Journal of Applied Physiology, Vol 47, Issue 5 1127-1130, Copyright © 1979 by American Physiological Society
ARTICLES |
P. M. Griffin and N. Zamel
A modification of the conventional volume-displacement whole-body plethysmograph was done using a low-resistance flow element and integrating the plethysmographic pressure signal. A flow element with a resistance of 0.036 cmH2O/(l/s), linear to 75 l/s was fitted to a 440-liter plethysmograph, resulting in an instrument with a volume-amplitude-frequency response flat to 15 Hz and in phase with mouth flow and esophageal catheter-balloon to at least 15 Hz without pressure compensation.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |