|
|
||||||||
1 Lovelace Respiratory Research Institute, and 2 New Mexico Resonance, Albuquerque, New Mexico 87108; and 3 Department of Physics, University of Pittsburgh, Pittsburgh, Pennsylvania 15260
| |
ABSTRACT |
|---|
|
|
|---|
We
partially obstructed the left bronchi of rats and imaged an inert
insoluble gas, SF6, in the lungs with NMR using a technique that clearly differentiates obstructed and normal
ventilation. When the inhaled fraction of O2
is high, SF6 concentrates dramatically in regions of the
lung with low ventilation-to-perfusion ratios (
A/
);
therefore, these regions are brighter in an image than where
A/
values are
normal or high. A second image, made when the inhaled fraction of
O2 is low, serves as a reference because the
SF6 fraction is nearly uniform, regardless of
A/
. The quotient of the first and second images displays the
low-
A/
regions and
is corrected for other causes of brightness variation. The technique
may provide sufficient quantification of
A/
to be a useful
research tool. The noise in the quotient image is described by the
probability density function for the quotient of two normal random
variables. When the signal-to-noise ratio of the denominator image is
>10, the signal-to-noise ratio of the quotient image is similar to
that of the parent images and decreases with pixel value.
perfusion; sulfur hexafluoride; lung; quotient; magnetic resonance imaging; probability density function; nuclear magnetic resonance
| |
INTRODUCTION |
|---|
|
|
|---|
REGIONS OF LUNGS WITH LOW alveolar ventilation-to-blood
perfusion ratios (
A/
)
are of great importance to understanding pathological lung function.
They are responsible for low oxygenation of arterial blood. In the
absence of pneumonia or other conditions that impose a diffusion
barrier between alveoli and blood capillaries, blood coming from areas
with normal or high
A/
is fully oxygenated as long as an air breather is near sea level.
Although there are methods of measuring overall
A/
heterogeneity in lungs (12, 23-25, 48, 49), there is a great
desire to spatially resolve the information. Ventilation images made
with gamma rays from inhaled 133Xe have low resolution, as
do perfusion images made with gamma rays from 99mTc in
aggregated albumin, which lodges in patent capillary beds.
The idea is to image an inert insoluble gas, such as SF6 or
C2F6, taking advantage of the fact that it
becomes concentrated where
A/
values are low (8,
14). The effect can be dramatic if the inhaled fraction of
O2
(FIO2) is high,
because in obstructed alveoli the soluble respiratory gases can come
close to equilibrium with venous blood, leaving the large
partial-pressure balance to the insoluble fluorinated gas. In contrast,
the gas in unobstructed alveoli is mostly O2. Figure
1 shows how the partial pressure
of SF6 in alveoli
(PASF6)
varies with
A/
(according to Refs. 8 and 14) with refinements (13) taking into account how
the solubility of O2 and CO2 in blood is
modified by the chemical environment of Hb (41). We assumed
that body temperature is 37°C, venous
PO2 is 40 Torr, venous
PCO2 is 46 Torr, base excess is
0, barometric pressure in our laboratory at 1,636-m elevation is 630 Torr, solubility of SF6 in blood is 0.0008, and
FIO2 is 75%. Figure
2 shows a comparison of
A/
discrimination for
different FIO2 values using the same calculations as in Fig. 1. For an
FIO2 of 75%, PASF6 is
doubled at a
A/
of 0.06 and tripled at a
A/
of 0.035 compared
with that for a normal
A/
of 0.84. Such two- and threefold changes can produce dramatic contrast in images.
|
|
Despite the overwhelming success of nuclear magnetic resonance (NMR) at imaging most soft tissues, lungs are problematic for two primary reasons. First, there is not much water in lungs to provide hydrogen nuclei to image. Second, water is diamagnetic, and the many gas-water boundaries make the magnetic field inhomogeneous (2, 10, 17). Spins precess at different frequencies and lose phase coherence within several milliseconds. Nonetheless, lung imaging has improved greatly in the last decade, with NMR pulmonary angiography approaching clinical utility (6, 7, 19, 22, 29, 42, 50, 51).
An alternative to imaging the water in lungs is to image a gas. Although the magnetic inhomogeneities are less severe in the gas spaces (1, 9, 10), imaging gases in lungs is still difficult because of the low density of nuclei. One way this problem has been overcome is to boost the signal by polarizing 3He or 129Xe to five orders of magnitude above the thermal equilibrium polarization (3, 4, 34). Because of prepolarization, one can make NMR images of 3He or 129Xe as rapidly as one can collect data and track gas during inhalation or exhalation to get information about ventilation. Recent images of emphysema look promising (30). However, the polarization is relaxed by O2 (39), lung tissue, and the imaging sequences (18, 26, 36) on the gases' journey to and from alveoli, and obtaining quantitative information is difficult.
