Adler, A., R. Amyot, R. Guardo, J. H. T. Bates, and Y. Berthiaume. Monitoring changes in lung air and liquid volumes with
electrical impedance tomography. J. Appl.
Physiol. 83(5): 1762-1767, 1997.
lung water; lung volume; thoracic impedance
Electrical
impedance tomography (EIT) uses electrical measurements at electrodes
placed around the thorax to image changes in the conductivity
distribution within the thorax. This technique is well suited to
studying pulmonary function because the movement of air, blood, and
extravascular fluid induces significant conductivity changes within the
thorax. We conducted three experimental protocols in a total of 19 dogs
to assess the accuracy with which EIT can quantify changes in the
volumes of both gas and fluid in the lungs. In the first protocol, lung
volume increments from 50 to 1,000 ml were applied with a large
syringe. EIT measured these volume changes with an average error of 27 ± 6 ml. In the second protocol, EIT measurements were made at end
expiration and end inspiration during regular ventilation with tidal
volume ranging from 100 to 1,000 ml. The average error in the EIT
estimates of tidal volume was 90 ± 43 ml. In the third protocol,
lung liquid volume was measured by instilling 5% albumin solution into
a lung lobe in increments ranging from 10 to 100 ml. EIT measured these
volume changes with an average error of 10 ± 10 ml and was also
able to detect into which lobe the fluid had been instilled. These results indicate that EIT can noninvasively measure changes in the
volumes of both gas and fluid in the lungs with clinically useful
accuracy.
0161-7567/97 $5.00
Copyright © 1997 the American Physiological Society
This article has been cited by other articles:
![]() |
J. A. Victorino, J. B. Borges, V. N. Okamoto, G. F. J. Matos, M. R. Tucci, M. P. R. Caramez, H. Tanaka, F. S. Sipmann, D. C. B. Santos, C. S. V. Barbas, et al. Imbalances in Regional Lung Ventilation: A Validation Study on Electrical Impedance Tomography Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 791 - 800. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hinz, P. Neumann, T. Dudykevych, L. G. Andersson, H. Wrigge, H. Burchardi, and G. Hedenstierna Regional Ventilation by Electrical Impedance Tomography: A Comparison With Ventilation Scintigraphy in Pigs Chest, July 1, 2003; 124(1): 314 - 322. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Adler, N. Shinozuka, Y. Berthiaume, R. Guardo, and J. H. T. Bates Electrical impedance tomography can monitor dynamic hyperinflation in dogs J Appl Physiol, February 1, 1998; 84(2): 726 - 732. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |