ARTICLES

A high accuracy linear rate meter

C. R. Wyss

By implementing analog computer techniques using digital circuits, an instantaneous rate meter was built with approximately 100 times the accuracy of commercially available rate meters. The circuit is accurate to within +/- 0.2 events/min over a range of rates of 0.2-900 epm. Modifications can be made to provide a digital display of rate. The circuit design techniques used in developing the rate meter may be used to generate a wide variety of functions of time with very high accuracy.