Journal of Applied Physiology, Vol 40, Issue 5 839-843, Copyright © 1976 by American Physiological Society
ARTICLES |
J. K. Li, A. G. van Brummelen and A. Noordergraaf
The performance of catheter-manometer systems for the measurement of pulsatile pressure has been evaluated by both experimental techniques and theoretical considerations. The former approach has shown, on occasion, multiple maxima in the amplitude response. The latter has been approached in a variety of ways, ranging from extreme lumping to application of transmission line theory while employing different configurations in the system's representation. Multiple maxima have also been seen, The present paper identifies the sources of the differences found and compares the relative merits of various theoretical approaches. It introduces the compliance of the system as a figure of merit and provides a simple first-order approximation formula for evaluation of the quality of a system. Damping and impedance matching to improve the system's frequency response were studied. It was found that they were not needed in a very stiff or a very compliant system, nor should one worry about the representation of such a system.
This article has been cited by other articles:
![]() |
J.-W. Lankhaar, N. Westerhof, T. J. C. Faes, K. M. J. Marques, J. T. Marcus, P. E. Postmus, and A. Vonk-Noordegraaf Quantification of right ventricular afterload in patients with and without pulmonary hypertension Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1731 - H1737. [Abstract] [Full Text] [PDF] |
||||