Journal of Applied Physiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 51: 584-591, 1981;
8750-7587/81 $5.00
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jansen, J. R.
Right arrow Articles by Versprille, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jansen, J. R.
Right arrow Articles by Versprille, A.

Journal of Applied Physiology, Vol 51, Issue 3 584-591, Copyright © 1981 by American Physiological Society


ARTICLES

Thermodilution technique for measurement of cardiac output during artificial ventilation

J. R. Jansen, J. J. Schreuder, J. M. Bogaard, W. van Rooyen and A. Versprille

The feasibility of using the thermodilution method to monitor cardiac output during artificial ventilation was studied in anesthetized pigs. Normal saline (0.5 ml) at room temperature was injected into the left ventricle or the right atrium. The dilution curves were detected in the aortic arch and the pulmonary artery, respectively. The ventilation rate was 10 cycles/min at end-expiratory pressures of 0, 5, 10, and 15 cmH2O. For each level, 50 measurements of cardiac output were performed at regular intervals over the ventilatory cycle. The order of measurements were randomly selected. The average of each series of 50 measurements showed excellent correlation with the estimates of cardiac output based on the direct Fick method for oxygen. The maximum difference between the values of cardiac output randomly measured by the thermodilution method was 40% for the left side of the heart and 70% for the right side. However, when the values of cardiac output were sorted according to the specific phases of the respiratory cycle, there was a systematic variation with a small random error. For the left side of the heart, a satisfactory moment of injection for estimation of mean cardiac output appeared to be at the end of the spontaneous expiration. On the other hand, the analysis of cardiac output values at the right side did not reveal any satisfactory moment for injectate administration under changing circumstances, e.g., positive end-expiratory pressure.


This article has been cited by other articles:


Home page
NEJMHome page
R. Rossaint, K. J. Falke, F. Lopez, K. Slama, U. Pison, and W. M. Zapol
Inhaled Nitric Oxide for the Adult Respiratory Distress Syndrome
N. Engl. J. Med., February 11, 1993; 328(6): 399 - 405.