Our technique uses inert fluorinated gases at thermal equilibrium polarization (31). Fluorinated gases were first imaged in lungs with NMR in the early 1980s (38). The way to obtain a good signal-to-noise ratio (SNR) is to choose gases with very rapid NMR relaxation, so that the magnetization recovers fast enough to allow extensive signal averaging. The present costs of 3He and 129Xe are two orders of magnitude higher than those of SF6. The cost of polarizing noble gases is declining. SF6, C2F6, CF4, cyclo-C4F8, and other good candidates for our methods are recoverable with commercial Freon liquefiers that allow O2 and N2 to escape. To recover 3He by condensation, one must liquefy or freeze O2 and N2 with more expensive apparatus.
An advantage of SF6, C2F6, and CF4 is that the spin-rotation relaxation is so rapid that the paramagnetism of O2 has no effect (16, 35, 37, 52). Alveoli are 25 times too large to cause the lengthening of relaxation that has been observed in small pores (33). The spin-lattice relaxation time constant T1 equals the spin-spin relaxation time constant T2 and is linearly related to the density of the gas mixture (16, 35, 37, 52); therefore, it is easy to calculate spin-density images and measure equilibrium gas concentrations in lungs.
To get quantitative information about
PASF6, we
need to correct for other causes of brightness variation in images,
such as radio frequency (RF) field inhomogeneities and gas-to-tissue
ratios. We made a reference image while
FIO2 was 20%; thus there was very little variation in
PASF6
regardless of
A/
(Fig. 1, curve B). By dividing an image made while
FIO2 was 75% by one made
while FIO2 was 20%, the
pixel values of the quotient image vary with
A/
as in Fig. 1,
curve C. The quotient image is bright where
A/
is low. The
nonlinearity implies that, if there are different
A/
values in
a region represented by one image pixel, the value of the pixel may
not be the average
A/
.
Fig 1, curve C, shows that there is discrimination in the range
of
A/
that
distinguishes obstructive lung diseases (5, 15, 47). How much a given
reduction in ventilation affects a mammal's gas exchange depends on
what fraction of the lung is obstructed. A criterion for whether the
lungs are exchanging gas adequately is whether the partial
pressures of O2 and CO2 in mixed arterial blood (PaO2 and
PaCO2, respectively) are normal. An
important regulatory mechanism is hypoxic pulmonary vasoconstriction
(e.g., Refs. 5, 44, 46). It reduces blood flow to obstructed regions with low PAO2, thereby
reducing the amount of hypoxic hypercapnic blood that contributes to
mixed arterial blood. Low PaO2 and
high PaCO2 at sea level result
when shifting blood flow to the better ventilated parts of lungs
fails to compensate. Thus a 40% reduction in ventilation to both lungs
will disrupt arterial blood gases, but complete obstruction of
one-tenth of a lung may not. Our method images
low-
A/
regions whether
or not they receive enough blood flow to lower
PaO2 and raise
PaCO2; however, if blood gases are so
altered, low-
A/
regions
are responsible, and they are what this technique is designed to image.
The quantitative relationships of obstruction,
A/
values, and blood
gases are subjects of ongoing research, and this technique
may be an additional noninvasive tool.
A matter that needs to be clarified is whether we can detect
A/
values that are low,
but higher than 0.1. Blood gases leaving a region with a
A/
of 0.5 are different
from normal and are of physiological relevance, but Fig. 1, curve
C, indicates a
A/
of 0.5 may be indistinguishable from a
A/
of 1. A factor that may be in our favor is that breathing a
high-FIO2 mixture will relax
hypoxic vasoconstriction, thus increasing blood flow and lowering
A/
values in obstructed
regions below their values for air breathing and into our detectable
range. Another possibility, yet untested, is that the high density of
SF6 may reduce ventilation to obstructed regions,
especially if it concentrates in them. In some cases of obstructive
lung disease, blood gases improve when patients breath low-density
He-O2 mixtures (20, 28, 43, 45). A possible mechanism is
that
A/
mismatch is improved by facilitating ventilation to obstructed regions. In any case, the relationship between air-breathing
A/
values vs. those for
high-FIO2 SF6
mixtures needs clarification and may serve to enhance the ability to
detect obstruction.
| |
METHODS |
|---|
|
|
|---|
Experimental obstructed ventilation.
We attempted to obstruct the ventilation in the left lungs of seven
sodium-pentobarbital-anesthetized, 350- to 450-g rats by inserting a
2.25 ± 0.03 mm outer diameter by 5.84 ± 0.003 mm length glass bead
into the left bronchus. The bead lumen contains a piece of
polypropylene tubing, and the combination has a resistance of 39 ± 3 Pa · s · cm
3
with a flow of air at 630-mmHg pressure and 21°C temperature. The
intent was to cause a nearly uniform obstruction of the left lung while
leaving the right lung unobstructed. This is difficult because the bead
must be lodged in the ringed portion of the left bronchus, which, in
most rats of this size, is the length of this bead. Distal to the
rings, the bronchus is more compliant, and the bead may not lodge in
one place or may block the openings of secondary bronchi. (The ringed
portion of the right bronchus is too short to avoid blocking the
secondary bronchi.) Although in prior practice procedures we could
reliably place the bead in the left bronchus, we had to wait until
after the imaging procedure to dissect the lungs and determine its
location. If it did not fit tightly in the ringed portion of the left
bronchus, it could dislodge and relocate. The bead was properly placed
in two of the seven attempts.
NMR imaging.
We imaged lungs by the methods of Kuethe et al. (31) with
modifications. In short, we collected free induction decays (FIDs) in
the presence of steady magnetic field gradients and varied the
direction of the gradient every other breath to obtain multiple one-dimensional projections. Initial experiments to develop techniques for imaging SF6, which has faster relaxation than the
C2F6 we had previously used, were performed in
I. J. Lowe's laboratory in Pittsburgh, PA, taking advantage of the
10-µs dead time between the end of RF transmission and beginning of
data collection. Short dead times and RF pulses are useful to minimize
the loss of imaging data and to reduce the loss of a rapidly decaying
signal. We then modified the NMR system in Albuquerque, NM, so that the
1.9-T magnet now has a Faraday shield that provides 30-dB attenuation to outside RF signals. The dead time, with a 400-g rat loading the
55-mm-inner-diameter quadrature bird cage coil (Morris Instruments, Gloucester, Ontario) is 8-12 µs, provided we bypass the audio stage filters in the Tecmag receiver, which requires digitizing the
signal at 1 MHz to capture the full bandwidth. Ninety-degree RF pulses
last 14-17 µs, depending on the size of the rat. We use a Miteq
AU-1114-BNC preamplifier, which has fast response and limits overload
voltage to the following Advanced Receiver Research 76-MHz narrowband
preamplifier, which filters out the upper sideband frequency. The noise
figure for the entire receiver chain is 1.19 F (dB), computed by
comparing the noise from a 76 K (liquid N2 cooled), 50-
resistor and a 294 K (room temperature) 50-
resistor.
Noise in the quotient image.
Because we are presenting a new technique for measuring relative gas
concentrations, it behooves us to quantify the measurement error. The
noise in typical NMR images is Gaussian for two reasons. 1) The
image is a discrete Fourier transform of data that have Gaussian noise
if the system is properly shielded. The Fourier transform of a Gaussian
function is Gaussian. 2) Even if the noise is non-Gaussian, the
Fourier transform sums large numbers of independent samples into each
pixel, which will produce essentially Gaussian noise. After the
magnitude of a complex image is taken, the noise associated with
low-value pixels is not Gaussian because the pixels are never negative.
However, masking magnitude images with a threshold value a few standard
deviations above zero to isolate the object ensures that the noise is
essentially Gaussian. To obtain the probability density function (PDF)
for the noise in a quotient of two images, let X be a Gaussian
random variable with mean µ1 and variance
21 and let Y be another
Gaussian random variable with mean µ2 and variance
22. Then X/Y has
PDF1
|
|
(1) |
exp (
t2)dt,
= µ1/µ2,
=
2/µ2, and
=
2/
1. If
1 =
2
|
|
(2) |
< 0.1, i.e., the
noise-to-signal ratio (NSR) of the denominator image is within reason.
Then the first (Cauchy or Lorentzian) term is essentially zero. In our
images,
is positive [for negative
, reverse right and left
in the rest of this paragraph]. The exponential factor of the
second term has the narrowest distribution and determines the variance.
It is essentially a Gaussian term with skew because of the variance
being a function of position. The error function (erf) factor rises
from 0 at z =
1/
2 to essentially 1 before z turns positive; thus it is essentially 1 over the
entire range of concern. The remaining factor is positive in the same region and has a broad distribution, peaking between z = 0 and z =
, so that its broad tail, descending
to the right, overlays the narrow exponential term, correcting some of
its skew to an essentially symmetric distribution centered around
,
with variance approximately
2(1 +
2
2)/
2 [or
1/µ22 (
21 + µ21/µ22
22)]. To get an accurate value, we integrate numerically to calculate the
expectation of the square. The standard deviation is the measurement error for individual pixels in our quotient images and is a function of
. It ranges from
/
to (
2 +
2/
2)1/2 (from
to

for
= 1) as our pixel
values range from 0 to 1. The distribution resembles a Gaussian
distribution, and the standard deviation quantifies variation like the
standard deviation of a Gaussian distribution in that 68-69% of
the area falls within ±1 SD and 95-96% falls within ±2 SD. In
our application, we typically know
and insert it to modify the
distribution. It is not easy to estimate
and
independently,
because the effect of
on the distribution is too similar to the
inverse of
. This paragraph's conclusions do not hold for at least
two cases of this PDF with large
.
