Journal of Applied Physiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 68: 1605-1614, 1990;
8750-7587/90 $5.00
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sartene, R.
Right arrow Articles by Durand, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sartene, R.
Right arrow Articles by Durand, G.

Journal of Applied Physiology, Vol 68, Issue 4 1605-1614, Copyright © 1990 by American Physiological Society


ARTICLES

Respiratory cross-sectional area-flux measurements of the human chest wall

R. Sartene, P. Martinot-Lagarde, M. Mathieu, A. Vincent, M. Goldman and G. Durand
Laboratoire de Physique des Solides, Universite Paris XI, Orsay, France.

A new device that utilizes the voltages induced in separate coils encircling the rib cage and abdomen by a magnetic field is described for measurement of cross-sectional areas of the human chest wall (rib cage and abdomen) and their variation during breathing. A uniform magnetic field (1.4 X 10(-7) Tesla at 100 kHz) is produced by generating an alternating current at 100 kHz in two square coils, 1.98 m on each side, parallel to the planes of the areas to be measured and placed symmetrically cephalad and caudad to these planes at a mean distance of 0.53 m. We demonstrated that the accuracy of the device on well-defined surfaces (squares, circles, rectangles, ellipses) was within 1% in all cases. Observed errors are due primarily to small inhomogeneities of the magnetic field and variation of the orientation of the coil relative to the field. Using a second magnetic field (80 kHz) perpendicular to the first, we measured the errors due to nonparallel orientation during quiet breathing and inspiratory capacity maneuvers. In 10 normal subjects, orientation effects were less than 2% for the rib cage and less than 0.7% for the abdomen. In five of these subjects, orientation effects at functional residual capacity in lateral and seated postures were generally less than or equal to 5%, but estimated tidal volume during spontaneous breathing was comparable to measurements in the supine posture. In five curarized patients, we assessed the linearity of volume-motion relationships of the rib cage and abdomen, comparing cross-sectional area and circumference measurements. Departures from linearity using cross-sectional areas were only one-third of those using circumferences. In seven normal subjects we compared cross-sectional area measurements with respiratory inductive plethysmography (RIP) and found comparable estimates of lung volume change over a wide range of relative rib cage contributions to tidal volume (-5 to 105%), with slightly higher standard deviations for the RIP (SD = 10% for RIP; SD = 4% for cross-sectional area).


This article has been cited by other articles:


Home page
ChestHome page
A. C. Halbower, S. L. Ishman, and B. M. McGinley
Childhood Obstructive Sleep-Disordered Breathing: A Clinical Update and Discussion of Technological Innovations and Challenges
Chest, December 1, 2007; 132(6): 2030 - 2041.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Respir. Crit. Care Med.Home page
ATS/ERS Statement on Respiratory Muscle Testing
Am. J. Respir. Crit. Care Med., August 15, 2002; 166(4): 518 - 624.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Visit Other APS Journals Online