Vol. 89, Issue 3, 1159-1164, September 2000
Effects of anodal vs. cathodal pacing on the
mechanical performance of the isolated rabbit heart
Anshul
Thakral,
Louis H.
Stein,
Mahesh
Shenai,
Boris
I.
Gramatikov, and
Nitish V.
Thakor
Department of Biomedical Engineering, Johns Hopkins School of
Medicine, Baltimore, Maryland 21205
 |
ABSTRACT |
Previous studies have suggested that
anodal pacing enhances electrical conduction in the heart near the
pacing site. It was hypothesized that enhanced conduction by
anodal pacing would also enhance ventricular pressure in the heart.
Left ventricular pressure measurements were made in isolated,
Langendorff-perfused rabbit hearts by means of a Millar pressure
transducer with the use of a balloon catheter fixed in the left
ventricle. The pressure wave was analyzed for maximum pressure (Pmax)
generated in the left ventricle and the work done by the left ventricle
(Parea). Eight hearts were paced with monophasic square-wave pulses of
varying amplitudes (2, 4, 6, and 8 V) with 100 pulses of each waveform delivered to the epicardium. Anodal stimulation pulses showed statistically significant improvement in mechanical response at 2, 4, and 8 V. Relative to unipolar cathodal pacing, unipolar anodal pacing
improved Pmax by 4.4 ± 2.3 (SD), 5.3 ± 3.1, 3.5 ± 4.9, and 4.8 ± 1.9% at 2, 4, 6, and 8 V, respectively. Unipolar anodal stimulation also improved Parea by 9.0 ± 3.0, 12.0 ± 6.0, 10.1 ± 7.7, and 11.9 ± 6.0% at 2, 4, 6, and 8 V,
respectively. Improvements in Pmax and Parea indicate that an anodally
paced heart has a stronger mechanical response than does a cathodally paced heart. Anodal pacing might be useful as a novel therapeutic technology to treat mechanically impaired or failed hearts.
anodal pacing; cardiac stimulation; pacing waveforms; myocardial
contraction
 |
INTRODUCTION |
PERMANENT
PACEMAKERS ARE THE standard treatment for a variety of
symptomatic bradycardias, such as sinus node dysfunction and
atrioventricular block. Cardiac stimulation and pacing have focused on
such parameters as stimulation site, waveform polarity, stimulus
strength, and pulse duration. Clinically, artificial pacemakers have
classically utilized standard unipolar cathodal (depolarizing)
waveforms (
1 ms in duration) because it is believed to be safer in
regards to inducing arrhythmia. Typically, unipolar anodal stimulation
has a shorter absolute refractory period than does unipolar cathodal
stimulation. It is presumed that this fact explains the smaller risk of
inducing ventricular fibrillation with cathodal pacing. Dekker
(4) showed that anodal currents can be used to excite the
myocardium; however, their threshold level for excitation is generally
higher than that of cathodal stimulation. The same author showed that,
although late diastolic stimulation thresholds are lower with cathodal
stimulation, with closely coupled stimuli, stimulation thresholds for
anodal pacing may become lower than cathodal thresholds. The threshold
levels may drop below those of unipolar cathodal pacing during the
relative refractory period. Cranefield et al. (3)
hypothesized that, for relative refractory tissue, anodal excitation
may be more efficient. Indeed, with shorter coupling intervals, during
the relative refractory period, anodal stimuli increase the amount of
sodium available for depolarization, accelerate repolarization, and increase depolarization amplitudes and upstroke. Previous work in
our laboratory showed that the increased thresholds of anodal pacing
can be brought down to the level of cathodal pacing by pacing with
equipolar biphasic pulses (18). Another recent study
(5) suggested that pulses of opposite polarity with some time delay in between might produce a similar effect.
Recent work done in humans with biventricular pacing, as a possible
treatment of congestive heart failure, has raised speculation that the
polarity of the stimulating pulse may have positive effects on the
contraction of the myocardium (1, 2,
17). Experimental work in our group demonstrated that
anodal pacing shows marked improvement in electrical conduction
velocities over cathodal pacing (14, 18). The
possible mechanism of enhanced electrical conduction may be associated
with an increased amount of sodium available and improvement in action
potential (AP) upstroke (16). In the present study, we
hypothesize that the improvements of electrical conduction and AP
upstroke due to anodal pacing will lead to a faster activation process
and hence to a possible improvement in the mechanical response of the
myocardium. A pilot study done in mongrel dogs first suggested that
anodal unipolar pacing might improve cardiac output (7). A
preliminary study done in our laboratory in isolated hearts appeared to
support this claim. We observed that changing the pacing waveform
slightly improved the ventricular pressure. We seek to examine the
statistical significance of this pacing therapy on the mechanical
response of the heart under different pacing conditions.
The central goal of this project was to test the hypothesis that anodal
stimulation would have a statistically significant, beneficial effect
on the contraction of myocardium and would show an improvement in
mechanical performance compared with standard cathodal
stimulation. We investigated the effects of the pacing waveforms on
intraventricular pressure by measuring the maximum pressure (Pmax)
generated by the left ventricle (LV) and the work performed by it
(Parea) in response to various waveforms.
 |
METHODS |
Eight New Zealand White rabbits, weighing between 2.5 and 3.0 kg, were used for this investigation. The animals were anesthetized intravenously with a 50 mg/kg dose of pentobarbital sodium with 2,000 USP units of heparin added as an anticoagulant. The heart was then
rapidly excised through median strenotomy, and the aorta was
cannulated. A modified HEPES buffer (in mM: 108 NaCl, 5 KCl, 5 HEPES,
20 sodium acetate, and 1 MgCl2) mixed with 32 ml of 0.5 M
CaCl2, 14.2 g dextrose, 4.4 g sodium pyruvate, 4 ml insulin, and up to 8 liters of deionized water was perfused in the
heart in a classic retrograde apparatus. Physiological pH of 7.4 and perfusate temperature of 37°C were maintained throughout the
experiment. Perfusion was delivered through a vertical column to ensure
that the mean aortic pressure was maintained at 80 mmHg. Once the
perfusion was stabilized, Ag-AgCl needle electrodes were inserted
epicardially into the myocardium for stimulation, as well as for
electrogram recording. The stimulation was delivered to the apex of the
LV with the reference kept at the base of the LV; thus the stimulating electrodes were separated by the distance of the LV, which was ~4 cm.
The heart was hanging freely from the apparatus to avoid mechanical
strain and was submersed into a perfusion bath connected to the pacing
reference and amplifier ground through a large-surface conductive plate
(10 cm2). Anodal and cathodal thresholds were determined
qualitatively to ensure that the myocardium was being stimulated
("capturing"). These thresholds were determined quantitatively in
previous studies done in our laboratory (18). A bipolar
electrogram was recorded by means of epicardial needle electrodes
inserted longitudinal to fibers in the middle of the right ventricle.
The electrogram was recorded for the purposes of monitoring electrical
activity and ensuring capture due to stimulation. A balloon catheter
was then inserted into the LV via the left atrium. Intraventricular pressure measurements were recorded from the balloon catheter by means
of a Millar pressure transducer (model SPC-470, Millar Instruments,
Houston, TX).
An NB-MIO-16L data-acquisition board, along with a customized program
written in LabVIEW (National Instruments, Austin, TX) running on a
Macintosh Quadra 650, was used for stimulation as well as data
acquisition. The program generated waveforms with varying amplitudes,
shapes, durations, and interstimulus coupling intervals. These
stimulation waveforms were then delivered through an end-stage
amplifier and Ag-AgCl needle electrodes to the heart. The program was
also set up to record the bipolar electrogram, as well as LV pressure,
simultaneously. The stimulus was also recorded so that pressure and
electrogram readings could be correlated with the characteristics of
the individual stimulus pulses.
Once excised and set up on the apparatus, the heart was stabilized for
a period of controlled pacing, which was adjusted in amplitude,
duration, and coupling interval, 5-10% shorter than spontaneous
R-R interval of the heart (generally between 400 and 600 ms), to ensure
capturing. The stimulus was delivered at 1-ms duration. The pacing
protocol was started 30 min after the heart was excised. It consisted
of controlled pacing with two different waveforms at four different
voltage levels. The stimulation was given in sets of 100 pulses of each
waveform. The order of the waveform was alternated between sets, i.e.,
anodal followed by cathodal and cathodal followed by anodal. The
acquisitions were stepped through for 2, 4, 6, and 8 V. The protocol
was repeated three to five times for each experiment. In
experiments 1 and 6, the heart was only able to
capture for two or three acquisitions per stimulation level.
The data-acquisition program recorded from three channels (stimulation,
bipolar electrogram, and pressure) at a sampling frequency of 1,000 Hz.
These data were then analyzed via a custom program written in MATLAB
(Mathworks, Natick, MA). The acquired pressure waves were analyzed for
two parameters: maximum systolic pressure generated (Pmax) and work
done (Parea). To ensure a "steady-state" condition, only the last
20 beats of each acquisition were analyzed; only captured stimulation
and corresponding pressure responses were considered. The Pmax of each
pressure wave was defined as the Pmax value after the onset of each
captured depolarization. Parea was calculated by estimating the
integral of the pressure wave from pressure wave onset to the time when
the pressure had fallen to 5% of Pmax.
Both parameters were calculated for each beat for both cathodal and
anodal pacing. The results were then represented as means ± SD
for every acquisition. This was repeated for each voltage level. Data
were then compared in terms of percent improvement due to anodal
pacing. The governing equation for each parameter was (A
C)/C × 100%, where A is anodal and C is cathodal.
For each experiment, the averages across all acquisitions for a
particular voltage and particular parameter were represented as
means ± SD and were then run through a paired t-test
to establish statistical significance. (A P value < 0.05 represented a statistically significant difference due to anodal
pacing.) An ANOVA test (repeated measures) was then run across all
experiments for each voltage level and each parameter to analyze the
variance due to two factors: waveform (anodal vs. cathodal) and subject
(experiment). The goal of the test was twofold: 1) to
analyze the impact of changing the waveform, and
2) to check whether differences among subjects were
significant. A P value < 0.05 indicates that the
variable was a statistically significant factor in the variance of data.
 |
RESULTS |
Figure 1 shows a typical recording
of average pressure waves resulting from a train of anodal (solid line)
and cathodal (dashed line) stimulation. It is important to note that
the averages for the pressure waves are taken for the last 20 beats of
the recording, thus avoiding variances due to the transient adaptation
resulting from switching pacing waveforms.

View larger version (12K):
[in this window]
[in a new window]
|
Fig. 1.
Typical recording of averaged pressure waves due to
anodal (solid line) and cathodal (dashed line) stimulation. Percent
improvement for maximum pressure (Pmax) at 4 V is shown for 1 subject.
Averages are taken for all captured pulses for the last 20 beats of
each pacing sequence.
|
|
The overall averages for Pmax and Parea across all experiments
for every voltage level are shown in Tables
1 and 2,
respectively. The data shown are in three parts: mean percent
improvement, SD, and P value (for pairwise
t-test). In general, we found that a higher Pmax was
generated due to anodal stimulation, as well as a larger integral under
the pressure wave. From Tables 1 and 2, respectively, we see that Pmax
is, on average, ~4-6% greater with anodal pacing, and Parea is
~10-12% greater with anodal pacing. Figure
2 gives insight into the statistically
significant range of improvement due to anodal pacing for Pmax at 4 V. The data displayed are the average improvements for all the
acquisitions in each experiment with the corresponding SD (represented
in percentages), as well as the P value results of the
pairwise t-test. The pairwise t-test failed to
show significance in experiments 1 and 6 of Fig. 2. This can be attributed to the fact that too few data points were
evaluated by the t-test in these experiments, which is due to the fact that the heart was only able to capture for two or three
acquisitions per stimulation level. During the experiments, capture and
noncapture were determined qualitatively for obvious cases, and, during
data analysis, our software detected noncapture.

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 2.
Average percent improvement [(A C)/C, where A is anodal
and C is cathodal] in Pmax due to anodal pacing at 4 V in 8 separate experiments. Shown are averages of each experiment at 4 V. The
averages, SDs, and P values for pairwise t-test
are also shown. An average improvement of 3-5% at 4 V can be
seen. P < 0.05 implies statistically significant
differences due to anodal pacing. SDs are due to weakening of the heart
over time during the experiment.
|
|
Variances in the data were analyzed by means of the ANOVA test
(repeated measures) for two factors: waveforms (within experiments) and
subject (between experiments). Waveform was found to be a statistically
significant factor, and the results of this test are shown in the last
columns of Tables 1 and 2 for Pmax and Parea, respectively. The data
shown in these columns are represented as the average percent
improvement across all experiments for each voltage level, the SD, and
the P values resulting from the ANOVA test discussed above.
One can see that waveform is in fact a statistically significant factor
for all parameters at all voltages (P < 0.05). These
data are shown graphically in Figs. 3
(Pmax) and 4 (Parea). The results of
the ANOVA test showed that subject was not a significant factor. We
conclude from these tests that the improvements in Pmax and Parea
reported in Tables 1 and 2, respectively, are not due to the variances
in subjects or experimental conditions but, rather, are due to the
changing of the pacing waveform.

View larger version (18K):
[in this window]
[in a new window]
|
Fig. 3.
Comparison of Pmax improvement [(A C)/C] due to
anodal pacing ranging from 2 to 8 V. Averages over all experiments for
each stimulus voltage level are shown. An average of 3-5%
improvement in Pmax for all experiments regardless of stimulus voltage
level was observed. Multivariated ANOVA test was run for the following
factors: subjects and waveforms. It was shown that subject (differences
between experiments) was not a significant factor (P
0.05). P values for waveforms (differences within
experiments) are shown. P < 0.05 implies statistically
significant differences due to anodal pacing. P < 0.05 for the ANOVA test proves that the pacing waveform is the factor
responsible for change in Pmax and not the differences in between
experiments. SDs are a result of variation in heart strengths across
the different subjects.
|
|

View larger version (19K):
[in this window]
[in a new window]
|
Fig. 4.
Comparison of pressure wave area improvement [(A C)/C]
due to anodal pacing. Averages over all experiments for each stimulus
voltage level are shown. An average of 3-5% improvement in Pmax
for all experiments regardless of stimulus voltage level was observed.
Multivariated ANOVA test was run for the following factors: subjects
and waveforms. It was shown that subject (differences between
experiments) was not a significant factor (P 0.05).
P values for waveforms (differences within experiments) are
shown. P < 0.05 implies statistically significant
differences due to anodal pacing. P < 0.05 for the
ANOVA test proves that the pacing waveform is the factor responsible
for change in work performed by the left ventricle and not the
differences in between experiments. SDs are a result of variation in
heart strengths across the different subjects.
|
|
 |
DISCUSSION |
Most present day pacemakers use either unipolar cathodal (negative
polarity) or bipolar pacing, where the cathode is closer to the
myocardium and thus is responsible for most of the stimulation. It is
also known that anodal pacing has higher excitation thresholds and a
potential for proarrhythmia. A period of vulnerability for arrhythmia
begins at the end of the refractory period and terminates later in the
cardiac cycle, and Mehra et al. (12) showed that anodal
stimulation has a shorter refractory period than cathodal stimulation.
Thus the period of vulnerability is longer for anodal stimulation. It
has been indicated that differences between arrhythmia vulnerability to
various stimulation waveforms are dependent on their excitability
characteristics, such as strength-interval curves (11,
13).
Another area of active interest in applying pacing and pressure
waveforms is the multisite-pacing experiment in which the anode is used
as the stimulus side. In congestive heart failure models, some signs of
increase in effectiveness of myocardium contraction have been shown
(1, 2, 7, 18).
However, not much has been done to correlate pacing to the heart's
electromechanical coupling, contractility, and other mechanical work.
This study presents new data on the effect of pacing waveform on the
heart's mechanical response. The results from our study show that the Pmax generated and the Parea vary with changing waveforms. On average,
a 4-6% improvement is seen in the Pmax generated by the LV due to
anodal pacing, and a 10-12% improvement is seen in the average
Parea due to anodal pacing.
The mechanism of anodal pacing-based stimulation and mechanical
performance enhancement in hearts is not fully understood. The question
arises as to how anodal stimulation can lead to stronger myocardial
contraction. At rest, cardiac myocytes exhibit closed sodium and
calcium channels that are slightly inactivated. However, during
hyperpolarization, these channels become less inactivated, thereby
permitting a larger current flow after stimulation (6). Presumably, faster upstroke allows for stronger contraction as more
calcium channels open sooner and for a longer period of time, promoting
the contraction cascade (8). Also, because intracellular calcium is at very low concentrations, small improvements are largely
amplified by the calcium-triggered calcium-release mechanism. Thus,
despite larger activation thresholds, it is conceivable that anodal
stimulation provides a measurable advantage in myocardial excitation-contraction coupling. Furthermore, because calcium is a
sensitive intracellular signal, accumulation over longer periods of
time may modulate contractility (8).
Extensive prior studies have shown that, during anodal stimulation by
bipolar electrodes, a "dog bone" shaped region of the cardiac
tissue under the stimulating electrodes becomes hyperpolarized. At the
same time, the tissue lying in the convexity of the dog bone becomes
depolarized and is commonly referred to as "virtual cathodes"
(10, 15, 19). Generally, it is
believed that the excitation wavefront of anodal stimulation propagates
from these virtual cathodes. Therefore, differences in anodal and
cathodal responses can be attributed in part to such virtual electrode stimulation in the case of bipolar pacing. This might be applicable in
unipolar pacing as well. Employing optical imaging, Knisley (9) showed in rabbit hearts that unipolar anodal
stimulation consistently produced early excitation at spots away from
the electrode on the fast (fiber longitudinal) propagation axis. At those spots, polarization was found to be positive for anodal stimulation, which is the opposite of polarization immediately adjacent
to or under the electrode. It is possible that such a virtual electrode
might alter conduction not only locally, but globally as well. Also,
our group previously showed that anodal stimulation enhances the
conduction velocities of the myocardium (18). An increase
in conduction velocity invariably derives from a faster upstroke due to
stronger inward ionic current. This observation was verified in our
earlier studies, in which APs recorded with the use of floating
microelectrodes showed a faster upstroke (dV/dt) for anodal
vs. cathodal stimulation (18). A faster upstroke may
result from increased intrinsic sodium channel activity, decreased
outward K+ current unitary conductance, or a recruitment or
decruitment of novel ionic channels (15,
16). A recent study presented the existence of a
novel hyperpolarization-induced potassium channel. Computer simulations
incorporating this channel kinetics do indeed show altered conduction
spread as a result of anodal pacing (14). We hypothesize
that these faster conduction velocities resulting from anodal
stimulation would alter the mechanical contraction of the heart. We
suggest here that increased conduction velocity may lead to higher
contractility and thus an increase in mechanical response produced by
the anodal stimulation. However, we do not have any data supporting the
hypothesized link between conduction and the observed mechanical
response. Considerable fundamental research remains to be done to
connect these possible relationships.
Limitations.
The studies reported here, as well as our previous preliminary studies,
were done in isolated, Langendorff-perfused hearts. Several
experimental conditions may have contributed to the variances in the
data. Some factors causing variances in our data included the inherent
condition of some hearts compared with others, because some animals
were in better physical condition than others. Over the course of the
experiment, we observed that the absolute Pmax generated by the LV
varied between 10 and 25 mmHg. This suggests the possibility that the
heart was weakening over time or previous damage could have occurred in
the heart. In some experiments, toward the end of the study,
the heart would be expected to be ischemic. We also noticed that anodal
pacing still showed a significant improvement in performance. This is
an intriguing subject for future studies to show the effects of anodal
pacing under ischemic conditions.
We have so far reported electrophysiological and pressure recordings
only from isolated hearts. Studies in in situ anesthetized, as well as
nonanesthetized, animals need to be done next. More comprehensive
measurements, including cardiac output, under different loading
conditions are warranted. Implications of our work will be most
relevant if benefits can be shown in diseased, ischemic, or failed hearts.
This study would be much more comprehensive if it examined the precise
depolarization patterns with the two stimulation waveforms studied.
Isochrone maps obtained through extracellular electrical recordings or
optical methods would shed light on the shape of the electric field
during stimulation and, later, the activation front during propagation.
The promising findings presented here create a need to review the
benefits of anodal pacing to enhance mechanical contractility. Benefits
of this therapy would be even more appreciated if electrical pacing is
shown to enhance heart performance in ischemia or heart failure. In
conclusion, use of stimulation pulses involving an anodal component may
offer a way for implanted pacemakers to enhance the electromechanical
response of the heart. This line of research seeks ultimately to expand
the potential applications of pacemaker therapy.
 |
FOOTNOTES |
Address for reprint requests and other correspondence: N. V. Thakor, Dept. of Biomedical Engineering, Johns Hopkins School of
Medicine, 720 Rutland Ave., Baltimore, MD 21205 (E-mail:
nthakor{at}bme.jhu.edu).
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Received 7 June 1999; accepted in final form 27 April 2000.
 |
REFERENCES |
1.
Bakker, P,
Sen KCA,
De Jonge N,
Klopping C,
Algra A,
De Medina R,
and
Bredee J.
Biventricular pacing improves functional capacity in patients with end-stage congestive heart failure (Abstract).
Pacing Clin Electrophysiol
18:
825,
1995.
2.
Bakker, PF,
Meijburg H,
De Jonge N,
Van Mechelen R,
and
Wittkampf F.
Benefical effects of biventricular pacing in congestive heart failure (Abstract).
Pacing Clin Electrophysiol
17:
820,
1994.
3.
Cranefield, PF,
Hoffman BF,
and
Siebens AA.
Anodal excitation of cardiac muscle.
Am J Physiol
190:
383-390,
1957.
4.
Dekker, E.
Direct current make and break thresholds for pacemaker electrodes on canine ventricle.
Circ Res
27:
811-823,
1970[Abstract/Free Full Text].
5.
Harttung, WM,
Lucet FH,
McTeague K,
Honeycutt C,
and
Langberg JJ.
Overlapping biphasic stimulation: a novel pacing mode with low capture thresholds (Abstract).
Circulation
90:
A365,
1994.
6.
Hoffman, BF,
and
Cranefield PF.
Electrophysiology of the Heart. Mount Kisco, New York: Futura, 1976.
7.
Hummel, JD,
Davis JH,
and
Mower MM.
Augmentation of cardiac output by anodal pacing (Abstract).
Circulation
90:
A366,
1994.
8.
Katz, AM.
Physiology of the Heart. New York: Raven, 1992.
9.
Knisley, SB.
Transmembrane voltage changes during unipolar stimulation of rabbit ventricle.
Circ Res
77:
1229-1239,
1995[Abstract/Free Full Text].
10.
Knisley, SB,
Hil BC,
and
Ideker RE.
Virtual electrode effect in myocardial fibers.
Biophys J
66:
719-728,
1994[Web of Science][Medline].
11.
Mehra, R,
and
Furman S.
Comparison of cathodal, anodal, and biopolar strength-interval curves with temporary and permanent pacing electrodes.
Br Heart J
41:
468-476,
1979[Free Full Text].
12.
Mehra, R,
Furman S,
and
Crump JF.
Vulnerability of the mildly ischemic ventricle to cathodal, anodal, and bipolar stimulation.
Circ Res
41:
159-166,
1977[Abstract/Free Full Text].
13.
Preston, TA.
Anodal stimulation as a cause of pacemaker induced ventricular fibrillation.
Am Heart J
86:
366-372,
1973[Web of Science][Medline].
14.
Rajasekhar, SSV
A study of cardiac pacing: effects of stimulation on electrical and mechanical performance of heart (Masters thesis).
In: Biomedical Engineering Department. Baltimore, MD: Johns Hopkins University, 1996, p. 194.
15.
Ranjan, R.
Mechanism of anodal stimulation (Ph.D. thesis).
In: Biomedical Engineering Department. Baltimore, MD: Johns Hopkins University, 1997, p. 107.
16.
Ranjan, R,
Chiamvimonvat N,
Thakor NV,
Tomaselli GF,
and
Marban E.
Mechanism of anode break stimulation in the heart.
Biophys J
74:
1850-1863,
1998[Web of Science][Medline].
17.
Saxon, L,
Natterson PD,
DeLurgio DB,
Stevenson WG,
Drinkwater DC,
Mower MM,
and
Thomas A.
Bi-ventricular pacing may provide inotropic support in refractory heart failure.
J Investig Med
43:
198A,
1995.
18.
Thakor, N,
Ranjan R,
Rajasekhar S,
and
Mower MM.
Effect of varying pacing waveform shapes on propagation and hemodynamics in the rabbit heart.
Am J Cardiol
79:
36-43,
1997[Medline].
19.
Wikswo, JP,
Wisialowski TA,
Altemeir WA,
Balser JR,
Kopelman HA,
and
Roden DM.
Virtual cathode effects during stimulation of cardiac muscle. Two-demensional in vivo-experiments.
Circ Res
68:
513-30,
1991[Abstract/Free Full Text].
J APPL PHYSIOL 89(3):1159-1164
8750-7587/00 $5.00
Copyright © 2000 the American Physiological Society