1 Clinic for Anaesthesiology, Ludwig-Maximilians University Munich, 81377 Munich, Germany; and Departments of 2 Physiology and 3 Paediatrics, Imperial College School of Medicine, London UB7, United Kingdom
| |
ABSTRACT |
|---|
|
|
|---|
We present a newly developed
electromechanical sensor with automated calibration for strain-gauge
plethysmography (filtrass) and compare it to a conventional
mercury-in-Silastic strain-gauge plethysmograph (MSG). Fluid filtration
capacity (Kf) and isovolumetric venous pressure
(Piv) of the limb were assessed noninvasively with both devices in 29 healthy volunteers. We found significantly higher Kf and
Piv values with MSG [4.6 ± 2.0 × 10
3
ml · min
1 · mmHg
1 · 100 ml tissue
1 (Kf units;
KfU) and 21.2 ± 8.1 mmHg for Pvi], than
with filtrass, giving values of 3.1 ± 0.8 KfU and 15.1 ± 7.1 mmHg. Because
Kf and Piv are profoundly influenced by the
calibration, we investigated the quality of the calibration signal and
its impact on the obtained values. We could show that the
reproducibility of repeated calibrations was higher with filtrass (58%
lower mean ± SD). The data were grouped according to the
quality of calibration, and we found no significant difference in
Kf and Piv between filtrass (3.0 ± 0.7 KfU and 15.9 ± 6.9 mmHg, respectively) and
MSG with good calibration signal (3.3 ± 0.8 KfU and 18.6 ± 7.1 mmHg, respectively; no
significant difference). However, we obtained significantly higher MSG values (5.6 ± 2.0 KfU and
23.1 ± 8.4 mmHg, respectively; P < 0.001) in the
group with a bad calibration signal. We suggest that the filtrass
sensor, which performs an automatic, standardized calibration procedure
and shows a linear signal response to stretch, gives highly
reproducible and reliable results and thus is more suitable for routine application.
fluid filtration capacity; isovolumetric venous pressure; noninvasive microcirculatory measurement
| |
INTRODUCTION |
|---|
|
|
|---|
THE STRAIN-GAUGE TECHNIQUE, first described by Whitney (28), utilizes a mercury-filled elastic tubing as a gauge [mercury-in-Silastic strain gauge plethysmograph (MSG)]. Changes in limb circumference are deduced from alterations in the resistance brought about by changes in diameter and length of the gauge. Although elegant and widely used for the noninvasive measurement of blood flow (2, 28), the metal-filled gauge system has some drawbacks, e.g., the toxicity of mercury, the variable lifetime of gauges, and the potential for producing an inaccurate calibration. The latter may be due to the apparent nonlinearity of the resistance change, most markedly obtained at the extreme ends of the strain-gauge stretch. We developed a new mercury-free strain-gauge design with an integrated automatic calibration device (filtrass 2001, DOMED Medizintechnik, Munich, Germany) that overcomes some of the major drawbacks inherent in the use of liquid-filled sensors for venous congestion plethysmography (VCP). Gamble et al. (11), using a modification of Whitney's original mercury-filled strain-gauge system (28), described a VCP protocol for determination of fluid filtration capacity (Kf) and isovolumetric venous pressure (Piv). The same system can also be used for the assessment of venous pressure (5) and factors influencing blood flow (10).
The present paper describes a novel electromechanical sensor, which
performs an automated calibration procedure. Most commercially available strain-gauge plethysmographs do not allow for a calibration of the strain-gauge response to stretch. Usually only the resistance of
the strain gauge is tested, thus ensuring the integrity of mercury-metal contact within the tubing of the sensor. We used the
system described by Gamble et al. (11), which enables a manual calibration. We investigated the quality of the calibration of
this mercury-filled gauge while applying two different calibration protocols and compared it to that obtained with the newly developed filtrass system. The studies were performed on a model limb and on
human legs. Moreover, we compared the values of
Kf and Piv obtained in 29 healthy volunteers
using the protocol described by Gamble et al. (11). We
were interested in establishing why different investigators have found
mean control values for Kf ranging from 2.6 to
5.6 × 10
3
ml · min
1 · 100 ml · tissue
1 · mmHg
1
(Kf units; KfU) (see
Table 1), using a similar VCP protocol and strain-gauge design to that described by Gamble et al.
(11).
|
| |
METHODS |
|---|
|
|
|---|
All subjects gave informed consent, and the study was approved by the local ethics committee.
Filtrass 2001.
The underlying principle of the filtrass 2001 sensor is fundamentally
different from that of the liquid metal-filled Silastic tubes of
conventional strain gauges, because no elastic recoil of a distensible
component is involved (Fig. 1). Changes
in limb circumference are detected with a passive inductive transducer, which has an accuracy of ±5 µm (Fig.
2). An inelastic but flexible plastic
line with a diameter of 0.5 mm spanning around the limb connects the
transducer to an in-built motor (Fig. 2) with a position accuracy also
of ±5 µm and a weight of 12 g. The plastic line glides on a
flexible zigzag band with a silicon-coated surface, ensuring low
friction and good fixation of the sensor to the limb (Fig. 1).
When the limb circumference increases, the passive transducer can be
pulled out to a maximum of 4 mm. This is followed by an outward drive
of the stepping motor if the length change exceeds 4 mm. This design
enables measurement of a maximum change in circumference of 22 mm,
which is certainly greater than the circumference changes that we have
experienced during our cumulative small pressure step studies
(11). With the filtrass sensor, the motor automatically resets to the initial position of the passive transducer before each
pressure step. This procedure ensures that the sensor is always
operating over the same sensing range, an advantage that is not shared
by the Silastic tube system.
|
|
MSG. MSG has been described in detail elsewhere (11, 28). Briefly, the gauge consists of tubing made of a silicon elastomer, filled with mercury, sealed with molybdenum pins, and balanced against an adjacent temperature compensation coil in a Wheatstone bridge. A double loop is formed around the limb to increase the length of the sensor and, therefore, the accuracy of measurement (28). The loops are positioned 1 cm apart on a calf with measured circumference and kept in place by means of plastic guides on ribbon with a low coefficient of friction. The friction component is decreased further by coating the guides, ribbon, and gauges with silicone lubricant (11). Changes in strain-gauge length cause alterations in electrical resistance of the gauge in the low-voltage circuit.
Calibration procedures and quality of calibration. A pre-tension giving a deflection equivalent to a 5-V vertical movement on the computer screen, i.e., a 50% vertical shift, was found to provide sufficient tension (11). The standard, known increase and decrease in gauge length were then delivered by means of five double turns of the calibration nut. The five up-and-down steps represented the change in signal attributable to known increases and decreases in gauge length, equivalent to ~0.8 V/mm change in length. At the end of the calibration procedure, the input length and voltage signals were translated into a calibration factor, with units of volts per millimeter, which was saved to disk for use during the subsequent off-line analysis.
All MSG data were sampled at 1 Hz for subsequent off-line analysis. Because one objective of this study was to investigate the influence of gauge pre-tensioning protocols on the results obtained, we quantified the quality of the calibration procedure by looking at the slope of the relationship between stretch and signal and by noting the deviation from the ideal linear relationship. All MSG calibration data were analyzed, off-line, to study the relationship between signal volts displayed on the computer screen and length (micrometers) due to each calibration pull of 5,000 µm. The slope of the relationship between volts and length in micrometers was defined as the calibration factor. A retrospective analysis of 109 previously performed MSG studies revealed a mean calibration factor of 0.83 ± 0.19. We defined a good calibration as lying within ±1.5 SD of this calibration factor. Hence a calibration was considered "good" when the volt-to-millimeter ratio was between 0.55 and 1.11 and "bad" when it lay outside this range. By that means, we excluded the extreme ends of the gauge stretch, in which the nonlinearity is most apparent. Referring to Hooke's law, a linear mechanical behavior of a coil can be expected only after an adequate pre-tensioning and before it is overstretched. These two situations would result in either a low or a high calibration factor, which were defined as bad calibrations.Study design. In the first study, we tested the reproducibility of the calibration signals of MSG and the filtrass sensors during repeated measurements. Both types of sensors were fixed on a model limb (circumference 420 mm), and the calibrations were performed 11 times. The deviation from the first run was compared for each run and expressed as standard deviation.
Reproducibility of the calibration in vivo. We tested the reproducibility of the calibration signal of both MSG and filtrass on seven healthy young (20-24 yr) volunteers (2 women, 5 men). Two MSG and two filtrass sensors were applied in randomized order on both calves [mean circumference 366 (range 320-400) and 364 (range 318-399) mm, respectively]. The calibration procedure was repeated 11 times for each of the four sensors. We calculated the deviation of every calibration signal from the first calibration run and determined the standard deviation.
Comparison of Kf and Piv obtained by filtrass and MSG sensors. Because the venous congestion protocol, which was common to all studies, has been described in detail elsewhere (11), we will only give an outline of the basic protocol used. Bilateral assessments of 29 healthy subjects, giving a total of 50 analyzable observations, were made (13 women, age 23.3 ± 2.3 yr, weight 58.4 ± 6.1 kg, height 170.3 ± 6.3 cm and 16 men, age 24.0 ± 3.5 yr, weight 76.4 ± 5.3 kg, height 180.1 ± 4.9 cm). The subjects lay supine for at least 20 min before the commencement of the study protocol. The legs were supported on a vacuum mattress with the midcalf points leveled with respect to the right atrium (one-third of the distance from sternal angle to the surface of the supporting bench). Congestion cuffs were fitted bilaterally at mid-thigh level, and the pressure was applied by the in-built filtrass pressure module. This allows a free selection of pressure steps ranging from 1 to 350 mmHg. In the present study, we applied seven cumulative pressure steps of 10 mmHg, which were maintained for 4 min each.
The circumference of each calf was measured before the strain gauge and the filtrass sensor were attached. Close to the maximum circumference of the calf, one MSG and one filtrass sensor were placed on each leg so that a total of four sensors were used. The position in relation to the knee was chosen in randomized order. The MSG sensors were set to the standard tensions and then calibrated at the start of each study, as described above.Determination of Kf and Piv. The determination of Kf and Piv follows the protocol described by Gamble et al. (11) and was used both in MSG and filtrass. Seven cumulative small pressure steps (10 mmHg) were applied without exceeding the diastolic blood pressure. The resulting volume changes were simultaneously measured with the two types of sensors and analyzed off-line. The ambient temperature was continuously recorded throughout the study period using the temperature sensor of filtrass, which is on the surface of the sensor holder.
Data analysis. When cuff pressure (Pcuff) is elevated above the existing venous pressure, a rapid increase in limb circumference attributable to altered vascular volume can be seen. A gradual increase in limb volume is observed when Pcuff exceeds filtration pressure at the microvascular interface. This change is caused by fluid filtration. Both the vascular volume and limb volume components can be obtained for each pressure step, because a series of pressure steps are applied. The relationship between Pcuff and limb volume is linear and represents the Kf (11). The intercept on the x-axis is the Piv, the Pcuff that needs to be exceeded to generate net fluid flux. A noninvasive measure of venous pressure at the level of the strain gauge is obtained by determining the intercept of the relationship between the vascular volume component and Pcuff (5). The data obtained with MSG and filtrass systems were blinded and analyzed independently by two investigators.
Statistical analysis. For data analysis we used Sigma Stat (SPSS, Chicago, IL). Values are given as means ± SD, and the mean values are compared by the paired Student's t-test; in cases of nonnormal distribution, we applied the Wilcoxon signed-ranked test. Significance was assumed at P < 0.05.
| |
RESULTS |
|---|
|
|
|---|
Reproducibility of the calibration signal (model limb). The regression analysis of the 11 repetitive measurements with 5 steps on the model limb revealed a r2 = 0.9751 for MSG, which was comparable to r2 = 0.9994 in filtrass. The deviation of 10 repeated manual calibrations from the first calibration of MSG gave a mean ± SD from the first run of ±99 µm compared with ±16 µm with filtrass. The maximum deviation observed in MSG was 395 µm, whereas the automated calibration procedure gave a maximum deviation of 43 µm.
Repetitive calibrations on human limbs.
The mean intraindividual standard deviation during 10 repetitive
calibration runs on human legs was ±132 (range 37-251) µm for
the MSG sensor compared with ±56 (range 29-109) µm for the filtrass sensor. The standard deviation during 280 individual calibration procedures in all volunteers ranged from 9 to 327 µm in
MSG but was only 13-150 µm for the filtrass sensor.
Representative examples of the calibration in vivo are given for MSG in
Fig. 3A and for filtrass in
3B. The obvious nonlinearity of the MSG calibration signal
and the wider deviation between the different calibration maneuvers can
be seen.
|
Kf and Piv values for 29 healthy volunteers. No significant changes in the temperature (either ambient or skin temperature) occurred during the study period. The in-built temperature probe of filtrass, which is located 5 mm above the skin level, representing a mixture of both skin and ambient temperature, was 28.6 ± 1.1°C at the start and 28.3 ± 1.0°C at the end of the study.
Measurement with the MSG plethysmograph gave significantly higher (P < 0.001) Kf values (4.6 ± 2.0 KfU) than with filtrass (3.1 ± 0.8 KfU). The standard deviation for these values was found to be more than double in MSG. Also Piv was measured significantly (P < 0.001) higher with MSG compared with filtrass, giving values of 21.2 ± 8.1 and 15.1 ± 7.1 mmHg, respectively. Because we could not readily explain these differences but already have shown the importance of the calibration, we grouped all MSG and their depending filtrass measurements according to their MSG calibration factor as defined above.Differences in Kf and Piv after grouping of the data
depending on the calibration factor.
The pre-tensioning procedures used gave rise to good calibration
factors in 21 measurements made in 16 volunteers and bad calibration
factors in 29 measurements made in 20 subjects. The Kf values found with MSG with a good calibration
signal were 3.3 ± 0.8 KfU, which was not
significantly different from the filtrass value of 3.0 ± 0.7 KfU (Fig.
4A). In contrast, the MSG
values obtained from a bad calibration signal were 5.6 ± 2.0 KfU, which was significantly (P < 0.001) greater than the value 3.3 ± 0.9 KfU measured with filtrass (Fig. 4B).
There was no significant difference in the Kf
values found with filtrass when the data were grouped according to the
quality of the MSG calibration, whereas the MSG-based
Kf was significantly higher in values with a
calibration factor < 0.55 or > 1.11 (P < 0.001).
|
| |
DISCUSSION |
|---|
|
|
|---|
This paper describes a novel, automatically calibrating, liquid metal-free measuring device for VCP. The studies showed that it gave more reproducible results than simultaneous studies using MSG sensors. In our opinion, several factors may have contributed to these observations, including the reproducibility of the manufacturing process of the sensor resulting from an industrial production, the automated calibration procedure, the absence of an elastic component, and the mercury-free design of the filtrass sensor.
The MSG holders and sensors used in the present and previous studies were custom made in our own laboratories; hence differences existed in the actual design and the mechanics for calibration. This may partially account for the different mean and median Kf values reported by the groups that have used this type of MSG (Table 1). The filtrass sensor, however, is an industrial product, with less variation in the sensor's mechanics. The manual calibration procedure of MSG has to be less accurate than that conducted by a motor with a position accuracy of ±5 µm. This expectation was confirmed by the calibration data obtained from the studies on both model and human limbs. The fact that filtrass utilizes the actual calibration curve as an "ideal standard," rather than a constant factor for the calculation of changes in limb volume as used in MSG, further adds to the higher accuracy of filtrass.
In MSG, silicon tubing is used in the construction of the gauges. Because the silicone material is an elastomer, it has recoil properties that, although they form an important part of the gauge function, are subject to changes attributable to altered manufacturing procedures, aging, and shelf life.
Other commercially available MSG systems do not give details on their manufacturing procedure for strain gauges. Because they are industrially produced, one assumes that the problems described above are less eminent. To our knowledge, no manufacturer, however, performs a further quality assurance besides a test for an electrical resistance change within the gauge, after delivery to the customer.
Quality of the MSG calibration. In the present study, we used filtrass as a control to test the working hypothesis that the large range of Kf values obtained in control studies on adults, in our own and other published studies, result from inappropriate pre-tensioning of the Silastic strain gauges and the nonlinearity of the resistance change in the gauge when stretched. The significance of pre-tensioning can best be explained in terms of Hooke's law, which states that, until the elastic limit of a material is reached, strain is proportional to the applied stress. Clearly, with the elastomer comprising the MSG, this relationship cannot be presumed until there is sufficient pre-tension to remove the inherent "kinks" in the tubing. Although an undertensioned gauge will give uniform steps, they will be inappropriately small. One of the main disadvantages of the MSG system is that the appropriate pre-tension and the calibration cannot be checked on-line, and the accuracy of the calibration and, therefore, the success of the study depend on the experience of the operator. In conclusion, we believe that one of the major sources of error when MSG is used for the determination of change in length is the failure to identify an optimal pre-tension.
In clear contrast to MSG, the filtrass sensor seems not to be hampered by a variability in the calibration. Inaccuracies due to mercury-containing strain gauges are completely avoided with filtrass because the relationship of pull and response by the inductive transducer is linear and the motor always resets the measurement range to the optimum value. Moreover, because it utilizes the actual response of the inductive transducer to a defined pull by the motor, a deviation from this linearity, due for example to increased friction, can be detected. This would result in an increased time delay or a deviation from the ideally linear relationship between the defined pull and the response of the inductive transducer. The time delay is tested for each calibration, and only if a maximum of 500 ms is not exceeded can the study be continued. Furthermore, because three calibration runs are always performed and the deviation between two calibrations has to be <200 µm, we assumed that more reproducible results should result from the use of the filtrass system. This contention is supported by the results obtained in the present study.Differences in Kf and Piv. The significant differences of Kf and Piv values and especially the wider spread of Kf obtained with the MSG plethysmograph lead to more inaccurate findings and therefore will provide less likely detectable differences when VCP is used for microcirculatory monitoring. The calibration signal of the MSG plethysmograph, both in the one used as well as in all other commercially available strain-gauge systems, is not analyzed in routine use. This leads to an inaccuracy of VCP measurements, which can only be in part overcome with great experience of the investigator. However, it is reassuring that, given an experienced investigator and a good calibration, no significant differences in the mean values for Kf and Piv were found in the present study. Moreover, the Kf values reported here are in good agreement with recent data published by our group and by colleagues using a similar MSG design and pressure protocol (3, 4, 7-9, 11, 13-15, 21, 22). They are also within the range reported by other investigators (16, 17, 24-27), who gave values ranging from 2.6 to 4.8 KfU. The results from the present studies support the notion that the differences in control Kf values that we have noticed in MSG studies (see Table 1) are most likely due to an insufficient pre-tensioning. This will result in a low calibration ratio and give rise to significantly higher Kf values. The use of a consistent and appropriate protocol has been shown to give rise to good longitudinal reproducibility of Kf values with a coefficient of variation of 10.6% (9). Because all investigators have used a consistent pre-tensioning protocol, it may be less surprising that the standard deviation in the results of each of these studies is small, on the order of 5-10% (see Table 1).
All of these data contrast markedly with those of Lundvall and co-workers (18, 20) who, using volume plethysmography, reported Kf values 10 times higher than previously reported values. Moreover, these workers applied venous compression pressures of 1.5-3.4 mmHg and argued that pressure steps >5 mmHg could activate the arteriovenous regulation mechanisms first described by Henriksen and Sejrsen (12). We found no evidence for such an activation, provided that the venous compression pressure steps were
10 mmHg (11). Indeed, we have found
that venous congestion pressures, when applied in small cumulative
steps, can be increased to within 20 mmHg of mean arterial blood
pressure without causing a decrease in limb blood flow
(10). It is interesting to note that Lundvall and
Länne (20) determined Kf
after the application of hydrostatic pressures that were smaller than
the expected capillary pressure, i.e., ~20 mmHg (1, 23),
which does raise questions about the validity of these results.
Temperature. Both skin and ambient temperature are possible sources of error when, in mercury-containing systems, changes in limb circumference are deduced from the change in resistance (28). An increase in temperature decreases the resistance of the mercury, thus having a major influence on the results obtained. In the MSG design used by our group, a temperature compensation is included in which a copper wire is in contact with both skin and ambient room air. Temperature changes occurring during a study, however, cannot be considered, because the temperature is adjusted for only at the beginning of a study. In our previous control studies, we always provided for constant skin temperature by means of a temperature control unit (11). This, however, is not always possible, particularly in clinical settings, e.g., on an intensive care ward (6). The filtrass sensor is independent of ambient temperature and free of this possible setback. Nevertheless, changes in ambient temperature should be avoided, because they have a profound influence on the measured microcirculatory parameters (19, 25).
Comparison to other commercially available strain-gauge systems. In this study no comparisons to other commercially available strain-gauge systems were made; hence no conclusions can be drawn with respect to the data given. However, one might speculate that the problems of MSG described in the present paper are of relevance to all MSG systems, in particular because either no calibration is performed or the quality of the calibration is not tested for. MSG plethysmographs are widely used for the assessment of limb blood flow and venous dynamics. In a yet-unpublished study, we found a significantly lower standard deviation of blood flow values with filtrass (10.3% during three repeated measurements) than with a commercially available strain-gauge system (22.3%), thus confirming our assumption that filtrass may be superior to MSG because of the improved calibration procedure and the avoidance of mercury.
Possible drawbacks of filtrass. The specific design of the filtrass sensor may add one source of error not present in MSG. Before any movement of the inductive transducer is initiated, the sticking friction of the plastic wire and of the friction within the inductive transducer has to be overcome. During any movement the gliding friction is always present. In contrast, the MSG behaves like a spring with very little sticking friction but, however, with a similar gliding friction. The volume responses obtained with MSG closely follow the stress-strain function of the elastic component of the silicon tubing, whereas in filtrass a step function may be seen, in particular when the sticking friction is relatively high. Although this is a potential drawback, in reality it is of minor practical importance because, during the calibration, the system automatically tests for the presence of friction, which is reflected as an increased time delay during the calibration procedure. If the time delay is too great, the calibration is rejected and an appropriate reason is given; this usually indicates the necessity to change the zigzag band.
In conclusion, we have developed a new, automated venous compression plethysmography system and compared its quality of calibration and resulting data to those obtained using our earlier conventional MSG plethysmography. We found that the new device had smaller standard deviation during repetitive calibrations on both model and human limbs. Using the previously described protocol to determine Kf and Piv, we found that the new device gave similar values for Kf when the calibration procedure of MSG was performed within 1.5 SDs of the optimum; however, incorrectly higher values were obtained with MSG when outside of this range. Because the standard deviations during repetitive measurements using the filtrass device were smaller and also because the new device utilizes the deviations of calibration curve rather than an "ideal standard" value for the calculation of changes in limb volume, we conclude that it has a higher accuracy than conventional MSG systems.| |
ACKNOWLEDGEMENTS |
|---|
The authors thank Daniel Oser for technical expertise regarding filtrass and Peter Christ, who conducted some of the experiments presented.
| |
FOOTNOTES |
|---|
F. Christ as an advisor and cofounder of DOMED Medizintechnik would like to disclose his commercial interest in filtrass.
This publication contains part of the MD thesis of A. Bauer.
Address for reprint requests and other correspondence: F. Christ, Clinic for Anaesthesiology, Ludwig-Maximilians Univ. Munich, 81377 Munich, Germany (E-mail: frank.christ{at}ana.med.uni-muenchen.de).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 18 November 1999; accepted in final form 8 May 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Aukland, K,
and
Reed RK.
Interstitial-lymphatic mechanisms in the control of extracellular fluid volume.
Physiol Rev
73:
1-78,
1993
2.
Benjamin, N,
Calver A,
Collier J,
Robinson B,
Vallance P,
and
Webb D.
Measuring forearm blood flow and interpreting the responses to drugs and mediators.
Hypertension
25:
918-923,
1995
3.
Christ, F,
Dellian M,
Goetz AE,
Gamble J,
and
Messmer K.
Changes in subcutaneous interstitial fluid pressure, tissue oxygenation, and skin red cell flux during venous congestion plethysmography in men.
Microcirculation
4:
75-81,
1997[Medline].
4.
Christ, F,
Gamble J,
Baranov V,
Kotov A,
Gartside IB,
Moser C,
Abicht JM,
Nehring I,
and
Messmer K.
Changes in microvascular fluid filtration capacity (Kf) during 120 days of 6° head down tilt (Abstract).
J Vasc Res
35:
29,
1998.
5.
Christ, F,
Gamble J,
Baschnegger H,
and
Gartside IB.
Relationship between venous pressure and tissue volume during venous congestion plethysmography (VCP) in man.
J Physiol (Lond)
503:
463-468,
1997[ISI][Medline].
6.
Christ, F,
Gamble J,
Gartside IB,
and
Kox WJ.
Increased microvascular water permeability in patients with septic shock, assessed with venous congestion plethysmography (VCP).
Intensive Care Med
24:
18-26,
1998[ISI][Medline].
7.
Gamble, J,
Anim N,
Smith JE,
Gartside IB,
and
Steer PJ.
Changes in peripheral microvascular parameters in pre-eclamptic (PET) patients (Abstract).
Int J Microcirc Clin Exp
16, Suppl 1:
197,
1996.
8.
Gamble, J,
Christ F,
and
Gartside IB.
The influence of venous congestion pressure (VCP) step size on the estimation of limb vascular permeability (Abstract).
Int J Microcirc Clin Exp
16, Suppl 1:
222,
1996.
9.
Gamble, J,
Christ F,
and
Gartside IB.
The effect of passive tilting on microvascular parameters in the human calf: a strain gauge plethysmography study.
J Physiol (Lond)
498:
541-552,
1997[ISI][Medline].
10.
Gamble, J,
Christ F,
and
Gartside IB.
Human calf precapillary resistance decreases in response to small cumulative increases in venous congestion pressure.
J Physiol (Lond)
507:
611-617,
1998
11.
Gamble, J,
Gartside IB,
and
Christ F.
A reassessment of mercury in silastic strain gauge plethysmography for microvascular permeability assessment in man.
J Physiol (Lond)
464:
407-422,
1993
12.
Henriksen, O,
and
Sejrsen P.
Local reflex in microcirculation in human skeletal muscle.
Acta Physiol Scand
99:
19-26,
1977[ISI][Medline].
13.
Jaap, AJ,
Shore AC,
Gamble J,
Gartside IB,
and
Tooke JE.
Capillary filtration coefficient in type II (non-insulin- dependent) diabetes.
J Diabetes Complications
8:
111-116,
1994[Medline].
14.
Jaap, AJ,
Shore AC,
and
Tooke JE.
Differences in microvascular fluid permeability between long-duration type I (insulin-dependent) diabetic patients with and without significant microangiopathy.
Clin Sci (Colch)
90:
113-117,
1996[Medline].
15.
Jeal, S,
Brown MD,
Bryant J,
and
Gamble J.
Differences in microvascular parameters in the limbs of endurance-trained, strength-trained and sedentary humans (Abstract).
J Physiol (Lond)
506:
23P,
1998.
16.
Krogh, A,
Landis AM,
and
Turner AH.
The movement of fluid through the human capillary wall in relation to venous pressure and to the colloid osmotic pressure of blood.
J Clin Invest
11:
63-95,
1932[Medline].
17.
Landis, EM,
and
Gibbon JH.
The effects of temperature and of tissue pressure on the movement of fluid through the human capillary wall.
J Clin Invest
12:
105-138,
1933[Medline].
18.
Länne, T,
Edfeldt H,
Quittenbaum S,
and
Lundvall J.
Large capillary fluid permeability in skeletal muscle and skin of man as a basis for rapid beneficial fluid transfer between tissue and blood.
Acta Physiol Scand
146:
313-319,
1992[ISI][Medline].
19.
Levick, JR,
and
Michel CC.
The effects of position and skin temperature on the capillary pressures in the fingers and toes.
J Physiol (Lond)
274:
97-109,
1978
20.
Lundvall, J,
and
Länne T.
Much larger transcapillary hydrodynamic conductivity in skeletal muscle and skin of man than previously believed.
Acta Physiol Scand
136:
7-16,
1989[ISI][Medline].
21.
Mahy, IR,
Lewis DM,
Shore AC,
Penney MD,
Smith LD,
and
Tooke JE.
Disturbance of peripheral microvascular fluid permeability by the onset of atrioventricular asynchrony in patients with programmable pacemakers.
Heart
75:
509-512,
1996
21a.
Mahy, IR,
Shore AC,
Smith LD,
and
Tooke JE.
The peripheral microcirculation in atrial fibrillation: preservation of capillary pressure and filtration coefficient.
Cardiovasc Res
28:
1555-1558,
1994[Medline].
22.
Mahy, IR,
Shore AC,
Smith LD,
and
Tooke JE.
Disturbance of peripheral microvascular function in congestive heart failure secondary to idiopathic dilated cardiomyopathy.
Cardiovasc Res
30:
939-944,
1995[ISI][Medline].
23.
Michel, CC.
Fluid movements through capillary walls.
In: Handbook of Physiology. The Cardiovascular System. Microcirculation. Bethesda, MD: Am. Physiol. Soc, 1984, sect. 2, vol. IV, pt. 1, chapt. 9, p. 375-409.
24.
Michel, CC,
and
Moyses C.
Fluid filtration with increments of venous pressure in the feet and forearms of human subjects (Abstract).
J Physiol (Lond)
362:
38P,
1985.
25.
Michel, CC,
and
Moyses C.
The measurement of fluid filtration in human limbs.
In: Clinical Investigations of Microcirculation, edited by Tooke JE,
and Smaje LH.. Boston, MA: Morhuis Nyhoff, 1987, p. 103-126.
26.
Michel, CC,
and
Moyses C.
Fluid filtration into the feet of human subjects during quiet sitting (Abstract).
J Physiol (Lond)
371:
74P,
1994.
27.
Seem, E,
and
Stranden E.
Transcapillary filtration in lower limbs with deep venous thrombosis; the role of the capillary filtration coefficient.
Scand J Clin Lab Invest
50:
331-336,
1990[Medline].
28.
Whitney, RJ.
The measurement of volume changes in human limbs.
J Physiol (Lond)
121:
1-27,
1953.
This article has been cited by other articles:
![]() |
A. O. Stirban and D. Tschoepe Cardiovascular Complications in Diabetes: Targets and interventions Diabetes Care, February 1, 2008; 31(Supplement_2): S215 - S221. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A.J. Struijker-Boudier, A. E. Rosei, P. Bruneval, P. G. Camici, F. Christ, D. Henrion, B. I. Levy, A. Pries, and J.-L. Vanoverschelde Evaluation of the microcirculation in hypertension and cardiovascular disease Eur. Heart J., December 1, 2007; 28(23): 2834 - 2840. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pickkers, M. J. Dorresteijn, M. P.W.J.M. Bouw;, J. G. van der Hoeven, and P. Smits In Vivo Evidence for Nitric Oxide-Mediated Calcium-Activated Potassium-Channel Activation During Human Endotoxemia Circulation, August 1, 2006; 114(5): 414 - 421. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Dorresteijn, L. T. van Eijk, M. G. Netea, P. Smits, J. G. van der Hoeven, and P. Pickkers Iso-osmolar prehydration shifts the cytokine response towards a more anti-inflammatory balance in human endotoxemia Innate Immunity, October 1, 2005; 11(5): 287 - 293. [Abstract] [PDF] |
||||
![]() |
S. S. Dhillon, K. Mahadevan, V. Bandi, Z. Zheng, C. W. Smith, and R. E. Rumbaut Neutrophils, Nitric Oxide, and Microvascular Permeability in Severe Sepsis Chest, September 1, 2005; 128(3): 1706 - 1712. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Gamble, M. J. Joyner, N. M. Dietz, and J. T. Shepherd A restrospective perspective J Appl Physiol, February 1, 2005; 98(2): 762 - 763. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Leslie, T. Attina, E. Hultsch, L. Bolscher, M. Grossman, M. A. Denvir, and D. J. Webb Comparison of two plethysmography systems in assessment of forearm blood flow J Appl Physiol, May 1, 2004; 96(5): 1794 - 1799. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Anim-Nyame, S. R Sooranna, M. R Johnson, J. Gamble, and P. J Steer A longitudinal study of resting peripheral blood flow in normal pregnancy and pregnancies complicated by chronic hypertension and pre-eclampsia Cardiovasc Res, June 1, 2001; 50(3): 603 - 609. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |