J Appl Physiol 86: 1075-1080, 1999;
8750-7587/99 $5.00
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mürtz, P.
Right arrow Articles by Urban, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mürtz, P.
Right arrow Articles by Urban, W.
Vol. 86, Issue 3, 1075-1080, March 1999

SPECIAL COMMUNICATION
LMR spectroscopy: a new sensitive method for on-line recording of nitric oxide in breath

Petra Mürtz1, Lars Menzel1, Wilhelm Bloch2, Alexander Hess3, Olaf Michel3, and Wolfgang Urban1

1 Institut für Angewandte Physik der Universität Bonn, D-53115 Bonn; and 2 Anatomisches Institut I and 3 Klinik und Poliklinik für Hals-Nasen-Ohrenheilkunde der Universität Köln, D-50931 Cologne, Germany


    ABSTRACT
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

Laser magnetic resonance spectroscopy (LMRS) is a sensitive and isotope-selective technique for determining low concentrations of gaseous free radicals with high time resolution. We used this technique to analyze the nitric oxide (NO) concentration profile while simultaneously measuring the flow and expired volume during several single breathing cycles. Eight healthy, nonallergic volunteers were investigated. An initial NO peak was found in all breathing cycles before the NO concentration dropped to a relatively stable plateau in the late phase of expiration. The nasal NO peak was significantly higher than the oral NO peak. The nasal NO plateau was always higher than the oral NO plateau. The height of the initial nasal and oral NO peak rose with increasing duration of breath hold, whereas the late expiratory NO plateau changed only little for either the nasal or the oral breathing cycles. Our findings demonstrate, in line with other reports using other techniques, that the nose is the primary source for NO within the airways.

laser magnetic resonance; nitric oxide concentration; exhaled air; nitric oxide synthase


    INTRODUCTION
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

SINCE NITRIC OXIDE (NO) has been detected in exhaled air (9), the physiological role of NO within the airways and its exact origin are still a matter of controversy, especially during normal breathing cycles. A variety of different techniques for the detection of NO in exhaled air has been used. The standard method for measuring NO concentration in breath so far has been the O3/NO3- chemiluminescence method. In several studies it was found that, in healthy subjects, the upper airways, especially the nose and the nasopharynx, are the main source for NO in human breath (3, 5, 8, 13-15, 22). However, under pathological conditions the site of NO generation may vary. Indeed, patients suffering from various lung diseases exhale more NO, which is released mainly by the epithelium of the lower airways (1).

NO is generated by NO synthases (NOS). All three known isoforms of this family of enzymes have been detected in the nasal mucosa (7). The inducible form (iNOS) is primarily expressed under inflammatory conditions and, on expression, can produce large quantities of NO (4). In healthy humans, iNOS is found in the epithelium of the paranasal sinuses and is also partly present in the epithelium of the nasal cavity (9, 16). The other, constitutive NOS isoforms, endothelial and neuronal NOS, are found preferentially in subsuperficial structures. In consideration of the short half-life of NO in tissues as well as its inhibition by steroids, it is most likely that iNOS is the main source for luminal NO generation (5).

The purpose of the present work was to establish laser magnetic resonance spectroscopy (LMRS) as a new, highly sensitive technique that is suitable for on-line detection of low NO levels in exhaled air. Because this method uses the magnetic properties of the NO molecule, it uniquely detects this particular molecule and is not affected at all by other absorptions possibly present in the wavelength region near 5 µm. The established method of measuring NO in breath with the O3/NO3-chemiluminescence technique has two main disadvantages: limited specificity (28) and lack of selectivity of isotopomeres. Thus signal contributions from molecules like ethylene or H2S may arise (see Ref. 5a). Moreover, quenching of chemiluminescence by water in the exhaled air cannot be excluded. LMRS has a fast time response, permitting continuous recording of changes in NO concentration during the whole normal breathing cycle. It has to be mentioned that on-line recording with a chemiluminescence analyzer is also possible (2). In the present work we show data of on-line NO measurements in eight healthy subjects by using LMRS. Furthermore, we prove that the nose is the main source of NO generation without using invasive techniques, e.g., intubation.


    MATERIALS AND METHODS
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

Experimental Design

Subjects. Eight young volunteers (4 male and 4 female students, nonsmokers), aged between 21 and 30 yr, were selected according to the following requirements: 1) the anamnesis showed no hint of an atopic predisposition; and 2) subjects were free of diseases, especially acute and chronic alterations of the upper and lower respiratory tract, such as rhinitis and bronchial asthma. Furthermore, subjects had no medication for at least 3 days before the experiments.

Breathing cycles. The experiments were performed at the Institute of Applied Physics, University of Bonn. The subjects performed defined breathing cycles through the nose and the mouth by using NO-free synthetic air (see Fig. 1). Cycles of mouth breathing were performed with subjects not wearing noseclips, because this led to slightly increased NO content during the whole exhalation.


View larger version (25K):
[in this window]
[in a new window]
 
Fig. 1.   Measurement of nitric oxide (NO) concentration and breath flow for nasal respiration (A) and oral respiration (B). Positive flow values, exhaled air; negative flow values, inhaled air. ppb, Parts/billion.

The duration of inhalation and exhalation was kept constant to 2.5 s, and the duration of breath hold was set at 0, 5, 10, and 15 s.

An additional measurement was performed with ambient air with a cycle equal to a breath hold of 5 s.

The flow was measured with a pneumotachometer (modified rhinomanometer, model A440, Allergopharma, Hamburg, Germany) with an analog-to-digital converter; the data were recorded digitally by means of a personal computer. Portions of the exhaled air were directed to LMRS via a lateral adapter. The subjects carried out five test cycles before the measurement, controlling their times for inhalation and exhalation and breath hold on a computer display. These test cycles served the dual purpose of practice and washout of the exogene NO because the NO-free air was used.

Technical Principles and Arrangement of the LMRS

The LMRS method is well established for spectroscopy of paramagnetic molecules (for principles and reviews, see Refs. 6 and 18). The advantage of this technique is that NO is detected selectively and at the same time with extreme sensitivity [sub-parts/billion (ppb) level]. The NO measurement is not interfered with by any other gas, and different NO isotopomeres can be distinguished. Moreover, a time resolution adequate for time-resolved measurements of breathing cycles can be achieved.

For detection of NO molecules, a CO laser (25, 26) serves as the light source. The well-known resonance between X2 Pi 3/2 R(1.5), v = 1left-arrow 0 rovibrational transition of the 14N16O molecule, and the CO laser transition P(13) v = 9right-arrow8 (1,884.349 cm-1, 400 mW laser power, single mode) at a magnetic flux density of 0.1490 Tesla was used. More details about the corresponding signal, Zeeman component, and line shape are given elsewhere (16). The extreme sensitivity is achieved by polarization detection, which is one of the most sensitive spectroscopic methods in the midinfrared wavelength region. This particular technique called "Faraday LMRS" uses the Faraday effect to produce the signals and was first developed in our laboratory (11, 27).

A block diagram of the LMRS setup is given in Fig. 2. The central part of the apparatus is a liquid-nitrogen-cooled CO flow laser with internal single-line selection. It is stabilized on the gain maximum of a particular laser line by means of a standard frequency-modulation technique (uncertainty: ±10 MHz). The laser output is deflected into the bore of the magnet (solenoid, flux density: 0.1490 ± 0.0030 T). The total length of the detection zone is 200 mm. This is the homogeneous part of the magnetic field; it is surrounded by a small coil for Zeeman modulation (at 8 kHz). A glass tube with glass fittings defines the NO-sample volume (active volume is some 10 cm3; working pressure inside the cell is 27 mbar). A polarizer in front of the magnet entry ensures that the incoming laser beam is well polarized. The sample volume is sealed by plane CaF2 windows. The polarization of the emerging beam is analyzed by a Rochon prism that is almost crossed with respect to the polarization of the incoming laser beam. With this setup, the Faraday rotation of a linearly polarized laser beam is detected. The infrared radiation is detected by a liquid nitrogen-cooled InSb detector and fed into a lock-in amplifier. The signal output is processed by a personal computer for averaging and quantitative signal-strength analysis. To avoid undefined reactions between NO and the wall material, such as adsorption or desorption, the gas-handling system was made of stainless steel, perfluoroalcoxy, and glass. Moreover, the experiments were done under flow conditions. The flow rate of the gas stream through the cell was ~800 ml/min. Under these conditions, no significant change in the NO signal because of surface effects was observed.


View larger version (41K):
[in this window]
[in a new window]
 
Fig. 2.   Schematic of experimental setup to measure NO in breath with laser magnetic resonance spectroscopy (LMRS). PZT, piezoelectric transducer; REF, reference; SIG, signal; OUT, output; Delta P, pressure difference; U, voltage.

The system was calibrated each day before and after a series of measurements by detecting the signal strength for different defined gas mixtures of NO calibration gas (1 part/million NO in N2, Messer-Griesheim) and NO-free synthetic air (Messer-Griesheim, NO <4 parts/trillion1) at constant total gas pressure. We used a calibration gas with NO at the parts per million level because these mixtures remain stable for months. Within the tested range (0-500 ppb NO), the relationship between signal strength and NO content is linear. Measurement errors of about ±3% are caused by instabilities of laser power. The drift rate of the system is less than some ppb during 1 h. However, the signal drift is corrected during each breathing cycle with the help of the signal level corresponding to the inhalation of the NO-free synthetic air (diffusion of NO from the mouth is negligible). We achieved a sensitivity of ~1 ppb at a lock-in time constant of 0.3 s. The corresponding response time, which is determined by both the lock-in time constant and by the gas flow conditions, was 0.8 s. Thus the contribution to the response time by both effects is on the same order. Faster lock-in time constants (average signal times) decrease the sensitivity. For example, it is possible to choose a lock-in time constant of 0.03 s (corresponding response time is 0.4 s) if a sensitivity of only 3 ppb is required. The gas dynamic time constant is principally determined by the connecting tubing and the gas flow rate. The gas exchange might be improved by using thinner and shorter tubing. Flow rates >800 ml/min did not improve the response time of the system.

Statistical Methods and Analysis of Data

To gain information about the site of prominent NO release and for statistical reasons, the NO concentration was referred to the physiological volumes of the compartments of the respiratory tract. Therefore, the expiratory volume was divided into 150-ml fractions (see Fig. 3). The NO concentration (in ppb) in each 150 ml was analyzed statistically. The data from the first (150 ml) and third (450 ml) fractions were transferred to a table (Table 1). Normally, inhalation and exhalation volume did not exceed 1,000 ml. Thus only the NO data of the first 750 ml were taken. To investigate the significance of the data, the paired Student's t-test was chosen.


View larger version (23K):
[in this window]
[in a new window]
 
Fig. 3.   NO concentration (top) dependence on expiratory volume (bottom) calculated from measured breath flow. Values for 150 and 450 ml were further analyzed (see Table 1).

                              
View this table:
[in this window]
[in a new window]
 
Table 1.   NO in exhaled air from humans


    RESULTS
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

In all breathing cycles a sharp initial NO peak was observed within the first 150 ml of expired volume. A relatively stable plateau of lower NO concentrations (by a factor of 1.2-2.6) was found within the 300- to 450-ml expiratory volume.

Nasal NO peak was significantly higher (by a factor of 3.3-7.6 depending on the type of breathing) than the oral NO peak (see Table 1). For example, for a 10-s breath hold, the nasal NO peak (126.6 ± 80.3 ppb) was 7.6-fold higher (P < 0.001) than the oral NO peak (16.6 ± 8.0 ppb). A strong correlation between the peak concentration and breath-holding time was observed. Five subjects showed saturation of the NO concentration after a 10-s breath hold.

Similar results were found for the plateau concentrations. The nasal NO plateau was always higher (by a factor of 2.9-4.3) than the oral NO plateau, but the height of the plateau concentration scarcely changed with increasing durations of breath hold.

An increase in NO concentration in the late phase of expiration (450- to 750-ml total expiratory volume) could be observed in some cases; this was always accompanied by a decrease in breathing flow.

All evaluated NO data for breathing with ambient air showed slightly, but insignificantly, increased levels compared with the same breathing type in synthetic air (see Table 1). A slight increase in NO is expected with ambient air, because the ambient air in our laboratory had a NO content of 2-7 ppb during the experiments.


    DISCUSSION
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

Our results demonstrate for the first time that LMRS is an adequate method for NO measurements in breath. This method fulfills all expected requirements of sensitivity, specificity, and high time resolution. Indeed, on-line recording of breathing flow and NO concentration could be done simultaneously. Thus LMRS is an interesting alternative to the commonly used chemiluminescence analyzer. For direct comparison, the properties of both analyzers are given in Table 2. All of the specifications fulfilled by an up-to-date, state-of-the-art chemiluminescence analyzer, especially sensitivity, response time, and accuracy, are also performed by LMRS. The response time of LMRS in its present setup is of the same magnitude2 as the Sievers NOA 280 NO analyzer (22a) but may be reduced by a factor of ~2 if a sensitivity of 3 ppb is required. However, there are two main advantages of LMRS over the chemiluminescence technique. First, LMRS is completely specific for NO and is not interfered with by any other gas. Chemiluminescence analyzers may have signal contributions from molecules like ethylene or H2S (see Ref. 5a). In addition, with this technique, quenching problems because of water in the exhaled air may arise. Second, LMRS can distinguish between different NO isotopomers (e.g., 14N16O and 15N16O), which is impossible with chemiluminescence analyzers. This property permits investigations of a special reaction path by using enriched pharmaceuticals.

                              
View this table:
[in this window]
[in a new window]
 
Table 2.   Specifications of the LMRS in comparison with a present state-of-the-art chemiluminescence analyzer like the Sievers NOA 280 

Taken together with the above considerations, our study shows that high NO concentrations were found within the first 150 ml of expiratory volume, which is considered to represent the main static volume of the nose, including paranasal sinuses, the nasopharynx, the trachea, and the noncontractile parts of the bronchial branch. Through the difference between nose and mouth breathing, these data give evidence that, under physiological conditions, the majority of NO is released in the nasal as well as nasopharyngeal cavity. This finding is in line with previous reports using the chemiluminescence analyzer (3, 5, 8, 13, 14, 22). An increase in breath-hold duration leads to higher NO concentrations throughout the whole breathing cycle, in agreement with Persson and co-workers (21).

Significant NO production has been postulated also for the distal airways, whereas the lowest NO release has been proposed for the alveoli (5, 15, 24). The studies yielding these results were performed by using invasive techniques, e.g., intubation, tracheostomy, closure of the nasopharynx, or bronchoscopy, to abolish addition of nasal air. In contrast, our experiments were performed without any of these invasive techniques and without subjects wearing noseclips, because this led to a slight increase in NO content during exhalation through the mouth. This indicates that there is a relevant mixture with nasal air in the nasopharynx under conditions of increased resistance in the nasal airways.

NO uptake in the lungs generated by conducting airways has recently been proposed (23). Although we cannot reach any conclusions in our present study with regard to such NO uptake by lung tissue, it is likely that NO diminishes the tone of smooth muscle cells of bronchioli and of blood vessels (12, 19, 20). This may be an important mechanism to counteract pulmonary hypertension. We are therefore planning for a future investigation in which breathing is modified with inspiration through the nose and expiration through the mouth. The comparison with data on mouth breathing may allow us to gain insight into the consumption of NO in the lower airway tract. Thus we may be able to evaluate changes in NO concentration under physiological and pathophysiological conditions in different parts of the airway tract during normal breathing cycles. Because of the capability to quantify the rare isotopomere 15N16O selectively (16) and to detect the very low biological 15N16O concentration of 0.3% of 14N16O in natural abundance, LMRS may be the method of choice to elucidate the kinetics of NO metabolism via application of 15N-labeled NO donators.


    ACKNOWLEDGEMENTS

We thank R. F. Curl for carefully reading the manuscript.


    FOOTNOTES

The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.

1 This was measured with a parts/trillion chemiluminescence analyzer (ECO Physics CLD 770 AL).

2 The electronic response time of the LMRS system is worse than that of chemiluminescence analyzers by a factor of 2-3. However, gas-flow conditions, which limit response time of the systems in practice, are better for LMRS because of better gas exchange due to a higher flow rate through the analyzing cell (800 ml/min at 27 mbar). Chemiluminescence analyzers work at atmospheric pressure and at a gas flow of ~100-300 ml/min. Under these conditions, electronic response time of the system is reduced by a factor of ~3 at least, if no external tubing is connected (22a). LMRS permits much higher flow rates of the gas sample through the cell because the system requires only a low working pressure (only a fraction of the exhaled air flows through the cell of the LMRS appartus, even at high flow rates).

Address for reprint requests: P. Mürtz, Institut für Angewandte Physik der Universität Bonn, Wegelerstrasse 8, D-53115 Bonn, Germany (E-mail: muertz{at}imsdd.meb.uni-bonn.de).

Received 30 April 1998; accepted in final form 10 November 1998.


    REFERENCES
TOP
ABSTRACT
INTRODUCTION
MATERIALS AND METHODS
RESULTS
DISCUSSION
REFERENCES

1.   Alving, K., E. Weitzberg, and J. M. Lundberg. Increased amount of nitric oxide in exhaled air of asthmatics. Eur. Respir. J. 6: 1368-1370, 1993[Abstract].

2.   Byrnes, C. A., S. Dinarevic, C. Busst, A. Bush, and E. A. Shinebourne. Is nitric oxide in exhaled air produced at airway or alveolar level? Eur. Respir. J. 10: 1021-1025, 1997[Abstract].

3.   Byrnes, C. A., S. Dianrevic, E. A. Shinebourne, P. J. Barnes, and A. Bush. Exhaled nitric oxide measurements in normal and asthmatic children. Pediatr. Pulmonol. 24: 312-318, 1997[Medline].

4.   Cunha, F. Q., J. Assreuy, D. W. Moss, D. Rees, L. M. Leal, S. Moncada, M. Carrier, C. A. O'Donnell, and F. Y. Liew. Differential induction of nitric oxide synthase in various organs of the mouse during endotoxaemia: role of TNF-alpha and IL-1-beta. Immunology 81: 211-215, 1994[Medline].

5.   Dillon, W. C., V. Hampl, P. J. Shultz, J. B. Rubins, and S. L. Archer. Origins of breath nitric oxide in humans. Chest 110: 930-938, 1996[Abstract/Free Full Text].

5a.   ECO Physics. CLD 770 AL ppt Chemiluminescence Analyzer. Dürnten, Switzerland: ECO Physics, 1993, chapts. 7.2 and 7.3.

6.   Evenson, K. M., R. J. Saykally, D. A. Jennings, R. F. Curl, and J. M. Brown. Chemical and Biochemical Application of Lasers, edited by C. B. Moore. New York: Academic, 1980, p. 95-138.

7.   Furukawa, K., D. G. Harrison, D. Saleh, H. Shennib, F. P. Chagnon, and A. Giaid. Expression of nitric oxide synthase in the human nasal mucosa. Am. J. Respir. Crit. Care Med. 153: 847-850, 1996[Abstract].

8.   Gerlach, H., R. Rossaint, D. Pappert, M. Knorr, and K. J. Falke. Autoinhalation of nitric oxide after endogenous synthesis in nasopharynx. Lancet 343: 518-519, 1994[Medline].

9.   Gustavsson, L. E., A. M. Leone, and M. G. Persson. Endogenous nitric oxide is present in the exhaled air of rabbits, guinea pigs and humans. Biochem. Biophys. Res. Commun. 181: 852-857, 1991[Medline].

10.   Hess, A., W. Bloch, J. Huverstuhl, and O. Michel. Alterated expression of inducible form of NO-Synthase (iNOS) in human nasal tissue after stimulation with cytokines. (Abstract). XVI Cong. Eur. Rhinolog. Soc. and VIII Cong. Internatl. Rhinolog. Soc., Ghent, Belgium, 8-12 Sept., 1996, p. 48.

11.   Hinz, A., J. Pfeiffer, W. Bohle, and W. Urban. Mid-infrared Faraday and Voigt effects for sensitive detection. Mol. Phys. 45: 1131, 1982.

12.   Kelm, M., and J. Schrader. Control of coronary vascular tone by nitric oxide. Circ. Res. 66: 1561-1575, 1990[Abstract/Free Full Text].

13.   Kharnitonov, S. A., and P. J. Barnes. Nasal contribution to exhaled nitric oxide during exhalation against resistance or during breath holding. Thorax 52: 540-544, 1997[Abstract].

14.   Kimberly, B., B. Nejadnik, G. D. Giraud, and W. E. Holden. Nasal contribution to exhaled nitric oxide at rest and during breathholding in humans. Am. J. Respir. Crit. Care Med. 153: 829-836, 1996[Abstract].

15.   Kirsten, A. M., R. A. Jörres, D. Kirsten, and H. Magnussen. Vergleich der nasalen und bronchialen Produktion von Stickstoffmonoxid bei Gesunden und Patienten mit Asthma. Pneumologie 51: 359-364, 1997[Medline].

16.   Koch, M., X. Luo, P. Mürtz, and W. Urban. Detection of small traces of 15N2 and 14N2 by Faraday LMR spectroscopy of the corresponding isotopomers of nitric oxide. Appl. Phys. B64: 683-688, 1997.

17.   Lundberg, J. O. N., T. Farkas-Szallasi, E. Weitzberg, J. Rinder, J. Lidholm, A. Änggard, T. Hökfeld, J. M. Lundberg, and K. Alving. High nitric oxide production in human paranasal sinuses. Nat. Med. 1: 370-373, 1995[Medline].

18.   McKellar, A. R. W. Mid-infrared laser magnetic resonance spectroscopy. Faraday Discuss. Chem. Soc. 71: 63, 1981.

19.   Moncada, S., R. M. J. Palmer, and A. E. Higgs. Nitric oxide: physiology, pathophysiology and pharmacology. Pharmacol. Rev. 43: 109-142, 1991[Medline].

20.   Palmer, R. M., A. G. Ferrige, and S. Moncada. Nitric oxide release accounts for the biological activity of endothelium-derived relaxing factor. Nature 88: 411-415, 1987.

21.   Persson, M. G., N. P. Wiklund, and L. E. Gustafsson. Endogenous nitric oxide in single exhalations and the change during exercise. Am. Rev. Respir. Dis. 148: 1210-1214, 1993[Medline].

22.   Robbins, R. A., A. A. Floreani, S. G. von Essen, J. H. Sisson, G. E. Hill, I. Rubinstein, and R. G. Townley. Measurement of exhaled nitric oxide by three different techniques. Am. J. Respir. Crit. Care Med. 153: 1631-1635, 1996[Abstract].

22a.   Sievers Instruments. NOA 280 NO Analyzer. Boulder, CO: Sievers Instruments, 1998.

23.   Suzuki, H., and J. A. Krasney. Nitric oxide in single-breath exhalation in humans. Jpn. J. Physiol. 47: 335-339, 1997[Medline].

24.   Tsujino, I., K. Miyamoto, M. Nishimura, H. Shinano, H. Makita, S. Saito, T. Nakano, and Y. Kawakami. Production of nitric oxide (NO) in intrathoracic airways of normal humans. Am. J. Respir. Crit. Care Med. 154: 1370-1374, 1996[Abstract].

25.   Urban, W. The carbon monoxide laser as spectroscopic source. Laser Optoelektronik 23: 56-61, 1991.

26.   Urban, W. Physics and spectroscopic applications of carbon monoxide lasers, a review. Infrared Phys. Technol. 36: 465-473, 1995.

27.   Urban, W., A. Hinz, W. Bohle, and D. Zeitz. Laser 85, edited by W. Weidelich. Berlin: Springer, 1986, p. 120-123.

28.   Van der Mark, T. W., E. Kort, R. J. Meijer, D. S. Postma, and G. H. Koeter. Water vapour and carbon dioxide decrease nitric oxide readings. Eur. Respir. J. 10: 2120-2123, 1997[Abstract].


J APPL PHYSIOL 86(3):1075-1080
8570-7587/99 $5.00 Copyright © 1999 the American Physiological Society



This article has been cited by other articles:


Home page
Am. J. Respir. Crit. Care Med.Home page
S. A. KHARITONOV and P. J. BARNES
Exhaled Markers of Pulmonary Disease
Am. J. Respir. Crit. Care Med., June 1, 2001; 163(7): 1693 - 1722.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar