|
|
||||||||
1 Department of Respiratory
Medicine, The Rigshospital, DK-2200 Copenhagen N; Departments of
2 Medical Physiology and
3 Biostatistics, In obstructive
lung disease the annual change in lung function is usually estimated
from serial measurements of forced expiratory volume in 1 s
(FEV1). Frequent measurements in
each patient may not improve this estimate because data are not
statistically independent; i.e., the measurements are autocorrelated.
The purpose of this study was to describe the correlation structure in
time series of FEV1 measurements.
Nineteen patients with severe
spirometry; periodicity; Fourier analysis; power spectrum
THE INTRODUCTION IN RECENT YEARS of small handheld
spirometers with timer and data-storage capability has made it possible to monitor pulmonary function daily by patient-administered serial spirometry (15). The accumulation of such information has made it
relevant to describe in more detail the temporal structure of repeat
spirometric tests. Previous studies from other organ systems,
especially the cardiovascular system, have revealed the presence of
significant fluctuations in many physiological variables. Both the
heart rate and the arterial blood pressure show considerable variability even under conditions in which the animal or the individual is at rest (10, 16). The fluctuations appear to be due to the
intrinsic, complex dynamics of the regulatory systems that govern a
given physiological variable and thus cannot be explained simply as the
result of measurement errors.
It is well known that when observations are made far apart in time they
will, in general, not be correlated; i.e., they are independent.
However, as the sampling frequency increases, correlations will be
present between the consecutive measurements, and they can no longer be
treated as independent. Such correlations between repeated measurements
made on the same subject are technically known as autocorrelations. The
surprising result that has emerged from many studies in the
cardiovascular system is that the time until measurements can be
regarded as independent can be very long (on the order of many hours)
(10, 16). The consequence of such long-range correlations is a
considerable persistence in the pattern of the fluctuations. Thus, if a
given measurement is found to be above its mean value at a given time,
there will be a high probability that the same will be the case for
measurements made during the following hours. The mechanisms underlying
the long-range correlations seen in many physiological variables are presently not well understood (10).
Besides being of fundamental physiological interest, the
autocorrelation structure of longitudinal data also has implications for statistical data analysis. Although the autocorrelation structure may be less critical when parameters such as the slope of decline in
pulmonary function are estimated, it has substantial impact on
parameters like the SE of the slope, which will be underestimated if
the autocorrelation is ignored and the observations are treated as
independent (18, 20).
In this study we have analyzed time series consisting of
patient-administered serial spirometry data from a controlled clinical trial in which emphysematous patients were followed with daily spirometric measurements for several years. Furthermore, two coauthors of this study (A. Dirksen and A. Kok-Jensen) performed
daily self-administered spirometry for 14-24 mo.
Twenty-two patients with severe
For home spirometry the Vitalograph R model, a direct-writing, 7-liter,
dry-wedge spirometer (Vitalograph, Buckingham, UK) was used during the
first half of the study (1991-1993) (15), and during the last half
(1993-1995) we used a handheld turbine spirometer with timer and
data-storage capability (DiaryCard, MicroMedical, Rochester, UK) (6).
The above-mentioned coauthors used the handheld turbine spirometer
only. As previously described (6, 15), the spirometers were calibrated
every 4 wk when the patients came for infusions.
At inclusion, patients were carefully instructed in spirometry for ~1
h, and they received written information on how to perform spirometry
at home. Spirometry was to be performed every morning and evening
throughout the study. A measuring sequence included at least three
maximal forced vital capacity maneuvers. The individual spirometry
results were presented in a blind fashion to ensure that the
participants' expiratory efforts and number of trials would not be
affected by previous results. To obtain blinding, Vitalograph charts
without preprinted grids were produced, and the software of the turbine
spirometer was modified to suppress the continuous display of results.
Every 4 wk when the patients came for infusions, results of the last
4-wk spirometries were displayed graphically and presented to the
participant to promote motivation. Instructions in spirometry were
repeated as needed, and the importance of a maximal inspiration at the
start of the test was emphasized regularly. Quality control was
performed by visual inspection of time-volume curves (for the
Vitalograph) or flow-volume curves (for the DiaryCard), and only
measurements fulfilling strict criteria were accepted for further
analysis; this meant that the highest
FEV1 of three technically
satisfactory attempts and the chosen
FEV1 did not exceed the next
highest measurement by >5% or 100 ml, whichever was greater
(17). For the majority of the subjects, the number of
discarded measurements was <5% of the total.
At inclusion and every 3 mo throughout the study, patients visited the
respiratory laboratory in the morning. Pulmonary function testing was
performed according to American Thoracic Society recommendations (1). A
constant-volume body plethysmograph and a dry rolling seal spirometer
(SensorMedics 2800 and 2450, Anaheim, CA) were applied. Fifteen minutes
after bronchodilatation (5 mg nebulized terbutaline), with the patient
seated and wearing a noseclip, a slow vital capacity maneuver was
obtained first, followed by a forced vital capacity maneuver, from
which the maximal flow-volume loop and
FEV1 were derived. Static lung
volumes (i.e., total lung capacity, functional residual capacity, and
residual volume) were measured by both the closed-circuit
helium-dilution technique and plethysmography, in which lung volumes
and airway resistance were determined during panting, with a frequency
of <1/s. The carbon monoxide diffusion constant
(KCO) was
measured by the single-breath technique, and because the hemoglobin was
always within normal limits, the values were not corrected for
hemoglobin. The diffusion capacity
(DLCO)
was calculated as the product of
KCO and the
alveolar volume. The latter was obtained from the dilution of helium
during the single-breath maneuver. All measurements were performed in triplicate except for helium dilution. Gas volumes are reported with
BTPS corrections, and results are
expressed in absolute values and as percentages of predicted values
that were calculated according to European reference equations (4, 17).
Data analysis.
For estimation of autocorrelations, the
FEV1 measurements were detrended
by ordinary linear regression for each subject separately. The
standardized residuals were used for calculation of the correlations (Pearson) between measurements made at a given interval, ranging from 1 day to 30 mo. To minimize the effect of the circadian variation in
FEV1, autocorrelations were
calculated separately for the morning and evening measurements. This
resulted in two autocorrelation functions for each subject, which were
then averaged so as to yield one autocorrelation function. Calculation
of the autocorrelations in a given individual was terminated when the
number of paired observations fell below 50 for either the morning or
evening measurements. The above-mentioned procedure was chosen because
it allowed for the presence of missing data while making use of all
available data. The normalized power spectrum of the fluctuations in
the FEV1 was obtained as the
Fourier transform of the mean of the empirical autocorrelation
functions by use of the fast Fourier algorithm.
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
Appendix
References
1-antitrypsin deficiency
(phenotype PiZ) and moderate to severe emphysema and two subjects with
normal lungs were followed for several years with daily
self-administered spirometry. FEV1
measurements fulfilling standard criteria were detrended, and the
autocorrelation profile and the power spectrum were calculated. On
average the subjects were followed for >3 yr and performed >1,000
acceptable spirometries. The autocorrelation of
FEV1 measurements in the
emphysematous patients was ~0.35 for short intervals and decreased
almost exponentially with a half time of 38 days. Between 3 and 4 mo,
the autocorrelation function became negative. It reached a minimum of
0.1 at ~8 mo and then increased toward zero over the following
12 mo. The autocorrelation function in the two normal subjects showed a
similar pattern, but with a faster decay toward zero. In the patients,
the power spectrum had a peak at 1 cycle/wk and showed a
1/f pattern, where f is frequency, with a slope of
0.88 at lower frequencies. We conclude that serial spirometric
measurements show long-range correlations. The practical implication is
that FEV1 need not be measured
more often than once every 3 mo in studies of the long-term trends in
lung function.
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
Appendix
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
Appendix
References
1-antitrypsin deficiency
(phenotype PiZ, verified by isoelectric focusing) (9) and moderate to
severe emphysema [30% predicted < forced expiratory volume in
1 s (FEV1) < 80%
predicted] participated in a randomized trial of augmentation
therapy with
1-antitrypsin (250 mg/kg) every 4 wk vs. a placebo. All refrained from smoking at least 6 mo before entering the study, and urinary cotinine was checked every 4 wk during the trial. All patients gave their informed consent to participate in the study, which was approved by the local medical ethics committee of the city of Copenhagen. Of the 22 patients, 3 were excluded because of poor cooperation and too few
measurements (<500). Furthermore, the two coauthors of this study
mentioned above performed daily home spirometry for a period of
14-24 mo.
where
k and
a are constants and
i is the interval between measurements
in days (see APPENDIX for details).
The parameters were estimated by the method of least squares, using a
quasi-Newtonian algorithm.
(1)
| |
RESULTS |
|---|
|
|
|---|
Table 1 shows that the 19 patients with
1-antitrypsin deficiency had
moderate to severe emphysema. On average they were followed for >3 yr
and performed >1,000 acceptable spirometries.
|
Raw data (FEV1) from two of the emphysematous patients and one normal subject are delineated in Fig. 1. In absolute terms the diurnal variations are of approximately the same magnitude in all three individuals, but seasonal fluctuations were larger in the subject with normal lung function. Gaps in the tracings indicate missing data, e.g., measurements not taken during holidays. The subject with normal lungs was appendectomized in September 1994.
|
The mean autocorrelation structure for the emphysematous patients is
shown in Fig.
2A. The
best fit for the model given in Eq. 1
was achieved with the values k = 0.3556 and a = 0.9910874 for the two
constants. With these values for the constants, the model explained
74% of the variance in the data
(R2 = 0.74). The
correlation between measurements taken on consecutive days was found to
be ~0.35, with a slow, approximately exponential decline (the half
time was ~38 days). Between 3 and 4 mo, the autocorrelation
function became negative. It reached a minimum of
0.1 at
~8 mo and then increased toward zero over the following 12 mo. The
mean autocorrelation profile shows moderate peaks at 12 and 24 mo,
which were not accounted for by the model given in Eq. 1.
|
The autocorrelation profiles of the two subjects with normal lungs (Fig. 2B) showed a more rapid decline during the first months. Seasonal variations were large in one of these subjects (raw data for this subject are delineated in Fig. 1), and weekly fluctuations were pronounced in both of them. Although weekly variations were less obvious in the time series from the emphysematous patients, power spectral analysis reveals a distinct peak at 1 cycle/wk even in this group (Fig. 3).
|
| |
DISCUSSION |
|---|
|
|
|---|
The results of the present study show that there are substantial fluctuations in the FEV1 when measured over several years. Initially, the autocorrelation profile of the fluctuations (Fig. 2) declines almost exponentially, with a half time of ~1 mo. After ~100 days, the autocorrelation function becomes negative, reaching a minimum at ~8 mo. Thereafter, it approaches zero over the following year. The implications of this finding are that fluctuations in lung function have a considerable temporal persistence and that they are dominated by fluctuations with a very low frequency. When plotted as the power spectrum against frequency (f), the relationship between the frequency and the contribution of each frequency (V) to the total variance could be expressed approximately as V = 1/fx, or "one over f spectrum," where x is the slope of the line in the log-log plot, and x is ~0.88 (Fig. 3). Thus in the patients the fluctuations were dominated by the slow frequencies (f > 1 cycle/wk).
Presently, we have no explanation for the irregular fluctuations in FEV1 observed in this study. Longitudinal variations in pulmonary function may be due to either biological variation or measurement error (21). Because it was the chief purpose of the present study to explore the long-range fluctuations in pulmonary function tests, we only determined the FEV1 twice daily. It is well known that pulmonary function undergoes significant diurnal variation (2, 11), and, with a sampling frequency of two measurements per day, the circadian variability of FEV1 may have been grossly underestimated in the present study (5). To minimize any possible effects from diurnal variation, all measurements were made at the same time of day. This is in accordance with the recommendations of both the American Thoracic Society and the European Respiratory Society (1, 17). In addition, the autocorrelation profiles for morning and evening values were calculated independently. Thus it appears unlikely that the observed variability can be explained by the diurnal variation in lung function.