The SNR of the quotient image is obtained from
, the NSR of the
denominator reference image. In addition to this
information, the PDF provides a good procedure for estimating
:
Signal average one-half as much for the reference image to obtain two
sets of data instead of one. From these two identically sampled sets of data, make two images, A and B. Use B to
determine the threshold mask, and apply the mask to A. Make a
histogram of the pixel values of A/B, and fit the histogram
with Eq. 2 to estimate
A/B and
A/B (
A/B will be very
close to 1). The
of the complete denominator image, made from the
combined data, is
A/B/
.
This procedure can be used for SNR calculations of images in general.
It has an advantage over the method of taking the mean of pixels in an
object and dividing it by the root-mean-square value of a region
containing no object. The object-free region of images often has
different noise than the object-containing region. The above procedure
estimates the noise in the region of the object.
| |
RESULTS |
|---|
|
|
|---|
Figure 3 is an image of obstructed
ventilation that has been corrected by using a reference image. The
image that detected the obstruction via
PASF6 was
made from 1.5 h of data collection while a rat breathed 75%
O2-25% SF6. It was divided by an appropriately scaled reference image made from 1 h of data collection when the same rat breathed 20% O2-80% SF6. Before
division, we used a mask to set the obstruction-detecting image to 0 in
the 301,226 out of 426,240 pixels where the reference image had pixel
values <330 (19.64% of the maximum pixel value 1,680.3), which
eliminated the pixels outside the lung.
|
Figure 4 shows a histogram of the pixel
values in Fig. 3. We eliminated the one pixel value, 1.037, out of 125,014 that fell >1.0; thus the color scale and histogram
represent values from 0 to the next highest pixel value, 0.9992. The
histogram is bimodal, with a narrow peak centered around 0.34, representing the unobstructed ventilation in the right lung, and a
broader peak centered around 0.52, representing the obstructed
ventilation in the left lung. The dotted and dashed lines are
histograms made from only the left or right lung, respectively.
|
To quantify the noise in Fig. 3, we first computed
for the
reference image and then used it to compute the standard deviation of
the PDF, Eq. 1, which gave us the standard deviation for the pixels in Fig. 3. By separating the data for the reference image into
two sets, thus making two images, and dividing one by the other and
vice versa, we obtained two estimates of
(0.05075 and 0.05061;
average 0.05068), the NSR of the reference image. Because we used three
sets (1.5 h) of data for the obstruction-detecting image and two sets
(1 h) for the reference image,
in Eq. 1
is
/
.
The standard deviation, from numerical integration of Eq. 1,
for
= 0.05068,
= 1.2247, and mean pixel value
= 0.342 is
0.0450, and for the same
and
and mean pixel value
= 0.520 is 0.0493. The standard deviation of the right lung histogram is
0.0487, slightly higher than 0.0450, indicating that most, but not all,
of its variation is measurement noise. The right lung histogram fits Eq. 1 better than a Gaussian, log normal, or Cauchy
(Lorentzian) distribution. The standard deviation in the left lung
histogram is 0.0990, twice as high as 0.0493, indicating much of the
variation is due to variation in SF6 concentration.
One can get a quick estimate of the image noise by examining the
left-hand tail of the histogram. Provided there is a substantial amount
of lung with near-normal
A/
values, it will all
theoretically translate to the same low pixel value. In other words,
physiology predicts a sharp left edge in the histogram. The left tail
of the histogram is then an estimate of the noise in measurement.
The image in Fig. 3 has some relaxation weighting because the
repetition time is ~1.7 times T1 (to maximize
SNR) and T1 is a function of gas density. To obtain
the correct spin-density information, we computed the dependence of
T1 on gas composition by assuming that the measured
T1 of SF6 for a normal rat applies to the gas composition calculated for alveoli with a normal
A/
. Then we
calculated the other T1 values from the
T1 values for two different inhaled gas
compositions, assuming T1 is a linear function of
gas density. We obtained a revised version of Fig. 1, curve C,
by using the steady-state magnetization to calculate pixel values from
gas composition. Projecting the pixel values through the revised curve
provides the second abscissa in Fig. 4. This
A/
scale also appears
under the color scale in Fig. 3, showing which colors represent which
A/
. The net
effect of the correction is that high
A/
is
represented by pixel value 0.353 instead of 0.313, which would be the
case without relaxation weighting.
The standard deviation from Eq. 1 describes the measurement
error for individual pixels. The SNR for groups of pixels increases with the square root of the number of pixels, making comparison of
groups more powerful. A familiar way to compare means of two groups of
pixels is Student's two-sample t statistic, t =
here µL is the mean of n
pixels from the left lung, µR is the
mean of m pixels from the right lung, and s is a
weighted average standard deviation, defined by s2 = Ns2L + Ms2R/N + M, where s2L is the sample
variance of the left lung, s2R
is the sample variance of the right lung, N = n
1, and M = m
1. If the left lung pixels and
right lung pixels were both normally distributed, then under the
hypothesis µL = µR, t has
Student's t distribution with N + M degrees of
freedom. The t-test is sufficiently robust that, if the
underlying distributions resemble normal, as ours do, the operating
characteristics of the test are essentially the same as if the
underlying distributions were strictly normal. We have the advantage of
being able to calculate s = 0.0682 using all 48,586 pixels from
the left lung and all 76,428 from the right, so that t has
125,012 degrees of freedom, and t is essentially Gaussian with
mean = 0 and SD = 1. Two pixels from the left lung with mean = 0.52 and
two from the right with mean = 0.342 give a t of 2.6; thus
µL > µR at the P < 0.005 level. Four pixels from each lung yield a t of 3.7 and P < 0.0001. Eight pixels from each lung yield a t
of 5.2 and P < 10
7. When
comparing both lungs, t is 451, and we cannot conveniently calculate the significance level because it is too small, but a
t of merely 8.2 implies significance at the
10
16 level.
In Fig. 3, groups of pixels that are different by two color bars are statistically different. The width of the color bars shows the measurement error for individual pixels of each color. Note that they grow wider toward the right side of the color scale, reflecting how the measurement error is a function of pixel value.
The results from all seven rats support the method's ability to detect obstructed ventilation. In two rats, the bead was correctly lodged in the proximal left bronchus. One of these images is shown in Fig. 3; the other was similar with the left lung brighter than the right. On dissection, the left lungs appeared darker in color than the right, but both appeared patent. In two rats, the bead was further in the left bronchus, where it did not fit as tightly but blocked secondary bronchial openings. There was partial atelectasis of the left lung, and the image showed a wedge-shaped bright region in the left lung. In one rat, the bead was similarly situated in the right bronchus, there was a similar region of partial atelectasis in the right lung, and the image showed a similar bright wedge-shaped area in the right lung. In one rat, the bead was found loose in the distal region of the left bronchus. The left and right lungs appeared similar on dissection, and the image showed no bright regions. In two rats, the bead was found loose near the carina, the lungs were similarly undistinguished on dissection, and the images showed no bright regions.
Even though the situation in which the bead reduced the ventilation to the entire left lung was repeated only once, the other bead placement and imaging results are in one-to-one correspondence. Where there was anatomic evidence that ventilation was obstructed, there is a bright region in the image; when the anatomic evidence is lacking, there is no bright region. This correspondence and the small significance levels associated with regions of images that appear different from one another are strong indications that the images display obstructed ventilation.
| |
DISCUSSION |
|---|
|
|
|---|
We do not have an independent measurement of
A/
;
therefore, the extent to which Fig. 3 displays
A/
values
relies on the theory leading to Fig. 1. The theory is well accepted by
respiratory physiologists and forms the basis of related techniques
that measure
A/
distributions (12, 23-25, 48, 49). If
A/
of the right lung
were normal, the predicted mean pixel value would be 0.364. The actual
value, 0.342, is within 6%, which is good agreement considering that
we were not monitoring blood gases to prevent overventilation and the
gas mixtures may have each varied by 4%.
Considering the difficulties of imaging gas in lungs with NMR, we are
very pleased that a quotient image has pixel values that fall within
6% of expectation and that we could quantify the noise at a reasonable
level. At present, we are secure that we have imaged obstructed
ventilation, and it appears that the detectable
A/
values are
in a physiologically relevant range. It is difficult to control
mechanical ventilation obstructions in rats, but this would be much
easier in larger mammals, which have bronchi that will accommodate
bronchoscopes, flowmeters, and controllable obstruction devices. We are
optimistic about the future quantitative development of this technique.
Scaling the technique from rats to human-sized mammals provides a signal-to-noise advantage. The SNR should increase by approximately the square root of the number of spins (11). Because lung volume scales with body mass (40), SNR increases by ~13 times in going from a 450-g rat to a 76-kg mammal. As a crude estimate, an image of human lungs similar to Fig. 3 with the same number of pixels would require a data collection time of 2.5 h/132 or 53 s. The resolution would be 2.92 mm instead of 0.529 mm. As with the rat, the data collection would occur during only 32% of the respiratory cycle, when the lungs are most full. People breathe 8-20 times/min at rest. Using the number of 15 breaths/min, one could acquire the obstruction-detecting image during 7 breaths and the reference image during 6 breaths. This scheme would imply collecting ~163 FIDs/breath, with each averaged over four or five acquisitions and separated by minor changes in the applied magnetic field gradient, corresponding to ~4.5° changes in direction. Alternatively, because only 53 s · 0.32/2 = 8.5 s of data are required for each image, the obstruction-detecting and reference image could each be acquired during one breath hold. However, people with obstructive lung disease often have trouble holding their breath.
Another scheme would purposely make the imaging process longer to make higher quality images by signal averaging, cardiac gating, and retrospective respiratory sorting. For example, assume that one eliminates data from one-fourth of spontaneous breaths in favor of those breaths that best match one another, collects data during only one-half of the heart cycle, and signal averages four times as much, the data collection would require 53 s · 1.33 · 2 · 4 = 10.6 min, but the image would have twice the SNR and fewer motion artifacts.
A limitation of our technique is lack of discrimination of high
A/
values, which are
useful, e.g., in detecting pulmonary embolisms. However, magnetic
resonance pulmonary angiograms and lung perfusion measurements are
improving rapidly. Because they may provide information sensitive to
embolisms, they may be particularly useful in filling a gap in our
technique. It has been common to measure
A and
of the
A-
-
A/
trio when evaluating lung function, but
A/
and
is also a useful combination.
| |
ACKNOWLEDGEMENTS |
|---|
Thanks go to Albert J. Olszowka, State University of New York at Buffalo, for providing computer programs of gas exchange and to Michael P. Hlastala and Thomas F. Hornbein, Univ. of Washington, Seattle, for criticism and advice in respiratory physiology. Thanks go also to Don Williams, Elena Simplaceanu, Maryann Butowicz, and Donald Bennett at Pittsburgh NMR Center for Biomedical Research (PNMRCBR) for helping D. O. Kuethe during visits to Pittsburgh. Steve Altobelli of New Mexico Resonance provided technical help in NMR, Jack Loeppky of Lovelace Respiratory Research Institute (LRRI) helped with and gave perspective in respiratory physiology, and Vicki White of LRRI helped with rats.
| |
FOOTNOTES |
|---|
This work was supported by National Institutes of Health Grants 1R29HL57967-01 (to D. O. Kuethe) and RR-03631-04 (to the PNMRCBR).
Present address of H. M. Gach: University of Pittsburgh Medical Center, Pittsburgh, PA 15213.
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
1
Algebra from expression in Ref. 27 or rederive
with theorem
h(z) = 


|x|f(zx)g(x)dx, where f and g are PDFs of X and Y,
respectively, and formula (21)
|
Address for reprint requests and other correspondence: D. O. Kuethe, NMR, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108-5127 (E-mail: dkuethe{at}NMR.org).
Received 19 April 1999; accepted in final form 20 January 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Ailion, DC.
Introduction to magnetic resonance imaging techniques used in lung.
In: Application of Magnetic Resonance to the Study of Lung., edited by Cutillo AG.. Armonk, NY: Futura, 1996, p. 33-48.
2.
Ailion, DC.
Lung and mediastinum: a discussion of the relevant physics.
In: Encyclopedia of Nuclear Magnetic Resonance, edited by Grant DM,
and Williams RK.. New York: Wiley, 1996, vol. 5, p. 2867-2873.
3.
Albert, MS,
Cates GD,
Driehuys B,
Happer W,
Saam B,
Springer CS,
and
Wishnia A.
Biological magnetic resonance imaging using laser-polarized 129Xe.
Nature
370:
199-201,
1994[Medline].
4.
Bachert, P,
Schad LR,
Bock M,
Knopp MV,
Ebert M,
Grossman T,
Heil W,
Hofmann D,
Surkau R,
and
Otten EW.
Nuclear magnetic resonance imaging of airways in humans with use of hyperpolarized 3He.
Magn Reson Med
36:
192-196,
1996[ISI][Medline].
5.
Barbara, JA,
Riverola A,
Roca J,
Ramirez J,
Wagner PD,
Ros D,
Wiggs BR,
and
Rodriguez-Roisin R.
Pulmonary vascular abnormalities and ventilation-perfusion relationships in mild chronic obstructive pulmonary disease.
Am J Respir Crit Care Med
149:
423-429,
1994[Abstract].
6.
Bergin, CJ,
Hauschildt J,
Rios G,
Belezzuoli EV,
Huynh T,
and
Channick RN.
Accuracy of MR angiography compared with radionuclide scanning in identifying the cause of pulmonary arterial hypertension.
Am J Roentgenol Radium Ther
168:
1549-1555,
1997.
7.
Bergin, CJ,
Pauly JM,
and
Macovski A.
Lung parenchyma: projection reconstruction MR imaging.
Radiology
179:
777-781,
1991
8.
Canfield, RE,
and
Rahn H.
Arterial-alveolar N2 gas pressure differences due to ventilation-perfusion variations.
J Appl Physiol
10:
165-172,
1957
9.
Chen, XJ,
Chawla MS,
Cofer GP,
Hedlund LW,
Moller HE,
and
Johnson GA.
Hyperpolarized 3He NMR lineshape measurements in the live guinea pig lung.
Magn Reson Med
40:
61-65,
1998[ISI][Medline].
10.
Durney, CH,
Ailion DC,
and
Cutillo AG.
Susceptibility effects (tissue-induced inhomogeneous broadening) in lung magnetic resonance imaging.
In: Application of Magnetic Resonance to the Study of Lung, edited by Cutillo AG.. Armonk, NY: Futura, 1996, p. 141-176.
11.
Edelstein, WA,
Glover GH,
Hardy CJ,
and
Redington RW.
The intrinsic signal-to-noise ratio in NMR imaging.
Magn Reson Med
3:
604-618,
1986[ISI][Medline].
12.
Farhi, LE.
Elimination of inert gas by the lung.
Respir Physiol
3:
1-11,
1967[ISI][Medline].
13.
Farhi, LE,
and
Olszowka AJ.
Analysis of alveolar gas exchange in the presence of soluble inert gases.
Respir Physiol
5:
53-67,
1968[ISI][Medline].
14.
Fenn, WO,
Rahn H,
and
Otis AB.
A theoretical study of the composition of alveolar air at altitude.
Am J Physiol
146:
637-653,
1946.
15.
Ferrer, A,
Roca J,
Wagner PD,
Lopez FA,
and
Rodriguez-Roisin R.
Airway obstruction and ventilation-perfusion relationships in acute severe asthma.
Am Rev Respir Dis
147:
579-584,
1993[ISI][Medline].
16.
Finney, RJ,
Wolfe M,
and
Jonas J.
NMR study of angular momentum relaxation in fluids. I. Compressed CF4.
J Chem Phys
67:
4004-4011,
1977.
17.
Ganesan, K,
Ailion DC,
Hackmann A,
Laicher G,
Chan PH,
and
Cutillo AG.
NMR relaxation and water self-diffusion mechanisms in lung.
In: Application of Magnetic Resonance to the Study of Lung, edited by Cutillo AG.. Armonk, NY: Futura, 1996, p. 115-140.
18.
Gao, J-H,
Lemen L,
Xiong J,
Patyal B,
and
Fox PT.
Magnetization and diffusion effects in NMR imaging of hyperpolarized substances.
Magn Reson Med
37:
153-158,
1997[ISI][Medline].
19.
Gefter, WB,
Hatabu H,
Holland GA,
Gupta KB,
Henschke CI,
and
Palevsky HI.
Pulmonary thromboembolism: recent diagnosis with CT and MR imaging.
Radiology
197:
561-574,
1995
20.
Gluck, EH,
Onorato DJ,
and
Castriotta R.
Helium-oxygen mixtures in intubated patients with status asthmaticus and respiratory acidosis.
Chest
98:
693-698,
1990
21.
Gradshteyn, IS,
and
Ryzhik IM.
Table of Integrals, Series, and Products, edited by Jeffrey A.. New York: Academic, 1965, p. 338, formula 3.462.5.
22.
Hatabu, H,
and
Gefter WB.
MR imaging of the pulmonary vasculature: basic concepts and clinical applications.
In: Application of Magnetic Resonance to the Study of Lung, edited by Cutillo AG.. Armonk, NY: Futura, 1996, p. 287-322.
23.
Hlastala, MP.
Multiple inert gas elimination technique.
J Appl Physiol
56:
1-7,
1984
24.
Hlastala, MP,
and
Domino KB.
Roles of hypoxia and blood flow in modulating
A/
heterogeneity in the lungs.
Adv Exp Med Biol
345:
67-73,
1994[Medline].
25.
Hlastala, MP,
and
Robertson HT.
Inert gas elimination characteristics of the normal and abnormal lung.
J Appl Physiol
44:
258-266,
1978
26.
Johnson, GA,
Cates G,
Chen XJ,
Cofer GP,
Driehuys B,
Happer W,
Hedlund LW,
Saam B,
Shattuck MD,
and
Swartz J.
Dynamics of magnetization in hyperpolarized gas MRI of the lung.
Magn Reson Med
38:
66-71,
1997[ISI][Medline].
27.
Kamerud, DB.
The random variable X/Y, X, Y, normal.
Am Math Monthly
85:
206-207,
1978.
28.
Kass, JE,
and
Castriotta RJ.
Helium therapy in acute severe asthma.
Chest
107:
757-760,
1995
29.
Kauczor, HU.
Contrast-enhanced magnetic resonance angiography of the pulmonary vasculature: a review.
Invest Radiol
33:
606-617,
1998[ISI][Medline].
30.
Kauczor, H,
Ebert UM,
Kreitner KF,
Nilgens H,
Surkau R,
Heil W,
Hoffmann D,
Otten EW,
and
Thelen M.
Imaging of the lungs using 3He MRI: preliminary clinical experience in 18 patients with and without lung disease.
J Magn Reson Imaging
7:
538-543,
1997[ISI][Medline].
31.
Kuethe, DO,
Caprihan A,
Fukushima E,
and
Waggoner RA.
Imaging lungs using inert fluorinated gases.
Magn Reson Med
39:
85-88,
1998[ISI][Medline].
32.
Kuethe, DO,
Caprihan A,
Lowe IJ,
Madio DP,
and
Gach HM.
Transforming NMR data despite missing points.
J Magn Reson
139:
18-25,
1999[ISI][Medline].
33.
Lizak, MJ,
Conradi MS,
and
Fry CG.
NMR imaging of gas imbibed into porous ceramic.
J Magn Reson
95:
548-557,
1991.
34.
Middleton, H,
Black RD,
Saam B,
Cates GD,
Cofer GP,
Guenther R,
Happer W,
Hedlund LW,
Johnson GA,
Juvan K,
and
Swartz J.
MR imaging with hyperpolarized 3He gas.
Magn Reson Med
33:
271-275,
1995[ISI][Medline].
35.
Mohanty, S,
and
Bernstein HJ.
Fluorine relaxation by NMR absorption in gaseous CF4, SiF4, and SF6.
J Chem Phys
53:
461-462,
1970.
36.
Moller, HE,
Chen XJ,
Chawla MS,
Driehuys B,
Hedlund LW,
and
Johnson GA.
Signal dynamics in magnetic resonance imaging of the lung with hyperpolarized noble gases.
J Magn Reson
135:
133-143,
1998[ISI][Medline].
37.
Revokatov, OP,
and
Parfenov SV.
Spin-lattice relaxation time in SF6 in the critical temperature region.
JETP Lett
15:
103-105,
1972.
38.
Rinck, PA,
Petersen SB,
and
Lauterbur PC.
NMR imaging of fluorine-containing substances. 19-Fluorine ventilation and perfusion studies.
ROFO Fortschr Geb Rontgenstr Nuklearmed
140:
239-243,
1984.
39.
Saam, B,
Happer W,
and
Middleton H.
Nuclear relaxation of 3He in the presence of O2.
Physiol Rev
52:
862-865,
1995.
40.
Schmidt-Nielsen, K.
Scaling: Why Is Animal Size So Important? Cambridge, UK: Cambridge Univ. Press, 1984, p. 241.
41.
Severinghaus, JW.
Blood gas calculator.
J Appl Physiol
21:
1108-1116,
1966
42.
Shioya, S,
Christman R,
and
Ailion DC.
An in vivo NMR imaging determination of multiexponential Hahn T2 of normal lung.
Magn Reson Med
16:
49-56,
1990[ISI][Medline].
43.
Shiue, S-T,
and
Gluck EH.
The use of helium-oxygen mixtures in the support of patients with status asthmaticus and respiratory acidosis.
J Asthma
26:
177-180,
1989[ISI][Medline].
44.
Swensen, ER,
Robertson HT,
and
Hlastala MP.
Effects of inspired carbon dioxide on ventilation-perfusion matching in normoxia, hypoxia, and hyperoxia.
Am J Respir Crit Care Med
149:
1563-1569,
1994[Abstract].
45.
Swidwa, DM,
Montenegro HD,
Goldman MD,
Lutchen KR,
and
Saidel GM.
Helium-oxygen breathing in severe chronic obstructive pulmonary disease.
Chest
87:
790-795,
1985
46.
Voelkel, NF.
Mechanisms of hypoxic pulmonary vasoconstriction.
Am Rev Respir Dis
133:
1186-1195,
1986[ISI][Medline].
47.
Wagner, PD,
Laravuso RB,
Uhl RR,
and
West JB.
Continuous distributions of ventilation-perfusion ratios in normal subjects breathing air and 100% O2.
J Clin Invest
54:
54-68,
1974.
48.
Wagner, PD,
Salzman HA,
and
West JF.
Measurement of continuous distribution of ventilation-perfusion ratios.
J Appl Physiol
36:
588-599,
1974
49.
West, JB.
Ventilation-perfusion relationships.
Am Rev Respir Dis
116:
919-943,
1977[ISI][Medline].
50.
Wielopolski, PA,
Haacke EM,
and
Adler LP.
Three-dimensional MR imaging of the pulmonary vasculature: preliminary experience.
Radiology
183:
465-472,
1992
51.
Wielopolski, PA,
Haacke EM,
and
Adler LP.
Three-dimensional MR imaging of the pulmonary vascular system.
In: Application of Magnetic Resonance to the Study of Lung, edited by Cutillo AG.. Armonk, NY: Futura, 1996, p. 323-386.
52.
Wolfe, M,
Arndt E,
and
Jonas J.
NMR study of angular momentum relaxation in fluids. II. Mixtures of CF4 with inert gases.
J Chem Phys
67:
4012-4018,
1977.
This article has been cited by other articles:
![]() |
M. F. V. Melo, R. S. Harris, J. D. H. Layfield, and J. G. Venegas Topographic Basis of Bimodal Ventilation-Perfusion Distributions during Bronchoconstriction in Sheep Am. J. Respir. Crit. Care Med., April 1, 2005; 171(7): 714 - 721. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. F. Vidal Melo, D. Layfield, R. S. Harris, K. O'Neill, G. Musch, T. Richter, T. Winkler, A. J. Fischman, and J. G. Venegas Quantification of Regional Ventilation-Perfusion Ratios with PET J. Nucl. Med., December 1, 2003; 44(12): 1982 - 1991. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.U. Kauczor, X.J. Chen, E.J.R. van Beek, and W.G. Schreiber Pulmonary ventilation imaged by magnetic resonance: at the doorstep of clinical application Eur. Respir. J., May 1, 2001; 17(5): 1008 - 1023. [Abstract] [Full Text] [PDF] |
||||
| ||