|
|
||||||||
Academic Medical Center, Department of Physiology, University of Amsterdam, 1105AZ Amsterdam, The Netherlands
| |
ABSTRACT |
|---|
|
|
|---|
We tested the reliability of noninvasive cardiac output (CO) measurement in different body positions by pulse contour analysis (COpc) by using a transmission line model (K. H. Wesseling, B. De Wit, J. A. P. Weber, and N. T. Smith. Adv. Cardiol. Phys. 5, Suppl. II: 16-52, 1983). Acetylene rebreathing (COrebr) was used as a reference method. Twelve subjects (age 21-34 yr) were studied: 1) six in whom COrebr and COpc were measured in the standing and 6° head-down tilt (HDT) postures and 2) six in whom CO was measured in the 30° HDT, supine, 30° head up-tilt (HUT), and 70° HUT postures on a tilt table. The COrebr-to-COpc ratio in (near) the supine position during rebreathing was used as the calibration factor for COpc measurements. Calibrated COpc (COcal sup) consistently overestimated CO in the upright posture. The drop in CO with upright posture was underestimated by ~50%. COcal sup and COrebr values did not differ in the 30° HDT position. Changes in the COrebr-to-COpc ratio are highly variable among subjects in response to a change in posture. Therefore, COpc must be recalibrated for each subject in each posture.
Finapres; stroke volume; blood pressure; tilt; standing
| |
INTRODUCTION |
|---|
|
|
|---|
THERE IS A GROWING INTEREST in monitoring physiological parameters noninvasively, both in research and in clinical situations. Blood pressure can now be measured noninvasively and continuously in the finger by using Finapres, which is based on the volume-clamp method of Peñáz (15, 22). One of the attractive derivatives of this method is the calculation of stroke volume (SV) on a beat-to-beat basis, a technique originally developed for use with the intra-arterial blood pressure wave (aorta or brachial artery). Several models (1, 7, 17, 19, 23) can be used to calculate SV from pulse contour analysis. We have many years of experience with the method of Wesseling and co-workers (23), which is based on a transmission line model of the circulation. In this model, corrections are used for age-dependent, dynamic changes in the aortic impedance due to changes in arterial pressure and heart rate (HR), and also for reflections of the pressure waveform from the periphery. Any pulse contour method will give only relative changes in SV. Calibration against an absolute method is necessary when absolute values are required.
In an earlier study (18), in which cardiac output (CO) was influenced by changes in volume status of the test subjects (by intravenous saline loading and lower body negative pressure), we demonstrated that the method of Wesseling et al. (23) can be successfully applied to the noninvasively measured pressure wave of the finger in a close-to-supine [6° head-down tilt (HDT)] position. Acetylene (C2H2) rebreathing (COrebr) was used as the absolute reference method for CO measurement. The study showed the calibration factor (COrebr-to-COpc ratio), where COpc is CO measurement by pulse contour analysis, to be stable over a period of weeks.
A major drawback of earlier validation studies (10, 18, 23) is that comparisons are made in only one, mostly supine, body position. Finapres is used in the upright and supine positions during the same session, e.g., to track beat-to-beat blood pressure changes during orthostatic tolerance testing (9, 20). Because body position tends to alter CO and the pulse wave at the same time, an evaluation explicitly testing the reliability of pulse contour analysis at different body positions seemed indicated. We therefore tested the pulse contour method applied to the Finapres waveform during orthostasis, using COrebr as a reference.
We used two different protocols: 1) standing vs. a close-to-supine (6° HDT) position and 2) four different tilt angles on a tilt table.
| |
METHODS |
|---|
|
|
|---|
Subjects
The standing vs. 6° HDT protocol was performed in six healthy male volunteers (subjects S1-S6, age 21-34 yr) during a 27-day HDT experiment (10 days in 6° HDT position) at the Deutsche Forschungsanstalt für Luft und Raumfahrt (DLR; Cologne, Germany). Written informed consent was obtained from all subjects after approval of the study by the DLR Institutional Research Review Committee.Six healthy male volunteers (age 23-30 yr) participated in the tilt-table study (subjects T1-T6). Written informed consent was obtained from all subjects. The study was approved by the Medical Ethics Committee of the Academic Medical Center of the University of Amsterdam.
Study Protocols
Standing vs. 6° HDT. Data were collected from subjects in the standing position and during a brief (20-min) episode of 6° HDT before the actual 10-day, extended HDT period, and on the first and fifth day of recovery. Rebreathing measurements were performed, using a bag-in-box system with a 3-liter rubber anesthesia bag. A gas mixture of 1.5 liters [0.3% C18O-0.5% C2H2-1.0% He-98.2% air (78.1% N2-21% O2-0% CO2-0.9% inert gases)] was used. The protocols for the experiments performed at DLR have been described elsewhere (18) in detail. In all subjects a rebreathing frequency of 24 breaths/min was used. Rebreathing measurements were performed in duplicate both in the standing and in the 6° HDT position. The washout period between duplicates was at least 2.5 min.
Tilt table. Rebreathing measurements were performed, using a bag-in-box system with a 2.3-liter rubber anesthesia bag. A gas mixture of 2 liters [1% C2H2-10% He-89% air (78.1% N2-21% O2-0% CO2-0.9% inert gases)] was rebreathed for 30 s. Fifteen seconds before rebreathing, a metronome was started for pacing the respiration rate. Then, after a normal expiration, the test subject changed from breathing normal air to breathing the bag-gas mixture by using a hand switch. Gas partial pressures (N2, O2, CO2, He, and C2H2) were monitored at the mouth by means of a mass spectrometer (Centronic 200 MGA). Gas partial pressures were corrected to a sum of 99.2%. Argon (not measured) was assumed to account for 0.8%. Respiratory flow was measured by using an Alveo Test (Erich Jaeger, Würzburg, Germany) flowmeter. In subject T1, a rebreathing frequency of 24 breaths/min was used, which was lowered in subjects T2-T6 to 15 breaths/min.
Body positions were changed by using a computer-controlled tilt table. Four different tilt angles were used,
30, 0, 30, and 70°,
with the subject supported by a shoulder support or footboard, respectively. A total of 12 measurements was performed in each subject,
each angle in triplicate. The 12 measurements were made in random
order. Between measurements, subjects were in the supine position for 5 min. After a new tilt angle was imposed, an adaptation period of 2 min
was allowed before the COrebr
measurement was started. In this protocol, the available washout period
for
C2H2 between measurements was at least 7 min.
During both the standing vs. 6° HDT and the tilt-table experiments,
the noninvasive finger arterial pressure wave was measured by a
Finapres (TNO-BMI, model 5). The finger was held at heart level, both
in the frontal and transverse plane to avoid hydrostatic level errors
during the maneuvers. All analog data were stored on tape for later
off-line analysis. Before the start of all measurement series, subjects
were familiarized with the equipment and the protocol.
Data Analysis
All analog data on tape were analog-to-digital converted at a sample rate of 100 Hz for computer analysis.Rebreathing. Data were evaluated by using the continuously
ventilated two-compartment lung model of Hook and Meyer (8). In this
model, the equilibration of partial pressures of a soluble perfusion-limited gas
(C2H2)
in two compartments (rebreathing bag and alveolar space) is described
by a biexponential process with a fast
(k1) and a
slower (k2)
rate constant, where
k1 represents the
bag-alveolar gas mixing and
k2 the
alveolar-lung capillary blood-gas transfer. Both
k1 and
k2 are derived
from a semilogarithmic plot of the end-expiratory partial pressures
against time of rebreathing. The time
0 intercept of
k2 is used for
calculating the initial C2H2
concentration. Knowing the alveolar-to-blood conductance of
C2H2,
the pulmonary capillary blood flow can be derived according to
|
c is pulmonary
capillary blood flow,
g and
b are the capacitance
coefficients of
C2H2
for gas and blood respectively, VR
is the rebreathing bag volume, SA is
the zero intercept of the slower alveolar exponential component, and
k1 and
k2 are rate constants of the fast and slow alveolar compartments, respectively. Model calculations showed the model to be relatively insensitive to
changes in respiration rate, dead space, alveolar volume, tidal volume,
and the ratio of tidal volume to bag volume.
Pulse contour analysis.
Beat-to-beat values for systolic, diastolic, and mean pressure and HR
were derived from the pressure wave. To calculate SV from the pressure
wave (SVpc), the method of
Wesseling et al. (23) was used, which relates the pulsatile systolic
area (PSA; area under the pressure curve from the start of the upstroke
to the incisura) and SVpc by the
effective characteristic impedance of the aorta (Zao). In
this method, corrections are used for age-dependent, dynamic changes in
Zao because of changes in arterial pressure and HR, which are due to
changes in aortic cross section and compliance (both pressure and age
dependent) and to reflections of the pressure waveform from the
periphery (which is influenced by HR). SV for each beat was calculated
from the PSA with the correction formula
|
|
Statistics
We estimated the separate roles of body position and measurement methods by two-way ANOVA (model with interaction). If different, Student's t-test was applied to compare the categories within a group (with Bonferroni correction).| |
RESULTS |
|---|
|
|
|---|
Standing Vs. 6° HDT
Changes in COrebr in response to the change in body position were dependent on the moment in the HDT study. When the 6° HDT is taken as the reference position, the decrease in COrebr during standing was largest just after the extended (10-day) HDT period and smallest before the extended HDT period [
1.4 l/min (
23%) and
1.0 l/min (
16%)], respectively (Table
1). This is most likely caused by the
change in cardiovascular filling state. Standing COrebr was significantly
different from the 6° HDT position for each measurement
period.
|
We calculated COcal sup in both standing and 6° HDT positions, before and during rebreathing. On average, the decrease in COcal sup in response to standing was one-half that of COrebr. Standing COcal sup was significantly different from the 6° HDT position before and on the first day after the HDT period, both during and just before the rebreathing measurement.
As shown in Fig.
1A, when
we took 6° HDT as the reference position, we found a large range in
the individual change of Rrebr/pc (the ratio between the methods) during standing.
|
Subjects S1 and S6 showed almost no change (<6%), and in subject S4 there was only a small decrease (8%) in Rrebr/pc between HDT and standing. In contrast, in subjects S2 and S5 there was a moderate-to-large decrease (14-34%) in Rrebr/pc. In subject S3 there were large changes (>20%) due to standing, but the direction of the change was inconsistent among the three measurement periods.
In the 6° HDT reference position, Rrebr/pc was constant within 9% in subjects S1-S5 and within 12% in subject S6 over the total of six measurements in each subject (3 days, duplicate measurements). In the standing position, differences were larger and ranged from 8 to 15%, with one outlier of 34% in subject S3.
When body position was changed from 6° HDT to standing, average
Rrebr/pc for six subjects changed
from 1.04 ± 0.09 to 0.96 ± 0.13 (P < 0.05) before HDT, from 1.03 ± 0.10 to 0.90 ± 0.10 (P < 0.01) on the first day of recovery, and from 1.01 ± 0.10 to 0.87 ± 0.16 (P < 0.01) on the fifth
day of recovery. Proportional changes in
Rrebr/pc when standing were
7 (range +13 to
27%),
12 (range 0 to
28%), and
14 (range +4 to
33%), respectively.
The difference between COrebr and
COcal sup for the 6° HDT
and standing postures are shown in the Bland-Altman (4) scattergrams of
Fig. 2A.
The overall differences (all subjects, all measurement days) are 0.0 ± 0.2 l/min in 6° HDT and 0.7 ± 0.8 l/min in the standing
position.
|
The MAP difference between the 6° HDT and standing position was
smallest before the HDT period (14%) and largest on the first day of
recovery (31%) (Table 2). The average
response in HR was somewhat larger after the HDT period. HR increase
during rebreathing compared with prerebreathing was only marginally
different and independent of the measurement day.
|
Tilt-Table Maneuvers
In four of six subjects (T1-T4), COrebr declined with increasing positive [head-up tilt (HUT)] tilt angle, whereas in two subjects (T5 and T6) COrebr decreased when the tilt angle was changed from 0 to 30° but did not decrease further when the tilt angle was changed to 70°. Results at 30 and 70° HUT were significantly different from the supine position. When tilted down to
30°, five of six subjects reacted with a
small-to-moderate (3-15%) decrease in
COrebr, whereas in
subject T3
COrebr increased by 0.8 l/min
(15%). Results at
30° were not significantly
different from the supine position. Average values for six subjects are presented in Table 3.
|
We calculated COcal sup in
all four positions, both before and during rebreathing. During
rebreathing, three of six subjects showed decreases in
COcal sup after being tilted
to angles of 30 and 70°. In these subjects the decreases in
COcal sup were, on average,
one-half those of COrebr. In two
subjects COcal sup did not
change, and in one subject it even increased. Also during tilt, we
found a large range in the individual change in
Rrebr/pc when using supine as the
reference position (Fig. 1B).
Average Rrebr/pc for six subjects
changed from 1.27 ± 0.34 in the supine position to 1.06 ± 0.34 (P < 0.01) at 30° and to 0.96 ± 0.39 (P < 0.01) at 70°.
Proportional changes in Rrebr/pc
during tilt were
17% (range
13 to
29%)
and
27% (range
11 to
44%), respectively. Thus in
all subjects there was an overestimation of
COcal sup in the HUT
positions compared with supine, but of different magnitude in each subject.
Bland-Altman scattergrams (4) of differences between COrebr and COcal sup are shown in Fig. 2B for each tilt position. The overall differences (all subjects, triplicate measurements) are 0.0 ± 0.6, 0.3 ± 0.5, 0.9 ± 0.6, and 1.5 ± 0.8 l/min for supine, 30° HDT, 30° HUT, and 70° HUT, respectively.
In all subjects MAP increased significantly during tilt from supine to
30 and 70° HUT. During tilt from supine to
30°, MAP increased slightly, but this increase was not significant (Table 4).
|
In five of six subjects HR increased with tilt angle (
30 to
70°). Only in subject T3 did HR
decrease with tilt angles from
30 to 30° and increase at
70° but did not reach the value at
30°. Averaged values
are presented in Table 4. The HR increase during rebreathing compared
with prerebreathing was affected only slightly by the tilt angle: 7, 6, 10, and 7 beats/min, respectively, for the four tilt angles.
Prerebreathing Pulse Contour Measurements
On the basis of analysis of the pulse contour before and during the rebreathing maneuver, we estimate that the rebreathing maneuver itself tends to increase CO.In the standing vs. 6° HDT experiments, COcal sup increased between 0.6 and 0.7 l/min (10-15%) due to the rebreathing maneuver. Results were similar for both the 6° HDT and standing postures (Table 1). In the 6° HDT position this increase was primarily caused by the increase in HR, whereas in the standing position both HR and SVcal sup increased; however, again, differences among subjects were large.
In the tilt experiments, the increase in COcal sup during the rebreathing period was largest in subject T1, whose rebreathing was performed at a higher respiration rate than in the other five subjects. On average COcal sup increased by 0.5 l/min (9.4%), 0.4 l/min (7.8%), and 0.4 l/min (9.1%) at tilt angles of 30° HDT, 0°, and 30° HUT, respectively, and by 0.9 l/min (19.0%) at 70° HUT (Table 3). This increase was primarily due to an increase in HR compared with prerebreathing, but it also occurred at 70° HUT in subjects T1 and T2 partly by an increase in SVcal sup.
| |
DISCUSSION |
|---|
|
|
|---|
In the present study we made a direct comparison of CO measurements by using pulse contour analysis at different body positions with an absolute, and noninvasive, reference method.
To investigate the rebreathing maneuver itself, we searched the
literature for studies that compared
C2H2
rebreathing to other CO measurement methods in different body postures.
We found none. One might reason that in the HUT compared with supine
position a change in distribution of the ventilation-perfusion ratio
(
A/
) is to be expected (24). In a study using
C2H2
rebreathing in dogs, Friedman et al. (6) did not find a significant
difference in estimation of pulmonary blood flow when
A/
was changed over a large range by occlusion of either one pulmonary
artery or one main bronchus. Using a three-compartment computer model, Burma and Saidel (5) found a small overestimation (<10%) of pulmonary blood flow in the upright position compared with supine, when
the
A/
distribution in the upright position as described by West (24) was
used. A small underestimation was found by Petrini et al. (16), who
used a two-compartment model with changing
A/
.
Verbanck and Paiva (21), in a model study, found only a very small
influence of
A/
on CO estimation. From these studies we concluded that the rebreathing
method can be used very well as a reference CO measurement method even
under conditions where
A/
is changing.
In the standing vs. 6° HDT experiments, we used the 6° HDT position as a reference. This position was part of the HDT study measurement protocol in which the supine position was not planned. As our later tilt experiments indicate only a small but not significant difference between supine and 30° HDT, we consider it admissible to use the 6° HDT position as a substitute reference position. In three of six subjects (S1, S4, S6), differences between the pulse contour and rebreathing methods were small or absent, but in the other three subjects we found a substantial overestimation of COcal sup compared with COrebr in the standing position. Differences between the duplicate measurements within each subject were small.
In the tilt experiments, we found COpc in the HUT position to overestimate CO in all six subjects. Among subjects, variability in responses to tilt was high using the pulse contour method, but within each subject the ratio between pulse contour and rebreathing was consistent within the triplicate measurements. In contrast to the first part of the study, in which duplicate measurements were made shortly after the initial measurement and without an in-between change of body position, triplicate measurements during the tilt experiments were made during a random sequence of 12 measurements in 4 positions. This, and the lower rebreathing frequency used, may explain the larger SDs found within the triplicate measurements in the tilt experiments (Fig. 1B).
In the supine position, the calibration factors in four of six subjects (T1-T4) were within the range found in our earlier study: 0.91-1.24 (18). In the other two subjects the calibration factors were 1.58 and 1.77 (Fig. 2B). COrebr values in the supine position for these subjects were not different from those for the other four subjects, but the variance in Rrebr/pc was larger at all tilt angles. This was due to a larger variance in the rebreathing measurements. Pulse contour measurements were consistent within each triplicate measurement. Also, in these two subjects there was no further decrease in Rrebr/pc from 30 to 70° HUT.
As already found in our earlier study (18), the rebreathing procedure itself changes the cardiovascular status of the subject: both HR and COcal sup increased compared with prerebreathing values. In the standing vs. 6° HDT study, which followed the same protocol, all measurements were made by using a rebreathing frequency of 24 breaths/min. The increase in COcal sup due to the rebreathing maneuver was ~0.7 l/min in both positions, but, especially in the standing position, differences among subjects were large. During the tilt experiments, COcal sup increased less during rebreathing in the supine reference position compared with the 6° HDT reference position (0.4 vs. 0.7 l/min) but slightly more (0.9 vs. 0.7 l/min) at 70° HUT compared with standing.
Tilt experiments were started in subject T1 by using a rebreathing frequency of 24 breaths/min. We decreased the rebreathing frequency to 15 breaths/min in the other five subjects to minimize the effect of rebreathing on COcal sup, although a less accurate estimation of the k2 rate constant can then be expected, because of less available data points for the curve fit. Despite the different breathing frequency and the consequently larger increase in COcal sup due to the rebreathing maneuver, subject T1 produced differences between COcal sup and COrebr similar to those in subjects T2-T6. Therefore, we included the results from this subject in this study. When subject T1 is excluded, the average increase in COcal sup during rebreathing is reduced to 0.3, 0.2, 0.2, and 0.7 l/min for the tilt angles of 30° HDT, 0°, 30° HUT, and 70° HUT, respectively.
In the 6° HDT and standing positions the differences between
COcal sup prerebreathing and
during rebreathing were approximately equal (0.6-0.7 l/min, Table
1). Therefore, the difference between pulse contour and rebreathing of
0.5-0.9 l/min in the standing position cannot be attributable to
the rebreathing maneuver itself. The same is true for the supine,
30° HDT, and 30° HUT positions during the tilt experiments,
where the difference between
COcal sup prerebreathing and
during rebreathing was 0.4 l/min in all three cases (Table 2). At
70° HUT, however, the increase in
COcal sup due to rebreathing
was approximately twice as much (0.9 l/min) as in the supine position.
Part of the 1.5-l/min difference between the methods at
70° HUT may therefore be attributable to the larger influence of
the rebreathing maneuver on the parameter to be measured (SV) in some
of the subjects (T1 and
T2). This does not fully explain the
large interindividual range of change in
Rrebr/pc (
11 to
44%) at this tilt angle, as both extreme values were found in
subjects in which SVcal sup
was not influenced by the rebreathing maneuver.
Why does Rrebr/pc change when the position of subjects is changed from the (near) supine to the upright position? First, this is likely not to be caused by using Finapres as an arterial pressure measurement device. Jellema et al. (11) found no difference between noninvasive finger arterial and intrabrachial COpc responses during tilt experiments. However, in both cases the method has been applied to a pressure wave more peripheral than aortic pressure, for which the method was originally developed.
Overestimation of CO in the upright position by pulse contour can be caused by a too-large systolic area (PSA) and/or a too-small estimated characteristic Zao. In the original model study of Wesseling et al. (23), hydrostatic pressure gradients as a result of the upright position are not taken into account. MAP is used in the model corrections to adapt the estimation of Zao for changes in compliance. A possible consequence is that an increasing pressure in the thoracic and abdominal aorta could mean a change toward the properties of a more rigid system with an increase in pulse-wave velocity (2) and/or a change in mismatch of characteristic impedance at the primary reflection sites (12, 13). Consequently, reflections may play a more dominant role in the systolic part of the pulse wave, also at higher HRs when ejection time is shorter. A change in the location of the primary reflection site will have the same effect. When a larger part of the PSA is due to reflections (14, 25), this has to be corrected in the model by assuming a less compliant arterial system, which results in an increase in the estimated Zao, leading to a lower estimated SV.
In our earlier validation study (18), in which saline infusion was used
to enlarge SV, only a small increase in HR (+4.1%) and no change in
MAP (+1.3%) was seen. During lower body negative pressure, which was
used to reduce SV, HR increased at most by +24% (at
40 mmHg
lower body negative pressure), whereas MAP decreased by 2.2-4.1%.
Furthermore, all measurements were performed in the same body position
(6° HDT). In the present study, responses in HR and MAP due to HUT
and standing were much larger, and both parameters increased. Maximal
responses during standing were +62% for HR and +31% for MAP; at
70° HUT they were +35% for HR and +17% for MAP. By using
Wesseling's correction formula (23), these changes in HR and MAP will
result in a decrease in the estimated Zao (and hence in a larger
SVpc), which in this situation
is probably a correction in the wrong direction.
In conclusion, the present model used to adapt the estimation of Zao to different hemodynamic parameters, i.e., HR, blood pressure, and the influence of age, does not account for situations in which a hydrostatic pressure gradient exists. Responses of COcal sup to changes in body position are widely different in our subjects and are sometimes nonphysiological. In our earlier study (18), the reproducibility of the pulse contour measurements in one body position was good over a long time period and over a large range of COs. Also, in the present study, pulse contour measurements in any one body position were well reproducible. However, pulse contour analysis using the Finapres waveform needs additional calibrations when changes in body position are involved.
| |
ACKNOWLEDGEMENTS |
|---|
The authors gratefully acknowledge the help of the Deutsche Forschungsanstalt für Luft und Raumfahrt (Dr. F. Baisch) in obtaining the rebreathing data from the HDT-88 study. These data were the result of the original contribution of Drs. H. Schulz and A. Hillebrecht in the team of Prof. M. Meyer for this international study. In that study, Dr. A. D. J. ten Harkel assisted Dr. Karemaker in the Finapres data collection. A full account of the HDT-88 results is found elsewhere (3).
| |
FOOTNOTES |
|---|
This study was supported by Space Research Organization Netherlands Grant MG020.
Address for reprint requests and other correspondence: W. J. Stok, Academic Medical Center, Department of Physiology, University of Amsterdam, Meibergdreef 9, 1105AZ Amsterdam, The Netherlands (E-mail: w.stok{at}amc.uva.nl).
Received 26 November 1997; accepted in final form 2 August 1999.
| |
REFERENCES |
|---|
|
|
|---|
1.
Alderman, E. L.,
A. Branzi,
W. Sanders,
B. W. Brown,
and
D. C. Harrison.
Evaluation of the pulse-contour method of determining stroke volume in man.
Circulation
46:
546-558,
1972
2.
Alexander, J., Jr.,
D. Burkhoff,
J. Schipke,
and
K. Sagawa.
Influence of mean pressure on aortic impedance and reflections in the systemic arterial system.
Am. J. Physiol.
257 (Heart Circ. Physiol. 26):
H969-H978,
1989
3.
Baisch, F.,
L. Beck,
J. M. Karemaker,
P. Arbeille,
F. A. Gaffney,
and
C. G. Blomqvist.
Head-down tilt bedrest. HDT'88-an international collaborative effort in integrated systems physiology.
Acta Physiol. Scand. Suppl.
604:
1-12,
1992.
4.
Bland, J. M.,
and
D. G. Altman.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1:
307-310,
1986[Medline].
5.
Burma, G. M.,
and
G. M. Saidel.
Pulmonary blood flow and tissue volume: model analysis of rebreathing estimation methods.
J. Appl. Physiol.
55:
205-211,
1983
6.
Friedman, M.,
S. A. Wilkins, Jr.,
A. F. Rothfeld,
and
P. A. Bromberg.
Effect of ventilation and perfusion imbalance on inert gas rebreathing variables.
J. Appl. Physiol.
56:
364-369,
1984
7.
Gratz, I.,
J. Kraidin,
A. G. Jacobi,
N. G. deCastro,
P. Spagna,
and
G. E. Larijani.
Continuous noninvasive cardiac output as estimated from the pulse contour curve.
J. Clin. Monit.
8:
20-27,
1992[Medline].
8.
Hook, C.,
and
M. Meyer.
Pulmonary blood flow, diffusing, capacity and tissue volume by rebreathing: theory.
Respir. Physiol.
48:
255-279,
1982[Medline].
9.
Imholz, B. P.,
J. J. Settels,
A. H. van der Meiracker,
K. H. Wesseling,
and
W. Wieling.
Non-invasive continuous finger blood pressure measurement during orthostatic stress compared to intra-arterial pressure.
Cardiovasc. Res.
24:
214-221,
1990
10.
Jansen, J. R.,
K. H. Wesseling,
J. J. Settels,
and
J. J. Schreuder.
Continuous cardiac output monitoring by pulse contour during cardiac surgery.
Eur. Heart J.
11, Suppl. I:
26-32,
1990.
11.
Jellema, W. T.,
B. P. Imholz,
J. van Goudoever,
K. H. Wesseling,
and
J. J. van Lieshout.
Finger arterial versus intrabrachial pressure and continuous cardiac output during head-up tilt testing in healthy subjects.
Clin. Sci. (Colch.)
91:
193-200,
1996[Medline].
12.
Latham, R. D.,
N. Westerhof,
P. Sipkema,
B. J. Rubal,
P. Reuderink,
and
J. P. Murgo.
Regional wave travel and reflections along the human aorta: a study with six simultaneous micromanometric pressures.
Circulation
72:
1257-1269,
1985
13.
McDonald, D. A.
Blood Flow in Arteries. Baltimore, MD: Williams & Wilkins, 1974, p. 254, 309, 351, 392.
14.
Murgo, J. P.,
N. Westerhof,
J. P. Giolma,
and
S. A. Altobelli.
Manipulation of ascending aortic pressure and flow wave reflections with the Valsalva maneuver: relationship to input impedance.
Circulation
63:
122-132,
1981
15.
Peñáz, J. Photoelectric measurement of blood
pressure, volume and flow in the finger (Abstract).
Dig. Int. Conf. Med. Biol. Eng. 104, 1973.
16.
Petrini, M. F.,
B. T. Peterson,
and
R. W. Hyde.
Lung tissue volume and blood flow by rebreathing theory.
J. Appl. Physiol.
44:
795-802,
1978
17.
Starmer, C. F.,
P. A. McHale,
F. R. Cobb,
and
J. C. Greenfield, Jr.
Evaluation of several methods for computing stroke volume from central aortic pressure.
Circ. Res.
33:
139-148,
1973
18.
Stok, W. J.,
F. Baisch,
A. Hillebrecht,
H. Schulz,
M. Meyer,
and
J. M. Karemaker.
Noninvasive cardiac output measurement by arterial pulse analysis compared with inert gas rebreathing.
J. Appl. Physiol.
74:
2687-2693,
1993
19.
Tajimi, T.,
K. Sunagawa,
A. Yamada,
Y. Nose,
A. Takeshita,
Y. Kikuchi,
and
M. Nakamura.
Evaluation of pulse contour methods in calculating stroke volume from pulmonary artery pressure curve (comparison with aortic pressure curve).
Eur. Heart J.
4:
502-511,
1983
20.
Ten Harkel, A. D.,
J. J. van Lieshout,
and
W. Wieling.
Effects of leg muscle pumping and tensing on orthostatic arterial pressure: a study in normal subjects and patients with autonomic failure.
Clin. Sci. (Colch.)
87:
553-558,
1994[Medline].
21.
Verbanck, S.,
and
M. Paiva.
Theoretical basis for time 0 correction in the rebreathing analysis.
J. Appl. Physiol.
76:
445-454,
1994
22.
Wesseling, K. H.
Finapres, continuous noninvasive finger arterial pressure based on the method of Peñáz.
In: Blood Pressure Measurement, edited by W. Meyer-Sabellek,
M. Anlauf,
R. Gotzen,
and L. Steinfeld. Darmstadt, Germany: Steinkopff Verlag, 1990, p. 161-172.
23.
Wesseling, K. H.,
B. De Wit,
J. A. P. Weber,
and
N. T. Smith.
A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification.
Adv. Cardiol. Phys.
5, Suppl. II:
16-52,
1983.
24.
West, J. B.
Ventilation, Blood Flow and Gas Exchange. Oxford, UK: Blackwell Scientific, 1990.
25.
Westerhof, N.,
P. Sipkema,
G. C. van den Bos,
and
G. Elzinga.
Forward and backward waves in the arterial system.
Cardiovasc. Res.
6:
648-656,
1972[Medline].
This article has been cited by other articles:
![]() |
B. Verheyden, H. Ector, A. E. Aubert, and T. Reybrouck Tilt training increases the vasoconstrictor reserve in patients with neurally mediated syncope evoked by head-up tilt testing Eur. Heart J., June 2, 2008; 29(12): 1523 - 1530. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Fakler, Ch. Pauli, G. Balling, H.P. Lorenz, A. Eicken, M. Hennig, and J. Hess Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery J. Thorac. Cardiovasc. Surg., January 1, 2007; 133(1): 224 - 228. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. W. J Bogert and J. J van Lieshout Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger Exp Physiol, July 1, 2005; 90(4): 437 - 446. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Gisolf, B. E. Westerhof, N. van Dijk, K. H. Wesseling, W. Wieling, and J. M. Karemaker Sublingual nitroglycerin used in routine tilt testing provokes a cardiac output-mediated vasovagal response J. Am. Coll. Cardiol., August 4, 2004; 44(3): 588 - 593. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. P. van Orshoven, P. L. Oey, L. J. van Schelven, J. M. M. Roelofs, P. A. F. Jansen, and L. M. A. Akkermans Effect of gastric distension on cardiovascular parameters: gastrovascular reflex is attenuated in the elderly J. Physiol., March 1, 2004; 555(2): 573 - 583. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Mahajan, A. Shabanie, J. Turner, M. J. Sopher, and J. Marijic Pulse Contour Analysis for Cardiac Output Monitoring in Cardiac Surgery for Congenital Heart Disease Anesth. Analg., November 1, 2003; 97(5): 1283 - 1288. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Cerutti, M. P. Gustin, P. Molino, and C. Z. Paultre Beat-to-beat stroke volume estimation from aortic pressure waveform in conscious rats: comparison of models Am J Physiol Heart Circ Physiol, September 1, 2001; 281(3): H1148 - H1155. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |