Vol. 84, Issue 3, 862-867, March 1998
Evaluation of a forced oscillation method to measure thoracic
gas volume
R.
Peslin and
C.
Duvivier
Unité 14 de Physiopathologie Respiratoire, Institut National
de la Santé et de la Recherche Médicale, Université
H. Poincaré Nancy I, 54500 Vandoeuvre-les-Nancy, France
 |
ABSTRACT |
The purpose
of this study was to test a plethysmographic method of measuring
thoracic gas volume (TGV) that, contrary to the usual panting method,
would not require any active cooperation from the subject. It is based
on the assumption that the out-of-phase component of airway impedance
varies linearly with frequency. By using that assumption, TGV may be
computed by combining measurements of total respiratory impedance (Zrs)
and of the relationship between the plethysmographic signal (Vpl) and
airway flow (
) during forced
oscillations at several frequencies. Zrs and Vpl/
were measured at 10 noninteger multiple frequencies ranging from 4 to
29 Hz in 15 subjects breathing gas in nearly
BTPS conditions. Forced oscillation
measurements were immediately followed by determination of TGV by the
standard method. The data were analyzed on different frequency ranges,
and the best agreement was seen in the 6- to 29-Hz range. Within that
range, forced oscillation TGV and standard TGV differed little
(3.92 ± 0.66 vs. 3.83 ± 0.73 liters,
n = 77, P < 0.05) and were strongly
correlated (r = 0.875); the
differences were not correlated to the mean of the two estimates, and
their SD was 0.35 liter. In seven subjects the differences were
significantly different from zero, which may, in part, be due to
imperfect gas conditioning. We conclude that the method is not highly
accurate but could prove useful when, for lack of sufficient
cooperation, the panting method cannot be used. The results of computer
simulation, however, suggest that the method would be unreliable in the
presence of severe airway inhomogeneity or peripheral airway
obstruction.
respiratory mechanics methods; plethysmography; respiratory
impedance
 |
INTRODUCTION |
ALTHOUGH SOME SYSTEMATIC errors have been demonstrated
in patients with severe airway obstruction (2, 25, 26), the reference
method to estimate thoracic gas volume (TGV) is that proposed by DuBois
et al. (6), namely, the plethysmographic measurement of body volume
changes (Vpl) during breathing efforts against a closed airway. A
drawback of the method, however, is that it requires some degree of
active cooperation from the subject to correctly perform the maneuver.
Several successive inspiratory and expiratory efforts of uniform
amplitude are, in our experience, necessary to detect the presence of a
drift in Vpl and correct the data for it (20). Thus it may be difficult
to obtain satisfactory recordings in young children, in patients with
dyspnea and, more generally, in uncooperative subjects. The purpose of
this study was to assess the reliability of an alternative method that
would not require any active cooperation from the subject. It is based on the combination of alveolar pressure
(PA) measurements by body plethysmography and of measurements of total respiratory mechanical impedance (Zrs) by forced oscillation; it also uses an assumption about
mechanical properties of airways. The method was tested in healthy
subjects by reference to the standard plethysmographic technique.
 |
PRINCIPLE |
When a subject rebreathes BTPS gas in
a body plethysmograph, Vpl is proportional to
PA and to TGV (6)
|
(Eq. 1)
|
where
K is the absolute pressure of dry gas
in the alveoli [barometric pressure
water vapor pressure
(PH2O)].
On the other hand, in the frequency domain, airways impedance (Zaw) is
the complex ratio of the pressure drop from the airway opening (Pao) to
the alveoli (Pao
PA) to gas flow
(
). Taking PA
from Eq. 1
|
(Eq. 2)
|
where
Pao/
is respiratory input impedance (i.e., Zrs), the
relationship between a pressure input at the airway opening and the
resulting flow. Zaw is a complex variable that includes a resistive
component [airway resistance (Raw)], and an out-of-phase component or reactance (Xaw). Provided airway walls are relatively stiff and airway properties are homogeneous, Xaw at comparatively low
frequencies (<50 Hz in humans) is expected to be dominated by gas
inertance (Iaw) and to increase linearly with frequency
|
(Eq. 3)
|
where
is the circular frequency (
= 2 ·
· f with f the
frequency). Such a linear relationship has indeed been observed in
studies in which Zrs has been partitioned into its airway and tissue
components (17, 22). The method described herein relies heavily on
Eq. 3 and assumes that Xaw/
is
constant (i.e., independent of
). By using that assumption and
considering the out-of-phase components of Pao/
(total total respiratory reactance; Xrs) and of Vpl/
(plethysmographic reactance; Xpl) in Eq. 2
|
(Eq. 4)
|
On
the basis of Eq. 4,
measurements of Xrs and Xpl at two frequencies
(indexes 1 and
2) are sufficient to compute TGV
|
(Eq. 5)
|
If
measurements at more than two frequencies are available, the value of
TGV that minimizes the variance of Xaw/
may be obtained by
|
(Eq. 6)
|
where
index i designates the measurement,
n is the number of measurements,
Ai = Xrsi/
i
and Bi = Xpli/
i. One may point out that the method does not impose any
specific condition on Xpl and on the reactance of the
tissues. It should be clear, however, that the very
definition of Zaw as (Pao
PA)/
contains the implicit assumption that
PA is homogeneous. The
consequences of mechanical inhomogeneity are examined in
DISCUSSION.
 |
MATERIALS AND METHODS |
The method was tested by using measurements of Xrs and Xpl obtained in
15 healthy subjects for a different purpose [separation of airway
and tissue components of Zrs by using the standard plethysmographic TGV
(TGVst)(16)]. The subjects
(10 men, 5 women) were aged 25-66 yr and were recruited among the
laboratory staff.
Equipment.
The subjects were seated in a 370-liter constant-volume body
plethysmograph and connected, via a mouthpiece, shutter, and Fleisch
no. 2 pneumotachograph, to a forced oscillation setup placed inside the
box. Pressure variations at the airway opening were applied by a
loudspeaker (Pioneer TS-W201, Pioneer Electronic), and the subject
rebreathed through a side tube from a low-impedance bag in which the
gas was maintained at BTPS by a
thermostated water bath. Box pressure (Pbox) and the pressure drop
across the pneumotachograph were measured with MP45 ±2-hPa Validyne
transducers (Validyne, Northridge, CA) and Pao with a DP15 ±50-hPa
Validyne transducer. All transducers were matched within 1% of
amplitude and 2° of phase up to 30 Hz. Pao was calibrated
by using a slanted fluid manometer and
by the
integral method that uses a 1-liter syringe. Pbox was calibrated in
terms of Vpl with a small reciprocating pump at frequency of ~2 Hz.
The input signal sent to the loudspeaker included 10 noninteger
multiple-frequency components ranging from 4 to 29 Hz. It was generated
at a rate of 320 Hz by a 486-type computer equipped with a 12-bit
analog-to-digital and digital-to-analog conversion board (PC-Lab,
Digimétrie, Perpignan, France). Pao,
, and
Vpl were sampled at the same rate by the computer after analog low-pass filtering with a corner frequency of 40 Hz.
Protocol.
The subjects supported their cheeks firmly with their palms during the
measurements. After Pbox had steadied, the forced oscillation data were
collected for 33 s. Then, the oscillations were stopped, the shutter
was closed and the subject was asked to pant for 5 s against the
occlusion to measure TGV by the usual approach (6). The flow signal was
sampled without interruption during the whole sequence so that any
difference between the mean lung volume during the forced oscillation
measurements and the volume when the shutter was closed could be
computed and corrected for. Five to six such measurements were made at
2- to 3-min intervals, the rebreathing bag being thoroughly washed with
fresh air between successive measurements.
Data analysis.
Because the mechanical time constant of the plethysmograph was kept
relatively short (3 s) to minimize the drift of Pbox and avoid
saturating the analog-to-digital converter, Vpl was first corrected for
it. This was done by adding to Vpl its integral divided by the time
constant (16). Then, forced oscillation signals were high-pass filtered
at 2 Hz to eliminate the main breathing components. The
Fourier coefficients of the signals were computed at the 10 frequencies
of interest and combined to obtain the in-phase and out-of-phase
components of Zrs and Vpl/
. Both were corrected for
the 2.1-ms time constant of the pneumotachograph. Vpl/
was also corrected for the gas compression in
the loudspeaker enclosure (0.8 liter); gas compression in the
rebreathing bag appeared negligible and was not corrected for. The loss
of flow through the extrathoracic airway walls (cheeks, mouth floor,
pharynx) was corrected for by dividing both Zrs and
Vpl/
by 1
Zrs/Zuaw, where Zuaw is the
impedance of upper airway walls. Zuaw was measured separately in
all subjects during Valsalva maneuvers according to Michaelson et al.
(13) by using the same equipment as above to record Pao and
. Finally, forced oscillation TGV
(TGVos) was computed from the
values of Xrs and Xpl according to Eq. 6.
TGVst was computed by linear
regression from the relationship between Vpl and Pao during the
occlusion by using an algorithm that minimized the influence of any
drift in Vpl (20). TGVst was
corrected for the difference between the mean lung volume during the
forced oscillation measurements and the volume when the shutter was
closed. All the TGV data, therefore, pertain to functional residual
capacity plus on-half the tidal volume.
The adequacy of the equipment and data analysis was tested by
performing measurements in a mechanical analog of the respiratory system made of an airway including a tube and resistive elements (fine-mesh metal screens) connected to a rigid box (5.0 × 23.6 × 48.3 cm), a wall of which presented a
530-cm2 opening covered by a
slightly stressed rubber membrane mimicking the tissues. TGV, computed
by assuming adiabatic compression (K = 1.4 barometric pressure), was highly reproducible but was slightly above the actual volume of gas (+7%), suggesting that gas compression in the analog was not quite adiabatic at our oscillation frequencies.
Of 79 measurements in 15 subjects, 2 measurements were rejected because
TGVst was unreliable (irregular
and looping Vpl-Pao relationship); although, in some instances, the
subjects breathed irregularly and Xrs or Xpl exhibited a marked
variability at low frequencies, no measurement was rejected on the
basis of the forced oscillation data.
 |
RESULTS |
An example of the raw reactance data is shown in Fig.
1. The main results are summarized in Table
1.
TGVos derived from the forced
oscillation data obtained over the entire frequency range was highly
correlated with TGVst but was
11.8% larger (4.28 ± 0.80 liters compared with 3.83 ± 0.73 liter, P < 0.001). The differences were much less when the data at the lowest frequencies were discarded, and they averaged +2.4% (P < 0.05)
and
1.6% (not significant) within the 6- to 29- and 9- to 29-Hz
frequency ranges, respectively. Discarding the highest frequencies did
not improve the agreement. It is also in the 6- to 29- and 9- to 29-Hz
ranges that the SD of the differences was the lowest. The 95%
"limits of agreement" [mean ± tSD of differences, with
t = 1.99 for 76 degrees of freedom (3)] were +0.79 to
0.61 liter for 6-29
Hz and +0.70 to
0.82 liter for 9-29 Hz. Variance analysis
showed that the differences varied significantly among subjects
(P < 0.001 within all frequency ranges). Mean individual data at 6-29 Hz are presented in Table 2: differences ranged from
0.33 to
+0.65 liter and were significantly different from 0 in 7 of 15 subjects. In terms of the unsigned relative differences
(100 · |TGVos
TGVst|/TGVst),
the 95% limits (mean ± 1.67 SD) were 17.3 and 18.7% for 6-29
and 9-29 Hz, respectively (Table 1). The relationship between
TGVos at 6-29 Hz and
TGVst, and the relationship
between their difference and their mean, are shown in Fig.
2. Whatever the frequency range, there was
no significant correlation between the differences and the means (Table
1).

View larger version (13K):
[in this window]
[in a new window]
|
Fig. 1.
Respiratory (Xrs) and plethysmographic (Xpl) reactance data as a
function of frequency ( f ).
Values are means ± SD (when larger than symbol) of 30 one-second
data blocks from a single measurement in a representative subject.
|
|
View this table:
[in this window]
[in a new window]
|
Table 1.
Forced oscillation TGV values derived from data over different
frequency ranges and compared with standard TGV values
|
|

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 2.
Forced oscillation thoracic gas volume
(TGVos) computed from 6- to
29-Hz data and standard TGV
(TGVst;
n = 77 measurements). Top: relationship between 2 estimates;
line of identity (dashed line) and regression line (solid line) are
shown. Bottom: relationship between
differences and mean of 2 estimates; mean of differences and 95%
limits of agreement are shown.
|
|
As noted in MATERIALS AND METHODS, the
above-mentioned data were obtained after Zrs and Vpl/
were corrected for the flow lost through upper airway walls. Omitting
that correction did not appreciably change the results. For instance,
within the 6- to 29-Hz frequency range, uncorrected
TGVos was 4.02 ± 0.61 liters (2.5% larger than with the correction), and the differences to TGVst averaged 0.19 ± 0.37 liter (95% limits of agreement: +0.92 to
0.54 liter).
 |
DISCUSSION |
Differences between estimates of a biological quantity by using two
methods include the systematic and random errors of both methods, as
well as biological variability when the measurements are not
simultaneous (3). In this study, a number of potential sources of
differences were absent or minimized: the two methods were based on the
measurements of the same variables with the use of the same transducers
so that static calibration errors would similarly affect
TGVos and
TGVst. The measurements were made
in close succession in subjects in the same posture, and any change in
lung volume was measured and corrected for. On the other hand, the
frequency ranges of the two measurements differed considerably, which,
although the frequency responses of the transducers and of the
plethysmograph appeared satisfactory, could be responsible for some
instrumental differences. The measurement of TGV by the method of
DuBois et al. (6), taken as a reference in this study, involves
substantial and easily measurable Pao and Vpl variations; moreover, our
linear regression method with drift correction ensures a highly
accurate determination of the slope of the Vpl-Pao relationship (20).
The physiological or pathological factors that may bias TGVst have been extensively
studied in the past (1, 2, 4, 5, 10, 25, 26). Bohadana et al. (4)
observed in healthy subjects that total lung capacities derived from
TGVst decreased by ~200 ml from
low (0.5-Hz) to medium (2.2-Hz) panting frequencies. Computer
simulation eliminated intrathoracic and extrathoracic airway wall
compliance as potential factors but suggested as a remote possibility a
combination of mechanical inhomogeneity and pleural pressure
nonuniformity. Abdominal gas compression could be responsible for some
over- or underestimation of TGV, depending on the relative phase of
abdominal pressure and PA
changes (5, 10). The error, however, should not exceed 150 ml at normal lung volumes (1). On the basis of the above considerations, we expect
the contribution of errors in
TGVst to the observed differences
not to exceed ±0.2 liter. Assuming the random errors in
TGVos associated with the
experimental noise to be on the same order of magnitude, one would
expect most of the differences between the two estimates to be within
±0.4 liter. Actually, the differences were outside that range in 20 of 77 measurements (6-29 Hz). In addition, differences between the
mean TGVos and the mean
TGVst (which should be little
affected by random errors) were above +0.4 liter in 3 of 15 subjects
(Table 2). This suggests the presence of some systematic cause of error
in TGVos in some subjects. Below we examine the factors that could be responsible for the discrepancies.
A crucial assumption in the computation of
TGVos is that Xaw increases
linearly with frequency (Eq. 3).
This will only be the case if Iaw is constant and if the effects of
other potential determinants of Xaw (airway wall elasticity, mechanical
inhomogeneity) are negligible.
The fluid inertance of a tube depends not only on its geometry and gas
density but also on the velocity profile of the gas: the fluid
inertance is 33% lower when the velocity profile is blunt than when it
is parabolic (14). Because the velocity profile tends to get blunter
when the frequency increases (8), one may expect some decrease in Iaw
with increasing frequency. To assess this factor, we measured the
inertance of 20-cm-long rigid tubes with internal diameters similar to
those of the large airways (1.23-1.90 cm). We observed a regular
decrease in inertance amounting to 6-9% from 5 to 30 Hz. We did
some computer simulation to evaluate the influence of such a decrease
on the estimation of TGV. The model included an airway characterized by
Raw and Iaw, alveolar gas elasticity
(TGV/K), and tissues with their
resistance (Rti), compliance (Cti), and inertance (Iti). Iaw was made
to decrease linearly by 10% from 4 to 29 Hz. Zrs and
Vpl/
were computed at the usual frequencies and
analyzed in the usual manner. The relative error in
TGVos was independent of Raw and
TGV, depended little on Iti, but decreased with increasing Rti and
increased with increasing Iaw and Cti. For reasonable values of Rti (1 hPa · s · l
1),
Cti (0.04 l/hPa), and Iaw (0.015 hPa · s2 · l
1
at 4 Hz) (19), TGVos was
overestimated by 3.2, 7.8, and 19.8% within the 4- to 29-, 6- to 29-, and 9- to 29-Hz frequency ranges, respectively. These figures are
inconsistent with our observations (Table 1), particularly with the
fact that TGVos tended to decrease when the lower end of the frequency range was increased. It suggests that the frequency dependence of Iaw in the airways is less than what
we observed in tubes, probably because the velocity profile is already
rather blunt in the large airways during spontaneous breathing (8, 9).
Mechanical inhomogeneity of the respiratory system may be responsible
for amplitude and phase differences between local
PA values. In that situation,
instantaneous Vpl, as measured by plethysmography, is proportional to
the local PA values weighted by
the local TGV values [Vpl = (
PAi · TGVi)/(K )],
where index i designates any parallel
lung compartment (15). This will compromise the validity of
Eqs. 2-6 and the linear increase
in the apparent Xaw with frequency. Some computer simulation was made
to evaluate the corresponding error in
TGVos. The model included two
compartments in parallel, each with its Zaw, TGV, and lung tissue
compliance (CL,ti); it also
included a common airway and a common chest wall, but they were seen to
influence the error negligibly. Starting from the homogeneous situation
(Raw = 2 hPa · s · l
1,
Iaw = 0.01 hPa · s2 · l
1,
TGV = 2 liters, CL,ti = 0.05 l/hPa in both compartments), increasing one of the Raw values by a
factor of two decreased TGVos by
3.8-5.2%, depending on the frequency range; the error reached
18-19% when the ratio of local Raw was raised to five. On the
other hand, changing the distribution of TGV between the compartments
(1 and 3 liters, respectively) and changing one of the
CL,ti values by a factor of 5 had very little or no effect. It therefore appears that substantial
mechanical inhomogeneity of airway resistance may lead to an
underestimation of TGVos.
Another potential cause of error is airway wall elasticity. Its
consequence is that some of the flow oscillation entering the airway is
shunted to the intrathoracic space so that the flow in the peripheral
airway is less than in the central airways. The shunt is negligible as
long as airway wall impedance is large compared with that of the
peripheral lung but may become important at high frequencies in the
presence of peripheral airway obstruction (12). We analyzed that
situation with a model including a central airway with a resistance (Rc = 1 hPa · s · l
1)
and an inertance (Ic = 0.015 hPa · s2 · l
1),
a resistive peripheral airway (Rp),
CL,ti (0.1 l/hPa), TGV (4 liters), and a resistive (Rw = 1 hPa · s · l
1),
compliant (Cw = 0.1 l/hPa), and inertive (Iw = 0.002 hPa · s2 · l
1)
chest wall. Airway wall properties were modeled as a compliance (elastic bridge; Cb), shunting Rp and
CL,ti (23). With a Cb of 0.003 l/hPa, as suggested from dead-space volume changes over the vital
capacity range (24) and a low value of Rp (0.2 hPa · s · l
1),
as expected in normal subjects, the error in
TGVos was <1 ml; raising Rp to 1 hPa · s · l
1
TGVos was underestimated by 2.2, 3.2, and 8.2% within the 4- to 29-, 6- to 29-, and 9- to 29-Hz
frequency ranges, respectively; the corresponding figures were 13, 26.5, and 48%, respectively, when Rp was raised to 2 hPa · s · l
1.
From these computer simulations, we conclude that the method should
provide reasonably accurate results for moderate degrees of mechanical
inhomogeneities and airway wall shunting but not in the case of severe
abnormalities; these factors are, therefore, unlikely to be responsible
for the differences seen in some of our subjects.
Finally, it is postulated in Eq. 1
that Vpl is only related to PA
variations. Other potential factors during breathing are the changes in
gas temperature and
PH2O
in the airways and the variations in the respiratory exchange ratio
during the respiratory cycle (7). These factors are minimized by having
the subject rebreathe in a circuit in which the gas is maintained at
BTPS (7, 11). This is what we
attempted to do in this study. However, we observed that the
low-frequency respiratory component of Vpl during breathing was not
always perfectly in phase with
, but included a small
component in phase with volume. This suggested that the gas
conditioning was not perfect and that the above factors had not been
completely eliminated. We measured by linear regression of Vpl vs.
volume the amplitude of that component (G, dimensionless) and found it
to range from
0.013 to +0.027: a negative value of G indicates
that the inspired gas is too hot and cools down in the airways, and a
positive value indicates that it warms up or increases its
PH2O.
The relationship between TGVos
TGVst at 6-29 Hz and
G is shown in Fig. 3. The differences were
loosely but significantly correlated with G
(r =
0.441,
P < 0.001); a higher correlation
(r =
0.561) was seen
within the 4- to 29-Hz frequency range. These observations are
consistent with the expected effect on
TGVos of a small thermal component
in the Vpl-
relationship. Computer simulation taking
into account a time constant of 80 ms (18) for the change in gas
condition shows that a negative value of G is associated with a
frequency-dependent overestimation of
TGVos: with G =
0.01 and
for reasonable values of tissue properties (Rti = 1 hPa · s · l
1,
CL,ti = 0.04 l/hPa, Iti = 0.002 hPa · s2 · l
1),
TGVos would be overestimated by
0.08 and 0.05 liter at 4-29 Hz and 6-29 Hz, respectively. The
effect depends very much on the thermal time constant and may become
much larger if the latter is shorter. Imperfect gas conditioning may
therefore explain some of the variance in the differences. It is not
clear whether it is also responsible for the overestimation of
TGVos at 4-29 Hz (11.7%);
indeed, on the basis of the positive mean value of G (0.0034 ± 0.0093), TGVos should rather be
slightly underestimated on average. We have no other explanation to
offer, however, for that finding.

View larger version (10K):
[in this window]
[in a new window]
|
Fig. 3.
Relationship between differences of
TGVos and
TGVst and slope (G; dimensionless)
of relationship between plethysmographic measurement of body volume
changes (Vpl) and airway flow ( ) during breathing.
Significant correlation (r = 0.44) suggests that a thermal factor is responsible for some of
variance of TGVos TGVst.
|
|
To summarize our results, this study showed a close relationship but
not a perfect agreement between the values of TGV derived from forced
oscillation measurements and those obtained with the usual panting
maneuvers in normal subjects. The results depended slightly on the
frequency range over which the data were analyzed, which may, in part,
be related to imperfect gas conditioning to BTPS. Even with the optimal frequency
range, differences >10% were seen in ~30% of the measurements;
95% limits of agreement were about ±0.7 liter (disregarding the
small systematic bias) or 17%. These values are obviously too large
for applications requiring a very accurate measurement of TGV. They are
not so large, however, compared with the 95% confidence interval of
predicted functional residual capacity or total lung capacity in adults (1.96 residual SD above 1 liter) (21). The method could, therefore, have some usefulness in situations in which, for some reason, the
method of DuBois et al. (6) cannot be used and in which severe
mechanical inhomogeneity and airway wall shunting are not expected.
 |
ACKNOWLEDGEMENTS |
The authors are grateful to B. Clement for typing the manuscript
and to M. C. Rohrer for preparing the figures.
 |
FOOTNOTES |
Address for reprint requests: R. Peslin, Unité 14 INSERM,
Physiopathologie Respiratoire, CO 10, 54511 Vandoeuvre-les-Nancy cedex,
France (E-mail: rpeslin{at}u14.nancy.inserm.fr).
Received 4 June 1997; accepted in final form 4 November 1997.
 |
REFERENCES |
1.
Bégin, P.,
and
R. Peslin.
Mesure pléthysmographique du volume gazeux intrathoracique. Un retour aux hypothèses.
Bull. Eur. Physiopathol. Respir.
19:
247-251,
1983[Medline].
2.
Bégin, P.,
and
R. Peslin.
Influence of panting frequency on thoracic gas volume measurements in chronic obstructive pulmonary disease.
Am. Rev. Respir. Dis.
130:
121-123,
1984[Medline].
3.
Bland, J. M.,
and
D. G. Altman.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
i:
307-310,
1986.
4.
Bohadana, A. B.,
R. Peslin,
B. Hannhart,
and
D. Teculescu.
Influence of panting frequency on plethysmographic measurements of thoracic gas volume.
J. Appl. Physiol.
52:
739-747,
1982[Abstract/Free Full Text].
5.
Brown, R.,
F. G. Hoppin,
R. H. Ingram,
N. A. Saunders,
and
E. R. McFadden.
Influence of abdominal gas on the Boyle's law determination of thoracic gas volume.
J. Appl. Physiol.
44:
469-473,
1978[Abstract/Free Full Text].
6.
DuBois, A. B.,
S. Y. Botelho,
G. N. Bedell,
R. Marshall,
and
J. H. Comroe, Jr.
A rapid plethysmographic method for measuring thoracic gas volume: a comparison with a nitrogen washout method for measuring functional residual capacity in normal subjects.
J. Clin. Invest.
35:
322-326,
1956.
7.
DuBois, A. B.,
S. Y. Botelho,
and
J. H. Comroe.
A new method for measuring airway resistance in man using a body plethysmograph: values in normal subjects and in patients with respiratory disease.
J. Clin. Invest.
35:
327-335,
1956.
8.
Finucane, K. E.,
S. V. Dawson,
P. D. Phelan,
and
J. Mead.
Resistance of intrathoracic airways of healthy subjects during periodic flow.
J. Appl. Physiol.
38:
517-530,
1975[Abstract/Free Full Text].
9.
Fredberg, J. J.,
and
J. Mead.
Impedance of intrathoracic airway models during low-frequency periodic flow.
J. Appl. Physiol.
47:
347-351,
1979[Abstract/Free Full Text].
10.
Habib, M. P.,
and
L. A. Engel.
Influence of the panting technique on the plethysmographic measurements of thoracic gas volume.
Am. Rev. Respir. Dis.
117:
265-271,
1978[Medline].
11.
Jaeger, M. J.,
and
A. B. Otis.
Measurement of airway resistance with a volume displacement body plethysmograph.
J. Appl. Physiol.
19:
813-820,
1964[Abstract/Free Full Text].
12.
Mead, J.
Contribution of compliance of airways to frequency-dependent behavior of lungs.
J. Appl. Physiol.
26:
670-673,
1969[Free Full Text].
13.
Michaelson, E. D.,
E. D. Grassman,
and
W. R. Peters.
Pulmonary mechanics by spectral analysis of forced random noise.
J. Clin. Invest.
56:
1210-1230,
1975.
14.
Pedley, T. J.,
R. C. Schroter,
and
M. F. Sudlow.
Gas flow and mixing in the airways.
In: Bioengineeering Aspects of The Lung, edited by J. B. West. New York: Dekker, 1977, vol. 3, chapt. 3, p. 163-265. (Lung Biol. Health Dis. Ser.)
15.
Peslin, R.
Theoretical analysis of airway resistances on an inhomogeneous lung.
J. Appl. Physiol.
24:
761-767,
1968[Free Full Text].
16.
Peslin, R.,
and
C. Duvivier.
Partitioning of airways and respiratory tissues mechanical impedances by body plethysmography.
J. Appl. Physiol.
84:
553-561,
1998[Abstract/Free Full Text].
17.
Peslin, R.,
C. Duvivier,
and
C. Gallina.
Total respiratory input and transfer impedances in humans.
J. Appl. Physiol.
59:
492-501,
1985[Abstract/Free Full Text].
18.
Peslin, R.,
C. Duvivier,
M. Vassiliou,
and
C. Gallina.
Thermal artifacts in plethysmographic airway resistance measurements.
J. Appl. Physiol.
79:
1958-1965,
1995[Abstract/Free Full Text].
19.
Peslin, R.,
and
J. J. Fredberg.
Oscillation mechanics of the respiratory system.
In: Handbook of Physiology. The Respiratory System. Mechanics of Breathing. Bethesda, MD: Am. Physiol. Soc., 1986, sect. 3, vol. III, pt. 1, chapt. 11, p. 145-178.
20.
Peslin, R.,
C. Gallina,
and
M. Rotger.
Methodological factors in the variability of lung volume and specific airway resistance measured by body plethysmography.
Bull. Eur. Physiopathol. Respir.
23:
323-327,
1987[Medline].
21.
Quanjer, P. Standardized lung function testing.
Bull. Eur. Physiopathol. Respir. 19, Suppl. 5: S45-S51, 1983.
22.
Rotger, M.,
R. Farré,
R. Peslin,
and
D. Navajas.
T model partition of lung and respiratory system impedances.
J. Appl. Physiol.
78:
938-947,
1995[Abstract/Free Full Text].
23.
Rotger, M.,
R. Peslin,
C. Duvivier,
D. Navajas,
and
C. Gallina.
Density dependence of respiratory input and transfer impedances in humans.
J. Appl. Physiol.
65:
928-933,
1988[Abstract/Free Full Text].
24.
Shepard, R. H.,
E. J. M. Campbell,
H. B. Martin,
and
T. Enns.
Factors affecting the pulmonary deadspace as determined by single breath analysis.
J. Appl. Physiol.
11:
241-244,
1957[Abstract/Free Full Text].
25.
Shore, S.,
J. Milic-Emili,
and
J. G. Martin.
Reassessment of body plethysmographic technique for the measurement of thoracic gas volume in asthmatics.
Am. Rev. Respir. Dis.
126:
515-520,
1982[Medline].
26.
Stanescu, D. C.,
D. Rodenstein,
M. Caubergh,
and
K. P. Van de Woestijne.
Failure of body plethysmography in bronchial asthma.
J. Appl. Physiol.
52:
939-948,
1982[Abstract/Free Full Text].
JAP 84(3):862-867
0161-7567/98 $5.00
Copyright © 1998 the American Physiological Society