Departments of 1Sports Medicine and of 2Cardiology, Máxima Medical Centre, Veldhoven; 3Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven; 4Rudolf Magnus Institute of Neuroscience, Section of Rehabilitation and Sports Medicine, University Medical Centre Utrecht, Utrecht; 5Department of Medical Physics, Máxima Medical Centre, Veldhoven; 6Department of Applied Physics, Eindhoven University of Technology, Eindhoven; and 7Department of Cardiology, University Medical Centre Utrecht, Utrecht, The Netherlands
Submitted 20 March 2008 ; accepted in final form 16 October 2008
| ABSTRACT |
|---|
|
|
|---|
cycle ergometry; circulatory dysfunction; hemodynamic response
The purpose of this study was to evaluate the accuracy of LiDCO and Physioflow at rest and during exercise, using the Fick method as reference. In addition, we investigated the sensitivity of both techniques by evaluating within-patient changes in cardiac output.
| METHODS |
|---|
|
|
|---|
Subjects
Ten male patients with stable CHF were recruited to participate in the study. Criteria for eligibility were CHF secondary to ischemic or dilated cardiomyopathy, New York Heart Association class II–III, left ventricular ejection fraction of 40% or less (assessed by echocardiography or radionuclide ventriculography), and sinus rhythm. Exclusion criteria were recent myocardial infarction (<3 mo prior), angina pectoris at rest, hemodynamically significant valvular heart disease, significant COPD and peripheral, vascular, neurological, or orthopedic disease limiting the ability to exercise. Subject characteristics are listed in Table 1. During the study period, there were no changes in medication use.
|
All patients performed exercise tests on two occasions using an electromagnetically braked cycle ergometer in an upright position (Corival, Lode, Groningen, The Netherlands). In both tests ventilatory parameters were measured breath-by-breath (Zan 680 USB, Oberthulba, Germany). Volume and gas analyzers were calibrated before each test. A 12-lead electrocardiogram was registered continuously. The mean duration between both occasions was 7.8 ± 4.9 days.
During the first session, patients underwent a symptom-limited incremental exercise test, using an individualized ramp protocol, with a duration of 8–12 min (42). The test was ended when a patient was not able to maintain the required pedaling frequency of 70 per minute. Peak oxygen uptake (
O2) and peak workload were defined as the average values of the last 30 s of the test. The
O2 at the ventilatory threshold was determined by the V-slope method, using the average value obtained by two independent observers (2). From this test the individual workload for the two constant-load tests was calculated (see below).
The second session was performed to evaluate the accuracy of the cardiac output assessment methods under different conditions: at rest, during light and moderate steady-state exercise, and maximal exercise. The session started with a resting period, in which cardiac output measurements were performed (see later). Thereafter, patients started exercising with 2 min of cycling at the lowest intensity of the ergometer (7 W), followed by two subsequent 5-min bouts at a light and moderate constant load of 30% and 80% of the workload of the ventilatory threshold (VT), respectively. Based on our previous experience with CHF patients, a workload of 30% of VT is the lowest workload at which significant changes in
O2 and cardiac output (
) occur, and a workload higher than 80% of VT does not ensure a steady state for all patients. Steady state was defined as a stable heart rate (increase <5%) during the last 2 min of exercise. Finally, a symptom-limited incremental exercise test was performed after a resting period of at least 15 min (average duration between tests: 81 ± 47 min) to assess cardiac output at maximal exercise.
Cardiac Output Assessment
Fick method. First, a 7.5-F fiber-optic catheter (CCOmbo, Edwards Lifesciences, Irvine, CA) was positioned in the right pulmonary artery under fluoroscopic guidance through the antecubital vein and connected to a hemodynamic monitor (Vigilance II, Edwards Lifesciences). A 20-gauge arterial catheter was then placed into the radial artery and connected to a pressure monitor (SC9000, Siemens Medical Systems, Erlangen, Germany), using a disposable pressure transducer (Safedraw, Becton-Dickinson, Franklin Lakes, NJ). In addition, a pulse oximeter (Onyx 9500, Nonin Medical, Plymouth, MN) was attached to the index or middle finger to monitor arterial oxygen saturation (SaO2) continuously.
Cardiac output (
) was calculated using the Fick equation:
![]() |
by heart rate.
For cardiac output assessments at rest, the direct Fick method (
dF) was applied as reference, using Hb and oxygen saturation (SO2) values obtained from blood samples drawn from the radial and pulmonary artery that were analyzed by an in vitro oximeter (ABL800 Flex, Radiometer America, Copenhagen, Denmark). These measurements were performed twice in an upright position before the start of the constant-load exercise tests (time interval
5 min) and repeated during the preceding resting period of the symptom-limited maximal test.
During exercise, the continuous Fick method was applied as a reference method, using continuously measured
O2, SaO2, and mixed venous oxygen saturation (SvO2) data. After the tests, these data were resampled into 0.5-s intervals and synchronized by using event markers. The second resting Hb measurement was used to calculate cardiac output (
cF). The continuous Fick method was used because it allows multiple paired comparisons with the other methods. Previous studies demonstrated that this method is feasible and accurate during exercise in patients with left ventricular dysfunction (31, 43).
Pulse contour analysis (LiDCO).
The pulse contour analysis method (LiDCO, London, UK) provides beat-to-beat changes in stroke volume, by calculating nominal stroke volume from a pressure-volume transform of the radial artery pressure waveform. Heart rate is calculated by the duration between subsequent pressure waveforms. To convert nominal stroke volume to absolute stroke volume, the system needs to be calibrated. An advantage of this method is that a calibration method (indicator dilution) is already incorporated. The indicator, lithium chloride, can be administered through a central or peripheral vein (13). The indicator dilution curve is generated by a flow-through cell containing a lithium-selective electrode, which is attached to the arterial line (16). Cardiac output is calculated from the dilution curve, using the following equation:
![]() |
After the patient was properly positioned on the ergometer, the calibration procedure started. First, the LiDCO plus monitor was connected to the SC9000 pressure monitor, which, in its turn, was connected to the radial artery catheter. Subsequently, 2 ml lithium chloride (30 mM LiCl) was injected into the right atrium through the pulmonary artery catheter. This measurement was repeated after 5 min and during the resting period preceding the symptom-limited exercise test. These resting measurements were used for comparison with the other cardiac output measurement methods. Immediately after each lithium dilution measurement, blood samples were drawn for determination of the reference
dF. Therefore, the direct Fick method and lithium dilution were not performed simultaneously (delay: 10–15 s). Normally, only one lithium dilution measurement is required for calibration. We used the second lithium dilution measurement for this purpose, because this was the last measurement before the start of the exercise protocol. After the calibration procedure, uncalibrated cardiac output values were automatically corrected to calibrated values.
During the exercise protocol, cardiac output (
Li) and stroke volume values (SVLi) were stored beat-to-beat for offline analysis.
Physioflow. Physioflow is an impedance technique based on the principle that variations in the impedance to a high-frequency (75 kHz) low-magnitude (1.8 mA) alternating current across the thorax during cardiac ejection result in a waveform from which SV can be calculated. Initially, stroke volume index is calculated at rest by evaluating 24 consecutive heart beats (autocalibration procedure), using measurements of the largest impedance difference during systole, the largest rate of variation of the impedance signal (contractility index), the thoracic fluid inversion time, heart rate, and the pulse pressure (= difference between systolic and diastolic arterial pressure) (8). Cardiac output is then calculated by multiplying the stroke volume index with body surface area and heart rate, which is obtained from the R-R interval determined on the ECG first derivative (8). Compared with the commonly used Sramek-Bernstein approach (5), Physioflow does not require estimations of baseline thoracic impedance or blood resistivity. Furthermore, as already stated, the position of the electrodes is not critical for the accuracy of the measurements.
After cleaning the skin, two pairs of electrodes (FS50, Skintact, Innsbruck, Austria) were positioned at the left base of the neck and along the xiphoid for transmitting and receiving electrical currents. Two electrodes were also placed on the chest (V1/V6 position) to obtain the ECG signal. The autocalibration procedure was started after a period of at least 5 min, in which patients were sitting immobile on the cycling ergometer. The SV (SVPh) and cardiac output (
Ph) values were stored beat-to-beat.
Data Analysis
All cardiac output data were imported into MATLAB (Mathworks, Natick, MA). After removal of outliers (30), data were synchronized by cross-correlating the heart rate data of the different assessment methods. Subsequently, data were averaged into 30-s intervals. Resting cardiac output and SV values obtained by LiDCO and Physioflow were compared with the direct Fick method at three different time moments, resulting in 30 measurements (10 patients x 3 measurements). During exercise, comparisons of simultaneous measurements were made during the last 2 min of constant-load exercise, resulting in four comparisons at both exercise intensities (i.e., a total of 80 measurements: 10 patients x 8 measurements), and during the last minute of the symptom-limited maximal test, resulting in two comparisons (i.e., a total of 20 measurements).
As outlined by Linton and Linton (25), the between-patient variability in cardiac output can have a substantial influence on the agreement between different assessment methods of absolute cardiac output values. One may expect that the presence of within-patient changes in cardiac output may be of even greater relevance for the clinical utility of a cardiac output assessment method. Therefore, as in other studies (23, 24), we evaluated within-patient changes in cardiac output by calculating relative rather than absolute changes (i.e., %increase relative to the second resting cardiac output value). An advantage of this approach is that relative changes are not influenced by the absolute resting cardiac output values.
Statistical Analysis
All data were analyzed using SPSS 12.0 statistical software (SPSS, Chicago, IL). Differences between continuous data were evaluated by the paired Student's t-test. Pearson's correlation coefficient (r) was calculated to quantify relations between variables. Differences between the assessment methods at rest and during exercise were evaluated by one-way ANOVA with Bonferroni post hoc analyses. The variation between measurements at rest and during steady-state exercise (repeatability) was evaluated by the coefficient of variation (SD of difference as a percentage of the mean). Agreement between methods was assessed by the approach described by Bland and Altman (6). The bias and limits of agreement (= bias ± 1.96 SD) among the methods are presented as a percentage of the mean values rather than as absolute values. This was done because it allows for a better evaluation of the relationship between cardiac output values and the size of the error (10). Because there was no difference in heart rate among the different methods, bias and limits of agreement are equal for cardiac output and SV and therefore are only provided for cardiac output. Data are presented as means ± SD. P values of <0.05 were considered significant.
| RESULTS |
|---|
|
|
|---|
dF measurements could not be analyzed because of blood clot formation in the syringe (n = 3) or technical problems (n = 2). The indicator dilution measurements were not performed in two patients owing to logistic reasons. Because of aberrant shapes of the dilution curves, 4 of the remaining 24 measurements could not be used for further analysis, with at least 1 valid measurement in each of the remaining eight patients. A total of 6 of 30
Ph measurements could not be analyzed because of an insufficient quality of the impedance signal. The mean coefficients of variation of the measurements at rest were 12.1% for the direct Fick method, 9.0% for lithium dilution, and 7.4% for Physioflow.
All subjects completed the exercise tests. During exercise, 8 of 100
cF measurements could not be used because of an unstable SO2 signal from the fiber-optic catheter. As stated before, indicator dilution measurements could not be performed in two patients, resulting in a total of 80 available calibrated LiDCO measurements (
Li). Of these measurements, 13 had to be excluded from further analysis due to damping or motion artifacts of the radial artery pressure tracing, with at least two remaining valid measurements per exercise intensity level in each patient. An insufficient quality of the impedance signal was observed in 24 of 100
Ph measurements. The mean coefficients of variation of consecutive measurements at light steady-state exercise were 4.5% for the continuous Fick method, 1.8% for LiDCO, and 2.4% for Physioflow. At moderate steady-state exercise these coefficients of variations were 5.9%, 2.3%, and 3.6%, respectively.
Table 2 summarizes cardiac output and SV values of the different methods at rest, during light and moderate constant-load exercise, and at peak exercise with corresponding
O2 and workload values. Considering the constant-load exercise tests, none of the subjects showed an increase in heart rate of more than 5% during the final 2 min of exercise (mean change: +0.9%, range: –4% to +4%). The mean change in
O2 during this period was +0.5% (range: –9% to +5%). This indicates that a steady state was attained in both constant-load tests.
|
ANOVA revealed overall differences in
and SV values at rest between the three assessment methods (P < 0.001); post hoc testing showed only significant differences in
and SV values between Physioflow and both other methods (P < 0.001). Table 3 summarizes the bias and limits of agreement among the different methods. Both
Li and SVLi were positively correlated with
dF and SVdF (r = 0.50, P = 0.04; and r = 0.71, P = 0.001, respectively). Neither
Ph nor SVPh was significantly correlated with values obtained by the other methods.
|
ANOVA revealed overall differences in
and SV values among the methods (P < 0.001); post hoc testing revealed that only
Ph and SVPh were significantly different from values obtained from LiDCO or the continuous Fick method (P < 0.001). The bias and limits of agreement between the methods at different exercise levels are summarized in Table 3, and Fig. 1 provides a graphic representation. A positive correlation was found between
Ph and
cF during exercise (r = 0.73, P < 0.001, Fig. 1B); however, SV values of these methods were not significantly correlated. There were highly significant positive correlations between
Li and
cF (r = 0.90, P < 0.001, Fig. 1A) and also between SVLi and SVcF (r = 0.76, P < 0.001). The lowest bias and narrowest limits of agreement were observed between
Li and
cF (0% ± 27%, Fig. 1A). Bias and limits of agreement between
Ph and both other methods were from the same order of magnitude (48% ± 52% for
Ph vs.
cF and 52% ± 50% for
Ph vs.
Li, Fig. 1, B and C).
|
When comparing relative within-patient changes in
, ANOVA showed no significant overall differences between the methods. However, significant differences between changes in SV values were observed (P = 0.02), with only a significant difference between Physioflow and LiDCO in post hoc analysis (P = 0.03). Correlations and Bland-Altman plots comparing within-patient changes of cardiac output are depicted in Fig. 2. The limits of agreement between changes of
cF and
Li ( ± 26%) were somewhat narrower than in the other comparisons (
cF vs.
Ph: ± 36%; and
Li vs.
Ph: ± 31%). In all comparisons the bias was close to zero (2%, –2%, and –3%, respectively). Significant correlations were observed among the different methods (Fig. 2, A–C). In addition, changes of both SVLi and SVPh showed positive correlations with changes of the reference SVcF (r = 0.67, P < 0.001; and r = 0.41, P < 0.001, respectively), as well as with each other (r = 0.51, P < 0.001). Considering the transition from "rest" to light steady-state exercise (i.e., the smallest increase in exercise intensity), there were no significant differences between relative changes among the methods. Bias and limits of agreement for changes at this transition were 2% ± 27% for
cF vs.
Li, –4% ± 42% for
cF vs.
Ph, and –4% ± 25% for
Li vs.
Ph.
|
| DISCUSSION |
|---|
|
|
|---|
O2 with respect to SaO2 and SvO2 (14) can be compensated. Therefore, it is questionable whether the Fick principle can be applied during peak exercise. Yamabe et al. (43), however, reported that cardiac output obtained by the continuous Fick method during ramp exercise showed good agreement with values during steady-state exercise at the same intensity in CHF patients, suggesting that the Fick principle can also be applied during peak exercise in these patients.
The accuracy of the continuous Fick method mainly depends on operator-related factors and the summated inaccuracies of the measurements of
O2, SaO2, SvO2, and Hb. According to the manufacturer's specifications, the gas exchange monitor for assessment of
O2 had a precision error of 1.5% (defined as 1 SD of the difference with the actual value). The used fiber-optic catheter had a precision error of 4% (37). Assuming precision errors of 3% for SaO2 (39) and 2% for Hb (11), we calculated a theoretical precision error of 11% for the continuous Fick method, with corresponding limits of error of 22% (28). Because we considered a measuring method clinically useful when its limits of error approximated the continuous Fick method, limits of agreement up to 31% were accepted (10). It should be recognized that these limits only apply to the comparison of absolute values and not to the changes of cardiac output, because the precision error of the continuous Fick method to detect these changes is not exactly known.
Although LiDCO has been shown to be useful to assess cardiac output in patients undergoing major surgery (15, 24), its accuracy during exercise conditions has not been evaluated previously. In a study using a different pulse contour analysis method, it was shown that this method is not suitable for measuring absolute cardiac output values during exercise in healthy individuals (33). However, it was also demonstrated that initial calibration with another method improved the accuracy substantially (38). As already stated, an advantage of our method is that an indicator dilution method as calibration is already incorporated. Although not all calibration measurements were successful in our study (failure of 17%), at least one valid measurement could be obtained in all patients. As only one calibration is required, this failure percentage does not have any impact on the potency of this technique in clinical practice. In agreement with other studies (19, 26), we observed an acceptable level of agreement with the reference method, and the repeatability of measurements was higher than the reference method (coefficient of variations: 9.0% and 12.1%, respectively). However, despite the high accuracy of the lithium dilution technique, there are some drawbacks to its clinical use. First, it is relatively time-consuming and expensive. Second, it requires an additional peripheral intravenous access for administration of lithium chloride (13). Therefore, it is valuable to consider the use of noninvasive methods for calibration of this technique, such as Doppler cardiography (35) or open-circuit inert gas techniques (4).
A major advantage of using LiDCO is that, once calibrated, it provides real-time beat-to-beat cardiac output data at rest and during exercise. However, an essential prerequisite remains a good quality of the radial artery pressure waveform. In our study, 16% of the calibrated measurements could not be used for further analysis because of the influence of damping or motion artifacts on the waveform during exercise. Yet, we could obtain two or more valid measurements at each exercise intensity level in all subjects, indicating that this technique is feasible during exercise in CHF patients. Considering the accuracy, we observed a low bias and acceptable limits of agreement between this technique and the continuous Fick method, both for the assessment of absolute values as well as for changes in cardiac output. However, this technique may somewhat overestimate cardiac output values exceeding 10 l/min (Fig. 1A), possibly due to an overestimation of resting cardiac output during the calibration procedure. This view is supported by the observation that an obvious overestimation of changes in cardiac output in the higher range is lacking (Fig. 2A). Another explanation for the possible overestimation of higher cardiac output values may be related to the fact that the algorithm that is used to transfer changes in arterial pressure to changes in volume is based on data obtained from healthy individuals. Yet, it was shown previously that exercise-induced changes in the relation between arterial pressure and flow (arterial input impedance) in CHF patients were not different from changes in healthy individuals (21, 22), suggesting that the algorithm is also suitable for CHF patients.
Despite the possible overestimation of cardiac output values exceeding 10 l/min, this study showed that LiDCO is accurate for assessment of absolute cardiac output values in a range that is normally obtained in CHF patients. Importantly, this technique was also able to detect accurately relatively small changes in cardiac output. Therefore, we postulate that pulse contour analysis is potentially useful for clinical purposes, such as assessment of prognosis or selection of candidates for heart transplantation (9), and for research purposes such as investigating the physiological determinants of exercise tolerance in CHF patients (17, 18, 29).
Impedance cardiography is a simple, inexpensive, and completely noninvasive method to assess cardiac output. The accuracy of impedance cardiography during exercise has been an issue of debate for many years, despite advancements in technology. Common problems that may compromise the quality of the impedance signal are movement and respiration artifacts (41), as well as the presence of pulmonary congestion (32). In the present study, 24% of the measurements obtained during exercise could not be analyzed because of insufficient quality of the signal. This relatively high percentage of failure may have been caused mainly by an irregular or oscillatory breathing pattern (12). As none of our CHF patients showed any clinical signs of pulmonary congestion, we believe that this did not significantly affect our study results. Moreover, pulmonary congestion has been shown to cause an underestimation of cardiac output by impedance cardiography (32), rather than an overestimation, which we observed. Despite the relatively high failure rate of Physioflow in our study, we obtained successful measurements at all exercise intensity levels in eight patients.
Considering the accuracy of Physioflow during exercise, previous studies in healthy individuals demonstrated a high level of agreement between Physioflow and the direct Fick method (8, 34). In contrast, the results of our study showed poor agreement with the reference method, with a systemic overestimation of SV by Physioflow. However, as shown by Bland-Altman analysis, the relative bias (i.e., expressed as a percentage of the mean value) between Physioflow and the reference method was almost similar under rest and exercise conditions (Table 3). This suggests that the discrepancy between both methods during exercise is mainly caused by an overestimation of SV during the calibration procedure. This notion is supported by the fact that the bias between Physioflow and the reference method for estimating relative changes is negligible (Fig. 2B). In a recent study with COPD patients, SV assessed by the Fick method was also overestimated by Physioflow (7). It is striking that the mean resting SV value in these COPD patients was comparable with the CHF patients in our study (60 and 57 ml, respectively). As no significant bias between Physioflow and the Fick method was observed in healthy individuals with higher SV values (8, 34), we postulate that the autocalibration algorithm of Physioflow is less suited for patients with relatively low SV values.
Despite the low accuracy of Physioflow for the assessment of absolute cardiac output values, we demonstrated a reasonable overall agreement between Physioflow and the reference method for the assessment of relative changes in cardiac output and SV. However, as shown by the Bland-Altman plot (Fig. 2B) and the relative wide limits of agreement between both methods for cardiac output changes from rest to light constant-load exercise ( ± 42%), the accuracy of Physioflow to detect subtle changes in cardiac output is limited. Yet, as greater changes in cardiac output (i.e., >200%) can be assessed with reasonable accuracy (Fig. 2B), Physioflow may still be useful for clinical purposes such as the evaluation of the cardiac output response to maximal exercise in mild to moderate heart failure, or the evaluation of exertional dyspnea during cardiopulmonary exercise testing.
To our knowledge, there are no studies that have evaluated reproducibility of LiDCO and Physioflow during exercise in CHF patients, and also our study design did not permit an assessment of day-to-day reproducibility of these methods. However, we evaluated the variation of subsequent measurements of the different methods. Both at rest and during constant-load exercise, the mean coefficients of variation of LiDCO and Physioflow were lower than those obtained from the reference method. Another limitation of this study is that only moderately impaired male patients were included. Consequently, our observations cannot be generalized to women or more severely impaired patients.
In conclusion, although the limits of agreement with the Fick method are pretty broad, the accuracy of LiDCO is clinically acceptable for the assessment of absolute values of cardiac output as well as within-patient changes during exercise in a range that is normally obtained by CHF patients. There may, however, be a slight overestimation by values exceeding 10 l/min. Whereas Physioflow offers the advantage of being completely noninvasive, relatively inexpensive, and easy to use, it systematically overestimates cardiac output during exercise, due to an overestimation of stroke volume at rest. However, provided that changes in cardiac output are not too small (i.e., <200%), this method may still be used to assess the relative increase in cardiac output during exercise in CHF patients. Although the failure rates of both LiDCO and Physioflow during exercise were relatively high (16% and 24%, respectively), the number of valid measurements was sufficient for these methods to be feasible in clinical practice.
| GRANTS |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
O2 determination at rest and during exercise in patients with heart failure. Am J Cardiol 99: 404–405, 2007.[CrossRef][Web of Science][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||