Diurnal variation is usually explained by a complex interaction of several coincident circadian rhythms (2). Possible biological mechanisms behind the variability we observed are numerous. Environmental exposures may influence pulmonary function. The weather is a universal, chaotic phenomenon, and thus it is obvious, as is often stated by patients, to suspect the weather as a possible source of variations in pulmonary function. Denmark is a small, flat country without mountains, and usually the weather is quite similar all over the country. However, in the present study, fluctuations in the patients were totally out of phase. The between-patient correlations were insignificant. Thus the weather was probably not a major source of variation. Another environmental exposure that may be considered is air pollution (7, 8). However, in Denmark the predominant western wind keeps outdoor air pollution low, making it an unlikely source of significant variability in pulmonary function.
A potential source of measurement error was the spirometers, but this was reduced to a minimum by careful quality assurance. The instruments were calibrated every 4 wk throughout the study. The variability in performance (coefficient of variation) was within 1%, and the calibration data showed no autocorrelation (6, 15). Another important source of measurement error is lack of comprehension by or cooperation of the participants. The spirometric maneuver requires a maximum physical effort and coordinated movement by a subject trained to perform it (14). To ensure that the participants continued to follow measurement procedures, the instructions in spirometry were repeated regularly, and the importance of a maximal inspiration at the start of the test was especially reemphasized (3).
As already mentioned above, one of the most striking features of the power spectrum was its 1/f pattern. The 1/f spectral pattern occurs widely in nature and is found in both animate and inanimate systems (13, 16, 19). For example, previous studies have shown 1/f spectral patterns in arterial blood pressure and heart rate (13, 16). To the extent that it has been possible to draw generalizations about the meaning of 1/f spectra, it appears that the spectra occur in systems that are spatially distributed and loosely coupled. The value of the FEV1 at any time is the result of all the processes that occur throughout the bronchial system. Examples of such processes could be environmental influences, infectious and/or inflammatory processes, or nervous stimulation of bronchial smooth muscle. Each particular process will have a characteristic time scale, and the finding of a 1/f pattern suggests that all the processes contribute to the variability, and that no single mechanism dominates. If the system was dominated by the action of a single mechanism, a single spectral peak would emerge because the system would be operating in a single mode.
Although we only followed the FEV1 in two normal subjects, it is striking that the autocorrelation function appears to decline toward zero faster in the normal subjects than in patients. The implication of this observation is that the fluctuations in FEV1 are more random in the normal subjects. The mechanism behind this difference is unknown, but a related phenomenon has been observed in patients with heart disease. Patients with heart disease tend to have a reduced heart rate variability, and it has been well documented that a reduced variability is an independent risk factor for sudden cardiac death (12, 22). Further studies are needed to determine whether the apparent change in the dynamics of FEV1 by itself has any clinical implications.
FEV1 is the most commonly used pulmonary function measurement in longitudinal studies of obstructive lung disease, and our findings have important implications for the efficient use of serial measurements for estimating the decline in lung function in individual subjects with either normal lungs or relatively stable lung function. When a phenomenon with high autocorrelation is being investigated, frequent sampling will not be worthwhile because measurements at short intervals will add little extra information.
The decline in lung function is often expressed as the slope of the regression line between FEV1 and the time of measurement. Standard formulas and most statistical software packages assume that the data points are randomly and independently distributed around the regression line. Basically, this assumption implies that if a measurement at one time is above the estimated regression line, there is a 50% chance that the next measurement will also fall above the regression line. This will not be the case for daily measurements of the FEV1, as demonstrated by the present results. Because of the high degree of autocorrelation between measurements, the probability that the next measurements will also fall above the regression line will be much greater than 50%. Indeed, the measurements need to be spaced more than 3 mo apart before this probability reaches 50%. This finding has significant implications for the precision of the estimated slopes. To quantify this issue, we investigated the precision of the estimated slope of decline in lung function as a function of sampling frequency.
The slope can essentially be estimated in two ways (see APPENDIX). Either the slope can be calculated as an ordinary least squares estimate (EOLS), i.e., ignoring the autocorrelation structure (assuming independence) or, more correctly, the slope can be obtained by using an appropriately estimated autocorrelation structure, derived as a maximum likelihood estimate (EML) in this model. As imaginary sampling schemes, we considered an observation period of 365 days, with sampling intervals of 1, 2, 3, 4, 5, 7, 14, 30, 60, 90, 120, 183, and 365 days, respectively. For each of these designs, we calculated the following three SEs: 1) the standard error of EML under the assumption of autocorrelation, i.e., by using Eq. A1 (see APPENDIX); 2) the standard error of EOLS under the assumption of autocorrelation, i.e., by using Eq. A2 (see APPENDIX); and 3) the standard error of EOLS under the assumption of independence, i.e., by using Eq. A3 (see APPENDIX).
In Fig. 4, the resulting SEs of EML (assuming the above-mentioned autocorrelation structure) are seen as the middle curve (solid line). It is obvious from Fig. 4 that the real gain in precision by frequent sampling is modest (reducing the SE by 1/3 from worst to best situation, which corresponds to 183 times more measurements). Even with an infinite number of measurements, it will not be possible within an observation period of 1 yr to bring the SE below ±80 ml/year, a value which is quite large compared with a predicted annual decline in FEV1 in normal subjects of ~30 ml/yr. However, provided we falsely assume the measurements to be independent (which is implicit in most statistical software packages), the resulting SEs of EOLS are shown as the bottom curve (dashed line) in Fig. 4. The estimated SEs are smaller than those of EML, and erroneously, it seems as if the gain from frequent sampling was large (a decrease of ~10 times). The top curve (dotted line) in Fig. 4 shows the SEs of EOLS, calculated in the appropriate correlation structure, and it can be seen that this curve is close to the curve for EML, indicating that the extra efficiency obtained by using the correct correlation structure for calculating the slope estimate is small. Thus, for estimation of the decline in lung function, EOLS may be used, as long as its true precision is calculated with due regard to the autocorrelation structure.
|
The practical implications for estimates of long-term trends in lung function are as follows: 1) it is not cost effective to measure FEV1 more often than once every 3 mo; and 2) provided measurements are performed more frequently than once every 3 mo, the autocorrelation structure should be taken into account when the SE of the decline in lung function in individual subjects is estimated.
To avoid any misconception, it should be emphasized that this recommendation of less frequent measurements applies to long-term trends in lung function only. The recommendation does not apply to monitoring of short-term changes in lung function, such as exacerbations of chronic obstructive pulmonary disease or asthmatic variations.
| |
APPENDIX |
|---|
|
|
|---|
In the investigation of the effect of sampling frequency on SE of the estimated lung function decline, we assumed that the FEV1 measurements, Yi, could be described by the linear model
|
|
denotes the parameter vector
(µ,
)', where µ is the intercept (level at the start
of the investigation period) and
is the slope of the regression
line (usually negative), and X is the
design matrix, reflecting the sampling frequency. Thus
X is an
N × 2 matrix, the first column
being only ones (corresponding to the initial level) and the second
column indicating N times of
measurements,
ti. In the
designs chosen, N varies between 2 and
365.
The
i notations in the model
formula above denote the random fluctuations of lung function around
the linear decline. We let
denote
the covariance matrix for these "errors," i.e., a matrix of
dimension varying from 2 × 2 to 365 × 365. In traditional
linear regression, this matrix is assumed to be diagonal, i.e.,
=
2I, where
I is the identity matrix (a matrix
with 1's in the diagonal and 0's elsewhere). This corresponds to
independence between observations taken at different times. The results
of the present study show that this is not the case for daily
FEV1 measurements, because an
appreciable amount of autocorrelation is present.
As an example of a reasonable covariance structure (reflecting the slowly decaying autocorrelation of moderate size), we chose for descriptive purposes a Laguerre function of the form
|
ij denotes the elements of
,
ti and
tj denote
observational times, and k and
a are constants chosen to make the
correlation pattern agree approximately with the one observed. The
Laguerre function was chosen because of its built-in exponential term.
The correlation
between the measurements taken 1 day apart can be
calculated as
|
= 0.354, and we find the correlation structure as shown in Fig. 1.
The empirical correlations are also shown in Fig. 1.
The parameter
in the regression
model can be estimated by either the maximum likelihood method (ML),
which takes the correlation structure,
, into account
|
|
|
(A1) |
|
(A2) |
is
determined under the assumption of independence between measurements,
the expression for its variance, Eq. A2, does not assume independence.
However, by ignoring the autocorrelation structure and using the
ordinary least square estimate
the traditionally quoted
variance estimate similarly assumes independence (
=
2
I), and therefore simplifies to
|
2 is estimated
(s2) from the actual observations,
Y, by
|
|
is the expected value, and tr denotes the
trace of a matrix. Note that
in
the equation above represents the "true" covariance structure of
the underlying stochastic process. The total expression for the
traditionally quoted variance of the conventional least square
estimator therefore becomes
|
(A3) |
| |
ACKNOWLEDGEMENTS |
|---|
The authors thank Dr. James K. Stoller, the Cleveland Clinic Foundation, for helpful comments on the manuscript.
| |
FOOTNOTES |
|---|
This study was supported by the National Danish Research Council for Public Health, the National Union for the Fight Against Pulmonary Diseases, the Danish Heart Association, and Astra Danmark A/S.
Address for reprint requests: A. Dirksen, The Respiratory Medicine Section ML 7721, The RHIMA Center, The Rigshospital, Tagensvej 20, DK-2200 Copenhagen N, Denmark (E-mail: ADi{at}RH.DK).
Received 28 August 1997; accepted in final form 27 February 1998.
| |
REFERENCES |
|---|
|
|
|---|
1.
American Thoracic Society.
Standardization of spirometry. 1994 Update.
Am. J. Respir. Crit. Care Med.
152:
1107-1136,
1995[Medline].
2.
Barnes, P. J. Circadian variation in airway
function. Am. J. Med. 79, Suppl. 6A: 5-9, 1985.
3.
Coates, A. L.,
K. J. Desmond,
D. Demizio,
and
P. D. Allen.
Sources of variation in FEV1.
Am. J. Respir. Crit. Care Med.
149:
439-443,
1994[Abstract].
4.
Cotes, J. E., D. J. Chinn, P. H. Quanjer, J. Roca, and J.-C. Yernault. Standardized lung function
testing. Standardization of the measurement of transfer factor
(diffusing capacity). Eur. Respir.
J. 6, Suppl. 16:
41-52, 1993.
5.
D'Alonzo, G. E.,
V. W. Steinijans,
and
A. Keller.
Measurements of morning and evening airflow grossly underestimate the circadian variability of FEV1 and peak expiratory flow rate in asthma.
Am. J. Respir. Crit. Care Med.
152:
1097-1099,
1995[Abstract].
6.
Dirksen, A.,
F. Madsen,
O. F. Pedersen,
A. M. Vedel,
and
A. Kok-Jensen.
Long term performance of a hand held spirometer.
Thorax
51:
973-976,
1996[Abstract].
7.
Dockery, D. W., and B. Brunekreef. Longitudinal
studies of air pollution effects on lung function.
Am. J. Respir. Crit. Care
Med. 154, Suppl.:
S250-S256, 1996.
8.
Dockery, D. W.,
and
C. A. Pope.
Acute respiratory effects of particulate air pollution.
Annu. Rev. Public Health
15:
107-132,
1994[Medline].
9.
Fagerhol, M. K.,
and
D. W. Cox.
The Pi polymorphism. Genetic, biochemical and clinical aspects of human alpha-1-antitrypsin.
Adv. Hum. Genet.
11:
1-62,
1981[Medline].
10.
Goldberger, A. L.,
D. R. Rigney,
and
B. J. West.
Chaos and fractals in human physiology.
Sci. Am.
262:
42-49,
1990[Medline].
11.
Hetzel, M. R.
The pulmonary clock.
Thorax
36:
481-486,
1981[Medline].
12.
Kleiger, R. E.,
J. P. Miller,
J. T. Bigger, Jr.,
and
A. J. Moss.
Decreased heart rate variability and its association with increased mortality after acute myocardial infarction.
Am. J. Cardiol.
59:
256-262,
1987[Medline].
13.
Kobayashi, M.,
and
T. Musha.
1/f Fluctuation of heart beat period.
IEEE Trans. Biomed. Eng.
29:
456-457,
1982[Medline].
14.
Krowka, K. J.,
P. L. Enright,
J. R. Rodarte, Jr.,
and
R. E. Hyatt.
Effect of effort on measurement of forced expiratory volume in one second.
Am. Rev. Respir. Dis.
136:
829-833,
1987[Medline].
15.
Madsen, F.,
C. S. Ulrik,
A. Dirksen,
K. K. Hansen,
N. H. Nielsen,
L. Frolund,
K. Viskum,
and
A. Kok-Jensen.
Patient-administered sequential spirometry in healthy volunteers, and patients with
1-antitrypsin deficiency.
Respir. Med.
90:
131-138,
1996[Medline].
16.
Marsh, D. J.,
J. L. Osborn,
and
A. W. Cowley, Jr.
1/f Fluctuations in arterial pressure, and the regulation of renal blood flow in dogs.
Am. J. Physiol.
258 (Renal Physiol. 27):
F1394-F1400,
1990
17.
Quanjer, P. H., G. J. Tammeling, J. E. Cotes, O. F. Pedersen, R. Peslin, and J.-C. Yernault.
Standardized lung function testing. Lung volumes and forced
ventilatory flows. Eur. Respir.
J. 6, Suppl. 16:
5-40, 1993.
18.
Sherrill, D., and G. Viegi. On modeling longitudinal
pulmonary function data. Am. J. Respir.
Crit. Care Med. 154, Suppl.: S217-S222, 1996.
19.
Voss, R. F.
Fractals in nature: from characterization to simulation.
In: The Science of Fractal Images, edited by H.-O. Peitgen,
and D. Saupe. New York: Springer-Verlag, 1988.
20.
Ware, J. H., and S. Weiss. Statistical issues in
longitudinal research on respiratory health. Am.
J. Respir. Crit. Care Med. 154, Suppl.: S212-S216, 1996.
21.
Weiss, S., and J. H. Ware. Overview of issues in
the longitudinal analysis of respiratory data.
Am. J. Respir. Crit. Care
Med. 154, Suppl.:
S208-S211, 1996.
22.
Woo, M. A.,
W. G. Stevenson,
D. K. Moser,
R. B. Trelease,
and
R. M. Harper.
Patterns of beat-to-beat heart rate variability and its association with increased mortality in advanced heart failure.
Am. Heart J.
123:
704-710,
1992[Medline].
This article has been cited by other articles:
![]() |
M. Schou, F. Gustafsson, A. Kjaer, and P. R. Hildebrandt Long-term clinical variation of NT-proBNP in stable chronic heart failure patients Eur. Heart J., January 11, 2007; (2007) ehl449v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Decramer, R Gosselink, M Rutten-Van Molken, J Buffels, O Van Schayck, P-A Gevenois, R Pellegrino, E Derom, and W De Backer Assessment of progression of COPD: report of a workshop held in Leuven, 11-12 March 2004 Thorax, April 1, 2005; 60(4): 335 - 342. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Dirksen Outcome measures in chronic obstructive pulmonary disease (COPD) Thorax, December 1, 2003; 58(12): 1007 - 1008. [Full Text] [PDF] |
||||
![]() |
American Thoracic Society/European Respiratory Society Statement: Standards for the Diagnosis and Management of Individuals with Alpha-1 Antitrypsin Deficiency Am. J. Respir. Crit. Care Med., October 1, 2003; 168(7): 818 - 900. [Full Text] [PDF] |
||||
![]() |
M. Decramer, P.N.R. Dekhuijzen, T. Troosters, C. van Herwaarden, M. Rutten-van Molken, C.P.O. van Schayck, D. Olivieri, I. Lankhorst, and A. Ardia The Bronchitis Randomized On NAC Cost-Utility Study (BRONCUS): hypothesis and design Eur. Respir. J., March 1, 2001; 17(3): 329 - 336. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Wencker, B. Fuhrmann, N. Banik, and N. Konietzko Longitudinal Follow-up of Patients With {{alpha}}1-Protease Inhibitor Deficiency Before and During Therapy With IV {{alpha}}1-Protease Inhibitor Chest, March 1, 2001; 119(3): 737 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. DIRKSEN, J. H. DIJKMAN, F. MADSEN, B. STOEL, D. C. S. HUTCHISON, C. S. ULRIK, L. T. SKOVGAARD, A. KOK-JENSEN, A. RUDOLPHUS, N. SEERSHOLM, et al. A Randomized Clinical Trial of alpha 1-Antitrypsin Augmentation Therapy Am. J. Respir. Crit. Care Med., November 1, 1999; 160(5): 1468 - 1472. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |