|
|
||||||||
1 Medical and 2 Surgical Intensive Care Unit and 3 Department of Clinical Physics, Institute for Cardiovascular Research, Free University Hospital, 1081 HV Amsterdam, The Netherlands
| |
ABSTRACT |
|---|
|
|
|---|
The purpose of this
study was to evaluate right ventricular (RV) loading and cardiac output
changes, by using the thermodilution technique, during the mechanical
ventilatory cycle. Fifteen critically ill patients on mechanical
ventilation, with 5 cmH2O of positive end-expiratory
pressure, mean respiratory frequency of 18 breaths/min, and mean tidal
volume of 708 ml, were studied with help of a rapid-response thermistor
RV ejection fraction pulmonary artery catheter, allowing 5-ml
room-temperature 5% isotonic dextrose thermodilution measurements of
cardiac index (CI), stroke volume (SV) index, RV ejection fraction (RVEF), RV end-diastolic volume (RVEDV), and RV end-systolic volume (RVESV) indexes at 10% intervals of the mechanical ventilatory cycle. The ventilatory modulation of CI and RV volumes varied from patient to patient, and the interindividual variability was greater for the latter variables. Within patients also, RV
volumes were modulated more by the ventilatory cycle than CI and SV
index. Around a mean value of 3.95 ± 1.18 l · min
1 · m
2 (= 100%), CI varied from
87.3 ± 5.2 (minimum) to 114.3 ± 5.1% (maximum), and RVESV
index varied between 61.5 ± 17.8 and 149.3 ± 34.1% of mean
55.1 ± 17.9 ml/m2 during the ventilatory cycle. The
variations in the cycle exceeded the measurement error even though the
latter was greater for RVEF and volumes than for CI and SV index. For
mean values, there was an inspiratory decrease in RVEF and increase in
RVESV, whereas a rise in RVEDV largely prevented a fall in SV index. We
conclude that cyclic RV afterloading necessitates multiple
thermodilution measurements equally spaced in the ventilatory cycle for
reliable assessment of RV performance during mechanical ventilation of patients.
right ventricular performance; ejection fraction catheter; critically ill; reliability of thermodilution
| |
INTRODUCTION |
|---|
|
|
|---|
IN CRITICALLY ILL PATIENTS, the measurement of right-sided cardiac output by using a pulmonary artery catheter and the thermodilution technique is commonly done (12). The catheter can also be equipped with a rapid-response thermistor, allowing bedside measurement of right ventricular (RV) ejection fraction and volumes (4, 6-10, 17, 25, 29, 35, 36). These latter measurements may also be of mechanistic, diagnostic, therapeutic, and prognostic significance in critically ill patients (4, 7, 12, 17, 29, 35, 36). For instance, RV end-diastolic volume may be a better predictor of preload-recruitable stroke volume by a fluid challenge than filling pressures so that a high volume may preclude a further rise in cardiac output with fluids, independently of filling pressures (3, 4, 7, 35, 36).
Nevertheless, the usefulness of pulmonary artery catheter insertion and thermodilution measurements of cardiac output and, especially, RV volumes has been doubted, partly because of insufficient accuracy, reproducibility, and predictive value for a response to fluid loading, particularly during mechanical ventilation (7, 30, 35, 36). Furthermore, there may be only partial agreement with RV volume measurements by other techniques in patients (9, 10, 17, 27, 36). Measurements are mostly performed at one phase in the mechanical ventilatory cycle, i.e., at the end of expiration, believed to be associated with the greatest reproducibility compared with injections at other phases of the ventilatory cycle, and outliers are usually excluded (1, 3, 4, 9, 10, 17, 20, 25-29, 31, 32, 35, 36). However, the timing of injectates in the ventilatory cycle is known to affect cardiac output measurements, irrespective of measurement errors, possibly via the ventilatory modulation of RV loading associated with cyclic changes in airway pressure and lung volume during mechanical ventilation (1, 10, 12, 14, 15, 17, 18, 24, 30, 32, 33).
Authors have recommended that, for a reliable estimation of mean thermodilution cardiac output, no specific phase in the ventilatory cycle should be selected, and that the best estimation resulted from averaging measurements at three or four equally spaced intervals in the cycle (1, 13-15, 24, 30, 33). Even though there are two (echocardiographic and thermodilution) studies suggesting changing RV volumes during the ventilatory cycle in mechanically ventilated patients, it is unclear how the modulation affects thermodilution RV ejection fraction (RVEF) and volumes (1, 18). Some authors advised measurement of RV volumes at apnea, even if this is not representative of RV performance during mechanical ventilation (1, 19). Finally, the potential mechanism, i.e., RV pre- or afterload changes, responsible for cardiac output modulations is unclear. In fact, lung inflation could increase RV afterload and volumes, or it could reduce RV preload and volumes and thereby cardiac ouput, as may occur during incremental positive end-expiratory pressure (PEEP) (10, 13, 16-18, 25, 34).
In consideration of the above data, we hypothesized that changes in RV loading, as assessed from thermodilution volume measurements, are responsible for the cardiac output modulation during the mechanical ventilatory cycle. Moreover, we wanted to quantify the effect and to assess its impact on reliable thermodilution measurements in the mechanical ventilatory cycle.
| |
METHODS |
|---|
|
|
|---|
Patients
. Informed consent was obtained from patients' relatives, and the
protocol was approved by the hospital Committee on Ethics. We
consecutively studied 15 critically ill patients who were in the
surgical intensive care unit and were on continuous volume-controlled positive-pressure ventilation (Siemens Servo 900B, Siemens Elema, Stockholm, Sweden) because of acute respiratory insufficiency. All
patients were on 5 cmH2O of PEEP, at an inspiratory time of 25%, an end-inspiratory hold of 10%, and an expiratory time of 65%
of the ventilatory cycle. All patients were sedated with continuous intravenous infusion of fentanyl and midazolam. The patients, without
known valvular incompetence, had sinus rhythm and were hemodynamically
stable. A radial artery catheter had been inserted for measurement of
the mean arterial blood pressure (mmHg). A thermodilution pulmonary
artery catheter, equipped with a rapid-response thermistor (model
93A-431H-7.5F, Baxter Edwards, Santa Ana, CA; response time 50 ms) and
intracardiac electrodes, was inserted percutaneously via the jugular or
subclavian vein until the inflated balloon wedged in a pulmonary artery
and the proximal injectate port recorded RV pressure. The catheter was
withdrawn thereafter to locate the injectate port just above the
tricuspid valve (31). The port was located 21 cm from the
tip. Hemodynamically significant tricuspid regurgitation was ruled out
in each patient on the basis of absence of V waves in the right atrial
pressure recording (27). The injectate temperature was
measured by an in-line temperature probe, distally from the injection
site (model 93-600 CO-set, Baxter Edwards). The rapid-response
thermistor, analog electrocardiograph signal, and the injectate
temperature probe were interfaced to the REF-1 computer for signal
processing according to in-built algorithms (Baxter Edwards; Refs. 7,
9). The heart rate (HR) was determined from the
electrocardiograph signal, and the computer detected the R waves. The
first-order exponential downslope of the thermodilution curve was used
by the computer to calculate the residual fraction (RF) from the
relationship between successive temperature plateaus synchronized to
the R wave (Fig. 1). The successive
values were averaged, and the RVEF was calculated as 1
RF.
Cardiac output was calculated by integrating the temperature change of
the blood. Cardiac index (CI) is cardiac output divided by body surface
area calculated from height and weight. Stroke volume (SV) index is
derived from CI/HR. The RV end-diastolic volume (RVEDV) index
(ml/m2) was calculated as SV index/RVEF. The RV
end-systolic volume (RVESV) index was calculated as RVEDV index
SV index.
|
Protocol
. Demographic and clinical features were recorded. The Simplified Acute
Physiology Score and Lung Injury Score (LIS) were assessed
(11, 22). The latter ranges from 0 (no acute
lung injury) to 4, with values above 2.5 indicating the adult
respiratory distress syndrome. The mean arterial pressure, right atrial
pressure, mean pulmonary arterial pressure, and pulmonary artery
occlusion pressure were measured with patients in the supine position,
with the midchest level as reference, and after calibration.
The tidal volume, respiratory rate, and peak and plateau airway
pressures were recorded. Total respiratory compliance was calculated as tidal volume/(plateau airway pressure
PEEP). Arterial blood was
obtained for determination of blood-gas values. Injection of 5 ml of
5% dextrose at room temperature was automatically performed by a phase
controller and a pneumatically driven syringe after manual start of the
cardiac output computer (14). The 5-ml bolus was injected
at 50 psi within 1 s, by using a power injector, through the
pulmonary artery catheter. After 12 s the syringe was
automatically refilled. The moment of injection was dependent on a
start signal given by the operator and the moment in the ventilatory
cycle set on the phase controller. The latter was derived from the
Siemens Servo 900B ventilator, which delivers 100 impulses during each
ventilatory cycle. The phase zero was at the start of inflation. The
injections (n = 11) were performed successively at 10%
intervals of the ventilatory cycle, at stable hemodynamics. In 2 patients, 10 successive 5-ml injections were done at 30% of the
ventilatory cycle to evaluate measurement error.
Calculations and statistical analyses . The means of the 11 thermodilution curve-derived parameters were calculated per patient to normalize results. The Wilcoxon rank-sum test was used to evaluate differences between group means at the phases and 100%. A group coefficient of variation (CV = SD/mean) was calculated for each variable, normalized by individual means, at each phase of the ventilatory cycle, to evaluate interindividual differences. ANOVA was used to evaluate phase and variable differences in CV. In each patient, a CV was also calculated for each variable, normalized by individual means, to evaluate intraindividual differences. ANOVA was done to compare variables with respect to the CV. When factors contributing to the observed variance (CVobs2) are independent from each other, then the observed variance is the sum of the contributing variances (21). The CV caused by the ventilatory cycle alone (CVresp) was thus assessed from CVobs2 = CVresp2+ CVmeas2, where CVmeas2 is the variance caused by measurement error. Measurement error (reproducibility) was assessed from pooled data obtained by 10 injections at 30% of the ventilatory cycle in 2 patients and by the injections at 0 and 100% of the cycle, denoting the same phase, in all patients. Linear correlation analysis on pooled normalized data was done to evaluate the contribution of RV volume changes to CI changes. A random generator was used to calculate, in each patient, the minimum number of random observations that need to be made from a series of observations to get a reliable estimate of the mean value of that series, with 95% or greater chance to be within 10, 20, and 30% limits of the mean value. Data are expressed as means ± SD, and P < 0.05 was considered statistically significant.
| |
RESULTS |
|---|
|
|
|---|
General
. Table 1 shows the main features of the
patients. All patients had acute lung injury, and one of them had adult
respiratory distress syndrome (LIS = 2.5). Table
2 describes baseline global hemodynamic
and respiratory variables.
|
|
|
|
|
0.84,
P < 0.001) and RVESV (r =
0.69,
P < 0.001) so that the latter two interrelated
positively (r = 0.69, P < 0.001). This
indicates that the ventilatory modulation of CI was largely caused by
changes in SV index, and that changes in both RVEF and RVEDV
contributed to changes in SV index. Hence, ventricular dilation, i.e.,
a rise in RVEDV prevented a fall in SV during a fall in the EF of the
RV contracting toward an increased RVESV index.
Error analysis
. At a measurement error shown in second column of Table
4, and the mean observed CVs of Table 3,
it was calculated (see formula in Calculations and statistical
analyses) that the mean CV of ventilatory modulation, independent
of measurement error, was 7.2% for CI, 6.4% for SV index, 13.7% for
RVEF, 13.5% for RVEDV and 24.9% for RVESV indexes.
Otherwise, the measurement error did not depend on the phase in the
mechanical ventilatory cycle. For all ventilatory phases together and
after correction for measurement error (second column of Table 4), the
mean CV for interindividual variation in ventilatory modulation was
6.5% for CI and 24.4% for RVESV index. The above data indicate that the ventilatory cycle modulated RV volumes more than CI and that interindividual differences in the modulation were also greater for the
former, irrespective of measurement error.
|
Prediction of minimum number of at-random measurements for reliable
RV volume assessments
. In Table 5, the minimum number of
at-random measurements for each variable during the ventilatory cycle,
necessary for an estimate within certain limits of the mean value, is
shown. Five to eight measurements are necessary, at minimum, to
reliably assess the RVEF, RVEDV, and RVESV and four for measurements of CI and SV index.
|
| |
DISCUSSION |
|---|
|
|
|---|
We show that the modulation of RV volumes and CI by the mechanical ventilatory cycle is greater than the measurement error and that the modulation of RV volumes is greater than that of CI. The difference in modulation between patients is greater for RV volumes than for CI. As judged from group means, RV afterload may rise during lung inflation and this may induce a fall in RVEF and rise in RVESV. A fall in SV is largely prevented by a rise in RVEDV, thereby explaining greater modulation of RV volumes than of CI in the mechanical ventilatory cycle. Because of ventilatory modulation of RV volumes and interindividual differences herein, assessment of RV performance by thermodilution requires multiple determinations at equally spaced intervals, or at least eight at random injections, in the ventilatory cycle.
Our results partly agree with those obtained by other investigators. The relatively large injectate volumes and high respiratory rates, which decrease the CI modulation, may partly explain less ventilatory modulation of CI in our study than in the studies by Assmann (1) and Jansen et al. (12-15). Our finding of the tendency for an expiratory fall in CI agrees with the literature (12-15, 19). Even though positive-pressure inflation generally decreases right-sided thermodilution cardiac output, the modulation may be dissimilar among patients in phase and amplitude so that in some patients CI may also increase during lung inflation (1, 12-15, 24, 30). Hence, no specific phase in the ventilatory cycle could be selected for a reliable estimation of mean CI over the ventilatory cycle in all patients. The phase and amplitude of the modulation by the ventilatory cycle may depend on the volume status and absolute blood flow on the one hand and on respiratory variables on the other (1, 12, 13, 15, 30). The fact that modulation may be affected by multiple factors may explain why we could not predict the interindividual differences in hemodynamic modulation by the ventilatory cycle in our patients.
Our data extend those obtained by Assmann et al. (1). They showed that the mechanical ventilatory cycle modulated RVEF and volumes more than CI, when assessed at 0.25 equally spaced fractions of the cycle (1). The assessment of RVEF and volumes by thermodilution was more reproducible during apnea than during mechanical ventilation, and the modulation was lower at higher respiratory rates (1). The intraindividual variability of RVEDV was 11.6% at a respiratory rate of 16 breaths/min (1). The higher variability at a comparable respiratory rate in our study can be explained by the larger number of measurements in the ventilatory cycle and lower injectate volumes (5 ml) than in their study (10 ml). The 5-ml boluses were used to limit fluid overload.
The measurement error for CI in our study may be somewhat lower than that reported before, in which repeated (manual) injections at the same phase in the ventilatory cycle were associated with a CV of 5-10%. Manual injections, however, may be more erroneous than automated ones (1, 10, 23, 29, 33). In vitro, the measurement error of CI and SV may be ~3%, whereas the error of RVEF, RVEDV, and RVESV ranges between 5 and 7% (8). In vivo also, repeated manual (phase-selected) injections in the mechanical ventilatory cycle have revealed greater error for RVEF and volumes than for CI measurements, in agreement with our results (1, 10, 20, 23, 26, 28, 29). In agreement with other investigators (1, 12, 14, 15, 30), we show that, if injections at equal intervals in the ventilatory cycle are impossible, averaging at least four random measurements in the cycle is an adequate strategy to estimate mean CI reliably in patients on mechanical ventilation. Because of greater modulation and measurement error, the minimum number of at-random determinations, necessary to yield an estimate of mean RVEF and volumes over the ventilatory cycle within a certain error, was higher than for CI assessments.
The group means over the ventilatory cycle suggest that lung inflation resulted in a rise in RVEDV and RVESV indexes after a rise in RV afterload, even though the actual thermodilution measurements of RV volumes may have taken place some time after inflation. A delay between lung inflation and actual measurements after injection implies that the rise in volumes could also have been caused by increased filling of the RV, after a reduced intrathoracic pressure and increased venous return, during expiration. This is unlikely, however, because increased filling would not decrease the ejection fraction and would tend to increase the SV (7, 10). This suggests that the volumes measured from injections in the inspiratory phase indeed reflected inspiratory events so that a transient rise in afterload resulted in a fall in RVEF and a rise in RVEDV and RVESV, attempting to maintain SV index. The latter agrees with the transmural pressure measurements and echocardiographic data obtained by Jardin et al. (17, 18) in mechanically ventilated patients. They observed that lung inflation was associated with a rise in RV transmural pressures and volumes after a rise in afterload. Conversely, the patterns of RV volume changes in our study resemble those during preload reductions with PEEP or increases with military antishock trouser inflation (7, 25), resulting in similar decreases and increases, respectively, in RVEDV and RVESV, thereby hardly affecting SV. We cannot exclude RV contractility fluctuations during the cycle, in the absence of end-systolic volume-transmural pressure relationships (7, 17, 20).
The cyclic changes in RV volumes cannot be explained by cyclic tricuspid regurgitation. The disparate rather than parallel fall of RVEF and CI in the inspiratory phase in our study argues against cyclic tricuspid regurgitation. Alternatively, an inspiratory fall in baseline pulmonary blood temperature of ~0.01-0.02°C could lead to a temporary overestimation of ~3% of thermodilution CI (12, 15). We did not observe fluctuations in the baseline temperature, and the CI did not rise during inspiration (Fig. 1). Finally, the distance between the injectate port and thermistor relative to the tricuspid and pulmonary valves, respectively, may affect the absolute values, but not the changes, of RVEF and RV volumes (31, 33).
Although experiments in animals generally show a predominant preload-lowering effect of positive-pressure ventilation, as evidenced by a fall in RV volumes (10), clinical studies, using echocardiography, nuclear angiography, or thermodilution, showed either a fall in RV preload (fall in volumes) or a rise in afterload (rise in volumes) during incremental PEEP ventilation. This seems independent of the measurement technique but dependent, in part, on baseline RV performance and thus underlying disease (4, 6, 17-19, 25). For instance, thermodilution RV volumes may decrease up to 25 cmH2O of incremental PEEP in some studies and may increase in other human studies, particularly when baseline RVEF was decreased and RVEDV increased after coronary artery disease or acute lung injury (1, 6, 17-19, 25). A rise in RV afterload during lung inflation, as suggested in our study, may not exclude a fall in preload during incremental PEEP ventilation, if the former is largely determined by an increased pulmonary air volume and vascular resistance and the latter by an increased intrathoracic pressure and decreased venous return (5). Finally, the effect of positive-pressure inflation may be time dependent, because an inspiratory hold in mechanically ventilated cardiac surgery patients may increase RVEDV and output only in the first 5 s (34).
In previous studies using the thermodilution method, SV, RVEF, and volumes were most often assessed at end expiration (9, 20, 25-29, 31-33, 36) and less often at end inspiration (10, 25, 26, 32). Our study indicates that this practice may have resulted in unpredictable under- and overestimations of RV performance in individual patients. It may also partly explain the reported controversy on effects of PEEP and on the value of RVEDV as a predictor of preload-recruitable SV (3, 7, 35). In fact, fluid loading and PEEP may alter the phase and amplitude of RV volume and output modulations by the ventilatory cycle (15) so that phase-selected assessments may preclude a reliable judgment of RV volume changes. The modulation by the mechanical ventilatory cycle may also explain why phase-selected thermodilution measurements did not always correlate well to other types of measurements of RV volumes and output (2, 15).
In conclusion, the mechanical ventilation in critically ill patients modulates RV afterloading in a cyclic manner, as assessed by thermodilution, so that the cyclic variation in RV volumes is greater than that of RV output. The cyclic modulation varies among patients. This has considerable impact on the timing of thermodilution measurements in the mechanical ventilatory cycle for the reliable assessment of RV performance.
| |
ACKNOWLEDGEMENTS |
|---|
We thank Dr. Jos Twisk for help with calculations and statistics.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: A. B. J. Groeneveld, Medical Intensive Care Unit, Free Univ. Hospital, De Boelelaan 1117, 1081 HV Amsterdam, The Netherlands (E-mail johan.groeneveld{at}azvu.nl or johangroeneveld{at}compuserve.com).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Received 27 October 1999; accepted in final form 24 February 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Assmann, R,
Heidelmeyer CF,
Trampisch H-J,
Mottaghy K,
Versprille A,
Sandmann W,
and
Falke KJ.
Right ventricular function assessed by thermo-dilution technique during apnea and mechanical ventilation.
Crit Care Med
19:
810-817,
1991[Web of Science][Medline].
2.
Cariou, A,
Monchi M,
and
Dhainaut J-F.
Continuous cardiac output and mixed venous oxygen saturation monitoring.
J Crit Care
13:
198-213,
1989.
3.
Chang, MC,
Blinman TA,
Rutherford EJ,
Nelson LD,
and
Morris JA.
Preload assessment in trauma patients during large-volume shock resuscitation.
Arch Surg
131:
728-731,
1996
4.
Cheatham, ML,
Nelson LD,
Chang MC,
and
Safcsak K.
Right ventricular end-diastolic volume index as a predictor of preload status in patients on positive end-expiratory pressure.
Crit Care Med
16:
1801-1806,
1998.
5.
Culver, BH,
Marini JJ,
and
Butler J.
Lung volume and pleural pressure effects on ventricular function.
J Appl Physiol
50:
630-635,
1981
6.
Dambrosio, M,
Fiore G,
Brienza N,
Cinnella G,
Marucci M,
Ranieri VM,
Greco MM,
and
Brienza A.
Right ventricular myocardial function in ARF patients. PEEP as a challenge for the right heart.
Intensive Care Med
22:
722-780,
1996[Web of Science][Medline].
7.
Dhainaut, J-F,
Pinsky MR,
Nouria S,
Slomka F,
and
Brunet F.
Right ventricular function in human sepsis. A thermodilution study.
Chest
1-12:
1043-1049,
1997
8.
Ferris, SE,
and
Konno M.
In vitro validation of a thermodilution right ventricular ejection fraction method.
J Clin Monit
8:
74-80,
1992[Web of Science][Medline].
9.
Globits, S,
Pacher R,
Frank H,
Pacher B,
Mayr H,
Neuhold A,
and
Glogar D.
Comparative assessment of right ventricular volumes and ejection fraction by thermodilution and magnetic resonance imaging in dilated cardiomyopathy.
Cardiology
86:
67-72,
1995[Web of Science][Medline].
10.
Groeneveld, ABJ,
Schreuder WO,
Vermeij CG,
and
Thijs LG.
Effect of incremental positive end-expiratory pressure ventilation on right ventricular function in anesthetized pigs: a thermodilution study.
J Crit Care
5:
218-227,
1990[Web of Science].
11.
Heffner, JE,
Brown LK,
Barbieri CA,
Harpel KS,
and
DeLeo J.
Prospective validation of an acute respiratory distress syndrome predictive score.
Am J Respir Crit Care Med
152:
1518-1526,
1995[Abstract].
12.
Jansen, JRC
The thermodilution method for the clinical assessment of cardiac output.
Intensive Care Med
21:
691-697,
1995[Web of Science][Medline].
13.
Jansen, JRC,
Schreuder JJ,
Boogard JM,
Van Rooyen W,
and
Versprille A.
Thermodilution technique for measurement of cardiac output during artificial ventilation.
J Appl Physiol
50:
584-591,
1981.
14.
Jansen, JRC,
Schreuder JJ,
Settels JJ,
Kloek JJ,
and
Versprille A.
An adequate strategy for the thermodilution technique in patients during mechanical ventilation.
Intensive Care Med
16:
422-425,
1990[Web of Science][Medline].
15.
Jansen, JRC,
and
Versprille A.
Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation.
Intensive Care Med
12:
71-79,
1986[Web of Science][Medline].
16.
Jardin, F,
Brun-Ney D,
Cazaux D,
Dubourg O,
Hardy A,
and
Bourdarias J-P.
Relation between transpulmonary pressure and right ventricular isovolumetric pressure change during respiratory support.
Cathet Cardiovasc Diagn
16:
215-220,
1989[Web of Science][Medline].
17.
Jardin, F,
Brun-Ney D,
Hardy A,
Aegerter P,
Beauchet A,
and
Bourdarias J-P.
Combined thermodilution and two-dimensional echocardiographic evaluation of right ventricular function during respiratory support with PEEP.
Chest
99:
162-168,
1991
18.
Jardin, F,
Delorme G,
Hardy A,
Auvert B,
Beauchet A,
and
Bourdarias J-P.
Reevaluation of hemodynamic consequences of positive pressure ventilation: emphasis on cyclic right ventricular afterloading by mechanical lung inflation.
Anesthesiology
72:
966-970,
1990[Web of Science][Medline].
19.
Jellinek, H,
Krafft P,
Hiesmayer M,
and
Steltzer H.
Measurement of right ventricular performance during apnea in patients with acute lung injury.
J Trauma
42:
1062-1067,
1997[Web of Science][Medline].
20.
Jensen, PJ,
Andersen PK,
and
Thøgersen C.
The reproducibility of measurement of right ventricular ejection fraction and cardiac output by the thermodilution technique in patients on mechanical ventilation.
Acta Anaesthesiol Scand
38:
486-489,
1994[Web of Science][Medline].
21.
King, RB,
Bassingthwaighte JB,
Hales JRS,
and
Rowell LB.
Stability of heterogeneity of myocardial blood flow in normal awake baboons.
Circ Res
57:
285-295,
1985
22.
Le Gall, JR,
Lemeshow S,
and
Saulnier P.
A new Simplified Acute Physiology Score (SAPS II) based on a European/North American Multicenter study.
JAMA
270:
2957-2963,
1993
23.
Lichtwarck-Aschoff, M,
Leucht S,
Kisc HW,
Zimmerman G,
Blümel G,
and
Pfeiffer UJ.
Monitoring of right ventricular function using a conventional slow response thermistor catheter.
Intensive Care Med
20:
348-353,
1994[Web of Science][Medline].
24.
Okamoto, K,
Komatsu T,
Kumar V,
Sanchala V,
Koubal K,
Bhadolia R,
and
Shibutani K.
Effects of intermittent positive pressure ventilation on cardiac output measurements by thermodilution.
Crit Care Med
14:
977-980,
1986[Web of Science][Medline].
25.
Pinsky, MR.
Effect of positive end-expiratory pressure on right ventricular function in humans.
Am Rev Respir Dis
146:
681-687,
1992[Web of Science][Medline].
26.
Rafferty, T,
Durkin M,
Hines R,
Elefteriades J,
Harris SN,
and
O'Connor TZ.
Thermodilution right ventricular ejection fraction measurement reproducibility
a study in patients undergoing coronary artery bypass graft surgery.
Crit Care Med
20:
1524-1528,
1992[Web of Science][Medline].
27.
Rafferty, T,
Durkin M,
Hines RL,
Elefteriades J,
and
O'Connor TZ.
The relationship between "normal" transesophageal color-flow Doppler-defined tricuspid regurgitation and thermodilution right ventricular ejection fraction measurements.
J Cardiol
7:
167-174,
1993.
28.
Safcsak, K,
and
Nelson LD.
Thermodilution right ventricular ejection fraction measurements: room temperature versus cold temperature injectate.
Crit Care Med
22:
1136-1141,
1994[Web of Science][Medline].
29.
Schreuder, WO,
Schneider AJ,
Groeneveld ABJ,
and
Thijs LG.
Effect of dopamine vs norepinephrine on hemodynamics in septic shock. Emphasis on right ventricular performance.
Chest
95:
1282-1288,
1989
30.
Snyder, JV,
and
Powner DJ.
Effects of mechanical ventilation on the measurement of cardiac output by thermodilution.
Crit Care Med
10:
677-682,
1982[Web of Science][Medline].
31.
Spinale, FG,
Zellner JL,
Mukherjee R,
and
Crawford FA.
Placement considerations for measuring thermodilution right ventricular ejection fractions.
Crit Care Med
19:
417-421,
1991[Web of Science][Medline].
32.
Stevens, JH,
Raffin TA,
Mihm FG,
Rosenthal MH,
and
Stetz CW.
Thermodilution cardiac output measurement. Effects of the respiratory cycle on its reproducibility.
JAMA
253:
2240-2242,
1985
33.
Thrush, DN,
and
Varlotta D.
Thermodilution cardiac output: comparison between automated and manual injection of indicator.
J Cardiothorac Vasc Anesth
6:
17-19,
1992[Medline].
34.
Van den Berg, PCM,
Grimbergen CA,
Spaan JAE,
and
Pinsky MR.
Positive pressure inspiration differentially affects right and left ventricular outputs in postoperative cardiac surgery patients.
J Crit Care
12:
55-65,
1997.
35.
Wagner, JG,
and
Leatherman JW.
Right ventricular end-diastolic volume as a predictor of the hemodynamic response to a fluid challenge.
Chest
113:
1048-1054,
1998
36.
Yu, M,
Takiguchi S,
Takanishi D,
Myers S,
and
McNamara JJ.
Evaluation of the clinical usefulness of thermodilution volumetric catheters.
Crit Care Med
23:
681-686,
1995[Web of Science][Medline].
This article has been cited by other articles:
![]() |
M. Engoren and D. Barbee Comparison of Cardiac Output Determined by Bioimpedance, Thermodilution, and the Fick Method Am. J. Crit. Care., January 1, 2005; 14(1): 40 - 45. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. MINEO, E. POMPEO, P. ROGLIANI, M. DAURI, F. TURANI, P. BOLLERO, and N. MAGLIOCCHETTI Effect of Lung Volume Reduction Surgery for Severe Emphysema on Right Ventricular Function Am. J. Respir. Crit. Care Med., February 15, 2002; 165(4): 489 - 494. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |