Journal of Applied Physiology Millar Instruments
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 82: 698-703, 1997;
8750-7587/97 $5.00
This Article
Right arrow Full Text Free
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Omboni, S.
Right arrow Articles by Mancia, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Omboni, S.
Right arrow Articles by Mancia, G.

Journal of Applied Physiology
Vol. 82, No. 2, pp. 698-703, February 1997
SYSTEMIC CIRCULATION AND FLUID BALANCE

special communication

Performance of the AM-5600 blood pressure monitor: comparison with ambulatory intra-arterial pressure

Stefano Omboni1, Gianfranco Parati1, Antonella Groppelli1, Luisa Ulian2, and Giuseppe Mancia2

1 Istituto Scientifico Ospedale S. Luca, Centro Auxologico Italiano, 20149 Milan; 2 Cattedra di Medicina Interna, Ospedale S. Gerardo, Monza, Universitá degli Studi di Milano, and Centro di Fisiologia Clinica e Ipertensione, Ospedale Maggiore di Milano, 20122 Milan, Italy

Received 29 March 1996; accepted in final form 26 September 1996.

Omboni, Stefano, Gianfranco Parati, Antonella Groppelli, Luisa Ulian, and Giuseppe Mancia. Performance of the AM-5600 blood pressure monitor: comparison with ambulatory intra-arterial pressure. J. Appl. Physiol. 82(2): 698-703, 1997.---The AM-5600 is a new device that simultaneously monitors electrocardiogram (ECG) and noninvasive blood pressure (BP) over a 24-h period. BP readings (Korotkoff sounds and cuff air pressure) are stored into the recorder, allowing the removal of BP artifacts after a visual check. In 12 subjects with essential hypertension, we compared BP values simultaneously provided by the AM-5600 and intra-arterial recordings. At rest, noninvasive systolic BP (SBP) values were lower (5.4 ± 4.9 mmHg) and diastolic BP (DBP) values were higher (7.3 ± 7.3 mmHg) than were intra-arterial values. In ambulatory conditions (9 subjects), between-method discrepancies were +0.8 ± 6.1 and +12.2 ± 7.4 mmHg for 24-h SBP and DBP, respectively. AM-5600 underestimated 24-h intra-arterial SBP and DBP SD, but it accurately tracked intra-arterial SBP and DBP changes. Editing removed 22.1% of total readings, slightly reducing between-method discrepancies. Thus the AM-5600 provides an accurate average estimate of resting and ambulatory SBP and, for DBP, a less accurate estimate that is slightly improved by editing. The AM-5600 allows accurate description of SBP and DBP profiles and thus may be suitable to describe the abrupt BP changes accompanying a number of clinical events.

24-hour ambulatory blood pressure monitoring; noninvasive blood pressure measurement; intra-arterial blood pressure measurement


0161-7567/97 $5.00 Copyright © 1997 the American Physiological Society




This article has been cited by other articles:


Home page
ANGIOLOGYHome page
E. Umana, W. Ahmed, M. A. Fraley, and M. A. Alpert
Comparison of Oscillometric and Intraarterial Systolic and Diastolic Blood Pressures in Lean, Overweight, and Obese Patients
Angiology, January 1, 2006; 57(1): 41 - 45.
[Abstract] [PDF]


Home page
BMJHome page
E. O'Brien, B. Waeber, G. Parati, J. Staessen, and M. G Myers
Blood pressure measuring devices: recommendations of the European Society of Hypertension
BMJ, March 3, 2001; 322(7285): 531 - 536.
[Full Text]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Visit Other APS Journals Online