The following is the abstract of the article discussed in the subsequent letter:
| |
ABSTRACT |
|---|
De Groote, Anne, Yves Verbandt, Manuel Paiva, and Pierre Mathys. Measurement of throacoabdominal asynchrony: importance of sensor
sensitivity to cross section deformations. J Appl
Physiol 88: 1295-1302, 2000.
Discrepancies in the assessment
of thoracoabdominal asynchrony are observed depending on the choice of
respiratory movement sensors. We test the hypothesis that these
discrepancies are due to a different dependence of the sensors on
cross-sectional perimeter and area variations of the chest wall. First,
we study the phase shift between perimeter and area
(
PA) for an elliptical model, which is
deformed by sinusoidal changes of its principle axes. We show that
perimeter and area vary sinusoidally in the physiological range of
deformations, and we discuss how
PA depends
on the ellipticity of the cross section, on the ratio of transverse and
dorsoventral movement amplitudes, and on their phase difference.
Second, we compute the relationship between perimeter, area, and the
output of the inductive sensor, and we proceed by comparing inductive
plethysmography with strain gauges for several cross section
deformations. We demonstrate that both sensors can provide different
phase information for identical cross section deformations and, hence,
can estimate thoracoabdominal asynchrony differently. Furthermore, the
complex dependence of the inductive sensor on perimeter and area warns against this sensor for the evaluation of thoracoabdominal asynchrony.
| |
LETTER |
|---|
Piezoelectric sensor vs. respiratory inductive plethysmograph
To the Editor: We read the paper by De Groote et al. (2) and found several errors of omission and commission. Thus they wrote in the introduction, "In these studies, strain gauges, inductive plethysmographs, magnetometers, and linear differential transducers are used without any detailed discussion of their characteristics." This statement contains errors of omission that should have been known to the authors or reviewers of the manuscript. Detailed discussion of validation of these sensors appeared in our review of noninvasive respiratory monitoring (4). Measurement of breath volumes among respiratory inductive plethysmograph, strain gauges, and impedance pneumograph, compared with the pneumotachograph, in babies revealed that the respiratory inductive plethysmograph was markedly superior to strain gauges and impedance pneumographs in estimating tidal volume in the prone posture. In this posture, waveforms were often absent from the strain gauges, because excursions were limited by compression of the baby's body and the transducer between the mattress, thereby giving spurious central apneas and false estimations of rib cage and abdominal excursions (1). In a study prompted by the work of Martinot-Lagarde et al. (3), bench testing of the respiratory inductive plethysmograph revealed that the inductive plethysmograph measures cross-sectional area (5). Stretching a standard inductive plethysmographic transducer band around wooden dowels, placed in holes on a pegboard, to form 23 curved and 5 rectangular shapes produced an output voltage that was linear for both curved and rectangular shapes and produced changes in cross-sectional area within the physiological range. Because De Groote et al. (2) indicated that their study was based on the work of Martinot-Lagarde et al. (3), they should have reconciled their mathematical model, using the approximations mentioned in their paper, with actual data from bench testing. Finally, the last error of omission relates to the piezoelectric strain gauges used as a standard of comparison to the inductive plethysmograph. We did not find "detailed discussion of their characteristics" (vide supra) in their present paper, references in that paper, or through a search of the relevant literature. Moreover, no validation of the piezoelectric sensor to a volumetric standard or description of its linearity and frequency response is provided. De Groote et al. (2) state that strain sensitivity of the piezoelectric film, combined with the belt elasticity, determines the perimeter measurement. Taking this statement at face value, validation of this transducer would be dependent on the linearity and frequency response for perimeter measurement of both the transducer and the belt. Alinearity in perimeter estimations or low-frequency response to measurement during breathing might distort the compartment waveforms and lead to errors in actual phase-angle measurements. Because "relatively stiff belts" surround the thorax and abdomen, how do the authors know that the piezoelectric sensor does not itself distort and restrict the compartments being tested for asynchrony? Finally, the authors did not consider, as a practical matter, the piezoelectric sensor estimation of phase angles during the shift from the supine to the prone posture.Although De Groote et al. (2) did provide a schematic of the inductive plethysmographic transducer in their Fig. 7 as the basis for the mathematical model, their single example of actual data from a 4-mo-old baby used an unvalidated example of this technology, i.e., "The inductive belts were incorporated in a pajama. ..." We believe that such an arrangement has the potential to distort the sinusoidal layout of the inductive plethysmographic transducer, thereby altering output linearity because of mutual coupling of self-inductance caused by a more square than sinusoidal shape of the coils of the stand-alone transducer. Of course, this situation may not have occurred; however, the authors should have been explicit on this point and have validated both the piezoelectric and inductive plethysmographic transducers applied to the body.
In conclusion, we do not believe that the theoretical evidence, unaccompanied by actual data, presented by De Groote et al. (2) justifies their warning against the use of the inductive plethysmographic sensor for the evaluation of thoracoabdominal asynchrony. Nor do we believe that the piezoelectric sensor was sufficiently tested as a reliable respiratory sensor to be utilized as a standard for comparison to inductive plethysmographic technology.
| |
ACKNOWLEDGEMENTS |
|---|
Conflict of interest disclosure: Marvin A. Sackner is Chief Executive Officer of Non-Invasive Monitoring Systems and owns 52% of its shares. No funding sources supported the writing of this letter.
| |
REFERENCES |
|---|
1.
Adams, JA,
Zabaleta IA,
Stroh D,
and
Sackner MA.
Measurement of breath amplitudes: comparison of three noninvasive respiratory monitors to integrated pneumotachograph.
Pediatr Pulmonol
16:
254-258,
1993[Web of Science][Medline].
2.
De Groote, A,
Verbandt Y,
Paiva M,
and
Mathys P.
Measurement of thoracoabdominal asynchrony: importance of sensor sensitivity to cross section deformations.
J Appl Physiol
88:
1295-1302,
2000
3.
Martinot-Lagarde, P,
Sartene R,
Mathieu M,
and
Durand G.
What does inductance plethysmography really measure?
J Appl Physiol
64:
1749-1756,
1988
4.
Sackner, MA,
and
Krieger BP.
Non-invasive respiratory monitoring.
In: Heart-Lung Interactions in Health and Disease, edited by Scharf SM,
and Cassidy SS.. New York: Marcel Dekker, 1989, p. 663-805.
5.
Watson, HL,
Poole DA,
and
Sackner MA.
Accuracy of respiratory inductive plethysmographic cross-sectional areas.
J Appl Physiol
65:
306-308,
1988
| Marvin A Sackner, | ||||||||||||
|
Jose A. Adams, Mt. Sinai Medical Center and Non-Invasive Monitoring Systems Miami Beach, FL 33139 E-mail: Artchive{at}msn.com |
To the Editor: Sackner and Adams express criticisms
that are the results of a misinterpretation of our paper (2). As
indicated in the title, we studied the importance of sensor choice for
the measurement of thoracoabdominal asynchrony, not for the evaluation of breath volume. Hence, the review of noninvasive respiratory monitoring, which is quoted by Sackner and Krieger (5), is irrelevant
to our study. Our mathematical model has been validated by bench
testing (1). In contrast to Watson et al. (7), we performed not only
static, but also dynamic, tests in a physiological range. Because the
analysis in our paper is based on a well-established theory (3, 4), we
did not find it necessary to include its experimental validation. If
Sackner and Adams had read our paper carefully, they would not have
stated that we utilized piezoelectric strain gauges as a standard of
comparison to inductive plethysmography. In the article, we consider
strain gauges in general, as a class of sensors designed to be
selectively sensitive to elongation. Specific piezoelectric strain
gauges are used only for the introduction example, which induced the
hypothesis of the study. Concerning the sensors used in the
introductory example, their gain and phase characteristics, with
respect to frequency as well as their sensitivity to perimeter changes,
were obviously evaluated (1).
Finally, we are surprised by the reservations of Sackner and Adams
concerning the use of inductive belts incorporated in a pajama, as Dr.
Sackner is the inventor of the LifeShirt technology, in which inductive
belts are incorporated in suits (http://www.lifeshirt.com). Because
the inductive plethysmographic transducer is already made of a wire
sewn into a piece of cloth, we do not think that our setup is
inappropriate. Such a technique has been successfully used in adults on
the ground and during the D-2 and Euromir 95 space missions (6) and,
more recently, during the Neurolab flight
(http://quest.arc.nasa.gov/neuron/photos/SS.html).
In conclusion, Sackner and Adams do not address any fundamental issues
about the validity of our warning against inductive plethysmography for
the measurement of thoracoabdominal asynchrony.
![]()
REPLY
| |
REFERENCES |
|---|
1.
De Groote, A.
Contribution à l'étude des mouvements respiratoires chez le nourrisson durant le sommeil, grâce à une instrumentation et un traitement de signal adaptés (PhD thesis). Brussels: Université Libre de Bruxelles, 1997.
2.
De Groote, A,
Verbandt Y,
Paiva M,
and
Mathys P.
Measurement of thoracoabdominal asynchrony: importance of sensor sensitivity to cross-section deformations.
J Appl Physiol
88:
1295-1302,
2000.
3.
Grover, FW.
Inductance Calculations: Working Formulas and Tables. New York: Van Nostrand, 1946, p. 8, 143.
4.
Martinot-Lagarde, P,
Sartene R,
Mathieu M,
and
Durand G.
What does inductance plethysmography really measure?
J Appl Physiol
64:
1749-1756,
1988.
5.
Sackner, MA,
and
Krieger BP.
Non-invasive respiratory monitoring.
In: Heart-Lung Interactions in Health and Disease, edited by Scharf SM,
and Cassidy SS.. New York: Dekker, 1989, p. 663-805.
6.
Wantier, M,
Estenne M,
Verbanck S,
Prisk GK,
and
Paiva M.
Chest wall mechanics in sustained microgravity.
J Appl Physiol
84:
2060-2065,
1998
7.
Watson, HL,
Poole DA,
and
Sackner MA.
Accuracy of respiratory inductive plethysmographic cross-sectional areas.
J Appl Physiol
65:
306-308,
1988.
| Anne De Groote, | ||||||||||||
| Yves Verbandt, | ||||||||||||
| Manuel Paiva, | ||||||||||||
|
Pierre Mathys, Biomedical Physics Laboratory and Dept. of Electronics Université Libre de Bruxelles 1050 Brussels, Belgium E-mail: adegroot{at}ulb.ac.be |
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||