Flow and pressure measurements were
performed in the ascending aortas of six pediatric patients ranging in
age from 1 to 4 yr and in weight from 7.2 to 16.4 kg. From these
measurements, input impedance was calculated. It was found that total
vascular resistance decreased with increasing patient weight and was
approximately one to three times higher than those of adults.
Conductance per unit weight was relatively constant but was
approximately three times higher than for adults. Strong inertial
character was observed in the impedance of four of the six patients.
Among a three-element and two four-element lumped-parameter models, the
model with characteristic aortic resistor (Rc) and
inertance in series followed by parallel peripheral resistor
(Rp) and compliance fitted the data best. Rp
decreased with increasing patient weight and was one to three times
higher than in adults, and Rc decreased with increasing patient weight and was 2 to 15 times higher. The
Rp-to-Rc ratio differed significantly between
infants and children vs. adults. The results suggested that
Rp developed more rapidly with patient weight than did
Rc. Compliance values increased with increasing patient
weight and were 3 to 16 times lower than adult values.
 |
INTRODUCTION |
EARLY PRESSURE AND FLOW MEASUREMENTS and subsequent
calculations of human adult aortic input impedance were made by Gabe et al. (3) and Patel et al. (11). Subsequent measurements were performed
by Mills et al. (5), Nichols et al. (8), Murgo et al. (7), and
O'Rourke and Avolio (9). Although considerable variability was found,
in many individuals the modulus fell to a shallow minimum of
8-10% of the zero frequency (Z0) resistance (~1,300
dyn · s · cm
5) at
4-5 Hz, and a peak occurred at ~7 Hz. Other
individuals exhibited modulus curves with dual minima. The phase of the
impedance dropped from zero at Z0 to about
40°
at ~1 Hz and then rose to a zero crossing at ~3-4 Hz and rose
gradually to ~25° at 5 Hz. Considerable individual variation also
existed in the phase.
Murgo et al. (7) found different impedance spectra in groups of
young (24 ± 2 yr), older (33 ± 3 yr), and oldest (40 ± 4 yr)
adults. In the young group, the modulus displayed a strong minimum at
low frequency, followed by relatively constant modulus at higher
frequency. The oldest group showed a minimum at higher frequency
followed by a maximum at twice that frequency, suggestive of a dominant
reflection site within the vasculature. Increased impedance phase was
also found in the oldest group. Both of these phenomena were explained
by the degeneration (stiffening) of the aorta with age, which would
increase wave speed and reduce compliance.
Several differences between infant and child vs. adult impedance were
expected. First, because of somewhat smaller aortic pulse pressure but
much smaller cardiac stroke volume, it was expected that infant/child
vascular systems would exhibit lower compliance. [Compliance may
be roughly estimated as stroke volume divided by pulse pressure. Lower
compliance is possible in spite of the increased tissue elasticity of
younger vessels (13), because compliance is a volumetric strain
response, whereas elasticity is a lineal strain response.] Lower
compliance would decrease the impedance phase, particularly at low
frequency. Second, because average aortic pressures are somewhat
smaller in infants and children, but average flow rates (cardiac
outputs) are much lower, total vascular resistance must be increased.
Because peripheral resistance (Rp) normally provides most
of the total vascular resistance, this factor would raise the modulus
not only at Z0 but across all frequencies. Third, because
of the shorter vessel lengths in infants and children, shorter wave
reflection times and, therefore, modulus minima at higher frequencies
were expected.
Several lumped-parameter models [with resistance (R), compliance
(C), and inertance (L) elements] have been applied to the adult
systemic vasculature (Fig. 1). The
two-element model (2) (termed RC) provides the important characteristic
of a drop in impedance modulus along with negative impedance phase with
increasing frequency, although these characteristics are accompanied by
an asymptotic modulus at high frequency of zero and an asymptotic phase
at high frequency of
90°, both of which are
nonphysiological. The three-element model (19) (termed RCR) provides a
significant improvement over the two-element model in that its modulus
has a nonzero high-frequency asymptote and its phase has a
high-frequency asymptote of zero. However, Burkhoff et al. (1) found
that, whereas the RCR model provided a reasonable representation of afterload for predicting integrated measures of cardiac performance, such as stroke volume, stroke work, and average systolic and diastolic aortic pressures, it did not provide realistic aortic pressure and flow
waveforms, and it significantly underestimated peak aortic flow. The
four-element model (14) (termed RLRC) provides potential for
improvement with a minimum modulus at intermediate frequency and
increasing modulus for higher frequency (whereas the RCR model has a
monotonically decreasing modulus) and positive phase at high frequency.
An alternative four-element model (17) (termed RLRC2) also provides a
minimum modulus at intermediate frequency but with a constant
high-frequency asymptote compared with the increasing asymptote of the
RLRC model. The RLRC2 model provides a transition from negative to
positive phase at intermediate frequency but with a high-frequency
asymptote of zero compared with the 90° asymptote of the RLRC
model. Five-element models have also been proposed and compared with
more simple models (14, 18).

View larger version (22K):
[in this window]
[in a new window]
|
Fig. 1.
Lumped-parameter models of the systemic circulation. R, resistance; C,
compliance; L, inertance; Rc, characteristic aortic
resistance; Rp, peripheral resistance. RC, 2-element model;
RR, 3-element model; RLRC, 4-element model; RLRC2, alternative
4-element model (see text).
|
|
This study provides what is thought to be the first report of
simultaneous high-fidelity measurements of aortic pressure and flow and
calculations of impedance in infants and children. The fits of the RCR,
RLRC, and RLRC2 models to the results were evaluated.
 |
METHODS |
Six patients at the Primary Children's Medical Center (Salt Lake City,
UT) were enrolled in this study between June 1996 and August 1997. Approval to conduct the investigation was obtained from the local
Institutional Review Board, and written informed consent was obtained
from the parents of the patients before enrollment in the
study. All patients were admitted for repair of simple cardiac defects and had normal aortic anatomy, stable cardiac function,
and no previous history of cardiac or thoracic surgery. The patients
ranged in age from 0.8 to 4.0 yr (2.2 ± 1.4 yr, mean ± SD) and in
weight from 7.2 to 16.4 kg (11.5 ± 3.9 kg). Five of the patients
underwent surgery for closure of atrial septal defects, and one
underwent surgery for pulmonary valvotomy and patch angioplasty of the
right pulmonary artery. The ages and weights of the patients, along
with average aortic pressures and flows, are presented in Table
1.
The patients, measurement procedures, and flow and pressure data are
the same as those used by Pantalos et al. (10) in the investigation of
intra-aortic balloon pump timing errors. Briefly, a transit time
ultrasonic perivascular flow probe (10 or 14 mm, 100-Hz frequency
response, Transonic Systems, Ithaca, NY) was positioned around the
aorta ~1 cm downstream of the aortic valve after surgical exposure of
the heart but before cannulation for cardiopulmonary bypass. A
high-fidelity catheter tip pressure transducer (5-Fr MPC-500, 5-kHz
frequency response, Millar Instruments, Houston, TX) was placed in the
aortic root through a hole, and purse-string suture, that was later
used for the cardioplegia infusion line at the time cardiopulmonary
bypass was initiated. The catheter was advanced down the ascending
aortic lumen so that the sensor was at the level of the flow probe.
Consequently, the pressure and flow measurements were made
simultaneously at the same location in the ascending aorta. Flow and
pressure waveforms were recorded on FM tape (MR-30 data tape recorder,
313-Hz frequency response; TEAC America, Montebello, CA).
Segments of the analog recordings were subsequently digitized (GW
Instruments MacADIOS analog-to-digital board in a Macintosh 8100 computer running Superscope 2.17 data- acquisition software) over 5 s
with 0.005-s sampling interval (10,000 samples). The beginnings of the
waveforms were selected from digitized data by identifying the sharp
rise in pressure between diastole and systole. The beginning and end of
each pressure cycle were similarly identified. Ten cycles were used,
except for patient 6, whose heartbeat frequency was low enough
that only eight cycles of digitized data were available in the 5-s
sampling window. A Fourier transform was calculated for each cycle of
pressure and flow. Input impedance was then calculated for each cycle
as the ratio of pressure and flow. The real and imaginary parts of the
impedance values were averaged, and standard deviations were
calculated. A threshold of quality for the resulting average impedance
values was established. Harmonics that had standard deviations of
either real or imaginary parts >20% of the zeroth harmonic modulus
were excluded, along with all higher harmonics for the same patient
data. The maximum number of harmonics was arbitrarily limited to 10. These impedance values were compared with the average impedance of five
adults (8).
The RCR model was fit to the impedance of each patient and to the adult
average. Two fitting procedures were evaluated. First, with the sum of
characteristic aortic resistance (Rc) and Rp
set equal to the measured modulus at Z0, Rc and
C were found iteratively such that the normalized root mean square
difference (NRMS) in measured and modeled impedance was minimized
|
(1)
|
where
Re and Im are the real and imaginary parts of the measured impedance
(Z) and the modeled impedance (A), and m is the number of
harmonics satisfying the accuracy criterion. Second, with
the sum of Rc and Rp again set equal to
Z0, Rc and C were found iteratively such that
the first harmonic modulus and phase were fit exactly. The NRMS was
also calculated for this second fitting method.
The four-element models were fit to the measured patient impedance
values and to the adult average by procedures similar to the first
method above. For the RLRC model, with the sum of Rc and
Rp set equal to Z0, Rc, C, and L
were found iteratively such that NRMS was minimized. For the RLRC2
model, with Rp set equal to Z0, Rc,
C, and L were found iteratively such that NRMS was minimized. For each
fit, the sensitivity of NRMS to parameter values was tested, including,
for the four-element models, variations in L and C with constant LC
product (constant crossover frequency; see DISCUSSION).
In addition, arterial compliance was also calculated for the patient
data by the area (4) and pulse pressure methods for comparison to the
results of the curve fits described above. These calculations were
based on the first complete cycle only. Mean systolic and mean
diastolic pressures (Table 1) were also calculated from the digitized
data for the first cycle only.
General correlation coefficients (r) and 95% confidence
intervals (CI95) were calculated for the correlations of
results with both patient weight and age. ANOVA comparisons were
performed for the results from pediatric patient vs. adults and for
results from infants (patients 1 and 3) vs. the group
of children (patients 2, 4, 5, and 6).
(Although patient 1 was not technically an infant, being
slightly older than 1 yr, the similarity of the response of patient
1 and true infant patient 3 promoted this grouping and
description of the groups for comparison purposes.)
 |
RESULTS |
Example pressure and flow waveforms used for impedance calculations are
shown in Fig. 2. The cycle periods were
constant for each patient within ±1 sampling interval except for
patient 5, for whom the variation was ±1.5 sampling
intervals, and patient 6, for whom the variation was ±2.5
sampling intervals. Mean pressures and flows for each 10-cycle (8-cycle
for patient 6) data set are given in Table 1. Although mean
flow increased with patient weight and was reasonably correlated
(r = 0.886, CI95 = 0.265 < r < 0.988), mean pressure was not well correlated (r = 0.214, CI95 =
0.723 < r < 0.874). Mean
resistance calculated by the ratio of mean pressure and mean flow
decreased with patient weight and was highly correlated as shown in
Fig. 3.

View larger version (30K):
[in this window]
[in a new window]
|
Fig. 2.
Example of aortic pressure (AOP; top) and aortic flow
(AOQ; bottom) waveforms (patient 2). Ticks on abscissas
indicate 1-s intervals.
|
|

View larger version (15K):
[in this window]
[in a new window]
|
Fig. 3.
Mean R calculated as the ratio of mean pressure and mean flow.
Correlation with patient weight is as follows: r = 0.988, 95% confidence interval (CI95) = 0.999 < r < 0.888.
|
|
Mean harmonics of pressure, flow, and impedance, along with standard
deviations, are provided in Table 2. The
impedance values listed are limited to the harmonics satisfying the
accuracy criterion (that standard deviation of either real or imaginary
parts of the impedance be no greater than 20% of the zeroth harmonic
modulus). The accuracy criterion corresponded in all but one case with
a modulus of flow <0.02 l/min or a modulus of pressure <0.15 mmHg. (The exception was patient 6, for whom the normalized standard deviations of the real and imaginary parts were 20.46 and 26.05, respectively, for a combination of a relatively low-pressure modulus of
0.21 mmHg and a relatively low-flow modulus of 0.028 l/min.) For
patient 1, the number of harmonics was arbitrarily limited to
10, although the accuracy criterion was not exceeded. Sampling frequency may also limit the resolution of the Fourier transforms; however, in these measurements, the sampling frequency (200 Hz) greatly
exceeded the frequency of the highest harmonic considered (25.4 Hz for
the 10th harmonic for patient 1).
Normalized impedance modulus and phase for all patients are shown in
Fig. 4. Two types of responses were found.
Type A had low inertance with constant high-frequency asymptote
of impedance modulus and a positive, but relatively small, impedance
phase at high frequency. Type B had high inertance with
impedance modulus that increased at high frequency and high-impedance
phase at high frequency. The type A response, which was also
facilitated by high Rc, occurred in the two youngest
patients (patients 1 and 3), and the type B
response, with lower Rc, occurred for the older patients
(patients 2, 4, 5, and 6). A comparison
of the fits of the lumped-parameter models, using an example of each
type of response, is shown in Fig. 5.


View larger version (2222K):
[in this window]
[in a new window]
|
Fig. 5.
Comparison of fits of lumped-parameter models to low-L,
high-Rc patient 1 (A) and high-L,
low-Rc patient 2 (B). , Patient data.
|
|
The total vascular resistances (Z0) of the models were
specified by the fitting procedures to be equal to the average
impedance moduli at Z0 (Table 2) for each
patient. The Z0 values, which resulted from
the Fourier analysis, thus were nearly identical to the average
resistances calculated from the average pressures and flows (Fig.
3). Z0, therefore, also decreased with patient weight and
was highly correlated (r = 0.988, CI95 = 0.890 < r < 0.999). Although Rc in the models showed a
general decrease with increasing patient weight, the correlations were
not strong (Fig. 6). Rp,
however, decreased with patient weight and was well correlated (Fig.
7). C increased with patient weight for all
models (Fig. 8). L decreased with patient
weight for the RLRC2 model, but no clear trend was evident for the
RLRC model (Fig. 9).

View larger version (18K):
[in this window]
[in a new window]
|
Fig. 6.
Characteristic aortic resistance for each model. Correlations with
patient weight: RCR, r = 0.824 and CI95 = 0.980 < r < 0.037; RCR2, r = 0.842 and CI95 = 0.982 < r < 0.097; RLRC, r = 0.819 and CI95 = 0.980 < r < 0.023; RLRC2, r = 0.494 and CI95 = 0.932 < r < 0.530.
|
|

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 7.
Rp for each model. Correlations with patient weight are as
follows: RCR, r = 0.954 and CI95 = 0.995 < r < 0.630; RCR2, r = 0.975 and CI95 = 0.997 < r < 0.785; RLRC, r = 0.953 and CI95 = 0.995 < r < 0.627; RLRC2, r = 0.988 and CI95 = 0.999 < r < 0.890.
|
|

View larger version (23K):
[in this window]
[in a new window]
|
Fig. 8.
Arterial C for each model. Correlations with patient weight are as
follows: RCR2, r = 0.907 and CI95 = 0.363 < r < 0.990; RLRC, r = 0.852 and CI95 = 0.131 < r < 0.984; RLRC2, r = 0.952 and
CI95 = 0.618 < r < 0.995. Statistics
for RCR model are not given because of inaccuracy of C values. For
reference, arterial C calculated by area (Ref. 4) and pulse pressure
methods are also shown: RC, area, incisura to end diastole; RC2, area,
peak after incisura to end diastole; RC3, pulse pressure. Correlations
with patient weight are as follows: RC, r = 0.748 and
CI95 = 0.161 < r < 0.971; RC2, r = 0.910 and CI95 = 0.377 < r < 0.990;
RC3, r = 0.961 and CI95 = 0.682 < r < 0.996.
|
|

View larger version (19K):
[in this window]
[in a new window]
|
Fig. 9.
Arterial L for the two 4-element models (RLRC and RLRC2). Correlations
with patient weight are as follows: RLRC, r = 0.222 and
CI95 = 0.876 < r < 0.719; RLRC2,
r = 0.691 and CI95 = 0.963 < r < 0.275.
|
|
 |
DISCUSSION |
Considerable variation among patients in pressure and flow was evident
in the waveforms (example in Fig. 2) and in their averages (Table 1).
Because the patients were selected on the basis of absence of left
ventricular and aortic abnormalities, the variability is not likely due
to abnormal cardiac or vascular performance. That all patients had
normal left ventricular performance appears to be supported by the
pressures and flows being in an acceptable range and by the reasonable
correlation of aortic flow with patient weight. The variability in
pressure may be due in part to subject variability, including response
to anesthesia. The anesthetic induction and maintenance agents used
with these patients (fentanyl, pancuronium bromide, isoflurane, and
nitrous oxide) are known to cause varying degrees of myocardial
depression and arterial vasodilation, which affect the stroke volume
and vascular resistance. The patient-to-patient variability in the
amount of agent needed to achieve the desired anesthetic plane and the
individual cardiovascular response to the agents, coupled with the
normal range of cardiovascular regulation, may have contributed to the
subject-to-subject variability found in this patient population.
The correspondence of high standard deviation (>20%) of impedance to
low pressure and flow moduli approaching the accuracy of the respective
measurements suggests that the quality of the impedance averages for
increasing frequency was limited at this level by the accuracy of the
pressure and flow measurements rather than by inherent variations in
the input impedance. Note that considerable fluctuations in impedance
(exceeding in some cases twice the value of the individual modulus)
nonetheless exist in the harmonics that satisfy the accuracy
criterion. These fluctuations may be representative of
primary changes in vascular response (due, for instance, to varying
vasoconstriction) or of changes in vascular response due to interacting
secondary functions (for instance, respiration) but may also derive in
part from inherent nonlinearities in vascular response.
For compliant vessels, greater pressure increases cross-sectional area
for flow, resulting in lower impedance to flow. The influence of this
effect was investigated by testing the correlation of Z0
with zeroth harmonic pressure for each of the cycles. Decreases in
Z0 with increasing pressure were found in all patients,
although correlations, other than for patient 1, were not
strong (patient 2: r =
0.153,
CI95 =
0.714 < r < 0.528; patient
3: r =
0.392, CI95 =
0.819 < r < 0.315; patient 4: r =
0.561,
CI95 =
0.880 < r < 0.107; patient
5: r =
0.312, CI95 =
0.787 < r < 0.395; patient 6: r =
0.365,
CI95 =
0.809 < r < 0.343). In the
most highly correlated subject (patient 1, Fig.
10), a maximum zeroth harmonic pressure
variation among cycles of ~10% corresponded to a maximum
Z0 variation of ~20%. Correlations between higher harmonics of impedance and pressure were not evident. These small differences do not explain the much larger fluctuations observed among
cycles in the higher harmonics of pressure, flow, and impedance but
suggest that linearity of response should not be assumed in cases in
which pressure varies more than in the present measurements (~10%).

View larger version (18K):
[in this window]
[in a new window]
|
Fig. 10.
Correlation of average zeroth harmonic impedance with average zeroth
harmonic pressure for patient 1: r = 0.900
and CI95 = 0.976 < r < 0.624.
|
|
The excellent correlation of mean resistance with patient weight (Fig.
3) appears to indicate that this measure of systemic vascular response
was normal despite the effects of anesthesia and other influences and,
furthermore, that patient weight was a good scaling parameter
(correlation with patient age was slightly lower: r =
0.928, CI95 =
0.993 < r <
0.471, as it was for most results). The average resistance of
the heaviest patient was only slightly higher than that of adults,
despite the child's weight being only about one-quarter that of
adults. On the other hand, the resistance of the lightest patient was
approximately three times that of adults. One might expect that
conductance (inverse of resistance) per unit weight may be constant,
and, indeed, this parameter varied only from 3.2 × 10
5 to 4.0 × 10
5
cm5 · dyn
1 · s
1 · kg
1
in the patients but was ~1.1 × 10
5
cm5 · dyn
1 · s
1 · kg
1
in adults. Conductance per unit weight was only weakly
correlated with weight among the patients [r = 0.222, CI95 =
0.719 < r < 0.876, and P = 0.493 between infants and children (ANOVA)] but was
significantly different in infants and children compared with adults
(P < 0.001).
Decreases in Rc (Fig. 6) and Rp (Fig. 7) with
patient weight were also seen. Because of the wide range of
Rc in the patients, the difference in Rc
between infants and children vs. adults was not significant
(RCR: P = 0.379, RCR2: P = 0.315, RLRC: P = 0.378, RLRC2: P = 0.196), but the difference between
infants and children was significant except for the RLRC2 model (RCR:
P = 0.015, RCR2: P = 0.059, RLRC: P = 0.018, RLRC2: P = 0.676). In the heaviest patient, Rp was
only ~10% higher than that of adults, whereas Rc was
still approximately three times higher.
The ratio of Rc to total vascular resistance also dropped
with increasing weight (except for the RLRC2 model, which exhibited an
Rc value for best fit that was influenced by its parallel
inertance). This trend is consistent with the obvious growth of the
arteries from childhood to adulthood; however, because Rc
represented a small fraction of the total vascular resistance, its
development had little impact on Z0. The normalized
Rc values (0.058-0.212) were higher than those for
adults [~0.028-0.056 from data by Nichols et al.
(8)], but, because of the wide range, the difference was not
statistically significant among these groups except for the RLRC2 model
(RCR: P = 0.323, RCR2: P = 0.222, RLRC: P = 0.321, RLRC2: P = 0.083). The difference between infants and
children, however, was significant (RCR: P = 0.006, RCR2:
P = 0.072, RLRC: P = 0.006, RLRC2: P < 0.090). The difference in the Rp-to-Rc ratio (Fig. 11) was significant for comparisons
of both infants and children vs. adults (RCR: P = 0.049; RCR2:
P = 0.028; RLRC: P = 0.046; RLRC2: P < 0.001)
and for infants vs. children (RCR: P = 0.018; RCR2: P = 0.093; RLRC: P = 0.017; RLRC2: P = 0.022).

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 11.
Ratio of Rp to Rc
( ). Correlations with patient
weight are as follows: RCR, r = 0.665 and CI95 = 0.319 < r < 0.959; RCR2, r = 0.582 and
CI95 = 0.436 < r < 0.946; RLRC,
r = 0.642 and CI95 = 0.355 < r < 0.956; RLRC2, r = 0.425 and CI95 = 0.919 < r < 0.590.
|
|
The reduction of Rp to a level near that of
adults suggests that the peripheral circulation develops
rapidly early in life, whereas the remaining difference between child
and adult Rc suggests that development of the proximal
arteries may follow the slower geometric growth of the body itself. The
difference in the development rates is supported by the above
comparisons of infant, child, and adult
Rp-to-Rc ratio. The reasons for this difference
in development rates remain to be investigated. One attractive
hypothesis is that Rp, as well as other parameters, may
decrease rapidly to reduce hydraulic power requirements. However, in
this data set, power increased with patient weight (r = 0.771, CI95 =
0.109 < r < 0.528) and even
increased slightly in terms of power per unit weight (r = 0.173, CI95 =
0.743 < r < 0.863).
The decrease in the Rc-to-Z0 ratio with
increasing weight is also evident in Fig. 4; however, other
distinguishing features in the impedance plots also stand out. With the
exception of patients 1 and 3, the patient data
exhibited increasing modulus and phase with increasing frequency at
high frequency. The phase increased to 60° or more for patients
2, 4, 5, and 6. This behavior, which is
characteristic of inertial flow, is much stronger than for adults.
Higher inertance, which scales with
l/A, where
is blood density, l is vessel length, and A is
cross-sectional area, might be expected in infants and/or children if
A scaled with l2. Indeed, the modeled
inertance values were approximately an order of magnitude larger than
those for adults for the RLRC model (Table 3), although the wide range of the
infant/child values made this difference statistically insignificant
(P = 0.340). Although no trend with increasing weight was
evident for L for the RLRC model, L in the RLRC2 model decreased with
weight (Fig. 9).
On the other hand, patients 1 and 3 exhibited no
significant trend in modulus for high frequency, and phase remained
below 35° for patient 1 and lower than 15° for
patient 3. Patient 3 was the youngest and lightest
subject, whereas patient 1 was the second youngest and third
lightest; however, the connection between age or weight and the less
inertial impedance behavior of these patients is unclear.
The RLRC model produced the best fits of the impedance data for all
patients. Furthermore, this model also produced the lowest values of
the Akaike information criterion (AIC) and the Schwarz criterion (SC;
see Ref. 17), two indexes that account in different ways for the
expectation that models with more parameters should provide better fits
(Table 3). A survey of the curve fits for each patient (examples in
Fig. 5) reveals the following characteristics.
First, the RCR model was severely limited in its ability to match
infant/child impedance because its phase cannot be positive. The curve
fits in this work were based on errors between modeled and actual
impedance at all harmonics satisfying the accuracy criterion, which
included high-frequency harmonics. These curve fits produced
unrealistic values of C because increasing compliance produced better
(though still poor) fits of the higher harmonics of impedance as the
modeled impedance phase approached zero. The best fits in several cases
would actually be obtained with infinite C; however, the tabulated C
values (Table 3) are those for which further increases in C produced
insignificant improvements in NRMS. Better estimates of C were obtained
with the RCR model by using fits based on the zeroth and first
harmonics only (designated RCR2 in Fig. 14). Yet another compliance
estimation method for the RCR model is based on a fit of the zeroth
harmonic, an estimate of Rp from an average of the
high-frequency harmonics of the impedance modulus and then a
calculation of C based on an approximate fit of the first harmonic
impedance modulus (18a). This method, however, depends on the first
harmonic modulus being larger than the estimated Rp, which
was not the case for the infants and children. The method resulted in
imaginary values for C.
The area method of Liu et al. (4), another compliance estimation method
based on the decay of pressure during diastole according to an RC model
of cardiac afterload, produced greatly different C values depending on
the interval of diastole chosen. C estimated by the area method is
given by
|
(2)
|
where
is the
integrated pressure from time 1 to time 2, and
P1 and P2 are the beginning and ending
pressures. If the interval from the incisura to the end of diastole was
used, the small pressure difference in the denominator of Eq. 2
produced large C values. If the interval from the peak in pressure
after the incisura to end diastole was used, the much larger pressure
difference produced much smaller C values. These two compliance values
(for models RC and RC2 in Fig. 8) essentially bracketed the compliances
resulting from the curve fits of the lumped-parameter models. C
estimated by the pulse pressure method (for model RC3 in Fig. 8) was
comparable to the higher area method results. C correlated strongly
with patient weight (Fig. 8) and was significantly different in infants and children relative to adults (RCR2: P = 0.024, RLRC:
P < 0.001, RLRC2: P < 0.001, pressure wave forms
were not available to calculate RC, RC2, and RC3 compliances, and
statistics for the RCR model are not given because of the inaccuracy of
the C values for this model). C in infants and children was
approximately 10 times lower than that in adults. C was not
significantly different between infants and children except for the RC3
model (RC: P = 0.383, RC2: P = 0.203, RC3: P = 0.053, RCR2: P = 0.218, RLRC: P = 0.506, RLRC2:
P = 0.188).
Second, for the patients with strong inertial character, the RLRC2
model produced impedance modulus curves with sharp minima, which
adversely influenced the fits. In addition, its transition from
negative phase to positive phase was abrupt and was followed by
decreasing phase toward an asymptote of 0° (see Fig. 5B).
This model did not appear to fit the character of these patients. For patients 1 and 3, the mismatch of the shape of the
modeled impedance to the patient data was less dramatic but still noticeable.
The poor performance of the RLRC2 model can be explained by examining
its impedance equation
|
(3)
|
which
can be nondimensionalized as
|
(4)
|
where
Z* = Z/Z0, n =
/
1, where
is
angular frequency and
1 is first harmonic angular
frequency, L* =
1L/Z0,
= Rc/Z0,
C* =
1RpC, and Z0 = Rp for this circuit. The imaginary part of
the impedance disappears (the phase crosses through zero) for small
nL*/
(L*/
is the ratio of aortic flow dissipation time to cycle period, indicating the relative impedance of
the parallel L and Rc flow pathways) and large nC*
(C* is the ratio of pressure dissipation time to cycle period,
indicating the relative impedance of the parallel Rp and C
flow pathways) when the inertance and compliance resonate for a
normalized angular frequency of approximately nX = (L*C*)
1/2. The crossover corresponds to
the first imaginary term X1 in Eq. 4, canceling the second
imaginary term X2. In the patients with inertial character, the
normalized crossover frequency was ~1.4-1.8, which was close to
the resonant frequency. At the crossover frequency,
nL*/
was small, allowing
the first real term R1 to be small. In addition, nC* was large,
allowing the second real term R2 also to be small. The impedance,
therefore, exhibited a precipitous drop near the crossover frequency
(Fig. 12).

View larger version (24K):
[in this window]
[in a new window]
|
Fig. 12.
Normalized modulus terms vs. normalized frequency (Z*; see Eq. 4) for patient 2 for RLRC model (A) and RLRC2
model (B). Thick line is resultant modulus. X1, X2, R1, R2: 1st
and 2nd imaginary and 1st and 2nd real terms of Eq. 4.
|
|
The physical interpretation of this modeled behavior is that the flow
substantially bypassed the resistance of the proximal arteries and
found a sufficiently large compliance downstream such that it offered
little back pressure. For the patients with inertial behavior studied
here, this behavior appears unrealistic. For patients 1 and
3, nC* was large enough, but
nL*/Rc* was not small enough at the
crossover frequency to result in a sharp minimum in modulus.
For the adults represented by Nichols et al. (8), nC* was
large, but nL*/Rc* was not small at the
crossover frequency; thus a shallow minimum results. In Eq. 4,
both R2 and X2 drop sharply with increasing frequency. When the falling
X2 matches the magnitude of the rising X1, these two terms cancel at
the crossover frequency nX. Similarly, R1 rises to
match the magnitude of the decreasing R2 at a normalized frequency of
nR =
1/4 × (L*C*)
1/2 for
small nL*/Rc* and large nC*. When
these frequencies are close to one another, which occurred for the
range of patient values of
of
0.2-0.5, the potential for a sharp minimum in modulus exists.
(Note in Fig. 12 that the R1-R2 match occurs near the X1-X2 crossover.)
This sharp minimum is prevented if the magnitude of R1 rises above that
of X1 before the crossover frequency. This matching frequency, called
the bypass frequency, is given by nb =
/L* and
corresponds to the frequency at which the impedance of the inertance
begins to exceed that of the Rc. Thus the ratio of the bypass and crossover frequencies controls the sharpness of the modulus
minimum. If n* = nb/nX
is greater than one, then a sharp minimum will occur. For patient
2, shown in Fig. 12, n* = 6.71, the bypass frequency
occurred beyond the maximum frequency on the graph, and a sharp minimum
resulted. The RLRC2 model provided reasonable fits for patients
1 and 3 and for the adult, for which n* = 1.68, 1.07, and
0.65, respectively, but excessively sharp minima for the other patients
with larger n* (n* = 6.7, 7.8, 5.2, and 7.8 for patients 2, 4, 5, and 6, respectively).
This sharp minimum, which depends on a delicate balance of inertial and
compliant impedance, does not occur with the RLRC model, which produced
the best quantitative fits and also provided qualitative shapes better
matched to the patient data. Its advantageous features for the patient
data included a more gradually and monotonically increasing phase with
increasing frequency at high frequency and a softer minimum in modulus.
The impedance of the RLRC model is given by
|
(5)
|
which
may be nondimensionalized as
|
(6)
|
where
the dimensionless parameters are the same as for the RLRC2 model,
except
= Rp/Z0 and Z0 = Rc + Rp for this circuit. For the RLRC model, the minimum
modulus is defined by
where the
imaginary terms cancel and where the second real term in Eq. 6
is small. (Note in Fig. 12 that R1 =
provides a floor for the
modulus.) In both the infant/child and adult data, the second real term
was only a few percent of
at the
crossover frequency, which was again near
(LC)
1/2.
The physical interpretation is that the aortic compliance is large
enough to accommodate the flow through the Rc, thereby decreasing impedance to near its minimum, before the frequency becomes
high enough for inertance to influence the impedance. In this model,
however, impedance is limited to a minimum of
by the series resistance.
Inertance serves only to increase impedance above the value of
with increasing frequency. This
concept is consistent with flow through relatively stiff conduits and branches, in which fluid inertia may grow to dominate viscous dissipation in the flow behavior, but viscous dissipation does not
disappear and is never bypassed.
The RLRC model performed better than the RCR and RLRC2 models for all
infant, child, and adult data in terms of lower NRMS, AIC, and SC
(Table 3). The RLRC2 model produced lower NRMS values than the RCR
model except for the adult data and produced lower AIC and SC except
for the adult data and patient 3. Stergiopulos et al. (17)
advocated the RLRC2 model over the RLRC and RCR models for fitting the
response of dogs and adult humans; however, the present results are, to
our knowledge, the first quantitative comparison of the fits of the two
four-element models to human data. Yoshigi and Keller (20) compared the
performance of 18 models with two to five elements (including RCR,
RLRC, and RLRC2) in matching the impedance of chick embryos and found
that the RLRC model and a three-element model resulting from the
elimination of the proximal resistor from the RLRC model produced the
best fits.
Westerhof et al. (18a) determined that, among many adult mammals, the
ratio of Rc to Z0
(Rc/Z0) was a constant. In
the six patients examined in this project,
Rc/Z0 dropped with increasing weight
for the RCR, RCR2, and RLRC models but increased for the RLRC2 model,
although none of the correlations were strong. The value of
Rc/Z0 of ~0.1 for the
heaviest subjects compared with the Westerhof et al. constant value of
0.055 and the value of 0.042 obtained from the Nichols et al. (8) data
suggests that this ratio may change during the development of humans.
Such development appears to include a rapid reduction in Rp
and a slow reduction in Rc, as discussed above.
Westerhof et al. (18a) also found that the ratio of pressure
dissipation time (RpC) to heart cycle period T
(RpC/T) was a constant, hypothesizing that a lower
limit of this ratio was necessary from a cardiovascular function
perspective to maintain diastolic pressure high enough to provide
adequate coronary flow. C* =
RpC (the same as
RpC/T except for a factor of 2
) increased for
all models, although correlations were not strong except for the RCR2 and RLRC2 models (Fig. 13). The strong
increase in C dominated the decrease in Rp and the increase
in T with weight to cause the increase in the ratio. The
differences in this parameter were strong between infants and children
for the RLRC and RLRC2 models (RCR2: P = 0.575, RLRC:
P = 0.073, RLRC2: P = 0.007) and between infants and
children vs. adults only for the RCR2 model (RCR2: P = 0.071, RLRC: P = 0.443, RLRC2: P = 0.156). The
value of RpC/T of ~3 for the heaviest subjects
compared with the values calculated from the Nichols et al. (8) data of
3.7 and 4.1 for the RCR2 and RLRC models, respectively, suggests that
this ratio may continue to rise into adulthood. However, the Westerhof
et al. (18a) value of 2.19 for adult humans suggests that this ratio
may reach a maximum at some intermediate age before falling to the
adult value. Such behavior would be consistent with a drop in
compliance in older individuals with constant Rp and heart
rate. More data are required to conclusively support either suggestion.

View larger version (23K):
[in this window]
[in a new window]
|
Fig. 13.
Ratio of pressure dissipation time to heart cycle period (C*).
Correlations with patient weight are as follows: RCR2, r = 0.991 and CI95 = 0.916 < r < 0.999; RLRC,
r = 0.310 and CI95 = 0.670 < r < 0.896; RLRC2, r = 0.930 and CI95 = 0.484 < r < 0.993. Statistics for RCR model are not given because of
inaccuracy of C values.
|
|
The behavior of compliance per unit weight depended strongly on the
model. Compliance per unit weight correlated strongly with weight only
for the RLRC2 model (RC: r = 0.707, CI95 =
0.246 < r < 0.965; RC2: r = 0.795, CI95 =
0.046 < r < 0.977; RC3: r = 0.807, CI95 =
0.013 < r < 0.978; RCR2:
r = 0.860, CI95 = 0.159 < r < 0.984;
RLRC: r = 0.469, CI95 =
0.553 < r < 0.928; RLRC2: r = 0.936, CI95 = 0.517 < r < 0.993). The mean compliance
per unit weight of the patients was nearly the same as in adults for the RLRC model but varied significantly between infants and children vs. adults for the RCR2 and RLRC2 models (RCR2: P < 0.001, RLRC: P = 0.931, RLRC2: P < 0.001). No significant
differences were found between infants and children except in the RC3
model (RC: P = 0.428, RC2: P = 0.265, RC3: P = 0.053, RCR2: P = 0.225, RLRC: P = 0.695, RLRC2: P = 0.188).
The ratio of flow dissipation time L/Z0 to heart cycle
period T (L*) was also evaluated (Fig.
14). This flow dissipation time represents the characteristic time for the inertia in the flow to be
dissipated by the total vascular resistance. No strong correlation was
apparent in this ratio with patient weight. However, L* varied significantly between infants vs. children but not between infants and
children vs. adults (RLRC: P = 0.165 and RLRC2: P = 0.880 for infants and children vs. adults; RLRC: P = 0.039 and
RLRC2: P = 0.063 for infants vs. children). L* was low in
infants, high in children, and low in adults. High L* was connected
with inertial behavior in the children.

View larger version (19K):
[in this window]
[in a new window]
|
Fig. 14.
Ratio of flow dissipation time to heart cycle period (L*). Correlations
with patient weight are as follows: RLRC, r = 0.270 and
CI95 = 0.693 < r < 0.887; RLRC2,
r = 0.662 and CI95 = 0.958 < r < 0.324.
|
|
Scaling arguments were used to compare wave reflection effects in the
pediatric patients to those in adults. Wave speed (W) may be estimated
by
|
(7)
|
where
Va is arterial volume. If
is assumed to be constant with patient
weight and Va is assumed to scale directly with patient weight, then
the 2-16 times smaller compliance in infants and children might
combine with the 4-10 times smaller volume to produce little
difference in wave speed. Indeed, when compliance per unit weight was
calculated from the pediatric patients and compared with values for the
average adult, no trend or strong correlation with weight was found for
the RLRC model. With the assumption that artery length scales with the
cube root of weight, wave reflection times estimated from the wave
speeds and artery lengths were 1.6-2.2 times shorter in the
infants and children than in the average adult. However, the heart
cycle periods in the patients were also 1.5-2.3 times shorter.
Thus no significant differences in wave reflection effects between
infants and children vs. adults were found. These estimates appear to
support the speculation that wave reflection time is regulated to
minimize cardiac work by avoiding reflection of high-pressure pulses
during the same systolic period (6).
In summary, these results provide the first calculations of aortic
input impedance in pediatric patients. For four of the six patients,
the results exhibited strong inertial character compared with adults,
with increasing modulus and large positive phase with increasing
frequency at high frequency. The RLRC lumped-parameter model reproduced
the character of the impedance data best. For this model, compliance
and Rp/Rc differed greatly from those of adults. Furthermore, Rc and Rp/Rc
varied considerably between infants and children, and Rp
and Z0 were well correlated with patient weight. These
comparisons demonstrate that the overall character of the circulation
is different in infants and children compared with adults and suggest
that significant changes are in progress during development from infant
to child.
We thank Transonic Systems (Ithaca, NY) for providing the
ultrasonic flow probes and meter for this project.
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests and other correspondence: M. K. Sharp,
Biofluid Mechanics Laboratory, Dept. of Bioengineering, Univ. of Utah,
50 S. Central Campus Dr., Rm. 2480, Salt Lake City, UT 84112 (E-mail:
m.k.sharp{at}m.cc.utah.edu).
Received 15 March 1999; accepted in final form 4 February
2000.