Vol. 87, Issue 1, 428-437, July 1999
Physiological effects of alveolar, tracheal, and
"standard" pressure supports
Jean-Luc
Diehl1,
Daniel
Isabey1,
Gilbert
Desmarais2,
Laurent
Brochard1,
Alain
Harf1, and
Frédéric
Lofaso1
1 Service de
Physiologie-Explorations Fonctionnelles, Institut National de la
Santé et de la Recherche Médicale Unité 492,
Hôpital Henri Mondor, 94010 Créteil; and
2 École
Supérieure d'Ingénieurs en Électrotechnique et
Électronique, 93160 Noisy le Grand, France
 |
ABSTRACT |
Pressure support
(PS) is characterized by a pressure plateau, which is usually generated
at the ventilator level
(PSvent). We have built a PS
device in which the pressure plateau can be obtained at the upper
airway level (PSaw) or at the
alveolar level (PSA). The effect
of these different PS modes was evaluated in seven healthy men during
air breathing and 5% CO2
breathing. Minute ventilation during air breathing was higher with
PSA than with PSaw and lower with
PSvent (16 ± 3, 14 ± 3, and 11 ± 2 l/min, respectively). By contrast, there were
no significant differences in minute ventilation during 5%
CO2 breathing (25 ± 5, 27 ± 7, and 23 ± 5 l/min, respectively). The esophageal
pressure-time product per minute was lower with
PSA than with
PSaw and
PSvent during air breathing (29 ± 26, 44 ± 44, and 48 ± 30 cmH2O · s, respectively) and 5% CO2
breathing (97 ± 40, 145 ± 62, and 220 ± 41 cmH2O · s,
respectively). In conclusion, during PS, moving the inspiratory
pressure plateau from the ventilator to the alveolar level reduces
pressure output, particularly at high ventilation levels.
control of breathing; positive inspiratory pressure; unloading
 |
INTRODUCTION |
AMONG THE DIFFERENT FORMS of partial ventilatory
support, pressure support (PS) ventilation is the most commonly used,
both in the early phase of acute respiratory failure and during weaning from mechanical ventilation (7, 9, 23, 27). During PS, each spontaneous
breath is assisted by a constant positive pressure applied by the
ventilator all along each inspiration. In the early study by Kacmarek
(21) of intensive care PS devices, some of the devices failed to reach
the prescribed pressure before the end of inspiration or failed to
maintain the PS level near the inspiratory peak flow
(
peak), i.e.,
at onset of inspiration. To minimize pressure instability during
inspiration, some manufacturers have developed devices in which
pressure is servo controlled to obtain a square pressure wave. Usually,
the pressure-measuring site used to servo control the PS device is
inside the ventilator. Because the inspiratory line of the breathing
circuit and the respiratory system generate substantial resistance to
flow, a pressure gradient exists between the ventilator, the airway
opening, and the alveoli (19). Therefore, when a square pressure signal is generated inside the ventilator, the pressure signal in the airways
is far from being constant throughout inspiration and is even highly
flow dependent at the alveolar level. We hypothesized that a PS device
capable of providing a square pressure signal at the airway opening
would reduce the work of breathing compared with a conventional PS
device producing a square pressure signal in the ventilator and that
this effect would be further increased if the PS device maintains the
square pressure signal at the level of the alveoli.
To test these hypotheses, we have built a ventilator in which the
location of the inspiratory positive-pressure plateau can be moved from
the ventilator to the alveoli. The site where the pressure is
controlled and regulated as a plateau can be inside the ventilator, at
the airway opening, or at the alveolar level. One of these three sites
is selected after estimation of the overall airway pressure (Paw) drop.
The effect of PS generated at the three different levels of the
ventilator-patient system (PSvent, PSaw, and
PSA) was physiologically
evaluated by measuring ventilatory parameters and the esophageal
pressure (Pes)-time product in normal subjects under normal conditions
(air breathing) and during breathing of a gas mixture enriched with 5%
CO2 to enhance ventilatory demand.
 |
METHODS |
Apparatus Tested
The PS device was similar (Fig. 1) to that
used in several previous studies (2, 9, 18, 23, 24), except the
generated pressure was servo-controlled. Total resistance of the
inspiratory line (1 m long) was ~2.5
cmH2O · l
1 · s.
To generate a positive inspiratory pressure, a jet of pressurized gas
(air or CO2-enriched gas mixture)
was blown into a chamber that was open to the atmosphere or connected
to a highly compliant balloon (100 liters) filled with the appropriate
CO2 mixture. This fluid system
created air entrainment and pressurization of the overall gas flow and
constituted a low-impedance positive-pressure generator that was not
limited within the physiological range of inspiratory flows
(
) used by the study subjects (19). The balloon
ensured that the compositions of the injected gas and the entrained gas
were similar.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 1.
Basic setup consisting of a 1-m-long inspiratory line, a 1-way valve,
and an expiratory valve that separated expiration from inspiration. To
this basic setup, a device that created positive inspiratory pressure
was added. However, when pressure support (PS) device was not
activated, total resistance of inspiratory apparatus was ~2.5
cmH2O · l 1 · s
and entire setup could be used to control measurements. To obtain a
positive inspiratory pressure, a jet of air or
CO2-enriched gas mixture from a
pressurized cylinder was blown into an injector, where it entrained a
flow of gas from a balloon (100 liters). Pressure was regulated by an
electrovalve upstream from jet injector. This electrovalve opened when
inspiratory flow rose to >17 ml/s and remained open until inspiratory
flow fell to <150 ml/s. Level of electrovalve opening was
servo-controlled to keep pressure constant at different levels of
ventilator-patient circuit. Site of pressure stability was located
inside ventilator (PSvent
regulation) when pressure was measured at this site, at airway opening
(PSaw regulation) when pressure
was measured near mouthpiece, or at level of alveoli
(PSA regulation) when a formula
was used to estimate alveolar pressure
(PA) from opening airway
pressure (Paw), airway resistance (Raw), and inspiratory flow
( ): PA = Paw ( · Raw).
|
|
To maintain the pressure plateau in the various modes of PS tested, a
servo valve was placed upstream of the jet injector, instead of the
simple on-off electrovalve used in the previous system. The servo valve
was programmed to open when
rose above 17 ml/s (1 l/min) and to remain open until
fell
below 150 ml/s, as in conventional inspiratory PS systems. Contrary to
these conventional systems, however, servo valve opening was commanded
by a computer that compared, at intervals of 2 ms, the measured
pressure [corrected or not corrected for airway resistance
(Raw)] with the desired plateau pressure. The difference between
these two pressures was minimized by using a second-order control loop
that was modified to take into account the constant delay due to dry
friction in the servo valve and to the self-inductance of the magnetic
coil. The loop regulation of the pressure is described in the
APPENDIX (11, 15). In practice, after
a
cycle, the servo valve opening was progressively
increased to compensate for 1) the
pressure drop through the pressure generator in the
PSvent mode,
2) the pressure drop through the PS
device inspiratory line in the
PSaw mode, and
3) a minimal estimated airway
pressure drop in the PSA mode.
With our computer-controlled PS device, the pressure plateau was
generated at the site where the reference pressure was measured (or
estimated), i.e., 1) inside the
ventilator (PSvent),
2) at the airway opening
(PSaw) when pressure was
measured near the mouthpiece, or 3)
at the alveolar level when alveolar pressure (PA) was estimated
(PSA regulation).
PA was estimated on the basis of
the Paw, Raw, and
according to the following
formula
|
|
Raw
was taken to be constant and equal to 80% of the resistance of the
respiratory system (Rrs). Rrs was measured in each subject with use of
the forced oscillation method (10).
Before using the above-described modified PS device in our study
subjects, we tested it in vitro using a lung model simulating patient
inspiratory effort (Fig. 2,
top). The lung model was a two-chamber Michigan test lung. One chamber was connected to and powered by a motor ventilator with a flow-controlled mode in such way
that the effort generated was not influenced by the connected system
(Cesar, Taema, Antony, France) (driving chamber), whereas the other
chamber (PS-pressurized chamber) was connected to the PS ventilator
under test. The two chambers were physically connected to each other by
a small metal piece that allowed the driving chamber to lift the
PS-pressurized chamber, thus mimicking the patient's contribution to
inspiration. With this system, the generation of positive pressure in
the driving chamber lowered the pressure in the PS-pressurized chamber
to subatmospheric levels, just as inspiratory muscle contraction
produces negative PA in vivo.
This effect was detected by the triggering system of the PS ventilators under test. Because the metal component was not secured to the PS-pressurized chamber, the latter, once effectively pressurized, could
rise above the driving chamber within a time interval dependent on the
mechanical properties (resistance and compliance) of the PS-pressurized
chamber. The compliance of this chamber was set at 80 ml/cmH2O. Positive end-expiratory
pressure (PEEP) was applied to the driving chamber at a level that
ensured synchronized motions between the two chambers at onset of
inspiration. A resistance of 4 cmH2O at 1 l/s, connected between
the device and the lung model, was used to mimic the patient's Rrs.
The motor ventilator was set to obtain a predetermined respiratory
frequency of 15 cycles/min, an inspiratory time
(TI) of 1.2 s, and a
decelerating flow with a
peak of 1.1 l/s.

View larger version (39K):
[in this window]
[in a new window]
|
Fig. 2.
Top: experimental setup. Circuits of
tested PS devices were connected to a 2-chamber Michigan test lung. One
chamber (driving chamber) of test lung was attached to and powered by a
ventilator; the other chamber (PS-pressurized chamber) was connected to
PS device under study. Driving chamber was able to lift PS-pressurized
chamber and induced an initial negative pressure until PS device
contributed to expansion of freely moving PS-pressurized chamber.
Pressure and flow were measured at end of PS device circuit. C,
compliance. Bottom: pressures obtained
inside ventilator (Pvent), at
mouth (Paw), and inside
pressurized chamber (PA)
during simulated spontaneous breathing with PS device disconnected (SB)
and during each condition of PS regulation:
PSvent regulation,
PSaw regulation, and
PSA regulation.
|
|
A Fleisch no. 2 pneumotachograph (Gould Electronic, Longjumeau, France)
was inserted between the lung model and the circuit of the tested
device. Pressure was measured inside the ventilator (Pvent), at the airway opening
(Paw), or inside the pressurized chamber
(PA) with a differential
pressure transducer (model MP45, ±70
cmH2O, Validyne, Northridge, CA)
used at each pressure site. Figure 2,
bottom, shows the
Pvent, Paw, and
PA signals for a PS set at 7 cmH2O during simulated spontaneous
breathing (SB) with the PS device disconnected and with each of the
three PS regulation conditions tested; the data clearly demonstrate the
efficiency of the pressure-regulated device used in our study.
Clinical Study
Experiments were performed in seven healthy men (age = 35 ± 5 yr,
weight = 73 ± 8 kg, height = 173 ± 3 cm).
Measurements.
The subjects were seated, wore a noseclip, and breathed via a
mouthpiece. Flow was measured using a pneumotachometer (Fleisch no. 2)
connected to a pressure transducer (model MP45, ±2
cmH2O, Validyne) and integrated to
yield tidal volume (VT). Paw
(model MP45, ±50 cmH2O,
Validyne) and PCO2 in respiratory air [end-tidal PCO2
(PETCO2); infrared analyzer, Gould, Ballainvilliers, France] were measured in the breathing tube close to the lips. Pes and gastric pressure (Pga) were recorded using a catheter-mounted transducer (Gaeltec, Dunvegan, Isle of Skye,
UK). The validity of the Pes measurements was checked by analyzing the
shape of the Pes curve after swallowing and by using the occlusion
technique (5).
All signals were digitized at 128 Hz and sampled for subsequent
analysis using an analog-numeric system (MP100, Biopac System, Goleta, CA).
The Pes-time products per breath, per minute, and per liter of
ventilation were calculated as previously described (23, 29). Briefly,
it was measured as the area enclosed within Pes and the chest wall
static recoil pressure (Pcw,st)-time curve over
TI, with inspiratory PEEP
(PEEPI) taken into account.
The Pcw,st-time curve was extrapolated to the predicted value of chest wall compliance (Ccw,st). Thus the slope of the Pcw,st-time
relationship was
(
VT/Ccw,st)/
TI.
Inspiratory work of breathing (WOB) per breath, per liter, and per
minute was computed from Pes-volume loops, as previously described (8).
Briefly, WOB was calculated from a Campbell diagram by computing the
area between the recorded Pes-volume curve during inspiration and the
static Pes-volume curve of the chest wall. The values for Pes at zero
flow instants were taken as the beginning and the end of inspiration.
The theoretical value of chest wall compliance, which theoretically
represents ~4% of the predicted value of the vital capacity per
cmH2O, was used to trace the
static Pes-VT curve of the chest
wall (8). This curve passed through the value for elastic recoil
pressure of the chest wall at end expiration, which was assessed by
measuring intrinsic PEEP on the Pes tracing. The beginning of
inspiration was thus separated from the elastic recoil pressure by an
amount equal to the intrinsic PEEP on the Campbell diagram.
Experimental protocol.
Two sessions were performed in each subject with two successive
inspired CO2 fractions: 0%
CO2 and 5%
CO2 to stress the respiratory system. Within each session, four 10-min measuring periods were performed, each with a different ventilatory mode: SB,
PSvent regulation,
PSaw regulation, and
PSA regulation. The order of the
different ventilatory modalities was different from one subject to
another and was determined using a randomization table. For each PS
ventilation mode, the level of plateau pressure was set at 7 cmH2O.
Data analysis.
The variables were recorded, after stabilization, from the 7th to the
10th min of each period. The following variables were read breath by
breath: TI as the onset of
to the onset of the expiratory flow,
TE as the remainder of the total
breath duration, VT from the
calibrated integrated flow signal,
PETCO2 as the peak of the
airway CO2 record, Pes,
Pga, Paw, and
peak.
These variables were used to calculate the following values breath by
breath: respiratory rate (RR = 1/TI
TE), total ventilation (
E = VT × RR), inspiratory
fraction
[TI/(TI
TE)], mean
inspiratory flow rate
(VT/TI),
Pes-time product per breath, Pes-time product per minute, Pes-time
product per liter, WOB per breath, WOB per minute, WOB per liter, and
dynamic intrinsic PEEP
(PEEPdyn,i) calculated from the
beginning of the negative deflection in the Pes tracing near the end of
expiration to the first point corresponding to zero or positive airflow.
Individual mean values were calculated for each variable in a given
session by averaging the breath-by-breath variables during the last 3 min of each recording period.
Statistical analysis.
Values are means ± SD. Statistical differences between the three
assisted ventilatory modes within each session were tested using the
nonparametric Friedman test because of the small number of subjects.
The 5% level was chosen as significant. When a significant difference
was observed, bilateral comparisons were performed using the Wilcoxon
signed-rank test. The Wilcoxon test was also used to assess bilateral
comparisons between air and 5%
CO2 breathing.
 |
RESULTS |
Effects of PS Mode on Breathing Pattern
Mean values are listed in Table 1. During
air breathing, RR was lower during SB than during ventilation with any
of the PS modes. No significant differences were observed among the PS
modes. During 5% CO2 breathing,
RR was lower during SB than during ventilation with any of the PS
modes, and RR was lower during
PSvent than during
PSaw and
PSA.
During air breathing, VT was
largest during PSA, followed by
PSaw,
PSvent, and SB. The differences
between each of the three PS modes and SB were significant. No
significant differences were found during 5%
CO2 breathing.
During air breathing,
E was largest
during PSA, followed by
PSaw,
PSvent, and SB. No significant
differences were observed during 5%
CO2 breathing. Similarly, we found
a trend to a decrease in
PETCO2 from
PSvent to
PSaw and
PSA
(P = 0.07) during air breathing,
whereas PETCO2 was identical
during 5% CO2 breathing with all
three PS modes. There was a tendency to a reduction in
TI during air breathing, with no
significant variation in
TI/TT.
During 5% CO2 breathing,
TI and
TI/TT
decreased significantly from
PSvent to
PSaw and
PSA.
Effects of PS Mode on Pes
Mean values of Pes-time product per breath, Pes-time product per liter,
Pes-time product per minute, WOB per breath, WOB per liter, WOB per
minute, and PEEPdyn,i are
displayed in Table 2. During air and 5%
CO2 breathing, Pes-time product
per breath, Pes-time product per minute, and Pes-time product per liter
were significantly higher during
PSvent than during
PSaw or
PSA. Similar results were observed
with the WOB indexes; however, the significances were less
systematically observed, except during 5%
CO2 breathing between
Psvent and
PSA.
The mean level of PEEPdyn,i was
<1 cmH2O with all ventilator
modes during air and 5% CO2
breathing. There were no significant differences among the four
ventilatory modes. Also, the shape of the Pga curves did not suggest
any significant abdominal expiratory activity (22).
Modifications in Paw,
VT/TI,
and
peak
Mean inspiratory Paw,
VT/TI,
and
peak are
shown in Table 3. Figure
3 shows recordings obtained with all
ventilatory modes tested and provides further evidence of the
effectiveness of the regulation device. As expected, mean inspiratory
Paw,
VT/TI,
and
peak
differed significantly among modes during air and 5%
CO2 breathing, with values higher
during PSaw and
PSA than during PSvent.

View larger version (75K):
[in this window]
[in a new window]
|
Fig. 3.
Tracing obtained in a representative patient during air breathing and
5% CO2 breathing and during 4 ventilatory modes: SB, PSvent
regulation, PSaw regulation, and
PSA regulation. Pes, esophageal
pressure;
FICO2,
fraction of inspired CO2.
|
|
Responsiveness to CO2
With all modes, we found significant differences in
E between air
breathing and 5% CO2 breathing
mainly because of a rise in VT.
A significant rise in RR, with a decrease in
TI, was observed only for
PSaw. However, this rise did not
exceed 23%, whereas the change in
VT was 53%. Mean inspiratory
Paw was lower during 5%
CO2 breathing than during air
breathing with SB and PSvent. The
opposite occurred for PSaw and
PSA, suggesting a more favorable partition between Paw applied by the ventilator and Pes-time product. This is illustrated in Fig. 4, which shows
the relationship between the patient's inspiratory effort and the
assistance provided by the machine.

View larger version (18K):
[in this window]
[in a new window]
|
Fig. 4.
Relationship between Pes-time product per breath (PTP/b), which
represents inspiratory effort, and Paw-time product during inspiration
per breath (PawTP/b), which represents assistance by machine during air
breathing ( ) and 5% CO2
breathing ( ).
|
|
 |
DISCUSSION |
PS ventilation is a pressure-targeted mode in which pressure is
delivered in a square-wave pattern that begins with the patient's inspiratory effort and terminates when a threshold of decreasing
is reached. However, a discrepancy between the
expected inspiratory plateau of pressure and the observed pressure
waveform is frequently observed in clinical practice (21). This
discrepancy is related to high
, high resistive
properties of the open-valve demand system and of the tubing between
the pressure generator and the patient, and/or differences in pressure
rise profiles. In extreme situations the result can be a lack of
positive-pressure assistance at the beginning of inspiration, which is
the time when resistive WOB is greatest.
Our laboratory has developed a regulation system that takes into
account Paw and Raw. We used this system to evaluate the physiological
effects of moving the pressure plateau from the generator to the mouth
and presumably to the alveolar level. The regulation system was
incorporated into a ventilator designed for noninvasive PS ventilation
in intensive care and currently used in clinical practice with good
results (9, 23). The efficacy of our regulation system was first
checked using a lung model, with satisfactory results.
Once a given inspiration is initiated, the PS ventilation device
delivers a high
, which decreases gradually
throughout the inspiration. Clearly, our servo-regulated system allows
adjustment of the initial flow to the value needed to reach the
appropriate preset pressure level at different sites in the airway
between the ventilator and the alveoli. As expected, we found that when the system was set to produce a pressure plateau in the airways rather
than in the ventilator, the initial ventilator-delivered pressure and
flow increased to overcome the resistance of the ventilator circuit.
When the system was set to produce a pressure plateau in the alveoli,
further increases in ventilator-delivered flow and pressure were
observed; this effect would theoretically improve the rate of alveolar
pressurization without increasing peak
PA.
WOB includes a predominant elastic component and a flow-resistive
component. Breathing through circuit tubing, connectors, a humidifier,
internal pneumatic circuitry, and ventilator valves necessarily
increases the resistive component. This may be crucial at the beginning
of inspiration, at the time where the prescribed inspiratory plateau
pressure has not yet been reached and the inspiratory flow demand is
the highest. Therefore, the pressure delivered by a conventional PS
ventilator (PSvent) at the
beginning of inspiration may undercompensate the resistive loading
imposed by the circuitry and airway even when the level of prescribed PS is high. This may result in a zero or negative
PA at the beginning of
inspiration, and therefore the elastic and resistive load imposed by
the respiratory system on the inspiratory muscles may not be reduced by
PS during a crucial part of inspiration. This phenomenon can
theoretically be minimized by prescribing higher levels of PS. However,
these high levels can cause respiratory system overdistension and
barotrauma as a result of high pressures being delivered at the end of
inspiration (17). Also, high levels of PS can produce desynchronization
between the patient's SB and the ventilator (12). Alternatively,
regulating the pressure at the alveolar level offers two advantages:
1) compensation for the resistive load due to the circuitry and airway at any level of ventilatory demand, resulting in positive PA
from the very beginning of inspiration, and
2) minimization of the risk of
barotrauma or patient-ventilator desynchronization, since in theory
there is no increase in peak PA.
Before discussing the possible advantages of this approach, we will
address several potential limitations or negative consequences.
The proposed approach, in which we apply a constant pressure at the
alveolar level, produces an abnormal flow pattern with an elevated
early peak flow. This very rapid flow acceleration is not necessarily
well tolerated at a subjective level, and the impact of this very high
sheer force at the beginning of inspiration may be significant when the
Rrs and/or the level of plateau pressure is high. Moreover, if the
total Raw is completely compensated during
PSA, with the assumption that gas
inertance is negligible, the time constant of the respiratory system
would decrease to zero. In this case, unless the subject is making a
substantial crescendo effort to maintain flow, flow will essentially
consist of a very brief spike, and the cycle will be immediately
terminated. This pattern, which should clearly not be an acceptable
outcome, was not observed in our study. In fact, extreme shortening of TI did not occur in our study,
possibly because we have compensated for only 80% of the Rrs during
PSA. Furthermore, the forced
oscillation resistance is estimated in the linear range and tends to
underestimate the actual resistance that is flow dependent during breathing.
Another limitation is the increase in the peak pressure at the upper
airway level, induced by regulating the pressure at the alveolar level,
which may facilitate a possible gastric inflation. To prevent this
gastric inflation during noninvasive ventilation, we voluntarily
limited the maximal Paw induced by our apparatus to 25 cmH2O. Thus our ventilator should
fail to maintain a Paw or PA if
the Rrs and/or the setting of the level of plateau pressure is too
high. In addition, we do not know the effect of
PSA in patients with airflow
obstruction in which pulmonary heterogeneities exist, but it is
probable that regional PA
differed with differences in regional time constants.
Therefore, the findings from this study cannot be generalized, and the
results may have been different at a higher level of the Rrs, at a
higher setting of the plateau pressure, and/or when pulmonary
heterogeneities predominate.
Previous studies have demonstrated that the site of pressure regulation
during continuous positive airway pressure (CPAP) influences the
imposed WOB (1, 3, 4, 13, 28). To eliminate the WOB imposed by the
breathing circuit and ventilator, many CPAP devices have been designed
with the goal of controlling the pressure level at the Y piece (1).
More recently, attempts have been made to overcome the work imposed by
the endotracheal tube, which is an additional resistive component of
the breathing apparatus, by relocating the site of pressure regulation
to the tracheal end of the endotracheal tube (3, 4, 28) or by estimating the tracheal pressure from the Paw and the endotracheal tube
resistance (13). These studies consistently found that for a given CPAP
level the WOB decreased when the pressure-controlling site was
physically moved closer to the ventilatory muscles. In keeping with
these studies, we observed that during PS ventilation inspiratory
activity decreased when the site of the preset pressure was moved from
the ventilator to the alveoli.
Varying the initial flow rate may interact with the breathing pattern
and/or the patient's effort. The respiratory control system has
options: 1) it can maintain the
central respiratory output, utilizing the additional unloading to
produce more ventilation, or 2) it
can maintain ventilation at the baseline level and use the assist to
reduce respiratory muscle work. Any combination of these two extremes
is also possible, and the result may vary with the ventilatory demand.
During air breathing in the resting condition, PS ventilation is known
to induce hyperventilation and hypocapnia as a result of an increase in
VT and also to decrease the
inspiratory drive (2, 18, 24, 30). Our results confirmed this
phenomenon and demonstrated that it was accentuated when the initial
flow rate was further increased by moving the site of regulation of the
appropriate preset pressure from the ventilator to the alveolar level.
The literature suggests that in acute conditions the additional
unloading is used primarily to reduce inspiratory activity (6, 23). In
a study of intensive care unit patients, Bonmarchand et al. (6) found
that an initial flow rate increase during PS was associated with a
decrease in diaphragmatic activity, although arterial
PCO2 and minute ventilation remained
unchanged over a broad range of flow rate variation. Similarly, we
recently compared several PS devices in critically ill patients
undergoing weaning from mechanical ventilation. We did not observe any
differences in arterial PCO2 and
minute ventilation, but we found that the inspiratory PS device that
provided the highest initial flow rate was more efficient in reducing
inspiratory activity than the inspiratory PS device that provided the
lowest initial flow rate (23). These findings suggest that intubated
patients with respiratory failure used the additional unloading
afforded by the increase in the initial flow rate to reduce respiratory muscle work rather than to produce more ventilation.
Noncontradictory results were observed in normal subjects when
inspiratory activity was constrained by
CO2 inhalation. During PS
ventilation and CO2 inhalation,
subjects have full control over the onset of inspiration.
TI and
VT were dependent on the level
of PS and the activity of inspiratory muscles.
TE was determined to be the time
when the subject triggered the next inspiration. Therefore, a
respiratory response to inhaled
CO2 is possible during conventional PS, and we previously found that PS shifted to the left
the relationship between
E and
PETCO2 and reduced inspiratory activity at any given
PETCO2 (14, 18, 30).
Therefore, these studies suggested that conventional PS induced a
nonchemical inhibition of inspiratory activity. Our study confirms the
findings of these previous studies and suggests that the nonchemical
inhibition is influenced by the shape of the initial flow rate induced
by the ventilator, since we observed that moving the site of regulation
of the appropriate preset pressure from the ventilator to the alveoli
was associated with a significant decrease in the inspiratory effort
indexes with no statistically significant differences for
E and
PETCO2. It is noteworthy that the reduction of inspiratory activity is not only due to the
reduction in duration of inspiratory effort, since significant reductions were also observed with WOB indexes, which do not depend on
TI.
Previous studies have shown correlations between the Pes-time product
and measurements of O2 utilization
and CO2 production by the
contracting respiratory muscles (8, 29). This led us to expect a
decrease in CO2 production with a
reduction in arterial PCO2 at a given
level of alveolar ventilation during
PSaw and
PSA compared with
PSvent. However, although
E was
similar during CO2 and air
breathing, PETCO2 values were virtually identical under all the PS ventilation conditions. This is probably ascribable to the fact that a number of phenomena have
opposite effects on arterial PCO2.
First, although a diminution in the work of breathing can lead to a
diminution in CO2 production, the
magnitude of this effect is probably limited. Second, the increase in
peak inspiratory pressure in the upper airways that occurs when the
site of pressure regulation is moved from the ventilator toward the
alveoli (Fig. 3) may increase the distension of the upper airways,
leading to an increase in the anatomic dead space. Third, variations in
the physiological dead space related to variations in
TI/TT
have been reported during controlled mechanical ventilation in
different circumstances, although their mechanism remains unclear (16,
25, 26). Rebreathing can probably be excluded on the basis of the
continuous analysis of the
PETCO2 records (Fig. 3).
Finally, some variation in
E was observed
among the three PS modes; although the differences were not
statistically significant in such a small group of subjects, we cannot
rule out variations in alveolar ventilation.
Despite similarities between CO2
stimulation in our study and that observed in most patients in clinical
practice in tonus of elevated inspiratory drive, major differences
exist because of differences in respiratory resistance and compliance.
For example, Jubran et al. (20) observed that although PS ventilation
reduced respiratory frequency in patients with chronic obstructive
pulmonary disease, several of these patients displayed expiratory
activity. The patients with expiratory activity were those with long
TI, which facilitated expiratory
muscle activity during lung inflation or shortened the time available
for lung emptying. Because our servo-regulated system allows
VT/TI
to increase by moving the site of preset pressure regulation from the
ventilator toward the alveoli, it also reduces
TI (as observed in our study)
and may therefore reduce the occurrence and/or the importance of
expiratory activity in these patients. Unfortunately, we were unable to
demonstrate this potential beneficial effect, because we did not
logically observe in our normal subjects any expiratory activity as
detected from the shape of Pga recordings in all studied conditions.
In conclusion, we have built a PS device allowing inspiratory pressure
regulation to be moved from the ventilator toward the alveolar level.
The efficacy of this system has been confirmed during experiments first
in a lung model, then in humans at rest and during stimulation with a
5% CO2-enriched mixture. We
observed a noticeable decrease in inspiratory activity when the site of the preset pressure was moved from the ventilator to the alveoli. This
beneficial effect was particularly marked when ventilation was
stimulated by CO2. These data
established in normal humans warrant a clinical evaluation.
 |
APPENDIX |
Loop regulation of the pressure assistance device.
The outline of the loop regulation used to control the pressure is
presented in Fig.
5A.
Preliminary experiments in the open-loop configuration have shown that
the delay between the command signal (Cmd) of the servo valve (PSV) and
the Paw had the same magnitude as the time constant (
) of the
overall system. This allowed us to describe the system with an equation
that approximates its response on the basis of the combination of a
constant delay (
) and a first-order mode of time constant
T (Broïda model)
where
p = j
(j2 =
1;
= pulsation) and Gs is direct-current gain between Paw and the command
signal of the servo valve.

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 5.
A: block diagram of pressure support
ventilation (PSV) device in which delivered pressure is computer
controlled. , mouth flow; Paw, airway pressure, which
is compared with reference (desired) pressure each 2 ms; Cmd, command
signal of servo valve. Gas source is at high pressure
( 0).
B: loop regulation circuit. In
circuit a
(top), structure of
C1 controller was actually
modified to simulate structure of a controller
(circuit b, bottom) that isolates
time delay ( ) externally to loop feedback (PI).
C: final structure of
C1 controller, which allows
effects of time delay ( ) and time constant
(T) to be separated by means of a
corrector called "Smith predictor." See APPENDIX for
definitions of other abbreviations.
|
|
Both
and T have a magnitude of 15 ms (sampling period: TE = 2 ms).
The desired performances of the system consisted of a global time
response <100 ms with an accuracy of the generated pressure within
±0.5 cmH2O. The ratio of
to T being close to unity, we could
not use a simple proportional-integral (PI) corrector, because the gain
of the closed-loop regulation could not be high enough to guarantee the
imposed global time response (circuit a in Fig. 5B).
Definition of the feedback controller.
We used a classic strategy to increase the gain of the loop regulation,
which consists of looking for a structure of the controller (C1 in Fig. 5B,
top) that artificially isolates the
externally to
the loop feedback (PI), as in circuit
b in Fig. 5B. Imposing identity of transfer functions between circuits
a and b, we obtained the following relationship between
C1 and
C2
The final structure of the C1
corrector is given in Fig. 5C. This
corrector is classically known as a "Smith predictor" (11, 14).
Such a structure greatly simplifies the numerical procedure, since
differential equations describing the transfer function G(p)
are expressed in terms of difference equations. Thus, taking the
transfer function that simulates the first-order part of the model
where
Sm is the value model output without delay. The corresponding
time-domain equations (t) can be expressed as
and
the difference equation becomes
where
k is sampling number and TE is sampling period. The
recurrent function of the model is
In
Fig. 5C the simulation of
is
performed by selecting from a list of the last values (Sm) of model
output. Moreover, the corrector C2
uniquely considers a transfer function
G(p),
but without time delay. Therefore, the model reduces to a simple PI corrector and takes into account the pure delay with sufficient stability and fast response. The classic PI equation is given by
where
KP is the direct-current gain of the PI controller. The time domain
equation is
Using
a simple algorithm to integrate
The
recurrent expression of the PI corrector becomes
 |
FOOTNOTES |
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests: F. Lofaso, Service de
Physiologie-Explorations Fonctionnelles, Hôpital Raymond
Poincaré, 92380 Garches, France (E-mail:
f.lofaso{at}rpc.ap-hop-paris.fr).
Received 9 July 1998; accepted in final form 18 February 1999.
 |
REFERENCES |
1.
Akashi, M.,
K. Sakanaka,
H. Noguchi,
and
Y. Takumi.
Flow-regulated continuous positive airway pressure to minimize imposed work of breathing.
Crit. Care Med.
18:
999-1002,
1990[Medline].
2.
Azarian, R.,
F. Lofaso,
F. Zerah,
H. Lorino,
G. Atlan,
D. Isabey,
and
A. Harf.
Assessment of respiratory compliance in awake subjects using pressure support.
Eur. Respir. J.
6:
552-558,
1993[Abstract].
3.
Banner, M.,
P. Blanch,
and
R. Kirby.
Imposed work of breathing and methods of triggering a demand-flow, continuous positive airway pressure system.
Crit. Care Med.
21:
183-190,
1993[Medline].
4.
Banner, M.,
R. Kirby,
C. Usafr,
and
P. Blanch.
Site of pressure measurement during spontaneous breathing with continuous positive airway pressure: effect of calculating imposed work of breathing.
Crit. Care Med.
20:
528-533,
1992[Medline].
5.
Baydur, A.,
P. K. Behrakis,
W. A. Zin,
M. Jaeger,
and
J. Milic-Emili.
A simple method for assessing the validity of the esophageal balloon technique.
Am. Rev. Respir. Dis.
126:
788-791,
1982[Medline].
6.
Bonmarchand, G.,
V. Chevron,
C. Chopin,
D. Jusserand,
C. Giraud,
F. Moritz,
J. Leroy,
and
P. Pasquis.
Increased initial flow rate reduces inspiratory work of breathing during pressure support ventilation in patients with exacerbation of chronic obstructive pulmonary disease.
Intensive Care Med.
22:
1147-1154,
1996[Medline].
7.
Brochard, L.
Non-invasive ventilation: practical issues.
Intensive Care Med.
19:
431-432,
1993[Medline].
8.
Brochard, L.,
A. Harf,
H. Lorino,
and
F. Lemaire.
Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation.
Am. Rev. Respir. Dis.
139:
513-521,
1989[Medline].
9.
Brochard, L.,
J. Mancebo,
M. Wysocki,
F. Lofaso,
G. Conti,
A. Rauss,
G. Simonneau,
S. Benito,
A. Gasparetto,
F. Lemaire,
D. Isabey,
and
A. Harf.
Noninvasive ventilation for acute exacerbation of chronic obstructive pulmonary disease.
N. Engl. J. Med.
333:
817-822,
1995[Abstract/Free Full Text].
10.
Brochard, L.,
G. Pelle,
J. de Palmas,
P. Brochard,
A. Carré,
H. Lorino,
and
A. Harf.
Density of frequency dependence of resistance in early airway obstruction.
Am. Rev. Respir. Dis.
135:
579-584,
1987[Medline].
11.
Dindeleux, D.
Régulation des systèmes stables à retard pur important. Limite de la régulation PID.
In: Technique de régulation industrielle D. Dindeleux. Paris: Eyrolles, 1981, p. 129-132.
12.
Fabry, B.,
J. Guttmann,
L. Eberhard,
T. Bauer,
C. Haberthür,
and
G. Wolff.
An analysis of desynchronization between the spontaneously breathing patient and ventilator during inspiratory pressure support.
Chest
107:
1387-1394,
1995[Abstract/Free Full Text].
13.
Fabry, B.,
C. Haberthür,
J. Guttmann,
R. Kuhlen,
and
R. Stocker.
Breathing pattern and additional work of breathing in patients with different ventilatory demands during inspiratory pressure support and automatic tube compensation.
Intensive Care Med.
23:
547-552,
1997.
14.
Fauroux, B.,
D. Isabey,
G. Desmarais,
L. Brochard,
A. Harf,
and
F. Lofaso.
Nonchemical influence of inspiratory pressure support on inspiratory activity in humans.
J. Appl. Physiol.
85:
2169-2175,
1998[Abstract/Free Full Text].
15.
Franklin, G.,
J. Powell,
and
A. Emami-Naeini.
Feedback control of dynamic system.
In: Feedback Control of Dynamic System. New York: Addison-Wesley, 1991, p. 448-449.
16.
Fuleihan, S.,
R. Wilson,
and
H. Pontoppidan.
Effect of mechanical ventilation with end-inspiratory pause on blood-gas exchange.
Anesth. Analg.
55:
122-130,
1976[Abstract/Free Full Text].
17.
Haake, R.,
R. Schlichtig,
D. Ulstad,
and
R. Henschen.
Barotrauma. Pathophysiology, risk factors and prevention.
Chest
91:
608-613,
1987[Free Full Text].
18.
Heyer, L.,
H. Lorino,
C. Delclaux,
D. Isabey,
A. Harf,
and
F. Lofaso.
Carbon dioxide response during positive inspiratory pressure in COPD patients.
Respir. Physiol.
109:
29-37,
1997[Medline].
19.
Isabey, D.,
and
A. Harf.
Basic physical principles for ventilators and ventilatory modes.
In: Acute Respiratory Failure in Chronic Obstructive Pulmonary Disease, edited by J. Derenne,
W. Whitelaw,
and T. Similowski. New York: Dekker, 1996, p. 685-708. (Lung Biol. Health Dis. Ser.)
20.
Jubran, A.,
W. Van de Graaff,
and
M. Tobin.
Variability of patient-ventilator interaction with pressure support ventilation in patients with chronic obstructive pulmonary disease.
Am. J. Respir. Crit. Care Med.
152:
129-136,
1995[Abstract].
21.
Kacmarek, R.
The role of pressure support ventilation in reducing work of breathing.
Respir. Care
33:
99-120,
1988.
22.
Lessard, M.,
F. Lofaso,
and
L. Brochard.
Expiratory muscle activity increase intrinsic positive end-expiratory pressure independently of dynamic hyperinflation in mechanically ventilated patients.
Am. J. Respir. Crit. Care Med.
151:
562-569,
1995[Abstract].
23.
Lofaso, F.,
L. Brochard,
T. Hang,
H. Lorino,
A. Harf,
and
D. Isabey.
Home versus intensive-care pressure support devices, experimental and clinical comparison.
Am. J. Respir. Crit. Care Med.
153:
1591-1599,
1996[Abstract].
24.
Lofaso, F.,
D. Isabey,
A. Harf,
and
P. Scheid.
Respiratory response to positive and negative inspiratory pressure in humans.
Respir. Physiol.
89:
75-88,
1992[Medline].
25.
Manthous, C.,
and
G. Schmidt.
Inverse ratio ventilation in ARDS improved oxygenation without autoPEEP.
Chest
103:
953-954,
1993[Abstract/Free Full Text].
26.
Mercat, A.,
L. Graini,
J. Teboul,
F. Lenique,
and
C. Richard.
Cardiorespiratory effects of pressure controlled ventilation with and without inverse ratio in the adult respiratory distress syndrome.
Chest
104:
871-875,
1993[Abstract/Free Full Text].
27.
Pennock, B.,
L. Crawshaw,
and
P. Kaplan.
Noninvasive nasal mask ventilation for acute respiratory failure.
Chest
105:
441-444,
1994[Abstract/Free Full Text].
28.
Sassoon, C.,
and
S. Gruer.
Characteristics of the ventilator pressure and flow-trigger variables.
Intensive Care Med.
21:
159-168,
1995[Medline].
29.
Sassoon, C.,
R. Light,
R. Lodia,
G. Sieck,
and
C. Mahutte.
Pressure-time product during continuous positive airway pressure, pressure support ventilation, and T-piece during weaning from mechanical ventilation.
Am. Rev. Respir. Dis.
143:
469-475,
1991[Medline].
30.
Scheid, P.,
F. Lofaso,
D. Isabey,
and
A. Harf.
Respiratory response to inhaled CO2 during positive inspiratory pressure in humans.
J. Appl. Physiol.
77:
876-882,
1994[Abstract/Free Full Text].
J APPL PHYSIOL 87(1):428-437
8570-7587/99 $5.00
Copyright © 1999 the American Physiological Society