Clinical Pharmacology Unit, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom
Submitted 3 June 2005 ; accepted in final form 30 August 2005
| ABSTRACT |
|---|
|
|
|---|
preejection period; pulse-wave transit time
An alternative technique involves measuring the transit time of the pulse pressure wave through the arterial tree. Measurement of pulse transit time involves detecting the pulse arrival at two separate arterial sites. This can be achieved easily distally using infrared photoplethysmography, and differential pulse transit time measured between finger and toe, as detected by photoplethysmography, has indeed been shown to satisfactorily reflect changes in pulse-wave velocity measured by Doppler ultrasound (23). The ECG R-wave is often used as a proximal timing point because it is simple to detect and tolerant of motion artifact. However, importantly, there is a considerable delay between the onset of electrical cardiac activity and the start of mechanical ventricular ejection (22). This delay is comprised of both the electromechanical delay and the period of isovolumic contraction, and is referred to as the preejection period (PEP).
It has been suggested that, because a near-linear correlation exists between transit time measured from the R-wave (rPTT) and BP, rPTT might be used as a surrogate marker of pressure (9). The use of rPTT in this way was originally described in the 1950s (34), and considerable research was subsequently performed on its application in the study of cardiovascular feedback, mainly in the field of psychophysiology (9, 19, 33, 37). There have been several studies that revisited the technique in recent years, probably due to the increasing ease with which signal analysis can be carried out using modern technology (1, 7, 14, 26). However, little work has been published on the effects of vasoactive drugs on this measurement in humans. In particular, studies have not been carried out quantifying PEP or comparing rPTT with invasive BP measurement.
This study used various vasoactive drugs to produce differing cardiac and vascular responses. The rationale was to compare changes in transit time measurements with the clinical "gold standard" for BP measurement over a wide BP range and under different conditions of vascular tone and cardiac contractility. We hypothesized that the relationship between intra-arterial BP and rPTT would vary following the administration of different vasoactive drugs due to differing effects on the vascular and cardiac components of rPTT.
| METHODS |
|---|
|
|
|---|
Studies were performed in a quiet, temperature-controlled (22 ± 2°C) environment after at least 1 h of acclimatization. Subjects were allowed a light breakfast not less than 4 h before attending and were requested to refrain from alcohol, caffeine, nicotine, or medications for the preceding 24 h. Studies were conducted with the subject lying supine. Continuous beat-to-beat measurements of BP, transit time, and PEP were made throughout the entire study protocol.
All drugs were administered via a 20-gauge intravenous cannula sited in the antecubital fossa of the dominant arm. The infusion rate was kept constant at 1 ml/min. After an initial 20-min 0.9% saline run-in period, four drugs were given, each for 15 min, with the dose increased every 5 min. A 25-min washout period followed each drug. Drugs given were glyceryl trinitrate (0.1, 1, 4 µg·kg1·min1), angiotensin II (2, 6, 12 ng·kg1·min1), norepinephrine (20, 60, 120 ng·kg1·min1), and salbutamol (albuterol, 0.4, 1.2, 2.4 µg·kg1·min1). Drug order was not randomized. Salbutamol was given last due to its relatively long half-life. Dose ranges and washout periods were based on previous studies and selected for their anticipated effects on BP and heart rate (HR) (10, 12, 15, 30, 38, 39). Glyceryl trinitrate was selected for anticipated decreases in BP (12, 15). Norepinephrine and angiotensin II increase mean arterial pressure (MAP) to a similar degree, but the former also increases peripheral pulse pressure (30, 38). The expected response to salbutamol was a fall in diastolic pressure (DBP) and a rise in systolic pressure (SBP) and HR (10, 39).
Intra-arterial pressure monitoring was used for all BP measurements. A 20-gauge 80-mm Vygon catheter was inserted under local anesthesia (1% lidocaine) into the nondominant radial artery using the Seldinger technique. A splint was used to minimize wrist movement. The cannula was connected by fluid-filled semi-rigid tubing to a TruWave disposable transducer (Edwards Life Sciences) positioned level with the right atrium and connected to a Diascope 2 monitor (S & W Medical). Transducers were factory calibrated and exceeded AAMI standards for performance interchangeability, with a sensitivity of 5 µV·V1·mmHg1 ±1% and nonlinearity of the greater of ±1.5% or ±1 mmHg. The natural frequency of the system was 40 Hz. SBP and DBP were taken as the maximum and minimum values of the waveform corresponding to the last measured R-wave, with mean pressure calculated as the average over the last pulse cycle. Custom hardware was used for determination of the transit time. The ECG was detected using a standard three-lead configuration (Lead II), with the signal sampled at 1 kHz. The pulse volume wave was detected on the dominant index finger using infrared transmission photoplethysmography, digitized at 200 Hz, with linear interpolation to 1-kHz accuracy. The time delay was calculated between R-wave peak and the base of the leading edge of the finger pulse wave. The pulse wave base was identified as the intersection of the tangent through the steepest part of the slope with the absolute minimum value of the pulse wave (8). PEP was determined from the B point of the first derivative of the transthoracic cardiac bioimpedance waveform, using an NCCOM3 Cardiodynamic monitor (BoMed Medical Systems). Bioimpedance has previously been validated for determination of systolic time intervals (17), and the NCCOM3 device has itself been compared favorably with both echocardiography (16) and mechanophonocardiography (35). Pilot work established that, over a 1-min resting period, the standard deviation of PEP measurements was 5.2 ms (mean 69 ms), with a mean beat-to-beat difference of 3.5 ms. The coefficient of variation in baseline PEP measurements was 8.9%. We consider these small variations in beat-to-beat PEP measurements obtained by our experimental technique to reflect satisfactory intrasubject reliability.
All data are expressed as means ± standard deviation (SD) unless otherwise stated. Changes from baseline, taken as the 2 min immediately before each infusion period, were assessed by ANOVA. Difference between each drug baseline was also compared by ANOVA. Measurements of transit time were taken both from ECG R-wave (rPTT) and from end of PEP (pPTT). The relationships between BP and different measures of transit time were evaluated by linear regression. Regression slopes and intercepts and Z-transformed Pearson correlation coefficients were compared for each drug infusion and washout period using repeated-measures ANOVA. Beat-to-beat variability was assessed for SBP, rPTT, and HR. Power spectra were calculated using a smoothed Lomb periodogram (18) for all three variables for each individual drug dose. Coherence is analogous to correlation coefficient in the time domain, ranging from 0 (no coherence) to 1, and was computed over the frequency ranges of 0.050.2 and 0.20.4 Hz. Comparison of coherence values was made by ANOVA.
| RESULTS |
|---|
|
|
|---|
|
|
|
The value of rPTT and pPTT as predictors of SBP and DBP, respectively, was assessed based on the assumption that it would be possible to obtain an ideal calibration slope for each individual equating to the average linear regression slope for all drugs. The 95% limits of agreement for predicted vs. actual BP were ±17.0 mmHg (SBP/rPTT) and ± 17.3 mmHg (DBP/pPTT). Percentage-predicted values falling within 5, 10, and 15 mmHg of actual value [based on British Hypertension Society system for assessing BP measurement accuracy (25)] were 44, 66, and 73%, respectively, for SBP, and 42, 64, and 72%, respectively, for DBP.
Average power spectra over all drugs are shown in Fig. 3. An example of the similarity in SBP and rPTT variability is given in Fig. 4. Mean coherence between SBP and rPTT variability was significantly (P < 0.001) greater at both lower and higher frequencies (0.58 ± 0.37 and 0.70 ± 0.33, respectively) than coherence of HR and rPTT variability (0.46 ± 0.41 and 0.52 ± 0.38, respectively). There was no significant difference in coherence between HR and either rPTT or SBP for lower (P = 0.33) or higher (P = 0.16) frequencies. Coherence was not significantly affected by drug type or dosage (P = 0.96) and is shown in Fig. 4.
|
|
| DISCUSSION |
|---|
|
|
|---|
The expected hemodynamic changes occurred with all four drugs, although the SBP response to salbutamol was mixed. rPTT had a negative correlation with SBP, which was relatively unaffected by different drugs in the population as a whole. rPTT also appeared to be useful as a marker of SBP variability. However, DBP and MAP were weakly correlated with rPTT, although more strongly related to pPTT.
SBP is dependent on both vascular function and ventricular contraction, and so it is perhaps unsurprising that rPTT, a composite measure of both vascular and cardiac activity, is correlated with SBP. However, although in the study population as a whole the correlation between rPTT and SBP appeared relatively unaffected by drugs, this finding must be treated with caution. There were slight differences in the rPTT response between norepinephrine and angiotensin II, despite similar BP profiles. Furthermore, it should be noted that four subjects in this study had positive correlations between rPTT and SBP during the administration of salbutamol. This drug has positive inotropic and chronotropic
2-adrenergic effects, as well as causing peripheral arterial relaxation. Although a fall in PEP is associated with an increase in cardiac inotropy, this does not necessarily relate to an increase in SBP, as any potential pressure risemay be offset by decreases in pressure augmentation by reflected waves or changes in aortic stiffness (24). It would therefore appear inappropriate to use rPTT as a predictor of SBP in all persons, particularly for assessing changes due to vasoactive drugs. Moreover, even using an idealized calibration slope, the limits of agreement between predicted and actual BP were wide, although similar inaccuracies have been described previously between sphygmomanometric and direct arterial pressure measurements (2, 36).
These data also show that PEP accounts for a substantial and variable proportion of rPTT, ranging from
12 to 35%. A number of relatively recent studies have employed rPTT as a marker of vascular function (3, 4), but this study demonstrates that the use of rPTT purely for the assessment of arterial stiffness is inappropriate and should be avoided, as PEP cannot be assumed to remain constant. Other devices, such as the Colin VP-1000 (Colin), have eliminated PEP by utilizing the phonocardiogram to time cardiac ejection. The phonocardiogram is regarded by many as the ideal way of determining systolic time intervals. The principal disadvantage, however, compared with bioimpedance, is that it requires accurate identification of two timing points rather than simply one: first, the end of cardiac ejection (the second heart sound); second, the left ventricular ejection period (measured by identifying the dicrotic notch using a proximal arterial pulse wave).
rPTT may nonetheless offer a potentially valuable means of detecting beat-to-beat changes in SBP. Indeed, with regular recalibration to standard oscillometric BP as suggested by Chen et al. (7), rPTT offers the opportunity to assess BP variability and detect sudden or transient hemodynamic changes. BP and HR variability are considered to offer important insights into vasomotor activity, have been associated with clinical outcomes, including cardiovascular death, and may be used in assessment of autonomic neuropathy (27). Sympathetic modulation of BP alters the HR through the actions of the sinoaortic baroreflex; coherence between these two measures therefore reflects baroreflex activity (20). rPTT shows beat-to-beat variability closer to that of SBP than HR and, therefore, may have a role in the assessment of vasomotor control and BP variability.
pPTT, but not rPTT, was strongly inversely correlated with DBP and MAP. Furthermore, the correlation was inconsistent between pPTT and SBP. These findings are both consistent with the fact that arterial stiffness, and therefore vascular pulse transit time (i.e., pPTT), is dependent on MAP rather than SBP. In many circumstances, SBP and DBP/MAP are positively associated with each other. This has led others to inappropriately use rPTT to predict both these variables (6, 14), with DBP calculated following adjustment for HR. However, the divergent SBP and DBP/MAP responses to salbutamol in eight subjects in this study have not been reported in previous published work in this field, and the present data suggest that rPTT cannot be used to predict DBP or MAP without a knowledge of PEP, regardless of the HR response.
This study has a few limitations. Baseline values of BP were not constant before each drug, tending to rise steadily over the course of the study, particularly after norepinephrine. Due to the short half-life of both pressor agents in particular, it seems unlikely that the rise in BP is entirely accounted for by direct drug effects. Randomizing drug order was not carried out because the much longer half-life of salbutamol necessitated its administration last and it was not justifiable to carry out the separate elements of the study on different days, because this would have required repeated arterial cannulation. The washout periods were also kept relatively short to minimize the duration of cannulation. Despite these points, the aim of the different drugs was to achieve a wide range of BP under varying conditions of vascular tone, and this was still achieved even if the hemodynamic effects of one drug had not completely resolved before the administration of the next. The use of fluid-filled manometer tubing introduces a degree of inaccuracy between pressure at the catheter tip and that at the more proximal transducer. However, this discrepancy was constant between subjects, and fluid-filled manometer tubes are nonetheless regarded as the gold standard in clinical practice. HR is a potential confounding factor (21) in the assessment of vascular stiffness and BP, although debate continues over whether reported increases in pulse-wave velocity with HR are genuine (13). Importantly, however, the large change in HR seen with salbutamol does not affect the interpretation of DBP being more important than SBP as a determinant of pPTT, because a high HR would, if anything, increase arterial stiffness and thus reduce vascular transit time.
In conclusion, this study demonstrates that rPTT has a negative correlation with SBP, which although relatively unaffected by vasoactive drugs in some persons is not reliable enough to enable rPTT to be a surrogate marker of SBP. Furthermore, the significant contribution of PEP to rPTT means that use of the latter parameter as a marker of purely vascular function should be avoided. However, rPTT may have a role in the assessment of BP variability and rapid pressure change. The association of pPTT with DBP/MAP means that use of rPTT as a predictor of diastolic or mean pressure is inadvisable.
| GRANTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|