|
|
||||||||
Department of Anesthesiology and Intensive Care Medicine, University Hospital of Freiburg, Hugstetter Strasse 55, D-79106 Freiburg, Germany
| |
ABSTRACT |
|---|
|
|
|---|
A newly designed gas-sampling device using end-tidal CO2 to separate dead space gas from alveolar gas was evaluated in 12 mechanically ventilated patients. For that purpose, CO2-controlled sampling was compared with mixed expiratory sampling. Alveolar sampling valves were easily controlled via CO2 concentration. Concentrations of four volatile substances were determined in the expired and inspired gas. Isoflurane and isoprene, which did not occur in the inspired air, had ratios of end-tidal to mixed expired concentrations of 1.75 and 1.81, respectively. Acetone and pentane, found in both the inspired and expired air, showed ratios of 0.96 and 1.0, respectively. Precision of concentration measurements was between 2.4% (isoprene) and 11.2% (isoflurane); reproducibility (as coefficient of variation) was 5%. Because the only possible source of isoflurane and isoprene in this setting was patients' blood, selective enrichment of alveolar gas was demonstrated. By using the new sampling technique, sensitivity of breath analysis was nearly doubled.
acetone; breath analysis; isoflurane; isoprene; pentane
| |
INTRODUCTION |
|---|
|
|
|---|
CONCENTRATIONS OF
VOLATILE substances in blood are reflected in their
concentrations in the exhaled air because blood concentrations relate
to alveolar concentrations by their blood-gas partition coefficient or
solubility. Constituents of patients' exhaled gas include volatile
anesthetics, acetone, n-pentane, and isoprene. Volatile
anesthetics such as isoflurane are exhaled for a long period after
exposure during anesthesia (2, 3, 8). Acetone is formed by
decarboxylation of acetoacetate, pentane arises from peroxidation of
n-6 polyunsaturated fatty acids (6), and
isoprene (2-methylbutadiene-1,3) is thought to be formed along the
mevalonic pathway of cholesterol synthesis (26).
Exhalation of acetone, pentane, and isoprene has been related to
metabolic disorders, inflammatory reactions (7, 9, 10,
12-14, 16, 17, 25, 28) and certain diseases such as acute
respiratory distress syndrome and pneumonia (24). Results,
however, are conflicting, and a systematic and reliable correlation
between the composition of patients' exhaled air and their clinical
status has not yet been established. Substance concentrations of
interest fall in the range of 10
12 to 10
9
mol/l. Because of these extremely low concentrations, quantitative chemical analysis of mixed exhaled gas may be affected considerably by
errors due to dilution and contamination with dead space gas. In
mechanically ventilated patients, dead space even comprises parts of
the respiratory tubing that may contain additional contaminants. To
analyze the relationship between chemical composition of exhaled gas
and patients' clinical conditions, substance concentrations in
alveolar gas must be determined because only alveolar concentrations reflect concentrations in blood. We therefore developed a sampling method using the expired CO2 to separate dead space gas
from alveolar gas.
The purpose of this study was 1) to evaluate the CO2-controlled sampling method in mechanically ventilated patients and to test it with respect to efficacy of alveolar gas sampling and 2) to assess the effect of CO2-controlled sampling on the measurement of exhaled substance concentrations. For this purpose, four volatile substances of different origins were analyzed in the exhaled gas. Two of the volatile substances came predominantly from the patients' alveoli; the other two had measurable concentrations in the inspiratory gas. Results of CO2-controlled sampling were compared with those obtained by means of a mixed sampling method.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Patients. After approval by the local ethics committee and after informed consent by the patient or next of kin had been obtained, 12 patients (7 men, 5 women) were chosen randomly among the patients of our surgical intensive care unit. Ten patients had received isoflurane as an anesthetic. All patients were under controlled or assisted mechanical ventilation.
Gas sampling.
A schematic drawing of the CO2-controlled sampling
device is shown in Fig. 1. The sensor
(cuvette) of a fast-responding infrared absorption mainstream
CO2 analyzer (930, Siemens-Elema, Solna, Sweden; Fig. 1,
no. 6) was inserted between the Y piece of the respiratory circuit and
the patient. A stopcock (Fig. 1, no. 7) was mounted through a bore into
the cuvette to prevent leakage of respiratory gas when the sampling
system was not connected. The stopcock joined the cuvette with the
electrically operated two-way valve (LFYA, Lee) and had an internal
dead space of 88 µl (Fig. 1, no. 8). Adsorption traps
(27) containing 80 mg of activated charcoal (Analyt,
Müllheim, Germany) were mounted onto the two outlets of the
valve. By means of a Y piece, the traps were connected to a roller pump
(Model 7553-75, Cole-Parmer Instruments, Niles, IL) working at a
constant flow of 200 ml/min (Fig. 1, no. 14). Volatile substances were
collected and concentrated by adsorption onto the activated charcoal.
Approximately 1 liter of exhaled gas was drawn through the traps by
means of the roller pump. Sample volume and flow were measured by
collecting the gas under water and recording the time of sampling. The
preconcentrating effect of the charcoal was ~1,500-fold. One thousand
milliliters of exhaled gas passed through the trap. Desorption took
place in 2 s, and the gas flow through the trap during this period
was 20 ml/min, i.e., the substances adsorbed on the charcoal were
liberated into a volume of 2 s × 20 ml/ 60 s = 0.67 ml.
Thus the cumulative sample volume-to-measured sample volume ratio
was ~1,500/l. Precision and reproducibility of the preconcentration
process have been evaluated in several studies (15, 22,
27). Relative standard deviations were
10%, and correlation
coefficients for calibration curves were always found to be >0.95.
|
|
|
|
|
|
|
|
Gas chromatographic analysis. Substances were desorbed from the activated charcoal by means of microwave energy (22), separated by gas chromatography, detected by flame ionization detection, and identified by mass spectrometry. In this study, two substances coming predominantly from the patient [isoflurane (23) and isoprene (24)] and two substances having measurable concentrations in the inspiratory gas [acetone and pentane (24)] were investigated. Substance concentrations were obtained from calibration curves by using flame ionization detection data. Measurements were made at least in duplicate. Time of sampling, ventilator settings, and the patient's diagnoses were recorded. The analytical method has been described in more detail elsewhere (24).
Variation of the measured values, limits of detection, and precision. In a previous study, we performed an extensive testing of the device in the laboratory with respect to precision of CO2 measurement, speed, and reproducibility of valve switching. Furthermore, we evaluated the relationship between the response time of the intake valve and cyclical pressure changes within the ventilator tubing system. The response time of the valve did not change up to a clinical relevant pressure of 100 mbar. The switching time of the intake valve is 18 ms.
Reproducibility of CO2-controlled gas sampling was assessed by analyzing the CO2 concentrations corresponding to the switching points of the valve. The coefficient of variation for the recognition of the switching thresholds was
2.5%. To assess the
reproducibility of the analytical method, the coefficient of variation
for expiratory isoflurane concentrations was calculated. For this
purpose, samples were collected in quadruplicate on different traps
from three patients. The mean coefficient of variation of the
isoflurane concentration was 5% (23). Limits of detection
were estimated on the basis of the signal-to-noise ratios obtained with
standards containing the compounds of interest at low concentration
levels. Standard mixtures were prepared and analyzed as described by
Schubert et al. (24). The limit of detection was defined
as that concentration of an analyte that produced a signal three times
greater than the baseline noise.
Five replicate measurements were performed by use of multicomponent
standards of the same concentration. Precision was assessed by
calculating the mean, standard deviation, and relative standard deviation of the observed values.
Statistical analysis. Because substance concentrations were not normally distributed, the Friedman test followed by a Student-Newman-Keuls test was applied for multiple comparisons within patient groups. A P value <0.05 was considered significant. Results are given as medians and 25-75% percentiles. CO2 concentrations determined from sensor recordings were normally distributed, and results are given as means ± SD.
| |
RESULTS |
|---|
|
|
|---|
Patients' demographic data, main diagnoses, ventilator type, and
mode of mechanical ventilation are listed in Table
1. A reliable identification of the
alveolar phase was observed by simultaneously analyzing the
CO2 concentrations measured at the Y piece of the
respiratory circuit and the electrical switching impulses of the valve
(Figs. 2 and 4). The valve was switched into expiratory position at a
mean expiratory CO2 concentration of 3.3 ± 0.1%
(mean ± SD). Switching into the inspiratory position occurred
when CO2 concentration had fallen by 90 ± 3% of the
immediately preceding maximum. In patients 6, 7,
and 12, estimates of isoflurane arterial concentrations were
derived. Ratios of end-tidal to arterial CO2 partial
pressure were 0.9, 1.0, and 0.6, respectively, in these patients. With
standard conditions (volume = 22.4 l/mol) assumed, isoflurane
blood concentrations were determined as 42.8, 25.7, and 135 nl/ml
blood, respectively (blood-gas partition coefficient = 1.4).
|
The maximum time necessary to sample one liter of alveolar gas was 22 min in a patient who was tachypneic (27 breaths/min).
Tables 2 and
3
summarize the results obtained by means of mixed expiratory sampling
and CO2-controlled sampling. In all patients, isoflurane
and isoprene concentrations were higher in the
CO2-controlled samples than in the mixed expired samples
(Table 2). Median expired isoflurane and isoprene concentrations were 1.75 (isoflurane) and 2 (isoprene) times higher in the
CO2-controlled samples than in the mixed expired samples
(Table 2). Expired acetone and pentane concentrations were not
different in the CO2-controlled and the mixed expired
samples (Table 3). Inspired isoprene and isoflurane concentrations were
at the limit of detection. Inspired pentane and acetone concentrations
were at 10-50% of the expired concentrations.
|
|
Limits of detection for n-pentane, isoprene, acetone, and isoflurane were 0.10, 0.20, 0.15, and 0.95 nmol/l, respectively; precision (as relative SD) was 4.05, 2.4, 9.2, and 11.2%, respectively. Calibration plots of all substances were linear in the investigated range. Correlation coefficients were between 0.943 and 0.997. Relative standard deviations of retention times on the gas chromatography column were <0.3% (n = 5).
| |
DISCUSSION |
|---|
|
|
|---|
We found that alveolar gas sampling was reliable and precise with our newly designed CO2-triggered "alveolar" sampling valve. Alveolar sampling valves were easily controlled via CO2 concentration. Alveolar gas sampling started at 3.3 ± 0.1 vol% CO2 (25 ± 0.8 mmHg), which is in the range of mean ± 2 SD of the preset value of 3.5 vol%. By choosing a high value of 3.5% expired CO2 as the starting point of alveolar sampling, errors due to admixture of dead space gas are minimized. In patients showing a slow rise of expired CO2 concentration (e.g., in obstructive lung disease) or in patients who have low expired CO2 concentrations (e.g., during hyperventilation), sampling times are increased due to the late onset of alveolar sampling. If alveolar CO2 concentration is <3.5%, the device will not collect alveolar gas. In these cases, the threshold would have to be lowered by changing the internal settings of the prototype used for this study. Therefore, an adjustable onset of alveolar sampling is planned for future realizations of the apparatus.
Because no standard analytical method exists in breath analysis against which to compare new findings, it is impossible to compare "tested" values with "known" values. Internal standards can only be used in the analytical process, not in the patient, i.e., it is not possible to produce known substance concentrations in patients' exhaled air. Therefore, we tested the analytical process by using each patient as his or her own control, comparing measurements taken simultaneously with and without the new sampling device in the same patient.
Estimates of arterial tracer gas partial pressures may be derived from the ratio of end-tidal (alveolar) to arterial CO2 partial pressures. Arterial blood concentrations could then be derived from blood-gas partition coefficients or solubility. These estimates, however, will only be correct for substances that have a similar solubility to that of CO2 because end-tidal to arterial partial pressure ratios vary with ventilation and pulmonary perfusion according to the solubility of the volatile substances.
Linearity, precision, and limits of detection of the microwave desorption/flame ionization detection method had been evaluated before (15). In a previous laboratory investigation, we did not find any influence of the ventilatory pattern on the performance of our CO2-controlled valve. However, in the present study, the number of patients was too small to analyze the systematic effects of the disease state on the performance of our setup.
Although the analytical settings were identical, only two of the four volatile substances analyzed in this study were enriched by the CO2-controlled sampling technique. The different behavior of these exhaled substances is due to different substance origins. As one would expect, alveolar concentrations are higher than mixed expired concentrations only for those substances originating predominantly from the patient, such as isoflurane or isoprene. Isoflurane is exhaled after exposure during anesthesia and could not be detected in the inspiratory gas of the ventilator systems (23). Isoprene is most probably generated along the mevalonic pathway of cholesterol synthesis (26), and concentrations in the gas delivery system were negligible (24). Substances having considerable concentrations in the inspiratory limb, such as acetone or pentane, were not enriched by CO2-triggered sampling compared with mixed expiratory sampling. Both substances were found in varying concentrations in background air, ventilator tubing, and gas delivery systems (11, 18, 19, 20,). Dead space concentrations of these substances add to alveolar concentrations, resulting in mixed expiratory concentrations equal to or even higher than those measured by CO2-triggered sampling. As a consequence, the origin of a substance (coming from patient's blood or from the dead space or the delivery system) can be determined by comparing mixed expiratory to alveolar concentrations.
A sampling of the pure inspiratory gas was not possible by means of the valve used for this study. Because of the mode of operation of the valve, the trap mounted in the inspiratory position collected a mixture of inspired and dead space gas. Therefore, the inspiratory gas had to be sampled from the inspiratory part of the circuit without using the valve. In a previous study, we demonstrated that contamination with expiratory gas did not occur at this location in the respiratory tubing (23). High inspiratory pentane and acetone concentrations are due to the fact that these substances occur in large amounts in the ambient air (24). The separation of inspiratory from dead space gas, planned for future uses of the apparatus, could provide more precise information on substance origins and may be used for assessment of pulmonary gas-mixing properties.
During the last decade, numerous trials have been undertaken to establish systematic and reliable relationships between the chemical composition of (human) exhaled air and patients' clinical conditions (8, 12-14, 24, 25). Interpretation of these studies, however, remains difficult because results are conflicting and the variation of measured substance concentrations is considerable. Analyzing mixed expired air collected in gas-sampling bags, Foster et al. (5) observed an increase in isoprene elimination in humans 19 h after exposure to ozone. The underlying mechanisms may be activation of cholesterol synthesis at the onset of repair processes after oxidative damage to fluid linings in the lungs.
Mendis et al. (14) reported an increase in isoprene elimination in patients with an acute myocardial infarction. They proposed a relationship between isoprene elimination and activation of neutrophils. In this case, end-expiratory sampling was used. The decreasing isoprene elimination found in patients with acute respiratory distress syndrome and pneumonia (24) is in contrast to the work cited above. In this study, mixed expiratory samples were taken from the expiratory limb in the respiratory circuit of mechanically ventilated patients. Using a modified end-expiratory sampling method, Philips et al. (21) reported increased pentane and disulfide concentrations in patients with schizophrenia. Pentane is considered a marker of in vivo lipid peroxidation (10). However, concentrations in exhaled gas reported in the literature vary from 10 (1) to 200 pmol/l (4, 11) up to 1.5 nmol/l (24) without a known relationship to patients' specific clinical conditions. Because online control of efficacy of alveolar gas sampling was not previously possible, dead space may have significantly confounded the results and may be the reason for the difficulty in clinical correlation of expired trace gas analysis.
Besides this issue, there is another problem in breath analysis that has not yet been solved. To improve clinical interpretation of the results, the question of how concentrations in the exhaled air can be converted into blood or tissue concentration needs to be addressed. To solve this problem, investigations of substance concentrations in blood and exhaled air must be undertaken, preferably in the controllable setting of an animal model.
In conclusion, the CO2-triggered valve enables reliable sampling of alveolar gas in mechanically ventilated patients. A distinction of those substances generated from within the patient from those coming out of the delivery system is possible by comparing mixed expired to alveolar substance concentrations. By use of the CO2-controlled enrichment technique, accuracy of breath analysis is enhanced, and substances occurring in very low concentrations are made accessible to chemical analysis.
| |
ACKNOWLEDGEMENTS |
|---|
We thank Dr. H. Pahl for valuable advice during the preparation of this manuscript and T. Berger for excellent technical assistance.
| |
FOOTNOTES |
|---|
Presented in part at the Annual Meeting of German Anesthesiologists (DAK 97), Hamburg, Germany, April 25, 1997.
Address for reprint requests and other correspondence: J. K. Schubert, Dept. of Anesthesiology and Intensive Care Medicine, Univ. Hospital of Rostock, Schillingallee 35, D-18055 Rostock, Germany (E-mail: jochen.schubert{at}medizin.uni-rostock.de).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 20 April 1999; accepted in final form 28 August 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Cailleux, A,
and
Allain P.
Is pentane a normal constituent of human breath?
Free Radic Res Commun
18:
323-327,
1993[Web of Science][Medline].
2.
Eger, EI II.
The pharmacology of isoflurane.
Br J Anaesth
56:
71-99,
1984.
3.
Eger, EI II,
and
Bahlman SH.
Is the end-tidal anesthetic partial pressure an accurate measure of the arterial anesthetic partial pressure?
Anesthesiology
35:
301-303,
1971[Web of Science][Medline].
4.
Euler, DE,
Davé SJ,
and
Guo H.
Effect of smoking on pentane excretion in alveolar breath.
Clin Chem
42:
303-308,
1996
5.
Foster, MW,
Jiang L,
Stetkiewicz PT,
and
Risby TH.
Breath isoprene: temporal changes in respiratory output after exposure to ozone.
J Appl Physiol
80:
706-710,
1996
6.
Frankel, EN.
Volatile lipid oxidation products.
Prog Lipid Res
22:
1-33,
1982.
7.
Habib, MP,
Dickerson F,
and
Mooradian AD.
Ethane production rate in vivo is reduced with dietary restriction.
J Appl Physiol
68:
2588-2590,
1990
8.
Heusler, H.
Quantitative analysis of common anaesthetic agents.
J Chromatogr A
340:
273-319,
1985.
9.
Kazui, M,
Andreoni KA,
Norris EJ,
Klein AS,
Burdick JF,
Beattie C,
Sehnert SS,
Bell WR,
Bulkley GB,
and
Risby TH.
Breath ethane a specific indicator of free-radical-mediated lipid peroxidation following reperfusion of the ischemic liver.
Free Radic Biol Med
13:
509-515,
1992[Web of Science][Medline].
10.
Kneepkens, CMF,
Ferreira C,
Lepage C,
and
Roy CC.
The hydrocarbon breath test in the study of lipid peroxidation: principles and practice.
Clin Invest Med
15:
163-186,
1992[Web of Science][Medline].
11.
Kohlmüller, D,
and
Kochen W.
Is n-pentane really an index of lipid peroxidation in humans and animals? A methodological reevaluation.
Anal Biochem
210:
268-276,
1993[Web of Science][Medline].
12.
Kokoszka, J,
Nelson RL,
Swedler WI,
Skosey J,
and
Abcarian H.
Determination of inflammatory bowel disease activity by breath pentane analysis.
Dis Colon Rectum
36:
597-601,
1993[Web of Science][Medline].
13.
Letteron, P,
Duchatelle V,
Berson A,
Fromenty B,
Fisch C,
and
Degott C.
Increased ethane exhalation, an in vivo index of lipid peroxidation, in alcohol-abusers.
Gut
34:
409-414,
1993
14.
Mendis, S,
Sobotka PA,
and
Euler DE.
Expired hydrocarbons in patients with acute myocardial infarction.
Free Radic Res
23:
117-122,
1995[Web of Science][Medline].
15.
Miekisch, W,
Schubert JK,
Müller WPE,
and
Geiger K.
Advances in Critical Care Testing. Schaffhausen, Switzerland: AVL Medical Instruments, 1999, p. 202-211.
16.
Morita, S,
Snider MT,
and
Inada Y.
Increased n-pentane excretion in humans: a consequence of pulmonary oxygen exposure.
Anesthesiology
64:
730-733,
1986[Web of Science][Medline].
17.
Ondrula, D,
Nelson RL,
Andrianopoulos G,
Schwartz D,
Abcarian H,
Birnbaum A,
and
Skosey J.
Quantitative determination of pentane in exhaled air correlates with colonic inflammation in the rat colitis model.
Dis Colon Rectum
36:
457-462,
1993[Web of Science][Medline].
18.
Phillips, M.
Method for the collection and assay of volatile organic compounds in breath.
Anal Biochem
247:
272-278,
1997[Web of Science][Medline].
19.
Phillips, M,
and
Greenberg J.
Ion-trap detection of volatile organic compounds in alveolar breath.
Clin Chem
38:
60-65,
1992
20.
Phillips, M,
and
Greenberg J.
Method for the collection and analysis of volatile compounds in the breath.
J Chromatogr A
564:
242-249,
1991.
21.
Phillips, M,
Saba M,
and
Greenberg J.
Increased pentane and carbon disulfide in the breath of patients with schizophrenia.
Clin Pathol
46:
861-884,
1994.
22.
Rektorik, J.
Sample Introduction in Capillary Gas Chromatography. Heidelberg, Germany: Alfred Huethig, 1985, p. 217-233.
23.
Schubert, JK,
Esteban-Loos I,
Geiger K,
and
Guttmann J.
In-vivo evaluation of a new method for chemical analysis of volatile components in the respiratory gas of mechanically ventilated patients.
Technol Health Care
7:
29-37,
1999[Medline].
24.
Schubert, JK,
Müller WPE,
Benzing A,
and
Geiger K.
Application of a new method for analysis of exhaled gas in critically ill patients.
Intensive Care Med
24:
415-421,
1998[Web of Science][Medline].
25.
Sobotka, PA,
Brottman MD,
Weitz Z,
Birnbaum AJ,
Skosey JL,
and
Zarling EJ.
Elevated breath pentane in heart failure reduced by free radical scavenger.
Free Radic Biol Med
14:
643-647,
1993[Web of Science][Medline].
26.
Stone, BG,
Besse TJ,
Duane WC,
Evans CD,
and
DeMaster EG.
Effect of regulating cholesterol biosynthesis on breath isoprene excretion in men.
Lipids
28:
705-778,
1993[Web of Science][Medline].
27.
Trinh, VD,
and
Cong-Khanh H.
Graphitized carbon black in quartz tubes for the sampling of indoor air nicotine and analysis by microwave thermal desorption-capillary gas chromatography.
J Chromatogr Sci
29:
179-183,
1991[Web of Science][Medline].
28.
Van Gossum, A,
and
Decuyper J.
Breath alkanes as an index of lipid peroxidation.
Eur Respir J
2:
787-791,
1989[Abstract].
This article has been cited by other articles:
![]() |
T. Perl, E. Carstens, A. Hirn, M. Quintel, W. Vautz, J. Nolte, and M. Junger Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry Br. J. Anaesth., December 1, 2009; 103(6): 822 - 827. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Grossherr, A. Hengstenberg, T. Meier, L. Dibbelt, B. W. Igl, A. Ziegler, P. Schmucker, and H. Gehring Propofol concentration in exhaled air and arterial plasma in mechanically ventilated patients undergoing cardiac surgery Br. J. Anaesth., May 1, 2009; 102(5): 608 - 613. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Cope, M. T. Watson, W. M. Foster, S. S. Sehnert, and T. H. Risby Effects of ventilation on the collection of exhaled breath in humans J Appl Physiol, April 1, 2004; 96(4): 1371 - 1379. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. von Basum, H. Dahnke, D. Halmer, P. Hering, and M. Murtz Online recording of ethane traces in human breath via infrared laser spectroscopy J Appl Physiol, December 1, 2003; 95(6): 2583 - 2590. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |