Journal of Applied Physiology, Vol 44, Issue 6 977-980, Copyright © 1978 by American Physiological Society
ARTICLES |
J. Prazma, D. Smith and W. J. Jochem
A highly sensitive, miniature, inexpensive circuit for the measurement of PO2 in vivo has been described. The circuit is constructed from a current-to-voltage convertor, clamping circuit, differential amplifier, and reverse voltage and overvoltage protector. The design of the circuit allows us to apply voltage bias to the measuring electrode while grounding the preparation. The clamping circuit holds the selected bias voltage constant while the differential amplifier subtracts this bias potential from the PO2 signal yielding an output voltage that is proportional to the current sensed by the oxygen electrode. The circuit is protected from reverse voltage and overvoltage.
This article has been cited by other articles:
![]() |
W. Stummer, A. Baethmann, R. Murr, L. Schurer, and O. S. Kempski Cerebral Protection Against Ischemia by Locomotor Activity in Gerbils : Underlying Mechanisms Stroke, August 1, 1995; 26(8): 1423 - 1430. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |