Vol. 85, Issue 3, 921-926, September 1998
Continuous measurement of tympanic temperature with a new
infrared method using an optical fiber
Manabu
Shibasaki1,
Narihiko
Kondo1,
Hirotaka
Tominaga1,
Ken
Aoki1,
Eiichi
Hasegawa2,
Yoshiyuki
Idota3, and
Toshimichi
Moriwaki3
1 Laboratory for Applied Human
Physiology, Faculty of Human Development, Kobe University, Kobe
657-8501; 2 Production Engineering
Laboratory, Shimadzu Company, Atsugi 243-0213; and
3 Faculty of Engineering, Kobe
University, Kobe 657-8501, Japan
 |
ABSTRACT |
The purpose of
this study was to investigate the utility of an infrared tympanic
thermometry by using an optical fiber for measuring tympanic
temperature (Tty). In the head
cooling and facial fanning tests during normothermia, right
Tty measured by this method
(infrared-Tty) and esophageal
temperature (Tes) were not
affected by decreased temple and forehead skin temperatures, suggesting
that the infrared sensor in this system measured the infrared radiation
from the tympanic membrane selectively. Eight male subjects took part
in passive-heat-stress and progressive-exercise tests. No significant
differences among infrared-Tty,
the left Tty measured by
thermistor (contact-Tty), and
Tes were observed at rest or at
the end of each experiment, and there was no significant difference in
the increase in these core temperatures from rest to the
end. Furthermore, there were no significant differences in
the core temperature threshold at the onset of sweating and slope (the
relationship of sweating rate vs.
infrared-Tty and vs.
contact-Tty). These
results suggest that this method makes it possible to measure
Tty accurately, continuously, and
more safely.
infrared sensor; device; core temperature; thermoregulation; humans
 |
INTRODUCTION |
CORE TEMPERATURE is one of the most important indexes
for clinical practice and thermoregulatory research in humans. It is necessary that the gold standard for measuring core temperature responds to rapid changes in the core temperature. Esophageal (Tes) and tympanic
(Tty) temperatures have been
used to estimate core temperature in clinical practice and
thermoregulatory research. Benzinger (1, 2) proposed that
Tty reflected the core temperature because of the proximity of the tympanic membrane to the carotid artery, where it flows into the hypothalamus. Although some researchers agree with his opinion (6, 7, 18, 24), others do not accept that
Tty is a true measurement of core
temperature because Tty is
affected by a decrease in skin temperature resulting from facial
cooling or fanning (5, 14, 15, 20). In a study directly measuring brain
temperature, Shiraki et al. (21) indicated that
Tty did not reflect the brain
temperature during hyperthermia. On the other hand, Mariak et al. (12)
suggested that the changes in brain temperature were better
approximated by Tty than by
Tes. This discrepancy may result
from the imperfections in the method used for measuring
Tty. Brinnel and Cabanac (6) and
Sato et al. (18) reported that when a probe was successfully placed on
the tympanic membrane, and the external canal was covered with cotton
or a thermal insulating material, then
Tty was not affected by ambient
temperature or by a decrease in skin temperature on the face or head.
Generally, the traditional methods for measuring
Tes and
Tty require that the thermometry
keeps in contact with the esophagus or tympanic membrane, respectively.
However, these methods have some risks (spread infection, injure the
esophagus or tympanic membrane, and cause subjects discomfort or pain).
Infrared tympanic thermometry has been developed to solve these
problems, as a method of measuring
Tty more safely and conveniently
(19). Terndrup noted that this method has been mentioned in many
publications since the late 1980s (see Ref. 23 for
review). However, the accuracy and reliability of existing
devices are not high enough to allow their use in clinical medicine and
thermoregulatory research (3, 10, 13, 17, 18, 22, 23). The reasons for
such inadequate performance in measuring
Tty are attributed to the characteristics of the infrared sensor and its size. An infrared sensor
measures the average temperature in its "field of vision." Existing devices may measure the infrared radiation from the auditory canal as well as the tympanic membrane. Furthermore, existing devices
are not suitable for continuous monitoring, although continuous information on the core temperature during experiments is valuable. We
have developed an infrared tympanic thermometry with an optical fiber
(optical fiber thermometry) to detect the limited amount of infrared
radiation from the tympanic membrane. A chalcogenide glass fiber was
used for the optical fiber, and this made it possible to detect the
temperature of the tympanic membrane selectively.
In this study, we investigated whether the optical fiber thermometry
accurately registered the temperature of the tympanic membrane during
different experimental maneuvers. Head cooling and facial fanning tests
were used to determine whether the temperature registered by the
optical fiber thermometry was affected by a decrease in skin
temperature. The passive-heat-stress and progressive-exercise tests
were used to compare Tty measured
by the optical fiber thermometry (infrared-Tty),
Tty measured with a thermistor
(contact-Tty), and
Tes measured by a
thermocouple. We also used the relationship between the
rate of sweating and the two core temperatures in these experiments as
an additional test of the utility of the optical fiber thermometry.
 |
MATERIALS AND METHODS |
Subjects.
Three healthy untrained male subjects participated in the head cooling
and facial fanning tests, and these three and five other male subjects
took part in the passive-heat-stress and progressive-exercise tests.
Their mean age, height, weight, and body surface area were 22.8 ± 0.9 (SD) yr, 1.70 ± 0.05 m, 64.5 ± 9.8 kg, and 1.74 ± 0.14 m2, respectively. Each subject was
informed of the nature of the procedures and the risks associated with
the experiments, and each signed an informed consent agreement when he
first visited our laboratory.
Procedures.
For the first group of tests, each subject sat on a chair in an
environmental chamber (SR-3000, Nagano Science, Osaka, Japan) maintained at an ambient temperature of 25°C and 50% relative humidity. After a 5-min baseline period, the subject's temple was
cooled by application of an ice pack or his forehead was cooled by air
blown from an electrical fan. Cooling lasted for 15-min followed by a
10-min recovery period. The two tests were performed on different days.
For the second group of tests, the subjects wore only shorts and
entered an environmental chamber maintained under the same conditions.
They remained seated for at least 1 h to reach equilibrium, during
which time the measuring devices were attached. After baseline values
were measured for 3 min, they then performed either the passive-heat-stress test or the progressive-exercise test. The passive-heat-stress test involved immersion of the subjects' lower legs in a hot water bath at 42°C for 60 min. The
progressive-exercise test involved pedaling on a cycle ergometer (model
M50, Combi) at a frequency set to 60 rpm. The subjects cycled at 50 W
for 1 min, and then the exercise intensity was increased by 1 W every 15 s for 30 min, for a total of 31 min. One subject was exhausted 27 min after the start of exercise, so only the first 27 min of data are
used for all of the subjects. These two tests were performed in a
random order, with at least 2 days between tests.
Measurement of Tty and
Tes.
In this study, two types of tympanic thermometry were used to measure
Tty. The first was an optical
fiber thermometry, consisting of an infrared sensor and an optical
fiber. The infrared sensor used was a pyroelectric infrared detector
(DLATGS, Shimadzu). This infrared sensor can detect
infrared energy independent of the wavelength of the radiation emitted
by an object (11), but the incident infrared radiation must be
interrupted at regular intervals. Therefore, this system is equipped
with a mechanical chopper, inserted between the infrared sensor and the
optical fiber (Fig. 1), which has two
blades that rotate at 120 Hz.

View larger version (21K):
[in this window]
[in a new window]
|
Fig. 1.
Schematic illustration of left external auditory canal and infrared
tympanic thermometry and optical fiber. Infrared rays from tympanic
membrane were transmitted to a pyroelectric infrared sensor along a
chalcogenide glass fiber. DC, direct current; A/D, analog to digital.
|
|
The optical fiber used was the most suitable for detecting human body
temperature. The core temperature in humans is normally between 36 and
40°C, and the wavelength of infrared radiation corresponding to
these temperatures is ~9.38-9.26 µm. A chalcogenide glass
fiber (NTEG, NOG) is the most suitable optical fiber for transmitting
infrared radiation at these wavelengths. This fiber has
been used in some machines: CO laser, thermal imaging, and so on. This
optical fiber transmits the infrared radiation from the tympanic
membrane better than do other optical fibers (16). The transmission
loss below 1 dB/m for the infrared wave with a wavelength of 9.35 µm
is matched to ~37°C. The fiber we used is 0.63 mm in diameter and
500 mm in length. NTEG makes it possible to detect the infrared
radiation from the tympanic membrane without the measurement being
influenced by the temperature of the surrounding ear canal wall and
without contacting the tympanic membrane.
The basic characteristics of the optical fiber thermometry, such as
stability (sensor and system drift) and linearity, were investigated by
using a blackbody temperature that was controlled by a Peltier element.
The optical fiber thermometry was calibrated with the blackbody
before each experiment.
To position the optical fiber in the extra-auditory canal, a
polyethylene sponge was glued to the fiber. This also insulated the
auditory canal from the ambient temperature (Fig. 1). Each subject compressed the polyethylene sponge and carefully inserted the
optical fiber into his right ear canal while he listened to the
spinning noise of the mechanical chopper that traveled through the
fiber and we monitored the temperature. The noise seems louder as the
fiber approaches the tympanic membrane, and the temperature of tympanic
membrane is maximal in the auditory canal. We used these parameters to
determine whether the optical fiber thermometry was properly positioned
to accurately measure Tty. The
temperature from the infrared sensor was sampled every 5 s and sent to
a personal computer (9801T, NEC).
The other method of thermometry was a thermistor sensor (Sensor
Technica), which measured the temperature of the contralateral tympanic
membrane
(contact-Tty). This
information was sent to the personal computer (9801NST, NEC) with a
data logger (K730, Techno Seven) every 4 s. The
thermistor probe was gently inserted into the left ear canal by the
test subject while we monitored the temperature. We used the sharp pain
felt by the subject and the scratching noise heard before and after
each experiment to determine that the sensor was in contact with the
tympanic membrane. The auditory canal was then filled with cotton wool
and taped to prevent movement of air inside the auditory canal.
Tes was measured by a
copper-constantan thermocouple with silicon coating at the tip. The
thermocouple was inserted through the nose to a distance equal to
one-quarter of each subject's height. The temperature was recorded
every second and was stored in the personal computer (9801RA, NEC) with
a data logger (HR2300, Yokogawa). Average values for
infrared-Tty,
contact-Tty, and
Tes were calculated for each
minute. The control and final temperatures were the average values for
the 3-min before the start and at the end of each experiment,
respectively. The change in the core temperatures
(
Tcore) was the difference
between the control and final values.
Measurement of skin temperatures, sweating rate, and heart rate.
During the tests of head cooling and facial fanning, the skin
temperature of the forehead and right temple was measured with the
copper-constantan thermocouple. These temperatures were sampled every
second. During the passive-heat-stress and progressive-exercise tests,
the local sweating rate and heart rate were measured. Sweating rate was
recorded continuously at two sites, the left side of the chest and the
forearm, by the ventilated capsule method. Dry nitrogen gas was pumped
through the capsules (chest: 8.54 cm2, forearm: 5.31 cm2) at a rate of 1.5 l/min. The
humidity of the nitrogen gas flowing out of the capsules was measured
with a capacitance hygrometer (HMP 133Y, Vaisala, Finland). The skin
temperatures and hygrometer output signals were stored in the same
system used for measuring Tes and
were calculated every minute. Heart rate was measured by using lead
V5 of an electrocardiogram.
Regression equations between sweating rate and the core temperature
were calculated for each subject during each test from the data sampled
every minute. Whenever the relationship showed a sigmoidal pattern, the
points representing the steepest portion of the curve were fitted by
regression analysis to obtain the slope.
Infrared-Tty,
contact-Tty, and
Tes thresholds for sweating were
the values for each core temperature after which sweating rate
increased above its equilibrium value.
Statistical analysis.
A one-way analysis of variance was performed to assess the difference
between calculated and measured values for the thermal parameters,
e.g., the control values, the final values,
Tcore, and the slope of
sweating rate vs. core temperature. Repeated two-way
analysis of variance with repeated measures was used to assess the
change in thermal parameters with time. Values of
infrared-Tty, Tes, and
contact-Tty sampled every 5 min
were used for the statistical analysis. In addition, Sheffé's
test was used to test treatment mean values among the parameters when
there were significant differences. Statistical significance was set at
P < 0.05 for all statistical tests.
 |
RESULTS |
The results of the stability and linearity tests of the optical fiber
thermometry, carried out with the blackbody, are shown in Fig.
2. The stability tests that used the
blackbody calibrated at 37°C were carried out at an ambient
temperature of 25°C for 60 min. There is no long-term drift in this
system, and the maximum difference between the temperature the optical
fiber thermometry measured and the temperature of the blackbody was
±0.2°C (Fig. 2A). The
temperature measured by the optical fiber thermometry shows a good
linear relationship with that of the blackbody (Fig. 2B).

View larger version (21K):
[in this window]
[in a new window]
|
Fig. 2.
Results of blackbody stability test at constant temperature
(A) and linear response between
infrared tympanic thermometry using an optical fiber (optical fiber
thermometry) and blackbody temperature
(B).
|
|
The head cooling and facial fanning tests were used to investigate
whether the optical fiber thermometry accurately measured only the
limited amount of infrared radiation from the tympanic membrane. The
skin temperatures on the right temple and forehead decreased remarkably
when the ice pack was pressed against the right temporal region (Fig.
3A) or
the electrical fan blew air in the subject's face (Fig.
3B).
Infrared-Tty and
Tes, however, are not influenced
by the decrease in these temperatures. The differences in
infrared-Tty between the control
and final values in the cooling and fanning tests are 0.0026 ± 0.0480 (SD) and 0.0411 ± 0.0731°C, respectively.

View larger version (24K):
[in this window]
[in a new window]
|
Fig. 3.
A: time course of tympanic temperature
measured by an infrared tympanic thermometry with an optical fiber
(infrared-Tty), esophageal
temperature (Tes), and
right-side temporal skin temperature
(Ttemple) during head cooling
test. Ice pack was positioned against right temporal area for 15-min
after a 5-min baseline period. Values of
infrared-Tty and
Tes were 36.88 ± 0.03 (SD) and
36.72 ± 0.06°C for 3-5 min before cooling and were 36.93 ± 0.05 and 36.71 ± 0.07°C for 18-20 min during
cooling. B: time course of
infrared-Tty, Tes, and
forehead skin temperature
(Tforehead) during facial
fanning test. Electrical fan with a 300-mm-diameter blade was placed
150 mm in front of the face. Facial skin was cooled for 15 min after a
5-min baseline period. Values of
infrared-Tty and
Tes were 36.75 ± 0.03 (SD) and
36.67 ± 0.05°C for 3-5 min before cooling and 36.75 ± 0.05 and 36.66 ± 0.06°C for 18-20 min during fanning.
|
|
No significant differences between
infrared-Tty,
contact-Tty, and
Tes are observed for the control
and final values and for
Tcore
in the passive-heat-stress and progressive-exercise experiments (Table
1). During both experiments,
there are no significant differences in the changes in
infrared-Tty,
contact-Tty, and
Tes over time (Fig.
4). In the progressive-exercise experiment,
there was a significant delay in the rise in
infrared-Tty and
contact-Tty compared with
Tes. There was no significant
difference in the onset of the elevation in
infrared-Tty,
contact-Tty, and
Tes in the passive-heat-stress
experiment (Table 1). The correlation between
infrared-Tty and
contact-Tty in both experiments is
statistically significant [correlation coefficient: 0.82 ± 0.10 (passive-heat-stress test) and 0.87 ± 0.07 (progressive-exercise test)].
View this table:
[in this window]
[in a new window]
|
Table 1.
Thermal parameters at rest and end periods of each test, increase in
core temperature from the rest period to the end period, onset time of
rising core temperature from the control value, slopes between
sweating rate and core temperature, and threshold for onset of sweating
during a passive-heat-stress and a progressive-exercise test
|
|

View larger version (32K):
[in this window]
[in a new window]
|
Fig. 4.
Comparison of core temperatures during passive-heat-stress
(A) and progressive-exercise
(B) tests. Time course of
infrared-Tty, tympanic temperature
measured with a thermistor
(contact-Tty), and
Tes during both experiments. One
subject became exhausted 27 min after the start of exercise, so data
plotted for all the subjects' data are for 27 min.
|
|
There are highly positive correlations, which range from 0.864 to
0.950, between sweating rate on the chest and forearm and the core
temperatures. There are no significant differences in the slopes of
sweating rate vs. infrared-Tty and
sweating rate vs. contact-Tty at
each site in both experiments. No significant differences in the
threshold values of infrared-Tty,
contact-Tty, and
Tes for sweating were observed for
either site.
 |
DISCUSSION |
Although many thermometries have been developed and used to measure
core temperature, use of these devices involves great risks, technical
difficulties, and controversies (4). Recently, a method of infrared
tympanic thermometry that did not involve contact with the tympanic
membrane was developed to measure
Tty safely and easily. The
accuracy of predecessors has been questioned (3, 10, 13, 17, 22). The
insufficient accuracy is caused by the fact that these devices may
measure the temperature of the auditory canal. To prevent this, we used
a chalcogenide glass fiber to detect infrared radiation solely from the
tympanic membrane (Fig. 1). Because the optical fiber thermometry
accurately measured the regulated temperature of a blackbody (Fig. 2),
it is believed that the optical fiber thermometry measurements are
stable and have a linear response, making the optical fiber suitable
for thermometry.
Because the tympanic membrane is near the point where the carotid
artery enters the hypothalamus,
Tty is believed to reflect the
brain and core temperatures (2). However, there were several problems
with methods of measuring Tty.
Some authors have reported that
Tty did not reflect the core
temperature, because it was influenced by changes in the ambient
temperature and by cooling of the head or face (4, 14, 15). Sato et al.
(18) reported that Tty fell by as
much as 0.6°C during head cooling by an ice pack when the sensor
was not in contact with the tympanic membrane. When the sensor was
carefully positioned and insulated,
Tty was not affected by the
decrease in head and face skin temperatures and was recommended as the
core temperature (6, 7, 18). Therefore, the head cooling and facial
fanning tests in normothermic subjects were used to determine whether
the optical fiber thermometry accurately measured
Tty. Figure 3 shows that
infrared-Tty was not affected by
the decreased skin temperatures during the head cooling or facial
fanning tests, when the same methods as those of Sato et al. were used.
The results of the head cooling and facial fanning suggested that the
tip of the optical fiber was close to the tympanic membrane, although
this system cannot determine the critical distance between the tympanic
membrane and the tip of the optical fiber. Calculating from the
diameter of the tympanic membrane (~1 mm) and the visual fields of
the optical fiber (~45°), the distance might be <2.5
mm. In the passive-heat-stress experiment, infrared-Tty,
contact-Tty and
Tes showed almost identical
responses (Fig. 4): no significant differences among
infrared-Tty,
contact-Tty, and
Tes are observed for the control
and final values,
Tcore, and
the onset of the rise in core temperatures (Table 1). In the
progressive-exercise experiment, the only difference among infrared-Tty,
contact-Tty, and
Tes was in the onset of the rise in core temperature. However, there were no significant differences in
the change in infrared-Tty,
contact-Tty, and
Tes over time in both experiments
(Fig. 4). Earlier studies (8, 9, 15) also reported
Tes increased before
Tty during exercise. In both experiments, infrared-Tty and
contact-Tty showed almost
identical responses, and the responses between
Tty
(infrared-Tty and
contact-Tty) and
Tes were similar to earlier
studies (8, 9, 15, 18). These results suggest that the temperature
measured by the optical fiber thermometry is the Tty and
that the optical fiber thermometry can be recommended as a device for
measuring Tty.
Because our system uses the infrared sensor, it was necessary to
determine whether the polyethylene sponge, attached to the optical
fiber and used to insulate the auditory canal, held the fiber close to
the tympanic membrane throughout exercise. Two subjects had a low
correlation between infrared-Tty
and contact-Tty during exercise.
These subjects reported that they felt that the fiber was slipping out
of their auditory canals. It is necessary to improve the method of
attaching the fiber, perhaps by fashioning an attachment similar to a
hearing aid. On the other hand, this system accurately measured
Tty, even in three subjects whose
heart rate reached 200 beats/min at the end of the exercise, suggesting that the optical fiber thermometry could accurately measure the Tty during intense exercise when
the fiber was positioned in external canal.
Thermoregulatory sweating responses are generally assessed by two
parameters, the core temperature threshold for the onset of sweating
and the slope of the relationship between core temperature and the rate
of sweating (i.e., sensitivity). We used these two parameters to
investigate whether infrared-Tty
could be used to evaluate the thermoregulatory sweating responses as
well as contact-Tty and
Tes. In this study, there was a
remarkable positive relationship between the core temperatures measured
by the three methods and sweating rate on the chest and forearm. In
both experiments, there were no significant differences in the
sensitivity of measurement determined by sweating rate vs.
infrared-Tty and sweating rate vs.
contact-Tty at each site.
Furthermore, there were no significant differences in
infrared-Tty,
contact-Tty, and
Tes thresholds for sweating in
both experiments (Table 1), although the increase in
infrared-Tty and
contact-Tty lagged behind the rise
in Tes during exercise (Fig. 4).
Consequently, these results also suggest that
infrared-Tty accurately reflects
the temperature that determines the thermoregulatory sweating response
and can be used in thermoregulatory research and clinical medicine.
In conclusion, the optical fiber thermometry performs extremely well,
and it accurately measures the Tty. Head cooling and facial
fanning tests proved that the optical fiber thermometry accurately
measured only the limited amount of infrared radiation from the
tympanic membrane. The changes in
infrared-Tty during the
passive-heat-stress and progressive-exercise experiments were the same
as those in contact-Tty.
Furthermore, the changes in infrared-Tty and
contact-Tty were similar to those
in Tes during both experiments.
There were also no significant differences between infrared-Tty and
contact-Tty and the changes in the
two parameters for monitoring sweating responses (threshold temperature
and sensitivity). Therefore, we can recommend the optical fiber
thermometry as a device for continuously measuring Tty.
 |
ACKNOWLEDGEMENTS |
The authors thank the test subjects for their wonderful cooperation
in our research.
 |
FOOTNOTES |
Address for reprint requests: M. Shibasaki, c/o N. Kondo, Laboratory
for Applied Human Physiology, Faculty of Human Development, Kobe
University, 3-11 Tsurukabuto, Nada-Ku, Kobe 657-8501, Japan (E-mail:
mshiba{at}kobe-u.ac.jp).
Received 1 December 1997; accepted in final form 28 April 1998.
 |
REFERENCES |
1.
Benzinger, T. H.
On physical heat regulation and the sense of temperature in man.
Proc. Natl. Acad. Sci. USA
45:
645-659,
1959[Free Full Text].
2.
Benzinger, T. H.
Heat regulation: homeostasis of central temperature in man.
Physiol. Rev.
49:
671-759,
1969[Free Full Text].
3.
Betta, V.,
F. Cascetta,
and
D. Sepe.
An assessment of infrared tympanic thermometers for body temperature measurement.
Physiol. Meas.
18:
215-225,
1997[Medline].
4.
Brengelmann, G. L.
Dilemma of Body Temperature Measurement. Springfield, IL: Thomas, 1987.
5.
Brengelmann, G. L.
Specialized brain cooling in humans?
FASEB J.
7:
1148-1152,
1993[Abstract].
6.
Brinnel, H.,
and
M. Cabanac.
Tympanic temperature is a core temperature in humans.
J. Therm. Biol.
14:
47-53,
1989.
7.
Cabanac, M.
Selective brain cooling in humans: "fancy" or fact?
FASEB J.
7:
1143-1147,
1993[Abstract].
8.
Cabanac, M.,
and
M. Caputa.
Open loop increase in trunk temperature produced by face cooling in working humans.
J. Physiol. (Lond.)
289:
163-174,
1979[Abstract/Free Full Text].
9.
Cabanac, M.,
M. Germain,
and
H. Brinnel.
Tympanic temperatures during hemiface cooling.
Eur. J. Appl. Physiol.
56:
534-539,
1987.
10.
Hansen, R. D.,
T. S. Olds,
D. A. Richards,
C. R. Richards,
and
B. Leelarthaepin.
Infrared thermometry in the diagnosis and treatment of heat exhaustion.
Int. J. Sports Med.
17:
66-70,
1996[Medline].
11.
Kita, J.,
H. Kishihara,
and
J. Kobayashi.
Development of a new pyroelectric infrared detector utilizing a single crystal of DLATGS.
Shimadzu Rev.
51:
205-210,
1994. (In Japanese)
12.
Mariak, Z.,
J. Lewko,
J. Luczaj,
B. Polocki,
and
M. D. White.
The relationship between directly measured human cerebral and tympanic temperatures during changes in brain temperatures.
Eur. J. Appl. Physiol.
69:
545-549,
1994.
13.
Matsumoto, T.,
M. Yamauchi,
A. Taimura,
E. Kaneda,
K. Tsuchiya,
M. Shimazu,
Y. Cao,
Z.-W. Luo,
J.-B. Lee,
N. Ohwatari,
K. Otomasu,
and
M. Kosaka.
Normal tympanic temperature measured with the infrared tympanic thermometer (Quick Thermo).
Jpn. J. Biometeor.
33:
123-130,
1996.
14.
McCaffrey, T. V.,
R. D. McCook,
and
R. D. Worster.
Effect of head skin temperature on tympanic and oral temperature in man.
J. Appl. Physiol.
39:
114-118,
1975[Abstract/Free Full Text].
15.
Nielsen, B.
Natural cooling of the brain during outdoor bicycling?
Pflügers Arch.
411:
456-461,
1988[Medline].
16.
Nishii, J.,
S. Morimoto,
I. Inagawa,
R. Iizuka,
T. Yamashita,
and
T. Yamagishi.
Recent advances and trends in chalcogenide glass fiber technology: a review.
J. Noncryst. Solids
140:
199-208,
1992.
17.
Rhoads, F. A.,
and
J. Grandner.
Assessment of aural infrared sensor for body temperature measurement in children.
Clin. Pediatr. (Phila.)
29:
112-115,
1990.
18.
Sato, K. T.,
N. L. Kane,
G. Soos,
C. V. Gisolfi,
N. Kondo,
and
K. Sato.
Reexamination of tympanic membrane temperature as a core temperature.
J. Appl. Physiol.
80:
1233-1239,
1996[Abstract/Free Full Text].
19.
Shinozaki, T.,
R. Deane,
and
F. Perkins.
Infrared tympanic thermometer: evaluation of a new clinical thermometer.
Crit. Care Med.
16:
148-150,
1988[Medline].
20.
Shiraki, K.,
N. Konda,
and
S. Sagawa.
Esophageal and tympanic temperature responses to core blood temperature changes during hyperthermia.
J. Appl. Physiol.
61:
98-102,
1986[Abstract/Free Full Text].
21.
Shiraki, K.,
S. Sagawa,
F. Tajima,
A. Yokota,
M. Hashimoto,
and
G. L. Brengelmann.
Independence of brain and tympanic temperatures in an unanesthetized human.
J. Appl. Physiol.
65:
482-486,
1988[Abstract/Free Full Text].
22.
Smith, D. L.,
and
P. C. Fehling.
Reliability of infrared-tympanic thermometry (ITT).
Aviat. Space Environ. Med.
67:
272-274,
1996[Medline].
23.
Terndrup, T. E.
An appraisal of temperature assessment by infrared emission detection tympanic thermometry.
Ann. Emerg. Med.
21:
1483-1492,
1992[Medline].
24.
Walpoth, B. H.,
J. Galdikas,
F. Leupi,
W. Muehlemann,
P. Schlaepfer,
and
U. Althaus.
Assessment of hypothermia with a new tympanic thermometer.
J. Clin. Monit.
10:
91-96,
1994[Medline].
J APPL PHYSIOL 85(3):921-926
8570-7587/98 $5.00
Copyright © 1998 the American Physiological Society