Knowledge of patient fluid distribution would be
useful clinically. Both single-frequency (SF) and impedance modeling
approaches are proposed. The high intercorrelation between body water
compartments makes determining the best approach difficult. This study
was conducted to evaluate the merits of an SF approach. Mathematical simulation was performed to determine the effect of tissue change on
resistance and reactance. Dilution results were reanalyzed, and
resistance and parallel reactance were used to predict the intracellular water for two groups. Results indicated that the amount
of intracellular and extracellular water conduction at any SF can vary
with tissue change, and reactance at any SF is affected by all tissue
parameters. Modeling provided a good prediction of dilution
intracellular and extracellular water, but an SF method did not.
Intracellular, extracellular, and total body water were equally
predicted at all frequencies by SF resistance and parallel reactance.
Extracellular and intracellular water are best measured through
modeling, because only at the zero and infinite frequencies are the
results sensitive only to extracellular and intracellular water. At all
other frequencies there are other effects.
bioimpedance spectroscopy; body cell mass; extracellular water; intracellular water; total body water
 |
INTRODUCTION |
KNOWLEDGE OF PATIENT FLUID distribution would be very
useful for guiding drug and renal replacement therapy as well as
nutritional support. Drug and renal replacement therapy to alleviate
excess fluid, as well as drug dosage, is presently prescribed using
only gross estimates of fluid distribution. Inaccurate information about fluid distribution can change drug disposition and requirement and cause serious acute and long-term complications for a dialysis patient. Nutritional status is generally assessed with often unreliable chemical and anthropometric methods. Dilution methods are tedious, and
magnetic resonance imaging, computer axial tomography, and whole body
counting are expensive and not suited for routine use. Some
investigators are attempting to use an X-ray-determined fat-free mass
to assess nutritional status, but this method cannot distinguish the
contribution of water to the estimate of fat-free mass. A loss in body
cell mass (BCM) with a concurrent increase in extracellular water (ECW)
(25, 36) would result in no detectable change in fat-free mass. A
simple, inexpensive, accurate, and reliable noninvasive method of
determining fluid distribution is needed. Impedance methods are
increasingly being used to fulfill this need.
Bioimpedance spectroscopy (BIS), which means fitting measured impedance
spectral data to a physical model, is a well-known analytic technique
(19). All the underlying principles used today in the impedance-body
composition field came from the use of this technique in biophysics
(8). In review, there is little conduction through skeletal muscle
tissue at low frequency (1-5 kHz), and the impedance is
principally a function of the ECW. As frequency increases, conduction
through the intracellular water (ICW) increases. At high frequency
(10-100 MHz) the ICW becomes fully conductive, and the impedance
is a function of both ECW and ICW (Fig.
1). This phenomenon is caused by cell
membrane capacitance (Cm) and named
-dispersion (4, 31). Thus at very low and very high
frequencies the overall impedance is essentially independent of the
Cm, whereas at
the mid- or characteristic frequency
(fc) the
dependence on the value of
Cm is at a
maximum. The fc
can also be defined as the frequency of maximum reactance. Two
different phenomena discovered at very low frequency (<1 kHz) and
very high frequency (>100 MHz) are named
- and
-dispersions,
respectively (31). Impedance spectral data measured on biological
tissue produce a semicircle with a suppressed center when resistance and reactance are plotted (Fig. 2). The
physical model most widely used since 1940 (23) to interpret this
phenomenon is the Cole model (3). The Cole model consists of the terms
resistance ECW (RE),
resistance ICW (RI),
Cm, and
the exponent
. It should be noted that the relationship
between RE and
RI and their respective volumes is
not simple, because of the nonlinear effects of the concentration of
nonconductor on the impedance (4, 41). Thus RE and
RI are only model terms, but for
simplicity RE and
RI are expressed as resistance ECW
and ICW, respectively. The Cole model is computed by using nonlinear
curve fitting to extrapolate the data to the low- and high-frequency
limits (19) (Figs. 1 and 2). These limits are known as resistance at
the zero frequency (R0), which
is the same as RE, and resistance
at the infinite frequency
(R
).
RI is computed as
1/RI = 1/R
1/R0. Although the Cole model is a
mathematical model and cannot be correctly represented as an equivalent
electronic circuit, an analog representation is shown in Fig.
3. Because biological tissue consists
of multiple parameters, modeling is considered essential, because it is
the only means of independently analyzing the different parameters
(19).

View larger version (65K):
[in this window]
[in a new window]
|
Fig. 1.
Frequency vs. resistance measured on 8 patients within 2 h after
cardiac surgery (time 3). Resistance
at zero (R0) and infinite
(R ) frequencies is
represented by 1 Hz and 100 MHz, respectively.
|
|

View larger version (22K):
[in this window]
[in a new window]
|
Fig. 2.
Resistance vs. reactance measured in 1 patient within 2 h after cardiac
surgery (time 3).
R0 and
R are represented by 1 Hz and
100 MHz, respectively.
fc,
Characteristic frequency (i.e., frequency of maximum reactance).
|
|

View larger version (12K):
[in this window]
[in a new window]
|
Fig. 3.
Equivalent electrical circuit analogous to Cole model.
RE, resistance extracellular water
(in ); RI, resistance
intracellular water (in );
Cm, membrane
capacitance (in F).
|
|
Thomasset was the first to attempt to apply this biophysics knowledge
clinically in 1963 (34). He proposed use of dual-frequency impedance
(i.e., 1 and 100 kHz) as a measure of ECW and total body water (TBW),
respectively (12, 34). An overwhelming number of different impedance
methods and multiple regression equations have since been introduced.
The field is now divided along three major lines of reasoning. The
first is the 50-kHz single-frequency impedance method that was
originally proposed as a measure of TBW (12) and fat-free mass (18,
26). With little scientific explanation, this method evolved into a
measure of both ECW and TBW (16) and now into a measure of TBW and BCM
(2, 14, 15, 24, 33). The second is prediction of ECW and TBW by low
(e.g., 1-5 kHz)- and high (e.g., 100-500 kHz)-frequency
impedance (5, 32). The third is use of a BIS or Cole modeling approach (4, 6, 11, 13, 38-40).
Because ECW and ICW are tightly regulated biologically and are parts of
the TBW, impedance measured at virtually any frequency in the 1-kHz to
1-MHz range equally predicts ECW, ICW, and TBW (40). This high
intercorrelation between the body water compartments has made
determining the best approach difficult. A plethora of single-frequency
impedance equations have been published, but cross validation and
detection of small changes in volume have generally been difficult. As
defined by the Cole model (3), biological tissue consists of four
parameters. A single-frequency impedance consists of two data (i.e.,
resistance and reactance). It may be that single-frequency impedance
equations do not cross validate well or detect small changes because
they rely on the tissue elements having relative uniformity between
individuals and when their tissue changes. There is also good reason to
believe that at any frequency other than zero and infinity the
impedance measurement would be affected, because a different amount of
volume would be measured when a change in
fc occurred. To
test these and other research questions, we conducted an evaluation.
Impedance spectral data measured before and after infusion during
cardiac surgery and before and after hemodialysis were compared with
the results of mathematical simulation. This determined the sensitivity of resistance and reactance to changes in the parameters of the Cole
model. Previous deuterium and sodium bromide results reported on
cardiac surgery patients (28) were reevaluated after use of different
apparent ECW and ICW resistivity constants in the volume equation we
developed from mixture theory (4). This analysis was conducted because
only the differences between methods had been reported (28), and it has
been discovered that the scaling of the apparent resistivity depends on
dilution method and protocol (4). To ascertain the validity of the
recently proposed parallel reactance
(XP) model (14, 15), we computed XP at multiple frequencies and
predicted the deuterium-sodium bromide space (i.e., ICW) measured on
the cardiac surgery patients. To test for sample dependency, total body
potassium (TBK) measured on a second sample was compared with TBK
predicted by XP at multiple frequencies.
 |
MATERIALS AND METHODS |
The cardiac surgery study was conducted at Henry Ford Hospital and
approved by the Henry Ford Human Research Committee; written informed
consent was obtained from each subject. Eight men, after elective
coronary artery bypass graft surgery, elected to participate. On the
patients' arrival at the intensive care unit, body weight was measured
to the nearest 1 kg with a standard balance, with the patient dressed
in a hospital gown. Body height was measured to the nearest 1 cm with a
stadiometer. Within 2 h of the patients' arrival, two blood samples
were taken to establish baseline deuterium and sodium bromide
concentrations. Patients then were measured preoperatively
(time 1) with a single-frequency
device (1 frequency at 50 kHz; model BIA 101, RJL Systems, Detroit,
MI). Patients were then immediately measured with a multiple-frequency
impedance device (44 frequencies logarithmically spaced between 1 kHz
and 1.348 MHz; model 4000B, Xitron Technologies, San Diego, CA).
Patients were placed in a supine position on their beds with limbs
slightly abducted. Skin current electrodes (model IS4000, Xitron
Technologies) were placed on the right dorsal surface of the hand and
foot at the metacarpals and metatarsals. Voltage-detector skin
electrodes were placed at the right pisiform prominence of the wrist
and between the medial and lateral malleoli at the ankle. Electrode placement was marked for subsequent measurements, and, when possible, electrodes were left in place for the duration of the study.
Within 2 h after surgery, the patients underwent bioimpedance testing
again (time 2). Patients also
received 10 g of intravenous solution of deuterium (9 ml of 99.9%
2H2O) and 30 mmol of
sodium bromide (10 ml of a 3 mmol/ml solution of sodium bromide) over 2 min through a central venous catheter. After 4 h of equilibration, two
additional blood samples were taken to determine postequilibration
concentrations (17, 29, 35). To predict ECW and TBW volumes,
bioimpedance was measured again immediately after blood sampling
(time 3). Blood samples were
collected in serum separator tubes, and the clotted samples were
centrifuged for 30 min. Mass spectroscopy (Metabolic Solutions, Boston,
MA) was used for deuterium analysis, and TBW was calculated using the
method of Schoeller et al. (30). Sodium bromide concentrations were
measured using an inductively coupled plasma mass spectrometer (PlasmaQuad2+, VG Elemental, Winsford, Cheshire, UK). Corrected sodium
bromide space was then estimated using the method of Price et al. (29),
whereby a correction is made for sodium bromide uptake into the red
blood cells and for Donnan equilibration. Dilution ICW was computed by
deuterium-sodium bromide.
The effects of isotonic saline infusion during surgery and after
recovery from surgery on the Cole model parameters were analyzed. Impedance measurements were taken on day
2, 24 h postdilution steady-state equilibration
(time 4), and on
day 3, 48 h postdilution steady-state
equilibration (time 5). Patients
received an average of 2.1 liters of infused fluid during surgery.
For the single-frequency method, ECW and TBW volumes were predicted by
applying the measured resistance and reactance measured at
time 3 to the same statistical
equations used by Patel et al. (28; see also Refs. 24 and 37). For the
BIS method a nonlinear curve-fitting program described previously (4)
was used to fit impedance and phase spectral data to the Cole model.
Cole model terms RE and
RI were used in a volume equation
we developed from mixture theory to predict ECW and ICW volumes,
respectively (4). TBW was computed as ECW + ICW. Recorded fluid inputs
and outputs were used to calculate estimated body weight at the time of
dilution steady state from baseline. As discussed previously (4), the
mixture-volume equation utilizes apparent ECW and ICW resistivity
constants. The constants used by Patel et al. (28) were previously
established from deuterium and sodium bromide data (38) and cross
validated (27). The male ECW and ICW constants used to predict ECW and
ICW were 214 and 824
· cm, respectively (28).
Also, these terms are only apparent resistivities, because they are
affected by geometry when a wrist-to-ankle measurement is made. (For a
full discussion see Ref. 4.) For the statistical analysis the Microsoft
Excel program was used. In addition to the descriptive statistics,
Pearson's product moment correlation and standard error of estimate
(SEE) statistics were computed for the relationship among dilution-,
single-frequency-, and BIS-determined ECW, TBW, and ICW volumes.
Resistivity constants computed from deuterium and sodium bromide data
collected by De Lorenzo et al. (4) were then used to repredict ECW and
ICW volume. The male values used for ECW and ICW were 174 and 1,177
· cm, respectively. The correlation, SEE, mean,
and mean differences were recomputed. New resistivity constants for the ECW and ICW were then computed by regressing the
samples computed by Cole model terms
RE and
RI against their dilution-determined ECW and ICW volumes, respectively. The equation and
methods we used to compute resistivity have been described previously
(4, 11). The ECW and ICW resistivity constants became 229 and 1,054
· cm, respectively.
Reactance at 5, 10, 49, 100, 204, 424, and 876 kHz measured at
time 3 was transformed into
XP by using the recently proposed equation: XP = reactance + resistance2/reactance (14, 15).
XP was then used in the male
multiple regression equations published by Kotler et al. (14) to
predict the dilution ICW: 254 * Ht2/XP + 1,493, where Ht is height (simple linear equation), 59.06 * Ht1.6/X0.5P
(base exponential equation), and 0.76 * 59.06 * Ht1.6/X0.5P + 18.52 * weight
386.66 (exponential prediction equation).
Although 49 kHz, rather than 50 kHz, was used, it is virtually
equivalent. Predicting dilution ICW with a TBK equation has validity,
because potassium is primarily distributed in the ICW and any
relationship between TBK and BCM is through the TBK-ICW relationship
(7). To test for intercorrelation between variables, dilution ICW was
also predicted by the simple
Ht2/resistance equation at all
frequencies. To further test for intercorrelation, dilution ECW and TBW
volumes were predicted at all frequencies by each of the above
XP equations. The descriptive
statistics, correlation and SEE, were computed.
To test for sample-dependent variation, we conducted the same
XP analysis on a second group of
48 healthy Italian men 26-57 yr of age. This study, which was
performed at University of Rome "Tor Vergata," was approved by
their Medical Ethical Committee. The subjects volunteered to
participate in the study (4), and written informed consent was obtained
from all participants. On the subjects' arrival in the morning in an
overnight-fasted state, body weight was measured to the nearest 0.05 kg
with a standard balance, with the subjects dressed in swimming clothes.
Body height was measured to the nearest 1 mm with a stadiometer. After
the measurement of weight and height and with the subjects still in the
fasting state, 40K was measured
with a whole body counter. The whole body counter was formed by a cell
2.5 m wide and 3 m high of 10-cm-thick lead bricks, the door of which
was formed by a 22-cm-thick iron slab. The room was continuously
ventilated. A single 20.3 × 10.2-cm thallium-activated sodium
iodine crystal was positioned above the subject, who was measured in a
sitting position and dressed only in paper pajamas. TBK was calculated
as 40K * 8,474.6 (7). The
coefficient of variation for TBK was 2-3%. ICW volume was
computed by assuming that TBK is present only in the ICW and that
potassium concentration in the ICW is 150 mM (7). After the measurement
of weight, height, and TBK, wrist-to-ankle (i.e., whole body) impedance
spectra, consisting of 21 frequencies ranging from 1 kHz to 1.348 MHz,
were measured with the same model multiple-frequency impedance device
described above. The measurements were taken within the first several
minutes after the subjects assumed a supine position. The measurements
were taken on the left side of the body with use of disposable
electrocardiograph electrodes (5 cm2, 3M, Minneapolis, MN) and in
accordance with the standard wrist-to-ankle protocol discussed above
(38). The impedance and phase spectral data were fit to the Cole model.
The equation described above (14, 15) was used to transform reactance
at 5, 10, 50, 100, 200, 500, and 748 kHz into
XP. Transformed
XP was used in the same equations
to predict the TBK-determined ICW of 48 healthy Italian men. Predicting
ICW with a TBK equation has validity, and any error in assuming the
potassium concentration in the ICW would only result in a scaling
difference (7). The frequencies differed from those used for the
cardiac patients, because different software, which requested a
different set of frequencies from the device, was used.
To investigate the effects of fluid and solute removal through
hemodialysis, impedance spectral data measured on 16 patients before
and after hemodialysis were fit to the Cole model and analyzed. This
study, which was conducted at the Karolinska Institute, was approved by
the Huddinge Hospital Medical Ethical Committee. The patients (10 men
and 6 women), ranging in age from 35 to 84 yr, volunteered to
participate in the study. Informed consent was obtained from all
participants. The patients were randomly selected from 28 hemodialysis
patients. The amount of fluid removed by ultrafiltration ranged from
0.08 to 3.96 liters. Treatment lasted for 4-4.5 h. Blood flow was
250-300 ml/min, dialysate flow was 500 ml/min, dialysate sodium
concentration was 141 mM, and bicarbonate concentration
was 34 mM. Dialysis machines (model AK 100, Gambro) and several types
of dialyzers (models AC 130 and 170, Baxter, and models GEF 15 and GEF
18, Gambro) were used. Temperature did not change during the procedure.
On the patients' arrival in the dialysis unit and after hemodialysis,
body weight was measured to the nearest 0.1 kg with a standard balance.
Body height was measured to the nearest 1 cm with a stadiometer. After
the patients were in a supine position on a bed for 20 min,
wrist-to-ankle (i.e., whole body) impedance spectra, consisting of 50 frequencies ranging from 5 to 500 kHz, were measured with the
multiple-frequency device described above. The measurements were taken
on the side opposite the side with the arteriovenous fistula by use of
skin electrodes (model IS4000, Xitron Technologies) and in accordance with the standard wrist-to-ankle protocol discussed above. Electrodes were left in place throughout the study, and patients remained in a
supine position throughout the treatment. After termination of
treatment, impedance spectra were measured again, and the data were fit
to the Cole model. Descriptive statistics were computed, and the
spectral data were graphically formatted. The difference in impedance
before and after dialysis at the low- and high-frequency limits (i.e.,
R0 and
R
) and at 5 and 500 kHz was
compared.
To simulate the sensitivity of resistance and reactance to changes in
the parameters of the Cole model, we computed resistance and reactance
at multiple frequencies in the 100-Hz to 10-MHz frequency range (i.e.,
-dispersion) (3, 31) for nominal data and with the modeled
parameters changed. Nominal data were Cole modeling results obtained
previously on one healthy adult woman (27) and previously reported
dispersion data outside the
-dispersion range on skeletal muscle
tissue (3, 31). At each frequency, resistance and reactance were
computed with the Cole model term
set at 0.63 and 1, after 5 liters
were added to the ICW and ECW, and for
fc changed from
40 to 100 kHz. The resistance and reactance data were then plotted vs.
log frequency with use of a Microsoft Excel Spreadsheet program.
To properly evaluate the single-frequency
XP proposal, we performed a
mathematical modeling evaluation (MATLAB, Math Works, Natick, MA). This
was performed to ascertain the interaction between XP and the biophysical parameters
expressed in the Cole model. Use of the Cole model had validity,
because it is the model most widely used to interpret impedance
measurements of biological tissue (23), and measured data in vivo fit
it with high precision (4). Impedance spectral data of one healthy
Asian man were fit to the Cole model. Nominal values were 594.9
for
RE, 935.6
for
RI, 2.85 nF for
Cm, and 0.7 for
. For frequencies of 5, 50, and 200 kHz and 1 MHz, we individually
varied RE,
RI,
Cm, and
±20% in 1% increments. The following equations were used to
compute XP at each 1% increment
for each frequency ( f). For the
Cole model, we have the impedance (Z) equation
|
(1)
|
where
= 2
f and
j =
.
Let RS and
XS be the series resistance and
reactance of the impedance, respectively; i.e.
|
(2)
|
From
Eqs. 1 and 2, one derives
|
(3)
|
and
|
(4)
|
where
|
(5)
|
Now
let RP and
XP be parallel resistance and
reactance of the impedance, respectively; i.e.
|
(6)
|
We
can obtain the relationships between series and parallel resistance and
reactance from Eqs. 2 and 6
|
(7)
|
|
(8)
|
Finally,
by replacing RS and
XS in Eqs.
7 and 8 with
Eqs. 3 and 4, we have the following equations for
RP and
XP
|
(9)
|
|
(10)
|
The results were then divided by the nominal value to express the
result uniformly as a ratio. Then the
XP ratio change was plotted vs.
the change in ratio for RE,
RI,
Cm, and
.
 |
RESULTS |
The physical characteristics of the cardiac surgery patients, as well
as their dilution-determined ECW, ICW, and TBW volumes, are shown in
Table 1. Table
2 displays their Cole modeling results. The
impedance data measured on the eight male cardiac surgery patients
corresponded extremely well to the Cole model. The results of Cole
modeling computed at five different time points before and after
cardiac surgery are shown in Table 3.
RE and
RI are not simply related to
volume (4). However, these data suggest that the infusion of fluid
affected both the ECW and ICW, since both
RE and
RI changed. The infused fluid
resulted in a decrease in RE and
RI and a concurrent increase in
fc.
As shown in Table 4, for the cardiac
surgery patients, single-frequency impedance predicted TBW well, with
little mean difference compared with dilution. The correlation between
single-frequency impedance and dilution ECW was reasonable, but the SEE
was quite high, and there was a mean difference of 1.7 liters
(Table 4). For single-frequency ICW, which was determined by
subtracting the predicted ECW from the predicted TBW, both
the correlation and SEE values were poor, and the mean difference was
large (Table 4).
View this table:
[in this window]
[in a new window]
|
Table 4.
Prediction of cardiac surgery patients' 2H2O
(TBW), NaBr (ECW), and 2H2O-NaBr (ICW)
spaces by use of single-frequency impedance and BIS approach with
use of different apparent ECW and ICW resistivity constants
|
|
Except for mean difference, the BIS prediction of the cardiac surgery
patients' dilution ECW, ICW, and TBW volumes was good for all three
sets of resistivity constants (Table 4). For ECW the correlation and
SEE values were the same for each constant (Table 4). For ICW,
correlation and SEE varied slightly. This establishes that ECW
resistivity, as used in our mixture-volume equation, is purely a scalar
term, having no effect on correlation or SEE. ICW resistivity is
effectively a scalar, since the nonlinearity was slight (Table 4).
Most noticeable about Table 4 is that the use of different resistivity
constants dramatically changed the mean difference between the BIS- and
dilution-determined volumes. The variation in scaling caused by the
dilution method equally affects the single-frequency method. These
results are considered further in the
DISCUSSION.
For the cardiac surgery patients, the strongest prediction of dilution
ICW with XP was achieved by using
Kotler's base exponential equation at all frequencies (14) (Table
5). Similar predictions of dilution ICW
were discovered at all the measured frequencies. Most interesting was
that the best prediction produced by
XP was not at 49 kHz but at 204 kHz. As shown in Table 6, the prediction of
dilution ICW with use of resistance alone, which is in the XP equation, was strong.
XP also predicted dilution ECW and
TBW with high correlation and reasonable SEE values at all frequencies measured (Table 7). The descriptive
characteristics of the 48 healthy Italian men used to further analyze
the XP model are shown in Table
8; their Cole modeling results are shown in
Table 9. As found with the cardiac surgery
patients, similar predictions of TBK were produced by
XP at all frequencies (Table
10).
XP computed at 50 kHz again did
not provide the best prediction of TBK. The best prediction was
provided by XP at 10 kHz, which is
different by a factor of 20 from that which predicted best for
the cardiac surgery patients (Table 5). The TBK predicted by resistance
alone was almost as good as that predicted by the inclusion of
reactance (Table 11).
View this table:
[in this window]
[in a new window]
|
Table 5.
Prediction of cardiac patients' 2H2O-NaBr
space (ICW) by use of base exponential Xp equation of
Kotler et al. (14) at various frequencies
|
|
View this table:
[in this window]
[in a new window]
|
Table 7.
Prediction of cardiac patients' dilution NaBr (ECW) and
2H2O (TBW) spaces by use of exponential
XP prediction equation of Kotler et al. (14) at various
frequencies
|
|
View this table:
[in this window]
[in a new window]
|
Table 10.
Prediction of TBK-determined ICW by use of base exponential
XP equation of Kotler et al. (14) at various
frequencies in healthy Italian men
|
|
View this table:
[in this window]
[in a new window]
|
Table 11.
Prediction of TBK-determined ICW by use of
Ht2/resistance at various frequencies in healthy Italian
men
|
|
The descriptive characteristics of the hemodialysis patients are shown
in Table 12; their Cole modeling results
before and after hemodialysis are shown in Table
13. As found with the cardiac surgery
patients and healthy Italian men, the measured impedance data fit the
Cole model well. Opposite to the results after infusion of fluid (Table
3), RE increased rather than
decreased with ultrafiltration (Table 13).
RI remained relatively stable,
suggesting that most of the fluid was removed from the ECW.
Cm changed
considerably, and
remained significantly lower than that found for
healthy subjects (Table 9). The effect Cole model term
is believed to represent a distribution of time constants caused by different cell
sizes and shapes (3). If this is the case, a decrease in
would
suggest a widening in cell size and shape. The
fc, rather than
increasing, as when fluid was infused, decreased when fluid was removed (Table 13).
The results of the mathematical modeling revealed virtually no change
in XP at 50 kHz with a ±20%
change in RE and a change of only
~2% at the extremes of frequency (Fig.
4). A ±20% change in
RI caused an ~15% change in
XP at 50 kHz and a very large
change (e.g., ~30%) at 1 MHz (Fig. 5). A
±20% change in
Cm caused an ~4% change in XP at 50 kHz, and
the effect was slightly nonlinear. At 1 MHz the effect of a ±20%
change in Cm
resulted in an ~12% change in
XP (Fig.
6). A ±20% change in
(from 0.56 to
0.84, ±20% from the nominal value of 0.7; Table 9) caused an
~26% change in XP at 50 kHz,
and the result was quite nonlinear. The second largest effect on
XP was caused by a change in
at 1 MHz (Fig. 7).
The frequency-resistance plot (Fig. 8)
suggests that at frequencies low enough to ensure that the measurement
is solely dependent on ECW (below a few hundred hertz), the effects of
a different phenomenon (
-dispersion) have become significant (4,
31). The resistance measured at the common 50-kHz frequency (2, 18, 33)
was only partially conducting through the ICW; thus it is clearly
dependent on both ECW and
fc. This also
confirms the criticism that 50-kHz measurements of TBW are dependent on high ECW-TBW intercorrelation (40). It is well known that ECW and ICW
are not fully measured until >10 MHz (4, 31). This is supported by in
vivo data (Fig. 1). Even if it were technically feasible to measure to
such high frequencies, the influence of another unwanted phenomenon
(
-dispersion) would still need to be removed. To predict TBW with
any single frequency requires the use of a TBW resistivity term
and the assumption that it is fixed. The resistivity of ECW is
different from the resistivity of ICW by a factor of 3-4 (4, 9);
thus the sensitivity of a single high-frequency measurement to changes
in ECW and ICW is different. A simple change in the ECW-to-ICW ratio
will significantly alter TBW resistivity and cause error, even when ion
concentration has not changed. This error can be removed by solving for
Cole model terms RE and
RI and determining independent
resistivities for the ECW and ICW, respectively. From both theoretical
(Fig. 8) and measured data (Fig. 1), it can be readily demonstrated that at any single frequency the amount of ECW and ICW conduction varies not only between subjects but when
fc changes. A
change in RE,
RI, or
Cm can cause a
change in fc.
When fc changes,
the amount of conduction through the ECW and ICW at any fixed frequency changes. Because
fc decreases when
fluid is removed through ultrafiltration, it was not surprising that,
between pre- and postdialysis, impedance at the low-frequency limit
(i.e., R0) changed 18.9% more
than at 5 kHz (Table 14). Nor was it
surprising that impedance changed 9.2% less at the high-frequency
limit (i.e., R
) than at 500 kHz (Table 14). Also not unexpected, impedance was greater at
R0 than at 5 kHz and less at
R
than at 500 kHz. These findings are considered more thoroughly in the
DISCUSSION.

View larger version (47K):
[in this window]
[in a new window]
|
Fig. 8.
Effect of changing from 0.63 to 1, adding 5 liters of intracellular
water (ICW), adding 5 liters of extracellular water (ECW), and changing
fc from 40 to 100 kHz on resistance.
|
|
The frequency-reactance plot (Fig. 9)
suggests that reactance is sensitive to all body composition parameters
in the frequency range around
fc and is
sensitive to the other dispersions (i.e.,
and
) for frequencies
significantly different from
fc. This indicates a tenuous relationship between reactance at any single frequency and any single body composition parameter, with any apparent
relationship being accentuated by the high correlation between
parameters (40). Furthermore, the sensitivity of reactance is extremely
dependent on the relationship between the frequency of measurement and
fc and is
symmetrical about
fc. Phase is a function of the ratio of resistance to reactance
[arctan(reactance/resistance)]; thus it is sensitive to all
the problems associated with both single-frequency resistance and
reactance.

View larger version (56K):
[in this window]
[in a new window]
|
Fig. 9.
Effect of changing from 0.63 to 1, adding 5 liters of ICW, adding 5 liters of ECW, and changing
fc from 40 to 100 kHz on reactance.
|
|
 |
DISCUSSION |
In review of Figs. 1, 8, and 9, the debate concerning the best single-
or dual-frequency impedance to use for measuring ECW and ICW appears
irrelevant (5, 15). Apart from the error caused by not accounting for
the different ECW and ICW resistivities, the proportion of current
conducting through the ICW at any single frequency varies. This is
supported by the hemodialysis results (Table 14). The change in
impedance before and after hemodialysis at
R0 was 18.9% greater than that at
5 kHz. With a decreased
fc after dialysis
(Table 13), conduction through the ICW at 5 kHz would be greater after
than before dialysis. More ICW conduction would alter the impedance,
because an increased volume (i.e., ICW) is being measured, making the
measurement less sensitive to changes in ECW. Similarly, because
ultrafiltration principally affects the ECW, one would expect the
impedance to change less at
R
than that at
R0, because both ECW and ICW are
being measured. As also shown in Table 14, between pre- and
postdialysis the impedance changed 9.2% more at 500 kHz than at
R
. Inasmuch as
fc decreases
after dialysis (Table 13), there is more conduction through the ICW at
500 kHz than before dialysis. By use of curve fitting to determine the
impedance at R0 and
R
, the measurement becomes
independent of
fc, and error
caused by the change in
fc is effectively
removed (Fig. 2). Although RI can
be computed by the impedance measured at other frequencies, the error
in calculating RI can be as high
as 200% by not using R
(13).
The previous conclusion that single-frequency impedance and BIS methods
provide similar ECW information appears tenuous. A 50-kHz resistance
had been used to predict TBW and fat-free mass, then the predicted
fat-free mass and reactance were used to predict ECW (28). There is
little relationship between reactance and ECW (32); thus, one datum
(i.e., resistance at 50 kHz) should not be used to predict two
variables. This is not intended as criticism, for such statistical
reasoning is widespread, but rather an example of how a statistical
equation that has no scientific basis can produce misleading results
(33). The lack of scientific reasoning and reliance on statistical
methods, despite high intercorrelation between variables (40), has
caused a great deal of confusion in the field.
The finding that the ECW and ICW resistivities are scalar or
multiplicative terms was replicated in this study (4, 22, 39) (Table
4). Different dilution methods (e. g., sodium bromide vs. sulfate)
produce significantly different-sized ECW and TBW spaces (7). It is now
believed that the size of the space measured by dilution varies for
even the same methods when different protocols and analysis techniques
are used (4, 39). This is further supported by this study, since both
deuterium and sodium bromide were used (Table 4). Had Patel et al. (28)
previously used the resistivity constants computed by De Lorenzo (4)
rather than those by Van Loan et al. (38), the mean TBW difference would have been very small, but then the mean difference for ECW would have been large (Table 4). Because each of these studies used deuterium and sodium bromide, it is difficult to determine which
scaling is physiologically correct.
Several findings emerge from this reassessment of the dilution results
on the cardiac surgery patients. The measured impedance spectral data
corresponded well to the Cole model (Table 2), and the impedance
spectroscopy approach provided a better prediction of ECW and ICW than
did single-frequency impedance (Table 4). The single-frequency TBW
result may appear adequate, but resistance at 50 kHz is a tenuous
prediction of TBW, because, as shown in Figs. 1 and 8, only a portion
of the ICW is measured at this frequency. Thus any 50-kHz prediction of
TBW would rely on high intercorrelation between ECW and TBW. The
further a subject's ECW and ICW volume deviated from the sample used
to regress the equation, the less precise the prediction would become.
This must be so, because the ECW and ICW resistivities are different by
a factor of 3-4 (3, 9), and
fc can change
when any tissue variable changes (Fig. 9). ECW can also be predicted by
a 50-kHz impedance by forcing a fit to an equation. That such an
approach is simply forcing an equation to fit the data, rather than a
measurement, is evidenced by the poor prediction of ICW (i.e.,
predicted TBW
predicted ECW) shown in Table 4. On the other
hand, similar predictions were obtained for ECW, ICW, and TBW when the
BIS approach was used (Table 4).
We previously stated that reactance can only be associated with
Cm, because ICW
is a resistive, not a reactive, medium (4). This is true, but the
results of mathematical modeling revealed that
XP is strongly influenced by
RI (Fig. 5). The Cole model predicts that it will be. The problem is that
XP is also highly sensitive to
changes in Cm and
(Figs. 6 and 7). Although
Cm and
are
considerably decreased in the cardiac surgery patients (~16 and 14%,
respectively; Table 2) compared with healthy men (Table 9),
Cm and
change
dramatically in theory (Fig. 9) and in practice (Tables 3 and 13).
Bestoso and Mehta (1) observed a mean 50% increase in
Cm after fluid
and solute removal by hemodialysis. We have discovered
Cm to be
considerably decreased in clinical populations (Tables 3 and 13)
compared with healthy subjects (Table 9) and on the individual level as
much as 75% (Table 15). The
exponent
was observed to be extremely decreased (i.e., 0.45) in
severely depleted patients compared with healthy subjects (i.e., 0.7;
unpublished observations). This study also discovered that
is lower
in clinical populations (Tables 3 and 13) and is considerably lower on
the individual level (Table 15).
Variation in Cm
and
was less in patients who had fluid infused (Tables 3 and
16) than in those who had fluid removed
through ultrafiltration (Tables 13 and 15). We suspect this is due to chemical changes that occur during hemodialysis. That
Cm increases with
fluid removal and decreases with overhydration is very interesting. Theoretically, Cm
changes only when there is a change in the thickness of the cell
membrane (10). What is interesting about Figs.
10 and 11
is that on recovery (time 5) the
frequency response approaches that of patients before cardiac surgery
(time 1). Most noticeable about
Figs. 12 and
13 is the higher
fc for this
dialysis patient than for healthy subjects (Table 9) and its change
from 124 kHz before to 71 kHz after dialysis (Table 15). It can be seen
in Fig. 10 that the data follow the common S curve that is easily fit.
However, in Fig. 12 the curvature is less discernible. We previously
reported (4) that modeling with both impedance and phase is essential,
because phase has a much broader range of sensitivity to change than
impedance (Figs. 11 and 13). Deurenberg et al. (5) stated that fitting
the model with impedance alone is adequate. However, with the data
shown in Fig. 12, the results of modeling by use of impedance alone
would be far less precise, because there would be no discernible curve
to follow without phase (19).
View this table:
[in this window]
[in a new window]
|
Table 16.
Cole modeling results of data measured at five time points on one
patient before and after cardiac surgery
|
|

View larger version (61K):
[in this window]
[in a new window]
|
Fig. 10.
Frequency vs. resistance measured on 1 patient at 5 time points before
and after cardiac surgery. R0 and
R are represented as 0.001 Hz
and 100,000 MHz, respectively. R0
and R were derived from
modeling.
|
|

View larger version (59K):
[in this window]
[in a new window]
|
Fig. 11.
Frequency vs. reactance measured on 1 patient at 5 time points before
and after cardiac surgery. R0 and
R are represented as 0.001 Hz
and 100,000 MHz, respectively. Reactance at
R0 and
R were set at zero.
|
|

View larger version (62K):
[in this window]
[in a new window]
|
Fig. 12.
Frequency vs. resistance measured on 1 patient before and after
hemodialysis. R0 and
R are represented as 0.001 Hz
and 100,000 MHz, respectively. R0
and R were derived from
modeling.
|
|

View larger version (58K):
[in this window]
[in a new window]
|
Fig. 13.
Frequency vs. reactance measured on 1 patient before and after
hemodialysis. R0 and
R are represented as 0.001 Hz
and 100,000 MHz, respectively. Reactance at
R0 and
R were set at zero.
|
|
Similar to a single-frequency resistance-predicted ECW or TBW, an
XP-predicted ICW is dependent on
the elements in the tissue having relative uniformity. The mathematical
modeling revealed that XP as
presented previously (14, 15) is merely a simplification of the Cole
model. Van Marken Lichtenbelt et al. (40) and others discovered that
resistance at any frequency predicts ECW, ICW, and TBW with virtually
equal precision. We have found that
XP also predicts ICW, ECW, and TBW
with equal precision at any frequency measured (Tables 5, 7, and 10).
The best prediction of ICW with XP
was not even produced by the proposed 49-50 kHz (14, 15) and was
sample dependent. Furthermore, RS
alone provided similar predictions (Tables 6 and 11). From the name of
this new theory, "parallel reactance," it would seem that
reactance should be providing most if not all the prediction, but the
opposite is the case. The correlation and percent SEE values reported
by Kotler et al. (14) for the
XP-predicted BCM were only 0.04 and 1.1% better, respectively, than those using
RS alone. The same resistance was also used to predict TBW (14). Lukaski (15), who is promoting an
XP BCM prediction, previously
promoted reactance as a measure of ECW (16), but confusingly it was
reported to be invalid by Kotler's laboratory, because reactance was
contributing virtually nothing to the prediction (32).
No theoretical basis for predicting BCM with
XP has been reported (14), and the
statement was made that "a major uncertainty in the theory
underlying BIA [bioelectrical impedance analysis] whether
the body's ionic circuit is arranged as a series or parallel circuit" (14). This statement should have been accompanied by a
reference. It was suggested that the improved correlation between XP and BCM was proof that
XP was superior to
XS for predicting BCM (14). It is
true that impedance measured at any single frequency can be interpreted
as a series or parallel circuit, with both resulting in two final
elements (resistance and reactance). The problem is that biological
tissue consists of more than two elements. No reference was given,
because there is no rigorous biophysical research to support this claim
or a single-frequency prediction of cell volume. Single biological
cells have been interpreted in biophysics since 1925 as a three-element
model, with an RE in parallel with
a series Cm and
RI (8). Cole (3) added an