demonstrated the satisfactory use of the
Respitrace, calibrated with QDC, for estimating VT in
healthy full-term newborns. The validation was achieved shortly after QDC, by comparing the mean volumes of sets of 10 breaths obtained from
calibrated signals with these measured by a pneumotachograph. The
relative contributions of rib cage and abdominal compartments to
VT were not examined, and isovolume-like situations
(obstructive apneas or simulated obstructive apneas) were not analyzed.
Hence, the accuracy of K was not studied independently from
the value of M. Furthermore, the values of K and
M were not indicated in the paper of Adams et al., which
prevents in-depth scrutiny of their results in the light of our analysis.
In contrast to the Adams et al. study (1), Brown et al.
(3) showed QDC to be unreliable for estimation of the
VT in anesthetized infants. To obtain sufficient accuracy
when evaluating VT by QDC, a minimum of 20 breaths needed
to be averaged in the analysis. In addition, the errors on the volume
reconstructed by QDC during imposed airway obstructions were very
large. Finally, the contribution of the rib cage to VT,
calculated by means of the QDC-calibration factor, was unexpectedly
high and inconsistent with published findings. The authors suggested
that, in anesthetized infants, the contributions of rib cage and
abdominal compartments during tidal breathing may not be sufficiently
variable to allow for the accurate derivation of K. However,
they could not verify this point due to the limitations of their
experimental setup. Our analysis of the model at quasi-constant
VT confirms this hypothesis by showing that a decreasing
variability in the contributions of thoracic and abdominal movements
can actually make the QDC method fail (Fig. 5).
In conclusion, QDC will only provide a correct calibration factor when
applied to an entire set of breaths with constant or quasi-constant
VT. More generally, physiological conclusions based on QDC
should be critically evaluated on a case-by-case basis.
 |
APPENDIX |
Demonstration of Eq. 4.
Using
Va = VT to consider tidal volumes, Eq. 2 can be rewritten as
|
(A1)
|
Because VT is constant in Sackner's method
(10), Eq. A1 represents a straight line with a
negative slope
K and intercept VT/M. In general, when pairs
(xi, yi) belong to a
straight line y = ax + b, the variance
(SD2) of the data sets X = {xi} and Y = {yi} are given by
|
(A2)
|
|
(A3)
|
where µ is the mean of the set
{xi} and n is the
number of data pairs (xi,
yi). Hence
|
(A4)
|
By analogy, in the case of Eq. A1
|
(A5)
|
and thus, as K is positive, is
|
(A6)
|
the solution of Eq. 2. (
K) is
interpreted geometrically as the slope of the line defined by the
uncalibrated thoracic and abdominal volumes plotted in a Konno-Mead representation.
If the pairs (xi,
yi) do not belong to a straight line
but are strongly correlated, a regression line can be calculated by a
least-square fitting. If y is considered as the dependent
variable, the slope is given by
|
(A7)
|
On the other hand, considering x as the dependent
variable yields
|
(A8)
|
Here CovXY is the covariance of
(xi,
yi) and r is the
correlation coefficient. By analogy
|
(A9)
|
and
|
(A10)
|
Equations A9 and A10 show that
(
K) calculated according to QDC is an approximation
of the slope of the regression lines calculated on the uncalibrated
thoracic and abdominal volumes plotted in the Konno-Mead
representation. The approximation is all the better when the points are
strongly negatively correlated, i.e., when r tends to
1.
 |
ACKNOWLEDGEMENTS |
We acknowledge Fernand Colin and Brigitte Dutrieue for helpful
suggestions on the manuscript.
 |
FOOTNOTES |
This study was supported by the Belgian Federal Office for Scientific
Affairs (PRODEX contract).
Address for reprint requests and other correspondence: A. De
Groote, Biomedical Physics Laboratory, cp 613/3, Route de Lennik, 808, Brussels, Belgium (E-mail: adegroot{at}ulb.ac.be).
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 17 July 2000; accepted in final form 25 September 2000.
 |
REFERENCES |
1.
Adams, JA,
Zabaleta IA,
Stroh D,
Johnson P,
and
Sackner MA.
Tidal volume measurements in newborns using respiratory inductive plethysmography.
Am Rev Respir Dis
148:
585-588,
1993[Medline].
2.
Banzett, RB,
Mahan ST,
Garner DM,
Brughera A,
and
Loring SH.
A simple and reliable method to calibrate respiratory magnetometers and Respitrace.
J Appl Physiol
79:
2169-2176,
1995[Abstract/Free Full Text].
3.
Brown, K,
Aun C,
Jackson E,
Mackersie A,
Hatch D,
and
Stocks J.
Validation of respiratory inductive plethysmography using the qualitative diagnostic calibration method in anaesthetized infants.
Eur Respir J
12:
935-943,
1998[Abstract].
4.
Chadha, TS,
Watson H,
Birch S,
Jenouri GA,
Schneider AW,
Cohn MA,
and
Sackner MA.
Validation of respiratory inductive plethysmography using different calibration procedures.
Am Rev Respir Dis
125:
644-649,
1982[ISI][Medline].
5.
Cohn, MA,
Rao ASV,
Broudy M,
Birch S,
Watson H,
Atkins N,
Davis B,
Stott FD,
and
Sackner MA.
The respiratory inductive plethysmograph: a new non-invasive monitor to respiration.
Bull Eur Physiopath Respir
18:
643-658,
1982[ISI][Medline].
6.
Konno, K,
and
Mead J.
Measurement of the separate volume changes of rib cage and abdomen during breathing.
J Appl Physiol
22:
407-422,
1967[Free Full Text].
7.
Loveridge, B,
West P,
Anthonisen NR,
and
Kryger MH.
Single-position calibration of the respiratory inductance plethysmograph.
J Appl Physiol
55:
1031-1034,
1983[Abstract/Free Full Text].
8.
Paiva, M,
Estenne M,
and
Engel LA.
Lung volumes, chest wall configuration and pattern of breathing in microgravity.
J Appl Physiol
67:
1542-1550,
1989[Abstract/Free Full Text].
9.
Revow, MD,
England SJ,
Stogryn HA,
and
Wilkes DL.
Comparison of calibration methods for respiratory inductive plethysmography in infants.
J Appl Physiol
63:
1853-1861,
1987[Abstract/Free Full Text].
10.
Sackner, MA,
Watson H,
Belsito AS,
Feinerman D,
Suarez M,
Gonzalez G,
Bizousky F,
and
Krieger B.
Calibration of respiratory inductive plethysmograph during natural breathing.
J Appl Physiol
66:
410-420,
1989[Abstract/Free Full Text].
11.
Sartene, R,
Dartus C,
Bernard JL,
Mathieu M,
and
Goldman MD.
Comparison of thoracoabdominal calibration methods in normal human subjects.
J Appl Physiol
75:
2142-2150,
1993[Abstract/Free Full Text].
12.
Stagg, D,
Goldman M,
and
Davis JN.
Computer-aided measurement of breath volume and time components using magnetometers.
J Appl Physiol
44:
623-633,
1978[Abstract/Free Full Text].
13.
Thompson, JR.
Qualitative diagnostic calibration technique.
J Appl Physiol
87:
869-872,
1999[Free Full Text].
J APPL PHYSIOL 90(3):1025-1030
8750-7587/01 $5.00
Copyright © 2001 the American Physiological Society