Journal of Applied Physiology, Vol 47, Issue 2 462-467, Copyright © 1979 by American Physiological Society
ARTICLES |
A. C. Jackson and A. Vinegar
A device and methodology is presented for testing the frequency response of pressure, volume, or flow transducers. Also reported are responses of selected transducers of all three types over the range of 2--120 Hz. Several pressure transducers tested had good frequency response when connected to the test system with a minimum of interconnecting fittings; others did not. Use of additional connectors degraded the response as did the addition of air-filled catheters. The frequency response of the pneumotachometers tested were influenced largely by the response characteristics of the associated pressure transducer and interconnecting fittings. These results emphasize the need to test the response characteristics of any transducer with specific connectors and fittings that are to be used to make the actual measurements of pressure, volume, or flow.
This article has been cited by other articles:
![]() |
ATS/ERS Statement on Respiratory Muscle Testing Am. J. Respir. Crit. Care Med., August 15, 2002; 166(4): 518 - 624. [Full Text] [PDF] |
||||
![]() |
P. Nassar, A. Jackson, and D. Carrier Entraining the natural frequencies of running and breathing in guinea fowl (Numida meleagris) J. Exp. Biol., May 1, 2001; 204(9): 1641 - 1651. [Abstract] [PDF] |
||||
![]() |
U. Frey, B. Reinmann, and J. Stocks The infant lung function model: a mechanical analogue to test infant lung function equipment Eur. Respir. J., April 1, 2001; 17(4): 755 - 764. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Reinmann, J. Stocks, and U. Frey Assessment of an infant whole-body plethysmograph using an infant lung function model Eur. Respir. J., April 1, 2001; 17(4): 765 - 772. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. MILLER, B. JONES, Y. XU, O. F. PEDERSEN, and P. H. QUANJER Peak Expiratory Flow Profiles Delivered by Pump Systems . Limitations due to Wave Action Am. J. Respir. Crit. Care Med., June 1, 2000; 161(6): 1887 - 1896. [Abstract] [Full Text] |
||||
![]() |
S. A. Tuck, J. C. Dort, M. E. Olson, and J. E. Remmers Monitoring respiratory function and sleep in the obese Vietnamese pot-bellied pig J Appl Physiol, July 1, 1999; 87(1): 444 - 451. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. Sobh, C. M. Lilly, J. M. Drazen, and A. C. Jackson Respiratory transfer impedance between 8 and 384 Hz in guinea pigs before and after bronchial challenge J Appl Physiol, January 1, 1997; 82(1): 172 - 181. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |