Journal of Applied Physiology Fuel your research with LabChart
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 95: 571-576, 2003. First published April 18, 2003; doi:10.1152/japplphysiol.00196.2003
8750-7587/03 $5.00
This Article
Right arrow Full Text Free
Right arrow Full Text (PDF) Free
Right arrow A corrigendum has been published
Right arrow All Versions of this Article:
95/2/571    most recent
00196.2003v1
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tang, Y.
Right arrow Articles by Baker, A. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tang, Y.
Right arrow Articles by Baker, A. B.

Calibration of pneumotachographs using a calibrated syringe

Yongquan Tang, Martin J. Turner, Johnny S. Yem, and A. Barry Baker

Department of Anaesthetics, University of Sydney, Royal Prince Alfred Hospital, Sydney, New South Wales 2050, Australia

Submitted 25 February 2003 ; accepted in final form 14 April 2003

Pneumotachograph require frequent calibration. Constant-flow methods allow polynomial calibration curves to be derived but are time consuming. The iterative syringe stroke technique is moderately efficient but results in discontinuous conductance arrays. This study investigated the derivation of first-, second-, and third-order polynomial calibration curves from 6 to 50 strokes of a calibration syringe. We used multiple linear regression to derive first-, second-, and third-order polynomial coefficients from two sets of 6–50 syringe strokes. In part A, peak flows did not exceed the specified linear range of the pneumotachograph, whereas flows in part B peaked at 160% of the maximum linear range. Conductance arrays were derived from the same data sets by using a published algorithm. Volume errors of the calibration strokes and of separate sets of 70 validation strokes (part A) and 140 validation strokes (part B) were calculated by using the polynomials and conductance arrays. Second- and third-order polynomials derived from 10 calibration strokes achieved volume variability equal to or better than conductance arrays derived from 50 strokes. We found that evaluation of conductance arrays using the calibration syringe strokes yields falsely low volume variances. We conclude that accurate polynomial curves can be derived from as few as 10 syringe strokes, and the new polynomial calibration method is substantially more time efficient than previously published conductance methods.

polynomial; linear regression; conductance; pneumotachometer



Address for reprint requests and other correspondence: M. J. Turner, Dept. of Anaesthesia, Royal Prince Alfred Hospital, Bldg. 89, Level 4, Missenden Rd., Camperdown, NSW 2050, Australia (E-mail: mjturner{at}mail.usyd.edu.au).




This article has been cited by other articles:


Home page
Br J AnaesthHome page
Y. Tang, M. J. Turner, and A. B. Baker
Systematic errors and susceptibility to noise of four methods for calculating anatomical dead space from the CO2 expirogram
Br. J. Anaesth., June 1, 2007; 98(6): 828 - 834.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Visit Other APS Journals Online
Copyright © 2003 by the American Physiological Society.