INNOVATIVE METHODOLOGY
1Department of Physiology, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands; and 2Department of Anesthesiology and Critical Care, University Hospital Tuebingen, Tuebingen, Germany
Submitted 14 October 2005 ; accepted in final form 14 December 2005
| ABSTRACT |
|---|
|
|
|---|
phosphorescence quenching; oxygen measurement; renal microvascular oxygenation; oxygen distribution
Pd-meso-tetra-(4-carboxyphenyl)-tetrabenzoporphyrin(Oxyphor G2) is a relatively new phosphor that is excellently suited for oxygen measurements in vivo (5). It is highly soluble in blood plasma, where it binds to albumin and confines to the circulation, as demonstrated recently for Oxyphor R2 in rat skeletal muscle (18). Oxyphor G2 has its absorptions maxima at 440 and 632 nm and its emission near 800 nm. The excitation wavelength of 632 nm combined with emission in the infrared spectrum allows much deeper tissue penetration than could be achieved with the most widely used phosphor to date, Pd-porphyrin [Pd-meso-tetra(4-carboxyphenyl)porphine]. A very recent example of the usability of Oxyphor G2 has been the measurement of the distribution of microvascular oxygen pressures in solid tumors by Wilson and coworkers (37).
Although the measurement of the distribution of microvascular oxygen pressures is a powerful means to gain insight into the heterogeneity in oxygen levels within a tissue or organ, we recently argued that the interpretation can be quite difficult (16). Especially in tissues with different anatomical layers, one can never be sure whether the found distribution reflects heterogeneity in microvascular PO2 (µPO2) within a layer or a difference in oxygenation between the two anatomical layers. Additional information could then be retrieved from measurements from different depths. A convenient way to do this is by using different excitation wavelengths (35). Other than the deeper penetration depth, another advantage of Oxyphor G2 is the huge spectral difference in excitation wavelengths. Considering the fact that 440-nm light in contrast to 632-nm light is highly absorbed by tissue, this would allow excellent discrimination between superficial and deeper tissue layers. Although the idea of using different wavelengths for obtaining information from different tissue depths in itself is not new (21, 35), it has not been applied for continuous and simultaneous monitoring of µPO2 at different tissue depths. If feasible, it could provide a very useful extension to the already powerful method of oxygen-dependent quenching of phosphorescence.
To profit from the above-mentioned possibilities, we designed and constructed a dual-wavelength phosphorimeter. The hypothesis was that application of this phosphorimeter in combination with Oxyphor G2 would allow the measurement of µPO2 in two different depths. This study was designed to validate and show the feasibility of dual-wavelength phosphorimetry to examine changes in oxygenation in two different anatomical layers: the kidney cortex and medulla. Therefore, four steps were undertaken. First, we developed a dual-wavelength phosphorimeter that allows exciting Oxyphor G2 at 440 and 632 nm near simultaneously. Second, in vitro calibration experiments were done to exclude hardware-induced interchannel differences. Third, the real penetration depth of both excitation channels was determined by ex vivo experiments in rat kidneys. Fourth, for evaluation of the accuracy of the oxygen measurements in vivo, experiments in rats were carried out in which the fractional inspired oxygen concentration (FIO2) was stepwise reduced. Finally, to demonstrate the usability of this method to detect changes in µPO2 in a pathophysiological relevant setting, we report on cortical and medullar PO2 values and oxygen distributions in rats during endotoxemia and resuscitation.
| MATERIALS AND METHODS |
|---|
|
|
|---|
800 nm) intensity decreases at a rate dependent on the surrounding oxygen pressure. The relationship between the measured decay time and the PO2 is given by the Stern-Volmer relation:
![]() | (1) |
is the measured decay time,
0 is the decay time at an oxygen pressure of zero, and kq is the quenching constant.
Equation 1 describes the relationship between the PO2 and the phosphorescence lifetime in case of a homogeneous oxygen distribution. However, large nonuniformities in oxygen pressure exist in vivo, resulting in a phosphorescence signal that in general can be described by an integral over an exponential kernel (6, 32, 33).
![]() | (2) |
) denotes the spectrum of reciprocal lifetimes that should be determined from the finite data set y(t). Several different approaches for retrieving information about the spectrum f(
) in the case of phosphorescence lifetime measurements have been described (6, 32, 33). In this paper, we use the approach published by Golub et al. (6) in which the heterogeneity in oxygen pressure is analyzed by fitting distributions of quencher concentration to the phosphorescence data. According to their work, the fitting function for a simple rectangular distribution with a mean PO2 Qm and a PO2 range from Qm
to Qm +
is
![]() | (3) |
is half the width of the rectangular distribution from which the standard deviation (
) can be calculated from the relationship
![]() | (4) |
resulting in the following fit equation:
![]() | (5) |
(wi
0 and
wi = 1).
The currently developed phosphors for oxygen measurements possess different excitation bands with a wide spread in excitation wavelengths. For example Oxyphor G2 has excitation bands of
440 and 632 nm (5). Combined with tissue optical properties, this feature can be used for achieving measurements with different penetration depths. Described in the following paragraph is a phosphorimeter that we developed to perform this type of measurements in a near-simultaneous fashion.
Description of the phosphorimeter.
The dual-wavelength phosphorimeter had to be able to measure phosphorescence lifetimes with near-simultaneous excitation at two different wavelengths in an automated fashion. There should especially be no need for the operator to physically have to change fiber-optic connections or filters during the experiment. The device and its construction should be simple and easily reproducible. Figure 1 shows schematically the constructed device. The central part of the phosphorimeter is a commercially available randomized trifurcated fiber optic (model no. 77536, Oriel, Stratford, CT). This fiber-optic light guide has a common bundle diameter of 5.5 mm. The fiber-optic light guide has a numerical aperture of 0.56 and an acceptance cone of 68°. The resulting light bundle is slightly diverging and, therefore, does not have spatial discriminating power in itself. Two branches of the fiber-optic light guide are used for the two excitation pathways. Excitation light is provided by two identical flash lamp modules (type C5605, Hamamatsu, Hamamatsu City, Japan), providing pulsed light with a light pulse duration of <5 µs. The output power at the end of the fiber-optic light guide does not exceed 1 µJ per pulse for both wavelengths, as measured with a power meter (Ophir model DGX, Ophir Optics, Jerusalem, Israel). Excitation light is filtered by two band-pass filters (Edmund Optics, York, UK) with central wavelengths of 440 and 632 nm for the respective excitation path. The remaining third branch of the fiber-optic light guide is used for returning the emitted phosphorescence back to the device. The signal is filtered by a 800-nm band-pass filter (Edmund Optics, York, UK). All band-pass filters have a full width at half maximum of 10 nm. The detector is an infrared-sensitive photomultiplier tube fitted in an ungated socket (type PS1250/12F, Thorn EMI Electron Tubes, Ruislip, UK). The current from the photomultiplier tube is voltage converted by a home-built, wide-band (30 MHz) transresistance amplifier. The whole device is under control of a computer containing a 1-MHz multifunction data-acquisition board (DAQ-board, PCI-MIO-16e1, National Instruments, Austin, TX). The data-acquisition board controls the triggers to the flashlamps and samples the phosphorescence signal with a sampling frequency of 1 MHz. The trigger pulse that triggers the flashlamp is internally directed to the data-acquisition clock (the used DAQ-board allows for this), meaning that data acquisition is started at the rising flank of this trigger pulse. A short delay of 3 µs exists before the flashlamp is actually fired. The resulting three data points without signal are omitted from the analysis. The used flashlamp units have a very stable time delay and moreover show only a time jitter of <300 ns between the triggering and the flash occurrence. For our purposes, we regard this as insignificant, and therefore we allowed ourselves to omit the optical trigger feedback system used in our laboratory's previously described phosphorimeter (15). To improve signal-to-noise ratio, the average signal of 50 flashes is used for analysis. The flashlamps fire with a frequency of 50 Hz, so the duration of one measurement on one wavelength is 1 s. The acquisition of the 632-nm excitation is started immediately after the acquisition of the 440-nm excitation has finished, so no more than a 1-s time interval exists between the two measurements. Before phosphorescence lifetime analysis, the adverse effect of the excitation pulse shape was removed using excitation pulse deconvolution, of which the details can be found in a previous study (15). For the excitation pulse deconvolution, both the blue and red backscattered flashlamp signals were recorded from the kidney before infusion of Oxyphor G2 into the animal. This results in a correction for a difference in flashlamp profile at 440 and 632 nm and also corrects for aging of the flashlamp. The phosphorescence lifetime analysis was performed by nonlinear monoexponential fitting (Eq. 1), nonlinear fitting of the rectangular distribution (Eq. 3), and the PO2 histogram recovery method from Eq. 5. Analysis software for the monoexponential and rectangular distribution fit procedures was written in LabView (version 6.1, National Instruments, Austin, TX). The PO2 histogram recovery was performed with GraphPad Prism (version 4, GraphPad Software, San Diego, CA) using a nonlinear fit procedure with constrained parameters (wi
0).
|
Ex vivo penetration depths experiments. To demonstrate that exciting Oxyphor G2 at 440 and 632 nm results in PO2 reading from different tissue layers in the rat kidney and can ultimately be used to gain information from the cortex and medulla, respectively, experiments were performed ex vivo. The idea was to determine the actual capability of exciting the phosphor at a certain depth for both wavelengths by placing slices of kidney tissue between the fiber-optic light guide and an Oxyphor G2-containing chamber. To this end, the vessels close to the hilus of the rat kidney were ligated, and the kidney was removed directly after the animal died. Either 500- or 1,000-µm longitudinal kidney slices were made by using a razor blade-like cutting device. For measurements of the penetration depth of the two different wavelengths (440 and 632 nm), a slice of kidney tissue was placed on top of a chamber filled with Oxyphor G2 (50 µM concentration in a 4% albumin solution). Zero-oxygen conditions within the chamber were created by an enzymatic oxygen removal by using ascorbate-oxidase and ascorbic acid (13). The experiments were performed at room temperature. The kidney slices were placed on top of each other following the anatomy from cortex toward the medulla. The actual thickness of the total reconstructed tissue was always measured using a micrometer. Phosphorescence returning through the kidney tissue after excitation with either 440 or 632 nm was measured for each additional piled-up slice to a total thickness of 5 mm.
To take into account the fact that the kidney slices contained no phosphor that would clearly contribute to the overall absorption, the extinction coefficient of Oxyphor G2 at 440, 632, and 800 nm was measured using a Specord UV VIS S10 absorption meter (Carl Zeiss, Oberkochen, Germany). The measured extinction coefficients were used to calculate estimations of additional attenuation of the overall signal due to the presence of the phosphor using Beer's law:
![]() | (6) |
Animal preparation for in vivo experiments. The experiments of this study were approved by the ethical committee for animal subjects of the Academic Medical Center at the University of Amsterdam. Care and handling of the animals were in accordance with the guidelines for Institutional and Animal Care and Use Committees. The measurements were performed in seven Wistar male rats (Charles River) with a body weight of 357 ± 35 g (mean ± SD).
Rats were anesthetized, mechanically ventilated, and monitored as described recently by our group (20). In short, induction of anesthesia was performed by intraperitoneal injection of a mixture of ketamine (90 mg/kg, Nimatek, Eurovet, Bladel, The Netherlands), medetomidine (0.5 mg/kg, Domitor, Pfizer, Capelle a/d Ijssel, The Netherlands), and atropine (0.05 mg/kg, Centrafarm, Ettenleur, The Netherlands), and anesthesia was maintained by continuous intravenous infusion of ketamine (50 mg·kg1·h1). To compensate for fluid loss, Ringer lactate (Freeflex, Fresenius Kabi, The Netherlands) was administered continuously at a rate of 15 ml·kg1·h1. Kidney preparation was performed via a 3-cm incision of the left flank. The kidney was exposed, decapsulated, and immobilized in a Lucite kidney cup (K. Effenberger, Geraetebau, Pfaffing, Germany), a perivascular flow probe (type 0.7 mm V-Series, Transonic Systems, Ithaca, NY) was placed around the left renal artery and connected to a flowmeter (T206; Transonic Systems). Throughout the entire experiment, the operation field was covered with Saran wrap (Downbrands) to prevent evaporation of body fluids. The temperature of the kidney surface was measured and kept at
37°C.
After surgery, the fiber-optic light guide for phosphorescence lifetime measurements was placed 1 mm above decapsulated kidney surface, and a 15-min intravenous infusion of Oxyphor G2 (1.2 ml/kg of a 5 mg/ml stock solution) was started. After a stabilization time of 50 min, the renal µPO2 was continuously measured during the entire experiment.
In vivo feasibility testing. For testing the reproducibility of renal oxygen measurements over a wide range of different PO2 in five animals, the FIO2 was reduced in three steps from 1.0 to 0.1. Changes in FIO2 have previously been shown to be an effective means to change µPO2 values (19). Each step comprised an equilibration time of 10 min.
In vivo experiment during lipopolysaccharide bolus injection. It is well known that lipopolysaccharide (LPS) infusion leads to marked reduction in blood flow to the kidney and possibly to redistribution of flow from cortex to medulla (8, 9). To test the relevance of simultaneous measurements in two different depths, we studied the response of cortical and medullary PO2 in a rat receiving an intravenous bolus injection of LPS (20 mg/kg; serotype 0127:B8, Sigma) to induce septic shock.
In vivo experiment during fluid resuscitation. For demonstration of the time course of the oxygen measurements under clinically relevant conditions, one rat received fluid resuscitation with 200-kDa hydroxyethyl starch (HAES-steril 6%, 20 ml/h, Fresenius Kabi, The Netherlands) directly after a 1-h intravenous LPS infusion was stopped (15 mg·kg1·h1; serotype 0127:B8, Sigma).
Data analysis. Labview 6.1 (National Instruments, Austin, TX) was used to develop a software environment to allow data acquisition and analysis of the phosphorescence decay curves. Oxygen histograms were analyzed using the GraphPad Prism package (version 4, GraphPad Software, San Diego, CA). Values are presented as means ± SD, unless otherwise indicated.
| RESULTS |
|---|
|
|
|---|
0 = 240 µs, which are in good agreement with previous reported values (5).
|
|
30 µM in full blood. Making the conservative assumption that 25% of the total kidney volume is comprised of blood, the overall concentration of Oxyphor G2 in the kidney is 7.5 µM. Because light in tissue is multiple scattered, the actual traveled path through the tissue at a certain depth is much longer than the depth itself (4, 34). Because for the two excitation wavelengths the optical properties of kidney tissue are not exactly known, we generated results using Eq. 6 with different values for a, creating path lengths ranging from 3 to 10 times the actual depth (Fig. 3B). The extinction coefficient for 800-nm light is sufficiently low to be ignored for our purposes. From Fig. 3B, it can be seen that for 440 nm at a depth of 0.5 mm and the assumption of a five times longer pathlength, the attenuation due to the phosphor is <30%. Even if we assume a 10 times longer pathlength, still 60% of the excitation light will reach a depth of 0.5 mm. And, for example, for 632 nm at 3 mm depth, the additional attenuation due to the phosphor is <50% if we assume the actual pathlength to be five times the penetration depth. These values show that, although there is obviously a nonnegligible contribution of the phosphor to the overall attenuation, the general outcome from the experiment in Fig. 3A is still valid. Figure 3, inset, shows a 500-µm slice of tissue with a millimeter scale to be able to link the measured penetration depth to an anatomical compartment. The kidney cortex was measured to be 1,200 µm in total thickness. The outer medulla started at a depth of 1,200 µm. This information allows concluding that the signal after excitation with 440-nm light is solely confined to the cortical compartment and that excitation with 632-nm light indeed gains information from the (outer) medullary compartment.
After the in vitro and ex vivo experiments, the dual-wavelength device was tested in vivo. Figure 4 shows a cartoon of the principle behind the measurement and examples of real phosphorescence traces measured in an anesthetized rat. Clearly, a difference in signal can be seen in lifetime as well as in overall phosphorescence intensity. The solid line is the resulting monoexponential fit. The lifetimes in this example are
= 54 and 69 µs for the blue and red channels, respectively. The corresponding PO2 readings are 53 and 38 Torr, indicating that indeed two different compartments were measured. Moreover, although the efficiency of excitation with 632-nm light is much lower than with 440-nm light (5), the phosphorescence yield in vivo appears to be much higher. Although this can partly be explained by the fact that a lower PO2 results in more phosphorescence yield, it is more likely that it indicates a much larger excitation volume in the case of the red channel. The fact that both channels are measuring different physical volumes is also underscored by the observation that, under certain experimental circumstances, the cortical µPO2 reading drops below the medullary µPO2. For example, application of the mitochondrial uncoupler carbonyl cyanide trifluoromethoxyphenylhydrazone (a single drop of a 10 mM stock solution) to the kidney surface led to a rapid decline in cortical µPO2 (higher oxygen consumption) and a much slower decline in medullary µPO2. Over 25 min, the cortical µPO2 dropped from 60 to 15 Torr, whereas the medullary µPO2 dropped from 45 to 30 Torr.
|
To test the dual-wavelength technique in vivo and to evaluate the different fitting models, the device was used in a simple experimental model in which the oxygenation of the rat kidney tissue was changed by reducing FIO2 stepwise from 1.0 to 0.1. The results of the monoexponential and rectangular distribution fit procedures are given in Fig. 5. Figure 5A shows a bar diagram of the mean cortical and mean medullar PO2 resulting from the rectangular distribution fit. During ventilation with 100% oxygen, the mean µPO2 of the cortex was 97 ± 10 Torr and the mean PO2 of the medulla was 68 ± 6 Torr. During ventilation with an FIO2 of 0.4, the average cortical oxygenation was 88 ± 7 Torr and the average medullary PO2 was 59 ± 4 Torr. Ventilation with room air (21% oxygen) resulted in a mean PO2 of 61 ± 9 Torr in the cortex and a mean PO2 of 39 ± 5 Torr in the medulla. During hypoxia (FIO2 = 0.1), the cortical oxygenation decreased to a mean PO2 of 33 ± 7 Torr and the medullary mean PO2 was 15 ± 2 Torr. The experiments were performed in n = 5 rats. Oxygen tensions of the rat cortex and medulla during stepwise reduction in FIO2 compare well with reported PO2 values in previous studies using invasive Clark-type oxygen electrodes (1, 14, 22).
|
). It is very interesting to see how the mean PO2 values and estimated standard deviations determined by the rectangular distribution fit correspond to the PO2 histograms recovered using Eq. 5. Figure 6 shows the PO2 histograms of FIO2 variations in one rat. The width of the bins was chosen to be 15 Torr so that 10 bins covered the PO2 region from 0 to 150 Torr. From each stable period after an FIO2 step, seven histograms were calculated. Each presented histogram contains the means and standard deviations of these seven histograms. The signal-to-noise ratio of both channels was well above 20, and the histogram recovery was remarkably reproducible, corresponding very well to the results obtained in computer simulations (6). From direct comparison with Fig. 5B, it is clear that the rectangular distribution fit gives adequate mean PO2 values and that the values for
give a good indication for the underlying heterogeneity. The monoexponential fit indeed has the tendency to report too-low values for the average PO2 and moreover gives no indication of heterogeneity at all. The histogram recovery, however, gives most detailed information. For example, the rectangular distribution fit gives a mean cortical PO2 of 33 Torr when the FIO2 is 0.1, whereas from the PO2 histogram it is clear that a large part of the measurement volume has a PO2 below 15 Torr.
|
|
| DISCUSSION |
|---|
|
|
|---|
To detect reliable heterogeneity in PO2 levels in the depth, it is obvious that in a homogeneous in vitro situation no differences in PO2 readings over the physiological relevant PO2 range should exist. It is already known that different excitation wavelengths per se do not cause differences in phosphorescence lifetime readings in Oxyphor G2 (5). Also, the excitation pulse deconvolution algorithm (15) should cancel out differences in measured lifetimes caused by variation in flashlamp decays. Our calibration experiment shows indeed that over a wide oxygen range no differences in PO2 readings in both channels exist. For the in vivo situation with heterogeneous µPO2 values, the rectangular distribution fit is clearly preferable over the monoexponential fit procedure because it lacks the tendency to bias toward low PO2 values and provides an indication of the level of heterogeneity. Furthermore, the PO2 histogram recovery method suggested by Golub et al. (6) appeared to be easy to implement and very robust with the signal-to-noise ratios achieved with our phosphorimeter. An alternative method would have been the maximum entropy method (33). The latter method allows the recovery of even more detailed PO2 histograms, but this comes at the expense of longer computation times (especially in the time domain), and it requires fairly high signal-to-noise ratios (33). The PO2 histogram recovery method reveals details that otherwise remain unseen. For example, in Fig. 6, some small contributions of high PO2 regions can be seen even at an FIO2 of 0.1. Because some PO2 values are higher than the possible arterial PO2, these values probably are the result of oxygen diffusion from air into the outer surface of the kidney or caused by small contamination of the kidney surface with Oxyphor G2-containing blood.
In a previous study (25), our laboratory calculated, using Monte-Carlo simulations, that green (520 nm) light results in a catchments depth of
0.5 mm in gut tissue. For near-infrared Oxyphor G2, a general penetration depth of several centimeters has been claimed (31). In the present study, we made an attempt to determine the catchments depth of the 440 and 632 nm in a direct manner since no sound estimations of the needed tissue optical properties of kidney (i.e., absorption coefficient, scatter coefficient, and anisotropy factor) at the specific wavelengths could be found, preventing Monte-Carlo simulations. We show that 440 nm is limited to a penetration depth of 700 µm and is therefore confined to the kidney cortex. In contrast, the 632-nm light easily reaches 3-mm depth and even excites phosphorescence at a depth of 4 mm, and therefore catchments volume includes the outer medullary region of the rat kidney. A drawback of our kidney-slice experiment is that the kidney slices contained no phosphor that would contribute to a higher overall absorbance and thus lower penetration depth. However, even very conservative assumptions about phosphor concentration and actual pathlength calculations show that the results of the kidney-slice experiment are a very reasonable estimation. This finding is confirmed by the fact that in vivo the red channel reports substantially lower PO2 values than the blue channel. The latter observation is known for the kidney as due to its anatomy; within the kidney, there is a diffusion of oxygen from arterial to venous vasa recta, thus leading to a relatively low PO2 within the renal medulla (1). In the superficial kidney cortex, PO2 ranges from 50 to 70 Torr (2), whereas the medullary PO2 from 10 to 20 Torr (3). These reported values, measured by Clark-type oxygen electrodes in rats (where penetration of the kidney is needed), are similar to our findings with the two-wavelength technique. Deviations (especially in the medullary PO2) may be explained by differences in FIO2, deeper penetration with oxygen electrodes, or even tissue destruction by oxygen electrodes.
Advantages and applications of this technique. The presented method has as its main advantage that it allows continuous and nondisruption measurements of the cortical and outer medullary microvascular oxygenation in the rat kidney. In addition to measurement of lifetime distributions (35, 37), it gives information about the in-depth heterogeneity of µPO2 levels, although not to such a detailed extent as two-photon excitation (16). The given examples in this paper demonstrate that the combination of dual-wavelength phosphorimetry and lifetime deconvolution (retrieval of PO2 histograms) provides a very powerful tool for in vivo oxygen measurements. Compared with oxygen electrodes (1, 14, 22), the obvious advantage is in its nondisruptiveness, which allows monitoring changes in time. In this study, the method allowed us to monitor different behavior in oxygenation in cortex and medulla. Oxygen measurements in the kidney in superficial and deep tissue using a method based on the same principal of using different excitation wavelengths have been reported by Rumsey et al. (21). For imaging of oxygen distributions, they had to inject two different oxygen probes in different rats. Although their PO2 readings were similar to our findings, our technique has the clear advantage of continuous and simultaneous measurements in two different depths. Furthermore, ease of handling and cost effectiveness are advantages of our method over the application of blood oxygenation level-dependent magnetic resonance image for the noninvasive detection of changes in intrarenal oxygenation in vivo, reviewed recently by Swartz and Dunn (27). The same counts for electron paramagnetic resonance with magnetic resonance image (11) where an additional advantage of our technique is that no injection of relatively large particles is needed (7).
Regarding the above-mentioned advantages, our method can be applied for PO2 readings not only in the kidney but also on different organs like the gut, heart, or skin. This is of special relevance as, for example, in the gut it has been shown that the mucosa responds differently to shock and resuscitation than the serosa (23). The calculation of PO2 histograms adds a certain complexity to the results, and straightforward interpretation of the distribution in terms of anatomical or microcirculatory compartments is not possible. However, the example of resuscitation after a period of endotoxemia clearly shows the additional value compared with only an average µPO2 reading. The increase in medullary µPO2 following resuscitation appears to be the result of only disappearance of very hypoxic regions instead of a right shift of the total histogram, probably indicating recruitment. In contrast, the drop in cortical µPO2 seems to be the result of the appearance of a hypoxic fraction that is actually induced by the treatment. In research concerning optimization of treatment strategies in septic shock, this could prove to be valuable information, since microcirculatory dysfunction is considered to be a major contributor to the pathophysiology of sepsis (10, 25).
Methodological limitations. The use of the near-infrared phosphor Oxyphor G2 allows a deeper penetration into the tissue than the conventional Pd-porphyrin dye. In this study, we could measure oxygenation at a depth of 4 mm in the outer medulla of the rat kidney. Although this is an improvement, this is much less than the originally acclaimed penetration depth of centimeters (31). Apparently in kidney with increasing depth of the measurement, a rapid decline of excitation and emission intensity occurs. For our current application, this means that, although readings of the outer medullary µPO2 can be performed, the inner medulla cannot be reached. Furthermore, in the recording of the deep signal, a substantial part of the total signal is superficial in origin. Nevertheless, our measured values for cortical and medullary PO2 compare well with previously reported values by other groups using oxygen electrodes (14) and are very comparable to the values we found using two-photon excitation phosphorescence lifetime measurements (16). Moreover, we demonstrate differences in the behavior of cortical and medullary PO2, indicating that the deeper signal is not overwhelmed by the superficial signal. This can also be seen from Fig. 3, where a substantially higher amount of phosphorescence is measured in the red channel, although excitation at 632 nm has lower efficiency than excitation at 440 nm. This also indicates a much larger catchment volume in the case of red excitation. However, since the contribution of the cortex to the medullary signal could be a potential source of error, we will investigate possibilities to improve the presented method either by changing the fiber geometry or by mathematical deconvolution algorithms.
One other general point of concern regarding phosphorescence lifetime measurements in vivo is the issue of phototoxicity. The collision of oxygen with an excited phosphor molecule results in the production of oxygen radials that can potentially harm the tissue under investigation (26). Therefore, the light dose and phosphor concentration used for oxygen measurements should be kept as low as possible. Although we cannot exclude phototoxicity in our application, no macroscopic alterations to the kidney surface are observable after hours of measurement. Moreover, in long-lasting time control experiments, no changes in µPO2 values are observed. Furthermore, it is very unlikely that the excitation light itself influenced the physiological function of the tissue because of the relatively low excitation energy used (<1 µJ per flash). The power density is <0.06 µJ/mm2, and this is in orders of magnitude less than the power density used in, for example, two-photon approaches (16).
Our results demonstrate that oxygen-dependent quenching of phosphorescence of Oxyphor G2 is excellently suited for quantitative µPO2 measurements in two different depths in vivo. Oxygen levels measured in the renal cortex and outer medullary region are consistent with previous reported values detected by oxygen electrodes in the rat kidney. This technique provides a new tool for quantitative measurements of oxygen levels under physiological and pathophysiological conditions with the characteristics of providing spatial information and a penetration depth of several millimeters.
| GRANTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
* T. Johannes and E. G. Mik contributed equally to this work. ![]()
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. G. Mik, T. Johannes, C. J. Zuurbier, A. Heinen, J. H. P. M. Houben-Weerts, G. M. Balestra, J. Stap, J. F. Beek, and C. Ince In Vivo Mitochondrial Oxygen Tension Measured by a Delayed Fluorescence Lifetime Technique Biophys. J., October 15, 2008; 95(8): 3977 - 3990. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Mik, T. Johannes, and C. Ince Monitoring of renal venous PO2 and kidney oxygen consumption in rats by a near-infrared phosphorescence lifetime technique Am J Physiol Renal Physiol, March 1, 2008; 294(3): F676 - F681. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Johannes, E. G. Mik, B. Nohe, K. E. Unertl, and C. Ince Acute decrease in renal microvascular PO2 during acute normovolemic hemodilution Am J Physiol Renal Physiol, February 1, 2007; 292(2): F796 - F803. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Kemp Detecting acute changes in oxygen: will the real sensor please stand up? Exp Physiol, September 1, 2006; 91(5): 829 - 834. [Abstract] [Full Text] [PDF] |
||||
| ||||||||