|
|
||||||||
1 Department of Physiology, University of Rome "Tor Vergata," 1-00173 Rome, Italy; and 2 Xitron Technologies, Inc., San Diego, California 92121
De Lorenzo, A., A. Andreoli, J. Matthie, and P. Withers.
Predicting body cell mass with bioimpedance by using theoretical methods: a technological review. J. Appl.
Physiol. 82(5): 1542-1558, 1997.
The body cell
mass (BCM), defined as intracellular water (ICW), was estimated in 73 healthy men and women by total body potassium (TBK) and by bioimpedance
spectroscopy (BIS). In 14 other subjects, extracellular water (ECW) and
total body water (TBW) were measured by bromide dilution and deuterium
oxide dilution, respectively. For all subjects, impedance spectral data
were fit to the Cole model, and ECW and ICW volumes were predicted by
using model electrical resistance terms RE and
RI in an equation derived from Hanai mixture theory,
respectively. The BIS ECW prediction bromide dilution was
r = 0.91, standard error of the
estimate (SEE) 0.90 liter. The BIS TBW prediction of deuterium space
was r = 0.95, SEE 1.33 liters. The BIS
ICW prediction of the dilution-determined ICW was
r = 0.87, SEE 1.69 liters. The BIS ICW
prediction of the TBK-determined ICW for the 73 subjects was
r = 0.85, SEE = 2.22 liters. These
results add further support to the validity of the Hanai theory, the
equation used, and the conclusion that ECW and ICW volume can be
predicted by an approach based solely on fundamental principles.
bioimpedance spectroscopy; extracellular water; total body
water
THE FIRST SUCCESSFUL VALIDATION of extracellular water
(ECW), TBW (total body water), and intracellular water (ICW) by using bioimpedance spectroscopy (BIS) methods was reported in 1992 (30). BIS,
which implies fitting complex impedance (Z) data measured at multiple
frequencies to a biophysical model, has been used extensively in
biophysics and is the technique from which derive all the underlying
theories for body impedance analysis (BIA) (48). Although a number of
studies have been reported that used the methods employed in this
study, the BIS principles and the rationale for the methods employed
remain poorly understood. Partly because of the complex nature of BIS,
there has been very little cross-transfer of BIS information to the
field of human body composition; furthermore, these methods have not
been fully reported by the investigators responsible for their
development (30). As a result, these methods have been only partially
reported with insufficient underlying discussion.
Several single-frequency Z studies have reported a prediction of body
cell mass (BCM; Refs. 3, 42), but the scientific bases of the
approaches used were not well supported, and the relationships reported
may have simply been a result of high intercorrelation between
variables (48). Results should be associated with ICW or total body
potassium (TBK) rather than BCM, because BCM is a concept that can only
be defined by ICW or TBK (33). Any relationship between Z and BCM can
only emerge from the relationship between Z and cell volume. What
becomes important then is the best approach of predicting cell volume
with Z.
This study reports the relationship between a BIS-predicted ICW and a
TBK-predicted ICW, and the relationship between a BIS-predicted volume
of ECW, ICW, and TBW compared with dilution-determined volumes. This manuscript also provides a description of
the principles of BIS and a discussion of the simplified single- and
dual-frequency methods relative to well-known BIS principles.
Physical Principles
If the complex Z [resistance (R) and reactance (X)] of
skeletal muscle tissue is measured, and the
f varies from low to high, a series of
values is derived that can be represented by complex points. The curve
formed by these points is called an impedance locus, and its shape is a
result of the electrical and structural characteristics of the tissue
(5). The mathematical model that is used most often to describe both
theoretical and experimental data on skeletal muscle tissue is known as
the Cole model. It produces a semicircular relationship between R and
X, with a depressed center when plotted (5) (Fig. 3).
Modeling is considered essential, because it is the only means of
independently analyzing the individual components of a heterogeneous
material system (5, 25). The Cole model can be viewed as the equivalent
electrical circuit shown in Fig. 4.
To accurately predict volume, the mixture effects need to be accounted
for, because the relationship between R and body water volume is
nonlinear (9, 17). These mixture effects are greater at LF because the
conductor (i.e., ECW) represents only 25% of the total body volume
compared with a concentration of nonconductor of 75%. At
HF, the concentration of nonconductor is much less (e.g., 40%). Good
examples of the mixture effects are the change in resistivity ( A group of 87 healthy Italian men (n = 77) and women (n = 10), ages
21-57 yr, volunteered to participate in this study. Written informed consent was obtained from all participants. The study protocol
was approved by the Medical Ethical Committee of the University of
"Tor Vergata" in Rome.
On arriving in the morning in an overnight-fasted state, subjects were
weighed in swimming clothes, and body weight (Wt) was measured with a
standard balance to the nearest 0.05 kg. Body height (Ht) was measured
with a stadiometer to the nearest 1 mm. After the measurements of Wt
and Ht were made, and still in the fasting state, all 87 subjects had
their 40K measured by a whole body
counter, formed by a cell 2.5 m wide and 3 m high of 10-cm-thick lead
bricks, the door to which was formed by a 22-cm-thick iron slab. The
room was continuously ventilated. A single 20.3 × 10.2 thallium-activated sodium iodine crystal was positioned above the
subject, who was measured in a sitting position and dressed in only
paper pajamas. TBK was calculated as
40K * 8,474.6 (11). The
correlation of variation (CV) for TBK was found to be between 2 and
3%. ICW was computed from TBK, assuming that potassium is only present
in the intracellular fluid and assuming a potassium concentration in
the intracellular fluid of 150 mmol/l (11).
For 14 of the men, out of the total sample of 87 men and women, TBW and
ECW were measured by dilution methods. Deuterium oxide (D2O) was used for the
determination of TBW, and sodium bromide (NaBr) was used for the
determination of ECW. While still in a fasted state and after the Z
measurements were taken, the subjects drank 50 g of a solution
containing 10 g of D2O (99.8%;
Carlo Erba) and 1.3 g of NaBr (Carlo Erba) in 38.7 g of tap water.
After a 3-h equilibration time, urine was collected for the
determination of D2O, and a venous
blood sample was taken for determining plasma NaBr. The subjects
remained in a fasted state throughout the equilibration period and
refrained from voiding. No baseline measurement samples of either urine
or plasma were taken. Enrichment of
D2O in the urine sample was
measured after sublimation with infrared spectrophotometry, as
described by Lukaski and Johnson (24). A correction of 5% for
nonaqueous dilution was used (11). The venous blood sample was
centrifuged (3,000 revolutions/min for 15 min) to separate the plasma.
The plasma was then stored in sealed plastic tubes at After the measurement of Wt, Ht, and TBK; before ingestion of the
D2O and NaBr solution; and with
subjects still in a fasted-state; single wrist-to-ankle (i.e., whole
body) complex Z measurements were taken in all 87 subjects by using a
BIS analyzer (model 4000B, Xitron Technologies, San Diego, CA). R and X
were measured, and the corresponding Z, phase ( The Z and The Excel program was used for the statistical analysis. In addition to
the descriptive statistics, the Pearson's product moment correlation
(r) and standard error of estimate
(SEE) statistics were computed. Bland-Altman plots were also
constructed to display the individual subject differences between the
BIS-predicted water volumes and those determined by TBK and dilution
methods.
Table 1 displays the physical
characteristics of the two subject groups. Table
2 provides the Cole modeling results for the two subject groups. Of the 87 subjects, according to the criteria rating the fit to the Cole model (APPENDIX a), 21 subjects were rated as 0, 64 were rated as 1, and the remaining 2 were rated as 2. The mean correlation of fit to the
Cole model, using scalar Z, was 0.998. With the use of the constants
for men from a previous study (47), the dilution ECW was predicted as
r = 0.91, SEE = 0.90 liters, with a
mean of 21.03 liters and a mean difference of 2.69 liters. Dilution TBW
was predicted as r = 0.95, SEE = 1.33 liters, with a mean of 41.02 liters and a mean difference of
Table 1.
Descriptive characteristics of TBK and dilution study groups
Table 2.
Cole modeling characteristics of TBK and dilution study groups
Table 3.
ECW, TBW, and ICW determined by D2O, NaBr, and BIS
-dispersion and is caused by cell
membrane capacitance
(Cm) (40) (Fig.
1). With direct current (DC), there is no
conduction through a capacitor. Thus, in the LF range of
-dispersion, there is minimal conduction through the cells because
of the high Z of the
Cm, and
conductivity is governed primarily by the properties of the ECW. As
f increases into alternating current
(AC), the Z of the
Cm decreases,
allowing more current to flow into the ICW compartment. Because of the
change in polarity that occurs with AC current, the cell membrane
charges and discharges the current at the rate of the
f. The Z decreases with
f, because the amount of conducting
volume is increasing. At higher frequencies (HF), the rate of charge
and discharge becomes such that the effect of the
Cm diminishes to
insignificant proportions, and the current flows through both the ECW
and ICW compartments in proportions dependent on their relative
conductivity and volumes (5) (Fig. 2).
Thus, at both very low and very high frequencies, the overall Z is
essentially independent of the
Cm, whereas at
the mid- or characteristic frequency
(fc),
the dependence on the value of the
Cm is at a
maximum.
Fig. 1.
Log dielectric constant of muscle tissue vs. log frequency in the
-,
-, and
-dispersion regions. Recreated in vitro data from H. Schwan, Electrical properties of tissue and cell suspensions. In:
Advances in Biological and Medical
Physics, edited by J. H. Lawrence and C. A. Tobias. New
York: Academic, 1957, vol. 5, p. 147-209. Used by
permission.
[View Larger Version of this Image (20K GIF file)]
Fig. 2.
Diagram representing high-frequency and low-frequency current
distribution in cell suspensions.
[View Larger Version of this Image (31K GIF file)]
Fig. 3.
Impedance locus: reactance vs. resistance in
-dispersion region.
R
and Ro, resistances at 100 MHz and 1 Hz, respectively; fc, characteristic
frequency.
[View Larger Version of this Image (7K GIF file)]
Fig. 4.
Equivalent electrical circuit analogous to Cole model.
RE is component value in ohms;
RI is aggregate component value
in ohms; and membrane capacitance
(Cm) is
aggregate component value in farads.
[View Larger Version of this Image (14K GIF file)]
)
that occurs with a change in hematocrit (14) and that plasma (i.e.,
ECW) is four- to sixfold more conductive than is skeletal
muscle tissue measured at 1 kHz (14). At 1 kHz, there is little
conduction through the cells; thus the
should be similar to that of
plasma. It is dramatically increased because the cells are
nonconductive at LF and restrict the flow of current. Hanai (17)
developed a theoretical equation that describes the effect on the
apparent conductivity of a conducting material having a restricting
concentration of nonconductive material in suspension. Hanai postulated
that the theory could be applied to tissues with nonconductive
concentrations ranging from 10 to 90%. To employ his theory, we have
constructed an equation that considers the ECW to be such a medium at
LF, where the ECW is the conductive material, and all remaining items
(including ICW because it is surrounded by cell membrane) in the body
are the restricting nonconductive material. At HF, the combination of both ECW and ICW forms the conductive medium, and all remaining items
in the body form the restrictive material. Previous results have
supported that this theory can be used in vivo to predict ECW, TBW, and
ICW volume (35, 47).
20°C
until analysis. Bromide enrichment was measured in the plasma by
high-pressure liquid chromatography (32). The accuracy of the NaBr
measurements by this technique is considered to be within 1%
(50). ECW was calculated by using a 10% correction for
nonextracellular distribution and 5% for the Donnan equilibration (11,
50). ICW was calculated as the difference between TBW and ECW.
), was computed from
R and X at 21 f ranging from 1 kHz to
1.248 MHz. The measurements were taken within the first
several minutes after the subjects assumed a supine position. No
correction was made for orthostatic fluid shifts. The measurements were
taken on the left side of the body, with the use of disposable
electrocardiogram electrodes (5 cm2; 3M, Minneapolis, MN) and in
accordance with the standard wrist-to-ankle protocol (47). Data were
transmitted directly from the analyzer into an ASCII format file via an
RS232 interface to a personal computer and controlled by the software
program supplied with the device.
spectra data were fit to an enhanced Cole model (5) to
account for any time delay (Td)
effects (see APPENDIX a), using the
nonlinear curve-fitting software developed for the device. The ECW and
ICW volumes were predicted by using modeled
RE and
RI values in equations
formulated previously (47) from Hanai mixture theory (17)
(APPENDIX b). The BIS TBW was
calculated as ECW + ICW. The constants used for
kECW (i.e., men = 0.306, women = 0.316) and
k
(men = 3.82, women = 3.40) had been scaled to
D2O and NaBr data collected in a
previous study (47). Although only the constants for men
were used in this study, the constants for women were included for
discussion purposes because a previous study discovered a gender
difference (47). When reference ECW and TBW data are available for only
one gender, the software automatically and arbitrarily scales the other
gender's constant by the same percentage difference discovered
previously. Further research needs to be done to determine whether
there truly is a gender difference in these terms. Expressing the above
kECW and
k
terms as
apparent ECW and ICW resistivity
(
ECW and
ICW, respectively), they become
214 and 206, and 824 and 797 for
ECW and
ICW in men and women,
respectively. These terms will be expressed as apparent
in the
remainder of the document. The male constants from a previous study
(47) listed above were used to predict the dilution ECW, TBW, and ICW
of the 14 men. The TBK-determined ICW of these 14 men was also
predicted by both BIS- and dilution-ICW volumes. Then, new constants
for men were computed from the dilution volumes measured on the sample
of 14 men, as described in APPENDIX b.
The new
ICW constant was then
used to predict the BIS-ICW volume for the other 73 men and women and
cross-validated against their TBK- determined ICW.
4.46 liters. Dilution ICW (as TBW
ECW) was predicted as r = 0.87, SEE = 1.69 liters, with a
mean of 19.99 liters and a mean difference of
7.14 liters (Table
3).
TBK
Dilution, Men
Men
Women
Combined
n
63
10
73
14
Age, yr
37.92 ± 9.60
25.33 ± 3.29
36.20 ± 9.99
28.57 ± 6.64
Height, cm
174.23 ± 7.43
164.55 ± 7.88
172.91 ± 8.20
175.25 ± 6.50
Weight, kg
74.99 ± 8.71
56.40 ± 4.86
72.44 ± 10.47
74.80 ± 8.83
TBK-ICW, liters
25.08 ± 3.19
17.54 ± 2.60
24.05 ± 4.06
28.35 ± 3.19
Potassium, g
147.13 ± 18.72
102.90 ± 15.23
141.07 ± 23.78
166.30 ± 18.71
ECW, liters
18.34 ± 2.04
TBW, liters
45.48 ± 3.87
ICW, liters
27.13 ± 2.63
Values are means ± SD. TBK, total body potassium; n, no.
of subjects; ICW, intracellular water; ECW, extracellular water; TBW,
total body water.
TBK
Dilution, Men
Men
Women
Combined
n
63
10
73
14
r of fit
0.9978 ± 0.001
0.9974 ± 0.001
0.9978 ± 0.001
0.9970 ± 0.001
RE,
582.09 ± 47.86
742.81 ± 44.56
604.13 ± 72.81
577.71 ± 44.87
RI,
1,109.04 ± 138.72
1,504.73 ± 209.42
1,163.20 ± 202.79
1,020.42 ± 81.33
Cm, nF
2.32 ± 0.43
1.27 ± 0.26
2.18 ± 0.55
2.24 ± 0.30
0.70 ± 0.02
0.68 ± 0.03
0.70 ± 0.03
0.68 ± 0.01
fc, kHz
57.02 ± 8.39
80.14 ± 17.22
60.19 ± 12.83
61.96 ± 5.95
Td, ns
20.66 ± 8.46
32.40 ± 9.43
22.26 ± 9.49
3.27 ± 4.37
Values are mean ± 1 SD. n, No. of subjects; r,
correlation of fit; RE and RI, Cole model
electrical resistance terms used to predict volume of ECW and ICW,
respectively; Cm, membrane capacitance;
, exponent;
fc, characteristic frequency; Td,
time delay.
Subject
Br
BIS-ECW
D2O
BIS-TBW
D2O-Br
BIS-ICW
1
20.61
22.93
49.29
43.96
28.68
21.03
2
16.47
18.99
43.9
38.09
27.43
19.10
3
17.86
18.84
42.7
40.36
24.84
21.52
4
22.81
24.45
48.84
45.91
26.03
21.45
5
16.91
20.39
44.33
41.05
27.42
20.66
6
17.41
19.96
43.71
40.18
26.30
20.21
7
16.22
19.53
42.60
38.46
26.38
18.93
8
20.21
23.73
49.40
44.70
29.19
20.97
9
17.19
20.38
43.43
37.87
26.24
17.49
10
15.23
17.83
40.28
35.05
25.05
17.23
11
20.52
22.31
51.42
45.26
30.90
22.95
12
18.71
20.79
41.19
36.00
22.48
15.21
13
17.25
20.97
42.81
40.15
25.56
19.18
14
19.41
23.33
52.78
47.30
33.37
23.97
Mean ± SD
18.34 ± 2.11
21.03 ± 2.02
45.48 ± 4.01
41.02 ± 3.84
27.13 ± 2.73
19.99 ± 2.34
BIS, bioimpedance spectroscopy.
The new
ECW and
ICW constants computed from the
dilution sample of 14 men were 174.32 and 1,177.94, respectively. For
discussion purposes, the computed contants for women became 167.8 and
1,139.34 for
ECW and
ICW, respectively. Using the
new
ICW constant for men, the
prediction of the TBK ICW for the 73 subjects with BIS ICW was
r = 0.85, SEE = 2.22 liters, with
virtually no mean difference (i.e., 0.08 liter). When the TBK ICW was
predicted by gender (men = 63, women = 10) the correlations and SEE
were similar to that of the total group, but there was a slight mean
difference (i.e., 0.15 and
0.38). For the 14 men, the
correlation and SEE values for the BIS- and dilution-predicted ICW and
the TBK-predicted ICW were r = 0.56 and 0.57, and SEE = 2.68 and 2.32 liters, respectively.
To determine the effect that scaling of
ECW and
ICW had on the correlation and
SEE, the new constants for men were used to repredict the dilution ECW,
TBW, and ICW on the 14 male subjects. The correlation and SEE remained
identical for ECW (i.e., r = 0.91, SEE = 0.90 liter) with no mean difference. For TBW, the correlation
decreased slightly (i.e., r = 0.95-0.94) and the SEE increased slightly (i.e., 1.33-1.41
liters) with no mean difference. For ICW, the correlation decreased
slightly (i.e., r = 0.87-0.80) and the SEE decreased very slightly (i.e., 0.01 liter) with no mean
difference. Thus,
ECW is purely
a scalar and has no affect on correlation or SEE for ECW. Similarly,
ICW is effectively a scalar,
since changing it only slightly alters the prediction of ICW because
the nonlinearity is slight. Figures
5-7 display the plotted differences between the BIS-predicted ECW, ICW, and TBW volumes and the dilution-predicted volumes. Figure 8
displays the plotted differences between the BIS and the TBK-predicted ICW. The ECW prediction was achieved by using the exponent 1.5 predicted by Hanai theory.
is
dependent on the concentration of nonconductor present in a
mixture, giving rise to an empirical exponent ranging from
1.43 for very small spheres to 1.53 for packed cylinders (7, 17). Hanai
theory predicts an exponent, 1.5. The exponent 1.5 was recently
confirmed in vitro in human blood (9). A linear equation computed from
multiple-regression analysis is not well suited to nonlinear effects.
Thus it did not seem prudent to solve for a five-dimensional nonlinear
biophysical model (i.e., Cole) and then use an overly simplistic volume
theory (i.e., Ht2/R) that assumes
only one material is being measured. Predicting volume with an equation
formed by scientific principles would enhance its utility and address
the error sources directly rather than accounting for them
statistically, which offers no scientific explanation.
Although the successful prediction of ECW and ICW volume with the
equation used in this study has been reported (35, 47), it had not been
reported that when we regressed an exponent against NaBr space (47),
the highest correlation was achieved by using the exponent 1.5 predicted by Hanai theory. This finding strongly suggests the presence
of mixture effects, and the strong predictions by using the exponent
1.5 (10, 35, 47) support the validity of Hanai's theory. As a
spherical theory developed in the emulsion sciences, the reasons why
Hanai's theory should not work are many, but they do not explain the
strength of prediction this theory provides or the emergence of the
exact theoretical exponent. In the absence of a more applicable theory,
we use the developed equation, because we believe the errors of not
accounting for mixture effects are greater than the inadequacy of the
theory.
bromide).
Heavy horizontal line, mean difference (bias); lighter lines, mean ± 2 times SD of differences.
Because the strength of the BIS prediction (r and SEE) is independent of scaling, and TBW is determined by ECW + ICW, a good prediction of a dilution TBW would suggest a strong ECW and ICW prediction. This would only not be the case if there were exactly offsetting errors in the ECW and ICW prediction. Thus it is probable that the BIS prediction of ICW was better than that of dilution because the r and SEE were better for the BIS prediction of TBW than that of ICW. The inaccuracies of determining ICW by dilution have not been adequately discussed, nor is it clear whether the errors are additive or propagated. For the 14 subjects, both the dilution and BIS-predicted ICW were poorly correlated to the TBK-ICW. That the TBK-ICW prediction improved substantially for the larger sample of 73 subjects suggests that the poor correlations were caused by outlying data in a small sample. That both BIS and dilution ICW were highly correlated with each other, as well as both poorly correlated to the TBK ICW in the 14 subjects, suggests that the lower correlation between BIS ICW and TBK ICW in the entire sample was due to the error in TBK rather than BIS. Nevertheless, the strength of the discovered relationships among the BIS ICW and dilution and TBK ICW suggests that BCM can be determined by BIS. The only other equation derived from Hanai theory, or any mixture theory for that matter, was never validated, and its sensitivity to change was poor (9). We believe the results achieved in this and other studies can be attributed to viewing the body as having three compartments (i.e., ECW, ICW, and the remainder) rather than only the two used previously (i.e., ECW, ICW) (9). We also believe this to be attributed to the fact that the Hanai equation describes the effect on conductivity of the material, not the overall conductance. Thus, its use is volume dependent. To apply mixture theory, total volume must be known and is provided by body Wt/body density (Db). Db varies between individuals, but the range is generally within 1-1.07 kg/l (20). The effect on
ECW in this range is ±1%,
because it is only dependent on the cube root of
Db. We do not use body Wt as a
fudge factor to improve the correlation (8) but rather as a
theoretically required term to measure of total body volume. Like
Db, body Wt is expressed in cube
root form; thus its contribution to the prediction of body water is
reduced by 2/3.
The fact that changing
ECW had
no effect on correlation or SEE for ECW, and only slightly for ICW when
ICW was changed demonstrates the scaling nature of these constants and that they have no effect on
the scientific relationship between the BIS and dilution volumes. Large
offsets were observed when the constants derived from a previous study
were used (47), but these constants, which were derived from
D2O and NaBr (47), have been
cross-validated (10, 35). As such, it is troubling that
ECW would decrease by 19% and
ICW increase by 30%. This
difference could have been caused by error in the BIS method. However,
the high correlations and low SEE values observed, and the fact that
the sample studied was similar to those samples used to calibrate (47)
and cross-validate (35) these terms, suggest otherwise.
Because subjects were healthy, it is unlikely that there were any major
differences in ion concentration. Even so, a 5 mmol change in ion has
been found to affect ECW only 1-2% and ICW only 4-5% (38).
Different dilution methods produce differently sized ECW and TBW
spaces, e.g., sulfate
(35SO4)
space being typically 20% smaller than NaBr space (11). Thus, a
ECW calibrated to NaBr would predict an ECW space scaled 20% larger than
35SO4
space. We have noted (28) that Van Marken Lichtenbelt et al.
obtained slightly higher
r2 and lower SEE values
by using the methods described in this study, but
D2O space was underpredicted by
6.3 liters, and the NaBr space overpredicted by 3.0 liters. The
NaBr-to-D2O space ratio was in the
expected range of 0.40 and 0.42 in this study (Table 1) and the study
reported by Van Marken Lichtenbelt (48), respectively. In
contrast, the BIS ECW-TBW ratio predicted in this study by the previous
constants was 0.51 (Table 3). This supports that the
ECW constant computed
previously (47) may be scaling ECW too large, but this
would equally occur if ICW were underestimated. That ICW, and thus TBW,
may have been scaled too low was supported by the finding that the
percent TBW of body Wt was 61% by dilution and 55% by BIS (Tables 1
and 3). Further evidence that the new
ICW constant may have validity
was that TBK-ICW, which was independent from dilution, was predicted
with very little mean difference. It is of concern that the previously
determined constants are predicting ECW and TBW with little mean
difference at some laboratories but not others. There is nothing
apparent in the dilution methods used in this study, by Van Marken
Lichtenbelt et al. (48) or Van Loan et al. (47). For
D2O, each used accepted protocols (e.g., fasted state, dosage, and equilibration time). Both we and Van
Loan et al. (47) analyzed D2O
enrichment with an accepted infrared spectrophotometry method and Van
Marken Lichtenbelt et al. (48) with an accepted isotope-ratio mass
spectrometry approach. All three laboratories corrected for isotope
fractionation. The only variable identified was that we did not make a
baseline measure of D2O, which
potentially could lead to an underprediction of TBW space. However,
this is unlikely to explain such a large scaling difference. We also
did not account for D2O lost in
the urine, but the subjects were measured in a fasted state and
refrained from drinking or eating; thus, this is unlikely to explain
such offset. Similarly, for NaBr, all three laboratories used accepted administration and analytical protocols, including 10% corrections for
nonextracellular distribution and 5% Donnan equilibration. As did Van
Marken Lichtenbelt et al. (48), we measured NaBr concentrations with
the accepted anion-exchange chromatographic method (32, 50), and Van
Loan et al. (47) used a fluorescent-excitation technique.
The only variables not accounted for were basal NaBr concentrations and
NaBr lost in the urine during the equilibration. However, over such a
short period of time and with the subjects being in a fasted state, the
loss in NaBr in the urine would be small, as would be the error caused
by not subtracting baseline NaBr from the plasma after administration
(32). There are other variations (such as hydration status and
metabolic rates) (11), but these also would not explain such large
offsets. As such, there was little difference in the methods used. The
offset could be attributed to the small sample size
(n = 24) originally used to compute
ECW and
ICW (47), but if correctly
determined there should not be such large deviations between
individuals or samples. The validity of these terms is supported by
their correspondence to biophysics results and cross-validation.
If BIS or even the same dilution methods have such variation, it will
be difficult to completely standardize the BIS prediction of ECW and
ICW. To establish whether this variability is BIS or dilution based,
ECW and
ICW should be computed from
D2O and NaBr collected from a
large, well-standardized, multiple-laboratory study. Later studies
could then use the same methods to judge how well these terms hold up.
However, if constants can be derived that allow ECW and ICW to be
predicted close to an accepted reality, the change in volume may become
the most relevant clinically. The change in BIS-predicted volume has
been reported to be quite good (10, 18). Despite the small gender
difference discovered in
ECW
and
ICW (35, 47), this may be
only a sample-specific phenomenon. Isolating
ECW and
ICW will allow investigation of the specific effects of temperature and ion concentration on ECW and
ICW, rather than using a gross single-frequency tissue
measurement.
Effects of geometry on
.
A wrist-ankle measurement would be inappropriate for patients with
ascites, but the vast majority of subjects do not have such conditions
(29). The good predictions of body water reported by this and many
other studies using a wrist-ankle measurement supports that body water
is evenly distributed in healthy subjects. If not, good predictions of
body water would not be possible. To evaluate the error caused by
making a wrist-ankle measurement and determine the validity of
ECW and
ICW computed previously (47),
we compared the values of these terms to results reported in biophysics
for plasma and ICW. First, we used standard anthropometric values for
the ratios of arm, leg, and trunk lengths and girths (45) in the
equation listed in APPENDIX c to
compute a geometry constant
KB and remove the
geometry effects on
. Albeit a rough approximation, because the
arms, legs, and trunk are not perfect cylinders and the fraction of ECW
and ICW is not constant between segments, an approximation should be
possible. The value for
KB was computed
to be 4.3. The longitudinal
of human skeletal muscle tissue
measured at 1 kHz has generally been reported to be 200-300
· cm. (14). These reported measures for apparent
were obtained from direct measurements on skeletal muscle tissue (14) and thus were corrected for the mixture effects caused by the ICW
contained in that muscle. Assuming the fluid distribution found in a
previous study (47) to be representative of healthy adults (45% ECW in
TBW and 73% TBW in FFM), the concentration of nonconductive material
in the skeletal muscle tissue was estimated to be 67.2%. By using
Eq. C4, this yielded a
for ECW of
nominally 250 · (1
0.672)1.5 or
ECW of 47
· cm. Our results indicated that
kECW was
nominally 0.311; by using a nominal
Db of 1.05 kg/l, this relates to a
ECW of 41
· cm, which is in reasonable agreement with both
the
calculated from skeletal muscle tissue measurements and to the
for pure ECW reported (50-60
· cm)(14).
Thus the distribution of ECW throughout the body was very consistent,
and there was no significant error caused by making a wrist-ankle
measurement. Similarly, the values found for
k
(i.e., 3.6)
and
(i.e., 0.7; Table 2) were in reasonable agreement with the
values of 0.3 and 0.6 previously reported in biophysics (5, 14),
respectively. These findings have been replicated in a pediatric sample
where the computed
ECW was
within 5% (49) of the value discovered in healthy adults (47). It is
uncertain how the mixture effects will be adequately accounted for with
a segmental measurement.
Principles of fitting data to a biophysical model.
Plots are used to construct an applicable physical or mathematical
model. Once a model has been constructed, computing the components of
the model becomes the focus (5, 25, 40). The Cole model can be computed
graphically or mathematically by drawing or fitting the best fitting
curve through the data and extrapolating each end of the
curve to where it intercepts the resistance axis (known as
R0 and
R
) (Figure 3).
R0 equals RE; thus, once
R0 an
R
are known,
RI can be determined by
1/R
1/R0 = 1/RI. Fitting is generally
performed mathematically because it is far more precise (25). Modeling
can also be performed either manually or mathematically by simply
fitting a circle through the measured R and X data (6, 43) but this
approach does not include f and thus
is two dimensional, using one-third less data to determine and
cross-check the best fit. This method also provides no estimate of
Cm but most
importantly does not allow for the effects of
Td to be removed. Network analysis
is a well-known analytical technique, and the common method used for
fitting data to a network model is to simultaneously fit
f and weighted Z and
data, using
nonlinear least squares curve fitting (25). The important data are not
at LF and HF but in the middle surrounding fc because these
data have greater certainty. To accurately fit and extrapolate a curve
requires adequate data on either side of
fc. Properly
weighting the raw data is essential for obtaining the most accurate fit
to the model (25) because measurements may have greater uncertainties
at LF and HF. By weighting, the data with more certainty (i.e., middle)
make a greater contribution to the overall fit to the model (25).
Determining what weights to use is not easily decided (25), and
simultaneously fitting a multidimensional nonlinear equation is an art,
with the raw data having a very complicated relationship to the final
fitted parameters. Evaluating the accuracy of fit should be determined by comparing the offset to fit to the expected measurement uncertainty at each f (25). Because of weighting
and the Cole model being multidimensional, evaluating the fit with a
two-dimensional statistical analysis [e.g., root mean square
error (RMSE) of R and X] would be meaningless (8, 27, 43).
Weighting can be determined by measuring the range of error at each
f, according to the expected error in
the measured quantities (e.g., accuracy specifications of the device),
or as we use by weighting the error rather than the data by comparing
the expected error with the actual error (25). Limitations must be
enforced to prevent the software from forcing a fit. As many
frequencies as possible should be used because solving for five
unknowns requires at least five data, and all data are potentially
contaminated with error (e.g., interference) and thus are uncertain
(25). The ability to delete data that are significantly decreasing the
overall accuracy of fit is a highly desirable capability (25). The
accuracy of resolving a model is a function of the square root of the
number of extra data pairs (i.e., Z and
) over the number of
variables in the model. A 16:1 increase in data provides a 4:1
improvement. However, processing time is effectively the square of the
number of data pairs. Thus, a 16:1 increase in data takes 256 times
longer to compute. We presently measure at 50 frequencies
logarithmically spaced from 5 kHz to 1 MHz to balance between accuracy
and processing time. It is important to space the frequencies
logarithmically to ensure a proper density of data. We fit with both Z
and
(rather than Z alone, or R and X) to ensure the best possible
accuracy of fit. If only Z is modeled (8), the overall accuracy is
reduced (25). Using Z and
provides twice the amount of data, and
is an extremely important discriminating variable because it has a
much broader range of sensitivity to change than Z. However, using
requires that the time delay
(Td) effects must be accounted for (37). We fit with Z and
vs. R and X because the weighting introduced by X would enhance the importance of frequencies furthest away from fc
and thus emphasize the opposite to what is needed. Furthermore, fitting against R and X is more complex and considerably slower because X is nonlinear.
Effects of f invariant
Td.
As published (30) and provided in Xitron's product literature since
1992, we recommend investigators extend the HF range of the measurement
by removing the effects of Td by
multiplying the Cole equation by the factor
e
jwTd, where
e is natural number, j is

1, and w is f in
radians/s. Despite the confusion this parameter has
caused (8, 43), all conductors exhibit a
Td that causes a linear
shift
with f. Conductor length would be an
obvious cause for Td (copper wire having a Td of ~1.2 ns/ft),
whereby an 8-ft conductor length (e.g., wrist to ankle) would produce
10 ns of delay (37). However, a longer
Td of 32.4 ns was observed in the
female subjects (Table 2). This is because
Td can also be caused by
interaction between contact R, stray capacitance and transmission line
effects, with the latter including conductor length (wrist to ankle)
and the conductor (i.e., body) position relative to ground (floor, bed, table, and so on) (15). Only conductor length is a true
Td effect, but in the 1 kHz-1 MHz
f range the other effects also give
rise to a linear
shift with f and
thus can be approximated as a Td. For simplicity, all effects will be labeled as
Td throughout the discussion. The
various causes of Td were
simulated by using the widely available SPICE circuit-simulation
software program (Intusoft, San Pedro, CA) and using the SPICE file
shown in Table 4.
Td were modeled by a high-Z
transmission line with conductor length set by
Td, and the body Z was modeled by
a simple three-element model (no
) using an RE = 680, RI = 900, and
Cm = 2.8 nF.
Figure 9 shows a plot of
vs.
f in the 1 kHz-1 MHz
f range for body Td of 0, 15, and 30 ns of
Td and a characteristic
transmission line impedance (Zc)
of 300
. Because the body is usually suspended some distance from a
ground plane, the body behaves like a transmission line (15). Figure
10 shows
vs.
f by using the same three-element values for RE, RI and
Cm, a fixed
conductor length Td of 15 ns, and
a Zc of 150, 300, 450, and 600
, respectively.
|
) vs. frequency ( f ), simulating
effects of time delay Td of 0 ns
(1), 15 ns
(2), 30 ns
(3) caused by different conductor
lengths on 3-element model consisting of RE in parallel
with series RI and
Cm and fixed
transmission line characteristic impedance of 300
.
vs. f, simulating
effects of Td caused by different
transmission line characteristic impedances of 150 (1), 300 (2), 450 (3) and 600
(4) on 3-element model consisting of RE in parallel with series
RI and
Cm and fixed
conductor length Td of 15 ns.
For convention,
in these plots is expressed as having positive
polarity. As shown in Figs. 9 and 10, there is a significant linear
shift with f caused by both conductor
length (wrist to ankle) and conductor (body) relative to ground.
Adjustment of the various parameters (e.g., distance from ground plane)
in the simulation circuit changed the magnitude and frequency at which Td effects emerged. As shown in
Figs. 11 and
12, there can be considerable variations
in Td between
subjects. Different devices, subjects, and environments will cause
different responses to the variables causing
Td (15). The interaction between
variables resulted in an effect larger than their sum and begin to
affect Z slightly over 1 MHz. As shown in Fig. 10, the effects caused
by conductor relative to ground can cause a negative
Td and opposite effect on
,
thus explaining the negative Td
shown in Table 2. Although less accurate than modeling, the
Td effects can be removed manually by the methods described in APPENDIX d.
vs. f of measured data (raw) and data fit to
Cole model on 1 subject with calculated
Td of
1.3 ns.
vs. f of measured data (raw) and data fit to
Cole model on subject with calculated
Td of 36.1 ns.
It is disappointing that several investigators would not use or even mention our suggested methods and information on Td, then cause undue confusion by reporting an intermediate result and incorrectly attributing the deviation in the raw X data from the Cole model to measurement error (8, 43), particularly when these methods had been disclosed and used to successfully predict ECW and ICW volumes for the first time using Cole model terms RE and RI (30, 47). The effects of Td are linear on
not on X, and
in the f range of interest
Td only significantly affects HF
shifts not Z (37). If Td were
not a valid term, the correlation of fit of Z (which is the
R2 + X2) would also be seriously
reduced, whereas it is not (i.e., 0.998; Table 2). Stroud et al. (43)
should have questioned their conclusions, because if the measurement
was poor, the data would not have corresponded so well to an electronic
circuit. Similarly, Deurenberg et al. (8) should have questioned their
conclusions when there was no deviation of Z from the model at HF (8),
and as stated, the final fit to the model included data up to 500 kHz.
To review APPENDIX a, it can readily
be seen that any f significantly affecting the overall fit will be deleted.
The practical reasons for modeling for
Td are simple. There are
Td effects in all raw data to
varying degrees (15); thus as much of the
Td effects as possible should be
removed from the analysis. Fortunately, in the
f range of interest, all the effects of Td cause a linear
shift
with f. Because the Cole model (i.e., Cm) causes a
nonlinear
shift with f, the
effects of Td can be effectively
modeled and significantly removed. Our modeling program and information
on Td were widely distributed in
1992. As such, it would not be difficult to adjust the various
parameters to push out to higher frequencies where the effects of
Td become dominant. However, as
discovered by Stroud et al. (43), without removing the effects of
Td, only useful data up to
500-600 kHz will be obtained (43). Table
5 is an output file (.MDL file) generated
from our fitting software on one of the subjects of this study. As
shown, few data were deleted from the final fit, and frequencies up to
1 MHz were included. Without removing the effects of
Td, inclusion of such HF data
would not be possible. The problem with not modeling for
Td is that there are variations in
the environment in which measurements will be performed,
and it is not uncommon, particularly in the clinical setting, to
observe fcs
>500 kHz. With usable data only to 500-600 kHz, it would not be
possible to accurately fit for
RI. Although the physics
underlying Td is important, what
is important to the prediction of ECW and ICW is to remove the effects
of Td so the highest possible
f range can be included. Even with
lower fc, the
closer the actual data are to
R
, the better the calculation
of RI will become.
|
-dispersion become dominant near 1 kHz
and must be avoided (Fig. 1). Most importantly, the proportion of
current conducting through the cells at "any" single
f is not fixed but varies with
fc, and
fc varies between
individuals, as well as in the same individual when RE,
RI, Cm, or
is altered
(5, 22, 26, 40; Figs. 1, 13, and 14). By
fitting the data to the R0 and
R
, the above error sources are
removed. Jaffrin (19) reported that the overestimation of
RI can be as high as 200% by not using
R
(19).
As shown in Figs. 1 and 13, a 50-kHz measurement is neither a measure of ECW or TBW but rather some of both. No single HF measurement is a measure of TBW, including R
and
Z at fc as
promoted by Cornish et al. (6, 27), but rather a measure of two
significantly different fluids. The
ECW has been reported to be
50-60
· cm (14) and the
ICW to be 200-300
· cm (5). It has been previously assumed that the
of TBW is constant. Obviously this assumption is invalid because a
simple change in the ECW-ICW ratio would dramatically change it. This
error can be reduced, as we have done, by using the measured
RE and RI with the previously established constants
ECW and
ICW to determine the actual
relative proportions of ECW and ICW. From this, one can establish the
of TBW. The equation shown uses a linear mixture effect; however,
in practice a nonlinear ECW-ICW mixture effect was used. The difference
is insignificant in healthy subjects or when small changes are not of
concern.
The prediction of ECW is inherently better than ICW and TBW for both
technical and theoretical reasons. The prediction of ECW is achieved
directly from model term RE, whereas ICW is predicted effectively by the difference between two large numbers
(R
and
R0). Thus, a 0.1% error in
R
is
0.5% error in the predicted ICW. Although there is a call for a return to a HF and LF
approach because R
is more
variable than a fixed HF (8), the above error sources can never be
resolved with a fixed-f nonmodeling
approach; thus, it is fraught with error. On the other hand, the repeatability and accuracy of solving for
R
and thus ICW is technical
rather than theoretical in nature. Improvements in the measurement
should reduce the variability in predicting ICW.
Parallel reactance, phase angle, and cell membrane capacitance.
Series reactance at 50 kHz (Xs)
had been proposed as a measure of ECW (41) and as a measure of the
extracellular mass/BCM ratio (42). A 50-kHz parallel X model
(Xp) has now been proposed as a
measure of BCM (3, 23). To support this proposal, Lukaski (23)
performed a progressive potato study to demonstrate the f dependence of biological tissue (23)
and drew on the statement by Foster et al. (12) that Z can be
interpreted as either a parallel or series circuit and both resulting
in two final elements (real and imaginary). The later is absolutely
true for any single f measurement, but
biological tissue consists of more than two elements. Z at any single
f can be interpreted as a parallel or series circuit, but the field is concerned with how to interpret the Z
of biological tissue. According to Fricke (13), Schwan (40), and Cole
(5), single biological cells can be represented as a series-parallel
network having three elements: RE in parallel with a series
Cm and
RI. Cole (5) added an exponent (
) to the model to
represent the distribution effects observed on biological cell
suspensions and tissues. The Cole model is used most often to interpret
Z measured on biological tissue and consists of four elements (31).
Based on the belief that how biophysicists interpret Z measurements has
merit, we use the Cole model. To do this, we fit all real and imaginary
data (i.e., corresponding Z and
) to the Cole model to discriminate
the component parts of the tissue.
If previous work in biophysics does have validity, the use of R and X
at any single f to predict ECW or ICW
would be an oversimplification and is dependent on the elements in the
tissue having relative uniformity between individuals. Any relationship
between BCM and X is likely a function of the relationship between X
and Cm because ICW is a resistive not capacitive medium. It has been suggested that as
the cell swells, the membrane becomes thinner and
Cm increases, and
the opposite occurs when the cell shrinks (16). However, X at any
single f is not merely
Cm, and
Cm can only be
computed by modeling for all the elements in the Cole model (5, 25, 40). Which variable is affecting X at any single
f cannot be determined. However, since
RE, RI, and
tend to be tightly regulated and vary within narrow limits, X would tenuously reflect
Cm and give rise
to a correlation to cell volume.
To investigate the strength of the relationship between
Cm and the
variables related to
Cm and BCM as
defined by TBK, we investigated their correlation to TBK. As shown in
Table 6, weight was strongly correlated to
TBK, but variables other than BCM can cause a change in weight.
Cm alone was
strongly correlated to TBK and improved when expressed as
Ht2 × Cm. There was a
poor direct relationship between
Xs and
Xp and TBK and
Cm, respectively.
Xs and
Xp were moderately correlated to
TBK when expressed as
Ht2/Xs
or
Ht2/Xp,
respectively, but Ht alone was highly correlated.
fC was also
correlated to TBK, but mathematically
fc is dominated
by Cm and
RE, and since this healthy population would have a narrow range in RE in relation to the other model parameters, the
relationship between
fc and TBK was
most likely dominated by
Cm. As shown in Table 6, there was a strong relationship between
fc and
Cm. Use of
fc cannot be
supported, because it is affected by all the variables in the model,
but as expected,
fc predicted TBK
better than X at 50 kHz. The mean
fc for this
healthy sample was
60 kHz (Table 2); thus, X would approximate
fc. However,
fc was more
highly correlated than X simply because
fc more closely
reflects Cm, whereas the strength of the relationship between X and
Cm varied with
fc, which ranged
in this sample from 43 to 110 kHz. The strong relationship between
Cm and TBK was
expected because
Cm is a function of cell surface area. The moderate relationship between
Cm and weight
reflects this dependence. However,
Cm is also
affected by the aspect ratio (length to cross-sectional area) of the
body's conductor. With an identical total cell volume, a greater
conductor length would cause less
Cm, whereas a
greater conductor cross-sectional area would cause higher
Cm. Although
error caused by aspect ratio is removed by
length2 × Cm, this is only
true for a uniform cylinder. Further refinements in
Cm would need to
be made by accounting for
KB
(APPENDIX c). As discussed above,
Cm is also
affected by the thickness of the cell membrane. With the errors caused
by aspect ratio and cell membrane thickness, it is unclear why a
surface measurement (i.e., Cm) would be
used to reflect what is inside the cells when it can be determined more
directly by a resistive-predicted ICW
(ICWR).
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||
, and an R-X graph have
been proposed as measures of fluid distribution and discriminating indexes of health and disease (2, 36). Recently,
-angle spectrum
analysis (that is,
vs. f) has
been proposed as descriptive of body water and body composition (4).
angle is a function of the ratio of R and X; thus, both
and an
R-X graph would be sensitive to the same errors and uncertainties as
any single frequency of R and X. The sensitivity of
is extremely
dependent on X, which in turn is extremely dependent on the
relationship between the frequency of measurement and
fc, and is
symmetrical about fc. X and
at
50 kHz change dramatically when
fc changes,
simply because 50 kHz is a fixed point on the changing curve (22, 26) (Fig. 14). X simply changes more than R because X is only 5% of the
total Z; thus, a slight change in
fc causes a
greater percentage change in X. In 1974 (Fig. 14; see Ref. 34), it was
observed that dialysis patients had a lower X and
measured at 50 kHz predialysis and that it returned to that observed in healthy
subjects postdialysis. Lofgren (22) attributed the cause of this to the change in fc 23 yr previously. The change in X and
with a change in fluid
distribution has given the incorrect impression that X and
are
somehow directly related to fluid distribution. A change in fluid
distribution does change
fc, which in turn
changes X and
, but this is principally caused by a change in ECW
(and Cm, as
discussed above). Again, the problem with using X,
,
or fc to reflect
any body composition parameter is that these variables are affected by
all the elements in the tissue (APPENDIX a). One can question the utility or need for X,
,
fc, or an R-X
graph when which variable is causing their change cannot be determined and they have no theoretical basis. On the other hand,
RE and RI at least relate in theory to a
physical object (i.e., ECW and ICW).
On the same individual,
Cm is determined
by total cell volume and membrane thickness and porosity (5). Thus, any
Ht2 × Cm relationship
would be a function of these three parameters. An
ICWR can be used to predict total
cell volume, and then if removed from the
KB corrected
Ht2 × Cm relationship
by using Ht2 × Cm/ICWR,
the remaining index would reflect cell membrane thickness and porosity.
Such a measure might have several applications. Scheltinga et al. (39)
observed that as the severity of sepsis increased,
Cm decreased to
the point where there was virtually no
- dispersion. This
corresponds to what Lukaski (23) observed on a cooked potato. It is
well known that with cell death or cell destruction, the cell membrane
loses its high resistive properties. During dialysis, ECW changes are
on the order of 20-30%, RI varies little, but both
Fc and
Cm can change by
as much as 2:1 (1, 19) (Fig. 14). As shown in Table 2,
the mean Cm for
the male subjects of this study was 2.32 nF. Bestoso et al. (1)
discovered a mean increase of
Cm in men from
pre- to postdialysis of 64% (1.64-2.47 nF), with the postdialysis
Cm being quite
close to that measured in the men in this study. Thus, if
Scharfetter's (38) estimates are correct that a 5-mmol change in ion
affects the ICW 4%, the error caused by ion on
ICWR, as well as the error in
predicting ICW, would be insignificant compared with the percentage change in Cm. Use
of Cm for
studying cell membrane health is an exciting area of research that
awaits further investigation.
In conclusion, there has been very poor crosstransfer of information
from the fields of physics and engineering to the field of human body
composition. It does not help that the principles of BIS and mixture
theory are rather complicated for many investigators. However, Z is an
engineering- and physics-based technique, and the principles and merits
of modeling and mixture theory have been known for a very long time.
Multiple-frequency devices safe for human studies and modeling programs
have now been available for >4 yr; thus, the lack of appropriate
equipment is no longer a valid reason for not using modeling. Due to
the high intercorrelation (48) among ECW, ICW, and TBW, one can always
correlate a limited single- or dual-frequency measurement to body
water, but this leads to population-specific equations and a reduced
sensitivity to change, which is why Z measurement has not yet reached
its full potential. Until investigators begin using the proven and accepted fundamental techniques used in other fields of science (i.e.,
modeling) that use Z measurements (e.g., biophysics, advanced materials
research, and chemical engineering), the use of Z in clinical medicine
will remain what it is today
a technique that generates many papers
but has no real clinical application.
Address for reprint requests: J. Matthie, Medical Dept., Xitron Technologies, Inc., 6295 Ferris Sq., Suite D, San Diego, CA 92121 or A. De Lorenzo, Dept. of Human Physiol., Univ. of Rome "Tor Vergata," via O. Raimondo 1, I-00173, Rome, Italy.
Received 23 May 1996; accepted in final form 13 November 1996.
Modeling
The Z and
spectra data were fit to the Cole-Cole model (5),
Eq. A1, using iterative nonlinear
curve-fitting software. The modeling program evaluated the weighted
least square error of both Z and
, where the weighting is
established by the published accuracy specifications of the instrument,
and removed any f that would
significantly decrease the total weighted least square error. In
addition to the correlation of fit using scalar Z, the program established the accuracy of fit to the model as follows:
1) Mean offset to fit <1/2 the instrument measurement specifications.
2) Mean offset to fit less than the instrument measurement specifications.
3) Mean offset to fit <2 × the instrument measurement specifications.
4) Mean offset to fit <5 × the instrument measurement specifications.
5) Mean offset to fit >5 × the instrument measurement specifications.
To prevent the program from deleting frequencies solely to "force" a fit to the model, the following limitations were enforced in the software
1) A maximum of 25% of the frequencies (f) may be deleted.
2) Within any 3:1 range of f, at least one f must remain.
3) Only
f whose Z and
lay more than the
instrument specification from the curve may be deleted.
4) Only one f is deleted per iteration of fitting.
5) A
f is only deleted if it results in the
maximum improvement in resultant fit; this is not necessarily the
f whose Z and
lay farthest from
the fit.
The Cole model was extended to allow for the
f invariant time delay
(Td) caused by the speed at
which electrical information is transferred through a conductor (15,
37). The error introduced by this fixed
Td was modeled as a
error that
increases linearly with f. This linear
error was mathematically modeled by multiplying Eq. A1 by the factor
e
jwTd. Thus the overall
modeled equation was
|
(A1) |
× f ); and
j is 
1.
fC was computed after the model
components (RE, RI, CM,
Td, and
) had been determined
by solving the equation
|
(A2) |
Theoretical Volume Equations
The ECW and ICW volumes were predicted from the modeled RE and RI by using equations formulated from Hanai's theory, which describes the effect that a concentration of nonconductive material has on the apparent resistivity (
) of the surrounding
conductive fluid, and is
|
(B1) |
is the apparent
of a conductive material;
0 is the actual
of a
conductive material; and C is volumetric concentration of the
nonconductive material contained in the mixture.
From Eq. B1, with the following assumptions, we derived a set of equations as follows
|
(B2) |
|
(B3) |
);
KB is a factor
correcting for a whole body measurement between wrist and ankle,
relating the relative proportions of the leg, arm, trunk, and height
(see APPENDIX c).
ECW is the
of extracellular
water (
· cm);
Db is body density (kg/l)
|
(B4) |
|
(B5) |
).
The following assumptions were made:
1) The volumetric concentration of nonconductive elements in the body at low frequencies (LF) is given by
|
2) The volumetric concentration of nonconductive elements in the body at high frequencies (HF) is given by
|
3) VTot is body Wt/Db.
4) The total volume of a body fluid can be described by
|
(B6) |
F is the
of the water; L is
body height; and R is the measured resistance between wrist and ankle.
5) The factors
Db,
KB, and
F can be considered largely
constant.
6) The Hanai equation is applicable at HF and LF to mixtures found in the human body.
By using Eqs. B2 and B4, predicted VECW and VICW were computed, from which predicted TBW was computed by using the following equations
|
(B7) |
Computing the Constants
kECW is established as the mean value of
|
is to
repetitively predict
VICW/VECW
and adjust k
until a minimum mean error between the predicted and measured ratio is
obtained.
Derivation of KB
It should be noted that the derivation for KB shown here is only an approximation for the purposes of confirming whether its use results in a
ECW value that is
within the range measured by other investigators.
The resistance (R) of a cylinder, measured longitudinally, is given by
|
(C1) |
is the resistivity of the material; L is the length of
the cylinder; and A is the
cross-sectional area of the cylinder.
Restating Eq. C1 in terms of the cylinder length and circumference
|
(C2) |
|
(C3) |
If we consider the body to be formed by five cylinders (the legs, the arms, and the trunk), then the volume of the body is given by
|
(C4) |
When we measure the Z between the wrist and the ankle, the measured value will be
|
(C5) |
|
(C6) |
Combining Eqs. C4, C5, and C6 yields
|
(C7) |
|
(C8) |
Removing the effects of Td
Td can be removed through modeling, but it is only applicable when the actual measured data are used as input to fitting a program against a model. In this case, an additional multiplicative term must be added to the model (Eq. A1), which yields no change in amplitude but instead yields a linear
shift with increasing f. As shown, the user will have one
more "constant" term (K in this
example) in the model
|
The modeling approach to removing
Td performs the best, allowing the
user to extend the usable f range up
to ~1 MHz, thus allowing for higher accuracy modeling. However, this
technique is the most complex and may take excessive computing
knowledge or time. The user can also approximate the observed
error
manually and recalculate the set of measurement data including a
correction for this effect. It should be noted that, without employing
the full model as outlined above, it is not possible to separately model this effect alone because the biological effects will
"skew" the results. The manual method does offer the advantage of
being very fast to implement, and it may be performed either by
computer or by hand. The technique does not offer the accuracy of this first method, however, because the correction is only optimized over
relatively few data points. These data do, however, increase the useful
f range of the measured data up to
several hundred kiloherz.
Because there is little biological
shift caused by
Cm at HF, choose
a HF and assume that all the observed
shift is caused by
Td. If 1-MHz data (as an example)
are selected, note the measured
shift at 1 MHz as
1MHz, then correct the measured
shifts by subtracting
1MHz × f/1,000 from all
measured
shifts. The measured Z data need no correction. The final,
corrected, resistance (R) and reactance (X) data may be computed as
follows
|
|
correction for the
f of the point used) or may wish to
try taking the average of several HF data points as shown in the
following example of averaging data collected at 700, 800, 900, and
1,000 kHz
|
|
or X data within a statistically produced
model. If only the Z or R is of significant interest, then these
corrections offer little because
Td only affects
and has no
effect on Z.
| 1. | Bestoso, J. T., and R. L. Mehta. Monitoring volume changes in hemodialysis (HD) with bioimpedance spectroscopy (BIS) (Abstract). J. Am. Soc. Nephrol. 4: 333, 1993. |
| 2. | Biasioli, S., R. Foroni, L. Petrosino, L. Cavallini, A. Zambello, G. Cavalcanti, and T. Taluri. Effect of aging on the body composition of dialyzed subjects: comparison with normal subjects. ASAIO Trans. 39: 596-601, 1993. |
| 3. | Chertow, G., E. Lowie, D. Wilmore, J. Gonzalez, N. L. Lew, J. Ling, M. S. Leboff, M. N. Gottlieb, B. Zebrowiski, J. College, and M. Lazzarus. Nutritional assessment with bioelectrical impedance analysis in maintenance hemodialysis patients. J. Am. Soc. Nephrol. 6: 75-81, 1995 [Abstract] . |
| 4. | Chumlea, W., and S. Guo. Bioelectrical impedance and body composition: present status and future directions. Nutr. Rev. 52: 123-131, 1994 [Medline] . |
| 5. | Cole, K. S. Membranes, Ions and Impulses: A Chapter of Classical Biophysics. Berkeley, CA: Univ. of Calif. Press, 1972. |
| 6. | Cornish, B. H., B. J. Thomas, and L. C. Ward. Improved prediction of extracellular and total body water using impedance loci generated by multiple frequency bioelectrical impedance analysis. Phys. Med. Biol. 38: 337-346, 1993 [Medline] . |
| 7. | De la Rue, R. E., and C. W. Tobias. The conductivity of dispersions. J. Electrochem. Soc. 106: 827-833, 1959. |
| 8. | Deurenberg, P., A. Andreoli, and A. De Lorenzo. Multi-frequency bioelectrical impedance: a comparison between the Cole-Cole modeling and Hanai equations with the classical impedance index approach. Ann. Hum. Biol. 6: 31-40, 1996. |
| 9. | De Vries, P. M. J. M., J. H. Meijer, K. Vlaanderen, V. Visser, P. L. Oe, A. J. M. Donker, and H. Schneider. Measurement of transcellular fluid shift during haemodialysis. Med. Biol. Eng. Comput. 27: 152-158, 1989. [Medline] |
| 10. | Finn, P. J., L. D. Plank, M. A. Clark, A. B. Connolly, and G. Hill. Progressive dehydration and proteolysis in critically ill patients. Lancet 347: 654-656, 1996 [Medline] . |
| 11. | Forbes, G. B. Human Body Composition: Growth, Aging, Nutrition, and Activity. New York: Springer, 1987. |
| 12. | Foster, K. R., and H. C. Lukaski. Whole body impedance-what does it measure? Am. J. Clin. Nutr. 64, Suppl.: 388S-396S, 1996. |
| 13. | Fricke, H. A mathematical treatment of the electrical conductivity and capacity of disperse systems. II. The capacity of a suspension of conducting spheroids surrounded by a non-conducting membrane for a current of low frequency. Phys. Rev. 26: 678-681, 1925. |
| 14. | Geddes, L. A., and L. E. Baker. The specific resistivity of biological material: a compendium of data for the biomedical engineer and physiologist. Med. Biol. Eng. 5: 271-293, 1967 [Medline] . |
| 15. | Gersing, E., M. Schafer, and M. Osypka. The appearance of positive phase angles in impedance measurements on extended biological objects. Innov. Tech. Biol. Med. 16: 71-76, 1995. |
| 16. | Gerth, W. A., L. D. Montgomery, and Y. C. Wu. A computer based bioelectrical impedance spectroscopic system for noninvasive assessment of compartmental fluid redistribution. In: Third IEEE Symposium on Computer Based-Medical Systems. Los Alamitos, CA: IEEE Computer Soc., 1990, p. 446-453. |
| 17. | Hanai, T. Electrical properties of emulsions. In: Emulsion Science, edited by P. H. Sherman. London: Academic, 1968, p. 354-477. |
| 18. | Ho, L. T., R. Kushner, D. A. Schoeller, R. Gudivaka, and D. M. Spiegel. Bioimpedance analysis of total body water in hemodialysis patients. Kidney Int. 46: 1438-1442, 1994 [Medline] . |
| 19. | Jaffrin, M. Y., M. Maasrani, B. Boudailliez, and A. le Gourrier. Extracellular and intracellular fluid volume monitoring during dialysis by multifrequency impedancemetry. ASAIO Trans. 42: M533-M538, 1996. |
| 20. |
Keys, A.,
and
J. Brozek.
Body fat in adult men.
Physiol. Rev.
33:
245-325,
1953.
|
| 22. | Löfgren, B. The electrical impedance of complex tissue and its relation to changes in volume and fluid distribution. Acta Physiol. Scand. 23, Suppl. 81: 2-51, 1951. |
| 23. | Lukaski, H. Biological indexes considered in the derivation of the bioelectrical impedance analysis. Am. J. Clin. Nutr. 64, Suppl.: 397S-404S, 1996. |
| 24. |
Lukaski, H. C.,
and
P. E. Johnson.
A simple, inexpensive method of determining total body water using a tracer dose of D20 and infrared absorption of biological fluids.
Am. J. Clin. Nutr.
41:
363-370,
1985
|
| 25. | MacDonald, J. R. Impedance Spectroscopy: Emphasizing Solid Materials and Systems. New York: Wiley, 1987. |
| 26. | Matthie, J., and P. Withers. The ambiguities of predicting total body water and body cell mass with a single frequency (50 kHz) measurement of bioimpedance (Letter to Editor). J. Am. Soc. Nephrol. 6: 1682-1684, 1995 . |
| 27. | Matthie, J. R., and P. O. Withers. Bioimpedance, the Cole model equation and the prediction of intra and extracellular water: science or marketing (Letter to Editor). Clin. Nutr. 15: 147-149, 1996; erratum, 15: 217-219, 1996. |
| 28. |
Matthie, J. R.,
and
P. O. Withers.
Impedance measurements of body-water compartments (Letter to Editor).
Am. J. Clin. Nutr.
61:
1167-1169,
1995
|
| 29. |
Matthie, J. R.,
and
P. O. Withers.
Segmental versus whole body multifrequency bioimpedance measurements (Letter to Editor).
J. Appl. Physiol.
79:
2177-2179,
1995
|
| 30. | Matthie, J. R., P. O. Withers, M. D. Van Loan, and P. L. Mayclin. Development of a commercial complex bio-impedance spectroscopic (CBIS) system for determining intracellular water (ICW) and extracellular water (ECW) volumes. In: Proceedings of the 8th International Conference on Electrical Bio-impedance Kuopio Finland 1992 Kuopio, Finland: University of Kuopio, 1992, p. 203-205. |
| 31. | McAdams, E. T., and J. Jossinet. Tissue impedance: a historical overview. Physiol. Meas. 16: A1-A7, 1995 [Medline] . |
| 32. |
Miller, M. E.,
J. M. Cosgriff,
and
G. B. Forbes.
Bromide space determination using anion-exchange chromatography for measurement of bromide.
Am. J. Clin. Nutr.
50:
168-171,
1989
|
| 33. | Moore, F. D., and C. M. Boyden. Body cell mass and limits of hydration of the fat-free body: their relation to estimated skeletal weight. Ann. NY Acad. Sci. 110: 62-71, 1963. |
| 34. | Nyboer, J., and J. A. Sedensky. Bioelectrical impedance during renal dialysis. Proc. Dial. Transplant. Forum 4: 214-219, 1974 . |
| 35. | Patel, R. V., J. R. Matthie, P. O. Withers, E. L. Peterson, and B. J. Zarowitz. Estimation of total body and extracellular water using single and multiple-frequency bioimpedance. Ann. Pharmacother. 28: 565-569, 1994 [Abstract] . |
| 36. | Piccoli, A., B. Rossi, L. Pillon, and G. Bucciante. A new method for monitoring body fluid variation by bioimpedance analysis: the Rxc graph. Kidney Int. 46: 534-539, 1994 [Medline] . |
| 37. | Ramo, S., J. Whinnery, and T. Van Duzer. Fields and Waves in Communication Electronics (2nd ed.). New York: Wiley, 1984. |
| 38. | Scharfetter, H., G. Wirnsberger, Z. Laszzlo, H. Holzer, H. Hinghofer-Szalkay, and H. Hutten. Influence of ionic shifts and postural changes during dialysis on volume estimation with multifrequency impedance analysis. In: Proceedings of the 9th International Conference on Electrical Bio-Impedance, Heidelberg, Germany 1995. Heidelberg, Germany: Univ. of Heidelberg, 1995, p. 241-244. |
| 39. | Scheltinga, M., E. Jorna, D. Chandi, M. Wust, E. Grep, D. Compas, and R. Wesdorp. Multiple frequency impedance measurements in critically ill patients. Clin. Nutr. 14: 42-44, 1995. |
| 40. | Schwan, H. P. Electrical properties of tissue and cell suspensions. In: Advances in Biological and Medical Physics, edited by J. Lawrence, and C. Tobias. New York: Academic, 1957, p. 147-209. |
| 41. |
Segal, K. R.,
S. Burastero,
A. Chun,
P. Coronel,
R. N. Pierson, Jr.,
and
J. Wang.
Estimation of extracellular and total body water by multiple-frequency bioelectrical-impedance measurement.
Am. J. Clin. Nutr.
54:
26-29,
1991
|
| 42. | Shizgal, H. M. Validation of the measurement of body composition from whole body bioelectric impedance. Infusionstherapie 17, Suppl.: 67-74, 1990. |
| 43. | Stroud, D. B., B. H. Cornish, B. J. Thomas, and L. C. Ward. The use of Cole-Cole plots to compare two multi-frequency bioimpedance instruments. Clin. Nutr. 14: 307-311, 1995. |
| 44. | Thomasset, A. Bio-electrical properties of tissue impedance measurements. Lyon Med. 209: 1325-1352, 1963. [Medline] |
| 45. | United States Army Natick Research, Development, and Engineering Center. Anthropometric Survey of U. S. Army Personal: Summary Statistics Interim Report, 1989. Tech. Report Natick/TR-89/027, AD-209-600. |
| 46. |
Van Loan, M. D.,
L. E. Kopp,
J. C. King,
W. W. Wong,
and
P. L. Mayclin.
Fluid changes during pregnancy: use of bioimpedance spectroscopy.
J. Appl. Physiol.
78:
1037-1042,
1995
|
| 47. | Van Loan, M. D., P. Withers, J. Matthie, and P. L. Mayclin. Use of bio-impedance spectroscopy (BIS) to determine extracellular fluid (ECF), intracellular fluid (ICF), total body water (TBW), and fat-free mass (FFM). In: Human Body Composition: In Vivo Methods, Models and Assessment, edited by K. J. Ellis, and J. D. Eastman. New York: Plenum, 1993, p. 67-70. |
| 48. |
Van Marken Lichtenbelt, W. D.,
K. R. Westerterp,
L. Wouters,
and
S. Luijendijk.
Validation of bioelectrical-impedance measurements as a method to estimate body-water compartments.
Am. J. Clin. Nutr.
60:
159-166,
1994
|
| 49. | Withers, P. Multi-frequency impedance measurements of extracellular fluid volume (Letter to Editor). Physiol. Meas. 16: 71-76, 1995 . [Medline] |
| 50. |
Wong, W. W.,
H. P. Sheng,
J. C. Morkeberg,
J. L. Kosanovich,
L. L. Clarke,
and
P. D. Klein.
Measurement of extracellular water volume by bromide ion chromatography.
Am. J. Clin. Nutr.
50:
1290-1294,
1989
|
This article has been cited by other articles:
![]() |
S. Ozturk, D. G. Taymez, G. Bahat, R. Demirel, H. Yazici, N. Aysuna, S. Sakar, and A. Yildiz The influence of low dialysate sodium and glucose concentration on volume distributions in body compartments after haemodialysis: a bioimpedance analysis study Nephrol. Dial. Transplant., November 1, 2008; 23(11): 3629 - 3634. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Kotanko, N. W. Levin, and F. Zhu Current state of bioimpedance technologies in dialysis Nephrol. Dial. Transplant., March 1, 2008; 23(3): 808 - 812. [Full Text] [PDF] |
||||
![]() |
A. L Gibson, J. C Holmes, R. L Desautels, L. B Edmonds, and L. Nuudi Ability of new octapolar bioimpedance spectroscopy analyzers to predict 4-component-model percentage body fat in Hispanic, black, and white adults Am. J. Clinical Nutrition, February 1, 2008; 87(2): 332 - 338. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Essig, B. Escoubet, D. de Zuttere, F. Blanchet, F. Arnoult, E. Dupuis, C. Michel, F. Mignon, F. Mentre, C. Clerici, et al. Cardiovascular remodelling and extracellular fluid excess in early stages of chronic kidney disease Nephrol. Dial. Transplant., January 1, 2008; 23(1): 239 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Earthman, D. Traughber, J. Dobratz, and W. Howell Bioimpedance Spectroscopy for Clinical Assessment of Fluid Distribution and Body Cell Mass Nutr Clin Pract, August 1, 2007; 22(4): 389 - 405. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Tatara and C. Tashiro Quantitative Analysis of Fluid Balance During Abdominal Surgery Anesth. Analg., February 1, 2007; 104(2): 347 - 354. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Raja, G. H. Raymer, G. R. Moran, G. Marsh, and R. T. Thompson Changes in tissue water content measured with multiple-frequency bioimpedance and metabolism measured with 31P-MRS during progressive forearm exercise J Appl Physiol, October 1, 2006; 101(4): 1070 - 1075. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Esposito, S. G. Thomas, L. Kingdon, and S. Ezzat Comparison of Body Composition Assessment Methods in Patients with Human Immunodeficiency Virus-Associated Wasting Receiving Growth Hormone J. Clin. Endocrinol. Metab., August 1, 2006; 91(8): 2952 - 2959. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Zhu, M. K. Kuhlmann, G. A. Kaysen, S. Sarkar, C. Kaitwatcharachai, R. Khilnani, L. Stevens, E. F. Leonard, J. Wang, S. Heymsfield, et al. Segment-specific resistivity improves body fluid volume estimates from bioimpedance spectroscopy in hemodialysis patients J Appl Physiol, February 1, 2006; 100(2): 717 - 724. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Bichara, A. Attmane-Elakeb, D. Brown, M. Essig, Z. Karim, M. Muffat-Joly, L. Micheli, I. Eude-Le Parco, F. Cluzeaud, M. Peuchmaur, et al. Exploring the role of galectin 3 in kidney function: a genetic approach Glycobiology, January 1, 2006; 16(1): 36 - 45. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A Kaysen, F. Zhu, S. Sarkar, S. B Heymsfield, J. Wong, C. Kaitwatcharachai, M. K Kuhlmann, and N. W Levin Estimation of total-body and limb muscle mass in hemodialysis patients by using multifrequency bioimpedance spectroscopy Am. J. Clinical Nutrition, November 1, 2005; 82(5): 988 - 995. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. C Papathakis, N. C Rollins, K. H Brown, M. L Bennish, and M. D Van Loan Comparison of isotope dilution with bioimpedance spectroscopy and anthropometry for assessment of body composition in asymptomatic HIV-infected and HIV-uninfected breastfeeding mothers Am. J. Clinical Nutrition, September 1, 2005; 82(3): 538 - 546. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Matthie Second generation mixture theory equation for estimating intracellular water using bioimpedance spectroscopy J Appl Physiol, August 1, 2005; 99(2): 780 - 781. [Full Text] [PDF] |
||||
![]() |
Y. C. Luiking, N. E. P. Deutz, M. Jakel, and P. B. Soeters Casein and Soy Protein Meals Differentially Affect Whole-Body and Splanchnic Protein Metabolism in Healthy Humans J. Nutr., May 1, 2005; 135(5): 1080 - 1087. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Strandberg and R. G. Hahn Volume kinetics of glucose 2.5% solution and insulin resistance after abdominal hysterectomy Br. J. Anaesth., January 1, 2005; 94(1): 30 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. J Foster and M. B Leonard Measuring nutritional status in children with chronic kidney disease Am. J. Clinical Nutrition, October 1, 2004; 80(4): 801 - 814. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Buchholz, C. Bartok, and D. A. Schoeller The Validity of Bioelectrical Impedance Models in Clinical Populations Nutr Clin Pract, October 1, 2004; 19(5): 433 - 446. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-C. Wu, Y.-P. Lin, W.-C. Yu, W.-S. Lee, T.-L. Hsu, P. Y.-A. Ding, and C.-H. Chen The assessment of fluid status in haemodialysis patients: usefulness of the Doppler echocardiographic parameters Nephrol. Dial. Transplant., March 1, 2004; 19(3): 644 - 651. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lof and E. Forsum Evaluation of bioimpedance spectroscopy for measurements of body water distribution in healthy women before, during, and after pregnancy J Appl Physiol, March 1, 2004; 96(3): 967 - 973. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bartok and D. A. Schoeller Estimation of segmental muscle volume by bioelectrical impedance spectroscopy J Appl Physiol, January 1, 2004; 96(1): 161 - 166. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. L. Cox-Reijven, B. van Kreel, and P. B Soeters Bioelectrical impedance measurements in patients with gastrointestinal disease: validation of the spectrum approach and a comparison of different methods for screening for nutritional depletion Am. J. Clinical Nutrition, December 1, 2003; 78(6): 1111 - 1119. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Davies, G. Woodrow, K. Donovan, J. Plum, P. Williams, A. C. Johansson, H.-P. Bosselmann, O. Heimburger, O. Simonsen, A. Davenport, et al. Icodextrin Improves the Fluid Status of Peritoneal Dialysis Patients: Results of a Double-Blind Randomized Controlled Trial J. Am. Soc. Nephrol., September 1, 2003; 14(9): 2338 - 2344. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-H. Chen, Y.-P. Lin, W.-C. Yu, W.-C. Yang, and Y.-A. Ding Volume Status and Blood Pressure During Long-Term Hemodialysis: Role of Ventricular Stiffness Hypertension, September 1, 2003; 42(3): 257 - 262. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bartok, R. L. Atkinson, and D. A. Schoeller Measurement of nutritional status in simulated microgravity by bioelectrical impedance spectroscopy J Appl Physiol, July 1, 2003; 95(1): 225 - 232. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Smith, B. Engel, A. M Diskin, P. Spanel, and S. J Davies Comparative measurements of total body water in healthy volunteers by online breath deuterium measurement and other near-subject methods Am. J. Clinical Nutrition, December 1, 2002; 76(6): 1295 - 1301. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Elliott, R. C. Backus, M. D. Van Loan, and Q. R. Rogers Evaluation of Multifrequency Bioelectrical Impedance Analysis for the Assessment of Extracellular and Total Body Water in Healthy Cats J. Nutr., June 1, 2002; 132(6): 1757S - 1759. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Elliott, R. C. Backus, M. D. Van Loan, and Q. R. Rogers Extracellular Water and Total Body Water Estimated by Multifrequency Bioelectrical Impedance Analysis in Healthy Cats: A Cross-Validation Study J. Nutr., June 1, 2002; 132(6): 1760S - 1762. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. A. Wilkins, R. D. Gleed, N. M. Krivitski, and A. Dobson Extravascular lung water in the exercising horse J Appl Physiol, December 1, 2001; 91(6): 2442 - 2450. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Yokoi, H. C. Lukaski, E. O. Uthus, and F. H. Nielsen Use of Bioimpedance Spectroscopy to Estimate Body Water Distribution in Rats Fed High Dietary Sulfur Amino Acids J. Nutr., April 1, 2001; 131(4): 1302 - 1308. [Abstract] [Full Text] |
||||
![]() |
H. Valensise, A. Andreoli, S. Lello, F. Magnani, C. Romanini, and A. De Lorenzo Multifrequency bioelectrical impedance analysis in women with a normal and hypertensive pregnancy Am. J. Clinical Nutrition, September 1, 2000; 72(3): 780 - 783. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Forro, S. Cieslar, G. L. Ecker, A. Walzak, J. Hahn, and M. I. Lindinger Total body water and ECFV measured using bioelectrical impedance analysis and indicator dilution in horses J Appl Physiol, August 1, 2000; 89(2): 663 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. STENVINKEL, B. LINDHOLM, F. LÖNNQVIST, K. KATZARSKI, and O. HEIMBÜRGER Increases in Serum Leptin Levels during Peritoneal Dialysis Are Associated with Inflammation and a Decrease in Lean Body Mass J. Am. Soc. Nephrol., July 1, 2000; 11(7): 1303 - 1309. [Abstract] [Full Text] |
||||
![]() |
G. Woodrow, B. Oldroyd, G. Stables, J. Gibson, J. H. Turney, and A. M. Brownjohn Effects of icodextrin in automated peritoneal dialysis on blood pressure and bioelectrical impedance analysis Nephrol. Dial. Transplant., June 1, 2000; 15(6): 862 - 866. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. P. Earthman, J. R. Matthie, P. M. Reid, I. T. Harper, E. Ravussin, and W. H. Howell A comparison of bioimpedance methods for detection of body cell mass change in HIV infection J Appl Physiol, March 1, 2000; 88(3): 944 - 956. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. G. Kemink, J. T. M. Frijns, A. R. M. M. Hermus, G. F. F. M. Pieters, A. G. H. Smals, and W. D. Van Marken Lichtenbelt Body Composition Determined by Six Different Methods in Women Bilaterally Adrenalectomized for Treatment of Cushing's Disease J. Clin. Endocrinol. Metab., November 1, 1999; 84(11): 3991 - 3999. [Abstract] [Full Text] |
||||
![]() |
R. Gudivaka, D. A. Schoeller, R. F. Kushner, and M. J. G. Bolt Single- and multifrequency models for bioelectrical impedance analysis of body water compartments J Appl Physiol, September 1, 1999; 87(3): 1087 - 1096. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. C.H. VAN DEN HAM, J. P. KOOMAN, M. H.L. CHRISTIAANS, F. H.M. NIEMAN, B. K. VAN KREEL, G. A.K. HEIDENDAL, and J. P. VAN HOOFF Body Composition in Renal Transplant Patients: Bioimpedance AnalysisCompared to Isotope Dilution, Dual Energy X-Ray Absorptiometry, andAnthropometry J. Am. Soc. Nephrol., May 1, 1999; 10(5): 1067 - 1079. [Abstract] [Full Text] |
||||
![]() |
P. R. Schloerb;, J. Matthie, G. Pan, P. Withers, B. Zarowitz, A. de Lorenzo, A. Andreoli, and K. Katzarski Bioimpedance Measurement of Extracellular Water J Appl Physiol, February 1, 1999; 86(2): 773 - 774. [Full Text] [PDF] |
||||
![]() |
J. Q. JAEGER and R. L. MEHTA Assessment of Dry Weight in Hemodialysis: An Overview J. Am. Soc. Nephrol., February 1, 1999; 10(2): 392 - 403. [Abstract] [Full Text] |
||||
![]() |
K. J. Ellis and W. W. Wong Human hydrometry: comparison of multifrequency bioelectrical impedance with 2H2O and bromine dilution J Appl Physiol, September 1, 1998; 85(3): 1056 - 1062. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Zhu, D. Schneditz, E. Wang, and N. W. Levin Dynamics of segmental extracellular volumes during changes in body position by bioimpedance analysis J Appl Physiol, August 1, 1998; 85(2): 497 - 504. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Matthie, B. Zarowitz, A. De Lorenzo, A. Andreoli, K. Katzarski, G. Pan, and P. Withers Analytic assessment of the various bioimpedance methods used to estimate body water J Appl Physiol, May 1, 1998; 84(5): 1801 - 1816. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. B. Mazess Letters to the Editor J Appl Physiol, January 1, 1998; 84(1): 396 - 396. [Full Text] [PDF] |
||||
![]() |
S. B. Heymsfield, C. Nunez, and A. Pietrobelli Bioimpedance Analysis: What Are the Next Steps? Nutr Clin Pract, October 1, 1997; 12(5): 201 - 203. [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |