J Appl Physiol 103: 867-874, 2007. First published June 7, 2007; doi:10.1152/japplphysiol.01106.2006
8750-7587/07 $8.00
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow All Versions of this Article:
103/3/867    most recent
01106.2006v1
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jarvis, S. S.
Right arrow Articles by Pawelczyk, J. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jarvis, S. S.
Right arrow Articles by Pawelczyk, J. A.

Simultaneous determination of the accuracy and precision of closed-circuit cardiac output rebreathing techniques

S. S. Jarvis,1 B. D. Levine,2 G. K. Prisk,3 B. E. Shykoff,4 A. R. Elliott,3 E. Rosow,5,6 C. G. Blomqvist,7 and J. A. Pawelczyk1,2

1Department of Kinesiology, Pennsylvania State University, University Park, Pennsylvania; 2Institute for Exercise and Environmental Medicine, University of Texas Southwestern Medical Center at Dallas and Presbyterian Hospital of Dallas, Dallas, Texas; 3Department of Medicine, University of California at San Diego, La Jolla, California; 4Department of Physiology, State University of New York at Buffalo, Buffalo, New York; 5Hartford Hospital, Hartford; 6Premise Development Corporation, Hartford, Connecticut; 7Cardiology Division, Department of Internal Medicine, University of Texas Southwestern Medical Center at Dallas, Dallas, Texas

Submitted 2 October 2006 ; accepted in final form 30 May 2007


    ABSTRACT
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
Foreign and soluble gas rebreathing methods are attractive for determining cardiac output (Qc) because they incur less risk than traditional invasive methods such as direct Fick and thermodilution. We compared simultaneously obtained Qc measurements during rest and exercise to assess the accuracy and precision of several rebreathing methods. Qc measurements were obtained during rest (supine and standing) and stationary cycling (submaximal and maximal) in 13 men and 1 woman (age: 24 ± 7 yr; height: 178 ± 5 cm; weight: 78 ± 13 kg; VO2max: 45.1 ± 9.4 ml·kg–1·min–1; mean ± SD) using one-N2O, four-C2H2, one-CO2 (single-step) rebreathing technique, and two criterion methods (direct Fick and thermodilution). CO2 rebreathing overestimated Qc compared with the criterion methods (supine: 8.1 ± 2.0 vs. 6.4 ± 1.6 and 7.2 ± 1.2 l/min, respectively; maximal exercise: 27.0 ± 6.0 vs. 24.0 ± 3.9 and 23.3 ± 3.8 l/min). C2H2 and N2O rebreathing techniques tended to underestimate Qc (range: 6.6–7.3 l/min for supine rest; range: 16.0–19.1 l/min for maximal exercise). Bartlett's test indicated variance heterogeneity among the methods (P < 0.05), where CO2 rebreathing consistently demonstrated larger variance. At rest, most means from the noninvasive techniques were ±10% of direct Fick and thermodilution. During exercise, all methods fell outside the ±10% range, except for CO2 rebreathing. Thus the CO2 rebreathing method was accurate over a wider range (rest through maximal exercise), but was less precise. We conclude that foreign gas rebreathing can provide reasonable Qc estimates with fewer repeat trials during resting conditions. During exercise, these methods remain precise but tend to underestimate Qc. Single-step CO2 rebreathing may be successfully employed over a wider range but with more measurements needed to overcome the larger variability.

foreign gas rebreathing; physiological gas rebreathing; direct Fick; thermodilution


CARDIAC OUTPUT (Qc), the volume of blood pumped by the heart each minute, can be measured several ways, both invasively and noninvasively. Each technique has advantages and limitations that may restrict its application during exercise. The direct Fick method has been implemented in humans in clinical environments since 1940 (21); together with the thermodilution method, they are considered "gold standard" Qc techniques. These methods, however, are expensive, require medical expertise, and are invasive, limiting their widespread utility (44). For instance, occasional reported complications associated with these gold standards include arrhythmias and pulmonary artery or right ventricle perforation (30, 40).

Rebreathing techniques, employing foreign or physiological tracers, are attractive alternatives because their relative risk is considerably less (44). Pulmonary blood flow may be estimated by rebreathing a gas mixture containing a foreign, soluble gas and measuring the rate of disappearance of this gas (often determined by mass spectrometry; Ref. 42). Foreign gases [such as acetylene (C2H2) and nitrous oxide (N2O)] are commonly used in rebreathing techniques because they are inert, soluble, and enter the blood stream via pulmonary diffusion but do not bind with hemoglobin (44). Often an inert insoluble gas (e.g., He, Ar, CH4, or SF6) is added to the gas mixture to determine system (bag/lung) volume and adequacy of gas mixing in the lungs (42). Alternatively, the rate of accumulation of a physiological soluble gas [carbon dioxide (CO2)] in the rebreathing mixture may be employed. In normal subjects, pulmonary blood flow is proportional to Qc (44). That is, if no diffusion limitation or physiological shunt is present, the rate of uptake of the foreign gas is a function of 1) the solubility coefficient of the gas, 2) pulmonary blood flow (cardiac output), and 3) lung tissue volume (37, 42).

Techniques based on these theories have been implemented in clinical settings (27), in exercise physiological research, during ground-based deconditioning studies (22), in low-earth orbit (39), and post spaceflight (34, 39). Previous studies have compared criterion methods against gas rebreathing techniques (13, 25, 32, 35) and others have examined these methods during exercise (23), but a simultaneous comparison of multiple rebreathing techniques has not been conducted over a wide range of postures and metabolic states. Thus the purpose of this investigation was to compare simultaneously obtained criterion and rebreathing data, during rest and exercise, to provide a determination of the accuracy and precision of these techniques.


    MATERIALS AND METHODS
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
Subjects

Thirteen men and one woman (age: 24 ± 7 yr; height: 178 ± 5 cm; weight: 78 ± 13 kg; VO2max: 45.1 ± 9.4 ml·kg–1·min–1; mean ± SD) volunteered to participate. All subjects were healthy, nonsmoking individuals that were recruited from the surrounding community. Most were active in recreational sports; no one was a competitive athlete at the time of the study. Volunteers received a physical exam, performed a maximal exercise test (VO2max), and performed 5–10 Qc rebreathing maneuvers during screening visits to become familiar with the procedures. Each subject was studied on a single day within 2 wk of screening. All aspects of this study were approved by the Institutional Review Board at the Presbyterian Hospital of Dallas and the University of Texas Southwestern Medical Center.

Criterion Techniques: Direct Fick and Thermodilution

Direct Fick.   A flow-directed thermistor-tipped pulmonary artery catheter was inserted from the antecubital fossa and advanced under fluoroscopic guidance into the main or proximal right pulmonary artery. Samples of pulmonary artery blood were withdrawn for determination of mixed venous oxygen content (CvO2) in the middle of the 1-min Douglas bag collection. Hemoglobin concentration and oxyhemoglobin saturation were determined using co-oximetry (IL 482, Instrument Laboratories). Qc was determined by using the Fick equation: Qc = VO2(CaO2 CvO2), where CaO2 is arterial oxygen content with SaO2 derived from ear oximetry (Ohmeda 3700 ear probe), and VO2 is oxygen uptake determined by open-circuit spirometry (Douglas bags).

Thermodilution.   Determinations were based on the temperature changes recorded at the pulmonary artery after 10 ml of 0°C isotonic saline was injected into the right atrium. Saline was stored in an ice bath until just before use. Careful attention was paid to minimize the time between removal of the syringe from the ice bath and injection into the right atrium. All curves were inspected visually, and only curves with apparent exponential decay were used for data analysis. The average of three measurements was used for each determination. Q was calculated by a Baxter cardiac output computer using the modified Stewart-Hamilton equation: Q = [V(TB – T1)K1K2]/{int}{Delta}TB(t)dt, where V is volume of isotonic saline injected; TB is temperature of blood; T1 is temperature of injectate; and K1 and K2 are constants.

Gas Rebreathing: Foreign and Physiological Soluble Gases

Five independent investigative teams developed six rebreathing methods: one-N2O, four-C2H2, and one-CO2 (revised single step). As the principle underlying these techniques is mass balance, they are fundamentally similar. In this investigation, the differences were the choice of soluble gas (CO2 vs. N2O vs. C2H2), the insoluble gas to determine system volume (Ar or He), and the analysis software independently developed by each team and used in their respective laboratories. To provide a simultaneous comparison, data were acquired using a gas mixture of O2 40%, CO2 0%, N2 balanced, Ar 5%, He 9%, N2O 2%, C2H2 0.600%. The gas mixture contained two inert insoluble gases since determinations with the C2H2 and N2O rebreathing methods use He as an insoluble gas, whereas the CO2 method used Ar. Each investigative team independently calculated Qc from the gas concentrations obtained during the same rebreathing maneuver. The rebreathing bag was prefilled so that subjects rebreathed 2.5 liters of the gas mixture during the resting conditions (supine and standing) and 3.0–3.5 liters during exercise for 25 s at a rate of 22 breaths/min. Gas concentrations were continuously analyzed via mass spectrometry (MGA-1100, Perkin Elmer). This combination of gas mixtures, bag volume, rebreathing time, and breathing frequency was devised to accommodate, to the greatest extent possible, all techniques within the experimental design. A turbine flowmeter (VMM-400, Interface Associates) was included between the subject's mouth and the rebreathing valve for those implementations requiring a volume/time signal. The catheter transit time was measured in response to a step change in gas concentration to align volume and gas concentration signals temporally.

N2O.   This method has been used previously to obtain measurements in astronauts, both pre- and in-flight, for evaluating the effects of microgravity on Qc (34). Qc measurements were determined by using the Sackner algorithm by fitting a line over all of the alveolar plateau regions from expiration after clearing the deadspace. Qc was calculated as: Qc = [(VA x 760)/{alpha}N2O(FAN2O intercept/F AN2O')(PA PH2O)] x [{Delta}ln(FAN2O/FAN2O'/{Delta}t], where Qc is pulmonary capillary blood flow; VA is alveolar volume (STPD) determined by He dilution; {alpha}N2O is the Bunsen solubility coefficient for N2O (0.407); FAN2O is the alveolar fraction of N2O, corrected for He dilution (FAN2O' is the initial fraction and FAN2O intercept is the intercept value corrected for timing); and PA PH2O is barometric pressure of the dry gas mixture (38).

The first three breaths of each maneuver were discarded for this method's analysis because thorough mixing of the bag-lung system had not been achieved. Thus, only the breaths containing stable He concentrations were used to determine Qc. N2O disappearance curves were calculated using a least-squares fit and the disappearance curve's intercept was shifted to time zero (3). Time zero was defined as the midpoint of the rapid rise in N2O concentration of the first test inspiration.

C2H2.   Similar to N2O rebreathing, Qc determinations with C2H2 rebreathing used He to compute alveolar volume by VA = I[He]/F[He] x VI, where I[He] is initial fraction of inspired He; F[He] is final He fraction; and VI is inspired volume in liters (STPD). The differences in determining Qc from each C2H2 rebreathing method are outlined below.

C2H2 (method 1).   Qc determinations for this method were identical to the N2O method described above. The equation used to calculate Qc was identical except that the solubility coefficient for C2H2 (0.768) was adjusted accordingly.

C2H2 (method 2).   This C2H2 rebreathing method has been previously used to determine Qc before and after spaceflight (31). It was the only fully automated method, providing valve switching (Hans Rudolph Series 2540) to connect the inspiratory circuit to the rebreathing bag. Since this method does not require a prefilled bag, the inspired fraction of C2H2 was calculated from the volume weighted integral of [C2H2] measured during the first inspiration of a known mixture. The rate of C2H2 absorption into blood was computed from the regression slope of the logarithm of regressed end-tidal C2H2 fractions. The end-tidal points were selected by the software based on Sackner's criterion, only for breaths where a stable [He] had been established (38). Time zero was defined from the initial rapid rise in C2H2 concentration (point at which inspired acetylene concentration exceeded 0.0025%), less the time required to inspire the sum of the measured mouthpiece and estimated physiological deadspace volume.

C2H2 (method 3).   This implementation was originally published by Triebwasser et al. (42) and used more recently in a study that examined cardiac atrophy during 6° head-down bed rest (5). C2H2 and He concentrations were averaged during the last 40 ms of each expiration, as determined by the zero crossing of the flowmeter. The slope of the change in [C2H2] over time was derived during each breath from the end respiratory gas concentrations (42). Since there is some uptake, albeit small, of C2H2 by the lung parenchyma upon the first inspiration, the intercept was back extrapolated to determine time zero by: intercept = (Vs)/{Vs + {alpha}t[(PA – PH2O)/(760)]} x Vt, where Vs is system volume (alveolar + bag); {alpha}t is Bunsen solubility coefficient for C2H2 (0.768); PA – PH2O is barometric pressure of the dry gas mixture; and Vt, which can be estimated by Vt = [Vs(1/intercept – 1)/(0.768)] x [760/(PA – PH2O)](42). Finally, to calculate Q, the following equation was used: Q = {[(Vs/intercept) x (760/ PA – PH2O)]/0.700} x (–m), where –m {ln[(FA2 x FHe1)/(FA1 x FHe2)]/(t2–t1)} was estimated over several breaths obtained by linear regression and the slope of the concentration ratio over time, FA1 and FHe1 and FA2 and FHe2 are the fractional C2H2 and He concentrations for the first and second samples, respectively; and t2 – t1 was the time in seconds between the two samples (42). Additionally, a partition coefficient of 0.700 for C2H2 between gas and blood was used (42).

C2H2 (method 4).   This method was developed as a prototype. Determinations made with this prototype method used a C2H2 rebreathing technique similar to the Sackner implementation described above.

CO2 (single step).   Although several different CO2 rebreathing methods have been developed (e.g., Refs. 10, 20), only the modified one-step method was employed in this investigation because its initial conditions and rebreathing technique were compatible with the other methods under study. This method has been used to determine Qc under a number of altered environmental conditions that include microgravity (39), water immersion, and head-down tilt (29). None of the other CO2 rebreathing methods could have been used in this investigation because they require a separate, steady-state measurement of CO2 production, and several rely on equilibration of CO2 in the gas and blood or unique respiratory maneuvers.

The modified one-step rebreathing maneuver is based on the fact that any CO2 leaving the blood during rebreathing enters either the gas phase or the tissue stores. For the ith breath once the rebreathing bag and lung gas are well mixed, the mass balance of CO2 states that Qc·(CvCO2 – CcCO2i)(ti+1 – ti) = Vgasi+1·FCO2i+1 Vgasi·FCO2i + Vt·(FCO2i+1 – FCO2i), where Vgas is the total gas volume, rebreathing bag + lung, FCO2i is the fraction of CO2 in the gas phase at time ti, and Vt is the volume of lung tissue expressed as equivalent gas volume.

A breath is defined to begin at the onset of the alveolar plateau. Because the rebreathing bag is empty at the end of inspiration, the gas sample is representative of the system. A fitted dissociation curve is applied to each value of PACO2 during the breath and the values are averaged to give CcCO2i; PACO2 is either the measured value during expiration, or the linear interpolation between end expiration and the start of the next alveolar plateau during inspiration. Vgas was calculated for each breath using the mass balance of N2 and Ar, and, in this study, corrected for the gases not usually in the rebreathing mix for this method.

Mixing of the bag-lung system was assessed from the concentration of Ar; in general, the gases were well mixed after the third breath. The same mass balance equation was applied also to the first, poorly mixed breaths in toto, with t0 as the end-tidal point before rebreathing. This yields a system of equations in the unknowns (Qc, CvCO2, and Vt) to be rearranged and solved by linear regression methods. In the usual implementation of this method, the breathing frequency and rebreathing times are set to obtain at least 12 breaths and nine equations per rebreathing maneuver, but in this multi-method trial we compromised with eight to nine complete breaths and six or seven equations.

The original one-step method (28) needs a rebreathing bag large enough that FCO2 remains below FCO20, the normal end-tidal value, for three or four breaths. To obviate the need to consider tissue stores of CO2, that method then measures the amount of CO2 added to the gas phase only in the time from the start of the hyperventilation of rebreathing at t0 to the time when FCO2 in the system again reaches FCO20. CcCO2 is obtained for the measurement period as in the modified method, but CvCO2 is obtained, using the dissociation curve, from a calculated equilibrium PeqCO2 obtained by extrapolation to the PCO2 at which the slope of the curve of end-tidal values would equal zero. However, in the modified one-step method used in these experiments, CvCO2 is estimated as the CO2 content remaining in the blood after its passage through the lungs, not as equilibrated CO2 content. The assumption is made that alveolar and mixed capillary PCO2 are the same, but CO2 is leaving the blood throughout the measurements and full oxygen saturation is ensured because the rebreathing starts with 40% oxygen.

Resting Conditions: Supine and Standing

After catheter placement, the subjects rested quietly for 30 min while under observation. Subjects were then taken to the lab to be instrumented with ECG and noninvasive blood pressure monitoring (Suntech 4240).

Qc was collected during supine rest, standing rest, submaximal exercise on a cycle ergometer at 100 W, and during maximal exercise. The supine resting protocol began with 2 min of allowing the subject to reach steady state, 1 min to obtain mixed venous samples, 1 min rebreathing, and then 1 min of thermodilution measurements. This cycle was repeated until four measurements were collected. The subject was then asked to move into the standing position and allowed 5 min to achieve steady state, after which the same protocol was repeated. Once the standing measurements were obtained, we allowed 15 min to prepare for the exercise protocol.

Exercise Conditions: Submaximal and Maximal

The subject exercised at 100 W on a cycle ergometer for 15 min, while Qc was collected four times, every 3 min, in the same manner as above. This workload was chosen because it allowed all subjects to maintain steady state for at least 15 min. To obtain the maximal exercise Qc measurement, the subject continued to exercise at the end of the submaximal bout while the workload was increased by 20 W/min until maximal effort was achieved. Qc was collected once during maximal exercise.

Statistical Analysis

Each investigative team acquired analog representations of gas concentrations for subsequent digital analysis with their own unique software. The derived values for Qc were then consolidated and analyzed with a condition x technique analysis of variance. When significant main effects were observed, comparisons between the criterion techniques (direct Fick and thermodilution) and the rebreathing methods were assessed using Tukey's honestly significant difference. Bartlett's test with post hoc analysis was used to determine whether significant differences in variance existed (47). Average fractional deviations were calculated to determine whether gas rebreathing techniques fell within an a priori determined acceptable margin of accuracy (10%) compared with our criterion methods. In all cases, differences with P < 0.05 were considered significant.


    RESULTS
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
Of the 182 Qc determinations for each method, 91% of direct Fick data were appropriate for analysis; that is, they did not fall outside of three standard deviations from the mean or were not missing data points. Similarly, 96% were used for thermodilution, 94–96% for C2H2 and N2O rebreathing, and 71% for CO2 rebreathing.

An identity plot is shown in Fig. 1 for direct Fick vs. thermodilution. The slope of this relationship was not significantly different from the line of identity. Bland-Altman analysis (Fig. 2) revealed no systematic differences between the two techniques.


Figure 1
View larger version (11K):
[in this window]
[in a new window]

 
Fig. 1. Identity plot of direct Fick vs. thermodilution. The dashed line represents the line of identity and the solid line represents the line of best fit by linear least squares regression. The slope of the line of best fit for direct Fick vs. thermodilution was not statistically different from the slope of the line of identity (slope = 1).

 

Figure 2
View larger version (15K):
[in this window]
[in a new window]

 
Fig. 2. Bland-Altman plot for direct Fick and thermodilution. These criterion methods show agreement during resting conditions, but tended to differ during submaximal exercise.

 
Table 1 and Fig. 3 portray the means compared in the analysis of variance (standing, supine, submaximal exercise, and maximal exercise). Significant differences were observed between all techniques (criterion and rebreathing; P < 0.05). In particular, the single-step CO2 rebreathing method was generally greater than the other rebreathing techniques. All foreign gas techniques (C2H2 and N2O) had a propensity to underestimate Qc during submaximal and maximal exercise, compared with direct Fick and thermodilution (P < 0.05). Single-step CO2 rebreathing was not different from thermodilution during exercise, while all C2H2 and N2O techniques were different from both criterion techniques. During submaximal exercise, direct Fick and thermodilution were significantly different.


View this table:
[in this window]
[in a new window]

 
Table 1. Comparison of cardiac output determined with criterion and soluble gas rebreathing methods

 

Figure 3
View larger version (13K):
[in this window]
[in a new window]

 
Fig. 3. Cardiac output during each condition, by method. ANOVA indicated that the interaction between condition and method was significant (P < 0.05). C2H2 and N2O tend to underestimate cardiac output, while single-step CO2 rebreathing tends to overestimate cardiac output compared with the criterion methods. This difference is especially clear during maximal exercise. *Different from direct Fick, P < 0.05; {dagger}different from thermodilution, P < 0.05.

 
Table 1 provides additional results where the within-method variability of each technique was compared. Bartlett's test indicated variance heterogeneity among the techniques (P < 0.05). Further inspection showed larger variance with the single-step CO2 rebreathing method, which is shown in Table 1 and Fig. 4. In general, during rest, the foreign gas rebreathing techniques had lower variance than did direct Fick (Table 1); however, this difference did not exist during exercise.


Figure 4
View larger version (11K):
[in this window]
[in a new window]

 
Fig. 4. 95% Confidence intervals for the standard deviations of each technique. Resting conditions provided the smallest variability in all methods. Single-step CO2 rebreathing showed the largest variability and, in most cases, foreign gas rebreathing had lower variance than did direct Fick. Refer to Table 1 for values. *Different from direct Fick, P < 0.05; {dagger}different from thermodilution, P < 0.05; {ddagger}different from CO2 rebreathing, P < 0.05.

 
Figures 5 and 6 illustrate the average fractional deviations, by method, from the criterion techniques. Resting conditions provide the most reasonable estimates of noninvasive cardiac output measurement, with most techniques (3 of 7 during supine rest and 6 of 7 during standing) falling within 10% of our criterion values for direct Fick. Similarly, 6 of 7 of the techniques during supine rest fell within 10% of criterion means and 6 of 7 during standing rest for thermodilution. During submaximal and maximal exercise conditions, however, all rebreathing techniques fell above or below the ±10% range, with the exception of single-step CO2 rebreathing.


Figure 5
View larger version (22K):
[in this window]
[in a new window]

 
Fig. 5. Average fractional deviations from direct Fick. Resting conditions provided the most reasonable estimates of noninvasive cardiac output (Qc) measurement, with most techniques falling within ±10% of the direct Fick values. During both exercise conditions there was less agreement between gas rebreathing and direct Fick, with the exception of single-step CO2 rebreathing, which fell within the defined as acceptable range.

 

Figure 6
View larger version (21K):
[in this window]
[in a new window]

 
Fig. 6. Average fractional deviations from thermodilution. Resting conditions provided the most reasonable estimates of noninvasive Qc measurement, with most techniques falling within ±10% of the thermodilution values. During both exercise conditions there was less agreement between gas rebreathing and thermodilution, with the exception of single-step CO2 rebreathing, which fell within the defined as acceptable range.

 

    DISCUSSION
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
The present study was designed to compare eight techniques of Qc measurement in humans during different resting (supine and standing) and exercise (submaximal and maximal) conditions. Direct Fick and thermodilution served as criterion techniques, where foreign and physiological soluble gas rebreathing techniques were systematically compared with our criterion methods.

Direct Fick vs. Thermodilution

Our data agree with the established values in the literature relative to oxygen consumption (see Table 1) for a given condition (rest or exercise; Refs. 1, 12, 18). We found that direct Fick and thermodilution were not statistically different during rest (standing or supine) or maximal exercise. However, we found that during submaximal exercise, direct Fick was statistically higher than thermodilution. Complete consensus does not exist among investigations that have examined the relationship between direct Fick and thermodilution. Some have reported agreement (15), while some investigations (13, 43) reported that thermodilution overestimates Qc compared with direct Fick and others reported that thermodilution underestimates Qc (11).

It has been suggested that thermodilution measurements may provide inaccurate estimates because this method relies on heat as a marker. During injection of a cold indicator, not all heat transfer can be accounted for; potential sources of error include heat gain during handling of the injectate (36), heat transfer through the vessel wall, or heat exchange with the air in the lungs (24). It is also possible that the development of tricuspid regurgitation during exercise (4) could compromise the measurement of Qc by thermodilution (8). It is most likely, however, that the differences seen during submaximal exercise in our study, for several reasons, are best explained by measurement error in the variables used to calculate direct Fick. Direct Fick presumes that steady-state conditions are maintained throughout the determinations, an assumption that is violated as exercise intensity increases (44). Additionally, although oxygen uptake during rest and exercise was consistent within each condition (refer to Table 1), erroneous measurements in oxygen content of arterial or mixed venous blood could affect Qc when using the direct Fick method (11). These errors generally are magnified when the a-vO2 difference is small, since the error in measurement of oxyhemoglobin saturation has the potential to be a large part of the a-vO2 difference. It should also be noted that we used pulse oximetry to measure arterial oxygen saturation in this study to avoid the requirement for arterial catheterization.

We acknowledge the potential error introduced by estimating arterial oxygen content from pulse oximetry. However, previous investigators have evaluated the accuracy of the Ohmeda 3700 pulse oximeter and have concluded it provides reasonable estimates of arterial oxygen saturation. Powers et al. (33) examined its usefulness during graded exercise on a cycle ergometer, finding that the Ohmeda 3700 was accurate during exercise in nonsmoking individuals. Additionally, Kagle et al. (19) exposed a group of healthy volunteers to steady-state hypoxic and rapidly declining hypoxic conditions, noting that the Ohmeda 3700 ear probe was accurate throughout an arterial oxygen saturation range of 90–100%. At rest and during submaximal exercise, any differences between pulse oximetry and "true" arterial saturation were likely to be small, although more significant differences could be present at higher levels of exercise (46). Moreover, if pulse oximetry data were inaccurate, the direct Fick Qc would be expected to be biased high, yet thermodilution and direct Fick were in agreement in all conditions except for submaximal exercise. Additionally, no subject was an endurance athlete, limiting the likelihood that true changes in arterial saturation would occur. Thus, on the basis of the analysis of the literature and our own methodology, the criterion Qc data are accurate.

C2H2 and N2O Rebreathing vs. Criterion Techniques

The results of C2H2 and N2O rebreathing were not statistically different from the criterion techniques during rest. Although previous studies support our findings (15, 16, 23), there is disagreement among the current literature, as others have claimed that foreign gas rebreathing is less accurate at rest (44). N2O and C2H2 rebreathing are widely used techniques because they are reliable, less expensive than the criterion techniques, and are easy to employ (44). Reported discomforts are transient and include dizziness from hyperventilation, dyspnea from buildup of CO2 (17), unpleasant taste from the gas mixture (26), and headache. Some have argued that C2H2 rebreathing is superior to N2O because N2O uptake may be influenced more than C2H2 rebreathing by factors such as blood lipid concentration (44). However, we did not find N2O rebreathing to be different from our criterion methods and others have also reported agreement by comparing N2O rebreathing against other gas rebreathing or dye dilution techniques (2, 48, 49). Finally, and most importantly, we did not find any evidence that the implementation of any method introduced systematic differences in these independently developed Qc estimates, suggesting that interlaboratory comparisons of Qc estimations are largely valid.

During exercise, however, we found that Qc as measured by C2H2 and N2O rebreathing was significantly lower than both criterion methods. This is in contrast to Liu et al. (23) who reported agreement between C2H2 and direct Fick from resting conditions up through maximal exercise. The time requirement needed to determine Qc during rebreathing is an unavoidable limitation. It has been suggested that the higher velocity of blood flow during exercise shortens the recirculation time of the rebreathing gas mixture, which may underestimate Qc by as much as 10–20% (6, 7, 40). Thus the available sampling time is inversely related to exercise intensity. Results from previous investigations have recommended that during maximal exercise, rebreathing measurements be made within 8 s to avoid recirculation of the rebreathing mixture (40, 49). We were unable to shorten our rebreathing time from 25 s because we needed to accommodate the requirements of the single-step CO2 method. Additionally, we kept the breathing rate low (22 breaths/min) to minimize the increase in Qc associated with hyperpnea (42). This limited the number of breaths before recirculation could occur and, on average, we were able to obtain three to five data points for regression that contained breaths with adequate gas mixing. Thus the first and most plausible explanation for the underestimation we observed with C2H2 and N2O rebreathing was because of recirculation of the rebreathing mixture. Limiting the analysis to the first few breaths after complete mixing has been achieved may help reduce the underestimation of the foreign gas rebreathing methods during exercise.

Second, we cannot exclude the possibility that the use of pulse oximetry to estimate arterial oxygen content introduced error. These differences should be small but a slight underestimation in oxygen saturation would lead to an overestimation in the criterion Qc for direct Fick. Special care was taken to follow exact procedures to ensure we obtained valid pulse oximetry data. The ear probe line was secured, head motion was kept to a minimum, and a real-time pulse oximetry tracing was monitored for any motion.

Last, it should be mentioned that gas rebreathing techniques estimate pulmonary blood flow, rather than left heart output (21). In healthy individuals without pulmonary shunts the difference between pulmonary blood flow and left heart output at rest is <3% (42).

Single-Step CO2 Rebreathing vs. Criterion Techniques

These data have been presented in part elsewhere (28, 29). Not surprisingly, the current literature includes many relationships between CO2 rebreathing and criterion methods similar to ours. In contrast to our findings, other investigators have reported that CO2 rebreathing underestimates Qc compared with direct Fick and thermodilution (9, 13, 25). We found that the modified single-step CO2 rebreathing method yielded higher results than the criterion techniques, except during submaximal exercise. The most plausible explanation for this overestimation appears to stem from inaccuracies in calculating end-capillary CO2 from end-tidal CO2. A fitted dissociation curve was applied to each value of PACO2 during expiration (or the linear interpolation between expiration and the next alveolar plateau) and averaged to obtain CCCO2i. The assumption that end-tidal CO2 accurately depicts end-capillary CO2 may introduce small errors, even in normal subjects (14).

During submaximal exercise, our data showed no significant differences between the single-step CO2 rebreathing method and thermodilution; during maximal exercise, the single-step CO2 method differed from thermodilution but not direct Fick. Previous studies support this finding and have indicated that during submaximal exercise the relationship between CO2 rebreathing and other methods was better than that observed at rest (35, 45). That conclusion should be interpreted with caution, however, since the single-step CO2 rebreathing method demonstrated the largest variability across all conditions.

Heterogeneity of Variance Between Methods

The intratechnique variance associated with CO2 rebreathing was statistically higher than all methods during standing rest, all but direct Fick during supine rest, and all other gas rebreathing techniques during submaximal and maximal exercise. Direct Fick had the next largest variance, showing significantly larger variances, compared with C2H2 and N2O rebreathing, during supine rest and submaximal exercise. Similarly, we found that during submaximal exercise, thermodilution had higher variance than C2H2 and N2O rebreathing. The standard deviations calculated from our direct Fick and thermodilution data compare favorably with past investigations (7, 13, 27, 41). Thus our data suggest a distinct advantage of foreign gas rebreathing techniques over CO2 rebreathing; since the variance of CO2 rebreathing was approximately double that of foreign gas rebreathing, approximately four times more trials would be needed to develop a statistically reliable estimate of Qc.

In conclusion, we found that rebreathing techniques to determine Qc provide similar results in healthy volunteers; however, there are limitations that should be kept in mind. Soluble gas rebreathing, as implemented, provided accurate data at rest, whether standing or supine. Although there were slight differences in how each of the C2H2 cardiac outputs were computed, these values differed little from one another, leading us to conclude that variations in the way these techniques have been implemented has little impact in determining Qc. This finding may be particularly important when Qc comparisons are made in multi-site or collaborative studies. Single-step CO2 rebreathing was accurate over a wider range (i.e., during exercise), but suffered from lesser precision. Based on these data, we suggest that foreign gas rebreathing can provide reasonable Qc estimates (within 10% of criterion) with relatively fewer repeat trials during resting conditions. During submaximal and maximal exercise these methods remain precise but tend to underestimate Qc. Single-step CO2 rebreathing may be successfully employed over a wider range (i.e., approaching maximal exercise) but with more measurements needed to overcome variability with the method.


    GRANTS
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
Supported by contract #NAS9-18440, NASA (to J. A. Pawelczyk).


    ACKNOWLEDGMENTS
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 
We acknowledge the following people for their outstanding technical assistance: Paul DeFrain, Robyn Etzel, Janelle Fine, Kevin Harper, Susie McMinn, Boyce Moon, Willie Moore, and Julie Zuckerman. We also appreciate the advice and support of Charles Sawin at the NASA Johnson Space Center.


    FOOTNOTES
 

Address for reprint requests and other correspondence: J. A. Pawelczyk, 107 Noll Laboratory, Pennsylvania State Univ., Univ. Park, PA 16802 (e-mail: jap18{at}psu.edu)

The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.


    REFERENCES
 TOP
 ABSTRACT
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 GRANTS
 ACKNOWLEDGMENTS
 REFERENCES
 

  1. Astrand PO, Cuddy TE, Saltin B, Stenberg J. Cardiac output during submaximal and maximal work. J Appl Physiol 19: 268–274, 1964.[Abstract/Free Full Text]
  2. Ayotte B, Seymour J, McIlroy MB. A new method for measurement of cardiac output with nitrous oxide. J Appl Physiol 28: 863–866, 1970.[Free Full Text]
  3. Begin R, Epstein M, Sackner MA, Levinson R, Dougherty R, Duncan D. Effects of water immersion to the neck on pulmonary circulation and tissue volume in man. J Appl Physiol 40: 293–299, 1976.[Abstract/Free Full Text]
  4. Bossone E, Rubenfire M, Bach DS, Ricciardi M, Armstrong WF. Range of tricuspid regurgitation velocity at rest and during exercise in normal adult men: implications for the diagnosis of pulmonary hypertension. J Am Coll Cardiol 33: 1662–1666, 1999.[Abstract/Free Full Text]
  5. Buckey JC Jr, Lane LD, Levine BD, Watenpaugh DE, Wright SJ, Moore WE, Gaffney FA, Blomqvist CG. Orthostatic intolerance after spaceflight. J Appl Physiol 81: 7–18, 1996.[Abstract/Free Full Text]
  6. Cander L, Forster RE. Determination of pulmonary parenchymal tissue volume and pulmonary capillary blood flow in man. J Appl Physiol 14: 541–551, 1959.[Abstract/Free Full Text]
  7. Chapman CB, Taylor HL, Borden C, Ebert RV, Keys A. Simultaneous determinations of the resting arteriovenous oxygen difference by the acetylene and direct Fick methods. J Clin Invest 29: 651–659, 1950.[Web of Science][Medline]
  8. Cigarroa RG, Lange RA, Williams RH, Bedotto JB, Hillis LD. Underestimation of cardiac output by thermodilution in patients with tricuspid regurgitation. Am J Med 86: 417–420, 1989.[CrossRef][Web of Science][Medline]
  9. Clausen JP, Larsen OA, Trap-Jensen J. Cardiac output in middle-aged patients determined with CO2 rebreathing method. J Appl Physiol 28: 337–342, 1970.[Free Full Text]
  10. Defares JG. Determination of PvCO2 from the exponential CO2 rise during rebreathing. J Appl Physiol 13: 159–164, 1958.[Abstract/Free Full Text]
  11. Dhingra VK, Fenwick JC, Walley KR, Chittock DR, Ronco JJ. Lack of agreement between thermodilution and fick cardiac output in critically ill patients. Chest 122: 990–997, 2002.[CrossRef][Web of Science][Medline]
  12. Ekblom B, Astrand PO, Saltin B, Stenberg J, Wallstrom B. Effect of training on circulatory response to exercise. J Appl Physiol 24: 518–528, 1968.[Free Full Text]
  13. Espersen K, Jensen EW, Rosenborg D, Thomsen JK, Eliasen K, Olsen NV, Kanstrup IL. Comparison of cardiac output measurement techniques: thermodilution, Doppler, CO2-rebreathing and the direct Fick method. Acta Anaesthesiol Scand 39: 245–251, 1995.[Web of Science][Medline]
  14. Farhi LE, Nesarajah MS, Olszowka AJ, Metildi LA, Ellis AK. Cardiac output determination by simple one-step rebreathing technique. Respir Physiol 28: 141–159, 1976.[CrossRef][Web of Science][Medline]
  15. Hoeper MM, Maier R, Tongers J, Niedermeyer J, Hohlfeld JM, Hamm M, Fabel H. Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension. Am J Respir Crit Care Med 160: 535–541, 1999.[Abstract/Free Full Text]
  16. Hsia CC, Herazo LF, Ramanathan M, Johnson RL Jr. Cardiac output during exercise measured by acetylene rebreathing, thermodilution, and Fick techniques. J Appl Physiol 78: 1612–1616, 1995.[Abstract/Free Full Text]
  17. Johnson BD, Beck KC, Proctor DN, Miller J, Dietz NM, Joyner MJ. Cardiac output during exercise by the open circuit acetylene washin method: comparison with direct Fick. J Appl Physiol 88: 1650–1658, 2000.[Abstract/Free Full Text]
  18. Jones NL, Campbell EJM, Edwards RHT, Robertson DG. Clinical Exercise Testing. Philadelphia: W. B. Saunders, 1975.
  19. Kagle DM, Alexander CM, Berko RS, Giuffre M, Gross JB. Evaluation of the Ohmeda 3700 Pulse Oximeter: Steady-state and transient response characteristics. Anesthesiology 66: 376–380, 1987.[CrossRef][Web of Science][Medline]
  20. Kim TS, Rahn H, Farhi LE. Estimation of true venous and arterial PCO2 by gas analysis of a single breath. J Appl Physiol 21: 1338–1344, 1966.[Free Full Text]
  21. Laszlo G. Respiratory measurements of cardiac output: from elegant idea to useful test. J Appl Physiol 96: 428–437, 2004.[Abstract/Free Full Text]
  22. Levine BD, Zuckerman JH, Pawelczyk JA. Cardiac atrophy after bed-rest deconditioning: a nonneural mechanism for orthostatic intolerance. Circulation 96: 517–525, 1997.[Abstract/Free Full Text]
  23. Liu Y, Menold E, Dullenkopf A, Reissnecker S, Lormes W, Lehmann M, Steinacker JM. Validation of the acetylene rebreathing method for measurement of cardiac output at rest and during high-intensity exercise. Clin Physiol 17: 171–182, 1997.[CrossRef][Web of Science][Medline]
  24. Mackenzie JD, Haites NE, Rawles JM. Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers. Br Heart J 55: 14–24, 1986.[Abstract/Free Full Text]
  25. Muiesan G, Sorbini CA, Solinas E, Grassi V, Casucci G, Petz E. Comparison of CO2-rebreathing and direct Fick methods for determining cardiac output. J Appl Physiol 24: 424–429, 1968.[Free Full Text]
  26. Nielsen OW, Hansen S, Christensen P, Gronlund J. Repeatability of the acetylene rebreathing method in measuring cardiac output: influence of acetylene concentration. Acta Anaesthesiol Scand 34: 354–357, 1990.[Web of Science][Medline]
  27. Nilsson LB, Eldrup N, Berthelsen PG. Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output. Acta Anaesthesiol Scand 45: 680–685, 2001.[CrossRef][Web of Science][Medline]
  28. Olszowka AJ, Shykoff BE, Pendergast DR, Farhi LE. Revised one-step method for determination of cardiac output. Respir Physiol Neurobiol 140: 99–109, 2004.[CrossRef][Web of Science][Medline]
  29. Olszowka AJ, Shykoff BE, Pendergast DR, Lundgren CE, Farhi LE. Cardiac output: a view from Buffalo. Eur J Appl Physiol 90: 292–304, 2003.[CrossRef][Web of Science][Medline]
  30. Patel C, Laboy V, Venus B, Mathru M, Wier D. Acute complications of pulmonary artery catheter insertion in critically ill patients. Crit Care Med 14: 195–197, 1986.[Web of Science][Medline]
  31. Pawelczyk JA, Zuckerman JH, Blomqvist CG, Levine BD. Regulation of muscle sympathetic nerve activity after bed rest deconditioning. Am J Physiol Heart Circ Physiol 280: H2230–H2239, 2001.[Abstract/Free Full Text]
  32. Peyton PJ, Thompson B. Agreement of an inert gas rebreathing device with thermodilution and the direct oxygen Fick method in measurement of pulmonary blood flow. J Clin Monit Comput 18: 373–378, 2004.[CrossRef][Medline]
  33. Powers SK, Dodd S, Freeman J, Ayers GD, Samson H, McKnight T. Accuracy of pulse oximetry to estimate HbO2 fraction of total Hb during exercise. J Appl Physiol 67: 300–304, 1989.[Abstract/Free Full Text]
  34. Prisk GK, Guy HJ, Elliott AR, Deutschman RA III, West JB. Pulmonary diffusing capacity, capillary blood volume, and cardiac output during sustained microgravity. J Appl Physiol 75: 15–26, 1993.[Abstract/Free Full Text]
  35. Reybrouck T, Amery A, Billiet L, Fagard R, Stijns H. Comparison of cardiac output determined by a carbon dioxide-rebreathing and direct Fick method at rest and during exercise. Clin Sci Mol Med 55: 445–452, 1978.[Web of Science][Medline]
  36. Runciman WB, Ilsley AH, Roberts JG. Thermodilution cardiac output—a systematic error. Anaesth Intensive Care 9: 135–139, 1981.[Web of Science][Medline]
  37. Sackner MA, Greeneltch D, Heiman MS, Epstein S, Atkins N. Diffusing capacity, membrane diffusing capacity, capillary blood volume, pulmonary tissue volume, and cardiac output measured by a rebreathing technique. Am Rev Respir Dis 111: 157–165, 1975.[Web of Science][Medline]
  38. Sackner MA, Markwell G, Atkins N, Birch SJ, Fernandez RJ. Rebreathing techniques for pulmonary capillary blood flow and tissue volume. J Appl Physiol 49: 910–915, 1980.[Abstract/Free Full Text]
  39. Shykoff BE, Farhi LE, Olszowka AJ, Pendergast DR, Rokitka MA, Eisenhardt CG, Morin RA. Cardiovascular response to submaximal exercise in sustained microgravity. J Appl Physiol 81: 26–32, 1996.[Abstract/Free Full Text]
  40. Simmons R, Shephard RJ. Measurements of cardiac output in maximum exercise. Application of an acetylene rebreathing method to arm and leg exercise. Int Z Angew Physiol 29: 159–172, 1971.[Web of Science][Medline]
  41. Sun XG, Hansen JE, Ting H, Chuang ML, Stringer WW, Adame D, Wasserman K. Comparison of exercise cardiac output by the Fick principle using oxygen and carbon dioxide. Chest 118: 631–640, 2000.[CrossRef][Web of Science][Medline]
  42. Triebwasser JH, Johnson RL, Burpo RP, Campbell JC, Reardon WC, Blomqvist CG. Noninvasive determination of cardiac output by a modified acetylene rebreathing procedure utilizing mass spectrometer measurements. Aviat Space Environ Med 48: 203–209, 1977.[Medline]
  43. Van GA, Ditchey RV, Groves BM, Wagner WW Jr, Reeves JT. Thermodilution method overestimates low cardiac output in humans. Am J Physiol Heart Circ Physiol 245: H690–H692, 1983.[Abstract/Free Full Text]
  44. Warburton DE, Haykowsky MJ, Quinney HA, Humen DP, Teo KK. Reliability and validity of measures of cardiac output during incremental to maximal aerobic exercise. Part I: Conventional techniques. Sports Med 27: 23–41, 1999.[CrossRef][Web of Science][Medline]
  45. Wigle RD, Hamilton PD, Parker JO. Measurement of cardiac output by carbon dioxide rebreathing. Can J Appl Sport Sci 4: 135–139, 1979.[Medline]
  46. Yamaya Y, Bogaard HJ, Wagner PD, Niizeki K, Hopkins SR. Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia. J Appl Physiol 92: 162–168, 2002.[Abstract/Free Full Text]
  47. Zar JH. Multiple comparisons. In: Biostatistical Analysis, edited by Ryu T. Upper Saddle River: Prentice-Hall, 1999, p. 208–230.
  48. ZeidiFard E, Davies CT. An assessment of a N2O rebreathing method for the estimation of cardiac output during severe exercise. Ergonomics 21: 567–572, 1978.[Medline]
  49. ZeidiFard E, Godfrey S, Davies EE. Estimation of cardiac output by an N2O rebreathing method in adults and children. J Appl Physiol 41: 433–438, 1976.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF) Free
Right arrow All Versions of this Article:
103/3/867    most recent
01106.2006v1
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jarvis, S. S.
Right arrow Articles by Pawelczyk, J. A.