Vol. 86, Issue 2, 647-650, February 1999
Infrared CO2 analyzer
error: an effect of background gas
(N2 and
O2)
R.
Arieli,
O.
Ertracht, and
Y.
Daskalovic
Israel Naval Medical Institute, Israel Defense Forces Medical Corps,
Haifa 31080, Israel
 |
ABSTRACT |
Three
infrared CO2 analyzers were tested
for the effect of background gases: the Ametek CD-3A (Ametek, Thermox
Instruments Division, Pittsburgh, PA), the Dräger Multiwarn P
CO2 (Dräger, Lübeck,
Germany), and the Servomex 1440 (Servomex, Crowborough, East Sussex,
UK). Various CO2 concentrations
were prepared with Wösthoff precision pumps (H. Wösthoff,
Bochum, Germany). Calibration with a different background gas
(O2 or
N2) caused a similar but systematic error in the CO2
readings of all three analyzers. When the
CO2 analyzers were calibrated with
N2 as the background gas, the
CO2 reading in an
O2-enriched atmosphere was 8%
lower than the true value. Conversely, calibration with
O2 as the background gas resulted
in a 10% overestimation of CO2
levels when N2 was the background
gas. This error may be important in a few fields of respiratory physiology.
carbon dioxide calibration; hyperoxia; alveolar-arterial
difference
 |
INTRODUCTION |
CARBON DIOXIDE
(CO2) is a very important gas in
human physiology and is extensively monitored in medical treatment,
physiological studies, and various environmental situations. Inspired
gas enriched with O2 is used in
many fields of human activity. For example, combinations of various
levels of O2 and
CO2 are common in diving and in
medical treatment. At our institute, where we practice hyperbaric
O2 therapy and study the
physiology of closed-circuit diving, combinations of various levels of
O2 and
CO2 are common. Infrared
CO2 analyzers are the most widely
used. In different situations, we came up with conflicting results by
using both our mass spectrometer and infrared
CO2 analyzer after calibration with certified commercial calibrating mixtures. Discrepancies in the
readings of the CO2 concentration
in a gas mixture and unacceptable respiratory ratios are two examples.
We therefore set out to study this conflict by using the Wösthoff
precision pumps (H. Wösthoff, Bochum, Germany) to produce various
precise concentrations of CO2 with
either N2 or
O2 as the background gas. These
precision pumps are known for their accuracy, and an analysis of a gas
mixture prepared by the precision pumps that we performed by using the
Micro-Scholander gas analyzer was proven accurate to within
0.01-0.02%. Infrared CO2
analyzers' instruction manuals do not indicate any effect of
N2 or
O2 on
CO2 sensing.
 |
METHODS |
Three infrared CO2 analyzers and a
mass spectrometer were used.
Ametek CD-3A (Ametek, Thermox Instruments Division, Pittsburgh,
PA).
The instruction manual for the Ametek CD-3A (1) suggests zero
calibration with either air or N2
and span calibration with a
CO2-containing mixture
(CO2 >4% and up to 15%). There
is no reference to interference by
O2 or
N2.
Servomex 1440 (Crowborough, East Sussex, UK).
The instruction manual for the Servomex PA 404 CO2 analyzer suggests
"narrow-band interference filters and a solid-state detector provide
a measurement which is inherently less affected by
cross-sensitivity" (10). The manual does suggest some background
effect of gases on the O2 sensor,
but there is no mention of the CO2
sensor. It is suggested that "these effects can be compensated by
either using the background gas as a zero or by offsetting the
N2 zero point by the amount of
error induced by the background gas" (10). There is no mention of
cross sensitivity of N2 or
O2 on
CO2 sensing.
Dräger Multiwarn P CO2
(Lübeck, Germany).
The instruction manual for the Dräger Multiwarn P
CO2 (6) suggests zero calibration
with pure N2 and span calibration with a CO2-containing mixture.
They suggest calibrating with a "concentration corresponding to the
typical values which are to be measured" (6). Nothing is mentioned
regarding cross sensitivity of N2
or O2.
The mass spectrometer used was the CaSE 9000T/BG (Biggin Hill, UK).
We used Wösthoff precision pumps to produce various
concentrations of CO2 with either
pure O2 or pure
N2. The
CO2 analyzer was calibrated with a
5% mixture coming from the pumps. The outflow (550 ml/min) from the
precision pumps passed through a rubber glove (to smooth out the
pressure fluctuations), from which it went on to the analyzer sensor
via a hole cut in one of the fingers. The values read for various
concentrations were recorded. To rule out asymmetry between the two
pumps, one was used for CO2 and the other for the background gas, and after a few readings they were
switched. The results of changing the pumps when
CO2 was read by using the Ametek
CD-3A proved symmetry. In the test, we calibrated the
CO2 analyzer with 0 and 5%
CO2 produced by the precision
pumps by using pure N2 as the
background gas. We then produced, in ascending and descending order,
various concentrations of CO2 with
the precision pumps, reading the
CO2 concentration on the
analyzer. We then changed the background gas to
O2 and read the
CO2 concentrations for various
combinations of the precision pumps' input. The background gas was
then changed back to N2, and we
read a few more concentrations to make sure that no drift had occurred
in the interim. After the completion of this series, we changed the
background gas. We calibrated the analyzer with O2 as the background gas and
proceeded with the previous protocol, interchanging
O2 and
N2.
 |
RESULTS |
The CO2 readings obtained from the
CaSE mass spectrometer (calibrated with a certified commercial mixture,
Scott Specialty Gases, Plumsteadville, PA), with either
N2 or
O2 as the background gas, are
shown in Fig. 1. In Figs. 1 and
2, we present the difference between the
analyzer readings of CO2 and the
precision pumps' input, on the
y-axis, as a function of the
percentage of CO2 in the precision
pumps' input. It can be seen that the
CO2 readings deviate only slightly
from the input of the precision pumps and are similar for both
background gases. When the Ametek CD-3A was calibrated by using
N2 as the background gas, other
CO2 readings showed small
deviations from the expected concentrations (Fig. 2A). However,
CO2 readings deviated to a greater
degree as a function of input CO2
concentration when the background gas was
O2. The slope of the decrease was
0.075 (%error/%CO2,
r2 = 0.99). There was no difference between new
CO2 readings and those from the
initial set, when the background gas was changed back to
N2. Calibration of the Ametek
CD-3A with O2 as the background gas gave the opposite results (Fig.
2B). The different
CO2 readings with
O2 as the background gas yielded
small deviations from the expected values, but replacing the
O2 with
N2 produced positive deviations
that are linearly related to the input
CO2 concentration (slope = 0.11, r2 = 0.98).
Results similar to those obtained with the Ametek CD-3A were derived by
using the Servomex 1440 (Fig. 2, C and
D). The slope for the deviation of
CO2 concentrations in
O2 from the input when the
calibration background gas was N2
was
0.080
(r2 = 0.99), and,
in N2 when the calibration
background gas was O2, was 0.10 (r2 = 0.99).
Similar results were obtained with the Dräger Multiwarn P
analyzer (Fig. 2, E and
F). When the background gas was the same as had been used for calibration, the
CO2 readings yielded a reclining
s-shaped response that crossed the zero at the calibration concentration. However, the deviation of
CO2 concentrations for the other
background gas (the gas that had not been used for calibration) was
similar to that seen in the Ametek CD-3A readings: underestimation when
the background gas was O2 (Fig.
2E) and overestimation when the
background gas was N2 (Fig.
2F). The deviations were corrected by subtracting the deviation for the calibration background gas from
the deviation of the noncalibration background gas (
in Fig. 2,
E and
F). The results gave a linear
relationship for the corrected deviation as slope =
0.085,
r2 = 0.97, and
slope = 0.09, r2 = 0.78 in Fig. 2, E and
F, respectively. The nonlinear
concentration readings of the Dräger analyzer may be inferred
from their instruction manual, which mentions a possible error of
±5% for the values 0 to the calibration value and ±10% from
the calibration value to 1.5 times this value (6). In all the
analyzers, the zero setting was not sufficient on recalibration with
another background gas; sensitivity had to be calibrated as well.

View larger version (21K):
[in this window]
[in a new window]
|
Fig. 1.
CaSE mass spectrometer readings of
CO2 concentrations. Difference
between CO2 concentration readings
and precision pumps' input is shown as a function of
CO2 input from H. Wösthoff
precision pumps. Both pure O2 and
pure N2 were used as background
gas for pure CO2 ( and ,
respectively).
|
|

View larger version (27K):
[in this window]
[in a new window]
|
Fig. 2.
Ametek CD-3A (A and
B), Servomex 1440 (C and
D), and Dräger Multiwarn P
(E and
F) readings of
CO2 concentrations. Difference
between CO2 concentration readings
and precision pumps' input is shown as a function of
CO2 input from H. Wösthoff
precision pumps. Right: results from
calibration with O2 as background
gas; left: results from calibration
with N2 as background gas. ,
CO2 readings when
N2 was background gas; ,
readings when background gas was
O2; , differences between
errors for N2 and
O2 with Dräger Multiwarn
P.
|
|
 |
DISCUSSION |
A CO2 infrared analyzer responds
differently, depending on whether the background is
N2 or
O2. Because most common commercial calibrating tanks contain a high percentage of
N2,
CO2 readings in an
O2-enriched atmosphere are
underestimated. For the three analyzers we tested (Ametek, Servomex,
Dräger), this underestimation was
0.075,
0.085, and
0.080, respectively, for each 1%
CO2. Therefore, calibrating an
infrared CO2 analyzer with
N2 as the background gas would
result in an 8% underestimation of
CO2 levels in
O2 as background. For example,
when the analyzer reading is 4.6%
CO2, the true concentration is
5%. Conversely, calibrating the three analyzers with
O2 as the background gas gave an
overestimation of 0.11, 0.09, and 0.10, respectively for each 1%
CO2 in
N2. Thus calibrating the infrared
CO2 analyzer with
O2 as the background gas will
result in a 10% overestimation of
CO2 levels in a
N2-rich atmosphere.
There are a few fields in respiratory physiology where an error in the
CO2 reading could seriously affect
the outcome. In hyperbaric physiology, a small change in
CO2 concentration can make a large
difference, because PCO2 is the
product of the fraction of CO2 and
pressure. CO2 is an important
factor in O2 toxicity (3), which
will be enhanced by an inspired PCO2 as small as 1 kPa (2). CO2 is an
important respiratory drive in hyperbaric conditions, and professional
divers tend to hypoventilate and retain
CO2 in their tissues (7). Diving
with breathing mixtures containing various concentrations of
O2 (Nitrox or closed-circuit diving) would affect the CO2
monitored with infrared analyzers. The alveolar-arterial
CO2 difference is negligibly small
in the healthy lung and is therefore used to calculate alveolar
PCO2 from arterial
PCO2 (5). An 8-10% error would
represent a serious flaw in the computation of alveolar-arterial gas
exchange. Various tests use the switch from an air- to an
O2-filled spirometer (4), and the
CO2 reading would be affected if
an infrared analyzer were used. In clinical medicine, if the analyzer
is calibrated with N2 as the
background gas, alveolar PCO2 may be
underestimated in patients breathing an enriched
O2 mixture. In artificial
respiration, for example, when the ventilator is set to give 5%
CO2 in the alveolar gas, the true
value may be close to 4.6%.
Lauber et al. (8) tested the accuracy of various infrared
CO2 analyzers as affected by an
array of background gases and conditions. They concluded that "all
tested analyzers were found to be safe for clinical use," allowing
for a 12% error. However, the error established by the present study,
which is probably related to the collision-broadening effect of
O2 (9), should be considered in
scientific research. Presently, only a gas mixture with the same
O2/N2
ratio as that used in the planned experiment can be used
as a calibrating mixture in infrared
CO2 analyzers.
 |
ACKNOWLEDGEMENTS |
The authors thank R. Lincoln for skillful editing.
 |
FOOTNOTES |
The opinions and assertions contained herein are the private ones of
the authors and are not to be construed as official or as reflecting
the views of the Israel Naval Medical Institute.
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests: R. Arieli, Israel Naval Medical
Institute, PO Box 8040, Haifa 31080, Israel.
Received 13 March 1998; accepted in final form 4 September 1998.
 |
REFERENCES |
1.
Ametek/Thermox.
Ametek CD-3A. Pittsburg, PA: Ametek/Thermox, 1986.
2.
Arieli, R. Hazards of CNS oxygen toxicity and the effects of
metabolic rate and CO2 production
(Abstract). Annu. Meet. Israel Physiol. Soc. Maale Hahamisha,
Israel, November 1997.
3.
Clark, J. M.
Oxygen toxicity.
In: The Physiology and Medicine of Diving (4th ed.), edited by P. B. Bennett,
and D. H. Elliott. London: Saunders, 1993, p. 121-169.
4.
Cunningham, D. J. C.,
P. A. Robbins,
and
C. B. Wolff.
Integration of respiratory responses to changes in alveolar partial pressures of CO2 and O2 and in arterial pH.
In: Handbook of Physiology. The Respiratory System. Control of Breathing. Bethesda, MD: Am. Physiol. Soc., 1986, sect. 3, vol. II, pt. 2, chapt. 15, p. 475-528.
5.
Dejours, P.
Principles of Comparative Respiratory Physiology. Amsterdam: North-Holland, 1975, p. 102.
6.
Dräger.
Dräger multiwarn p. Lübeck, Germany: Dräger, 1991.
7.
Lanphier, E. H.,
and
E. M. Camporesi.
Respiration and exertion.
In: The Physiology and Medicine of Diving (4th ed.), edited by P. B. Bennett,
and D. H. Elliott. London: Saunders, 1993, p. 77-120.
8.
Lauber, R.,
B. Seeberger,
and
A. M. Zbinden.
Carbon dioxide analysers: accuracy, alarm limits and effects of interfering gases.
Can. J. Anaesth.
42:
643-656,
1995[Medline].
9.
Raemer, D. B.,
and
I. Calalang.
Accuracy of end-tidal carbon dioxide tension analyzers.
J. Clin. Monit.
7:
195-208,
1991[Medline].
10.
Servomex.
Servomex 1440. Crowborough, UK: Servomex, 1986.
J APPL PHYSIOL 86(2):647-650
8570-7587/99 $5.00
Copyright © 1999 the American Physiological Society