Journal of Applied Physiology, Vol 51, Issue 4 1027-1032, Copyright © 1981 by American Physiological Society
ARTICLES |
J. W. Severinghaus
Combined transcutaneous PO2-PCO2 electrodes are described in which the interaction between the two electrodes due to OH- production at the O2 cathode has been eliminated. An anode of either anodized aluminum or platinum has been driven at a current equal to cathode current to force stoichiometric consumption of OH- at its rate of production. The AgCl reference electrode operates at zero current. O2 sensitivity was not significantly altered by electrolyte pH variation from 6.7 to 9.0 with variations by PCO2. These electrodes have been found stable both with and without spacers, and with electrolytes dissolved in 50-100% ethylene glycol. In 22 anesthetized patients, with electrode temperature of 43 degrees C (s refers to skin surface, a to arterial blood); PsO2 = 0.52PaO2 + 15 (range 54-300) (r = 0.66; Sy . x = 29.6; n = 46); and PsCO2 = 1.39PaCO2 + 2.1 (range 24-98) (r = 0.99; Sy . x = 2.28; n = 48).
This article has been cited by other articles:
![]() |
S. S. Torok, J. D. Leuppi, F. Baty, M. Tamm, and P. N. Chhajed Combined Oximetry-Cutaneous Capnography in Patients Assessed for Long-term Oxygen Therapy Chest, June 1, 2008; 133(6): 1421 - 1425. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Sahbaie, S. Madanlou, P. Gharagozlou, J. D. Clark, J. Lameh, and T. M. Delorey Transcutaneous Blood Gas CO2 Monitoring of Induced Ventilatory Depression in Mice. Anesth. Analg., September 1, 2006; 103(3): 620 - 625. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |