|
|
||||||||
1 Department of Human Physiology, School of Medicine, Flinders University of South Australia, and Departments of 2 Critical Care Medicine and 3 Biomedical Engineering, Flinders Medical Centre, Adelaide, South Australia 5042, Australia
| |
ABSTRACT |
|---|
|
|
|---|
The
application of impedance pneumography for monitoring respiration in
small animals has been limited by problems with
calibration. With improved instrumentation, we describe
the calibration of tidal volume in anesthetized rats. The detection of
changes in voltage, reflecting the electrical impedance variations
associated with respiration, was optimized by using disposable adhesive
silver-silver chloride electrodes, advanced circuitry, and
analog-to-digital recording instrumentation. We found a linear
relationship between change in impedance and tidal volume in individual
rats (R2
98%), which was strongly influenced by rat weight. Consequently, a
calibration equation incorporating change in impedance and rat weight
was derived to predict tidal volume. Comparison of the predicted and
true tidal volumes revealed a mean
R2
98%,
slopes of ~1, intercepts of ~0, and bias of ~0.07 ml. The
predicted volumes were not significantly affected by either frequency
of respiration or pulmonary edema. We conclude that impedance
pneumography provides a valuable tool for the noninvasive measurement
of tidal volume in anesthetized rats.
impedance pneumography; monitoring respiration; spontaneous breathing; frequency; edema
| |
INTRODUCTION |
|---|
|
|
|---|
DIRECT MEASUREMENT of respiration in small animals presents numerous problems, as the techniques employed are frequently invasive and often impose relatively large dead spaces. Impedance (Z) pneumography offers the potential for the continuous, noninvasive measurement of breathing.
The principle of this technique is based on the change in Z (
Z) that
occurs across transthoracic electrodes during breathing. Z is defined
in terms of an instantaneous ratio of voltage
(V) and current
(I) by Z = V/I
(26), taking into account phase angle, analogous to Ohm's law. In
biological tissues, the current is carried by ions, the concentration
of which is relatively constant. The Z modulates a carrier waveform
passed through the chest between two electrodes to produce an output
that varies with respiration (4). It has been proposed that the
Z
during respiration is directly related to tidal volume
(VT) (25). Indeed, numerous studies have shown that the
Z during respiration correlates with spirometry and quantitatively relates to the moment-to-moment variations in the spirometer (4, 14).
However, in some studies
Z not directly related to
VT have been reported (13, 15).
Consequently, this type of Z pneumography has not been widely used. In
reality, the deficiencies may be more a consequence of the
instrumentation rather than the principle itself (23).
The aim of this study was to describe a quantitative method for
determining VT in rats, taking
advantage of the advances in both analog-to-digital recording
technology and the improvement and affordability of disposable
electrodes. Because Z may be influenced by the pattern of breathing,
electrode position, and lung water, we have also compared the
Z in
rats ventilated at various frequencies (f) with different electrode
positions and with experimentally induced pulmonary edema.
| |
METHODS |
|---|
|
|
|---|
Design Elements of the Impedance Pneumograph and Recording System
Respiratory movements were monitored by using a modified constant-current impedance pneumograph (IPG) based on a design first reported by Geddes and associates (9) in 1962 (Fig. 1). The IPG was used in the fully floating mode with bipolar electrode connection.
|
Briefly, the complete unit is built on a single circuit board and
incorporates alternating-current or direct-current output coupling.
Voltage regulators provide stable direct-current power rails of ±15
V, which supply various circuit components. A single transistor
Colpitts oscillator produces a low-voltage, high-frequency (50 kHz)
current, via a toroidal transformer, between two transthoracic electrodes. Two 10-k
resistors in series with the transformer output
provide sufficient internal Z to operate the IPG in essentially constant-current mode. Field-effect transistor (FET) input amplifiers with very high-input Z present virtually no loading on the measured circuit. The frequencies around 50-100 kHz are commonly used, as
they are high enough to avoid stimulation of tissue, electrode polarization, and problems associated with high skin Z values and, yet,
not so high as to encounter problems with radio interference (7).
The signal is amplitude modulated by
Z across the thorax, buffered
by the two FET input amplifiers, amplified by a differential amplifier,
and passed through high- and low-pass filters in series, effectively
forming a band-pass filter centered on 50 kHz. The narrow band of
frequencies (44-54 kHz) allowed to pass eliminates unwanted
interference, leaving only the 50-kHz carrier waveform modulated by the
Z signal. The signal emerging from the filters is then demodulated by a
precision full-wave rectifier to remove the carrier, with the final
output representing the
Z. The original circuit incorporated an
electrocardiogram filter and output, and, although this was included in
the updated IPG, it was not used.
The modifications to the IPG, including a higher 50-kHz voltage at the transformer output, allows a doubling of the series resistors, affording near-perfect constant-current operation. Modern components, such as the FET input amplifiers, allow absolute minimum loading on the electrode circuit. The offsetting capability was not included because the high-pass filter is effectively alternating-current coupled and the use of a precision LM308 differential amplifier made the offsetting facility redundant. The original circuit omitted a feedback resistor around the final operational amplifier of the precision rectifier; this has been corrected.
The IPG was interfaced with a MacLab system 4 analog-to-digital instrument and Chart software (version 3.5.3; AD Instruments, Sydney, Australia). Data were recorded at 100 Hz.
Anesthesia
Male porton rats (118-480 g) were anesthetized with methohexital sodium (60 mg/kg ip; Brietal, Eli Lilly, Sydney, Australia) and pentobarbital sodium (40 mg/kg ip; Nembutal, Boehringer Ingelheim, Sydney, Australia). Prolonged anesthesia was maintained by occasional injections of Brietal (30 mg/kg ip) and Nembutal (20 mg/kg ip). The thorax and abdomen were shaved, and the remaining hair was completely removed by using hair removal cream (Veet; Reckitt & Colman Pharmaceuticals, Sydney, Australia). Body temperature was monitored by using a rectal thermocouple (Yellow Springs Instruments, Yellow Springs, OH) and maintained at 37°C by placing the rats on a thermostatically controlled pad.Electrode Selection and Placement
Preliminary experiments in spontaneously breathing rats examined the effect of electrode type and position on the amplitude of the Z signal. Adhesive, disposable solid-gel monitoring electrodes (25 and 50 mm), 3M red dot Ag-AgCl electrodes (3M Health Care, London, Canada), and Neonatal Hydrogel Ag-AgCl electrodes (3M Health Care) were compared with subcutaneous stainless steel needles (0.45 × 13 mm; Becton-Dickinson Medical Products) and stainless steel dental needles (0.40 × 38 mm; Halas Dental, Sydney, Australia). Unlike the subcutaneous electrodes, the disposable electrodes were easy to attach, detected a strong Z signal with high reproducibility, and afforded a high degree of tolerance to changes in rat position. Although all adhesive types proved similarly effective, the 25-mm 3M red dot electrodes were used for the remainder of the study because of their affordability and suitable size.The position of the electrodes on the chest is reported to influence
the magnitude of the signal detected (2, 18). We, therefore, tested
several electrode positions and found the strongest
Z with the least
noise when the electrodes were placed on either side of the chest in
the midaxillary line at the xiphoid level, with the outer edges of the
electrodes almost meeting along the sternum and the lower edges lining
up with the base of the rib cage. Although a ground electrode was
theoretically unnecessary in the fully floating mode, one was applied
to the lower abdomen to minimize drift. Potentially, this electrode
provides more defined capacitive paths from each electrode and
associated leads to Earth.
Ventilation and Monitoring Cardiorespiratory Variables
A caudal artery was cannulated for sampling arterial blood and monitoring systemic arterial blood pressure and heart rate via a disposable pressure transducer (Sorenson Trans-pac; Abbott Critical Care Systems, Chicago, IL) interfaced to the MacLab recording system. The catheter was flushed continuously (syringe pump, model 355, Sage Instruments, Cambridge, MA) with heparinized isotonic saline (2 IU heparin/ml; David Bull Laboratories, Melbourne, Australia) at a rate of 1 ml/h.Baseline cardiorespiratory variables were recorded for 1 min in the spontaneously breathing anesthetized rats in both the supine and prone positions. Rats were then tracheostomized, by using a 16-gauge cannula, and were mechanically ventilated, by using a sinusoidal waveform where the inspiratory time/total time for the waveform = 0.5 (flexiVent; SCIREQ Scientific Respiratory Equipment, Montreal, Quebec, Canada), in the prone position at a f of 60 breaths per minute (bpm) (22).
Arterial blood (~0.15 ml) was collected into heparinized syringes (PICO; Radiometer, Copenhagen, Denmark), and blood gases were measured on a blood-gas-pH analyzer (model ABL 5; Radiometer). VT and f were adjusted to maintain the arterial PCO2 between 35 and 45 Torr. Respiratory f was determined by using the Chart software ratemeter function.
Calibration
The IPG was calibrated by using both control (n = 20) and edematous (n = 19) rats. Before calibration, the rats were given an additional dose of Brietal (15 mg/kg) via the lateral caudal vein at the hilum of the tail to further suppress respiratory drive. Although cardiac output was maintained for some minutes, this dose was lethal to nonventilated rats. The mechanical ventilator was momentarily set to deliver 0 ml and then programmed to increase VT in 0.5-ml increments up to 4.5 ml. VT was held at each increment for six breath cycles and then progressively lowered in the same manner. The procedure was repeated three times. On each occasion, a f of either 30, 60, or 120 bpm was randomly used. The total procedure took 6 min/rat.Induction of Edema
A polyethylene catheter (0.2-mm internal diameter × 0.5-mm external diameter) was inserted ~4 mm below the tracheostomy and directed to a point just above the carina. Haemaccel (Behring Institute, Marburg, Germany), containing Evans blue (15 mg/ml) as a marker, was instilled intratracheally into ventilated rats at a rate of 3 ml/kg over 30 min (compact infusion pump, model 975; Harvard Apparatus). The rats were placed prone at a 30° angle with the head tilted a further 5° to prevent reflux of the instillate. Again VT and f were adjusted to maintain the arterial PCO2 between 35 and 45 Torr. After instillation, the rats were returned to the horizontal position and calibrated as described in Calibration.Lung Isolation
After calibration, the thorax was opened and the heart and lungs were removed (22). The lobes were resected, weighed, and lyophilized to determine the wet-to-dry lung weight ratio.Data Calculation
The Z signal was calibrated by using linear regression analysis, solving the change in millivolts (
mV) for the
VT delivered for each rat
(y = intercept + slope · x), where
the
mV represents the change of magnitude of Z computed directly by
using the "Max
Min" software function for the fourth
breath cycle at any given VT
|
(1) |
|
(2a) |
|
(2b) |
|
|
(3) |
Statistics: Validation of Model
For any given
mV, a predicted
VT was calculated by rearranging
the calibration equation
|
|
|
(4) |
To establish the validity of the model we used the "PRESS" technique, otherwise known as the "leave one out" approach, as described in Draper and Smith (8), to predict the VT for each rat, at each f, both with and without edema. This technique involves leaving out one observation from the data analysis; the rest of the data is then used to construct a model to predict the observation omitted. This process is repeated for every observation, and, as before, the mean R2, slopes, intercepts, and bias ± 95% CI of the predicted vs. actual VT were determined. Unless otherwise stated, results are expressed as means ± 95% CI. Student's t-test was used for all comparisons.
| |
RESULTS |
|---|
|
|
|---|
Z Variations
Spontaneous breathing.
A typical Z trace from a spontaneously breathing rat is
shown in Fig.
2A.
Inspiration was associated with increased Z, and expiration with a
decrease. Sighs were recorded as large upward deflections occurring at
a rate of about one sigh every 3 min. The sensitivity of the detection
is illustrated by the small pulsatile deflections, consistent with
cardiac oscillations (16), that were superimposed on the major
respiratory fluctuations. However, it is also possible that these
inflections are due to electrical activity during myocardial
contraction. Similarly, breathing impinged on cardiac output,
particularly during sighs. The Z signal remained stable with
involuntary movement of the rat's legs, head, or tail, and changes
were absent during periods of apnea.
|
Mechanical ventilation.
The
Z signal remained remarkedly constant at any
given VT (Table
1). In individual rats, the
Z at any
VT was the same regardless of
whether VT was being increased
or decreased (Fig. 2B). If the lung
was held at constant volume, there was little, if any, noise or drift.
However, the minor pulsatile deflections, reflecting either cardiac
oscillations or electrical activity of the heart, persisted. When the
rat was taken off the ventilator, the absolute Z decreased to the level
found at 0 ml VT. Presumably,
this corresponds to functional residual capacity.
|
Effect of Body Position
Body position, whether supine or prone, did not affect the
Z, as reflected over five breath cycles recorded after 30 s in either position [spontaneous breathing; supine: 1.65 ± 0.114 (SE) mV, prone: 1.59 ± 0.123 mV;
n = 39 rats].
Extent and Distribution of Edema
Intratracheal instillation of Haemaccel resulted in a relatively even distribution of Evans blue (Fig. 3). The wet-to-dry lung weight ratio increased from 4.8 ± 0.03 (SE), similar to that reported previously (20), to 6.1 ± 0.12 (P < 0.001).
|
Derivation of Calibration Equations
A direct, linear relationship was found in each rat between
mV and
VT (all
R2
97%). This
relationship was clearly dependent on weight, with the smallest rats
having the greatest
Z. Regression analysis was, therefore, used to
examine the effect of rat weight on the
mV for each given
VT (Table
2). Because multiple measures were performed on each rat (i.e., at
VT = 0.5, 1.0, 1.5, 2.0 ml,
etc.), each data point cannot truly be considered an independent event. Therefore, rather than using standard multiple linear regression analysis to derive the calibration equation, we first solved the millivolts for ventilated VT in
each rat and then solved the acquired slopes
(slope1, Eq. 2a) and intercepts
(intercept1, Eq. 2b) for rat weight (Table
3). Clearly, the slope was more dependent
on rat weight than was the intercept.
|
|
Consequently, the final slopes (slope2 and slope3) and intercepts (intercept2 and intercept3), which were used to derive the calibration equations (Eq. 3), included both VT and rat weight in their derivation.
Correlation and Agreement Between Predicted and Actual VT
When the calibration equation from each group was rearranged to predict VT (Eq. 4), a high degree of correlation was obtained for all groups (R2: 98-99%), irrespective of respiratory rate or edema (Table 4). The slopes were ~1, and their CI crossed 1. There was minimal offset, as the intercepts were between
0.05 and 0.00 mV and the upper and lower CI crossed zero. The
difference between the predicted and actual
VT or "bias" was minimal
and was between 0.14 and 0.17 ml for the control and between
0.01 and
0.03 ml for the edematous rats. The 95% CI of
the bias were ~0.09 ml in the control and ~0.05 ml in the edematous
rats.
|
Model Validation: The PRESS Technique
When the PRESS technique was applied to each data set (i.e., 30, 60, or 120 bpm; control or edema), comparable R2, slopes, intercepts, and bias ± 95% CI were obtained (Table 5) to those described above (Table 4). This was also the case when the equations (Eq. 4) derived from the edematous rats were applied to the control data and vice versa (Table 6).
|
|
| |
DISCUSSION |
|---|
|
|
|---|
Our results demonstrate that Z pneumography can be used to determine VT accurately in small animals independently of f or pulmonary edema. The technique is relatively noninvasive and can be used to monitor breathing in anesthetized animals over prolonged periods.
Improvements in Pneumography
Pneumograph design.
At least three IPG circuit designs have been reported (1, 23). Unlike
the other circuits, the constant-current IPG requires no balance
adjustment and has a linear response with high Z tolerance, and
calibration is independent of total subject Z. Because it measures the
Z, it is the only circuit suitable for volumetric determination. In
addition, the "floating system" used avoided subject-ground
artifacts (23).
Electrode selection and placement. TYPE. Several electrodes, including some of those previously reported (17, 21), were evaluated. The adhesive Ag-AgCl disposable electrodes produced stable, reproducible, and noise-free signals with little drift over periods of >10 h (data not shown). In contrast, the stainless steel subcutaneous needle electrodes produced an erratic baseline with drift and noisy characteristics, similar to that reported by others (17, 27). This has been attributed to the generation of electrode offset potentials (24). Baseline drift is normally the result of long-term changes in the electrode offset potential, whereas short-term changes appear as noise on the trace. Another problem with the stainless steel electrodes is that they are not chemically inert. Reactions between the skin or tissue and the electrodes can produce soluble metallic salts, which are toxic and irritant and also affect offset potentials.
Unlike the stainless steel electrodes that store offset potentials and may take as long as 6 min to recover to an acceptable baseline, the offset potential of the adhesive Ag-AgCl electrode is small and decreases exponentially (24), hence producing a stable trace. The electrical stability of the adhesive Ag-AgCl electrodes has also been considerably enhanced through the use of electrolyte gel. CONFIGURATION. Both bi- (2, 18) and tetrapolar (1, 19, 26) electrode systems have commonly been used in Z pneumography. Whereas the four-electrode system can distinguish between subject base Z and
Z because of volume
changes, this system probably measures wall geometry associated with
respiration (10) rather than actual VT. The two-electrode
configuration we used afforded large signal detection and high
resolution and was resistant to outside interference.
PLACEMENT.
Because Z is dependent on the type, quantity, and distribution of
tissues, the surface area and position of the electrodes on the chest
are critical. Bone, lung, heart, and connective tissues comprise a
relatively constant contribution to total transthoracic electrical Z
such that intrathoracic gas and biological fluids become the main
variables. Whereas ionized fluids have a comparatively low resistance,
fat and air are highly resistive (19). Z measurements are, therefore,
influenced not only by VT, but
also by the underlying fat. Therefore, as the position of subcutaneous
electrodes may critically change, e.g., with alterations in muscle
tension with depth of anesthesia, the large surface area and adhesive
properties of the 3M electrodes afforded relatively high tolerance to
positioning and provided good reproducibility.
Consistent with previous reports (2, 15), we found the greatest
amplitude of Z excursions with the electrodes placed on either side of
the chest in the mid-axillary line at the xiphoid level, with a ground
electrode applied to the lower abdomen.
Recording system. Technological advances in analog-to-digital instrumentation and software have led to the development of sensitive, fast multichannel data recorders capable of defining high-fidelity signals. The system used records and displays experimental data on-line with excellent resolution and also has the advantage of computer-based data handling and analysis. These critical advances allow data manipulation and greater sensitivity that was lacking in earlier Z research.
Z Variations in the Rat
Effect of weight and body position on Z.
The
Z was directly related to
VT. Consistent with the findings
of Baker et al. (2), Berman et al. (5), and Hamilton et al. (11), who
reported no change in the relationship between Z and
VT with body position in humans
(2, 11) and dogs (5), body position did not affect
Z in rats.
Analogous to humans, the
Z at any
VT was greatest in the smallest
animals (2, 3). Consequently, rat weight was included in deriving the
calibration equations.
Comparison of predicted and actual VT. We used regression analysis to investigate the strength of the relationship between the ventilated VT and that predicted from the calibration equation. The bias was summarized for each group (f and edema) by calculating the mean difference between the predicted and actual VT. The precision was reflected by their 95% CI (6).
Comparison of the predicted and true VT by linear regression analysis showed R2
98%, with slopes and intercepts of ~1 and ~0, respectively (Table
4). In other words,
98% of the observed
Z can be explained by the
calibration equation. The bias was small, and the CI was narrow. In a
200-g rat, this amounts to an error of ~3% at resting VT.
Effect of frequency.
Consistent with the work of Baker and Hill (3), we found that the
f
over a wide range of respiratory rates, 30-120 bpm, had no effect
on
Z or the relationship between predicted and actual
VT. However, at high f the
baseline did not immediately return to zero during the expiration phase
of the volume calibration maneuver. Presumably, this reflects dynamic
hyperinflation, although we have no way of knowing this.
Effect of alveolar edema.
In agreement with others (5, 12), the absolute level of Z decreased
slightly over the 30-min period of Haemaccel infusion. The decrease in
absolute Z may have resulted from changes in regional ventilation,
perfusion, and/or increased lung fluid; however, the
Z did
not change. Consequently, the correlation between predicted and actual
lung VT was not affected by
edema or f. Moreover, there was no deterioration in the bias or the
95% CI. Indeed, comparable R2, slopes,
intercepts, and bias ± 95% CI were obtained when the equations
derived from the edematous rats were applied to the control data and
vice versa.
Z to accurately predict VT
over a wide range of f. Predicted
VT was not affected by edema. The IPG provides a valuable tool for the continuous, noninvasive measurement of VT and f in small
anesthetized animals.
| |
ACKNOWLEDGEMENTS |
|---|
We are grateful to Andrew Irving for technical assistance and Amanda Richter for constructing the impedance pneumograph.
| |
FOOTNOTES |
|---|
This research was supported by National Health and Medical Research Council of Australia Grant #950054 and by the Australian and New Zealand Intensive Care Society.
Address for reprint requests: I. R. Doyle, Dept. of Human Physiology, Flinders Univ., Bedford Park, South Australia 5042, Australia.
Received 19 June 1997; accepted in final form 2 October 1998.
| |
REFERENCES |
|---|
|
|
|---|
1.
Allison, R. D.,
E. L. Holmes,
and
J. Nyboer.
Volumetric dynamics of respiration as measured by electrical impedance plethysmography.
J. Appl. Physiol.
19:
166-173,
1964
2.
Baker, L. E.,
L. A. Geddes,
and
H. E. Hoff.
A comparison of linear and non-linear characterizations of impedance spirometry data.
Med. Biol. Eng.
4:
371-379,
1966[Medline].
3.
Baker, L. E.,
and
D. W. Hill.
The use of electrical impedance techniques for the monitoring of respiratory pattern during anaesthesia.
Br. J. Anaesth.
41:
2-17,
1969
4.
Barker, A.,
and
B. H. Brown.
Simple impedance pneumograph and volume integrator.
Med. Biol. Eng.
11:
352-353,
1973[Medline].
5.
Berman, I. R.,
W. L. Scheetz,
E. B. Jenkins,
and
H. V. Hufnagel.
Transthoracic electrical impedance as a guide to intravascular overload.
Arch. Surg.
102:
61-64,
1971[Medline].
6.
Bland, J. M.,
and
D. G. Altman.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1:
307-310,
1986[Medline].
7.
Cooley, W. L.,
and
R. L. Longoni.
A new design for an impedance pneumograph.
J. Appl. Physiol.
25:
429-432,
1968.
8.
Draper, N.,
and
H. Smith.
Multiple regression and mathematical model building.
In: Applied Regression Analysis (2nd ed.). New York: Wiley, 1981, p. 419-421.
9.
Geddes, L. A.,
H. E. Hoff,
D. M. Hickman,
M. Hinds,
and
L. E. Baker.
Recording respiration and the electrocardiogram with common electrodes.
Aerospace Med.
33:
791-793,
1962.
10.
Grenvik, A.,
S. Ballou,
E. McGinley,
E. Millen,
W. L. Cooley,
and
P. Safar.
Impedance pneumography: comparison between chest impedance changes and respiratory volumes in 11 healthy volunteers.
Chest
62:
439-443,
1972
11.
Hamilton, L. H.,
J. D. Beard,
and
R. C. Kory.
Impedance measurement of tidal volume and ventilation.
J. Appl. Physiol.
20:
565-568,
1965
12.
Hemstad, J. R.,
B. D. Spiess,
T. L. Machioro,
and
G. Raghu.
Pulmonary artery and noninvasive haemodynamics during lung lavage in primary alveolar proteinosis.
Chest
105:
1605-1608,
1994
13.
Hill, R. V.,
J. C. Jansen,
and
J. L. Fling.
Electrical impedance plethysmography: a critical analysis.
J. Appl. Physiol.
22:
161-168,
1967
14.
Kira, S.,
Y. Hukushima,
S. Kitamura,
and
A. Ito.
Transthoracic electrical impedance variations associated with respiration.
J. Appl. Physiol.
30:
820-826,
1971
15.
Kubicek, W. G.,
E. Kinnen,
and
A. Edin.
Calibration of an impedance pneumograph.
J. Appl. Physiol.
19:
557-560,
1964
16.
Lemke, R. P.,
S. A. Al-Saedi,
R. E. Alvaro,
N. E. Wiseman,
D. B. Cates,
K. Kwiatkowski,
and
H. Rigatto.
Use of a magnified cardiac airflow oscillation to classify neonatal apnea.
Am. J. Respir. Crit. Care Med.
154:
1537-1542,
1996[Abstract].
17.
Lewis, F. J.,
V. K. Moses,
M. L. Quinn,
and
M. L. Petrovic.
Automatic measurement of tidal volume in post-operative patients.
J. Thorac. Cardiovasc. Surg.
56:
825-831,
1968[Medline].
18.
Logic, J. L.,
M. G. Maksud,
and
L. H. Hamilton.
Factors affecting transthoracic impedance signals used to measure breathing.
J. Appl. Physiol.
22:
251-254,
1967
19.
Luepker, R. V.,
J. R. Michael,
and
J. R. Warbasse.
Transthoracic electrical impedance: quantitative evaluation of a non-invasive measure of thoracic fluid volume.
Am. Heart J.
85:
83-93,
1973[Medline].
20.
Nicholas, T. E.,
and
H. A. Barr.
Control of release of surfactant phospholipids in the isolated perfused rat lung.
J. Appl. Physiol.
51:
90-98,
1981
21.
Nicholas, T. E.,
J. H. T. Power,
and
H. A. Barr.
Surfactant homeostasis in the rat lung during swimming exercise.
J. Appl. Physiol.
53:
1521-1528,
1982
22.
Nicholas, T. E.,
J. H. T. Power,
and
H. A. Barr.
Effect of pattern of breathing on subfractions of surfactant in tissue and alveolar compartments of the adult rat lung.
Am. J. Respir. Cell Mol. Biol.
3:
251-258,
1990.
23.
Pacela, A. F.
Impedance pneumography-a survey of instrumentation techniques.
Med. Biol. Eng.
4:
1-15,
1966[Medline].
24.
Strong, P.
Electrodes.
In: Biophysical Measurements. Beaverton, OR: Tektronix, 1973, p. 219-247.
25.
Valentinuzzi, M. E.,
L. A. Geddes,
and
L. E. Baker.
The law of impedance pneumography.
Med. Biol. Eng.
9:
137-163,
1971.
26.
Van de Water, J. M.,
B. F. Mount,
J. R. Barela,
R. Schuster,
and
F. S. Leacock.
Monitoring the chest with impedance.
Chest
64:
597-603,
1973
27.
Ward, H. E.,
and
T. E. Nicholas.
Effect of artificial ventilation and anaesthesia on surfactant turnover in rats.
Respir. Physiol.
87:
115-129,
1992[Medline].
This article has been cited by other articles:
![]() |
C. G. De Pasquale, A. D. Bersten, I. R. Doyle, P. E. Aylward, and L. F. Arnolda Infarct-induced chronic heart failure increases bidirectional protein movement across the alveolocapillary barrier Am J Physiol Heart Circ Physiol, June 1, 2003; 284(6): H2136 - H2145. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Davidson, A. D. Bersten, H. A. Barr, K. D. Dowling, T. E. Nicholas, and I. R. Doyle Lung function, permeability, and surfactant composition in oleic acid-induced acute lung injury in rats Am J Physiol Lung Cell Mol Physiol, December 1, 2000; 279(6): L1091 - L1102. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |