|
|
||||||||
1 Biobehavioral Research
Laboratory, A wearable, ambulatory impedance monitor
(AIM) has been developed to permit impedance cardiographic measurements
while patients and volunteers engage in normal daily
activities. The AIM system was developed for use with a
new hybrid tetrapolar spot-band electrode configuration and was
designed to be comfortable and inconspicuous. The
objective of the present study was to provide a preliminary evaluation
of AIM comparability with the widely validated Minnesota model 304B
impedance cardiograph with standard tetrapolar band electrodes.
Orthostatic challenge was used to systematically alter cardiac function
in a laboratory setting in 11 healthy men and women. Both while the
subjects were sitting and while they were standing, the AIM yielded
measures of cardiac function, including heart rate, preejection period,
left ventricular ejection time, and stroke volume, that were similar to
those acquired by using the reference Minnesota 304B system (all
Pearson R correlations > +0.87, all
P < 0.001). Cardiac responses to
postural shift, expressed as change measures from sitting to standing,
were also comparable for the AIM and Minnesota reference monitoring
systems. Potential applications, including the assessment of 24-h
hemodynamic profiles, are illustrated and discussed.
ambulatory monitoring; cardiac function; preejection period; stroke
volume; hemodynamics
IMPEDANCE CARDIOGRAPHY has been used to examine
cardiovascular responses to a wide range of stimuli. These include
responses to exercise (3), mental stress (15), sleep (1), orthostatic challenge (2), and a variety of pharmacological challenges (5). Studies
reported to date have been conducted almost exclusively in the
laboratory or clinical environment, because the typical measurement
equipment is of limited portability.
Measurement of blood pressure (BP) and the electrocardiogram (ECG) was
once also restricted to laboratory and clinical settings, but
technological advances have led to the availability of ambulatory BP
(ABP) and Holter monitors that record physiological data during the
patient's and volunteer's normal daily activities. Over the past 10 years, ambulatory studies have yielded new insights into both normal
physiological processes and pathophysiological mechanisms. Examples
from Holter monitoring include spectral analysis of heart rate (HR)
variability to index autonomic control (11) and ECG analysis of
myocardial ischemia in coronary heart disease (12). ABP
monitoring has led to the identification of "white coat
hypertension" (4), and ABP has been shown to correspond more closely
with end-organ damage than does clinic BP (8).
Impedance cardiography provides assessments of a number of indexes of
cardiac function, including HR, preejection period (PEP; an inversely
related index of myocardial contractility), left ventricular ejection
time (LVET), the Heather index (HI; an index of myocardial
contractility), and stroke volume (SV). Cardiac output (CO) may also be
derived from the HR and SV measurements, and simultaneous BP
measurement permits the derivation of systemic vascular resistance
(SVR). Indeed, a growing body of literature of laboratory and clinic
studies has united impedance cardiography with BP measurement to
document hemodynamic responses in terms of BP, CO, and SVR patterning
(15).
The primary objective of the present study was to assess how a newly
developed ambulatory impedance monitor (AIM) would compare with
established impedance cardiographic instrumentation. The AIM was
developed to be used with a recording electrode configuration designed
to maximize signal fidelity while simultaneously considering patient
comfort. The Minnesota model 304B with a standard tetrapolar band
electrode configuration was used as a reference impedance cardiographic
system, because it remains the most widely validated standard (13).
Moreover, the study's objective was not to revisit the validation of
impedance cardiography as a technique per se but rather to evaluate
whether the AIM system, with its unique hardware and electrode
configuration, would yield cardiovascular function indexes similar to
those that are obtained with standardized impedance cardiographic
methodology. Orthostatic responses associated with volitional relaxed
sitting and standing postures were used to systematically alter cardiac
function, providing an ecologically relevant manipulation suitable for
the laboratory assessment of a monitor designed for ambulatory
recording. In addition, one individual volunteered to wear the
instrument for 24 h on a typical workday, and his data are presented to
illustrate the potential value of ambulatory impedance cardiography.
Subjects
![]()
ABSTRACT
Top
Abstract
Introduction
Methods
Results
Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Methods
Results
Discussion
References
![]()
METHODS
Top
Abstract
Introduction
Methods
Results
Discussion
References
Measurements
Minnesota 304B. The Minnesota model 304B (IFM, Greenwich, CT) is a commercially available impedance cardiograph that has been the most widely used in research studies (13). This instrument utilizes a 4-mA constant-current source with 100-kHz oscillator frequency. The Minnesota 304B also includes ECG signal input and amplification. The tetrapolar band electrode configuration, first described by Kubicek and colleagues (6), was employed for impedance signal [thoracic impedance (Z0) and first derivative of pulsatile impedance (dZ/dt)] acquisition, as it represents the reference standard that has been most extensively validated (13). The inner two recording electrode bands were positioned around the base of the neck and around the thorax over the tip of the xiphoid process. The outer two current electrode bands were positioned to encompass the neck and thorax, at least 3 cm away from each of the recording electrodes. The ECG was recorded independently by using disposable ECG electrodes (Cleartrace, Conmed, Utica, NY) placed on the chest in a lead II configuration.
AIM. The AIM is a microcomputer-based wearable bioelectric impedance monitor and signal processing system designed for 24-h ambulatory measurement. It consists of a small 3 × 4 × 1.5-in. plastic enclosure that contains a credit-card-sized bioelectric impedance cardiograph, a credit-card-sized microcomputer and data logger, and a 9-V battery. The AIM generates an 80-kHz, 2-mA constant sine wave alternating current. The AIM computer section ensemble averages, analyzes, and stores the ECG, dZ/dt, and Z0 waveforms and the computed cardiac function indexes during each measurement sequence.
A tetrapolar combination of spot and band electrodes was specifically developed for use with the AIM system. The recording electrodes were Mylar band electrodes placed around the base of the neck and around the thorax over the tip of the xiphoid process, identical to that for the Minnesota reference configuration (in this validation study, it was, therefore, necessary to apply the inner two recording band electrodes only once, for use with both instruments). Disposable ECG spot electrodes (Cleartrace) were used as current electrodes, with one applied behind the right ear (over the base of the mastoid process) and the other over the lower right rib cage, 6 cm below the lower recording band electrode. ECG recording with the AIM system utilized only one additional disposable spot electrode, placed on the lower left rib cage; the two impedance current spot electrodes served simultaneously as sources for the ECG signal, approximating a lead II configuration. The AIM spot-band impedance electrode configuration and ECG electrode placement are illustrated in Fig. 1. The AIM was worn on a belt around the waist during the entire test protocol.
|
Procedure
The study was designed to assess the effects of posture on impedance-derived indexes of cardiovascular function, measured by the two impedance cardiographic systems. Measurements from the two systems could not be recorded concurrently because of electrical interactions between them. Therefore, sitting and standing measurements were recorded with one system and then repeated with the other, with order counterbalanced across subjects.The test procedure began by instrumentation of subjects with ECG and impedance electrodes. Next, subjects were randomized to one of two possible test orders, and either the AIM with two-band and two-spot electrode configuration or the Minnesota with four-band electrodes was set up for signal acquisition. Subjects were seated and asked to relax quietly for 5 min. In each of the last 2 min, a 30-s sequence of ECG and impedance signals was recorded and processed to derive HR, SV, PEP, LVET, and HI, which were representative of relaxed sitting values. Next, subjects were asked to stand and adopt a comfortable posture for 5 min. Again, in each of the last 2 min, ECG and impedance signals were recorded for 30 s and processed to derive HR, SV, PEP, LVET, and HI, which, this time, were representative of relaxed standing values. This procedure was then replicated exactly but using the impedance system (AIM or Minnesota) designated for testing in second order by the counterbalancing randomization.
Derivation of Cardiac Function Measures
The basal Z0, the dZ/dt, and the ECG waveforms from both the AIM and Minnesota impedance cardiographs were acquired, each at a 500-Hz sample rate, and were processed by using specialized ensemble-averaging software (COP, BioImpedance Technology, Chapel Hill, NC). This system utilizes ensemble averaging to filter noise and respiratory artifact from the impedance cardiographic waveforms and has been validated against the invasive thermodilution technique (14). In accordance with recommended standards (13), PEP was computed as the time interval (ms) between the ECG Q wave and the dZ/dt B point, LVET was computed as the time interval (ms) between the dZ/dt B point and X point, SV was derived by using the Kubicek equation (6), and HI was computed as maximum dZ/dt (dZ/dtmax) divided by the Q-wave-to-dZ/dtmax time interval.Data Analysis
The effects of the impedance monitoring system (AIM and Minnesota) and posture (sitting and standing) were evaluated by using repeated-measures analysis of covariance (ANCOVA) tests for each cardiovascular measure. The order of testing (AIM and Minnesota) was entered as a covariate. Correlation analyses (Pearson's R) were employed to evaluate the correspondence between the two measurement systems across individuals. The effects of posture were also defined as percent change for each cardiovascular measure from sitting to standing posture and were analyzed by repeated-measures (AIM and Minnesota) ANCOVA. An
level
of 0.05 was adopted.
| |
RESULTS |
|---|
|
|
|---|
Posture and Monitor Effects
Mean values for the five indexes of cardiac function measured while the subjects were sitting and standing with the AIM and Minnesota monitoring systems are shown in Table 1. Standing, compared with sitting, was associated with significantly higher HR (P < 0.05), longer PEP (P < 0.05), and shorter LVET (P < 0.01). There were also nonsignificant trends toward lower SV (P < 0.1) and smaller HI (P < 0.1) associated with standing compared with sitting. There were no significant differences for any of the measures associated with the impedance monitoring system used, nor was there any monitor by posture interactions.
|
Correlation Analyses
Pearson R values for the correlations between the AIM and Minnesota cardiac function indexes measured while the subjects were sitting and standing are shown in Table 2. As may be seen, all correlations are positive and statistically reliable, indicating good correspondence between the two measurement systems.
|
Postural Responses to Standing
For each index of cardiac function, the response to standing was computed as the difference between standing and sitting values, expressed as a percent change from the sitting value. The mean postural responses measured by using the AIM and Minnesota systems are illustrated in Fig. 2. As may be seen, the AIM and Minnesota systems indicated similar postural responses for all measured variables. Repeated-measures ANCOVAs confirmed that the two impedance systems indicated similar responses (all P = nonsignificant). Finally, correlations of percent change from sitting to standing were positive and significant for HR (R = 0.79, P < 0.01), SV (R = 0.68, P < 0.05), and HI (R = 0.83, P < 0.01) and were positive but nonsignificant for PEP (R = 0.38) and LVET (R = 0.47).
|
Example 24-h Hemodynamic Study
One subject, a 41-yr-old healthy man, volunteered to undergo 24-h testing on a typical workday. In addition to the AIM system, the subject was instrumented with the Suntech (Raleigh, NC) Accutracker ABP monitor, programmed to take a BP measurement every 15-min during waking hours and every 30-min during sleep. The ABP monitor activated the AIM device to initiate a 30-s ensemble average impedance data acquisition concurrent with every ABP measurement. The 24-h recording period tracked the course of a normal workday, evening relaxation at home, and overnight sleep. SVR was derived from each of the 84 simultaneously recorded BP and CO values, i.e., SVR (dyn · s · cm
5) = (mean BP/CO) * 80, where mean BP = [(systolic BP
diastolic BP)/3] + diastolic BP. As may be seen from the 24-h
hemodynamic profile obtained, illustrated in Fig.
3, systolic and diastolic BP showed a
circadian variation, with the lowest pressures occurring during
overnight sleep and the highest during the workday. Interestingly, SVR
was the more significant determinant of BP variations across the 24-h
period (systolic BP/SVR correlation = +0.35,
P < 0.01; diastolic BP/SVR
correlation = +0.50, P < 0.001),
with CO showing a nonsignificant positive correlation. However, posture
was associated with substantial alterations in SVR. Controlling for
posture, over the course of the daytime waking hours (6 AM to 11 PM),
CO was revealed to be strongly related to systolic BP variations (R = +0.61,
P < 0.01), whereas SVR maintained
its association with diastolic BP
(R = +0.32,
P < 0.05). It is of note that, in considering these associations, it should be remembered that the SVR
and BP measures are computationally dependent.
|
| |
DISCUSSION |
|---|
|
|
|---|
The AIM yielded measures of cardiac function that were comparable with those acquired by using the impedance cardiographic standard: the Minnesota 304B with a tetrapolar band electrode system. The AIM system incorporated a new hybrid spot-band electrode configuration, designed to be less conspicuous and more comfortable for 24-h monitoring. Nonetheless, the AIM generated SV, PEP, LVET, and HI values that were comparable with those measured by using the impedance standard. These findings were evident when individuals were tested both while sitting and while standing. Moreover, the cardiac responses to postural shift, expressed as change measures from sitting to standing, were also comparable for the AIM and Minnesota reference monitoring systems. These observations suggest that the AIM system is capable of measuring cardiac function with the same strengths and limitations that have been established for the impedance cardiographic technique, based on >100 previous validation studies employing the standard Minnesota system and a variety of invasive and noninvasive reference techniques (13). In other words, the AIM system should yield valid measures of systolic time intervals (PEP and LVET) in absolute terms and valid relative change measures for SV and CO (13).
The ensemble-averaging approach to signal acquisition and processing was a common aspect of both the AIM and Minnesota systems. This approach has been shown to effectively filter movement and respiratory artifact from the impedance cardiogram, making it especially well suited to monitoring during physical activity (10). These strengths of ensemble averaging are, therefore, likely to excel at optimizing impedance signal fidelity for the recording of individuals performing their normal daily activities in the ambulatory environment.
The observations over 24 h of hemodynamic monitoring made in one of the study subjects provide an illustration of how ambulatory impedance cardiography may present the opportunity for hemodynamic studies in the natural environment. One limitation of the 24-h hemodynamic data illustrated in the present report is that they are not validated. Nonetheless, it is of note that the impedance waveforms recorded were of surprisingly high fidelity, presenting no ambiguity in their interpretation and the computation of the various indexes of cardiac function. Therefore, there was no indication that the 24-h hemodynamic data lacked the validity generally attributed to impedance cardiographic data. Because there is, presently, no other technology available to make possible 24-h ambulatory CO measurements, it is likely to be difficult to establish the validity of the AIM system in the ambulatory environment.
Naturalistic studies of the hemodynamic mechanisms of BP regulation may help further our understanding of the etiology of cardiovascular disease. Laboratory studies have indicated that elevated SVR is a feature of vascular hypertrophy and hypertension (7), whereas recurrent surges in CO may promote vascular endothelial damage and atherosclerosis (9). The AIM system, by permitting the assessment of hemodynamic phenomena in the natural environment, may help further our understanding of the pathophysiology of cardiovascular disease.
In summary, the AIM holds the potential for providing insights into previously unexplored aspects of cardiovascular control during the activities of daily living. Its comparability to the Minnesota standard suggests that the data generated by the AIM should share the same strengths and limitations offered by impedance cardiography in general (13). Strengths include the noninvasive, unobtrusive, and risk-free nature of the impedance technique. Limitations include the prerequisite of a structurally normal heart and volumetric measures (i.e., SV and CO) that are generally considered to be limited in validity to relative change indexes. Just as ambulatory ECG and ABP assessment has significantly advanced medical knowledge, ambulatory impedance cardiography holds the promise of being a useful tool both in basic physiological research and in clinical investigations. In combination with ABP monitoring, the ability to perform 24-h hemodynamic assessments may be especially valuable for studying cardiovascular disease.
| |
ACKNOWLEDGEMENTS |
|---|
The authors express their gratitude to Kristy Johnson for technical support, Richard Lutz for computer support, and Ann Wilson for administrative support.
| |
FOOTNOTES |
|---|
This research was supported by National Heart, Lung, and Blood Institute Grants HL-49427 and HL-53724 and National Institute of Mental Health Grant 5T32-MH-19109.
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests: A. Sherwood, Box 3119, Duke Univ. Medical Center, Durham, NC 27710 (E-mail: sherw002{at}mc.duke.edu).
Received 18 March 1998; accepted in final form 21 August 1998.
| |
REFERENCES |
|---|
|
|
|---|
1.
Bunnell, D. E.,
W. C. Bevier,
and
S. M. Horvath.
Effects of exhaustive submaximal exercise on cardiovascular function during sleep.
J. Appl. Physiol.
58:
1909-1913,
1985
2.
Cybulski, G.
Influence of age on the immediate cardiovascular response to orthostatic manoeuvre.
Eur. J. Appl. Physiol.
73:
563-572,
1996.
3.
Du Quesnay, M. C.,
G. J. Stoute,
and
R. L. Hughson.
Cardiac output in exercise by impedance cardiography during breath holding and normal breathing.
J. Appl. Physiol.
62:
101-107,
1987
4.
Glen, S. K.,
H. L. Elliott,
J. L. Curzio,
K. R. Lees,
and
J. L. Reid.
White-coat hypertension as a cause of cardiovascular dysfunction.
Lancet
348:
654-657,
1996[Medline].
5.
Imhof, P. R.,
P. Muller,
and
R. Keller.
Pharmacological profiling of cardiovascular agents in healthy volunteers by means of non-invasive methods.
Methods Find. Exp. Clin. Pharmacol.
9:
829-832,
1987[Medline].
6.
Kubicek, W. G.,
J. N. Karnegis,
R. P. Patterson,
D. A. Witsoe,
and
R. H. Mattson.
Development and evaluation of an impedance cardiograph system.
Aerospace Med.
37:
1208-1212,
1966[Medline].
7.
Lund-Johansen, P.
Hemodynamic patterns in the natural history of borderline hypertension.
J. Cardiovasc. Pharmacol.
8:
8-14,
1986.
8.
Mancia, G.,
P. L. Gamba,
S. Omboni,
F. Paleari,
G. Parati,
R. Sega,
and
A. Zanchetti.
Ambulatory blood pressure monitoring.
J. Hypertens. Suppl.
14:
S61-S68,
1996[Medline].
9.
Manuck, S. B.,
M. F. Muldoon,
J. R. Kaplan,
M. R. Adams,
and
J. M. Polefrone.
Coronary artery atherosclerosis and cardiac responses to stress in cynomologus monkeys.
In: In Search of Coronary Prone Behavior: Beyond Type A, edited by A. W. Siegman,
and T. M. Dembroski. Hillsdale, NJ: Erlbaum, 1997, p. 207-227.
10.
Miyamoto, Y.,
J. Higuchi,
Y. Abe,
T. Hiura,
Y. Nakazono,
and
H. Mikami.
Dynamics of cardiac output and systolic time intervals in supine and upright exercise.
J. Appl. Physiol.
55:
1674-1681,
1983
11.
Panina, G.,
U. N. Khot,
E. Nunziata,
R. J. Cody,
and
P. F. Binkley.
Assessment of autonomic tone over a 24-hour period in patients with congestive heart failure: relation between mean heart rate and measures of heart rate variability.
Am. Heart J.
129:
748-753,
1995[Medline].
12.
Pasceri, V.,
G. A. Lanza,
G. Patti,
P. Pedrotti,
F. Crea,
and
A. Maseri.
Preconditioning by transient myocardial ischemia confers protection against ischemia-induced ventricular arrhythmias in variant angina.
Circulation
94:
1850-1856,
1996
13.
Sherwood, A.,
M. T. Allen,
J. Fahrenberg,
R. M. Kelsey,
W. R. Lovallo,
and
L. J. van Doornen.
Methodological guidelines for impedance cardiography.
Psychophysiology
27:
1-23,
1990[Medline].
14.
Sherwood, A.,
L. S. Carter, Jr.,
and
C. A. Murphy.
Cardiac output by impedance cardiography: two alternative methodologies compared with thermodilution.
Aviat. Space Environ. Med.
62:
116-122,
1991[Medline].
15.
Sherwood, A.,
and
J. R. Turner.
Hemodynamic responses during psychological stress: implications for studying disease processes.
Int. J. Behav. Med.
2:
193-218,
1995.
This article has been cited by other articles:
![]() |
J. A. McFetridge-Durdle, F. S. Routledge, T. Sampalli, R. Fox, H. Livingston, and B. Adams Hemodynamic Response to Postural Shift in Women with Multiple Chemical Sensitivities Biol Res Nurs, January 1, 2009; 10(3): 267 - 273. [Abstract] [PDF] |
||||
![]() |
T. G.M. Vrijkotte, L. J.P. van Doornen, and E. J. C. de Geus Overcommitment to Work Is Associated With Changes in Cardiac Sympathetic Regulation Psychosom Med, September 1, 2004; 66(5): 656 - 663. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Cybulski J Appl Physiol, April 1, 2000; 88(4): 1509 - 1510. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |