Journal of Applied Physiology AJP citation statistics
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Appl Physiol 47: 1131-1135, 1979;
8750-7587/79 $5.00
This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yoshiya, I.
Right arrow Articles by Tanaka, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yoshiya, I.
Right arrow Articles by Tanaka, K.

Journal of Applied Physiology, Vol 47, Issue 5 1131-1135, Copyright © 1979 by American Physiological Society


ARTICLES

Evaluation of a hot-wire respiratory flowmeter for clinical applicability

I. Yoshiya, Y. Shimada and K. Tanaka

A hot-wire flowmeter was evaluated for clinical applicability. 1) Calibration with a syringe could be done if emptying duration was 1-15 s. 2) Outputs linearly decreased with decreasing concentration of N2 in O2. Outputs with CO2 in O2 and N2O in O2 represented nonlinear convex relationships with varying concentrations of CO2 and N2O. Correction for each gas mixture to be measured is required. 3) Outputs linearly decreased with decreasing barometric pressure. 4) Stability assessed with a piston respirator was excellent (within +/- 2% of syringe volume) after 15 min warmup time. However, daily calibrations are recommended in clinical situations. 5) Nebulization, if not excessive, was acceptable if the expired gas was measured at the mouth. 6) Hot-wire burning, which occurred when it was partially in contact with materials whose specific heat differed with air, was successfully protected with a simple shutoff circuit. 7) The possibility of producing nitrogen oxides by the catalytic action of the platinum hot-wire was denied by colorimetric determination. Interchangeability and sterilizability of transducers and improved mechanical strength with platinum-rhodium alloy are also discussed.


This article has been cited by other articles:


Home page
Am. J. Respir. Crit. Care Med.Home page
ATS/ACCP Statement on Cardiopulmonary Exercise Testing
Am. J. Respir. Crit. Care Med., January 15, 2003; 167(2): 211 - 277.
[Full Text] [PDF]


Home page
ChestHome page
S. Miyoshi, T. Yoshimasu, T. Hirai, I. Hirai, S. Maebeya, T. Bessho, and Y. Naito
Exercise Capacity of Thoracotomy Patients in the Early Postoperative Period
Chest, August 1, 2000; 118(2): 384 - 390.