|
|
||||||||
1 The Krannert Institute of Cardiology and Department of Medicine, Indiana University School of Medicine, and 2 Department of Electrical Engineering, Indiana University-Purdue University at Indianapolis, Indianapolis, Indiana 46202
| |
ABSTRACT |
|---|
|
|
|---|
A force transducer with variable sensitivity and speed is described. Its moving element is a cantilever beam that projects vertically into a muscle bath. A brace constrains bending of the beam to a short, proximal "hinge." Rotation of the beam about the hinge is amplified 30-fold by an optical lever consisting of a laser diode beam reflected from a mirror on the cantilever to a photodiode pair. This design places the electrical components at a distance from the damp environment of the muscle bath. Large changes in sensitivity and speed can be obtained by substituting different cantilevers. Smaller changes can be made by varying the length of the hinge. A transducer with a 6-mm cantilever optimized for the study of single, skinned skeletal muscle fibers is described in detail. This device had a resonant frequency of 22 kHz and sensitivity such that the total root-mean-square noise in the circuit was more than 500-fold smaller than the expected maximum force. Variations of this device with orders of magnitude different sensitivities are also described.
tension measurement; cantilever beam; muscle contraction
| |
INTRODUCTION |
|---|
|
|
|---|
A MAJOR LIMITATION IN STUDIES of muscle contraction is the absence of an ideal force transducer; i.e., one with high sensitivity, low noise, high speed, low compliance, and linearity that is also rugged, easy to repair, and resistant to humidity. The need for high-speed transducers for use with intact skeletal or smooth muscle and with high sensitivity for skinned fibers, single fibers, or myotubules has led to many different designs (1-3, 6). The device described here integrates the above features of an ideal transducer to a significant degree. It has a cantilever beam that can be projected into a muscle bath. Rotational movement of the cantilever is amplified by an "optical lever" derived from a laser diode beam reflected from a mirror on the cantilever to a pair of diodes. This arrangement places the light source and sensor away from the damp environment of the bath. The dynamic properties of the device can be altered over a small range by adjusting the length of a compliant "hinge" at the proximal end of the beam, and over a larger range by substituting different beams. The principal description here is of a device for studying single skeletal muscle fibers that develop up to 1 mN force. Also described are cantilevers for studying muscles that develop orders of magnitude more and less force.
| |
METHODS |
|---|
|
|
|---|
Principle of the Design
The moving element is a cantilevered aluminum alloy beam braced to allow flexion only in a short "hinge" region at the proximal end (Fig. 1C). A laser diode beam is reflected from a stationary mirror (ms in Fig. 2B) to a mirror attached to the proximal end of the cantilever (mc in Fig. 2B) and then focused on a photodiode pair. The two mirrors are set at slightly more than a 90° angle, so that the beam is reflected back toward its origin but passes by the front of the laser diode. This arrangement greatly reduces sensitivity to movements that do not alter the angle between the mirrors, and it moves the light source and sensor away from the extension arm and from the muscle bath. The beam between the mirrors parallels the axis of the cantilever, so that the length of the hinge (dimension of segment b in Fig. 1A) can be altered without altering the position of the light beam striking the mirrors. Such alterations in hinge length allow the compliance of the beam, and therefore its sensitivity and resonant frequency, to be adjusted.
|
|
The cantilever principally described was made from sheet aluminum milled flat in the clamped portion and hinge (segments a and b in Fig. 1A) to ~0.8 mm thick × 0.8 mm wide. The thickness of the beam increases abruptly distal to the hinge (from 0.8 to 2.5 mm), to brace against flexion (segment c in Fig. 1A). Below this brace, the beam tapers both in thickness and width toward the distal tip. The back upper edge of the brace is angled at ~38° to the horizontal and serves to support a 2.5 × 4 × 0.2-mm front surface mirror. A more robust but slower cantilever for measuring isometric force in permeabilized whole muscle was made from titanium folded as indicated in Fig. 1B. The advantages of the titanium are that it can be folded and that it is more resistant to corrosion, but its greater density diminishes its speed.
The length of the free end of the cantilever is 6 mm, with the proximal 2 mm comprising the hinge, brace, and mirror support (segments b and c in Fig. 1A). This length is the minimum for our applications. It allows up to 2 mm to be projected into a solution bath and ~2 mm of protection against a rise in the level of the bath to reach the moving mirror (segment d in Fig. 1A). To attach muscles to the transducer, a hook made from an 8 mm × 100-µm stainless steel wire was cemented perpendicularly to the tip of the cantilever with a thin layer of water-resistant epoxy (see side views in Fig. 1). Muscle ends were held with aluminum T clips (5), and holes in the clips were passed over the hooks to connect the muscle to transducer and servomotor.
The body of the transducer is made in three sections (Fig. 2A) that can be adjusted relative to each other during alignment of the beam. The middle section is an aluminum block that holds the laser diode, serves as a heat sink, and is attached to the apparatus. A "front piece" mounted toward the muscle holds both the stationary mirror and the clamp that grips the cantilever beam (Fig. 2B). This section can be moved vertically, i.e., along the axis of the cantilever beam, to adjust the position of the laser beam on both mirrors. Small rotations of the front piece (about an axis perpendicular to the bearing surface) cause horizontal displacements of the laser beam reflected from the mirrors. Such rotation is used to center the beam horizontally on the photodiodes. A "backplate" holds the photodiode pair.
Not shown in the diagrams are a framework and wrap made of hydrophobic strips (Parafilm, M-type, American National Can, Chicago, IL) that cover but do not touch the moving element. The strips offer additional protection against floods and slow thermal changes caused by air drafts.
Components
Laser diode. A 5-mW, 635-nm-wavelength laser diode with constant-current driver and focusable lenses was obtained from Merideth Instruments (Glendale, AZ; part no. VDM21). The assembly is mounted in a 9-mm-diameter tubular housing. The lens diameter is 5 mm, and its focal length is 4.8 mm. The light beam exiting the lens has an approximately oval cross section, with major and minor axes 5 × 3 mm. This reduces to ~3 × 1.8 mm at the moving mirror and to its smallest size, ~300 µm vertical diameter, at the diode pair, 135 mm from the lens and 90 mm from the moving mirror. Because the angular change of the light beam is twice the angular change of the mirror, the ratio 15:1 of rotating light beam length (90 mm) to cantilever length (6 mm) creates a 30:1 ratio of light spot movement to tip movement.
Photosensor. A photodiode pair with total area measuring 1 × 3 mm divided along the long axis, so that each diode measured 0.5 × 3 mm, was obtained from Hamamatsu (US representatives in Park Ridge, IL; part no. S2722). Figure 2, C and D, shows, respectively, the circuit used to measure the separate outputs of the diode when characterizing the apparatus and the circuit used in normal operation to obtain the difference between the outputs of the two diodes. The circuits are mounted on the transducer, near the diode pair, to reduce noise along the connecting cables.
Back-plate. The photodiode pair is mounted on a 6.3-mm-wide pedestal on a larger Plexiglas plate. This pedestal fits into a 6.3-mm-wide slot in the back of the center section of the back-plate, and the Plexiglas plate is mounted on rods that allow the plate and diodes to be moved vertically. Springs on the lower ends of the rods force the plate upward against a screw that adjusts the vertical position of the diodes so that the focused laser beam falls equally on both diodes.
Front piece. The front piece is a 19 × 19-mm-square plate with an angular slot milled partially through its center (Fig 2A). The base of the slot, at 60° to the horizontal, holds the stationary mirror (ms in Fig. 2B). The laser beam strikes this mirror at 30° to the horizontal and is then directed vertically downward to meet the moving mirror (mc in Fig. 2B). The apex of the angle at the end of the slot is cut off ~0.5 mm from its end, and the front piece is made of steel to provide a rigid clamp for the cantilever. A brass front piece having a sharper apex was found to be too flexible. The upper front face of the front piece is milled away to clear the line of sight from microscope to muscle. Slots for the attachment screws provide both vertical adjustment and small rotational adjustment of the front piece.
Cantilever beam. Cantilever beams were made of aluminum alloy, type 7075, hardened to T-6 temper (obtained from Pioneer Aluminum, Wichita, KS). The metal was first cut and milled to have the exact thickness of hinge and clamped region and the rough outlines of the distal extension. The final shape of the element (Fig. 1A) was achieved by hand-grinding and filing. With practice, a new element could be made from the milled sheet metal in about an hour.
Calculations
The resonant frequency (f) of a mass oscillating on a spring of stiffness (s) is given by
|
(1) |
|
(2) |
| |
RESULTS |
|---|
|
|
|---|
Light Beam Width, Sensitivity to Movement, and Noise
The sensitivity of the photodiodes to movement of the light beam was measured by substituting a longer cantilever beam that had a more compliant hinge that could be flexed easily with a servomotor making ramp cycles to move the spot across the diode. The separate outputs from the two diodes, together with their sum, are shown in Fig. 3A. The difference between them, representing the sensitivity of the device to movement of the laser beam, is shown in Fig. 3B. The total vertical width of the light beam, ~0.3 mm, was estimated from the difference between the width of the diode pair, 1 mm, minus the width of the plateau in the sum of the diode currents, ~0.7 µm. This beam width should be less than the width of the diode pair, if the effect of light movement across their center is not to be offset by movement of light on and off their outside edges.
|
The data in Fig. 3 were obtained with the laser beam focused to as small a spot as possible at the diode pair. To assess whether the focus was critical, the maximum slope in the middle of the difference records (Fig. 3B) was measured when the laser diode was displaced along its axis in its holder. The slope increased by <5% when the diode was displaced up to 8 mm in either direction, indicating that the sensitivity of the device was not critically dependent on the focus of the laser beam.
The slope of the difference trace in Fig. 3B is 0.33% of the full signal per micrometer. When used with the ×10 amplifier, shown in Fig. 2D, and a slightly brighter laser diode, this slope corresponded to ~800 mV/µm movement of the laser spot at the diode. An optical lever ratio of 30:1 makes the sensitivity of the device to tip movement 24 V/µm.
The root-mean-square (RMS) noise of the electrical signal with the laser on was 400 µV (measured with a Tektronix TDS-340A digital oscilloscope.) Most of the peak-to-peak noise is contained within a band ~4 times the RMS noise, equivalent to 1.6 mV. A goal of the design was to have sufficient sensitivity for this band to be <1% of the maximum expected signal. For single skeletal muscle fibers that develop 1 mN, the desired sensitivity was at least 160 mV/mN or 1.6 V/g.
Sensitivity and Resonant Frequency
The sensitivity and resonant frequency are plotted as a function of hinge length in Fig. 4. As hinge length was increased from 0.2 to 1.5 mm, sensitivity increased from 2.2 to 5.2 V/g, whereas resonant frequency declined from 22 to 12 kHz. The plot of sensitivity against hinge length had a positive intercept and was not linear. This positive intercept suggests compliance of structures outside the hinge equivalent to a few tenths of a millimeter of additional hinge length. This compliance could be in either the clamp or adjacent parts of the cantilever. The lack of linearity is due, at least in part, to reduction of the optical lever ratio as the cantilever was lengthened.
|
At each hinge length, the output of the transducer was found to be linear for loads up to 2 g. At the shortest hinge length, four times the RMS noise corresponded to 7.3 µN of force or ~0.7% of the expected maximum force. Thus the ratio of maximum signal-to-noise was ~550, well above the stated goal of 400.
The resonant responses of the device to sudden unloading at the three
hinge lengths are shown in Fig. 5,
A-C. The dominant resonance at the shortest hinge length
was 22 kHz, with lesser resonances at higher frequencies. These other
resonances were less obvious at longer hinge lengths, where the
dominant resonance was of lower frequency and greater amplitude. The
sources of these higher resonances were not identified. The dominant
resonances seen after rapid unloading decayed with time constants
between 2.5 ms for the short hinges and 8.9 ms for the longest. As
described in Fig. 6, the steps typically
used with a muscle attached did not excite these resonances
appreciably. The dependence of resonant frequency on hinge length was
greater than expected on the basis of the changes in sensitivity and
compliance. This greater change was due to increased equivalent mass of
the cantilever element as the hinge was lengthened.
|
|
Equivalent Mass
The equivalent mass of the cantilever at its tip was determined from the ratio of frequencies observed with and without a 2.9-mg piece of solder glued to the tip. At the 0.5-mm hinge length, these frequencies were 11 and 18 kHz. At 1.5-mm hinge length, they were 9 and 12 kHz. Substituting the ratios of the frequencies and the weight of the lead (and its adherent glue) into Eq. 2 yielded an equivalent mass of 1.8 mg at the short hinge length and 3.1 mg at the long length. The relatively large difference in the equivalent mass of the cantilever at the two hinge lengths further suggests that a substantial part of the equivalent mass is in the proximal structures, including the brace, mirror, and mirror support. These structures move a proportionately greater distance from the fulcrum when the hinge is lengthened, so that they make a proportionately greater contribution to the overall equivalent mass of the cantilever.Similar calculations suggest that the equivalent mass of the cantilever was ~1.7 mg at the shortest hinge length.
Cantilever Designs for Other Force Ranges
The cantilever element described above was designed for the study of single, skinned skeletal muscle fibers that develop a maximum of 1 mN of force. Additional elements were made for two other preparations that develop very different levels of force. One was for studying cultured myotubes. This element had approximately the same length and width as that shown in Fig. 1A, but the hinge was approximately fourfold thinner, i.e., 0.2 mm thick. As with the element described in detail above, the cantilever was tapered in both length and width toward the tip. When the hinge length was as short as it could be, ~0.3 mm, the element was ~15 times more sensitive, i.e., 30 V/g, than the standard element set at the same hinge length, and its resonant frequency was 9 kHz.At the opposite end of the sensitivity spectrum, a cantilever was made for measuring isometric force in permeabilized muscle preparations that develop 10-30 mN force. Because ruggedness is more important than speed in this application, the cantilever was made of titanium, and its width was not tapered (Fig. 1B). To increase sensitivity, as well as to provide greater clearance above the solutions, the cantilever beam length was increased to 10 mm. With a hinge length of ~0.1 mm, the sensitivity was 3.1 V/g and the resonant frequency was 7 kHz, several thousand times greater than needed for this application.
Use of the Transducer in Muscle Experiments
The performance of the transducer in typical physiological experiments is shown in Fig. 6, which illustrates two types of muscle experiments done on the same strip of tracheal smooth muscle, which was 7.5 mm long and weighed 1.2 mg. Figure 6A shows a slow recording made with a digitization rate of 1 sample/100 ms to record a tetanus produced by 60-Hz electrical field stimulation lasting 12.3 s. This record has been spliced to exclude a period of rapid recording associated with an isotonic step. At 12 s after the onset of stimulation, the recording rate was increased to 1 sample/ms, and after 20 ms, servo-control was changed to keep force constant at about one-half of the isometric value for a period of 200 ms before the original length was restored. The lowest trace in Fig. 6B shows the time course of force. The upper pair of superimposed records are of overall muscle length (dotted trace) and central segment length (solid trace), monitored with a "spot follower device" (5). The gain of the records, indicated by calibration bars at either end, was adjusted so that the records superimposed over the last 100 ms of isotonic shortening. The main point to be made from this example is that there is no evidence of the lightly damped 22-kHz resonance of the transducer, either in the force response or in the performance of the servo-loop.The records in Fig. 6C illustrate a transient tension response to a rapid stretch complete within 200 µs. For this record, the servomotor was controlled from the central segment length sensor, and a stretch complete within 200 µs was applied to the otherwise isometric segment. Force rose in association with the stretch and then relaxed toward its isometric level with a complex time course. Again, even with this rapid step there is no suggestion of the inherent resonant frequency of the transducer.
| |
DISCUSSION |
|---|
|
|
|---|
A major advantage of the present device is that cantilever beams with very different characteristics can be substituted in the same apparatus. Such substitutions allow the other components of the device to be used for a wide range of experiments. Although the apparatus described had characteristics for use in single, skinned skeletal muscle fibers, which produce up to 1 mN of force, the device can be used with preparations that develop greater and lesser forces. When similar dimensions of cantilever are used, the resonant frequency of the device will vary in inverse proportion to the square root of sensitivity.
Another advantage is that the elements are very rugged. A general rule of force transducers is that they should operate near their breaking point (7) to make them as light and therefore as fast as possible. This proved to be difficult in the present structures because such delicate elements frequently broke during the grinding and filing required for fabrication. The elements developed here were found to be much stronger than the minimum required by the muscle preparations. Although it seems likely that we could have developed the skill to make narrower and therefore lighter cantilevers, the observation that a substantial fraction of the equivalent mass of the element was in essential proximal structures (braces, mirror, etc.) suggested that the gains were likely to be small.
Comparison with Other Transducers
All force transducers are "moving element" types, i.e., very small movements of a compliant element are sensed by a variety of techniques. These devices can be classified according to the type of movement and the method of sensing the movement. The movement can be categorized broadly as those devices that have an extension arm that can be projected vertically into a solution trough and those where the muscle pulls directly on the moving element. Transducers without extension arms can be very fast. For example, Huxley and Lombardi (6) have described a capacitance gauge transducer with a resonant frequency of 50 kHz. An additional advantage of the capacitance device is that air movement between the capacitor plates provides damping (5). The major disadvantage of these instruments is that they require that the muscle pull directly on one of the capacitor plates. When the muscles are mounted horizontally, as required for many experiments, the transducer must be mounted below the level of the solution bathing the muscle. In experiments on intact cells in stationary baths, this requirement is met by having a hook extending from the capacitor through a vertical meniscus formed by a slot in one side of the muscle chamber. Such an arrangement is not possible in experiments that require moving solutions. When we add an extension arm to capacitance transducers, for use in skinned-fiber experiments, the resonant frequency drops to a few kilohertz. This reduction occurs partly because of the large mass of the arm and its supports and partly because of compliance in the glue used to attach the arm to the capacitor plate (Y. L. Chiu and L. E. Ford, unpublished observations). With more attention to detail, Cecchi et al. (1) have made a similar capacitance gauge device with a short projection arm and a resonant frequency of 11 kHz, several times less than that for the armless device.Recently, De Winkel et al. (4) made a transducer with a 2-mm extension arm with a 70-kHz resonant frequency. Although this arm would be impracticably short for the types of experiments anticipated here, this device illustrates the increases in speed that can be achieved when the length of the projection arm is reduced toward zero.
The method of sensing movement of the compliant element in the transducer can be divided broadly into those in which the sensor is incorporated into the moving element (strain gauges, piezoelectric elements, capacitance devices) and those in which the moving element is independent of the sensor (usually photoelectric devices). The main advantage of the latter category is that the characteristics and design of the moving element are independent of the material in the sensing element. A further advantage of the photoelectric devices is that they can be used in a damp environment where variations in moisture might alter the sensitivity of other sensing devices (e.g., capacitance transducers) or short circuit the output entirely [e.g., piezoelectric devices such as that used by Chiu et al. (3)].
Another advantage of the present device is ease of fabrication. The three main sections of the device, independent of the cantilever beams, can be made to relatively low tolerances in a few hours of shop time. With practice, the cantilever beams can be made within an hour. We have previously described a photoelectric force transducer that had a resonant frequency of 6 kHz and in which a lever was glued to a twisting strip hinge (2). An aluminum foil vane attached to the moving element interrupted a light beam (from an incandescent lamp) falling on a pair of phototransistors, the output of which was used to signal force. By substituting quieter photodiodes of the type used here, we have been able to increase the resonant frequency of that device to ~15 kHz (C. Y. Seow and L. E. Ford, unpublished observations). With a brighter light source or quieter photodiodes, some of the resulting higher sensitivity might be traded for higher resonant frequency by using a stiffer twisting hinge. The reason for developing the present instrument instead was the greater ease in making and aligning the moving elements. The older moving elements were built of carbon fiber struts glued to a twisting phosphor-bronze strip hinge. Making an element took several hours, and frequently less-than-satisfactory elements resulted when glue joints were found to have unacceptable levels of hysteresis. The absence of glue joints in the present device (except for those holding the hook and mirror) ensures an absence of hysteresis and greatly simplifies fabrication.
A final advantage of the present device over earlier photoelectric devices is that the position of the photodiode pair can be adjusted easily to bring the laser beam to the center of the two diodes. Because of the large ratio of light beam movement to cantilever tip movement, the device is relatively insensitive to small movements of the photodiodes on their adjustable holder that might result from thermal changes.
In conclusion, the transducer described has at least four significant advantages. First, the same instrument can be adapted for the study of very different sizes of muscle through substitution of different moving elements. Second, alignment of the focused laser beam on the diode pair is greatly facilitated by the division of the main transducer body into three sections that can be moved independently. Third, the electronic components are located at a safe distance from the damp environment of the muscle bath. Finally, with a resonant frequency of 22 kHz, it is the fastest transducer known to us, with a signal-to-noise ratio adequate for single-fiber study and with an arm that can be projected several millimeters into a muscle bath.
| |
FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests and other correspondence: L. E. Ford, Krannert Institute of Cardiology, 1111 West 10th St., Indianapolis, IN 46202 (E-mail: lieford{at}iupui.edu).
Received 8 April 1999; accepted in final form 3 September 1999.
| |
REFERENCES |
|---|
|
|
|---|
1.
Cecchi, G.,
F. Colomo,
and
V. Lombardi.
A capacitance gauge force transducer for isometric muscle fibres.
J. Physiol. (Lond.)
292:
1P-2P,
1979.
2.
Chiu, Y. L.,
J. Asayama,
and
L. E. Ford.
A sensitive photoelectric force transducer with a resonant frequency of 6 kHz.
Am. J. Physiol. Cell Physiol.
243:
C299-C302,
1982
3.
Chiu, Y. L.,
S. Karwash,
and
L. E. Ford.
A piezoelectric force transducer for single muscle cells.
Am. J. Physiol. Cell Physiol.
235:
C143-C146,
1978
4.
De Winkel, M. E. M.,
T. Blange,
and
B. W. Treijtel.
The complex Young's modulus of skeletal muscle fibre segments in the high frequency range determined from tension transients.
J. Muscle Res. Cell Motil.
14:
302-310,
1993[Medline].
5.
Ford, L. E.,
A. F. Huxley,
and
R. M. Simmons.
Tension responses to sudden length change in stimulated frog muscle fibres near slack length.
J. Physiol. (Lond.)
269:
441-515,
1977
6.
Huxley, A. F.,
and
V. Lombardi.
A sensitive force transducer with resonant frequency 50 kHz.
J. Physiol. (Lond.)
305:
15P-16P,
1980.
7.
McLaughlin, R. J.
Systematic design of cantilever beams for muscle research.
J. Appl. Physiol.
42:
786-794,
1977
This article has been cited by other articles:
![]() |
A. V Smolensky, J. Ragozzino, S. H Gilbert, C. Y Seow, and L. E Ford Length-dependent filament formation assessed from birefringence increases during activation of porcine tracheal muscle J. Physiol., March 1, 2005; 563(2): 517 - 527. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Maass-Moreno, T. Burdyga, R. W. Mitchell, C. Y. Seow, J. Ragozzino, and L. E. Ford Simple freezing apparatus for resolving rapid metabolic events associated with smooth muscle activation J Appl Physiol, June 1, 2001; 90(6): 2453 - 2459. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |