1 Department of Anesthesiology and Intensive Care Medicine, 2 First Department of Surgery, and 3 Ludwig Boltzmann Institute of Economics of Medicine in Anesthesia and Intensive Care, Community Hospital of Lainz, A-1130 Vienna, Austria
| |
ABSTRACT |
|---|
|
|
|---|
To study the effect of positive airway pressure (Paw) on
the pressure gradient for venous return [the difference between
mean systemic filling pressure (Pms) and right atrial pressure
(Pra)], we investigated 10 patients during general anesthesia for
implantation of defibrillator devices. Paw was varied under apnea from
0 to 15 cmH2O, which increased Pra from 7.3 ± 3.1 to 10.0 ± 2.3 mmHg and decreased left ventricular stroke volume by 23 ± 22%. Episodes of ventricular fibrillation, induced for
defibrillator testing, were performed during 0- and
15-cmH2O Paw to measure Pms (value of Pra 7.5 s after onset
of circulatory arrest). Positive Paw increased Pms from 10.2 ± 3.5 to
12.7 ± 3.2 mmHg, and thus the pressure gradient for venous return
(Pms
Pra) remained unchanged. Echocardiography did not reveal
signs of vascular collapse of the inferior and superior vena cava due
to lung expansion. In conclusion, we demonstrated that positive Paw
equally increases Pra and Pms in humans and alters venous return
without changes in the pressure gradient (Pms
Pra).
mean systemic filling pressure; positive end-expiratory pressure; right atrial pressure; mechanical ventilation
| |
INTRODUCTION |
|---|
|
|
|---|
POSITIVE PRESSURE VENTILATION and positive
end-expiratory pressure (PEEP) reduce cardiac output in experimental
and clinical settings primarily because of a decrease in venous return
(3, 15, 21, 23). Total venous return is determined by a
driving pressure gradient and the resistance to venous return.
According to Guyton's concept (8, 27), the driving force is the
difference between mean systemic filling pressure (Pms), which is the
equilibrium pressure in the systemic vessels under the condition of no
flow (9), and right atrial pressure (Pra), which is the back pressure to venous return. Because PEEP increases Pra (22), it has been hypothesized that the PEEP-induced reduction in blood flow may result
from a decreased pressure gradient (Pms
Pra) (1, 27). However,
as suggested at first by Scharf et al. (22) and later demonstrated in
experimental studies (5, 19), PEEP also increases Pms, thus preserving
the pressure gradient for venous return and buffering the decrease in
cardiac output (5, 22). A PEEP-induced decrease in blood flow was,
therefore, attributed to an increase in the resistance to venous return
(2, 5). Furthermore, a PEEP-induced vascular collapse of the inferior
vena cava (IVC) was observed in canine studies (4), consistent with a
vascular waterfall (20), causing the back pressure to venous return to be located upstream of the right atrium. In humans, such a vascular collapse of the IVC has been observed by Nakhjavan et al. (18) in
hyperinflated patients with emphysema. It is unknown whether this may
occur in patients without lung disease during positive-pressure ventilation.
Many clinical studies were conducted to address the change in venous
return induced by positive airway pressure (Paw), but virtually no data
regarding Pms are available in humans (10). From experimental studies
there is concern that PEEP comparably increases Pra and Pms, but this
remains hypothetical for humans. The purpose of this study was to
measure Pms and to evaluate the changes of the gradient (Pms
Pra) caused by positive Paw in humans. Therefore, we investigated
patients undergoing surgical implantation of cardioverter-defibrillator
devices under general anesthesia. These devices are designed to
automatically detect and terminate life-threatening ventricular
tachyarrhythmias by countershock therapy. During the surgical procedure
of implantation, ventricular fibrillation is repeatedly induced for
threshold testing of the device, resulting in episodes of circulatory
arrest long enough to measure Pms (10). Moreover, the study was
conducted to address the effect of positive Paw on the dimensions of
the great intrathoracic veins, which was studied by echocardiography.
| |
METHODS |
|---|
|
|
|---|
Patients and procedure.
The study was performed on 14 patients without history of lung disease
scheduled for surgical implantation of an implantable cardioverter-defibrillator device (Medtronic, Minneapolis, MN). The
protocol was approved by the local ethics committee, and written informed consent was obtained from each patient. An additional nine
patients in whom echocardiographic observations were completed are
described in Echocardiography. Anesthesia was
induced with intravenous midazolam (2 mg), etomidate (0.2 mg/kg), and
fentanyl (2 µg/kg). After neuromuscular blockade with vecuronium
bromide (0.1 mg/kg), patients were intubated and mechanically
ventilated (Cicero, Draeger, Austria) (50% oxygen-50% air, tidal
volume 10 ml/kg, respiratory rate 12 breaths/min, PEEP 3 cmH2O). Anesthesia was maintained by propofol infusion (0.4 mg · kg
1 · h
1)
and fentanyl as required. Administration of vecuronium bromide (2 mg)
was repeated according to neuromuscular monitoring. After hemodynamic
stabilization, lactated Ringer solution was infused at a rate of 200 ml/h throughout the procedure. To optimize oxygen reserve during
artificial cardiac arrest, pure oxygen was given 1 min before induction
of each episode of ventricular fibrillation until stable hemodynamics
were reestablished. After transvenous placement of the endocardial
right ventricular defibrillation electrode via the subclavian route,
circulatory arrest was induced for defibrillation threshold testing by
eight rapid ventricular pacing stimulations followed by a countershock
within the T wave, resulting in ventricular fibrillation. From the
first rapid ventricular pacing beat, radial artery pressure (Pa) did
not show any evidence of effective stroke volume. If the
first countershock for threshold testing delivered at least 11 s after
induction of circulatory arrest was unsuccessful in terminating
ventricular fibrillation, a second internal countershock was applied.
If this was unsuccessful, an external rescue shock was delivered. The
defibrillation threshold was tested in a step-down procedure with a
minimum of two episodes of ventricular fibrillation in each patient.
Physiological monitoring. Paw, Pa, and Pra were simultaneously recorded on a chart recorder (model 2400S, Gould Instruments, Hainault, UK), connected with the transducers (Pressure Monitoring Kit, Baxter-Edwards, Santa Ana, CA) via analog pressure modules (model M1006A T for Pra measurements and model M1006B T for Pa and Paw measurements, Hewlett Packard, Palo Alto, CA). The monitoring devices allowed measurement of Pra with an accuracy of at least 0.5 mmHg. Paw was measured at the proximal end of the endotracheal tube with an air-filled catheter connected with the transducer, balanced at zero level against ambient air. The blood pressure transducers were referenced to the midthorax level. Pa was measured via a cannula in the right or left radial artery (20 gauge, Ohmeda, Swindon, UK) introduced under local anesthesia before induction of general anesthesia. Pra was measured with a catheter (7F, 60 cm, CS-17752, Arrow, Reading, PA) introduced under general anesthesia via the right or left femoral vein. Depending on the patient's height, the catheter was advanced through the common iliac vein into the IVC to a length of 25-35 cm so as to lie with the tip in the right atrium. Transesophageal echocardiography, performed in three patients, verified correct position of the catheter tip as anticipated by physical examination.
To assess changes in venous return, we used the pulse contour method, which takes relative changes in the stroke area under the systolic part of the Pa curve as a measure of relative changes in left ventricular stroke volume (LVSV) (13). This substitution seems justifiable because all measurements were made during apneic conditions. During apnea, venous return equals LVSV, and after a sudden change in Paw a new steady state is reached within seconds (27). Stroke area was determined by averaging the measurements from three consecutive heartbeats during the period from 10 to 15 s of apnea as described in the Experimental protocol.Experimental protocol.
The protocol started 15 min after induction of anesthesia to allow for
hemodynamic stabilization. All measurements were done during short
apnea. Before implantation of the defibrillator device, the effect of
Paw on the intact circulation was tested. Therefore, Pa, Pra, and
stroke area were recorded as Paw was varied for 15 s at 0 cmH2O and 15 cmH2O. For 0-cmH2O
Paw, patients were disconnected from the ventilator. For
15-cmH2O Paw, inspiratory-hold maneuvers were performed
manually under continuous observation of Paw. Subsequent evaluation of
the recorded charts revealed constant sustained Paw during all
inspiratory-hold maneuvers with only minor fluctuations of Paw within a
range of 14-16 cmH2O. Generally, during the
inspiratory-hold maneuvers for testing the effect of Paw on the intact
circulation, a plateau in Pra was observed between 5 and 15 s of apnea,
and the Pa curve showed steady-state conditions from 10 to 15 s of apnea. Repeated measurements at the two Paw levels resulted in reproducible hemodynamic effects. After implantation of the device, in
a second step of the protocol the effect of positive Paw on Pms was
evaluated. Therefore, episodes of cardiac arrest for threshold testing
were performed alternately at 0- and 15-cmH2O Paw. Apnea for circulatory arrest was started 5 s before induction of ventricular fibrillation and was sustained until termination of the arrhythmia, to
stable heart action, had occurred. The sequence of Paw levels was
varied. The longest apneic period was 32 s with an episode of
circulatory arrest of 24 s. All fibrillation episodes were performed
for testing the effectiveness of the defibrillator. No additional
episode was provoked for study purposes. A representative episode of
circulatory arrest is shown in Fig. 1. Data
collection was considered successful if Pra taken at
0-cmH2O Paw remained stable throughout the study period
within a range of
0.5 mmHg.
|
Measurement of Pms. During circulatory arrest, Pms is theoretically measured by taking the value of Pra after equilibration of blood pressures in the arterial and the venous compartment. However, as equilibration is completed, vasomotor reflexes may cause overestimation of Pms. In canine preparations, a duration of circulatory arrest of 5-7.5 s was appropriate to allow sufficient completion of blood flow and to reach a plateau in Pra before reflex vasoconstriction (5). In our study, we considered Pra to reflect Pms 7.5 s after the onset of circulatory arrest (Fig. 1). During this period, Pra reached a plateau that was considered to allow accurate measurement of Pms if Pra rose <1 mmHg from 7.5 to 10 s.
As observed in canine (5, 16) and human studies (10) of circulatory arrest, Pa significantly exceeds Pra at the time of Pms measurement, indicating that equilibration of blood pressures is incomplete, which may result in underestimation of Pms. Therefore, we corrected our data by considering the ratio of arterial to venous compliance. A given amount of blood that is transferred during cardiac arrest from the arterial to the venous side of the circulation causes a large drop in Pa and a small increase in Pra. The ratio between the change in Pa and Pra reflects the ratio of venous to arterial vascular compliance. Dividing the remaining pressure gradient (Pa
Pra) at the time
Pms is taken by the ratio of compliances will, therefore, amount to the
value of Pms, which would have further increased if equilibration was complete.
Echocardiography. For clinical reasons, transesophageal echocardiography was only available in three patients undergoing implantation of a defibrillator. To complete our echocardiographic observations, after approval of the ethics committee, we studied consecutively nine additional patients with coronary artery disease before cardiac surgery: four patients with left ventricular ejection fraction >40% and five patients with left ventricular ejection fraction <40%, without history of obstructive lung disease. None of the patients was in acute cardiac congestion, and their Pra matched the Pra of the study group. All patients were in stable sinus rhythm. After induction of anesthesia and hemodynamic stabilization, we performed apneic periods during 0- and 15-cmH2O Paw as described in Experimental protocol. Inspiratory holds for 10-15 s at 25-cmH2O Paw were performed in hemodynamically stable patients with immediate release of Paw if mean Pa decreased below 50 mmHg. During these periods we measured IVC and superior vena caval (SVC) dimensions in long- and short-axes views by two-dimensional transesophageal echocardiography (SONOS 2500, Hewlett Packard). Echocardiographic images were recorded on videotape and subsequently analyzed by using the integrated software package. The diameter of the IVC was measured between the junction of the hepatic vein and the point where it opens into the right atrium (3-10 mm distal to the right atrium). The diameter of the SVC was measured between the point where it crosses the right pulmonary artery and the point where it opens into the right atrium (10-15 mm distal to the right atrium). The largest venous dimensions throughout the cardiac cycle, usually occurring during early ventricular diastole, were taken for analysis. Care was taken to observe a maximum length of the intrathoracic veins and any signs of vascular collapse during all phases of the cardiac cycle (narrowing of the proximal part of the SVC during atrial contraction was not considered to reflect compression by the expanding lungs). Because of the small number of observations from the study group, echocardiographic data of those and of the cardiosurgical patients are presented as one group.
Statistical analysis.
Data are presented as means ± SD. For statistical analysis, mean
values for Pa and Pra were evaluated by planimetric assessment of the
recorded strip charts and by averaging the measurements from three
consecutive heartbeats during the period from 10 to 15 s of apnea. If
minor fluctuations in Pra were present at the time of Pms measurement
during circulatory arrest, the mean of Pra from 7 to 7.5 s was taken to
reflect mean Pms. If more than one episode of circulatory arrest was
performed at one or both Paw levels, data for Pms were averaged. The
gradient for venous return was calculated as Pms
Pra for each
Paw level. Hemodynamic and echocardiographic data at 0- and 15- cmH2O Paw were compared by using paired t-test.
Because no absolute values for LVSV were available from the pulse
contour method, data for 15-cmH2O Paw are expressed as
relative changes in LVSV, calculated with the stroke area at
0-cmH2O Paw set as 100%. Linear regression analysis was
used to test whether relative changes in LVSV were related to changes
in the gradient for venous return (Pms
Pra). P
0.05 was considered statistically significant.
| |
RESULTS |
|---|
|
|
|---|
Surgical implantation of the cardioverter-defibrillator device was
carried out successfully in all patients, who discharged to a coronary
care unit postoperatively within 2 h. Four patients were excluded from
analysis: two patients showed variations of Pra >0.5 mmHg because of
blood loss and volume replacement. One patient required inotropic
support because of hemodynamic instability. In one patient, atrial
contractions persisted during the episodes of ventricular fibrillation,
causing large
-waves in the Pra curve, making assessment of Pms
impossible. Data from 10 patients successfully completing
the protocol were taken for statistical analysis. In one of those
patients, arrhythmia frustrated assessment of LVSV. In three cases, a
second internal countershock was required to defibrillate the heart; no
external rescue shock was applied. Demographic data and left
ventricular ejection fraction, assessed by radionuclide scintigraphy or
ventriculography within 6 wk from the study, are given in Table
1.
|
During stable hemodynamic conditions before implantation of the
defibrillator, increasing Paw from 0 to 15 cmH2O tended to decrease Pa from 70 ± 12 to 66 ± 12 mmHg (P = 0.07),
increased Pra from 7.3 ± 3.1 to 10.0 ± 2.3 mmHg
(P < 0.001), and decreased LVSV by 23 ± 22% (P = 0.01). After implantation of the defibrillator, during threshold
testing we studied a total of 27 episodes of circulatory arrest.
Varying Paw during cardiac arrest from 0 cmH2O (n = 14) to 15 cmH2O (n = 13) increased Pms from 10.2 ± 3.5 to 12.7 ± 3.2 mmHg (P < 0.001). Thus the calculated
pressure gradient for venous return (Pms
Pra) remained
unaffected by positive Paw (2.9 ± 1.3 vs. 2.7 ± 1.3 mmHg for 0- and
15-cmH2O Paw, respectively; P = 0.40, Fig.
2). In individual patients, Paw-induced
relative changes in LVSV were not correlated with changes in the
gradient for venous return (n = 9; P = 0.74) or left
ventricular ejection fraction (Fig. 3).
|
|
The pressure gradient between Pa and Pra at the time Pms was measured
was 20 ± 7 mmHg at 0-cmH2O Paw and 18 ± 4 mmHg at
15-cmH2O Paw. The ratio of arterial to venous compliance
(calculated as the ratio between the increase of Pra and the decrease
of Pa during 7.5 s of circulatory arrest) was 1:17 at
0-cmH2O Paw and 1:18 at 15-cmH2O Paw.
Therefore, transfer of volumes sufficient to lower Pa to Pms would have
raised Pms by
(at 0-cmH2O Paw) and
(at 15- cmH2O Paw) of that remaining pressure gradient.
Therefore, correction of Pms data for incomplete equilibration of blood
pressures increased Pms at both Paw levels by a mean of 1.2 mmHg to
11.4 ± 3.6 mmHg during 0-cmH2O Paw and to 13.9 ± 3.4 mmHg during 15-cmH2O Paw. Increasing Paw did not change the
calculated pressure gradient (corrected Pms
Pra) for venous
return (4.1 ± 1.4 vs. 3.9 ± 1.6 mmHg for 0- and
15-cmH2O Paw, respectively; P = 0.43).
Echocardiography in a total of 12 patients enabled us to visualize the entire intrathoracic IVC down to a region distal to the junction of hepatic veins. The SVC was imaged for a length of 3.5-5 cm up to a region cranial to the crossing of the right pulmonary artery. Increasing Paw from 0 to 15 cmH2O did not change the IVC diameter (22.5 ± 3.8 vs. 21.8 ± 3.7 mm; P = 0.14) but significantly reduced the SVC diameter (18.2 ± 1.6 vs. 14.8 ± 2.6 mm; P < 0.001). Changes in the SVC diameter were inversely correlated with Pra (P < 0.05). No signs of venous collapse were observed due to compression by the expanding lungs under study protocol conditions. However, marked narrowing of the SVC to <5 mm in diameter (according to a reduction in the cross-sectional area by >90%) was observed in two patients with 25-cmH2O Paw, whereas IVC diameter was only slightly reduced at this Paw level.
| |
DISCUSSION |
|---|
|
|
|---|
The measurement of Pms under clinical conditions is limited by the
availability of planned, unavoidable circulatory arrests. During
surgical implantation of cardioverter-defibrillator devices, we found a
clinical setting meeting this requirement. Most values for Pms derived
from our patients were within the range reported in animal studies (5,
6, 19, 27). Patients with Pms values exceeding this range were, even
without clinical signs of decompensated congestive heart failure, in
end-stage heart disease. With our study in humans, we demonstrated that
positive Paw of 15 cmH2O increased Pms and Pra equally. As
a consequence, the pressure gradient for venous return (Pms
Pra) remained unaffected by positive Paw, thus confirming
the findings of several animal studies (5, 19, 22). Positive Paw
decreased LVSV even in the absence of changes in the pressure gradient
for venous return.
Measurement of Pms during cardiac arrest is based on the assumption that complete equilibration of blood pressures had occurred at the moment Pms is recorded. Starr and Rawson (24) reported static blood pressures from postmortem studies in patients with and without cardiac congestion of 21 and 9 cmH2O (mean values), respectively. Clinical assessment of Pms requires short episodes of cardiac arrest. Baroreflex activation may occur after seconds and results in overestimation of Pms. Although some changes in vascular tone may be seen within the short time frame studied, it is unlikely that this had major effects on our data because baroreflexes are blunted under anesthesia (7), and from 7.5 to 10 s from the onset of circulatory arrest the rise of Pra never exceeded 0.5 mmHg. On the contrary, early measurements during circulatory arrest may underestimate Pms because of incomplete equilibration of blood flow. To provide rapid equilibration of blood pressures for Pms measurement, pumping blood from the aorta into the great veins has been recommended by Guyton et al. (9). However, comparable values for Pms were derived without arterial-venous pumps (7), and, more recently, Pms was determined in a canine study after 5-7.5 s of cardiac arrest (5).
At the time of our Pms measurements, 7.5 s after the onset of circulatory arrest, Pa still significantly exceeded Pra. The pressure gradient persisted even after prolonged episodes of circulatory arrest up to 24 s and is most likely due to a Starling resistor mechanism (i.e., vascular waterfall) in the precapillary vascular bed (16). Such a waterfall effect will prolong the time blood flow requires to equilibrate blood pressures and might cause underestimation of Pms. We corrected our Pms data by dividing the remaining pressure gradient by the ratio of venous to arterial vascular compliance. This method is based on the assumption that, during circulatory arrest, a given blood volume is released from the arterial system and added into the venous vasculature, causing pressure changes according to the vascular compliances. However, this is a simplified model of tissue blood flow. If arterial inflow is stopped abruptly, venous outflow may continue distal to the Starling resistor (16), making changes in arterial and venous pressures independent of the ratio of their compliances. However, our method should be accurate enough to assess the dimension of underestimation of Pms due to incomplete equilibration of pressures. Mean underestimation of Pms was 1.2 mmHg at both Paw levels, which is a small value but represents ~30% of the observed pressure gradient for venous return. Fessler et al. (5) observed a mean pressure gradient for venous return of 3.89 ± 0.26 (SE) mmHg under control conditions without PEEP, which is comparable to the 4.1 ± 1.4 mmHg gradient observed by us.
During PEEP ventilation, reflex vasoconstriction increases Pms (5, 9), which is important to counterbalance the decrease in cardiac output (5). However, even after denervation of peripheral baroreceptors and during total spinal anesthesia, PEEP-induced increases in Pms are observed (5). In our study, measurements were made during 15-s periods of positive Paw, thus avoiding profound vasoreflex activation. Therefore, our data during positive Paw may reflect the hemodynamic effect of a prolonged mechanical inspiration rather than the circulatory effect of PEEP. This may account for the relatively small increase in Pms (and also Pra) induced by 15-cmH2O Paw in our study (2.5 ± 0.6 mmHg), compared with experimental studies reporting an increase in Pms by ~4 mmHg (data only presented as figure in Ref. 5) under 15-cmH2O PEEP (5) and by 3.4 mmHg under 10-cmH2O PEEP (19). This problem could have been avoided by ventilating our patients with 15-cmH2O PEEP for several minutes and to pause at end expiration when they were fibrillated. However, defibrillator devices are implanted in patients with severe heart disease, and threshold testing is, irrespective of its low mortality, a potentially life-threatening procedure. Ventilation with PEEP would profoundly reduce cardiac output for minutes and would, therefore, reduce the patients' metabolic reserves for subsequent prolonged episodes of circulatory arrest if defibrillation were unsuccessful. For ethical reasons, we refused a study design that would expose patients to an increased risk. Finally, because cardioverter-defibrillator devices are increasingly implanted without the use of general anesthesia (25), it seems questionable to prolong general anesthesia because of study purposes in this high-risk population.
A reduction in blood flow induced by PEEP (5, 19) and short increases
in Paw (our data) without a decreased driving pressure gradient (Pms
Pra) indicates an increase in the resistance to venous return
(5). Furthermore, an increase in lung volume may create vascular
waterfalls (20) at the diaphragmatic inlet of the IVC. In canine
studies, PEEP increased the critical downstream pressure of the IVC and
SVC, below which venous return was maximal (6), suggesting that the
condition of a vascular waterfall would occur when the pressure within
a great intrathoracic vein was below this critical pressure. In the
presence of such a waterfall, the effective back pressure to venous
return would be higher than Pra and located upstream of the right
atrium, thus making changes in venous return independent from changes
in Pra. Such a vascular collapse of the IVC has been demonstrated in
dogs, especially in the left lateral position (4). In this species a
substantial part of the IVC is located cephalad of the diaphragm within
the thorax. Thus the expanding lungs may directly compress the IVC, causing vascular collapse. By contrast, in humans the IVC commonly opens into the right atrium immediately at the thoracic inlet. In a
study in critically ill patients under mechanical ventilation with PEEP
up to 10 cmH2O and mean positive Paw up to 22 cmH2O, blood pressures in the common iliac vein and the SVC
were found to be interchangeable (11). In another study, Jardin et al. (14) measured IVC diameter distal to the junction of the hepatic vein
in healthy volunteers. During spontaneous inspiration, which enhanced
venous return to the right heart, they observed a decrease in vena
caval diameter. During continuous positive Paw breathing at 15 cmH2O the IVC diameter was increased, consistent with a reduction in venous return, but the percent inspiratory decrease of
vena caval diameter was unchanged (14). The presence of a vascular
waterfall between the IVC and the right atrium should have altered the
inspiratory change of caval diameter. In our study, no evidence of
intrathoracic venous collapse under study protocol conditions was found
by using transesophageal echocardiography. The IVC dimensions were
especially unaffected by inspiratory-hold maneuvers. By contrast,
inspiratory hold significantly reduced the SVC diameter. This finding
was associated with low Pra, but signs of vascular collapse were
definitely not seen during 15-cmH2O Paw. Therefore, a
PEEP-induced vascular waterfall creating important pressure gradients
within the great veins should not be a common phenomenon in humans.
However, Nakhjavan et al. (18) observed a pressure gradient within the
IVC during spontaneous inspiration in patients with emphysema. Abnormal
caval blood flow was attributed to severe hyperinflation of the lungs
with a depressed diaphragm, thus exposing parts of the IVC to the
lungs. In our study we observed subtotal compression of the SVC in two
patients at 25-cmH2O Paw. Thus, in humans,
positive-pressure ventilation could create a vascular waterfall in the
SVC, but this may be restricted to peak inspiratory pressure and low
central venous pressure.
Because mechanical inspiration changes abdominal pressure, which has important implications for splanchnic and nonsplanchnic IVC blood flow, our data have to be discussed with respect to the concept of abdominal vascular zone conditions (26). Even without measurement of abdominal pressure, we suggest that our patients started in a physiological abdominal vascular zone 3 condition at baseline, in which IVC pressure exceeds abdominal pressure and vascular waterfalls are absent. If the respiratory muscles are paralyzed, which was the case in our study, any positive inspiratory pressure must overcome the elastance of the lung and the diaphragm before transmitting the remaining force to the abdomen. The increase in abdominal pressure will therefore be less than the increase in pleural pressure. It is unlikely that mechanical inspiration may change zone 3 into zone 2 conditions because this would require an increase in abdominal pressure that exceeds the increase in IVC pressure. Therefore, based on the lack of compression of the IVC from our observations, we suggest that zone 3 conditions were still present during 15-cmH2O Paw and that mechanical inspiration significantly affected transhepatic resistance to venous flow. This is supported by experimental findings from Brienza et al. (2), who demonstrated that PEEP, besides an increase in the back pressure to flow, affects splanchnic flow because of an increase in liver venous resistance, resulting from direct mechanical compression of the liver by diaphragmatic descent. However, in patients with increased abdominal pressure and a zone 2 abdomen at baseline (i.e., vascular waterfall at the diaphragm), the findings of our study would be expected to be different.
A limitation of our study was the substitution of the measurement of venous return, which was impossible in the clinical setting, by the measurement of LVSV using the pulse contour method. Cardiac output measurement by thermodilution was not possible because a pulmonary artery catheter could potentially interfere with the implanted right heart endocardial defibrillation electrode. However, after an acute increase in Paw, a change in venous return is transmitted to the aorta within a few heartbeats, and a new hemodynamic steady state is reached within seconds (12, 27). Changes in LVSV should, therefore, adequately reflect changes in venous return during a quasi-steady-state condition after 10-15 s of inspiratory hold. Although the pulse contour method permits adequate estimation of LVSV (13), it is not generally accepted because it requires constant downstream resistance and compliance over the time period studied. In our patients, the inspiratory-hold maneuvers resulted in a wide range of hemodynamic response, which may be the range of ventricular function. However, no relationship was found between baseline ventricular function as measured by ejection fraction and the cardiovascular effect of positive Paw. The observed range of hemodynamic response may further result from our study design because we controlled Paw and not lung volume. Differences in lung compliance may, therefore, add to the individual response to increased Paw, even in subjects without history of lung disease (17).
Transfemoral placement of the central venous catheter for measuring Pra was chosen to avoid any interference with the surgeon's subclavian approach. Correct placement of the catheter tip proximal to the right atrium was verified by echocardiography in three patients. The absence of venous collapse as assessed by our echocardiographic studies and the absence of a pressure gradient between the SVC and the common iliac vein in mechanically ventilated patients (11) make it unlikely that small variations of the catheter position in the other patients would have influenced our Pra data.
In conclusion, we measured Pms in humans and confirmed values obtained
in animal experiments. We demonstrated that an acute increase of Paw
from 0 to 15 cmH2O increased Pms comparable to the rise in
Pra. Changes in venous return occurred independent of changes in the
pressure gradient for venous return calculated as (Pms
Pra).
Vascular collapse of great intrathoracic veins, which is a common
finding in canine studies, was not observed under those conditions.
| |
ACKNOWLEDGEMENTS |
|---|
We gratefully thank Michael Zimpfer from the Department of Anesthesia and General Intensive Care, University of Vienna (Vienna, Austria), who supported our study by providing the four-channel recorder. We also appreciate the help of Kurt Neumann, who provided expert advice regarding our statistical evaluation.
| |
FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests and other correspondence: H. Jellinek, Abteilung für Anaesthesie und Intensivmedizin, Krankenhaus der Stadt Wien Lainz, Wolkersbergenstrasse 1, 1130 Vienna, Austria (E-mail: jeh{at}ana.khl.magwien.gv.at).
Received 10 March 1999; accepted in final form 13 November 1999.
| |
REFERENCES |
|---|
|
|
|---|
1.
Braunwald, E.,
J. T. Binion,
W. L. Morgan,
and
S. J. Sarnoff.
Alterations in central blood volume and cardiac output induced by positive pressure breathing counteracted by metaraminol (Aramine).
Circ. Res.
5:
670-675,
1957
2.
Brienza, N.,
J. P. Revelly,
T. Ayuse,
and
J. L. Robotham.
Effects of PEEP on liver arterial and venous blood flows.
Am. J. Respir. Crit. Care Med.
152:
504-510,
1995[Abstract].
3.
Dhainaut, J. F.,
J. Y. Devaux,
J. F. Monsallier,
F. Brunet,
D. Villemant,
and
M. F. Huyghebaert.
Mechanisms of decreased left ventricular preload during continuous positive pressure ventilation in ARDS.
Chest
90:
74-80,
1986
4.
Fessler, H. E.,
R. G. Brower,
E. P. Shapiro,
and
S. Permutt.
Effects of positive end-expiratory pressure and body position on pressure in the thoracic great veins.
Am. Rev. Respir. Dis.
148:
1657-1664,
1993[ISI][Medline].
5.
Fessler, H. E.,
R. G. Brower,
R. A. Wise,
and
S. Permutt.
Effects of positive end-expiratory pressure on the gradient for venous return.
Am. Rev. Respir. Dis.
143:
19-24,
1991[ISI][Medline].
6.
Fessler, H. E.,
R. G. Brower,
R. A. Wise,
and
S. Permutt.
Effects of positive end-expiratory pressure on the canine venous return curve.
Am. Rev. Respir. Dis.
146:
4-10,
1992[ISI][Medline].
7.
Green, J. F.
Pressure-flow and volume-flow relationships of the systemic circulation of the dog.
Am. J. Physiol.
229:
761-769,
1975.
8.
Guyton, A. C.
Determination of cardiac output by equating venous return curves with cardiac response curves.
Physiol. Rev.
35:
123-129,
1955
9.
Guyton, A. C.,
D. Polizo,
and
G. G. Armstrong.
Mean circulatory filling pressure measured immediately after cessation of heart pumping.
Am. J. Physiol.
179:
261-267,
1955.
10.
Hiesmayr, M.,
H. Jellinek,
P. Perger,
W. Augustin,
P. Simon,
C. Kainz,
and
W. Haider.
Is there a systemic waterfall in humans? Evidence from haemodynamic measurements during artificial circulatory arrest (Abstract).
J. Cardiothorac. Vasc. Anesth.
6, Suppl. 1:
A2,
1992.
11.
Ho, K. M.,
G. M. Joynt,
and
P. Tan.
A comparison of central venous pressure and common iliac venous pressure in critically ill mechanically ventilated patients.
Crit. Care Med.
26:
461-464,
1998[ISI][Medline].
12.
Jansen, J. R.,
and
A. Versprille.
Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation.
Intensive Care Med.
12:
71-79,
1986[ISI][Medline].
13.
Jansen, J. R. C.,
K. H. Wesseling,
J. J. Settels,
and
J. J. Schreuder.
Continuous cardiac output monitoring by pulse contour during cardiac surgery.
Eur. Heart J.
11, Suppl. 1:
26-32,
1990.
14.
Jardin, F.,
J. C. Farcot,
P. Gueret,
J. F. Prost,
Y. Ozier,
and
J. P. Bourdarias.
Echocardiographic evaluation of ventricles during continuous positive airway pressure breathing.
J. Appl. Physiol.
56:
619-627,
1984
15.
Leithner, C.,
A. Podolsky,
S. Globits,
H. Frank,
A. Neuhold,
J. Pidlich,
E. Schuster,
T. Staudinger,
C. Rintelen,
M. Roggla,
D. Glogar,
and
M. Frass.
Magnetic resonance imaging of the heart during positive end-expiratory pressure ventilation in normal subjects.
Crit. Care Med.
22:
426-432,
1994[ISI][Medline].
16.
Magder, S.
Starling resistor vs. compliance. Which explains the zero-flow pressure of a dynamic arterial pressure-flow relation?
Circ. Res.
67:
209-220,
1990
17.
Mundie, T. G.,
D. Easa,
K. C. Finn,
E. L. Stevens,
G. Hashiro,
and
V. Balaramann.
Effect of baseline lung compliance on the subsequent response to positive end-expiratory pressure in ventilated piglets with normal lungs.
Crit. Care Med.
22:
1631-1638,
1994[ISI][Medline].
18.
Nakhjavan, F. K.,
W. H. Palmer,
and
M. McGregor.
Influence of respiration on venous return in pulmonary emphysema.
Circulation
33:
8-16,
1966
19.
Nanas, S.,
and
S. Magder.
Adaptions of the peripheral circulation to PEEP.
Am. Rev. Respir. Dis.
146:
688-693,
1992[ISI][Medline].
20.
Permutt, S.,
and
R. L. Riley.
Hemodynamics of collapsible vessels with tone: the vascular waterfall.
J. Appl. Physiol.
18:
924-932,
1963
21.
Quist, J.,
H. Pontoppidan,
R. S. Wilson,
E. Lowenstein,
and
M. B. Laver.
Hemodynamic response to mechanical ventilation with PEEP.
Anesthesiology
42:
45-55,
1975[ISI][Medline].
22.
Scharf, S. M.,
P. Caldini,
and
R. H. Ingram.
Cardiovascular effects of increasing airway pressure in the dog.
Am. J. Physiol. Heart Circ. Physiol.
232:
H35-H43,
1977
23.
Schuster, S.,
R. Erbel,
L. S. Weilemann,
W. Lu,
B. Henkel,
S. Wellek,
H. Schinzel,
and
J. Meyer.
Hemodynamics during PEEP ventilation in patients with severe left ventricular failure studied by transesophageal echocardiography.
Chest
97:
1181-1189,
1990
24.
Starr, I.,
and
A. J. Rawson.
Role of the "static blood pressure" in abnormal increments of venous pressure, especially in heart failure.
Am. J. Med. Sci.
199:
27-55,
1940[ISI].
25.
Stix, G.,
A. Anvari,
M. Grabenwoger,
C. Merhaut,
Z. Turel,
G. Laufer,
and
H. Schmidinger.
Implantation of a unipolar cardioverter/defibrillator system under local anaesthesia.
Eur. Heart J.
17:
764-768,
1996
26.
Takata, M.,
R. A. Wise,
and
J. L. Robotham.
Effects of abdominal pressure on venous return: abdominal vascular zone conditions.
J. Appl. Physiol.
69:
1961-1972,
1990
27.
Versprille, A.,
and
J. R. Jansen.
Mean systemic filling pressure as a characteristic pressure for venous return.
Pflügers Arch.
405:
226-233,
1985[ISI][Medline].
This article has been cited by other articles:
![]() |
J. R. Mitchell, W. A. Whitelaw, R. Sas, E. R. Smith, J. V. Tyberg, and I. Belenkie RV filling modulates LV function by direct ventricular interaction during mechanical ventilation Am J Physiol Heart Circ Physiol, August 1, 2005; 289(2): H549 - H557. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Schipke, G. Heusch, A. P. Sanii, E. Gams, and J. Winter Static filling pressure in patients during induced ventricular fibrillation Am J Physiol Heart Circ Physiol, December 1, 2003; 285(6): H2510 - H2515. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Malhotra, V. V. Muse, and E. J. Mark Case 12-2003 - An 82-Year-Old Man with Dyspnea and Pulmonary Abnormalities N. Engl. J. Med., April 17, 2003; 348(16): 1574 - 1585. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |