Journal of Applied Physiology, Vol 41, Issue 4 590-592, Copyright © 1976 by American Physiological Society
ARTICLES |
M. Sullivan, J. Paliotta and M. Saklad
The measurement of respiratory mechanics is frequently performed on intubated patients. This study shows the effect which the turbulent and inertial characteristics of an endotracheal tube have on the accuracy of the measurement. Two techniques are examined utilizing a mechanical model: the least-squares and the phasor technique. It is concluded that both techniques will give erroneous results if the characteristics of the tube are not accounted for. The error is markedly reduced if the turbulent and inertial terms are accounted for in the analysis.
This article has been cited by other articles:
![]() |
M. Lichtwarck-Aschoff, A. Helmer, R. Kawati, M. Lattuada, U. H. Sjostrand, N. Zugel, J. Guttmann, and G. Hedenstierna Good short-term agreement between measured and calculated tracheal pressure Br. J. Anaesth., August 1, 2003; 91(2): 239 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.-H. Jarreau, B. Louis, G. Dassieu, L. Desfrere, P. W. Blanchard, G. Moriette, D. Isabey, and A. Harf Estimation of inspiratory pressure drop in neonatal and pediatric endotracheal tubes J Appl Physiol, July 1, 1999; 87(1): 36 - 46. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Barnard, A. Shukla, T. Lovell, and J. Goldstone Esophageal-Directed Pressure Support Ventilation in Normal Volunteers Chest, February 1, 1999; 115(2): 482 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. STRAUS, B. LOUIS, D. ISABEY, F. LEMAIRE, A. HARF, and L. BROCHARD Contribution of the Endotracheal Tube and the Upper Airway to Breathing Workload Am. J. Respir. Crit. Care Med., January 1, 1997; 157(1): 23 - 30. [Abstract] [Full Text] |
||||
![]() |
B. Demers and R. S. Irwin Importance of the Nonelastic Resistance of Endotracheal Tubes J Intensive Care Med, January 1, 1990; 5(1): 3 - 4. [PDF] |
||||
![]() |
G. N. Plost, S. C. Campbell, R. T. Vagedes, and B. Y. Shon The Nonelastic Work of Breathing by Normal Humans Using Different Size Endotracheal Tubes J Intensive Care Med, January 1, 1990; 5(1): 23 - 25. [PDF] |
||||