Journal of Applied Physiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 56: 804-809, 1984;
8750-7587/84 $5.00
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Peslin, R.
Right arrow Articles by Begin, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Peslin, R.
Right arrow Articles by Begin, P.

Journal of Applied Physiology, Vol 56, Issue 3 804-809, Copyright © 1984 by American Physiological Society


ARTICLES

In-phase rejection requirements for measuring respiratory input impedance

R. Peslin, P. Jardin, C. Duvivier and P. Begin

Respiratory flow is commonly obtained by measuring the pressure difference across a pneumotachograph. When respiratory input impedance is studied, that pressure difference may be very small with respect to the absolute pressure swings inside the pneumotachograph. Then the in-phase rejection of the differential pressure transducer is expected to markedly influence the accuracy of the data. The problem was investigated by computer simulation and by measurements on a mechanical analog of the respiratory system made of a resistance, an inertance, and a compliance arranged in series. Both studies demonstrated that comparatively small differences in the volumes of the chambers or in the lengths or diameters of the connecting tubes led to artifactual frequency dependence of resistance and serious misestimation of compliance and inertance. Errors were larger when the resistance of the pneumotachograph was smaller and the impedance of the subject larger. In practice, with usual pneumotachographs accurate impedance measurements require using the most symmetrical transducers presently available (common-mode rejection ratio of about 70 dB at 30 Hz).


This article has been cited by other articles:


Home page
Eur Respir JHome page
E. Oostveen, D. MacLeod, H. Lorino, R. Farre, Z. Hantos, K. Desager, and F. Marchal
The forced oscillation technique in clinical practice: methodology, recommendations and future developments
Eur. Respir. J., December 1, 2003; 22(6): 1026 - 1041.
[Abstract] [Full Text] [PDF]


Home page
J. Appl. Physiol.Home page
A. Giannella-Neto, C. Bellido, R. B. Barbosa, and M. F. V. Melo
Design and calibration of unicapillary pneumotachographs
J Appl Physiol, January 1, 1998; 84(1): 335 - 343.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Visit Other APS Journals Online