1 Section of Experimental Anesthesiology, Department of Anesthesiology and Critical Care Medicine, University of Freiburg, 79106 Freiburg, Germany; and 2 Division of Clinical Physiology, Clinic of Cardiac and Thoracic Surgery, University Hospital, 4031 Basel, Switzerland
The interrupter technique is used to determine airway and tissue resistance. Their accuracy is influenced by the technical properties of the interrupter device and the compliance of the respiratory system. We investigated the influence of valve characteristics and respiratory system compliance on the accuracy of determining airway and tissue resistance by means of a computer simulation. With decreasing compliance we found increasing errors in both airway and tissue resistance determination of up to 34 and 71%, respectively. On this basis we developed a new occlusion valve, with special emphasis on rapid closing time and tightness in the closed state to improve the accuracy of resistance determination. The newly developed occlusion device greatly improves the accuracy of airway and tissue resistance determination. We conclude that respiratory system compliance is a limiting factor for the accuracy of the interrupter technique. To apply the interrupter technique in patients with extremely low respiratory system compliances, we need sophisticated technical devices.
computer simulation; initial pressure drop; interrupter technique; respiratory system mechanics
This article has been cited by other articles:
![]() |
N. Beydon, S. D. Davis, E. Lombardi, J. L. Allen, H. G. M. Arets, P. Aurora, H. Bisgaard, G. M. Davis, F. M. Ducharme, H. Eigen, et al. An Official American Thoracic Society/European Respiratory Society Statement: Pulmonary Function Testing in Preschool Children Am. J. Respir. Crit. Care Med., June 15, 2007; 175(12): 1304 - 1345. [Full Text] [PDF] |
||||
![]() |
G. Mols, V. Kessler, A. Benzing, M. Lichtwarck-Aschoff, K. Geiger, and J. Guttmann Is pulmonary resistance constant, within the range of tidal volume ventilation, in patients with ARDS? Br. J. Anaesth., February 1, 2001; 86(2): 176 - 182. [Abstract] [Full Text] [PDF] |
||||
![]() |
Q. LU, A. CAPDEROU, P. CLUZEL, E. MOURGEON, L. ABDENNOUR, J.-D. LAW-KOUNE, C. STRAUS, P. GRENIER, M. ZELTER, and J.-J. ROUBY A Computed Tomographic Scan Assessment of Endotracheal Suctioning-Induced Bronchoconstriction in Ventilated Sheep Am. J. Respir. Crit. Care Med., November 1, 2000; 162(5): 1898 - 1904. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |